Understanding Environmental Surveillance

attery-powered, direct-reading instruments are the selection of protective equipment necessary How Combustible Gas Monitors W ork
lassified as two groups—single-gas instruments for the tasks in the area. It may also dictate the
r multiple-gas instruments—typically duration of worker exposure to the environment To understand how portable combustible gas
monitoring one or a combination of the following of the space, or whether an entry will be made detection instruments work, it is first important
tmospheric conditions: at all. Substance-specific detectors should be to understand what is meant by the Lower
. oxygen deficiency or enrichment; used whenever actual contaminants have been Explosive Limit (LEL) and Upper Explosive Limit
identified. (UEL).When certain proportions of combustible
. the presence of combustible gas; and
vapors are mixed with air and a source of ignition
. the presence of certain toxic gases.
is present, an explosion can occur.The range of
Depending on the capabilities of the instrument, concentrations over which this reaction can occur
monitoring can be conducted simultaneously is called the explosive range.This range includes
or oxygen and combustible gas, or for oxygen, all concentrations in which a flash will occur or a
ombustible gas and toxic gases.These devices flame will travel if the mixture is ignited (see
re commonly referred to as 2-in-1, 3-in-1, 4-in-1 Figure 3).The lowest percentage at which this
r 5-in-1 alarms. can happen is the LEL; the highest percentage is
No matter which type of instrument is used to the UEL.
heck environmental gas concentrations, regular Most combustible instruments display gas
monitoring should be performed because a Figure 2 concentrations as a percentage of the LEL. Some
ontaminant’s level of combustibility or toxicity models have gas readouts as a percentage by vol-
might increase even if it initially appears to be ume and others display both percent of LEL and
ow or non-existent. In addition, oxygen percent combustible gas by volume.What’s the
eficiency can occur unexpectedly. difference? For example, the LEL of methane (the
major component in natural gas) is 5 percent by
Atmospheric Composition volume, and the UEL is 15 percent by volume. If
we slowly fill a room with methane, when the
o determine the composition of an atmosphere, concentration reaches 2.5 percent by volume, it is
eliable instruments should be used to draw air 50 percent of the LEL; at 5 percent by volume it is
amples. If possible, do not open the entry portal Figure 3 100 percent of the LEL. Between 5 and 15 percent
o the confined space before this step has been by volume, a spark could set off an explosion.
ompleted. Sudden changes in atmospheric com- Combustible Gases
osition within the confined space could cause
iolent reactions, or dilute the contaminants in In order for combustion to occur, there must be
he confined space, giving a false low initial gas three elements:
oncentration. 1.fuel
When testing permit-required spaces for accept- 2. oxygen to support combustion
ble entry conditions, always test in the following 3. heat or a source of ignition
rder: This is known as the fire triangle, but if you
. oxygen content remove any one of the legs, combustion will not
. flammable gases and vapors occur (see Figure 2). Figure 4

. potential toxic air contaminants The percentage of combustible gas in the air is Different gases need different percent by volume
important, too. For example, a manhole filled concentrations to reach 100 percent of the LEL
with fresh air is gradually filled by a leak of (see Figure 4). Pentane, for example, has an LEL of
combustible gas such as methane or natural gas, 1.5 percent. Instruments that measure in percent
mixing with the fresh air. As the ratio of gas to air of the LEL are easy to use because, regardless of
changes, the sample passes through three the gas, you are most concerned with how close
ranges: lean, explosive and rich (see Figure 3). In the concentration is to the LEL.
the lean range, there isn’t enough gas in the air to
burn. On the other hand, the rich range has too
much gas and not enough air. However, the Single-Gas Monitors for Oxygen Deficiency
gure 1 explosive range has just the right combination of Oxygen indicators measure atmospheric concen-
Comprehensive testing should be conducted in gas and air to form an explosive mixture. Care trations of oxygen. Concentrations are generally
arious locations within the work area. Some must be taken, however, when a mixture is too measured over a range of 0 to 25 percent oxygen
ases are heavier than air, and tend to collect at rich, because dilution with fresh air could bring in air, with readings being displayed on either
he bottom of a confined space. Others are the mixture into the flammable or explosive digital readout or an analog meter.
ghter, and are usually in higher concentrations range. An analogy is the automobile that won’t
Oxygen indicators are calibrated with uncontami-
ear the top of the confined space. Still start on a cold morning (a lean atmosphere
nated fresh air containing a minimum of
thers are the same molecular weight as air, because the liquid gasoline has not vaporized
20.8 percent oxygen.With some models, an
o they can be found in varying concentrations sufficiently), but can be flooded with too much
alarm is activated when oxygen levels drop below
hroughout the space.This is why test samples gasoline (a rich atmosphere with too much
19.5 percent.
hould be drawn at the top, middle and bottom vaporization). Eventually,when the right mixture of
f the space to pinpoint varying concentrations of gas and air finally exists (explosive), the car starts.
ases or vapors (see Figure 1).The results of the
tmospheric testing will have a direct impact on
Single-Gas Monitors for Combustible Gases Multiple-Gas Monitors for Oxygen and Detector Tube Sampling Systems
Combustible Gas
Detector tube-type devices are recommended for
In applications where it is necessary to determine conducting quick evaluations of potential hazards
oxygen and combustible gas levels simultaneous- that cannot otherwise be measured.With detec-
ly, 2-in-1 diffusion-type devices can be used. tor tubes, a known volume of air is drawn
Sensors measure 0 to 100 percent of the LEL and through the tube, using a manually operated or
oxygen from 0 to 25 percent. Remote sampling battery-powered sampling pump. If gas or vapor
requires either a pump module or an aspirator is present in the air, chemically treated granules in
bulb adapter. the tube are stained a different color. By measur-
ing the length of the color stain within the tube,
Figure 5 users can determine concentration levels.
Multiple-Gas Monitors for Oxygen,
Single-gas instruments for monitoring Combustible and Toxic Gases Most tubes available today are made of glass,
combustible gases and vapors are generally have break-off tips, and are filled with treated
Toxic gases and vapors, which can be inhaled or chemical granules.They generally have a shelf life
calibrated on pentane and are designed for gen- absorbed through the skin, are frequently found
eral-purpose monitoring of hydrocarbon vapors. of 24 to 30 months.
in confined spaces. Sometimes, these atmospher-
Such instruments operate by the catalytic action One type of pump frequently used with a
ic hazards can also displace oxygen and may inca-
of a heated platinum filament in contact with detector tube is a compact, bellows-type device.
pacitate the body’s ability to maintain respiration.
combustible gases (see Figure 5).The filament Accurate and repeatable sample flows can be
Some toxic gases and vapors can also cause long-
is heated to operating temperature by an electric assured by a shaft that guides the bellows during
term physical damage to the body in cases of
current.When the gas sample contacts the heat- compression. Some models feature an end-of-
repeated exposure.
ed filament, combustion on its surface raises the stroke indicator that lets the user know when a
temperature in proportion to the quantity of A number of instruments are available to assist in full air sample has been drawn. Models with an
combustibles in the sample. A Wheatstone detecting toxic gas. Pocket-size monitors operate integral stroke counter eliminate the tedious
bridge circuit, incorporating the filament as one by diffusion or an aspirator bulb. Larger instru- recording of multiple pump strokes.
arm, measures the change in electrical resistance ments with built-in pumps draw samples from
due to the temperature increases.This change the immediate area or from outside the confined
space work area when used with sampling lines. Personal Sampling
indicates the percentage of combustible gas
present in the sample. Diffusion-type instruments are available for Personal sampling is used to determine the con-
simultaneously measuring the LEL of combustible centration of airborne contaminants. Personal
gases, oxygen levels and toxic levels (in parts per sampling pumps are designed to measure
Single-Gas Monitors for Toxic Gases
million) of H2S, CO and other toxic gases. Alarms individual workers’exposures, so they typically
Compact, battery-powered devices can be used to also alert the user to low and high oxygen levels. are lightweight, belt-mounted, battery-powered
measure levels of such gases as carbon monoxide Remote sampling pump adapters are available to devices.
(CO) or hydrogen sulfide (H2S), depending on the convert these diffusion-type instruments into The process of sampling entails drawing a prede-
model selected.Toxic gas monitors use electro- pump-style instruments. termined volume of air through a filter designed
chemical cells (see Figure 6). If the gas of interest to trap contaminants.The filter is contained in a
enters the cell, the reaction produces a current plastic cassette, which is attached by plastic
output proportional to the amount of gas tubing to a sampling pump calibrated to draw a
in the sample.With these instruments, audible specific, known volume of air into the filter.
and visible alarms sound if the gas concentration After air samples are drawn, the filters are sent
exceeds a preset level.These devices are well suit- to a laboratory where they are examined to
ed for use in confined spaces containing motors determine the level of exposure.
or engines, which can generate large quantities of
CO, as well as in sewers, waste treatment plants Personal sampling determines the concentration
and “sour crude”processing stations which tend found in the “breathing zone”or the area near the
to have hazardous volumes of H2S. Figure 7 worker’s face, which is usually measured at or
near the collar or lapel.
Photoionization Devicesfor Toxic Gases
and Vapors Calibration
A photoionization detector, featuring micro- To ensure the accuracy of all monitoring and
processor technology, uses ultraviolet light to detection equipment, calibration should be
ionize molecules of chemical substances in a performed regularly. If the instrument reading
gaseous or vaporous state (see Figure 7). differs significantly from the values of the known
A real-time digital readout allows the user to standard, the instrument should not be used until
make an immediate determination of gas it has been adjusted or, if necessary, repaired.
Figure 6 and vapor concentrations. Depending upon
calibration input, gas and vapors are measured
over a 0.1 to 10,000 ppm scale. Some instruments
automatically compensate for signal loss due to
humidity, which is inherent in all PID detectors.
FiveStar® Alarm
he FiveStar Alarm provides the flexibility to meet your gas
etection needs, yet is incredibly simple to use.The durable,
igh-impact case makes it long-lasting, while the advanced
rgonomic design and software reduce worker load and
implify operation.This means less time spent training and
ess time out of service. Over its life, the FiveStar Alarm’s ease
f use, safety features, and long-term money savings combine
o provide the best in gas detection available today.

The FiveStar Alarm s simplicity is obvious The FiveStar Alarm saves money:
rom the start
: • Unbeatable warranty and service.
Three well-marked, easy-to-use keys. • Long-life combustible sensor.
Display faces user. • One-button calibration saves training and gas.
Industry leading RFI protection. • 5-gas capability means one detector can do the job of two detectors.
Flexible smart sensors. • Battery “gas gauge”reduces down time.
Weatherproof—IP54-rated. • 20-hour NiMH battery option means one battery can do the job of two.
Gas sensors available—you can order the FiveStar Alarm with sensors to detect:
The FiveStar Alarm s advanced softwarefeatures • Combustible gas (0–100% LEL or 0–5% methane)
mean safer operation: • Oxygen (0–25%)
Limits on sensor zeroing. and up to three of the following gases:
Limits on one-button calibration. • Ammonia 0–100 ppm • Nitric oxide 0–100 ppm
User acknowledges configuration changes. • Carbon monoxide 0–2000 ppm • Nitrogen dioxide 0–100 ppm
Data recording is standard. • Chlorine 0–20 ppm • Phosphine 0–5 ppm
• Chlorine Dioxide 0–5 ppm • Sulfur dioxide 0–100 ppm
or more complete information, please refer to
• Hydrogen sulfide 0–500 ppm
Bulletin 0816-06-MC.
Moregas detection capabilities arecontinually int roduced. Callfor availability of other gas
detection capabilities. Some configuration limitations apply.

iveStar Alarms can be ordered in one of two convenient ways. Either select from the most popular kit combinations below or refer to MSA’s highly effective
ustom-ordering process,“Assemble-to-Order.”

ap mp
Pu ket ase on Ki
t
FiveStar Alarm Kit Options r y er C
tte r
ge ar
g
tion
ec
k
ne be Jac ng C ti
Ba Char h a ch Li o r
the rryi
a
t C alibr e r
tP r
10
Cd ls ft lib
LEL O2 CO H2
S Ni AC Fas C Pu 1f Lea Ca Ca
eluxe, 4-gas 10018002 • • • • • • • • • • • •
eluxe, 3-gas, CO 10018003 • • • • • • • • • • •
eluxe, 3-gas, H2S 10018004 • • • • • • • • • • •
dustrial, 4-gas 10018005 • • • • • • • • • •
dustrial, 3-gas, CO 10018006 • • • • • • • • •
dustrial, 3-gas, H2S 10018007 • • • • • • • • •
conomy, 4-gas 10018008 • • • • • • •
conomy, 3-gas, CO 10018009 • • • • • •
conomy, 3-gas, H2S 10018030 • • • • • •

Assemble-to-Order (ATO) System: You Make the Choices


he ATO System makes it easy to “custom order” the FiveStar Alarm, configured exactly the way you want it.You can choose from an
xtensive line of base instrument components and accessories. See the ATO chart on the following page to make your selections.

or a full descripton ofthe Fi


veStar Alarmsaccessories,please refer to pages 67—68.

ee pages 82—86 for calibration accessories.


FiveStar® Alarm Assemble-to-Order (ATO) Options

A. Electrochemical Sensor Options Sensors (Select 4) Code H. Sampling Lines Sampling Lines Code
Please list the code number for the toxic None 0 MSA’s high-speed sampling lines are of the None 0
gas sensors desired in the spaces to the Oxygen (0-25%) 1 highest quality materials guaranteed to not 5-foot coiled 1
right. In order to properly process your Carbon Monoxide (0-2000 ppm) 2 alter sample concentration. Screw fittings 10-foot 2
order, list them in descending order. Use Hydrogen Sulfide (0-200 ppm) 3 with O-Ring couplings at each end ensure 15-foot 3
zeros to complete any of the four selections Nitric Oxide (0-100 ppm) 4 that there are no leaks when the system is 25-foot 4
where no sensor is desired (e.g. 2100). operating.
Nitrogen Dioxide (0-100 ppm) 5
Add the price of each sensor to obtain the
sensor sub-total pricing. Sulfur Dioxide (0-100 ppm) 6 I. Sampling Probes Sampling Probe Code
Chlorine (0-20 ppm) 7 MSA’s sampling probes are durable and None 0
Phosphine (0-5 ppm) 8 comfortable to use. The built-in WaterStop® 8" probe 1
Ammonia (0-100 ppm) 9 filter in the handle of the probe ensures your 1-foot straight 2
Chlorine Dioxide(0-5 ppm) A sampling system is not contaminated with 3-foot straight 3
dirt and debris if the probe tip is dipped 3-foot, holes near tip 4
B. Combustible Gas Sensor Options Sensors Code under water. You must order a probe if you
None 0 3-foot, holes near handle 5
If a combustible gas sensor is desired, order a sampling line as we strongly
select type of display. There are two options % CH 4 (Methane) M recommend you operate the instrument Hot-gas sampling probe 6
available: 0-5% Methane by vloume (%CH4) % LEL Pentane L with this filter in place.
or 0-100% Lower Explosive Limit (%LEL
Pentane)
J. Carrying Attachments Belt Clip Code
While a high strength, molded plastic belt Molded plastic belt clip 1
C. Page Options Page Options Code clip is offered with the instrument at no Swivel belt loop 2
Select your desired combination of the No option pages active 0 charge, there are three other options for Pistol grip handle 3
following pages: Datatagging only 3 handling the instrument. The swivel belt Stainless steel belt clip 4
Peak Peak, STEL &TWA A loop allows quick release from your belt
Short Term Exposure Limit (STEL) All four option pages active B with a half turn. The pistol grip kit is helpful
Time Weighted Average (TWA) for longer stints of hand held sampling.
Datatagging K. Instrument Jacket Instrument Jacket Code
D. Alternate Languages Language Code To offer the highest levels of environmental None 0
The FiveStar Alarm speaks your worker’s English only 0 and rough handling protection for your Leather Case 1
language. Select the alternative languages FiveStar Alarm, you can order either a leather Cordura Nylon Jacket 2
English, Spanish & French 1
option and the instrument will display a or Cordura Nylon jacket. The cases are
screen at battery attach which asks user to custom designed to fit either the diffusion
select between three languages for screen instrument or the instrument with the
displays. pump.
E. Battery Pack Options Battery Pack Code L. Calibration Kits Calibration Accessories Code
There are multiple, flexible options to None 0 MSA provides two options for verifying None 0
power the instrument. The standard NiCd Standard NiCd Rechargeable A instrument calibration. The RP kit, with a Model RP Calibration Kit 1
and heavy-duty NiMH rechargeable battery Replaceable “AA” B constant flow regulator and replaceable Squirt Gas Check Kit 2
packs have a built-in battery gauge to cylinders provides the greatest accuracy.
Codes A & B C
provide the remaining hours of run time The Squirt® Gas Check Kit’s aerosol canisters
NiMH D provides an economical means to verify
for the instrument. The NiMH battery Codes B & D E
provides up to 20 hours of continuous instrument calibration.(Only available in ATO
operation for long-running jobs. The for Combustible, O 2,CO,H 2S other gases must
convenient replaceable pack can operate be purchased separately.)
with either alkaline or lithium batteries. M. Data Communications Kit Data Communications Code
F. Battery Charger Options Battery Charger Code The FiveStar Link Communications Kit None 0
None 0 contains the software and adapter neces- FiveStar Link 1
From wall adapters to vehicular chargers,
8-24VDC (Vehicle) 1 sary to download the instrument datalog
you can recharge your batteries wherever
110 VAC 2 or change the instrument setup options.
you need. Also, the unique Fast Charger
offers the best technology available in safe, 220 VAC 3 N. Packaging Options & Carrying Cases
rapid charging of your NiCd battery packs. Packaging & Carrying Cases Code
110 VAC and Vehicle 4
For a small investment, the Fast Charger You can further protect your FiveStar Alarm Std shipping container only 0
Fast Charger (NiCd only) 5
not only fully charges in less than five with optional carrying cases. For ease of Black plastic case 1
Fast Charger (NiMH and NiCd) 6 transport and handling, these cases can be
hours, it also pays for itself by ensuring the Heavy duty case with video 2
longest possible life from your rechargeable even be ordered to accommodate calibration Large, heavy duty case with 3
packs. accessories.With the premium cases, a sim- video & space for cylinder
ple to understand training videotape is
G. Sampling and Calibration Methods Sampling/Cal. Methods Code Orange plastic case 4
included at no extra charge.
The FiveStar Alarm’s sampling systems None 0
provide the highest degree of protection Aspirator Pump Module A O. Instrument Case Selection Instrument Cases Code
available. The patented PulseCheck® Pump Calibration Cap C The FiveStar Alarm is offered in two case Nickel-plated 0
constantly checks the flow system for PulseCheck Pump P housings: standard, nickel-plated ABS or Stainless filled 1
changes in operation and will provide an stainless steel fiber-filled black ABS. The
alarm to the user if it senses a major standard nickel-plated version provides high-
change. You do need to have at least one er levels of Radio Frequency Interference (RFI)
pump, aspirator or calibration cap in your protection, whereas the stainless-filled ver-
“fleet” to calibrate your instruments. sion provides resistance against corrosion in
harsh chemical environments.
Your Model Part Number
Please write in numbers from selections above in appropriate boxes and
contact your Safety Products Distributor to place your order.

Selection A A A A B C D E F G H I J K L M N O
A-5ST- - 1 - - -
Orion® Multigas Detector
he Orion Multigas Detector is a Features
ow-cost, reliable, and easy-to-use
• One-button calibration • TIM ® System compatibility
portable instrument for detecting
• Long-life battery • Large, easy-to-read display
he presence of O2, H2S, CO and
ombustible gas. It will stand up to • Carbon-filled nylon case • Durable sensors
he roughest handling in even the • Tough, durable construction • Link™ Software compatibility
oughest environments. • Built-in pump

Approvals
he Orion Multigas Detector has been designed to meet intrinsic
afety testing requirements in certain hazardous atmospheres.The
Orion Multigas Detector has been approved by UL and cUL for use in
Class I, Division I, Groups A, B, C, and D.

Kit se

ase
Bla tion a Ca
Orion Multigas
y

yC
al Cap
ter

ra dur
ger ry

tL u m p
Detector Kits

om
at
ar tte

Bo con
Ca e Co
Ca ne B

Int atio

Or be
Ch a

1f e
& HB

n
ro

E
i
ali

g
r
ern

an
lib

tP

ck
lib

ot
M
S

Alk

10f
LEL

Ni
CO

H2
2

PartNo.
O

eluxe 4-gas, datalog • • • • • • • • • • • • 10030399


eluxe 4-gas • • • • • • • • • • • • 10025932
eluxe 3-gas H2S • • • • • • • • • • • 10025935
eluxe 3-gas CO • • • • • • • • • • • 10025938
ndustrial 4-gas • • • • • • • • • • 10025931
ndustrial 4-gas, less jacket • • • • • • • • • • 10030420
ndustrial 3-gas H2S • • • • • • • • • 10025934
ndustrial 3-gas CO • • • • • • • • • 10025937
conomy 4-gas, NiMH • • • • • • 10030398
conomy 4-gas • • • • • • 10025930
conomy 3-gas H2S • • • • • 10025933
conomy 3-gas CO • • • • • 10025936

Assemble-to-Order (ATO) System:You Make the Choices


he ATO System makes it easy to “custom order” the Orion Multigas Detector, configured
xactly the way you want it.You can choose from an extensive line of base instrument
omponents and accessories. See the ATO chart on the following page to make your
elections.

or a full descripton ofthe Fi


veStar Alarmsaccessories,please refer to pages 67—68.

ee pages 82—86 for calibration accessories.


Orion® Multigas Detector Assemble-to-Order (ATO) Options

A. Combustible Sensor Options Combustible Sensor Code I. Sampling Probes Sampling Probe Code
To choose to have MSA’s own long-life 20L None 0 MSA’s sampling probes are durable and None 0
combustible sensor installed in your new % LEL (Pentane, 0–100%) 1 comfortable to use. The built-in Water Stop 1-foot straight 1
Orion Multigas Detector, choose selection filter ensures your sampling system is not 3-foot straight 2
1. For no combustible sensor, enter a “0.” contaminated with dirt and debris if the 3-foot, holes near tip 3
Oxygen Sensor Code probe tip is dipped in water. You must order 3-foot, holes near handle 4
B. Oxygen Sensor Option a probe if you order a sampling line as we
None 0 Hot-gas sampling probe 5
To choose to have an oxygen sensor strongly recommend you operate the
installed in your new Orion Multigas Oxygen (0–25%) 1 instrument with this filter in place.
Detector, choose selection 1. For no oxygen
sensor, enter a “0.” J. Carrying Attachments Belt Clip Code
While a high-strength, nylon clip is offered Standard nylon 0
C. Carbon Monoxide Sensor Option Carbon Monoxide Sensor Code with the instrument at no charge, a swivel Swivel belt loop 1
To choose to have a carbon monoxide sen- None 0 belt loop kit, which allows quick release
sor installed in your new Orion Multigas Carbon monoxide (0–999 ppm) 1 from your belt with a half turn, is also
Detector, choose selection 1. For no carbon available.
monoxide sensor, enter a “0.” Instrument Jacket Code
K. Instrument Jacket
D. Hydrogen Sulfide Sensor Option Hydrogen Sulfide Sensor Code For the highest levels of environmental and None 0
To choose to have a hydrogen sulfide sensor None 0 rough handling protection, you can order Black Rubber Boot B
installed in your new Orion Multigas Hydrogen sulfide (0–200 ppm) 1 from a variety of jackets and boots for your Black Leather Case L
Detector, choose selection 1. For no hydro- Orion Multigas Detector. Orange Cordura Jacket C
gen sulfide sensor, enter a “0.”
E. Battery Pack Options Battery Pack Code L. Calibration Kits Calibration Accessories Code
You can choose between MSA’s NiMH Standard NiMH rechargeable A You can choose one of two calibration None 0
rechargeable batteries which will run your (comes with standard AC charger) options for your Orion Multigas Detector. Model RP Calibration Kit 1
Orion Multigas Detector for up to 16 hours The flexible RP kit provides a constant flow Squirt Gas Check Kit 2
Replaceable alkaline B
in pump mode and 20 hours in diffusion regulator that is reused and provides the
Both NiMH and alkaline C greatest accuracy. The squirt gas test kit
mode, or you can choose the convenience
of alkaline batteries. For ultimate flexibility, provides a very economical means to verify
choose “both!” instrument calibration.

F. Battery Charger Options Battery Charger Code M. Data Communications Kit Data Communications Code
If you order your Orion Multigas Detector None/Standard AC charger 0 If you need to track the readings and None 0
with a rechargeable battery pack, the Vehicular charger 1 operation of your Orion Multigas Detectors, Orion link module 1
charger is included in the price. If you want consider ordering them with the optional Module and download 2
the optional vehicular charger adapter, built-in datalogging module. This option software
choose selection “1.” also provides STEL and TWA capabilities.
G. Sampling Methods Sampling Methods Code N. Packaging Options & Carrying Cases Packaging & Carrying Cases Code
The Orion Multigas Detector’s patented None D Std shipping container only 0
You can further protect your Orion Multigas
Pulsecheck™ pump technology leads the Integral, Pulsecheck pump P Detector with optional carrying cases. For Black plastic case w/space 1
industry. If you chose the built-in pump, Manual aspirator module A ease of transport and handling, these cases for cylinder, with video
operating directly off the instrument bat-
can be even be ordered to accommodate Heavy duty case with video 2
teries, you are assured of an audible pump
calibration accessories.With the premium Large, heavy duty case with 3
fault alarm and you do not need to main-
cases, a simple to understand training video and space
tain a second set of batteries. If you are
videotape is included at no extra charge. for 2 cylinders
ordering diffusion instruments, a calibra-
tion cap will be included to calibrate your Orange plastic case 4
instruments. w/space for cylinder,
with video
H. Sampling Lines Sampling Lines Code
materials guaranteed to not alter sample None 0
concentration. Screw fittings with O-Ring 3-foot coiled 1
couplings at each end ensure there are no 5-foot coiled 2
leaks when the system is operating.
10-foot 3
15-foot 4
25-foot 5

Your Model Part Number


Please write in numbers from selections above in appropriate boxes and contact your Safety Products Distributor to place your order.

Selection A B C D E F G H I J K L M N
A-ORION- - - - - - - - - - - - - -
Solaris™ Multigas Detector ne w!
he Solaris Multigas Detector is an affordable, durable, reliable, easy-to-use portable instrument for detecting the
resence of O2, H2S, CO and combustible gas. It is designed to withstand rough handling in harsh environments.
est of all, it delivers MSA’s commitment to quality at the smallest size and, even smaller price.

Features
rgonomic Design Lightweight TIM System Compatible
or maximum comfort. Under 8 oz., compact size that Compatibility for computer
makes it easy to wear. automated calibration and
Alarm System record-keeping.
Durable
World-class triple alarm system Sampling Pump Option
with 100+ decibel audible alarm, Case with rubberized armor
multiple high-intensity visual provides superior protection Optional powered sampling
larm and strong vibrating alarm. against liquid and dust ingress pump with extensive contaminant
(IP 65 rated. filtering system.
DisplayFeatures
Long-life Battery W arranty
uperior display features simulta-
eous gas concentration display, Rechargeable lithium ion battery Outstanding 2-year, all-inclusive
acklighting for easy reading and provides 14+ hours of continuous warranty.
lphanumeric message bar for run time and delivers best-in-
asy use. class performance in extreme
conditions.

Solaris Kits
Basic Kits Industrial Kits (includes Econo-cal calibration kit)
-Gas Instrument 10048148 4-Gas Instrument 10049293
-Gas (LEL, O2, CO) 10048150 3-Gas (LEL, O2, CO) 10049294
-Gas (LEL, O2, H2S) 10048211 3-Gas (LEL, O2, H2S) 10049295
-Gas (LEL, O2) 10048213 2-Gas (LEL, O2) 10049296

Standard Kits (includes datalogging option)* Deluxe Kits (includes Econo-cal calibrationkit, datalogging &
-Gas Instrument 10048147 sampling pump)*
-Gas (LEL, O2, CO) 10048149 4-Gas Instrument 10049297
-Gas (LEL, O2, H2S) 10048212 3-Gas (LEL, O2, CO) 10049298
-Gas (LEL, O2) 10048214 3-Gas (LEL, O2, H2S) 10049299
2-Gas (LEL, O2) 10049300

Assemble-to-Order (ATO) System:


You Make the Choices
he ATO System makes it easy to “custom order”
he Solaris Detector, configured exactly the way
ou want it.You can choose from an extensive line
f base instrument components and accessories.
ee the ATO chart on the following page to make
our selections.

or a full descripton ofthe Solaris Detector s


ccessories,please refer to pages 67—68.

ee pages 82—86 for calibration accessories.


Solaris™ Multigas Detector Assemble-to-Order (ATO) Options

A. Datalog Options Catalog Options Code


The datalog option enables the Solaris Multigas Detector to store exposure data Standard Instrument S
for easy retrieval at a later time. To download the Solaris data the MSA datalog kit Datalogging option selected D
(P/N 710946) must also be purchased.

B. Combustible Sensor Options Combustible Sensor Code


If a combustible gas sensor is desired, two options are available: 0-100% LEL or None 0
0-5% CH4. 0-100 % LEL 1
0-5 % CH4 2

C. Oxygen Sensor Option Oxygen Sensor Code


To choose to have an oxygen sensor installed in your new Solaris Multigas Detector, None 0
choose selection 1. O2 Sensor 1
For no oxygen sensor, enter a “0”.

D. Carbon Monoxide Sensor Option Carbon Monoxide Sensor Code


To choose to have a carbon monoxide sensor installed in your new Solaris Multigas None 0
Detector, choose selection 1. For no carbon monoxide sensor, enter a “0”. CO Sensor 1

E. Hydrogen Sulfide Sensor Option Hydrogen Sulfide Sensor Code


To choose to have a hydrogen sulfide sensor installed in your new Solaris Multigas None 0
Detector, choose selection 1. For no hydrogen sulfide sensor, enter a “0”. H2S Sensor 1

F. Battery Charger Options Battery Charging Options Code


A Solaris-specific battery charger must be used to recharge the None 0
batteries in the Solaris unit. Global battery charger comes with North American outlet-compatible N
several international plug outlet configurations for countries Global outlet-compatible G
outside of North America.

G. Instrument Approvals Battery Charger Code


It is now possible to select the specific type of approval that a Solaris instrument Underwriter Laboratory (UL) U
possesses. Default approval for North America will be granted by Underwriters Mine Safety & Health Admin
Laboratory (UL). The Solaris UL mark carries both United States and Canadian (MSHA) M
approvals. MSHA approval is needed for mining applications that are specifically
regulated by MSHA procedures.

H. Sampling Pump Sampling Pump Code


It is now possible to combine the sampling pump and probe for the Solaris unit into None 0
one easy-to-use accessory. The pump probe is very versatile and can be attached to Universal pump probe 1
the Solaris instrument in seconds. The built-in Water Stop filter helps to ensure the
sampling system is not contaminated with dirt and debris during use.

I. Calibration Kits Calibration Kits Code


MSA provides two options for performing instrument calibration. The cost-effective None 0
Econo-Cal (34 liter) cylinder and fixed flow regulator provides an economical means Econo-Cal Calibration Kit 1
to calibrate the Solaris instrument. The Model RP calibration kit also has a fixed flow Model RP Calibration Kit 2
0.25 lpm regulator and provides the user with a larger capacity (58 liter) calibration
gas cylinder.

Your Model Part Number


Please write in numbers from selections above in appropriate boxes and contact your Safety Products Distributor to place your order.

Selection A B C D E F G H I
Watchman® Multigas Monitor
he Watchman Multigas Monitor is a durable, portable instrument used to detect
nd monitor combustible gases, oxygen, and toxic gases in workplace atmospheres,
specially in confined spaces such as manholes, storage tanks, tank cars, vaults, mines,
nd sewers. Designed for rugged handling, the monitor incorporates the state-of-the-art
echnology in a strong aluminum housing.The Watchman Monitor is ideal for long jobs in
eavy industrial environments, as the instrument will run up to 16 hours on a single charge.
or more complete information, see Data Sheet 08-02-39.

Approvals
he Watchman Multigas Monitor meets intrinsic safety testing requirements for use
n Class I Division 1, Groups A, B, C and D; Class II Division 1, Groups E, F, G; and Class III
azardous locations.

Watchman Multigas y
g
/Ta r op
Monitor Kits isp
la
/T WA arge Line
h e er St rap
T
10 Pro W a Line
D S ak EL C C ft b t
LEL O 2 CO H2 Pe ST A PartNo. Notes
ndustrial 4-Gas • • • • • • • • • 711132
ndustrial 4-Gas • • • • • • • • • • 711406
ndustrial 3-Gas • • • • • • • • • 711407 Replaces Model 360
ndustrial 3-Gas • • • • • • • • • 711408 Replaces Model 361
ndustrial 2-Gas • • • • • • • • 711409 Replaces Model 260

Assemble-to-Order (ATO) System: You Make the Choices


he ATO System makes it easy to “custom order” the Watchman Multigas Monitor, configured
xactly the way you want it.You can choose from an extensive line of base instrument components
nd accessories. See the ATO chart on the following page to make your selections.

or further information on W atchman Multigas Montior accessories,please refer to pages 67—68.

ee pages 82—86 for calibration accessories.


Watchman® Multigas Monitor Assemble-to-Order (ATO) Options

A. Electrochemical Sensor Configuration Sensors (Select 4) Code H. Sampling Lines Sampling Lines Code
Options Sulfur Dioxide (0-100 ppm) 6 The Watchman Monitor’s sampling lines are None 0
Please list the code number for the toxic gas Nitrogen Dioxide (0-100 ppm) 5 of the highest quality materials guaranteed 5-foot 1
sensors desired in the spaces to the right. In Nitric Oxide (0-100 ppm) 4 to not alter sample concentration. The 10-foot 2
order to properly process your order: Hydrogen Sulfide (0-200 ppm) 3 heavy-duty synthetic rubber material is 15-foot 3
• list them in descending order Carbon Monoxide (0-2000 ppm) 2 ideally suited to industrial applications 25-foot 4
• use zeros to fill any remaining Oxygen (0-25%) 1 where heat and chemical resistance are 50-foot 5
selections (example: 2100) crucial to instrument performance. Screw
None 0
• add the price of each sensor to obtain fittings with O-Ring couplings at each end
the sensor sub-total pricing help reduce the possibility of leaks when the
monitor is operating.
B. Combustible Gas Sensor Options Sensors Code
If a combustible gas sensor is desired, None 0 I. Sampling Probes Sampling Probes Code
select type of display. There are two options 0–5% CH 4 (Methane) M MSA’s sampling probes are durable and easy None 0
available: 0–100% LEL Pentane L to use. An assortment of probe types is avail- 18-inch plastic 1
0-5% Methane (%CH4) or able to suit your applications. You can select 3-foot brass 2
0-100% Lower Explosive Limit (%LEL) between heavy-duty nickel plated brass 3-foot plastic 3
probes or phenolic plastic probes where 4-foot solid brass 4
C. Peak Page Peak Page Code electrical conductivity might be a concern.
Allows user to view the peak gas readings Not enabled 0
J. Line Traps Line Traps Code
for the entire time the instrument is run- Peak display enabled 1
ning. The user can access this page by press- The Watchman Monitor is offered with two None 1
ing simply the “PAGE” button on the tester types of sample line protection from water Water Stop 2
keypad. ingress: a water stop filter and a water trap. Line Trap Assembly 3
The convenient, easy-to-use in-line water Both 4
D. STEL Page STEL Page Code stop filter is a membrane filter that will
Enables page and alarms for 15-minute- Not enabled 0 prevent liquids from being drawn into the
weighted-average gas exposures. To enable STEL display enabled 1 instrument. The water trap provides a water
the STEL page, enter a “1” in the box to the collection bulb with a stop to automatically
right, otherwise, enter a “0”. cut off the pump flow when the bulb fills.
The water trap provides a visual indication
E.TWA Page TWA Page Code you are drawing water into your sampling
Enables page and alarms for 8-hour- Not enabled 0 lines.
weighted-average gas exposures. To enable TWA display enabled 1
the TWA page, enter a “1” in the box to the K. Remote Alarms Remote Alarms Code
right, otherwise, enter a “0”. The Watchman Monitor’s remote alarms None 0
F. Datatagging Page TAG Page Code provide an additional level of audible and 90-dB horn / 3-foot cable 1
Allows user to enter a text “tag” using the Not enabled 0 visual alerts to your gas monitoring 90-dB horn / 15-foot cable 2
keypad to denote a location or note with a program. Available in low- and high-sound 105-dB horn / 3-foot cable 3
Datatagging pages enabled 1
time stamp. To enable the datatagging output and with an optional 15-foot exten-
sion cable, the remote alarms ensure your 105-dB horn / 15-foot cable 4
page, enter a “1” in the box to the right,
otherwise, enter a “0”. team is notified in high-noise areas.

G. Battery Charger Options Battery Charger Code L. Packaging Options Packaging Options
From wall adapters to vehicular chargers, None 0 You can further protect your Watchman Standard packaging 0
you can recharge your batteries wherever Vehicle, 8–24VDC 1 Monitor with a high-strength, weatherproof Premium field case with 1
you need. Once recharged, your Watchman Single-unit 110 V 2 carrying case. For ease of transport and compartments for calibration
monitor is ready to provide you with up to handling, the premium case accommodates accessories and bonus
Single-unit 220 VC 3
15 hours of continuous gas monitoring. the instrument along with its sampling and training video
Five-unit, 110/220 VC 5
calibration accessories. A simple-to-under-
stand training videotape is included at no
extra charge. (Standard packaging shown.)

Your Model Part Number


Please write in numbers from selections above in appropriate boxes and
contact your Safety Products Distributor to place your order.

Selection A A A A B C D E F G H I J K L
A-WTM- - - - - - -0
List sensors in descending
numerical order.
Passport® Personal Alarm
MSA’s Passport Personal Alarm is the world’s top-selling For more complete information, please
multigas detector, with over 50,000 shipped to customers refer to Data Sheet 08-00-39.
worldwide. Its durability is field-proven in the toughest
ndustries around the globe. As the first gas detector with Approvals
mplified three-pushbutton operation, the Passport
The Passport Personal Alarm has been
ersonal Alarm set the standard for simplicity, and users
approved for use in Class I Division 1,
ypically require less training to operate it.
Groups A, B, C and D; Class II Division 1,
he Passport is packed with the features needed to help Groups E, F, G; and Class III hazardous
nsure worker safety. Capable of monitoring up to five locations.
ases simultaneously, it is equipped with bright, multidi-
ectional alarm lights and a loud alert horn.The display
aces the user, allowing hands-free operation when worn Assemble-to-Order (ATO) System:
n the belt or on a strap around the neck.The Passport You Make the Choices
larm’s optional high-efficiency pump is powered by the The ATO System makes it easy to “custom order” the Passport Personal Alarm, configured exact-
nstrument, which provides audible and visual alerts in the ly the way you want it.You can choose from an extensive line of base instrument components
vent of flow faults.The pump can be removed to provide and accessories. To obtain a copy, request Bulletin 0808-03-MC. To obtain a copy of the ATO via
iffusion-mode operation.The instrument also records gas FAX, call MSA QuickLit Information Service at 1-800-672-9010. At the prompt, request QuickLit
xposure readings for up to 12 hours for later retrieval if Document #2390 (ATO for MicroGard Portable Alarm).
ecessary.
be Forafull
y ro
Passport Alarm Kits tP ket t
g er er p descripton
Kit
A / harg e Ba Pu m lPu ne/1 Jac acke Cas ase Test
Ta tt p m f e
y ofthe
la W
isp T C in n ra Li ge J ge C
D S ak TEL/ iCd/ lkal dd-o nteg 10ft ran lack ran lack Bump Fi
veStar
LEL O 2 CO H2 Pe S N A A I O B O B PartNo.
Alarms
ndustrial 4-Gas • • • • • • • • 711403 accessories,
ndustrial 4-Gas • • • • • • • • • • 711404 please refer
ndustrial 4-Gas • • • • • • • • 803912 to pages
ndustrial 3-Gas • • • • • • • 711405 79—80.
ndustrial 3-Gas • • • • • • • 803913
See pages
ndustrial 3-Gas • • • • • • • 803894 96—100 for
ndustrial 4-Gas • • • • • • • • • • • 711129 calibration
ndustrial 4-Gas • • • • • • • 801092 accessories.
ndustrial 3-Gas • • • • • • • • • • • 711131
ndustrial 3-Gas • • • • • • 803892
ndustrial 3-Gas • • • • • • • • • • • 711130

MicroGard® Portable Alarms


MicroGard Portable Alarms are The MicroGard Alarm features For more complete information,
ocket-sized instruments for state-of-the-art circuitry coupled see Data Sheet 08-00-40.
multaneously monitoring with proven MSA combustible gas
ombustible gas and oxygen and oxygen sensors for reliable Approvals
oncentrations in workplace protection. In the diffusion mode, All MicroGard Portable Alarms have
tmospheres and confined spaces, the instrument is ideal for been approved for use in hazardous
uch as manholes, storage tanks, handheld, pocket, or belt use. locations, Class I, Division 1, Groups
ank cars, and shipyards. A, B, C, and D.

MicroGard Portable Assemble-to-


y
y y r y atter
Order (ATO)
e
Alarm Kits la la tt B
isp Disp Ba line p
be ne ket
System:
D 4 Cd a m
LEL CH O 2 Ni Alk Pu Pro Li Jac PartNo.
ndustrial Kit #1 • • • 711410 You Make
ndustrial Kit #2 • • • • • • 711411 the Choices
711412 The ATO System makes it easy to “custom order” the
ndustrial Kit #3 • • •
MicroGard Alarm, configured exactly the way you want
ndustrial Kit #4 • • • • • • 711413 it.You can choose from an extensive line of base instru-
ndustrial Kit #5 • • • • 711414 ment components and accessories. To obtain a copy,
ndustrial Kit #6 • • • • • • • 711415 request Bulletin 0808-03-MC. To obtain a copy of the
ATO via FAX, call MSA QuickLit Information Service at
ndustrial Kit #7 • • • • 711416
1-800-672-9010. At the prompt, request QuickLit
ndustrial Kit #8 • • • • • • • 711417 Document #2390 (ATO for MicroGard Portable Alarm).
Passport® PID II Organic Vapor Monitor
and Passport® VOC 2000 Monitor Passport® PID II Monitor

Passport PID II Organic Vapor Monitor


The Passport PID II Organic Vapor Monitor is a convenient, portable instrument for
detecting low concentrations of volatile organic compounds (VOCs).The Passport
PID II Monitor’s compact design supports a variety of monitoring applications.
Used primarily in connection with hazardous materials spills, soil remediation,
and underground tank leakage, the Passport PID II Monitor can also be
used for emissions monitoring and general leak detection. For more
complete information see Data Sheet 08-01-39.

Passport VOC 2000 Monitor


The Passport VOC 2000 Monitor integrates an additional
Approvals
electrochemical oxygen sensor to extend its usefulness for many
confined space applications. For more complete information, see MET (an OSHA-NRTL); Class I, Division 1,
Bulletin 0816-19-MC. Groups A, B, C, and D; Class II, Division 2,
Passport® VOC 2000 Monitor Groups E, F and G; and Class III, Division
PassportPIDIIKits 1 and 2.
Kit includes Passport PID II, NiCd battery pack, 110V charger,
10-ft coiled Teflon sampling line, 1-ft probe, PVC case 711401
Kit includes Passport PID II, alkaline battery pack, pistol grip, Common Compounds Detectable with
8-in probe, PVC case 711402 Passport PID II Monitor
Acetone Octane
PassportVOC 2000 Kits Benzene Propylene
Kit includes Passport VOC 2000, NiCd battery pack, 110V charger, Chlorobenzene Phosphine
10-ft coiled Teflon sampling line, 1-ft probe, stainless steel belt clip, Cyclohexane Toluene
standard packaging 10010864 Heptane Trichloroethane
Kit includes Passport VOC 2000, NiCd battery pack, 110V charger, Hexane Many Others
8-in probe, pistol grip, standard packaging 10010865 Methyl Isobutyl Ketone (MIBK)
Methyl Tetrabutyl Ether (MTBE)
Assemble-to-Order (ATO) System:You Make the Choices
The ATO System makes it easy to “custom order” the Passport PID II Organic Vapor
Monitor or Passport VOC 2000 Monitor, configured exactly the way you want it.
You can choose from an extensive line of base instrument components and
accessories.To obtain a copy of the Passport PID II Organic Vapor Monitor ATO,
request Bulletin 0816-04-MC.To obtain a copy of the Passport VOC 2000 Monitor
ATO, request Bulletin 0816-20-MC.

To obtain a copy of either ATO via FAX, call MSA QuickLit Information Service at
1-800-672-9010. At the prompt, request QuickLit Document #2391 (ATO for
Passport PID II Organic Vapor Monitor), or request QuickLit Document #2245
(ATO for Passport VOC 2000 Monitor).

For further information on PassportPIDII and VOC 2000 accessories,please refer to Passport® PID II Monitor in action
pages 67—68.

See pages 82—86 for calibration accessories.


Orion® G Multigas and Leak Detector
he Orion G Multigas and Leak Detector is the only instrument you need for
etecting combustible gas leaks on appliances, flues and underground pipes.
nd with its internal O2, H2S, CO and LEL sensors, it also is great for confined
pace applications.

Features
Multi-Function Flexibility Tough, Durable Construction
indoor or outdoor (or both) • excellent resistance to penetration
configure any combination of gases from water and dust
including O2, CO and H2S • weather-resistance rating of IP54
for low-level leaks, 0-100% LEL or
CH4 0-100% by volume Large, Easy-To-Read Display
• one of the largest displays of any
ar-Hole Probe Filter Trap multigas detectors on the market
bar-hole probe filter trap is designed to • backlighting provides exceptional Orion® G Multigas
keep out water and dirt contrasts and Leak Detector
"quick connect" feature allows user to with Flexible Probe
attach probe or sampling line while TIM¤ Compatibility
minimizing the possibility of air leaks
• TIM® (Total Instrument Manager)
System allows users to automatically
uperior Design calibrate their instruments and track
calibration records.
sensor is located on the end of a flexible Orion® G Multigas and Leak
gooseneck probe, thereby allowing Detector with Coiled Probe
sensor to be placed closer to item being Built-InPump
tested
• built-in pump for remote sampling of the
super-fast response time
environment and fast response to leaks.
fits easily and comfortably in the hand
intrinsically safe.
k n e se
ac k io Ca

e
de ide ry P Pac er t ct Lin

as
i e ry
ar

C
ong-Life Battery ox ulf te y rg ec n
n
ng Ki
t
OrionG Multigas and on n S Bat tter ha nn Co pli e

a
t n ing tic C

ur
o
accommodates either a NiMH battery Leak Detector Kits
o

or
M e a C o ed m b o t i g s
L gen on rog line h B cle k C ad Sa ro k B ra og Pla

C
pack or alkaline batteries ib al
Flexible Probe LE xy arb yd lka im ehi uic hre 0 ft ft P lac

w
al at ed

lo
provides up to 16 hours of continuous

l
O C H A N V Q T 1 1 B C D R

Ye
%
instrument operation Deluxe 4 gas 10041504 • • • • • • • • • • • • • •
Deluxe 3 gas (H2S) 10041505 • • • • • • • • • • • • •
urable Sensors
Deluxe 3 gas (CO) 10041506 • • • • • • • • • • • • •
MSA Series 20L sensor provides up to
Industrial 4 gas 10041507 • • • • • • • • • •
50 times the silicone poison tolerance,
considerably expanding sensor Industrial 3 gas (H2S) 10041508 • • • • • • • • •
service life
Industrial 3 gas (CO) 10041509 • • • • • • • • •
fast-reacting, solid-state 20G sensor in
tip of probe Economy 4 gas 10041510 • • • • • • • •
all sensors are easy to maintain Economy 3 gas (H2S) 10041511 • • • • • • •
Economy 3 gas (CO) 10041512 • • • • • • •
One-Button Calibration
Supereconomy 4 gas 10041513 • • • • • •
calibration is accomplished with
Supereconomy 3 gas (H2S) 10041514
push of a single button (with appropriate • • • • •
calibration gas) Supereconomy 3 gas (CO) 10041515 • • • • •

OrionG Multigas and Leak Detector Kits Coiled Probe


Deluxe 4 gas 10048330 • • • • • • • • • • • • • •
Deluxe 3 gas (H2S) 10048331 • • • • • • • • • • • • •
Deluxe 3 gas (CO) 10048332 • • • • • • • • • • • • •
Industrial 4 gas 10048333 • • • • • • • • • •
Industrial 3 gas (H2S) 10048334 • • • • • • • • •
Industrial 3 gas (CO) 10048335 • • • • • • • • •
Economy 4 gas 10048336 • • • • • • • •
Economy 3 gas (H2S) 10048337 • • • • • • •
Economy 3 gas (CO) 10048338 • • • • • • •
Supereconomy 4 gas 10048339 • • • • • •
Supereconomy 3 gas (H2S) 10048340 • • • • •
Supereconomy 3 gas (CO) 10048341 • • • • •
Assemble-to-Order (ATO) System: You Make the Choices
The ATO System makes it easy to “custom order” the Orion G Multigas and Leak Detector, configured exactly the way you want it.You can choose from an
extensive line of base instrument components and accessories. See the ATO chart below to make your selections.

For further information on Orion G Detector accessories,please refer to pages 67—68.See pages 82—86 for calibration accessories.

Orion® G Multigas and Leak Detector Assemble-to-Order (ATO) Options

A. Combustible Sensor Options Combustible Sensor Code I. Sampling Probes Sampling Probe Code
To choose to have your MSA 20L % LEL (Methane, 0–100%), 0 MSA’s sampling probes are durable and None 0
combustible sensor gas reading factory Flexbile probe comfortable to use. The built-in Water Stop 1-foot straight 1
set to be displayed as %LEL or %Volume, % Volume (0–100%), Flexbile probe 1 filter ensures your sampling system is not 3-foot straight 2
enter a selection for display reading and contaminated with dirt and debris if the Bar hole probe 3
probe type. % LEL (Methane, 0–100%), 2 probe tip is dipped in water. You must order
Coiled probe 8-inch Stix 4
a probe if you order a sampling line as we Hot gas probe 5
% Volume (0–100%),Coiled probe 3 strongly recommend you operate the
instrument with this filter in place.
B. Oxygen Sensor Option Oxygen Sensor Code Sampling connection method chosen above
To choose to have an oxygen sensor None 0 determines type of fitting supplied with
installed in your new Orion G Multigas and Oxygen (0–25%) 1 sampling line.
Leak Detector, choose selection 1. For no
oxygen sensor, enter a “0.” J. Carrying Attachments Belt Clip Code
While a high-strength, nylon clip is offered Standard nylon 0
C. Carbon Monoxide Sensor Option Carbon Monoxide Sensor Code with the instrument at no charge, a swivel Swivel belt loop 1
To choose to have a carbon monoxide sen- None 0 belt loop kit, which allows quick release Pistol grip 2
sor installed in your new Orion G Multigas Carbon monoxide (0–999 ppm) 1 from your belt with a half turn, is also
and Leak Detector, choose selection 1. For available. For easier one-hand operation,
no carbon monoxide sensor, enter a “0.” choose the pistol grip option.
D. Hydrogen Sulfide Sensor Option Hydrogen Sulfide Sensor Code K. Instrument Jacket Instrument Jacket Code
To choose to have a hydrogen sulfide sen- None 0 For the highest levels of environmental and None 0
sor installed in your new Orion G Multigas Hydrogen sulfide (0–200 ppm) 1 rough handling protection, you can order a Black rubber boot B
and Leak Detector, choose selection 1. For jacket or boot for your Orion G Multigas Yellow cordura jacket C
no hydrogen sulfide sensor, enter a “0.” and Leak Detector. Both boot and jacket D
E. Battery Pack Options Battery Pack Code
You can choose between MSA’s NiMH L. Calibration Kits Calibration Accessories Code
Standard NiMH rechargeable A
rechargeable batteries which will run your You can choose one of two calibration None 0
(comes with standard AC charger)
Orion G Multigas and Leak Detector for options for your Orion G Multigas and Leak Model RP Calibration Kit 1
Replaceable alkaline B
up to 16 hours, or you can choose the con- Detector. The flexible RP kit provides a (includes gas cylinder)
venience of alkaline batteries. Also avail- Replaceable alkaline with C
constant flow regulator that is reused and Squirt Gas Check Kit 2
able in easy-to-change alkaline with thumbscrew attachment
provides the greatest accuracy. The squirt (includes gas cylinder)
thumbscrew attachment. gas test kit provides a very economical
F. Battery Charger Options Battery Charger Code means to verify instrument calibration.
If you order your Orion G Multigas and None/Standard AC charger 0 M. Data Communications Kit Data Communications Code
Leak Detector with a rechargeable battery Vehicular charger 1 If you need to track the readings and None 0
pack, the charger is included in the price. If operation of your Orion G Multigas and Orion link module 1
you want the optional vehicular charger Leak Detectors, consider ordering them Module and
adapter, choose selection “1.” with the optional built-in datalogging download software 2
Sampling Connection Methods Code module. This option also provides STEL and
G. Sampling Connection Methods TWA capabilities.
The Orion G Multigas and Leak Detector Threaded A
offers the flexibility of using standard Quick-connect B N. Packaging Options & Carrying Cases Packaging & Carrying Cases Code
threaded connections or the new Quick- You can further protect your Orion G Std shipping container only 0
Connect air-line fittings. Multigas and Leak Detector with optional Black plastic case with video 1
carrying cases. For ease of transport and and space for 2 cylinders
H. Sampling Lines Sampling Lines Code handling, these cases can be even be Heavy-duty case with video 2
MSA’s high-speed sampling lines are of the None 0 ordered to accommodate calibration acces- and space for 2 cylinders
highest quality materials guaranteed to 3-foot coiled 1 sories.With the premium cases, a simple to
not alter sample concentration. Threaded 5-foot coiled 2 understand training videotape is included
fittings with O-Ring couplings at each end 10-foot 3 at no extra charge.
ensure there are no leaks when the system 15-foot 4
is operating. Also available in a Quick-
25-foot 5
Connect version for easy attaching and
5-foot 6
tight seal. Sampling connection method
chosen above determines type of fitting
supplied with sampling line.

Your Model Part Number


Please write in numbers from selections above in appropriate boxes
and contact Safeware to place your order.

Selection A B C D E F G H I J K L M N
A-ORION G- - - - - - - - - - - - - -
Combustible Gas Indicators
Gascope® Combustible Gas Indicators and
Tankscope® Combustible Gas Indicator, Model 62T
Gascope Combustible Gas Indicators are portable instruments for use in detecting,
measuring, and pinpointing leaks of combustible gases and vapors. Each unit is easily
arried with integral neck and waist straps, leaving hands free for climbing, operating
he instrument, or carrying additional equipment.
hree Gascope Combustible Gas Indicator models are available.The Model 60 is
esigned for use by gas utility companies in routine testing for methane-in-air
oncentrations in manholes, sewers, curb boxes, and other street openings.The
Model 62S, also suitable for use by gas utility companies, is designed for reading
–100% LEL methane-in-air and 0–100% by volume methane-in-air.The Model 62 is
or general industrial use and is designed for reading 0–100% LEL pentane-in-air. For
more complete information, see Data Sheet 08-01-04.
he Tankscope Combustible Gas Indicator, Model 62T, is a portable instrument
Gascope® Combustible Gas Indicator
pecifically designed to detect combustible gas leaks in inerted shipboard oil holds.
ased on the highly successful Gascope Combustible Gas Indicator, the portable
ankscope Indicator has an easy-to-read analog display and is very simple to use. Approvals
Gascope Combustible Gas Indicators are listed by MET for use in
Gascope/Tankscope Indicators Class I, Division 1, Groups C and D hazardous locations defined by the
Complete with carrying straps and batteries,less sampling line. National Electric Code.
Utility Model 60 465475
Model 62 465681
Utility Model 62S 468410
Model 62T,Tankscope 711258

Explosimeter® Combustible Gas Indicator


he Explosimeter Combustible Gas Indicator detects and measures concentrations of combustible
ases or vapors in the air.The instrument can be used in the immediate environment or, with
ampling lines and probes, it can draw samples from remote areas.The unit is housed in a cast
luminum case.

pprovals
Model 2A is listed by Underwriters’Laboratories (UL) for use in hazardous locations as defined by
he National Electrical Code. UL approval is for Class I, Group D, Divisions 1 and 2; and Class I, Groups
, B and C, Division 2 (Division 1 excluded), hazardous atmospheres. Explosimeter® Combustible Gas Indicator

or more complete information, see Data Sheet 08-00-03.

xplosimeter Indicator
omes complete with carrying straps, less sampling line.
escription Applications Part No.
Model 2A General testing for combustible gases or vapors in the air 89220

or further information on Gascope and Explosimeter accessories,please refer to pages 67—68.

ee pages 82—86 for calibration accessories.


Multigas Instrument Accessories
Replacement Sensors
Replacement Sensors FiveStar Orion Passport Orion G Watchman Solaris MicroGard
Combustible 710317 10024247 478537 10024247 478537 10046947 478537
Oxygen 815333 10025940 480566 10025940 480566 10046946 480566
Carbon Monoxide 710656 711306 636240 711306 636240 10046944 —
Hydrogen Sulfide 710657 711307 636241 711307 636241 10046945 —
Nitric Oxide 710662 — 808350 — 808350 — —
Nitrogen Dioxide 710663 — 807477 — 807477 — —
Sulfur Dioxide 710664 — 807476 — 807476 — —
Chlorine 10012282 — 807478 — — — —
Phosphine 10011530 — — — — — —
Ammonia 10012286 — 637577 — — — —
Methane Leak (external) — — — 10040784 — — —
Prefilter for CO sensor — — — — — 10047967 —
Protective felt for sensor inlets — — — — — 10044927 —
Cover label for empty sensor location — — — — — 10049052 —

Batteries
FiveStar FiveStar Passport,
Batteries Black ABS Ni-Plated Orion PID II,VOC 2000 Watchman MicroGard Orion G
Rechargeable NiCd 710427 10008606 — 496990 814127 478510 —
Heavy-Duty NIMH 10008642 10013145 10031091 800527 — — 10031091
Replaceable “AA” 710955 10008607 10031092 800526 — 482245 10031092
Replaceable, w/thumbscrews — 711041 10042020 — — — 10042020
Battery Pack Jack Cover — — — — — — 10030901

Chargers Sampling Pumps


FiveStar, Passport, Sampling Pumps FiveStar Orion Passport MicroGard Solaris
Watchman, Orion, Calibration Cap 710824 10020550 497367 478530 10044994
Chargers PID II,VOC 2000 Orion G MicroGard Solaris
Calibration Cap Ass’y — — — — 10044995
Charger, 110V 494716 10020551 633548 —
Pump Module 710790 — 497430 802830 —
Charger, 220V 495965 10020551 — —
Integral Pump — — 814105 — —
Charger,Vehicle 710423 10026502 478490 10049410
Aspirator Module 710959 10020545 809964 485800 —
Charger, 5-Unit 801759 — 478540 —
Pump, Chlorine
Charger Stand* 710570 — — — Compatible† 10018395 — 811719 — —
Fast Charger, NiCd* 710989 — — — Universal Pump Probe
Fast Charger, NiMH-NiCd* 10018394 — — — (UL approved) — — — — 10046528
Charger Cradle Assembly — — — 10048185 Universal Pump Probe
North American Power Supply — — — 10047342 (MSHA approved) — — — — 10047595
Global Power Supply — — — 10047343
*Use with P/N 478530 †Other restrictions apply. Call MSA Customer Service for details.
*FiveStar only.

Sample Lines Probes


FiveStar, Orion, Passport, Watchman, Gascope, FiveStar, Orion, Passport, Watchman, Gascope,
Sample Lines MicroGard, PID II,VOC 2000 Orion G Explosimeter Probes MicroGard, PID II,VOC 2000 Explosimeter Orion G
5-foot, coiled 710465 710465 — 1-foot, plastic 800332 — 800332
5-foot 497332 497332 11354 18-inch, plastic — 486934 —
10-foot 497333 497333 11955 3-foot, plastic 800333 73743 800333
15-foot 497334 497334 11912 3-foot, brass — 11961 —
25-foot 497335 497335 11913 3-foot, bar hole 803561 — 10040572
50-foot — — 11958 3-foot, solid 800562 — —
5-foot,Teflon 800972 — Hot gas 803848 — 803848
10-foot,Teflon 811187 — —
3-foot, coiled — 10018118 —
Sampling System Filters
FiveStar, Orion,
Passport Watchman Application
Water Trap — 710459 Removes water before it enters instrument
n-Line Water Stop Kit — 711257 Blocks water in sample line
Replacement Water Stop Filters 801582 (10 ea.) 710445 (5 ea.)
harcoal Filter Holder 801291 14273 Helps eliminate organic vapors or differentiate methane from other combustible gases
Replacment Charcoal Filters 14318 14318
ine Scrubber Holder 711039 — Helps eliminate erroneously high CO readings for unsaturated hydrocarbons (e.g., ethylene)
Line Scrubber Cartridges 711052 —
Dilution Tube, 1:1 813514 — Helps allow sampling in inert atmospheres
n-Line Dust Filter 813409 —
Dust Filter Replacements 813971 (10 ea.) —
nd-of-line Dust/Water Filter 806160 — Helps prevent drawing dust and water into the sampling probe
Replacement FIlters 806162 (5 ea.) —

Protective Rubber Boots for Passport, PID II Downloading Software Kit


attery Type Black Boot Orange Boot Passport,Watchman,
tandard NiCd 804955 802806 Description FiveStar, Orion PID II,VOC 2000 Solaris
eavy-Duty NiCd 806751 806750 Downloading Kit, complete 710946 804679 710946
lkaline 806749 806748 Datalogging software only — — 710988

Protective Instrument Jackets


Diffusion Only FiveStar Solaris Orion Passport PID II MicroGard
Orange Nylon 711242 — 10020486 812701 — 711255
lack Vinyl — — 10020485 806715 — 478533
lack Leather 710884 — — — — —
Add-On Pump
Orange Nylon 711120 — 10020486 812701 — —
lack Vinyl — — — 806715 — 805283
lack Leather 711119 — 10020485 — — —
ntegral Pump
ellow Nylon — — — 812833 812833 —
lack Vinyl — — — 711022 711022 —
ordura Jacket — 10049053 — — — —

Remote Alarms
he Remote Alarm gives added audible alarm capability to many MSA portable gas monitors (listed below).
n alarm condition triggers both the Remote Alarm and the gas monitor’s alarm. In addition, two red LEDs on the
emote Alarm will light.The alarm is for workers wearing hearing protection or to alert standby personnel outside a
onfined space.Two versions are available: 90-decibel where hearing protectors are not worn, and
05-decibel if hearing protectors are worn (dB levels measured at one foot). Operates on one 9V alkaline battery.
emote Alarms forPassport
, PID I,IMicrogard, and W atchman Instruments
0dB Horn 800992
05dB Horn 800991
5-ft extension cable (total length must not exceed 300 ft) 800365
dapter cable (required) for Passport, Passport PID II, and MicroGard 812706
PULSAR™+ Single-Gas Detector
The PULSAR+ Single-Gas Detector is reliable, easy-to-use, and
durable.With sensor options for CO, H2S and O2 , this unit will provide industry-leading
performance and stand up to the roughest handling in the toughest environments.

Industry-Leading Features
Easy-to-Read Display Replaceable Sensor
The PULSAR+ Single-Gas detector offers MSA-built, patented Button™ Sensors
the simplest, easiest-to-read display. Its are easily replaceable.The stainless steel,
large numeric display—one of the largest solid electrolyte construction sensors
on the market—features a backlight that contain no liquid, eliminating the
provides excellent contrast.The easy-to- possibility of sensor leakage.The
understand display icons help the user to extremely small and thin sensor profile
easily identify STEL,TWA and peak gas allows the PULSAR+ Single-Gas Detector
concentration levels at a glance. to deliver high performance in a very
compact package. Specifications
One-Button Autocalibration Gas Type CO H2S O2
Durability Range 0-500 ppm 0-200 ppm 0-25% O2
Intuitive, completely automatic calibra- Resolution 1 ppm 1ppm 0.1% O2
tion with the press of a button helps to The PULSAR+ Single-Gas Detector is
Response Time T90<60 sec T90<60 sec T90<35 sec
prevent improper calibration. No sepa- designed to withstand even the harshest
rate, easy-to-lose accessories are needed conditions and usage.The interlocking
for gas-response testing or instrument gasket that provides water resistance also
calibration. Simply attach the tubing to absorbs some of the shock when Pulsar+ Single-Gas Detector
the sensor inlet and apply gas. dropped, its solid-state and rugged case Instrument Type Part Number
design increase impact resistance and CO 10036171
Distinctive Alarm System minimize the chance for breakage. CO with vibrating alarm 10036174
H2S 10036172
A triple alarm system leaves no doubt for W ater & Dust Resistance H2S with vibrating alarm 10036175
the user in an alarm situation.The audio
alarm is piercing and distinctive so that it With its IP54 rating, the PULSAR+ Single- O2 10036170
cannot be confused with other common Gas Detector is highly water- and dust- O2 with vibrating alarm 10036173
workplace sounds.The best-in-class visual resistant.The simple, yet effective design
alarm features four multi-directional, features an interlocking single-gasket
super-bright LED displays located on the suspension system, providing a tight seal
top of the instrument so it can be seen and making it one of the most water- and
from all angles.The alarm status is also dust-resistant instruments in its class.
shown in the numeric display. An optional
vibrating alarm emits a strong, pulsing User Configuration
vibration that further calls attention to
the user when in alarm mode.The vibra- The PULSAR+ Single-Gas Detector
tor is strong enough to be felt through features fully adjustable warning, peak,
heavy work clothes. STEL and TWA alarm levels. STEL and TWA
function can be deactivated. All alarm
Clip Options and calibration gas value settings are
password protected.
With options for suspender- and cell
phone-style clips and lanyards, MSA’s Large Rubber Button
unique clip designs keep the unit
attached during even the roughest use. Large rubber buttons make the PULSAR+
Single-Gas Detector easy to operate, even
with gloves.The button labels feature
English text and international icons.
Pulsar™ Single-Gas Detector
he MSA PULSAR Single-Gas Detector provides
maintenance-free, 24-month monitoring for the detection of CO, H2S, or O2.With a
nique design combined with a competitive price, it is the clear choice in single-gas
etectors.
PULSAR Single-Gas Detector Assemble-To-Order System
elect instrument options and mark them in the boxes to the right.

1 = CO without vibrating alarm


2 = CO with vibrating alarm
3 = H2S without vibrating alarm Keyfeaturesof the PULSAR SIngle-Gas Detector include:
4 = H2S with vibrating alarm
• The PULSAR requires no field calibration, battery or sensor
5 = O2 without vibrating alarm replacement.
6 = O2 with vibrating alarm
• Spring and suspension attachment clips are standard—keeping the
unit attached during even the roughest use.The unit can be worn
A = CO 25 ppm & 100 ppm
inside or outside a pocket, as well as on the belt or helmet.
B = CO 30 ppm & 60 ppm
C = CO 35 ppm & 100 ppm • A long-life battery offers 33% more capacity than the competition.
D = CO 35 ppm & 400 ppm • A numeric display with characters 4x larger than the competition
E = CO 50 ppm & 200 ppm allows users to count down service life from 24 months with
F = CO 100 ppm & 300 ppm accuracy and confidence.
G = CO 150 ppm & 200 ppm • It’s designed to survive a 6-ft drop and to meet an environmental
protection rating of IP54 to help ensure that the unit will be virtually
L = H2S 10 ppm & 15 ppm impervious to water and dust ingress.
M = H2S 10 ppm & 20 ppm • Because the PULSAR Single-Gas Detector has a built-in gas-delivery
N = H2S 10 ppm & 25 ppm adapter, no easy-to-misplace accessories are needed to conduct
O = H2S 15 ppm & 25 ppm gas-response tests.
W = O2 19.5 % & 18.0 % • Newly designed, MSA-built, stainless steel sensors with solid elec-
trolyte construction contain no liquid, eliminating the possibility of
X = O2 22.5 % & 19.5 %
sensor leakage.
• A triple alarm system, complete with audio, vibrating, and visual
Your Model Part Number
alarms leaves no doubt for the user in an alarm situation.
Please write in numbers and letters from selections
above in appropriate boxes and contact your Safety A-PULSAR- - For more complete information, see Bulletin 0802-18-MC.
Products Distibutor to place your order.

Titan™ Combustible Gas Detector


The Titan Combustible Gas Detector is a hand-held instrument used for the detection of combustible gases and vapors in
air.The unit is equipped with an 85 dB audible alarm and one of the biggest displays available.The intuitive one-button
operation allows simplicity of use.Simply switch on, and the Titan warns you of the presence of dangerous levels of
combustible gases or vapors. It will stand up to the roughest handling in even the toughest environments. And, it is part
of the MSA Stellar series of portable gas detection instruments.

TITAN Combustible Gas Detector


with alkaline batteries with Nimh batteries with vibrating battery pack
entane-calibrated 10033144 10033387 10033388
Methane-calibrated 10039414 10039415 10039416
ropane-calibrated 10042385 10042386 10042387

ccessories
attery charger and stand 10040783
arrying case with strap 10035638
Model RP cylinder, pentane 804532
Model RP cylinder, methane 491041

or more complete information see Bulletin 0802-20-MC.


Responder® Alarms
MiniCO® Responder Alarm
Compact and lightweight, the MiniCO Responder
Alarm provides fast, accurate sampling of CO
concentrations from 0 to 999 parts per million.
Ideal for firefighters, gas utility emergency crews, and workers in a steel mill,
the MiniCO Responder identifies hazardous situations immediately.When
lower levels of CO are detected, the MiniCO Responder Alarm gives a slow,
intermittent beep or “low alarm.”When higher levels of CO are detected, the
unit sounds a “high alarm,”emitting fast shrill beeps.

MiniH2S® Responder Alarm


In pulp and paper mills, water and waste treatment facilities, and oil refining
or exploration sites, the presence of hydrogen sulfide can be hazardous to MiniCO and MiniH2S Remote Responder Alarms
safe working conditions.The MiniH2S Responder Alarm helps workers in MiniCO Responder with on/off capabilities 710424
these facilities assess their environment.
MiniCO Responder without on/off capabilities (steel-mill style) 710510
Like the MiniCO Responder Alarm, audible and visible warnings or alarms MiniCO Responder with vibrating alarm and on/off capabilities 10008671
are activated when gas concentrations exceed two pre-set levels. A
MiniCO Responder with vibrating alarm without on/off capabilities 10008673
“Warning”(W) or low alarm is activated when a low concentration of
MiniH2S Responder with on/off capabilities 710850
hydrogen sulfide is detected. A distinctive pulsing alarm is activated when a
higher concentration is detected.This “High Alarm”(A) can be cleared only MiniH2S Responder with vibrating alarm and on/off capabilities 10008672
when the concentration drops below the alarm set point.The measuring
range is 0–500 ppm H2S. MiniOX, MiniOX Remote, Mini Cl2 and MiniClO2 Responder Alarms
MiniOx Responder with on/off capabilities 710965
MiniOX® & MiniOX® Remote Responder Alarms MiniOx Remote Responder with 10-ft cable, with on/off capabilities 10001923
MiniCl2 Responder 10020744
For detecting low levels of oxygen in confined spaces or in any location
MiniCl2 Responder with vibrating alarm 10020747
suspected of oxygen deficiency or enrichment, the MiniOX Responder
MiniClO2 Responder 10020745
Alarm and the MiniOX Remote Responder Alarm are the units for immediate
assessment. A peak (lowest) reading of the total percentage of oxygen MiniClO2 Responder with vibrating alarm 10020749
detected is displayed at the push of a button. An intermittent beep warns
when oxygen enrichment is detected. However, when the MiniOX Responder
Alarms detect insufficient oxygen levels, a high-rate alarm is activated and
MiniCO® Responder Kit
locks out any attempt to reset the unit until it is removed to an environment The MiniCO Responder Kit is designed
where the oxygen level exceeds the warning level.The MiniOX Responder specifically for firefighters or utility service
Alarm has an internal sensor; the MiniOX Remote Responder Alarm has personnel who must respond quickly to
an external sensor attached to a 10-foot cable.The measurement range CO calls.The kit contains a MiniCO
is 0–100% O2. Responder with a protective jacket and
neck strap, a Squirt® Gas Bump Tester or
MiniCl2 and MiniClO2 Responder Alarms RP size calibration gas cylinder, and 10
MiniCO Carbon Monoxide Detector Tube Cards, all packaged for convenience
The Mini Cl2 and Mini ClO2 Responder Alarms offer convenient low-level in a durable, compact carrying case.The Squirt Gas Bump Tester provides a
chlorine and chlorine dioxide monitoring, respectively. Both alarms are quick and certain means of calibrating the MiniCO Responder in the field.
perfect for use in many industries, including chemical, food & beverage, The Detector Tube Cards offer homeowners an easy and affordable means of
HazMat, petrochemical, pharmaceutical, pulp and paper, semiconductor checking their exposure to CO levels after the service person or firefighter has
and water/ wastewater treatment.The MiniCl2 Responder Alarm has a left the home. The Responders are ETL-approved for use in Class I Division 1,
0–15 ppm measuring range; the MiniClO2 Responder Alarm’s measuring Groups A, B, C and D locations.
range is 0–6 ppm.
Responder Alarm Kits
With only 3 buttons, both the MiniCl2 and Mini ClO2 Responder Alarms
MiniCO Responder Kit with bump test gas kit 710551
are very easy to use.Weighing less than 8 ounces and running on a
commercically available nine-volt battery, they each offer the user the ability MiniCO Responder Kit with calibration gas, regulator, and tubing 710550
to determine peak readings with the press of a button. For user safety, the
Alarms have a specially designed, time-delay power button that must be
pressed for 5 seconds to turn the power off.This feature helps prevent
accidental shutoff during use.
Both Alarms feature a lifetime warranty on most components.

Approvals
ETL (an OSHA-NRTL); Class I; Division 1; Groups A, B, C and D.
For more complete information, see Bulletin 0802-16-MC.
Detector Tube Pumps
Toximeter™ II Automatic Kwik-Draw®
Detector Tube Pump Detector Tube Pumps
The Toximeter II Automatic Detector Tube MSA’s Kwik-Draw and Kwik-Draw
Pump makes the sampling process easier, Deluxe Pumps can be used with an
allowing the user to preset the number of assortment of MSA’s detector tubes to
pump strokes (from 1 to 250 pump spot-test atmospheres for a wide vari-
trokes). Intrinsically safe, the automatic ety of toxic substances.They are
pump works with all MSA detector tubes. designed for one-hand operation and
t can also be used as a sampling pump. consistent delivery of a sample draw
volume of 100 milliliters (ml).
oximeter II Automatic
Detector Tube Pump 655585 For more complete information, see
Data Sheet 08-00-02.
or more complete information, see Data
Sheet 08-01-02. Kwik-Draw Basic Pump, with remote sampling adapter
and carrying pouch 488543
Kwik-Draw Deluxe Pump, with remote sampling adapter,
carrying pouch and end-of-stroke indicator 487500
Accessories
Gas-Tester™ II H
Sampling line, 10 ft 73067
Detector Tube Sampling line, 25 ft 73068
Pump Sampling line, 50 ft 73069
The Gas-Tester II H Pump is Remote Sampling Adapter, required for above sampling line 87970
et for action by compress- Solvent-Resistant Sampling Line, 25 ft, with reel 488780
ng the bellows. A pump Tube Holder, required for 488780 line above 488872
troke is started by press- Flue Gas Kit 470321
ng the release button.When the sample (100ml) is drawn through the tube,
he end-of-stroke indicator changes color. An accurate measurement is
obtained because the sample draw procedure itself is controlled only by the Detector Tube Handbook
pecifications of the pump and the flow resistance of the detector tube.
Detector Tube Handbook 813929
Gas-Tester II H Pump 696944

or more complete information, see Data Sheet 08-00-02.

Detector Tube Industry Action Sets


Agricultural Set Petroleum Set Semiconductor Sets
x Phosphine 2 x Hydrogen Sulfide Chemical Vapor Deposition
x Sulfur Dioxide 2 x Carbon Monoxide Process
x Methyl Bromide 2 x Qualitest 3 x Ammonia tubes
x Hydrogen Cyanide 4 x Benzene 3 x CO2 tubes
Pulp & Paper Set 2 x Hexane 3 x Nitrous fumes tubes
x Chlorine Dioxide Mining Set 3 x H2S tubes
x Sulfur Dioxide 3 x Carbon Monoxide Etching Process
x Hydrogen Sulfide 2 x Nitrogen Dioxide 4 x Ammonia tubes Agricultural Set 655864
x Chlorine 3 x Nitrous fumes 4 x Chlorine tubes
Pulp and Paper Set 655868
x Ozone 4 x Natural Gas/ 4 x Hydrogen Chloride
Methane Indoor Air Quality Set 710981
ndoor Air Quality Set tubes
Pharmaceutical Set 655869
x Carbon Monoxide Synthetics Manufacturing Epitaxy Process
x Carbon Dioxide Set Petroleum Set 655865
3 x CO tubes
x Formaldehyde 2 x Toluene Mining Set 655867
3 x Hydrogen Chloride tubes
x Ozone 4 x Qualitest 3 x Phosphine tubes Synthetics Set 655866
x H2O (relative humidity) 2 x Ethanol 3 x Phosgene tubes Chem.Vapor Depos. Set 655933
Pharmaceutical Set 2 x Vinyl Chloride Etching Process Set 655934
Crystal Growth Process
x Acetic Acid 2 x Trichloroethylene Epitaxy Process Set 655935
6 x Phosgene tubes
x Aromatic HC 6 x Phosphine tubes Crystal Growth Proc. Set 655931
x Qualitest
x Hydrogen Cyanide Note: All sets are boxes of 12 tubes
Chemical Warfare Agent Detector Tubes
Detect a wide range of nerve, blister, blood, and choking agents.
• Quick, simple to use; easy to interpret results
• “Go/No Go”style
• Can be used with MSA’s full line of pumps
• All sold in boxes of 10 tubes
• Important: Sold only in the U.S.

Nerve (Sarin,VX, GA, GD, etc.) 10007654


Blister (Mustard Gas - HD only) 10007653
Blister (Lewisite) 10007650
Blister (Mustard Gas - HD and HN) 10007652
Blood and Choking (CG, DP, AC, CK) 10007651

Chemical Warfare Agent Detector Tube Kits


Chemical Warfare Kits are available in a Single CWA Tube Sampler and a Multi CWA Tube
Sampler.The Single CWA Tube Sampler will only sample one CWA tube at a time.The Multi
CWA Tube Sampler allows the user to simultaneously sample up to four CWA tubes at the
same time without making adjustments.The kit is supplied with an MSA Escort ELF pump,
which maintains constant flow with a built-in “Run/Hold”feature that allows the user to
start/stop the pump and still maintain an accurate sampling timer.The four-port manifold
tube holder is pre-set for the proper flow rate for CWA tube sampling.

Multi CWA Tube Sampler 10049765


Single CWA Tube Sampler 10044998

Multi CWA Tube Sampler

Single CWA Tube Sampler


Detector Tubes
Detector Part No. Measuring range Threshold Limit Value
ubstance Measured Tube applicable (1 pkg. of 10 tubes) (ppm) 1998 ACGIH (ppm)
cetaldehyde Formaldehyde-0.1 497649 5–50 25 (ceiling)
cetic Acid Acetic Acid-1 804138 1–80 10
cetone Acetone-100 804141 100-10,000 500
Qualitest QL 497665 n/a
cetylene dichloride, cis and trans Trichloroethane-5 487343 10–500 200
,2 Dichloroethylene) Qualitest QL 497665 n/a
cetylene tetrabromide Trichloroethane-5 487343 5–200 1
,1,2,2-Tetrabromoethane) Qualitest QL 497665 n/a
cetylene tetrachloride Trichloroethane-5 487343 50–1000 1
,1,2,2-Tetrachloroethane) Qualitest QL 497665 n/a
mmonia NH3–2 804405 2–500 25
NH3–20 800300 20–1000
NH3–0.1% 804406 0.1–10 Vol.-%
-Amyl chloride (1-Chloropentane) Trichloroethane-5 487343 5–550 –
enzene Aromatic HC 804132 5–500 .5
C6H6-1 807024 1–100
C6H6-5 804411 5–100
Qualitest QL 497665 n/a
C6H6-0.5 655837* 0.5–10
romine C12-0.2 803944 0.2–3 0.1
romobenzene Aromatic HC 804132 30–720 –
romoethane (Ethyl bromide) Trichloroethane-5 487343 15–400 5
romoform (Tribromomethane) Trichloroethane-5 487343 7–200 0.5
romomethane (Methyl bromide) Trichloroethane-5 487343 20–270 1
3-Butadiene Ethylene-50 804428 100–1200 2
Qualitest QL 497665 n/a
-Butane Propane-200 804418 200–3800 800
Qualitest QL 497665 n/a
-Butanol (Butyl Alcohol) Ethanol-100 804136 100–3900 50 (ceiling)
ec. Butanol (sec-Butyl Alcohol) Ethanol-100 804136 300–5100 100
-Butene (1-Butylene) Ethylene-50 804428 100–5000 –
Qualitest QL 497665 n/a
-Butylene, cis and trans Ethylene-50 804428 200–5000 –
-Butylene) Qualitest QL 497665 n/a
utyl Alcohol (n-Butanol) Ethanol-100 804136 100–3900 50 (ceiling)
ec-Butyl Alcohol (sec-Butanol) Ethanol-100 804136 300–5100 100
-Butylamine Triethylamine-5 804134 2–28 5 (ceiling)
o-Butylamine Triethylamine-5 804134 3–36 –
ec-Butylamine Triethylamine-5 804134 2–18 –
Butylamine Triethylamine-5 804134 2–14 –
-Butylchloride (1-Chlorobutane) Trichloroethane-5 487343 5–170 –
Qualitest QL 497665 n/a
-Butylene (1-Butene) Ethylene-50 804428 100–5000 –
Qualitest QL 497665 n/a
-Butylene (2-Butene, cis and trans) Ethylene-50 804428 200–5000 –
Qualitest QL 497665 n/a
-Butyl mercaptan Ethylmercaptan-0.5 804589 1.5–15 0.5
Butyl mercaptan Ethylmercaptan-0.5 804589 0.8–5 –
arbon Dioxide CO2-100 497606 100–3000 5000
CO2-0.1% 487333 0.1–7.0 Vol.-%
CO2-1% 804419 1–20 Vol.-%
arbon Disulfide CS2-2 492514 2–300 10
Qualitest QL 497665 n/a
arbon Monoxide CO-0.001 % 804421 0.001–0.3 Vol.-% 25
CO-5 803943 5–1000
CO-10 487334 10–3000
CO-3000 815507 3000–70000
CO-0.1% 804423 0.1–1.0 Vol.-%
CO-0.3% 487335 0.3–7.0 Vol.-%
CO-10/color 47134 10–1000
(special orifice assembly 497652
for CO-10/color)
Qualitest QL 497665 n/a
hlorine Cl2-0.2 803944 0.2–30 0.5
ClO2-0.05 804133 1–46
Cl2-50 655862 50–500
Detector Part No. Measuring range Threshold Limit Value
Substance Measured Tube applicable (1 pkg. of 10 tubes) (ppm) 1998 ACGIH (ppm)
Chlorine dioxide ClO2-0.05 804133 0.05–15 0.1
Chlorobenzene Aromatic HC 804132 40–610 10
Chlorobromomethane Trichloroethane-5 487343 5–180 200
1-Chlorobutane (n-Butylchloride) Trichloroethane-5 487343 5–170 –
Qualitest QL 497665 n/a
Chloroethane(Ethyl chloride) Trichloroethane-5 487343 50–800 100
Chloroethylene (Vinyl chloride) VC-1 803950 1–70 5
Trichloroethane-5 487343 20–550
Chloroform (Trichloromethane) Trichloroethane-5 487343 8–100 10
1-Chloropentane (n-Amylchloride) Trichloroethane-5 487343 5–550 –
1-Chloropropane (l-Propylchloride) Trichloroethane-5 487343 5–220 –
2-Chloropropane (2-Propylchloride) Trichloroethane-5 487343 8–1700 –
Cycloheptane Hexane-20 497664 80–3300 –
Cyclohexane Hexane-20 497664 20–3400 300
Qualitest QL 497665 n/a
Cyclohexylamine Triethylamine-5 804134 7–38 10
Cyclooctane Hexane-20 497664 20–2100 –
Cyclopentane Hexane-20 497664 80–2700 600
n-Decane Hexane-20 497664 50–500 –
1,2-Dibromoethane (Ethylene dibromide) Trichloroethane-5 487343 25–700 –
Dibromomethane (Methylene dibromide) Trichloroethane-5 487343 9–200 –
1,1-Dichloroethane (Ethylidene chloride) Trichloroethane-5 487343 8–300 100
1,2-Dichloroethane (Ethylene dichloride) CH2Cl2-50 804416 30–720 10
1,1-Dichloroethylene Trichloroethane-5 487343 10–600 5
(Vinylidene chloride)
1,2-Dichloroethylene Trichloroethane-5 487343 10–500 200
(Acetylene dichloride, cis and trans)
Dichloromethane (Methylene chloride) CH2Cl2-50 804416 50–1000 50
1,2-Dichloropropane (Propylene dichloride) Trichloroethane-5 487343 5–440 75
1,3-Dichloropropane (Trimethylene dichloride) Trichloroethane-5 487343 5–220 –
Diesel Oil Qualitest QL 497665 n/a
Diethylamine Triethylamine-5 804134 3–27 5
Dimethylamine Triethylamine-5 804134 3–27 5
2,2-Dimethylbutane Hexane-20 497664 100–4900 –
Ethanol (Ethyl Alcohol) Ethanol-100 804136 100–6000 1000
Qualitest QL 497665 n/a
Ethene (Ethylene) Ethylene-50 804428 50–5000 –
Qualitest QL 497665 n/a
Ethyl Alcohol (Ethanol) Ethanol-100 804136 100–6000 1000
Qualitest QL 497665 n/a
Ethylamine Triethylamine-5 804134 4–55 5
Ethyl benzene Tol.-5 803947 5–1800 100
Ethyl bromide (Bromoethane) Trichloroethane-5 487343 15–400 5
Ethyl chloride (Chloroethane) Trichloroethane-5 487343 50–8000 100
Ethylenediamine Triethylamine-5 804134 5–27 10
Ethylene dibromide (1,2-Dibromoethane) Trichloroethane-5 487343 25–700 –
Ethylene dichloride (1,2-Dichloroethane) CH2Cl2-50 804416 30–720 10
Ethylidene chloride (1,1-Dichloroethane) Trichloroethane-5 487343 8–300 100
Ethyl mercaptan Ethylmercaptan-0.5 804589 0.5–80 0.5
Formaldehyde Formaldehyde-0.1 497649 0.1–55 0.3 (ceiling)
Formic Acid Qualitest QL 497665 n/a 5
Acetic Acid-1 804138 2–160
Furfuryl alcohol Phenol-1 813778 on request 10
Gasoline Gasoline-30 492870 30–6000 300
Qualitest QL 497665 n/a
n-Heptane Hexane-20 497664 20–2600 400
n-Hexane Hexane-20 497664 20–3200 50
Hydrogen Chloride HCl-1 803948 1–30 5 (ceiling)
Qualitest QL 497665 n/a
Hydrogen Cyanide HCN-2 803945 2–50 4.7 (ceiling)
Hydrogen Fluoride HF-1 804142 1–50 3 (ceiling)
Hydrogen Sulfide H2S-1 487339 1–200 10
H2S-100 487340 0100–4000
H2S-0.1% 655932 0.1–4 Vol. %
Qualitest QL 497665 n/a
Detector Part No. Measuring range Threshold Limit Value
ubstance Measured Tube applicable (1 pkg. of 10 tubes) (ppm) 1998 ACGIH (ppm)
sobutane (Methylpropane) Propane-200 804418 200–4200 –
sobutanol Ethanol-100 804136 100–2900 50
Isobutyl Alcohol, 2-Methylpropyl Alcohol)
sobutene Ethylene-50 804428 400–2600 –
Isobutylene, Methylpropene)
so-Butylamine Triethylamine-5 804134 3–36 –
sobutylene Ethylene-50 804428 400–2600 –
Isobutene, Methylpropene)
sobutyl Alcohol Ethanol-100 804136 150–2900 50
Isobutanol, 2-Methylpropyl Alcohol)
sobutyl Methyl Ketone MEK-50 813334 50–6500 –
so Octane Hexane-20 497664 100–3000 –
sopropanol Ethanol-100 804136 200–5000 400
Isopropyl Alcohol, 2-Propanol) Qualitest QL 497665 n/a
sopropyl Alcohol Ethanol-100 804136 200–5000 400
Isopropanol, 2-Propanol) Qualitest QL 497665 n/a
sopropylamine Triethylamine-5 804134 5–30 5
sopropyl mercaptan Ethylmercaptan-0.5 804589 0.5–5.5 –
Kerosene Qualitest QL 497665 n/a –

Ketones Qualitest QL 497665 n/a –


iquified Petroleum Gases Gasoline-30 492870 Semiquantitative –
Qualitest QL 497665 n/a
Mercury Hg–0.1 mg/m3 497663 0.1–0.8 mg/m3 0.025 mg/m3
(0.01–0.08 ppm) (inorganic)
Methane Natural Gas 655789 Semiquant. 5000+ –
Methanol (Methyl Alcoho1) Ethanol-100 804136 100–2350 200
Methyl Alcohol (Methanol) Ethanol-100 804136 100–2350 200
Methylamine Triethylamine-5 804134 4–55 5
Methyl benzene (Toluene) Tol.-5 803947 5–1000 50
Methyl bromide (Bromomethane) Trichloroethane-5 487343 9–200 5
MeBr-200 710544 200–8000 ppm
MeBr-2 710391 2–100 ppm
2-Methyl butane Hexane-20 497664 50–3000 –
Methyl chloroform Trichloroethane-5 487343 5–1500 350
1,1,1-Trichloroethane) Qualitest QL 497665 n/a
Methylcyclohexane Hexane-20 497664 80–4900 400
Methylcyclopentane Hexane-20 497664 150–3700 –
Methylene chloride (Dichloromethane) CH2Cl2-50 804416 50–1000 50
Methylene dibromide (Dibromomethane) Trichloroethane-5 487343 9–200 –
Methyl Ethyl Ketone (MEK) MEK-50 813334 50–4000 200
Qualitest QL 497665 n/a
Methyl mercaptan Ethylmercaptan-0.5 804589 0.5–5 0.5
2-Methyl pentane Hexane-20 497664 150–4500 –
3-Methyl pentane Hexane-20 497664 100–3700 –
Methylpropane (Isobutane) Propane-200 804418 200–4200 –
Methylpropene (Isobutylene, Isobutene) Ethylene-50 804428 400–2600 –
2-Methylpropyl Alcohol Ethanol-100 804136 150–2900 50
Isobutanol, Isobutyl Alcohol)
Nitrogen Dioxide NO2-0.5 487341 0.5–50 3
NO2-2 804435 2 –140
Nitrous Fumes Nitr.-0.5 487336 0.5–50 –
Nitr.-2 804425 2–140
Nitr.-10 803946 10–300
Nitr.-50 804426 50–3000
n-Nonane Hexane-20 497664 50–2800 200
n-Octane Hexane-20 497664 50–3000 300
Ozone Ozone-0.05 804140 0.05–5 0.05 (ceiling)
Pentachloroethane Trichloroethane-5 487343 10–300 –
n-Pentane Hexane-20 497664 50–3900 600
Qualitest QL 497665 n/a
Perchloroethylene Per-5 804429 5–200 25
Tetrachloroethylene) Per-10 487337 10–500
Qualitest QL 497665 n/a
Phenol Phenol-1 813778 1–25 5
Qualitest QL 497665 n/a
Phosgene Phosgene-0.1 803949 0.1–20 0.1
Detector Part No. Measuring range Threshold Limit Value
Substance Measured Tube applicable (1 pkg. of 10 tubes) (ppm) 1998 ACGIH (ppm)
Phosphine PH3-0.05 497101 0.05–3 0.3
PH3-0 1 485680 0.1–100
PH3-50 489119 50–2000
Propane Propane-200 804418 200–4000 2500
Qualitest QL 497665 n/a
n-Propanol (Propyl Alcohol) Ethanol-100 804136 100–3000 200
Qualitest QL 497665 n/a
2-Propanol Ethanol-100 804136 200–5000 400
(Isopropanol, Isopropyl Alcohol) Qualitest QL 497665 n/a
Propene (Propylene) Ethylene-50 804428 20–5000 –
Qualitest QL 497665 n/a
Propyl Alcohol (n-Propanol) Ethanol-100 804136 100-3000 200
Qualitest QL 497665 n/a
n-Propylamine Triethylamine-5 804134 2–28 –
1-Propylchloride (1-Chloropropane) Trichloroethane-5 487343 5–220 –
2-Propylchloride (2-Chloropropane) Trichloroethane-5 487343 8–1700 –
Propylene (Propene) Ethylene-50 804428 20–5000 –
Qualitest QL 497665 n/a
Propylene dichloride (1,2-Dichloropropane) Trichloroethane-5 487343 5–440 75
n-Propyl mercaptan Ethylmercaptan-0.5 804589 0.7–8.0 –
Styrene Styrene-10 804135 10–300 20
Qualitest QL 497665 n/a
Sulfur Dioxide SO2-1 487338 0.5–25 2
SO2-5 497662 5–120
SO2-100 497661 100–4000
Sulfur Hexaflouride SF6 Decomposition 804433 0.5–15 1000
decomposition products Products
1,1,2,2-Tetrabromoethane (Acetylene tetrabromide) Trichloroethane-5 487343 5–200 1
1,1,2,2-Tetrachloroethane Trichloroethane-5 487343 50–1000 1
Tetrachloroethylene Per-5 804429 5–200 25
(Perchloroethylene) Per-10 487337 10–500
Qualitest QL 497665 n/a
Tetrahydrofuran Ethanol-100 804136 Semi-quant 200
Toluene (Methyl benzene) Tol.-5 803947 5–1000 50
Qualitest QL 497665 n/a
Tribromomethane (Bromoform) Trichloroethane-5 487343 7–200 0.5
1,1,1-Trichloroethane Trichloroethane-5 487343 5–1500 350
(Methyl chloroform) Qualitest QL 497665 n/a
1,1,2-Trichloroethane (Vinyltrichloride) Trichloroethane-5 487343 10–170 10
Trichloroethene (Trichloroethylene) Tri-5 487342 5–250 50
Trichloroethylene (Trichloroethene) Tri-5 487342 5–250 50
Trichloromethane (Chloroform) Trichloroethane-5 487343 8–100 10
1,2,3-Trichloropropane Trichloroethane-5 487343 10–1200 10
Triethylamine Triethylamine-5 804134 5–30 1
Trimethylamine Triethylamine-5 804134 5–30 5
Trimethylene dichloride Trichloroethane-5 487343 5–220 –
(1,3-Dichloropropane)
2,2,4-Trimethylpentane Hexane-20 497664 100–3000 –
Vinyl Chloride (Chloroethylene) VC-1 803950 1–70 5
Qualitest QL 497665 n/a
Vinylidene chloride Trichloroethane-5 487343 10–600 5
(1,1-Dichloroethylene)
Vinyltrichloride (1,1,2-Trichloroethane) Trichloroethane-5 487343 10–170 10
Water Vapor H2O-10 655863 10–100% RH –
0-Xylene (1,2-Xylene) Tol.-5 803947 5–2500 100
Qualitest QL 497665 n/a
m-Xylene (1,3-Xylene) Tol.-5 803947 5–2500 100
Qualitest QL 497665 n/a
p-Xylene (1,4-Xylene) Tol.-5 803947 5–1200 100
Qualitest QL 497665 n/a
Escort ELF® and Escort® LC Sampling Pumps
Escort ELF Sampling Pumps Escort LC Sampling Pumps
he patented Escort ELF Sampling Pump can be used for The Escort LC Sampling Pump can be used with a
ersonal and area sampling.The state-of-the-art electronic variety of personal and area sampling devices to
aminar flow sensor, consisting of a laminar flow element and collect such airborne contaminants as asbestos
ressure sensor, provides constant flow (volume) control, with fibers, toxic gases, vapors, particulates, fumes, and
2.5% regulation of flow rate (from 1 to 3 lpm) and automatic mists. It can also be used to sample silica dust, coal
ompensation for changes in battery voltage, temperature, dust, and organic vapors.
ltitude, and sample load. The Escort pump is exceptionally compact, light-
n internal secondary standard calibrates the pump continu- weight, and quiet in operation. Engineered for use
usly and needs to be checked against a primary standard only in “hostile”environments, the unit can be sprayed
nce a month (or every 200 hours for coal mine dust sampling). with water while it is operating without being
damaged.
pprovals
scort ELF Sampling Pumps are UL approved as intrinsically safe Approvals Escort ELF® Sampling Pump
or use in hazardous locations—Class 1, Groups A, B, C, D; Class II, Escort LC Sampling Pumps are UL approved as
roups E, F, and G; and Class III, Division I locations. NIOSH-certi- intrinsically safe for use in hazardous locations—
ed for coal mine dust sampling (TC-74-030). MSHA certified as Class I, Groups A, B, C, D; Class II, Groups E, F, and G;
ntrinsically safe for underground use (Approval No. 2G-3924-1). and Class III, Division I locations.

EscortELF Kits EscortLCKit


scort ELF Pump with flow fault indicator, Escort LC Pump with 110V
10V single-unit charger, Gemini® Twin Port single-unit charger, sampling line
ampler, and standard packaging 805559 and standard packaging 711400
scort ELF Pump with flow fault indicator,
10V single-unit charger, sampling line,
nd standard packaging 805560
eplacement Parts Escort® LC Sampling Pump
attery Pack with O-ring 497702
One Replacement Inlet Water Stop Filter 802897
scort Pump Overhaul Kit—common components for
outine maintenance 802922
nlet Dust Filter (pkg of 5) 808935

Assemble-to-Order (ATO) System: Replacement Parts and Accessories


You Make the Choices
he ATO System makes it easy to “custom-order”Escort Pumps, configured
xactly the way you want them.
ou can choose from an extensive line of base instrument components and
ccessories. See the ATO Chart on the following page.
o obtain a copy of the ATO via FAX, call MSA QuickLit Information Service at
800-672-9010. At the prompt, request QuickLit Document #2346 (ATO for Battery Chargers MSA Omega¤ Chargers
scort and Escort ELF and Escort LC Sampling Pumps). 120 VAC 50/60 Hz 494716
220 VAC 50/60 Hz 495965
120/240 VAC 50/60 Hz (five-unit) 801759
MSA Sampling Pump Accessories
Sampling Equipment
MSA sampling pump accessories and air sampling equiment allow
monitoring of many different contaminants in various applications. Sampling line only 456226
ampling pump accessories can be used in personal and area sampling for a 10mm Cyclone Assembly to separate respirable dust
wide variety of airborne contaminants such as asbestos fibers, toxic gases, from non-respirable dust 456243
apors, particulates, mists, and fumes. Gemini Twin-Port Sampler for sorbent tubes 497697
Accessories for MSA sampling pumps include filter media, a cyclone assem-
bly, preweighed filter cassettes, filter holder assemblies, impingers, the Carrying Accessories Jacket
Gemini® Twin-Port Sampler, sorbent tubes, and calibrators. Leather jacket 811741
or more information, see Data Sheet 08-09-01.
Escort ELF® and Escort® LC Sampling Pump Assemble-to-Order (ATO) Options
A. Number of Pumps Number of Pumps Code
It’s your choice: You can order one pump at a time, or the convenient “5-Pump Kit,” which includes a durable PVC carry- One 1
ing case and bonus training video. Five 5
NOTE:When you order the 5-Pump Kit through this ATO Ordering Selection Matrix, any accessories (such as single-unit
chargers, Gemini Twin-Port Samplers, cyclones or instrument jackets) you order will also be provided in five-packs.
B. Pump Type Pump Type Code
No matter whether you choose the Single-Pump Option or the 5-Pump Kit Option in Section A, you have a choice of Escort ELF Pump with 90-second ELF1
three different types of pumps: Flow-Fault Indicator
•The Escort ELF Pump with 90-second Flow-Fault Indicator. In the event that a flow blockage occurs, the Flow-Fault Escort ELF Pump without ELF2
Indicator circuitry illuminates an LED to alert the user.Within 90 seconds, the pump automatically shuts off, but the 90-second Flow-Fault Indicator
cumulative elapsed time remains on the display. To select this option, write “ELF1” in the box at right. Escort LC Pump ELFLC
•The Escort ELF Pump without Flow-Fault Indicator. To select this option, write “ELF2” in the box at right.
•The Escort LC Pump. To select this option, mark “ESLC” at right.

C. Battery Chargers Battery Chargers Code


Which type of battery charger do you want? Select from the following choices and mark your selection in the box at No battery charger 0
right (all operate from 50/60 Hz):“0” for no battery charger,“1” for 120-Volt charger,“2” for 220-Volt charger, or “5” for 120-Volt charger 50/60 Hz 1
the 5-Unit charger. 220-Volt charger 50/60 Hz 2
5-Unit charger 120/240 V, 50/60 Hz 5

D. Sampling Accessories Sampling Accessories Code


Sampling Accessories include the following: Sample Line only 1
• Sample Line 10 mm Cyclone Assembly* (P/N 456243) 2
• 10 mm Cyclone Assembly, which separates respirable dust particles from non-respirable particles Gemini Twin-Port Sampler 3
• The Gemini® Twin-Port Sampler, which streamlines sampling with sorbent tubes by enabling collection of two or No Sampling Line 4
more samples simultaneously from a single sampling pump. 10 mm Cyclone Assembly IH* (P/N 10044015) 2

E. Protective Jackets Protective Leather Jacket Code


A protective leather jacket is offered as an option to protect the instrument from everyday wear, tear and scuff marks. None 0
To select the jacket, mark “1” in the box at right. Protective Leather Jacket for 1
Sampling Pump

Your Model Part Number Selection A B C D E * Cyclone Assembly 456243


is for use with MSA
-
Please write in numbers from selections above in appropriate boxes
and contact Safeware to place your order. A-PMP- - - - pre-weighed cassettes (MSHA);
Cyclone Assembly 10044015
is for use with non-MSA
cassettes.

42 CFR Oil-Mist Sampling Kits General-Purpose Cassette


The NIOSH respirator certification standard, The general-purpose cassette is a 5-µ pore size PVC
42 CFR Part 84, has introduced new ways to select sample cassette that can be used for silica sampling.
non-powered air-purifying particulate filter This cassette can also be used for all airborne partic-
respirators. One of the primary concerns of the ulate collection not limited to quartz or silica.
standard is oil-mist exposure. According to this
This cassette is individually packaged in a sealed
NIOSH standard, there are 3 levels of respiratory
plastic bag. Each cassette is weighed by a robot to
protection: N = Not resistant to oil; R = Resistant
.01 mg at 70°F +/- 5° and 45 RH +/- 10%. Each
to oil; P = Oil-Proof. But how do you know whether Escort ELF® Kit
cassette is attached to a Dust Data Card which
oil mist is present in the workplace? MSA’s 42 CFR
includes the cassette serial number and actual
Sampling Pump Kits can help you answer this question by providing
weight.The cassette inlet contains a tamper-proof
convenience and flexibility in oil-mist sampling.
deflector plate, and the outlet contains an anti-
Kit includes: blow-back check valve. Both the inlet and outlet
• Sampling Pump (Escort ELF or LC) • 110V charger openings are capped (sealed) to ensure cleanness.
• 3-piece 37-mm MCE filter cassette, • Sampling line The perimeter of the cassette housing is sealed with
0.8 micron pore size, 10 filters • Garment clip tamper-proof sealing tape.
per kit • Instruction manual Because of the extreme accuracy of the MSA weighing process, users of
this cassette can sample the atmosphere for dust particulate and have the
analytical analysis performed with complete confidence in its accuracy.
Escort ELF 42 CFR Sampling Pump Kit 711585
Escort LC 42 CFR Sampling Pump Kit 711586 General-Purpose Cassette 711361
Sampling Pump Accessories
Sorbent Tube Sampling Cassette Sampling Miscellaneous Accessories

Couplers and Supplementary Parts Kit


Filter Media

Coal Dust Cassettes -


P/N 803462

Gemini¤ Twin-PortSampler General Purpose Filter


Cassettes
or low-flow control when sorbent
ubes are used, the patented Gemini
win-Port Sampler is a valved Asbestos Sampling Filter Cassettes
mechanism that allows flow adjust-
ment down to 1 mlpm (0.001 lpm to GeneralPurpose FilterCassettes
00 mlpm total between both Fi
ftycomplete 3-piecepreloaded filter cassettes with MCE Impinger assemblies and accessories
ubes). U.S. Patent No. 5,370,004. filters.
s an added benefit, the Gemini With 25mm, 0.8-µ pore size 695677
ccessory permits simultaneous With 37mm, 0.8-µ pore size 695676 Silica Dust Cassette
ampling from sorbent tubes, with P/N 711361
ndependently controlled flow rates FilterDiscsfor Nuisance Dust Sampling
f each. Dual sampling means two MSA’s complete line allows users to load sampling
Used with 37mm cassettes.
ke sorbent tubes can be attached cassette cases and select your own combinations of
PVC, 0.5-µ pore size, 50/pack 459733
or simultaneous sampling at differ- sampling media.Your MSA distributor can help you
nt flow rates, or two different tubes PVC, 0.8-µ pore size, 50/pack 812805 select the accessories that best suit your application
an be used to sample two types of PVC, 5.0-µ pore size, 50/pack 625413 requirements.
ubstances at once. Glass Fiber 463784
Filter Cassette Cases, 2-piece, 25mm, pack of 50 695681
emini Twin-Port Sampling Kit— M CEFilterDiscsfor Asbestos and Nuisance Dust Filter Cassette Cases, 2-piece, 37mm, pack of 12 625412
ncludes Gemini Sampler,
25mm, 0.8-µ pore size, 100/pack 695674 Filter Cassette Cases, 3-piece, 37mm, pack of 10 449347
ube protectors,Y-connector,
ips and carrying case 497697 37mm, 0.8-µ pore size, 50/pack 463797 50mm-long Cowl for 25mm Filter Cassette Case,
37mm, 0.45-µ pore size, 50/pack 463796 pack of 5 695683
harcoal Sampling Tubes
150mg), 50 tubes 697169 25mm Support Pad (felt backup disc), pack of 100 695684
harcoal Sampling Tubes Silver Membrane FiltersforSilica, Coke Oven 37mm Support Pad (felt backup disc), pack of 25 449375
600mg), 50 tubes 697170 Emissions and Carbon Black 25mm Cellulose Bands, jar of 60 484683
lica Tubes (225 mg), 50 tubes 697171 Silver Membrane, 0.8-µ pore size, 50/pack 464324 37mm Cellulose Bands, jar of 60 625415
lica Tubes (600 mg), 50 tubes 697172 37mm Stainless Steel Coupler for 3-piece filter
mberlite XAD-2, 50 tubes 697175 cassette case used with Cyclone Assembly 457392
Asbestos Sampling FilterCassette
opcalite, 50 tubes 697176 Plastic Coupler for preweighed filter
All units include 50complete filter cassettes preloaded with
cassette sampling used with Cyclone Assembly 457391
arbotrap, 25 tubes 697174 25mm MCEfilters.
With 0.8-µ pore size and 50mm anti-static cowl 695679 Sampling Line Assembly (used with all pumps
arbosieve, 25 tubes 697173
and filter cassette cases) 456226
enax/CMS 491165 With 0.45-µ pore size and 5.0-µ pore size and
50mm anti-static cowl 696172 25mm Filter Cassette Case Sampling
Line Coupler, package of 10 695685
Gas Sampling Bags Coal Dust and Silica FilterCassette 37mm Filter Cassette Case Sampling
Line Coupler, package of 3 459743
Preweighed filter cassette with 5-µ pore size
PVC filter and Mine Data Card - Coal Dust 803462 Charcoal Filter Tubes—for use in in-line
sampling to protect pumps from vapor damage 804403
Preweighed filter cassette with 5-µ pore size
PVC filter and Data Card - Silica 711361 Supplementary Parts Kit—includes
3 stainless steel support screens, small brush,
tweezers, and press/pry tool 456246
Respirable Dust Sampling w/Cyclone Assembly
All-Glass Impinger Assembly—includes
10mm Cyclone Assembly, fritted-glass flask, inlet cap, nozzle, and cap 10008396
used with MSA pre-weighed cassettes 456243
Bubbler for Fritted-glass Flask 10008397
10mm Cyclone Assembly, IH version
eflon Sample Bag Assembly 471677 Flask Holster 10008398
used with 37mm cassettes 456243
edlar Sample Bag Assembly 472992 Tubing for Flask 93495
Sampling Pump Calibration Check Devices
A primary calibration device, the DigiCal™ PrimaryCalibration Devices
Calibrator provides instantaneous calibration for DigiCal Calibrator 655101
instruments like the MSA Escort LC or the sec-
Air Inlet Caps, pkg of 2 655102
ondary flow standard inside the Escort ELF
110V Charger with A/C Adapter 655112
Sampling Pumps. Just press the plunger and the
DigiCal Calibrator does the work. Its unique flow Bubble Solution, 4-oz bottle 655273
cell replaces conventional bubble tubes and Sub C Battery Pack 655169
makes calibration easier.
The DigiCal Calibrator achieves extreme accuracy SecondaryCalibration Devices
by utilizing a computerized flow meter that pro- Flowmeter for use with Flow-Lite and
vides instantaneous flow readouts on a digital dis- Escort Sampling Pumps, 0.2 to 4 lpm 490197
play. Accurate measurements are possible within Flowmeter for use with Flow-Lite and The DigiCal Calibrator makes sampling pump
calibration a stress-free, one-step procedure.
±0.5 percent at any altitude. Escort Sampling Pump with Gemini
Twin-Port Sampler, 30 to 370 mlpm 490198
For more complete information, see Bulletin
0810-34-MC.

Ventilation Smoke Tube Kits Airborne Compounds Sampling Chart


MSA’s Ventilation Smoke Tube This new MSA/SupelCo guide to sor-
Kits are for use where controlled bent tube sampling lists MSA part
generation of a visible smoke is numbers. It is a very comprehensive
desired in order to determine guide to all types of air sampling using
the velocity of slow-moving air personal sampling pumps.
currents and establish their
direction and flow patterns in MSA/SupelCo Sampling Chart 814473
shafts, mines and tunnels.They
can also be used in commercial
buildings and industrial process-
ing plants to determine velocity Ventilation Smoke Kit, including
and flow patterns of heating, aspirator bulb, two rubber plugs,
ventilating, and air-conditioning and six smoke-producing tubes
systems. contained in plastic carrying case 458481
Ventilation Smoke Kit, including
For more complete information, aspirator bulb, six tube caps, and
see Data Sheet 08-00-20. two glass smoke-producing tubes
contained in plastic carrying case 5607
Ventilation smoke tubes,
pkg of 12 458480
Glass smoke tubes, pkg of 10 5645
TIM® System – The Total Instrument Manager
alibrateFour Instruments Simultaneously were bump-tested or calibrated on the TIM
System.This easy-to-use and easy-to-read report
he TIM System software is designed to calibrate menu is designed with the user in mind.Whether
p to four instruments at the same time— with you need a quick reference to any particular in-
nly one computer and four TIM Control Centers. strument and its calibration results,or a complete
Whether one, two, three, or four instruments are detailed master report of all instruments,catego-
eing calibrated, the TIM System ensures that each rized and printed in hard-copy form, the TIM
nstrument is fully calibrated and ready to go! System offers you this capability at your fingertips.
mmediateCalibration Results Ergonomic Design; Loaded with Features
fter an instrument has been bump-tested or cali- Designed for use where space is limited, the com-
rated, the TIM System immediately displays and pletely portable TCC measures only 11" x 9" x 6" calibrated: a Passport, Gasport,Watchman,
rints the results of the bump test or calibration. and weighs less than 7 pounds.The housing is FiveStar Alarm, or Orion instrument.The TIM
he user can set up the system for a bump test or a .090" aluminum, painted with rugged wear-resis- System automatically identifies the instrument
ull calibration.The display on the TCC (TIM Control tant epoxy paint.The unit features a sloped face serial number and model type, as well as the actu-
enter) unit not only shows whether an instrument panel, four non-skid mounting feet, and an easy- al gas sensors installed in the instrument.The TIM
assed or failed the bump test or calibration,but to-read lighted display which facilitates operation System is so advanced that if an instrument is not
so displays the results for each sensor. in low light areas. In addition, a membrane-acti- in proper working condition, it will let you know
vated, alpha/numeric, panel-mounted keyboard what is wrong. Beyond simply calibrating a given
assword Availability
provides easy touch-button data entry. instrument, here is what the TIM System’s
he TIM System allows password identification, Setup is a snap! In less than 15 minutes, your TIM “smarts”will do for you:
desired.The password identification maintains System is completely operational. All necessary • Automatically generate test protocol based on
entification of individual users, maintains identi- connectors and power connectors are supplied. instrument sensor configuration.
cation of instrument serial numbers, and records And don’t worry…we have designed the software • Adjust run time for instrument and pump type
he user date, time and type of calibration (bump so that it’s easy to install. to ensure shortest test time.
est or full calibration) performed. Passwords are • Check instrument software and sensor compati-
maintained and assigned by the system manager. Auto-Recognition Capability bility.
• Notify user if gas for instrument is not assigned
toresAllCalibration Records This unique feature can be found only in the TIM to TCC.
System. Our unique auto-recognition feature
he TIM System electronically stores all bump test • Synchronize date and time of instrument with
eliminates the need for changing setups between personal computer.
nd calibration records.With a click of the mouse, a tests.The TIM System will automatically recog-
rop-down screen displays all recorded calibrations • Print computer-generated test receipt.
nize the MSA instrument that is intended to be
f any particular instrument. Each record is stored • Much more.
ndividually by date, time, and results of each
nstrument calibration. Each calibration record
ncludes the test results of each instrument test or TIM System Kits
alibration. Depending on your need, you can choose from any one of 11 already-assembled TIM System kits
nstrumentRecords: designed to calibrate your specific instrument.
TIM System Kit Part No.
Instrument and pump type
10010742
10010151
10010152
10010153
10010154
10010155
10010156
10010157
10010158
10010150
10030771
10046769
10046768
10046767
Model number (sensors installed)
Owner and owner information
Instrument location TIM System Components
User comments TIM Control Center (P/N 710338) • • • • • • • • • • •
Test results Software CD-ROM (P/N 710460) • • • • • • • • • • •
Instrument assignment Fresh Air Inlet Filter (P/N 633697) • • • • • • • • • • •
Instruction Manual (P/N 710548) • • • • • • • • • • •
omputer Requirements
6-foot Length of Tubing (P/N 603806) • • • • • • • • • • •
o ensure that the TIM System can be easily en- Warranty Card (P/N 10006796) • • • • • • • • • • •
anced for years to come, it is designed to operate PC Computer, Pentium (P/N 10006402) •
n a standard IBM-compatible personal computer. Data Docking Module (P/N 804679) • •
Minimum requirements are a Pentium processor, Module Bracket,Watchman (P/N 10006711) • •
6 MB RAM, a 30 MB hard drive, and a CD-ROM
Gas Miser Model RP Regulator (P/N 710288) • • • • • •
rive.The software is compatible with all Windows
Gas Miser Model BD Regulator (P/N 710289) • • • •
perating Systems (all data can be downloaded to
he network). A computer is required to operate the Holster Assembly, Integral Pump (P/N 710658) • •
ystem at each local site, but a single computer can Holster Assembly,Add-On Pump Module (P/N 711391) • •
un up to four TIM Control Centers. Holster Assembly, FiveStar (P/N 10010164) • •
Holster Assembly, Orion (P/N 10028009) •
eportCapability
Deluxe Docking Stand,Orion/Orion G (P/N 10042686) •
he TIM System’s data software includes a Deluxe Docking Stand, Orion only (P/N 10043139) •
ser-friendly archive—with both view and print Deluxe Docking Stand, Orion G only (P/N 10043138) •
apability—of all calibrations of all instruments that
Assemble-To-Order for the TIM System Accessories
If one of MSA’s 14 pre-packaged TIM System kits does not exactly meet your requirements,
The TIM System requires one ofthe holster assemblies or the data
docking module described below tocommunicate with the instru-
you can custom order a complete system that will meet your needs.To configure your TIM ment.Other accessories are optional.
System, select instrument options and mark them in the appropriate boxes below.Then Holster assembly for use with Orion. Includes jet-eye
contact your MSA Safety Products Distributor to place your order. As with all MSA communication module and gas feed connection. 10028009
Assemble-To-Order requests, your order will usually be shipped the same day it is received. Holster assembly for use with Passport integral pump
1—Basic TIM System and Gasport-style instruments. Includes built-in
data docking module and gas feed connection. 710658
All TIM Systems come complete with this basic configuration (“A-TIM”).The only other
Holster assembly for use with Passport add-on pump.
components needed are a PC computer and a holster assembly or data docking module, Includes built-in data module and gas feed connection. 711391
which can be ordered as an accessory.The Basic TIM System includes:
Holster Assembly for use with FiveStar.
Description Part No. Description Part No. Includes jet eye infrared communication module
TIM Control Center (TCC) 710338 Instruction Manual 710548 and gas feed connection. 10010164
Software CD ROM 710460 Gas Miser Regulator, Model RP 710288 Holster Assembly for use with Orion and Orion G.
Fresh Air Inlet Filter 633697 6-foot Calibration Tubing 603806 Includes jeteye infrared communication module
and gas feed connection 10028009
2—Data Docking Module Holster Assembly for use with Solaris.
Includes jet eye infrared communication
Select only option. Enter number of selection in position 2.
and pumped gas delivery 10049419
0 = None Docking Stand, deluxe, heavy-duty,
1 = Add-on pump holster assembly (P/N 711391) 6=Data Docking Stand Orion (P/N 10028009) for use with Orion and Orion G.
2 = FiveStar holster assembly (P/N 10010164) 7=Data Docking Stand Orion (P/N 10043139) Includes jet eye infrared communication module
and gas feed connection. 10042686
3 = Data docking module (P/N 804679) 8=Data Docking Stand Orion G (P/N 10043138)
Docking Stand, deluxe, heavy-duty,
4 = Integral pump holster assembly (P/N 710658) 9=Data Docking Stand Orion (P/N 10049419)
for use with Orion only.
5 = Orion holster assembly (P/N 10028009) Includes jet eye infrared communication module
and gas feed connection. 10043139
3—Gas Regulator Type
Docking Stand, deluxe, heavy-duty,
This option allows your system to be ordered with multiple regulators. Enter letter of selec- for use with Orion G only.
tion in position 3. Includes jet eye infrared communication module
0 = None and gas feed connection. 100443138
A= Gas Miser, Model BD (P/N 710289) Data docking module for use with Gasport,
B = Gas Miser, Model RP (P/N 710288) Passport and Watchman. Includes MSA link
data logging software. 804679
C = Natural Gas Regulator (P/N 710545)
A/B Switchbox (6" x 6" x 2V") 10010168
D= Gas Miser BD and Gas Miser RP
Connecttwo holster assembly or data docking modules for use with
E = Gas Miser RP and Natural Gas Regulator one TCC-TIM Cont rolCenter.Shipped complete with inter-connecting
F = Gas Miser BD and Natural Gas Regulator cable and dual feed flow system. Switch allow s usage with either
connected holster/module assembly.Only one instrumentcan be
G= All: Gas Miser BD, Gas Miser RP, and Natural Gas Regulator calibrated ata time.
4—Personal Computer TCC Printer (6" x 6" x 9") 10010167
Shipped complete with 5 rollsofprinter paper and required cables.
This option allows users to order a personal computer with the TIM System.The computer Connectsdirectly to the TCC and will automatically printa 3"x 5"
will have the TIM System software installed and the icon initialized. Enter number of test receipt of instrumenttest.Idealforconfined space ent ry permits.
selection in position 4.
0 = None 1 = Pentium computer (software installed)
The computer supplied will be configured with the following minimum requirements.
However, because of rapid advancements in computer technology, MSA reserves the right Available for Chlorine and Ammonia use!
to ship a computer of a larger size and faster processor speed. Limited to FiveStar Instruments.
Contact Customer Service for detailed information.
Minimum requirements:
• IBM compatible 333 • 32x CD-ROM drive • mouse
MMX processor • Windows 98 • keyboard
• 6.4 GB hard drive • 17" monitor
• 32 MB RAM • 3V" 1.44MB floppy drive

A - TIM - -
1 2 3 4
Calibration
Regulators
25/1.5LPM Flow Cont
rol Natural Gas Demand Regulator
Model RP Cylinder .25 lpm* 467895
This low-pressure Demand Regulator is designed to
Model RP Cylinder 1.5 lpm 467896
be used with MSA Instrument Calibration Systems
to supply 100% natural gas.The Demand Regulator
Can be used with chlorine and ammonia; replaces
will automatically supply 100% natural gas to the
N 809945.
calibrator as required. No adjustments are necessary or provided. Like the Gas
Miser Regulator, the Natural Gas Regulator features an automatic ON/OFF
valve that releases gas only on demand. Comes complete with (barb type)
Model RP outlet fitting and easily connects to natural gas supply with W" NPT thread.
Recommended for use with pressure of 0.3 to 5.0 psig.
Gas Miser¤ Demand Regulator: Natural Gas Demand Regulator 710545
he IntelligentRegulator
he Gas Miser Regulator is the most advanced, yet simple- Model RPCombination Regulators
o-use calibration gas delivery system available. Made of
ickel-plated brass and polished aluminum (especially Two regulators in one.Trigger activation allows a
mportant for reactive gases), the Gas Miser Regulator can bump test or calibration check in the “SQUEEZE”
upply gas flow from 0.1 lpm to 3.0 lpm. But that’s only the mode or fixed-flow calibration in the up or
eginning. Designed for use with any MSA Model RP Gas Miser ® “LOCKED”position. Designed for use with all
alibration Cylinder except chlorine and ammonia*, the Model RP Model RP and Econo-Cal calibration cylinders.
as Miser Regulator features an automatic ON/OFF valve Supplied with calibration tubing and special
hat releases gas only on demand. So only the amount of fitting.
as needed to calibrate your MSA instrument is delivered
Model RP Combination Regulator, .25 lpm 711175
ccurately.When the calibration is complete, the Gas Miser
egulator shuts off automatically. And, for user conve- Model RP Combination Regulator, 1.5 lpm 711174
ience, the regulator can stay connected to the cylinder
here it remains in a ready state—making it ideal for such Gas Miser¤ Manifold
xed installations as work benches and calibration stations.
The Gas Miser Manifold is not only a cylinder
he Gas Miser Regulator also eliminates the need to Gas Miser® Model holder but it also incorporates a 4-station “bump
hange regulators to accommodate different instruments BD-20
test”or calibration manifold. It works with Model
r flow rates. It’s not only intelligent, it’s cost-efficient. RP and Econo-Cal cylinders and Gas Miser regula-
upplied with calibration tubing and special fitting. tor on any pumped or aspirated instrument.The
A Gas Miser Regulator for chlorine and ammonia is available for Model RP cylinders — manifold will supply the correct amount of gas
N 10034391 required and will bump test or calibrate up to four instruments
Model RP 710288 simultaneously. Gas Miser regulator sold separately.
Model RP, Chlorine and Ammonia 10034391 4-station Manifold 710274
Model BD-20, with CGA 590 fitting 710289

Model RP Calibration Check Kits


Model RP Check Kits consist of a regulating valve which includes a gauge
o measure container pressure, an adapter hose, sensor adapter (where
pplicable), instructions, and a case fitted with room for two Model RP
ylinders of calibration check gas.

heck Kit, Model RP, with 1.5 lpm regulator—complete 477150


heck Kit, Model RP, with 0.25 lpm regulator—complete 477149

Calibration Gas
ach MSA Calibration Gas Cylinder is shipped with an MSA certifies that the gas mixture in calibration gas cylinders was prepared gravimetrically,
ndividual copy of a material safety data sheet (MSDS) using NIST traceable weights.The lot number and nominal value of the gas constituents in
nd an individual copy of a certificate of analysis. percent by volume, percent by mass, PPM, or volume are specified on the cylinder.The
uncertainty statement of the specified nominal value is also listed.
Model RP and Model R Calibration Cylinders Non-Reactive Gases,Steel Cylinder,100 Liters
100 LTRS 19 LTRS 34 LTRS
1000 PSI 300 PSI 500 PSI
Gas Fill Gas Mixture Background Model RP Model R Econo-Cal
Air air zero (THC < 1 ppm) — 801050 — —
Carbon 400 ppm carbon monoxide air 806255 — —
Monoxide 300 ppm carbon monoxide air 473180 461769 —
200 ppm carbon monoxide air 809243 — —
100 ppm carbon monoxide air 809242 — —
50 ppm carbon monoxide air 809241 — —
100 ppm carbon monoxide nitrogen 806734 — —
Hydrogen 0.8% hydrogen air 803102 — —
Isobutylene 100 ppm isobutylene air 494450 — 10048279
Model RP
Methane 2.5% methane air 491041 459942 —
6.6% methane nitrogen 801049 — — Model RP Cylinder
Oxygen 20.8% oxygen nitrogen 479857 468248 — Contents:Pressure 1000
5.0% oxygen nitrogen 493580 476302 — psig; approximately 100
Nitrogen 100% nitrogen — 481317 — — liters at atmospheric
Nitrous Oxide 10 ppm nitrous oxide nitrogen 806736 — — pressure
Size: 13-3 4" x 3"
Pentane 0.75% pentane (50% LEL) air 804532 — —
Weight: 2 lb 13 oz
Propane 0.6% propane air 493579 — —
Material: Steel
Combination 0.35% pentane, 19.0% oxygen, 100 ppm carbon monoxide nitrogen 10007047 — —
0.6% propane, 15% oxygen, 60 ppm carbon monoxide nitrogen 801051 — —
1.45% methane, 300 ppm carbon monoxide, 15% oxygen nitrogen 10010162 — —
2.5% methane, 60 ppm carbon monoxide, 15% oxygen nitrogen 813718 — —
2.5% methane, 300 ppm carbon monoxide, 15% oxygen nitrogen 10040291 — —
1.45% methane, 15% oxygen nitrogen 478192 — —
1.45% methane, 15% oxygen, 60 ppm carbon monoxide nitrogen 478191 — —

Model RP and Model Econo-Cal¤ Calibration Cylinders Reactive Gases,Aluminum Cylinder,58 & 34 Liters
58 LTRS, 500 PSI
Gas Fill Gas Mixture Background Model RP Econo-Cal
Ammonia 25 ppm ammonia nitrogen 814866 711078
Chlorine 10 ppm chlorine nitrogen 806740 711066
Hydrogen Sulfide 40 ppm hydrogen sulfide nitrogen 467897 711062
15 ppm hydrogen sulfide nitrogen 806253 711064
10 ppm hydrogen sulfide nitrogen 467898 711060
Nitric Oxide 50 ppm nitric oxide air 812144 711074
Nitrogen Dioxide 10 ppm nitrogen dioxide air 808977 711068
Model RP Econo-Cal
Phosphine 0.5 ppm phosphine nitrogen 710533 711088
Hydrogen Chloride 40 ppm hydrogen chloride nitrogen 710210 711080 Model RP Cylinder
Hydrogen Cyanide 10 ppm hydrogen cyanide nitrogen 809351 711072 Contents: Pressure 500
psig; approximately 58
Sulfur Dioxide 10 ppm sulfur dioxide air 808978 711070
liters at atmospheric
Combination 1.45% methane, 15% oxygen, 300 ppm carbon monoxide, nitrogen 804770 711058
pressure
Gas Cylinders 10 ppm hydrogen sulfide
Size: 13X" x 3"
1.45% methane, 15% oxygen, 300 ppm carbon monoxide, nitrogen 10048788 10048790 Weight: 1 lb 11 oz
20 ppm hydrogen sulfide
Material: Aluminum
1.45% methane, 15% oxygen, 60 ppm carbon monoxide, nitrogen 10045035 10048280
20 ppm hydrogen sulfide
Econo-Cal Cylinder
Contents: Pressure 500
1.45% methane, 15% oxygen, 10 ppm hydrogen sulfide nitrogen 804769 711056
psig; approximately 34
1.45% methane, 15% oxygen, 300 ppm carbon monoxide nitrogen 813720 711076 liters at atmospheric
10 ppm hydrogen sulfide
pressure
1.45% methane, 15% oxygen, 60 ppm carbon monoxide nitrogen 10048890 10048981 Size: 13X" x 3"
20 ppm hydrogen sulfide Weight: 1 lb 1 oz
2.5% methane, 15% oxygen, 20 ppm hydrogen sulfide nitrogen 10048889 10048888 Material: Aluminum
0.35% pentane, 19% oxygen, 100 ppm carbon monoxide, nitrogen 10007049 —
35 ppm hydrogen sulfide

Note: Reactive gases have an expiration date listed on each cylinder.This is to ensure the highest quality and accuracy for instrument calibration.Most cylinders
have an expiration date of 12 months.Check with MSA Customer Service for the exact shelf life of a particular calibration cylinder.
Model RPCalibration Cylinder Steel, Model BD-20 Calibration Cylinders
100% Methane Contents: Pressure 2200 psig; approximately 552 liters (20 cu ft) of gas at
ontents: Pressure 1000 psig; approximately 58 atmospheric pressure
ters of pure methane with odorant added to Size: 25" x 4W" Weight: 10 lb 9 oz Material: Steel
mell like natural gas in all commercial gas lines
ze: 8V" x 3" Weight: 1 lb 13 oz Gas Fill Gas Mixture Background Part No.
Material: Steel Methane 2.5% Methane, 15% Oxygen, 60 ppm Carbon Monoxide Nitrogen 710566
1.45% Methane, 15% Oxygen, 60 ppm Carbon Monoxide Nitrogen 710565
as Fill Gas Mixture Part No.
Air, zero Air, zero,THC<1 ppm — 710776
Methane 100% Methane 711014

Squirt® Gas Bump Tester

quirt Gas Bump Test Kit, less cylinder, but with required fittings and adapters, complete with instructions 813411

Cylinders for SquirtGas


Combustible
Squirt Gas Pentane Carbon Hydrogen
strument Cylinder Methane Simulant Oxygen Monoxide Sulfide Isobutylene Balance
xplosimeter 815307 2.5% See cylinder — — — — Air
asport 814350 2.5% — 15% O2 60 ppm CO — — Nitrogen
asport 814349 2.5% — 15% O2 300 ppm CO 35 ppm H2S — Nitrogen
assport, FiveStar 814497 1.3% 50% LEL 15% O2 60 ppm CO — — Nitrogen
assport, FiveStar 814559 1.3% 50% LEL 15% O2 300 ppm CO 35 ppm H2S — Nitrogen
MiniCO** Responder 814978 — — — 60 ppm CO — — Air Contents: Pressure 155
MiniH2S*Responder 814979 — — — — 35 ppm H2S — Nitrogen psig; approximately 11
MicroGard** 815308 1.3% 52% LEL 15% O2 — — — Nitrogen liters at 70°F
assport PID II 815704 — — — — — 100 ppm Air

Shelf life item. See note page 103. ** Requires calibration adapter.

Cylinder-Holders Calibration Cylinders Are Recyclable!


ingle Portable Cylinder-Holder MSA has affiliated with the Association of Retarded Citizens,
Butler County—a nonprofit organization that employs mentally
he Single Portable Cylinder-Holder is
retarded citizens—and established a Cylinder Recycling Center.
esigned for use with all MSA Model R
The calibration cylinders are not refilled, but salvaged for scrap.
nd Model RP cylinders. It fits neatly on a
All money generated from selling the scrap cylinders is placed in
workbench or shelf, and its unique design
the ARC operating fund for salaries and operating costs.
nsures that the calibration cylinder (with
he regulator attached) always stays Calibration cylinders are considered hazardous unless the cylinders
where you put it. are empty. MSA has developed special devalving tools that will
ensure the cylinder is empty and render the cylinder non-fillable.
ngle Cylinder-Holder, 6" wide x 13" long x 4" high 710386 For your convenience, MSA also offers a specially pre-addressed
shipping box (also recyclable) which helps you pay the lowest
Double Cylinder-Holder,W all-Mounted shipping costs available.
The MSA Recycling Center limits the
he Wall-Mounted Double Cylinder-Holder holds all return of calibration cylinders to
MSA Model R and Model RP cylinders. A molded base MSA-logoed cylinders only. Non-MSA
nd holding straps keep the cylinders securely in cylinders will be rejected. Contact the
osition, yet both cylinders and/or regulators can be Customer Service Center for additional
asily changed.The holder is ideal for field station details at 1-800-MSA-2222.
alibrations and can be easily mounted for workbench
pplications. Devalving Tool, Model “RP”& Econo-Cal Cylinders 711228
ouble Cylinder-Holder, 17" high, 10" wide, 3V" deep at base 710483 Devalving Tool, Model “R”cylinders 711229
Shipping Box, Special, Pre-addressed 711227