Professional Documents
Culture Documents
Huminator
humidity generator
slide 1
Huminator – humidity generator from TIS
technical data:
slide 2
Huminator – humidity generator from TIS
general important information
- never move or transport the Huminator with filled water tube / tank;
the water may destroy the motherboard and get into the evaporator
slide 3
Huminator – humidity generator from TIS
Drying container and drying reagent
- it is necessary to dry the drying reagent from time to time
But be careful: do only dry the reagent, not the complete container!
Regeneration
slide 4
picture: broken container
Huminator – humidity generator from TIS
Water in the chamber
- could occur in every generator running in high humidity and/or temperature ranges over a long time
- to avoid this, do not cool from high to low temperatures in a fast way!
slide 5
Huminator – humidity generator from TIS
adjustment of the Huminator
Adjustment at 11.3%RH and 75.3%RH at +25°C:
- it is also very important to use a reference system for the adjustment to generate exactly
11.3%RH respectively 75.3%RH at +25°C
75.3%RH
Button
slide 6
Huminator – humidity generator from TIS
calibration procedure in the Huminator
- regulate the temperature first until it is stable,
then turn on the humidity regulation to the desired value
slide 8
TIS calibrators – Huminator and Pneumator
Pneumator
pressure calibrator and
measuring instrument
slide 9
Pneumator – pressure calibrator and measuring instrument
calibrate differential pressure – in the lab or on-site
- differential and gauge pressure probes with low Pascal up to 1 bar range
- air velocity/volume flow: not directly calibratable but differential pressure calibration
e.g. testo 521, 512 or any other differential pressure measuring instrument
- choose between „CTRL“-mode (step-by-step) or „AUTO-mode “ (defined sequence)
- very short adjustment time (just a few seconds)
- high mobility: useable for 8 hours without mains supply (incl. pump function)
or
slide 10
Pneumator – pressure calibrator and measuring instrument
highly precise differential pressure measuring
- differential pressure in cleanrooms
- filter pressure and ventilator pressure
- mode „MESS“
- accuracy +/- 0.3 % of final value at 0..100 Pa!
- accuracy +/- 0.1 % of final value at > 100 Pa up to 1 bar!
slide 11
Pneumator – pressure calibrator and measuring instrument
highly precise measuring of air velocity and volume flow
v (m/s) = s * (2 * ∆p / rho)
V (m3/s) = A * s * (2 * ∆p / rho)
p1 = static pressure
p2 = dynamic pressure
∆p = p2 – p1
slide 12
Pneumator – pressure calibrator and measuring instrument
for example: Prandtl Pitot tube testo 0635.2145
- length 300 mm
- diameter 4 mm
- tube factor = 1,0 (Prandtl)
slide 13
Pneumator – pressure calibrator and measuring instrument
the measuring cell inside the Pneumator: inductive and innovative
works
completely
frictionless
concentric
waved
membrane
movement of the
cell causes
inductance
changes on both
sides
Magnetically
coated on
both sides
slide 14
Pneumator – pressure calibrator and measuring instrument
the control elements
slide 15
Pneumator – pressure calibrator and measuring instrument
Ordering information
slide 16
Pneumator – pressure calibrator and measuring instrument
slide 17
Pneumator – pressure calibrator and measuring instrument
EUR 390
slide 18
Pneumator – pressure calibrator and measuring instrument
flexibel, precise, mobile – the features at a glance
- English/German menu
slide 19