Humidity Sensors
Humidity sensor is for reading humidity level in incubator. When the water
level is low in the autoclavable water reservoir, an audible and visual Low
www.okuman.com.tr
Temperature and
Humidity alarm occurs. Humidity Sensor
Backup Battery
OKM 801 is equipped with backup battery, once power supply is off, the
backup battery supports the air circuit and alarm system to keep working
at least 30 minutes. Also external monitor with haemodynamic parameters
will have a separate power backup.
8’’ Color LCD Touchscreen display provides all incubator parameter
Touchscreen
users to operate soft buttons as well as touchscreen
04
OKM 801 Infant Incubator
Can be operated from both side of hood Ergonomic designed handle with perfect height
Optional
Height Adjustment
Haemodynamic
Monitor
Air Curtain
• High Skin Temperature • Low Skin Temperature • High Skin 1 Temperature • High Skin 2 Temperature
• Skin Probe Disconnect • High Air Temperature • Low Air Temperature • Electronic circuit fault as High Temperature Cut Out
the data collected are true and accurate. Alarms are triggered when any errors related to the operation of the sensors are
detected.
• Humidity Heater Failed • Add water • Low Humidity (moisture) • Reservoir Out Of Position • Oxygen Cell Difference
• Low Oxygen Ratio • High Oxygen Ratio • Check O2 Supply • Servo O2 System Fail • Change Oxygen Cells
• High Spo2 alarm • Low Spo2 alarm • External display failure alarm • Separate Haemodynamic parameters failure alarms
05
Working Condition Monitoring Alarms
In order to keep proper running and provide the best user ecperience, OKM 801 infant incubator monitors the working
condition of each aspect.
• Motor Failed • Power Failure • Stuck Key • Sensor Module Failure • Sensor Disconnect • Low Air Flow
• Air Flow Probe Failed • Battery Disconnect • Sensor Out of Position • Access Panel Open
• Heater Failed • Air Probe Failed • Skin1 Probe Fail Alarm • SpO2 sensor Fail Alarm • Watchdog Failed
GENERAL SPECIFICATIONS
Height 126.1cm to 146.1cm Tubing Access Ports 10
Width 99.7 cm Access Door Size 18x13 cm (Max. 6 doors)
Depth 57.7cm Mattress to Hood Height 48.0 mm
PERFORMANCE
Air Flow Velocity Across Mattress < 10 cm/sec
Temperature Rise Time At 22 °C (72 °F) Ambient <20min (From 22 °C, 50% RH, to 35 °C)
Temperature Variability < 0.5 °C
Temperature Overshoot < 0.5 °C Maximum
Temperature Uniformity With A Level Mattress < 0.8 °C
Operating Noise Level In Hood ≤45dBa
Carbon Dioxide (Co2) Level <0.5%
SCALE (OPTIONAL)
Weight Range 300 g to 8 kg
Weight Display Resolution 1g
Weight Accuracy ±10 g
Operating Environment 20 ° to 30 °C
Temperature 10 to 95% RH
Air Velocity Up to 0.1 m/sec
STORAGE/SHIPPING INFORMATION
Temperature -25 ° to 60 °C
Humidity 0 to 95 % RH
www.okuman.com.tr
848 848
442
479.9
479.9
07
188.5
188.5
1461
1390.9
477.6 872
1261
872
251
282
282 381