You are on page 1of 17

HOME OXYGEN CONCENTRATOR

USER MANAUL

Model No.:DE-1LW
In order to use this product safely, please read the user manual before use
CONTENTS

Safety Precautions 02

Product Features 03

Key Technical Indexes 04

Operating Instructions 05

Cleaning, Care and Maintenance 10

Troubleshooting 13

Working Principle 15

01
Safety precautions

WARNING
This device is not medical instrument NOT INTENDED to be
life sustaining or life supporting.

Pre-Installation
● Upside down operation prohibited.
● If the power supply has unstable voltage, or is beyond
AC220±22(110±10)volt, please connect the voltage regulator before using
the machine.
● Please use standard socket with safety qualification and terminal board
with electrical safety certification.
● Don’t open the case non professionals.

Laying Place
● The machine should be located in a well-ventilated space and avoid
direct sunlight. The distance between the machine and the wall or other
objects should be no less than 10cm.
● Do not put other things or liquid vessel on the machine.
● Stacking sundries is prohibited on the back and bottom vent of the
machine, to prevent shutdown or decrease in oxygen concentration due to
high temperature cause by blocked exhaust.

Using
● WARNING: Oxygen is a combustion-supporting gas, so do not smoke
and keep away from combustion source while using the machine for fear
of fire accident.
● Do not switch on & off frequently. You can restart it after you have
switched it off for 3-5 minutes, considering its service life.
● Avoid oil or grease-like material for any part of the machine. So as not
to pollute the oxygen machine and cause fire.

02
Feature of DEDAKJ Home Oxygen Concentrator
Overview
The scope of product application: The DEDAKJ 1LW Oxygen Concentrator
is designed in accordance with standard of Home usage appliances, It
supplies a high concentration of oxygen.

Operating principle
This product uses AC220V/110V as power source, air as raw material and
adopts the pure oil-free air compressor and qualified molecular sieve
imported with original packaging, so as to generate high purity oxygen by
PSA at normal temperate and supply oxygen.

Function Feature
● Intelligent Control: LED display, timing function, easy to use and
simple to operate.
● Overheating protection: The pure oil-free air compressor with
overheat protector, will stop working automatically when the compressor's
working temperature is too high to protect the compressor and the whole
part of the product.

Accessory List
Oxygen concentrator
Remote control
Power cord
User Manual
Warranty card ( attached at the end of the manual)
Certificate of conformity (attached to the end of the manual)
Quick start manual

03
Key Technical Indexes
● Product Name: Home Oxygen concentrator
● Model: DE-1LW
● Oxygen Generation Method: PSA
● Rated Power: 120VA
● Rated Voltage: AC220±22V 50 ± 1Hz/AC110±10V 60Hz
● Oxygen Concentration: 30% -90%
● Oxygen Flow Rate: 1-7L / min adjustable
● Oxygen Pressure: 0.03-0.06MPa
● Carbon Dioxide: Pass the test according to the prescribed method
● Carbon Monoxide: Pass the test according to the prescribed method
● PH: Pass the test according to the prescribed method
● Morphological Identification: Colorless gas, odorless and tasteless,
combustion-support
● Operating Noise: ≤ 45dB (A)
● Dimension: 340 x 180 x310mm
● Net Weight: 6.0kg
● Classification: Class II
● Display Control: LED light display, intelligent chip control, infrared
remote control

● Safe Working Conditions:


Environmental Temperature : 5℃-40℃
Relative Humidity: ≤ 80
Atmospheric Pressure: 86kpa-106kpa
● When the temperature is lower than 5℃, the device should be placed at
normal working environmental temperature for more than four hours
before use.

04
Operating Instructions
Inspection
Check the appearance of the device and accessory list before use, check
the quantity as if consistent with the manual. Please contact with the
distributor or manufacture soon as if any defective.

Appearance schematic diagram

Oxygen Outlet
Led Display
Infrared receptor Nebulizer Outlet
ON/OFF Switch
Stall Key Knob
Timing +
Timing -

Humidifier Water Tank

Inlet Filtration
Power Switch
Fuse Muffler
Power Interface

05
Preparation Work

Grip the top handle of the humidification water


tank by hand, pull out the water tank after
pressing.

Rubber plug
Filling the Humidifier Water Tank
Open silica gel plug at top and fill pure water to the
highest water level line.
(Cool boiled water, distilled water and so on) *Do not
exceed water level line.

Make sure to cover the rubber plug before using, align


the positioning groove at the bottom of water tank and
then the outlet, put the humidification water tank back
to the main body.
Max water level

Air Filter Inspection


Before switching the device ,Please check if the air filter is in place and
keep it clean.

Connect Power Cord


Plug the output power cable into power supply sock of the oxygen,and
plug the other end of the power cord into the main socket.

06
Connect Oxygen Tube
Connect Oxygen Tube
Connect the large-diameter end of
“oxygen tube” with “oxygen outlet
mouth”and connect the other end
with oxygen inhaler. If there are too
many drops of water in the water trap
during or after oxygen inhalation,
open the water trap and pour it clean.

Inhaler Wearing
Wear oxygen inhaler as shown in the picture. When using the earhook
inhaler, adjust the position of the oxygen outlet and point it at the nostril. If
using the inhaler nasal, just align the two outlets with the nostril and
tighten the rubber ring.

Nebulizer Function Using Method( Liquid Medicine from the Doctor)

1. Before using the atomization function, please knob the nuts of atomiza-
tion.
port and place the nut in the hand-held groove position.
2. Connect the atomization port to the bottom of the atomization cup
through.
catheter, the atomizing cup can be poured into liquid medicine or diluted
liquid medicine as needed.( The maximum capacity of atomization cup is
8ml. Do not exceed the maximum scale), wear a mask or mouth contain-
ing type as required.

07
3. Then press the oxygen button of the machine, transfer to 2L stall
oxygen.
state, Change the oxygen generator function into atomization function
4. After atomization, the liquid is injected into the human respiratory tract
through the atomization device.
5. After do atomization for approximately 30-35 minutes, then finished
atomization, press the oxygen key to shut down, removed the atomization
device, and clean the atomization cup, after dried and put it in a cool dry
pla.
6. Finally, re-twist the nut back to the atomization port, and tighten.

Starting Oxygen Concentrator


Turn on the power switch, press the oxygen key, the machine starts to
make oxygen, adjust the flow meter, and then you can inhale oxygen
(please see the operation instructions on the panel or remote control for
details).

DE-1LW LED Screen Display Instruction

1. Negative ion display


2. Preset time
3. Oxygen flow rate
4. Oxygen concentration
5. Cumulative time
6. Cooling fan
7. Sleep display
8. Fault alarm display:
compressor, speed motor, cooling fan

08
Oxygen Key
Press “oxygen key”, display is fully lit ,and machine is in oxygen and
negative ion working condition,while the water tank light show blue. At
same time, the screen displays preset time and accumulated time. Under
working state, press the "oxygen key" to enter the standby state, and the
fan will be turned off after a delay of 3 minutes.

Flow Regulation
Adjust the flow rate by pressing “flow regulation key”. Customer can
control the oxygen flow rate according to their need. There are 1-7 stall
points, it’s corresponding to l-7L / min flow, oxygen flow increases and
oxygen concentration decreases.Switch on in power off condition, the
default flow value is 1L ,restart on standby. The value of the flow file will
retain the setting of the flow file when the device is turned off.

Timing Plus, timing minus key


Press“timing +” key to increase the timing each 15 minutes by one-way
cycle setting, maximum setting is 180 minutes. After 3 seconds of inactivi-
ty, the timing will start to count down until the timing automatically shuts
down.When press "timing -" key, the time will be setting decreased as
each 15 minutes from total 180 minutes.

Note: When operate functional key, it will be different at the statue of


display“ 000”
1.In the highlighted state, press functional key will directly reflect operat-
ing result.
2.In the power saving state, press functional key once to activate highlight
mode , press it again , then it will reflect operating result.

Press any key each time , the buzzer sound once. It will be in power
saving mode after 1 minute without operation.

09
Sleep Display
Press “sleep ” key while in working , the display switches to sleep mode
after the sleep display symbol flashing 3 times. After pressing any key or
touch any function key on the panel , display waked up again.

Accumulation function
This machine is designed with accumulated time function. Record the
usage time every 20 minutes. After one hour, the display will increase by
one hour.

Automatic fault detection function


If cooling fan fails during use, the display will show a red fan icon, accom-
panied by a buzzer alarm and the machine will stop working.
For other function, please refer to Quick Start Manual

DE-1LW Remote control instrument

1. ON/OFF Oxygen
2. Stall Key
3. Timing Plus Key
4. Timing Minus Key
5. Negative ion display
6. Sleep display

Cleaning, Care and Maintenance


Cleaning housing unit
The surface of the housing should be cleaning at least once every month.
Disconnect the power cord before you clean it. Wipe it with a clean, mild,
damp cloth soaked with disinfectant.Do not let the liquid penetrate into the
machine.

10
Cleaning water tank
To keep the water tank clean, the water used should be pure water,
distilled water or cold boiled water. we suggest to change water every day.
The water tank should be cleaned every week. First use mild detergent
and shake the tank to wash. Then use clean water to rinse it,
After cleaned the water tank, please remember to put water tank plastic
plug back to avoid leakage of oxygen from the water inlet, caused no
oxygen output from oxygen outlet mouth when the machine is in normal
working.

Air filter cleaning and replacement


The white air filter at the back of the machine can reduce noise, isolate
dust and purify the air. It should be cleaned around every ten days. First
use mild detergent to wash, and then rinse it with clean water. Pay special
attention that it must be dry to be positioned back, or the function of the
molecular sieve could be damaged, which will lead to decreased oxygen
concentration.

Cleaning oxygen Inhaler and atomizing components


Oxygen inhaler should be cleaned after use, and the nasal suction head
on the nasal inhaler should be cleaned and disinfected after use.
Soaking the inhaler in the solution of 5% potassium ferric acid for 5
minutes, cleaning with water, or wiped with alcohol. Oxygen tube as
consumable should be replaced regularly.

Fuse Replacement
After confirmed the power cord plug is out of power suppler outlet,using a
slotted screwdriver to pry off the fuse holder cap, take out the defective
glass fuse tube ,replace new one, put the cap back and tighten it.

11
Use a small slotted screwdriver to pry off the fuse
holder cap

Spare Parts

Using

Replace a new fuse Put the fuse cap back to socket

Long-term Desposite process

Long-term Deposits
If oxygen concentrator is in unfrequency of use, please do these:
● Take out humidified water tank, pour water out,and put it back in place
● Cut off power supply, collect power cord.
● Keep oxygen concentrator storage in a normal temperature and dry
place without directly sunlight and pollution.

Please do as following if reuse the the oxygen concentrator


● Check if the electric is damaged.
● Check if the oxygen tube is disconnnected.
● Cleaning humidifier water tank.

12
Troubleshooting
NO. Phenomenon Cause analysis Solution
In the initial stage of using, The odor that comes from
there is smell for the silicone tube and ABS rubber
connection tube and other parts is normal and non-toxic by
plastic parts. It is normal. inspection.The odor will
1 Odor in air. disappear as time goes by.
The humidifier water tank
Clean the oxygen tube and
hasn't been cleaned for a long
humidifier water tank.
time or the water hasn't been
changed.

The water level of the


humidifier water tank is beyond
the highest water level, so the Keep the water lower than the
airflow pressed the bubble into highest water level.
Water flows from the the oxygen tube.
2 oxygen outlet. pour the water in water tank.
The machine has been
used for a long time, so that And add water into the water
the steam in the airflow tank when there is no water
gathers on the tube wall. coming out from oxyhen outlet.

The machine lelse lt is the sound of nitrogen


3 "puff" sound every discharging which is normal
several seconds and is asign of fine working.
The flowmeter is off or the Turn up the oxygen flow.
value istoo small
After startup, the
machine operates The oxygen tube is twisted or
Straighten the oxygen tube.
4 well, the flowmeter blocked.
displays "0" or is The inner connection tube of Disconnect the powercord and
too small the machine is disconnected or contact professionals to repair it.
broken.

After startup, there The silica gel stopper of the Remove the humidifying water
is bubble in the humidifying water tank is not tank to plug the silica gel
water tank,butthere properly positioned. stopper.
5
is no oxygen
getting out from the
oxygen outlet.

13
NO. Phenomenon Cause analysis Solution
Disconnect the power cord and
Compressor malfunction. contact professionals to repair
it.
After startup, the
The temperature is so high lt is a normal case.When the
in dicator light works,
6 but the machine does
that the self protection mode temperature lowers down, the
of the oilfree compressor is machine will restart automati-
not(the sound is
activated. cally.
irregular).
Disconnect the power cord and
Separation valve mulfunction contact professionals to repair
it.

The inner conection tube of Disconnect the power cord and


the machine is disconnected contact professionals to repair
or broken. it.

If it still does not work as you have followed the solutions listed above,
do not disassemble the machine. Please consult the manufacturer for
assistance.

14
Working Principle
The air comes into the compressor after compressed by air inlet filter unit,
and then it enters into molecular sieve tower to separate the oxygen and
nitrogen.The oxygen successfully goes through the molecular sieve tower
into the fine sieve tower.In contrast, the nitrogen is absorbed by the
molecular, then to be discharged into the atmosphere through the sepa-
rating valve. The oxygen concentration is enhanced in the fine sieve tower
and controlled flow size by flowmeter.Then the oxygen is humidified by
the humidifying water tank when it comes to it. Finally , the oxygen
reached the user through oxygen supply tube.

15