You are on page 1of 8

www.okuman.com.

tr

OKM 801INFANT INCUBATOR


OKM 801
INFANT
INCUBATOR
Premature and / or infants which might have serious health problems after
delivery they need close caring of the physician and the nurses because they are
not capable to keep their body temperature at the required level and stability by
their own. Besides the medical cure, in order to improve health conditions; patients
02
need a fresh air in a steril environment with constant temperature and appropriate
humidity as well as silence which do not interrupt their sleep. Also have optional
OKM 801 Infant Incubator

Haemodynamic monitoring features alongwith additional external display. Simple


viewing of all modes and applicable set values of the parameters.

OKM 801 Infant Incubator is specially designed with high technology to


meet all necessary requirements and makes difference in NICU.
Stable
With ambient temperature probe OKM 801 infant incubator is able
to compare the difference between ambient and setting temperature,
and changing heating power output accordingly in order to provide more
stable inside hood air temperature. The micro processor controlled unit has Temperature 1
automatic Self test function. Connector

• Skin temperature monitoring through Reusable & Disposable sensor Temperature 2


• Air circuit temperature monitoring Connector

More accurate temperature control and safer system

Minimize Temperature Uniformity


OKM 801 incubator’s air circuit system is designed based on massive
experimentation. With plenty aerodynamics knowledge the temperature
uniformity is improved as much as possible in order to give infant the
safest micro-environment. Allows the temperature to be monitored on
two different (T1 and T2) points of the body for more accuracy to maintain
targeted temperature level.

Super Quiet Design


It is approved that in addition to the direct auditory effects of noise, noise
also influences the cardiovascular system, the respiratory system, sleep
patterns, and stress levels of neonates.

OKM 801 provides newborn a super quiet environment,


hood sound is reduced to 45dB.

Detachable Sensor Module


Sensor module is connected with incubator main body 20 pin connector.
During maintenance and service it can be removed easily.

Three Channel Temperature Sensors


OKM 801 with External Screen/
There are three channel of temperature sensors in the module box. One Haemodynamic Monitor
channel is for reading hood temperature to remain the temperature at
setting value. Another temperature sensor is used to compare the reading
with other two sensors. If the reading of three sensors differ from 0.8°C,
alarm is triggered.

Humidity Sensors
Humidity sensor is for reading humidity level in incubator. When the water
level is low in the autoclavable water reservoir, an audible and visual Low
www.okuman.com.tr

Temperature and
Humidity alarm occurs. Humidity Sensor

Oxygen Sensors (Optional)


Oxygen
Sensor
The oxygen control system adjusts the flow of oxygen within the incubator (Optional)
hood with a valve and an oxygen sensor module. The sensor module has two
independent oxygen fuel cells. When the sensor module is outside of the
hood during Oxygen Control mode, audible and visual alarms are enabled,
and the flow of oxygen is interrupted. 03

Backup Battery
OKM 801 is equipped with backup battery, once power supply is off, the
backup battery supports the air circuit and alarm system to keep working
at least 30 minutes. Also external monitor with haemodynamic parameters
will have a separate power backup.
8’’ Color LCD Touchscreen display provides all incubator parameter
Touchscreen
users to operate soft buttons as well as touchscreen

04
OKM 801 Infant Incubator

Can be operated from both side of hood Ergonomic designed handle with perfect height

1,5 liter water reservoir enables


OKM 801 to keep running for 36 hours. Big storage drawers

Detachable water reservoir, Three auxiliary power output


easy to clean and sterilize

Optional
Height Adjustment

Haemodynamic Monitor with Tente castor with lock


essential Incubators parameters.
Other Features / Air Curtain
While the front or back access panels open, the air circuit motor rotation rate
is accelerated automatically. Therefore more warm air is delivered to the hood.
It effectively prevent heat lose and provide infant stable micro-environment.

Haemodynamic
Monitor

Double Wall Design

Air Curtain

Alarm Audio & Visual Display


Temperature Monitoring and Protection Alarms
OKM 801 infant incubator monitor and control temperature on various points on heating system to
ensure accuracy and safety. The alarms will be actived if any unmoral phenomenon is monitored.

• High Skin Temperature • Low Skin Temperature • High Skin 1 Temperature • High Skin 2 Temperature
• Skin Probe Disconnect • High Air Temperature • Low Air Temperature • Electronic circuit fault as High Temperature Cut Out

Sensor Monitoring Alarms


OKM 801 incubator monitors the condition of air, humidity, oxygen sensors, air flow probes and sensor box to ensure all
www.okuman.com.tr

the data collected are true and accurate. Alarms are triggered when any errors related to the operation of the sensors are
detected.

• Humidity Heater Failed • Add water • Low Humidity (moisture) • Reservoir Out Of Position • Oxygen Cell Difference
• Low Oxygen Ratio • High Oxygen Ratio • Check O2 Supply • Servo O2 System Fail • Change Oxygen Cells
• High Spo2 alarm • Low Spo2 alarm • External display failure alarm • Separate Haemodynamic parameters failure alarms

05
Working Condition Monitoring Alarms
In order to keep proper running and provide the best user ecperience, OKM 801 infant incubator monitors the working
condition of each aspect.

• Motor Failed • Power Failure • Stuck Key • Sensor Module Failure • Sensor Disconnect • Low Air Flow
• Air Flow Probe Failed • Battery Disconnect • Sensor Out of Position • Access Panel Open
• Heater Failed • Air Probe Failed • Skin1 Probe Fail Alarm • SpO2 sensor Fail Alarm • Watchdog Failed
GENERAL SPECIFICATIONS
Height 126.1cm to 146.1cm Tubing Access Ports 10
Width 99.7 cm Access Door Size 18x13 cm (Max. 6 doors)
Depth 57.7cm Mattress to Hood Height 48.0 mm

Weight 80 kg Soft Bed Mattress Size 3100 cm2 standard


(Bed safety mechanism (upto 77 x 40cm/
active when pulled
Package Dimensions 107*68*155 cm 3080cm2 on request)
outside)
X-Ray Tray 41 x 38.9 cm Trendelenburg ±12°
Hood cover opening 180° (Both Front & back)
External Display
for Incubator & 8, 10 or 12 inches Display of Parameters Active & set values
Haemodynamic colour TFT Trend Display 2,4,8,12,24 Hours
parameters (Optional)
Air Filter (Replacable) Min. 0.3 to Max. 0.5 macron
Power Supply and 100-240 VAC and
Frequency 50/60 Hz

06 TEMPERATURE CONTROL MODES


Temperature Control Modes Skin and Air Temperature Control Modes
OKM 801 Infant Incubator

Air Mode Control Temperature Range 20 °C-37°C


Air Mode Control Override Temperature Range 39°C-40°C
Air Mode Control Accuracy ≤±1.0°C
Skin Mode Control Temperature Range 34°C-37.5°C
Heat Cut of Temperature 40° C ± 0.8 ° C
Skin Mode Control Override Temperature Range 37.5°C-39°C
Skin Temperature Accuracy ±0.3°C
Dual-Skin Temperature Monitoring Yes

PERFORMANCE
Air Flow Velocity Across Mattress < 10 cm/sec
Temperature Rise Time At 22 °C (72 °F) Ambient <20min (From 22 °C, 50% RH, to 35 °C)
Temperature Variability < 0.5 °C
Temperature Overshoot < 0.5 °C Maximum
Temperature Uniformity With A Level Mattress < 0.8 °C
Operating Noise Level In Hood ≤45dBa
Carbon Dioxide (Co2) Level <0.5%

SERVO HUMIDITY (OPTIONAL)


With AutoClean Function of Autoclavable Chamber
Humidity Control Range 30% - 95% RH
Humidity Control Operating Time Without Refilling 24 hours
Humidity Control Reservoir Capacity 1500 ml
Humidity Display Accuracy ±5% RH
Humidity Control Accuracy ±5% RH
DIGITAL OXYGEN DISPLAY / SERVO CONTROLLED OXYGEN (OPTIONAL)
Oxygen Control Range 21% to 65%
Oxygen Display Accuracy Of Full Scale ± 2%
Oxygen Control Accuracy (100% Calibration) ± 3%
Oxygen Control Accuracy (21% Calibration) ± 5%
Oxygen Display Resolution 1%

SCALE (OPTIONAL)
Weight Range 300 g to 8 kg
Weight Display Resolution 1g
Weight Accuracy ±10 g
Operating Environment 20 ° to 30 °C
Temperature 10 to 95% RH
Air Velocity Up to 0.1 m/sec

SpO2 with HAEMODYNAMIC MONITOR (OPTIONAL)


Parameters SpO2, PR, PI, PVI, SpCO, SpOC, SpMet, SpHb
Mesurement Range 0 to 100 %
Perfusion Index 0.02 – 20%
Haemo parameters on External screen PR, NIBP, SpO2, Temp, EtCO2

STORAGE/SHIPPING INFORMATION
Temperature -25 ° to 60 °C
Humidity 0 to 95 % RH
www.okuman.com.tr

848 848
442
479.9
479.9

07
188.5
188.5

1461
1390.9

477.6 872
1261

872
251

282

282 381

876 540 876


© 2018 OKUMAN Medikal Sistemler A.S. All rights reserved.
Technical specifications are subject to change by OKUMAN without prior notice.

OKUMAN Medikal Sistemler A.S.


Kazım Karabekir Caddesi 95/95
06060 Iskitler Ankara / TURKEY
T: +90 312 384 05 20
F: +90 312 384 19 75
info@okuman.com.tr
www.okuman.com.tr Catalog No: OKM 801-003-04-2018-EN

You might also like