You are on page 1of 7

Mã số/ Code: LAB 08/WI-LAB-03

BẢNG KIỂM TRA CƯỜNG LỰC GĂNG TAY THEO PO


Lần ban hành/ Issue: 01
PO GLOVES TENSILE RECORD
Ngày/ Date: 23/12/21

Tensile Test Report


Specification: EN 455-02

Customer: HARZ KAUR Test date: 21/10/2022


PO: Temperature: 24.9oC
Type: BLUE NITRILE-TEXTURED Moisture : 50%
Weight of glove: 3.0G Device Code: 1.TS.1
Device Code Digital thickness gage: 1.TG.1

Elongation at Ultimate
Thickness Area Force at break Tensile strength
No. Size break elongation
mm mm2 N Mpa
mm %
1 M
2 M
3 M
4 M
5 M
6 M
7 M
8 M
9 M
10 M
11 M
12 M
13 M
RESULT #VALUE! #VALUE! #VALUE! #VALUE! #VALUE! #VALUE!

Test by : HIẾU Manager : HUY


/ Code: LAB 08/WI-LAB-03
an hành/ Issue: 01
Date: 23/12/21

ate: 21/10/2022
erature: 24.9oC
ure : 50%
e Code: 1.TS.1

Stress at 300%

#VALUE!
Mã số/ Code: LAB 08/WI-LAB-03
BẢNG KIỂM TRA CƯỜNG LỰC GĂNG TAY THEO PO
Lần ban hành/ Issue: 01
PO GLOVES TENSILE RECORD
Ngày/ Date: 23/12/21

Tensile Test Report


Specification: EN 455-02

Customer: HARZ KAUR Test date: 21/10/2022


PO: Temperature: 24.9oC
Type: BLUE NITRILE-TEXTURED Moisture : 50%
Weight of glove: 3.0G Device Code: 1.TS.1
Device Code Digital thickness gage: 1TG1

Force at Elongation at Ultimate Tensile


Thickness Area
No. Size break break elongation strength Stress at 300%
mm mm2
N mm % Mpa
1 M 0.058 0.174 5.926 151.058 604.233 34.057 4.713
2 M 0.054 0.162 5.821 151.765 607.058 35.932 4.500
3 M 0.056 0.168 5.989 140.900 563.601 35.649 4.724
4 M 0.058 0.174 5.980 144.395 577.580 34.368 4.418
5 M 0.062 0.186 5.955 143.891 575.563 32.016 4.359
6 M 0.053 0.159 5.932 143.229 572.917 37.308 4.369
7 M 0.057 0.171 5.971 144.097 576.388 34.918 4.345
8 M 0.052 0.156 5.902 142.132 568.528 37.833 4.654
9 M 0.055 0.165 5.907 141.194 564.777 35.800 4.464
10 M 0.053 0.159 5.875 143.860 575.439 36.950 4.426
11 M 0.062 0.186 5.930 146.436 585.745 31.882 4.541
12 M 0.054 0.162 5.811 151.941 607.764 35.870 4.337
13 M 0.061 0.183 5.806 146.329 585.317 31.727 4.475
RESULT 0.056 0.168 5.926 144.097 576.388 35.649 4.464

Test by : HIẾU Manager : HUY


Mã số/ Code: LAB 08/WI-LAB-03
BẢNG KIỂM TRA CƯỜNG LỰC GĂNG TAY THEO PO
Lần ban hành/ Issue: 01
PO GLOVES TENSILE RECORD
Ngày/ Date: 23/12/21

Tensile Test Report


Specification: EN 455-02

Customer: HARZ KAUR Test date: 22/10/2022


PO: Temperature: 24.6oC
Type: BLUE NITRILE-TEXTURED Moisture : 51%
Weight of glove: 3.0G Device Code: 1TS.1
Device Code Digital thickness gage: 1TG1 AGING

Force at Elongation at Ultimate Tensile


Thickness Area
No. Size break break elongation strength
mm mm2
N mm % Mpa
1 M 0.055 0.165 5.862 136.241 544.963 35.527
2 M 0.057 0.171 6.168 145.560 582.239 36.070
3 M 0.054 0.162 5.855 141.978 567.910 36.142
4 M 0.063 0.189 6.041 145.287 581.147 31.963
5 M 0.055 0.165 6.054 139.915 559.659 36.691
6 M 0.062 0.186 6.015 146.698 586.793 32.339
7 M 0.053 0.159 6.086 147.378 589.513 38.277
8 M 0.057 0.171 5.850 136.746 546.984 34.211
9 M 0.062 0.186 5.911 143.387 573.549 31.780
10 M 0.056 0.168 6.162 141.072 564.286 36.679
11 M 0.060 0.180 6.057 143.171 572.685 33.650
12 M 0.057 0.171 6.193 135.112 540.449 36.216
13 M 0.052 0.156 5.816 140.847 563.386 37.282
RESULT 0.057 0.171 6.041 141.978 567.910 36.070

Test by : HIẾU Manager : HUY


số/ Code: LAB 08/WI-LAB-03
ban hành/ Issue: 01
y/ Date: 23/12/21

t date: 22/10/2022
mperature: 24.6oC

ice Code: 1TS.1

Stress at
300%

5.370
5.013
5.488
5.353
5.459
4.924
4.956
5.003
5.107
5.369
5.256
4.823
4.816
5.107

nager : HUY
POWDER CONTENT WORKSHEET
BẢNG KIỂM TRA HÀM LƯỢNG BỘT

Date / Ngày: 21/10/2022


Product / Sản phẩm: Powder free glove
Batch No:
Brand:

DESCRIPTION / Mô tả

Weight of filter paper,W1(g)


Khối lượng giấy lọc
Weight of powder + filter paper, W2(g) Khối lượng
bột + giấy lọc
POWDER CONTENT

Powder content A = (W2 – W1) (g)


(Hàm lượng bột)

Weight of filter paper, W3(g)


Khối lượng giấy lọc
Weight of filter paper (Blank sample), W4(g) Khối
lượng giấy lọc (mẫu trắng)
Weight of blank sample B=(W4 – W3)
(g) khối lượng mẫu trắng

Powder comtent: ( A  B ,) (mg/glove)


x1000
5
Hàm lượng bột

Checked by / Người kiểm tra: HIẾU

Disposition / Quyết định:


Đạt

Remarks / Ghi chú:


Remarks / Ghi chú:
Mã Số/ Code: LAB 09/WI-LAB-04
Lần ban hành/ Issue: 02
Ngày/ Date: 27/02/22

Line:
Size: M - TEXTURED
Device Code: 1.S5.1

SPECIFICATION SAMPLE 1
Tiêu chuẩn Mẫu 1

0.5518

0.5543

0.0025

0.5421
≤ 2mg/glove
0.5436

0.0015

0.20

Verified by / Người xác nhận: HUY

Không Đạt