Software Info DB: cobas_e_411 24th of March 2009 10:51:48 AM

List of Records for System/s: cobas e 411

Date of previous/last Publication: 14-Mai-2008 / 24-Mrz-2009

Color of new Records since previous Publication


Color of modified Records since previous Publication

Sort Order: LastUpdate,DESC

Click the record no. to get the complete information !

Record Version Section Created LastUpdate SysPart Category Title Status Status Ver
AE-000150 </=V02-01 Hint 06-Mar-2009 24-Mar-2009 cobas e 411 Others DVD RAM's can be ordered from Roche directly only for information
AE-000151 V02-01 Hint 09-Mar-2009 24-Mar-2009 cobas e 411 Troubleshooting The system information is included completely in the Online Help only for information
AE-000149 </=V02-01 Hint 06-Mar-2009 24-Mar-2009 cobas e 411 QC Specification of how many reagent lots of one test can be viewed on the "QC Install Edit" screen only for information
AE-000154 V01-02 Issue 17-Mar-2009 24-Mar-2009 cobas e 411 QC Standby bottle QC not displayed correctly is solved in -> V02-01
AE-000155 V01-02 Issue 17-Mar-2009 24-Mar-2009 cobas e 411 QC The control PC THYROD2 is not displayed in the Run status screen is solved in -> V02-01
AE-000157 </=V02-01 Hint 18-Mar-2009 24-Mar-2009 cobas e 411 Reagent Reagent packs/tests have been disappeared from the reagent screen in a exception state only for information
AE-000158 V02-01 Hint 23-Mar-2009 24-Mar-2009 cobas e 411 Specification Rack Assignment are limited to 100 Rack ID's for calibrators and controls only for information
AE-000153 V02-01 Issue 10-Mar-2009 23-Mar-2009 cobas e 411 QC Wrong prioritization of a tube/cup with "control barcode label" as opposed to "position assignment" is planned to be solved in -> >V02-01
AE-000127 V02-01 Issue 29-Jan-2009 17-Mar-2009 cobas e 411 System Overview Exception within the Sample Tracking screen no change planned
AE-000156 V01-02 Issue 17-Mar-2009 17-Mar-2009 cobas e 411 Host Result upload not possible under investigation
AE-000126 V01-02 Issue 29-Jan-2009 12-Mar-2009 cobas e 411 QC QC in status screen not saved is solved in -> V02-01
AE-000133 V02-01 Hint 30-Jan-2009 12-Mar-2009 cobas e 411 Operation Calib Mask bar is displayed in addition with Reagent Mask bar under specific circumstances only for information
AE-000125 </=V02-01 Hint 28-Jan-2009 12-Mar-2009 cobas e 411 Operation Workaround to use the so called "keep function" simular to E2010 only for information
AE-000144 V02-01 Issue 02-Feb-2009 12-Mar-2009 cobas e 411 Alarm If a RP and CalSet of two different lots are placed on board an alarm occurs no change planned
AE-000147 </=V02-01 Hint 06-Mar-2009 12-Mar-2009 cobas e 411 Configuration In total 60 tests can be installed only for information
AE-000148 </=V02-01 Hint 06-Mar-2009 12-Mar-2009 cobas e 411 Reagent Correct handling with SysWash only for information
AE-000039 </=V02-01 Issue 22-Jan-2007 12-Mar-2009 cobas e 411 Dilution If a "diluent ratio" is selected for a test by mistake, no alarm message occurs no change planned
AE-000130 V02-01 Issue 30-Jan-2009 12-Mar-2009 cobas e 411 Configuration Special keyboard setting no change planned
AE-000132 V02-01 Hint 30-Jan-2009 12-Mar-2009 cobas e 411 Operation Only 6 digits for the reagent lot no. can be displayed in the Test Review screen no change planned
AE-000152 V02-01 Issue 10-Mar-2009 12-Mar-2009 cobas e 411 QC Deletion of only 1 "cumulative QC-Chart" causes deletion of all "QC-Charts" from same reagent lot no change planned
AE-000140 V02-01 Issue 02-Feb-2009 12-Mar-2009 cobas e 411 Result Data with Review By Exception flags are displayed with"H" instead of "I" for incomplete is planned to be solved in -> >V02-01
AE-000135 V02-01 Issue 30-Jan-2009 12-Mar-2009 cobas e 411 System Overview The colour of Stand by QC is displayed as valid, although it has not been measured no change planned
AE-000136 V02-01 Issue 02-Feb-2009 12-Mar-2009 cobas e 411 Calibration No "change over" calibration request is displayed for "Stand by" reagent packs no change planned
AE-000053 </=V02-01 Issue 30-Mar-2007 12-Mar-2009 cobas e 411 Alarm The alarm messages will be displayed in all language versions improvements are planned in -> >Int. V 2.1.4
AE-000119 V01-02 Issue 09-Dez-2008 12-Mar-2009 cobas e 411 QC Control results did not appear on screens is planned to be solved in -> V02-01
AE-000141 V02-01 Issue 02-Feb-2009 12-Mar-2009 cobas e 411 Alarm Rack- and position information is displayed and printed incorrectly is planned to be solved in -> >V02-01
AE-000124 V02-01 Hint 28-Jan-2009 12-Mar-2009 cobas e 411 Operation How to "switch off/on the system"... only for information
AE-000117 V01-02 Issue 25-Sep-2008 12-Mar-2009 cobas e 411 Result Test order is mixed up in Data Review screen after changes from "Name" to "Test No." settings is planned to be solved in -> V02-01
AE-000121 </=V02-01 Hint 27-Jan-2009 12-Mar-2009 cobas e 411 Troubleshooting In case there is any issue please open a Clarify case as usual with the corresponding information only for information
AE-000116 V01-02 Issue 25-Sep-2008 12-Mar-2009 cobas e 411 Reagent The status for QC values from PC U1/U2 for FT4 were displayed as "Stand by" R/P's incorrectly is planned to be solved in -> V02-01
AE-000134 V02-01 Issue 30-Jan-2009 12-Mar-2009 cobas e 411 Calibration The volume on the calibrators are not calculated if Standby RP's are used no change planned
AE-000122 </=V02-01 Hint 28-Jan-2009 12-Mar-2009 cobas e 411 QC QC Violation settings is not possible for more than one control lot only for information

created by 'Software Information Tool' [V 1.0] for Roche-internal use only ! 1


Software Info DB: cobas_e_411 24th of March 2009 10:51:48 AM

List of Records for System/s: cobas e 411

Date of previous/last Publication: 14-Mai-2008 / 24-Mrz-2009

Color of new Records since previous Publication


Color of modified Records since previous Publication

Sort Order: LastUpdate,DESC

Click the record no. to get the complete information !

Record Version Section Created LastUpdate SysPart Category Title Status Status Ver
AE-000011 V02-01 Hint 22-Jan-2007 06-Mar-2009 cobas e 411 Troubleshooting Different options to store system and run data to an internal or external device only for information
AE-000098 </=V02-01 Hint 16-Jan-2008 06-Mar-2009 cobas e 411 Others FBT with Standard Cup Adapter can be used on rack system only only for information
AE-000035 </=V02-01 Hint 22-Jan-2007 06-Mar-2009 cobas e411 Documentation The print out of the "Application Parameter Report" is incorrectly displayed or printed no change planned
AE-000005 V02-01 Hint 22-Jan-2007 06-Mar-2009 cobas e 411 Result The " Rerun" button in Workplace/DataReview screen marks the sample ID only as a rerun only for information
AE-000034 </=V02-01 Hint 22-Jan-2007 06-Mar-2009 cobas e 411 QC The "Control Comment button" under QC/Individual/Chart is limited with 20 characters no change planned
AE-000006 </=V02-01 Hint 22-Jan-2007 06-Mar-2009 cobas e 411 Configuration Yellow colour is displayed for a test if "Reagent Level Check" setting fall short only for information
AE-000080 </=V02-01 Hint 10-Aug-2007 06-Mar-2009 cobas e 411 Host Host communication trace can only be printed no change planned
AE-000106 </=V02-01 Hint 17-Apr-2008 06-Mar-2009 cobas e 411 Troubleshooting Additional information to avoid "Slow Screen Response Time" only for information
AE-000027 </=V02-01 Issue 22-Jan-2007 06-Mar-2009 cobas e 411 Operation The status of an "empty" reagent/pre-treatment pack do not change correctly is planned to be solved in -> >V02-01
AE-000077 </=V02-01 Hint 24-Jul-2007 06-Mar-2009 cobas e 411 Troubleshooting Install the Roche CD Version before starting settings and routine measurements only for information
AE-000056 </=V02-01 Hint 04-Mai-2007 06-Mar-2009 cobas e 411 Operation System water consumptions per day information only
AE-000041 </=V02-01 Hint 22-Jan-2007 06-Mar-2009 cobas e 411 Reagent The "data file" for reagent packs can be manage up to 300 reagent packs only for information
AE-000086 V02-01 Hint 24-Okt-2007 06-Mar-2009 cobas e 411 QC QC results is not displayed in chronological order or they are not printed only for information
AE-000128 V02-01 Hint 30-Jan-2009 26-Feb-2009 cobas e 411 Documentation Reagent lot number is not displayed on the result print out only for information
AE-000093 </=V02-01 Hint 11-Dez-2007 26-Feb-2009 cobas e 411 Configuration The setting "Rack ID Digit 5" should be used for all rack ranges assignment only for information
AE-000137 V02-01 Hint 02-Feb-2009 26-Feb-2009 cobas e 411 Documentation Misleading information on result printout for Calculated tests no change planned
AE-000059 </=V02-01 Issue 04-Mai-2007 26-Feb-2009 cobas e 411 QC Printout for selected QC charts is planned to be solved in -> >V02-01
AE-000089 </=V02-01 Hint 24-Okt-2007 26-Feb-2009 cobas e 411 Dilution The sample volume dilution remains after it has been programmed only for information
AE-000075 </=V02-01 Issue 24-Jul-2007 26-Feb-2009 cobas e 411 QC There is no data link between Workplace/DataReview/QCView and QC-Individual screen no change planned
AE-000103 </=V02-01 Hint 18-Feb-2008 28-Jan-2009 cobas e 411 QC Both systems (Rack and Disk) can not deal with barcode label for Non-Roche controls only for information
AE-000067 </=V02-01 Hint 29-Jun-2007 28-Jan-2009 cobas e 411 Troubleshooting Backup data can not be done, if the external DVD-RAM drive is assigned with the wrong drive letter only for information
AE-000111 </=V02-01 Hint 24-Apr-2008 28-Jan-2009 cobas e 411 Configuration The "Windows Time Zone" should set to the default settings "Tokyo" independents of the countries only for information
AE-000112 </=V02-01 Hint 29-Apr-2008 28-Jan-2009 cobas e 411 Troubleshooting How to ...copy "data files" from the computer to the DVD Writer only for information
AE-000049 </=V02-01 Hint 26-Mar-2007 28-Jan-2009 cobas e 411 Others Abbreviation of cobas only for information
AE-000066 </=V02-01 Hint 25-Jun-2007 28-Jan-2009 cobas e 411 Alarm If the system is rebooted sound setting OFF is changed to ON only for information
AE-000105 </=V02-01 Hint 07-Mar-2008 28-Jan-2009 cobas e 411 Host Manufacturer Specific Records should not be send to Host only for information
AE-000094 </=V02-01 Hint 11-Dez-2007 28-Jan-2009 cobas e 411 Configuration Calculated test screen can be used partly only for information
AE-000073 </=V02-01 Hint 24-Jul-2007 28-Jan-2009 cobas e 411 Reagent Place the Universal Diluent reagent pack/barcode label into correct reagent rotor position only for information
AE-000074 </=V02-01 Hint 24-Jul-2007 28-Jan-2009 cobas e 411 Operation Keyboard appears in help screen only for information
AE-000071 </=V02-01 Hint 03-Jul-2007 28-Jan-2009 cobas e 411 Configuration The setting OFF on the "Automatic Printout" for Routine Sample is more efficiently only for information
AE-000078 </=V02-01 Hint 06-Aug-2007 28-Jan-2009 cobas e 411 Operation Consumption of ProCell during "Finalization" process only for information
AE-000068 </=V02-01 Hint 03-Jul-2007 28-Jan-2009 cobas e 411 Operation Display of "sample status" is changed, if STAT port or routine tray is used wrongly only for information

created by 'Software Information Tool' [V 1.0] for Roche-internal use only ! 2


Software Info DB: cobas_e_411 24th of March 2009 10:51:48 AM

List of Records for System/s: cobas e 411

Date of previous/last Publication: 14-Mai-2008 / 24-Mrz-2009

Color of new Records since previous Publication


Color of modified Records since previous Publication

Sort Order: LastUpdate,DESC

Click the record no. to get the complete information !

Record Version Section Created LastUpdate SysPart Category Title Status Status Ver
AE-000058 </=V02-01 Hint 04-Mai-2007 28-Jan-2009 cobas e 411 Operation System reagent consumption in the sample reception mode only for information
AE-000008 </=V02-01 Hint 22-Jan-2007 22-Jan-2009 cobas e 411 Operation Dealing with STAT samples only for information

created by 'Software Information Tool' [V 1.0] for Roche-internal use only ! 3


Software Info DB: cobas_e_411 24th of March 2009 10:54:27 AM

List of ARCHIVED Records for cobas e 411

Date of previous/last Publication: 14-Mai-2008 / 24-Mrz-2009

Color of new Records since previous Publication


Color of modified Records since previous Publication

Sort Order: LastUpdate,DESC

Click the record no. to get the complete information !

Record Version Section Created LastUpdate SysPart Category Title Status Status Ver
AE-000101 V01-01 Issue ### ### cobas e 411 Operation Software remains in CF Access while performing sample results backup is solved in -> V02-01
AE-000083 </=V01-02 Issue ### ### cobas e 411 Reagent The "Reagent Load List" button in the System Overview turns yellow is solved in -> V02-01
AE-000036 </=V02-01 Issue ### ### cobas e 411 QC Current QC selection does not work correctly if both Stand by and Current QC's are selected is solved in -> V02-01
AE-000114 V01-02 Issue 29-Mai-2008 ### cobas e 411 QC Manual selection for a control can not be performed on Rack systems only is solved in -> V02-01
AE-000026 </=V01-02 Issue ### ### cobas e 411 Alarm The number of messages on the printout of "Cumulative Alarm Trace is displayed incorrectly is solved in -> V02-01
AE-000109 V01-02 Issue 17-Apr-2008 ### cobas e 411 QC An "unknown error" appeared, if a QC for a test will be installed, but reagent pack is not on board is solved in -> V02-01
AE-000024 </=V01-02 Issue ### ### cobas e 411 Configuration Sample processing works incorrectly, if a test number for a test is not stored in "refer dat file" no change planned
AE-000120 V01-02 Issue ### ### cobas e 411 QC Control was not displayed for Stand by rackpack is solved in -> V02-01
AE-000118 V01-02 Issue 08-Dez-2008 ### cobas e 411 QC QC's are not shown or printed in Cumulative Statistics is solved in -> V02-01
AE-000029 </=V01-02 Issue ### ### cobas e 411 Reagent Status of "Current" diluent reagent pack is changed to "Stand by" is solved in -> V02-01
AE-000115 </=V02-01 Hint ### ### cobas e 411 QC QC chart with the cumulated data (in light blue) still appears after deleting the QC individual Data only for information
AE-000129 V02-01 Issue ### ### cobas e 411 Host The system send a "Cancel" message to the Host in addition to the valid result is planned to be solved in -> >V02-01
AE-000095 </=V01-02 Issue 11-Dez-2007 ### cobas e 411 Operation Unexpected software crash pressing start button from the sampling stop status is solved in -> >V01-02
AE-000007 </=V01-02 Issue ### ### cobas e 411 QC QC after calibration is only valid for a test when the control directly follows the calibrator vials is planned to be solved in -> >V01-02
AE-000113 V01-02 Issue 26-Mai-2008 28-Mai-2008 cobas e 411 QC QC individual chart was printed incorrectly is solved in -> >V01-02
AE-000108 V01-02 Issue 17-Apr-2008 13-Mai-2008 cobas e 411 QC Different control results are displayed under the QC Individual list even only one QC is used is solved in -> >V01-02
AE-000081 </=V01-02 Issue ### 17-Apr-2008 cobas e 411 Maintenance Delete Calibration Data under Service part does not work is solved in -> >V01-02
AE-000107 V01-02 Issue 17-Apr-2008 17-Apr-2008 cobas e 411 QC QC values were displayed incorrectly in the chart with the date "15.02.2008" is solved in -> >V01-02
AE-000102 V01-02 Issue ### 17-Apr-2008 cobas e 411 Operation Unexpected software crash after pressing start button is solved in -> Int. V 1.2.3
AE-000085 </=V01-02 Issue 24-Okt-2007 17-Apr-2008 cobas e 411 Host Screen freeze while uploading results is solved in -> >V01-02
AE-000025 </=V01-02 Issue ### 17-Apr-2008 cobas e 411 QC Sometimes QC Violation is displayed when the setting of the QC-Violation is OFF is planned to be solved in -> >V01-02
AE-000031 </=V01-02 Issue ### 17-Apr-2008 cobas e 411 Operation Position assignment for calibrators /controls is reset after measurment directly improvements are planned in -> >V01-02
AE-000090 </=V01-02 Hint 10-Dez-2007 17-Apr-2008 cobas e 411 Calibration The sorting criteria of calibration causes are according to cobas 6000 specification only for information
AE-000084 V01-02 Issue 24-Okt-2007 17-Apr-2008 cobas e 411 QC Program error after control loading is solved in -> >V01-02
AE-000099 </=V01-02 Issue ### 17-Apr-2008 cobas e 411 Alarm Alarm 41-01-05 "CF has been accessed > 100.000 times" is solved in -> >V01-02
AE-000042 </=V01-02 Issue ### 17-Apr-2008 cobas e 411 QC Individual QC list and chart are displayed incorrectly in same cases is planned to be solved in -> >V01-02
AE-000087 </=V01-02 Issue 24-Okt-2007 17-Apr-2008 cobas e 411 QC Cumulative QC for CA 15-3 TM1and TM2 is solved in -> >V01-02
AE-000096 </=V01-02 Issue 11-Dez-2007 17-Apr-2008 cobas e 411 QC QC violation is indicated after a successful calibration is solved in -> >V01-02
AE-000097 </=V01-02 Issue 11-Dez-2007 17-Apr-2008 cobas e 411 Troubleshooting Restore function does not work for "system specific data" is solved in -> >V01-02
AE-000012 </=V01-02 Issue ### 16-Apr-2008 cobas e 411 Troubleshooting In case of a software trouble following informations are requested for investigations improvements are planned in -> >V01-02
AE-000018 </=V01-02 Issue ### 16-Apr-2008 cobas e 411 Operation Sample Tracking screen is deleted after pressing the Start button improvements are planned in -> >V01-02
AE-000104 V01-02 Issue ### ### cobas e 411 Host A typo error was noticed on the Communication Trace Screen in the software. is planned to be solved in -> V01-02

created by 'Software Information Tool' [V 1.0] for Roche-internal use only ! 4


Software Info DB: cobas_e_411 24th of March 2009 10:54:27 AM

List of ARCHIVED Records for cobas e 411

Date of previous/last Publication: 14-Mai-2008 / 24-Mrz-2009

Color of new Records since previous Publication


Color of modified Records since previous Publication

Sort Order: LastUpdate,DESC

Click the record no. to get the complete information !

Record Version Section Created LastUpdate SysPart Category Title Status Status Ver
AE-000100 V01-01 Issue ### ### cobas e 411 QC Software crashed when trying to edit a new control in the QC menu is solved in -> >V01-02
AE-000022 </=V01-02 Issue ### ### cobas e 411 QC The control results with commented and/or excluded status are not displayed with different colors is solved in -> >V01-02
AE-000091 </=V01-02 Hint 11-Dez-2007 12-Dez-2007 cobas e 411 Configuration User Trainings Guide for V.01-01 is available now only for information
AE-000092 </=V01-02 Hint 11-Dez-2007 12-Dez-2007 cobas e 411 Configuration A new printer "LaserJet P2015" is recommended, refer to SB no.6 only for information
AE-000054 </=V01-02 Issue 03-Mai-2007 11-Dez-2007 cobas e 411 QC Non-Roche control values changes automatically when a new RP is placed on board is solved in -> >V01-02
AE-000063 </=V01-02 Issue 21-Mai-2007 11-Dez-2007 cobas e 411 QC Memory capacity of Routine View and QC View in Workplace/DataReview improvements are planned in -> >V01-02
AE-000028 </=V01-02 Issue ### 11-Dez-2007 cobas e 411 Alarm No yellow warning is displayed, if the amount of system reagent set 1 is empty is planned to be solved in -> >V01-02
AE-000052 </=V01-02 Issue ### 11-Dez-2007 cobas e 411 Configuration The button in the "Retry function setup screen" will be changed from "Start" to "Ok" improvements are planned in -> >V01-02
AE-000082 </=V01-02 Issue ### 11-Dez-2007 cobas e 411 Host Communication failure occurrs sporadically is planned to be solved in -> >V01-02
AE-000019 </=V01-02 Issue ### 11-Dez-2007 cobas e 411 Maintenance Maintenance button on the Workflow Guide does not change colour is planned to be solved in -> >V01-02
AE-000048 </=V01-02 Issue ### 11-Dez-2007 cobas e 411 Calibration A "change over" request is not displayed, if registration time of a reagent pack is done earlier is planned to be solved in -> >V01-02
AE-000050 </=V01-02 Hint ### 11-Dez-2007 cobas e 411 QC The 3 SD control ranges are displayed on the QC printout only no change planned
AE-000038 </=V01-02 Issue ### 11-Dez-2007 cobas e 411 Reagent The display of system reagent on the System Overview is display incorrectly improvements are planned in -> >V01-02
AE-000023 </=V01-02 Issue ### 11-Dez-2007 cobas e 411 Operation Indication for a STATsample ID is not implemented on the Disk system. improvements are planned in -> >V01-02
AE-000055 </=V01-02 Hint 04-Mai-2007 11-Dez-2007 cobas e 411 Configuration Application parameter report can only be done for all the tests only for information
AE-000021 </=V01-02 Issue ### 11-Dez-2007 cobas e 411 Operation The test keys colours "dark grey" is not changed after result is available improvements are planned in -> >V01-01
AE-000070 </=V01-02 Issue 03-Jul-2007 11-Dez-2007 cobas e 411 Configuration Incorrect handling of some Documentation and Printer Settings is solved in -> >V01-02
AE-000045 </=V01-02 Issue ### 11-Dez-2007 cobas e 411 Documentation Printout of "Worklist" is not implemented the for Rack system no change planned >V01-02
AE-000079 </=V01-02 Hint ### 11-Dez-2007 cobas e 411 QC QC after Calibration on cobas e411 only for information
AE-000069 </=V01-02 Issue 03-Jul-2007 11-Dez-2007 cobas e 411 Operation During operation it is not possible to change the Operator ID with "Logoff" button is planned to be solved in -> >V01-02
AE-000017 </=V01-02 Issue ### 11-Dez-2007 cobas e 411 Documentation Print View and Print function does not work correctly is planned to be solved in -> >V01-02
AE-000032 </=V01-02 Issue ### 11-Dez-2007 cobas e 411 Result It is not possible to print out the results with "Review by Exception" flags improvements are planned in -> >V01-02
AE-000033 </=V01-02 Issue ### 11-Dez-2007 cobas e 411 Result Reagent information on the Result Report is missing improvements are planned in -> >V01-02
AE-000030 </=V01-02 Issue ### 11-Dez-2007 cobas e 411 Host Cobas host protocol type is not available yet is planned to be solved in -> >V01-02
AE-000037 </=V01-02 Issue ### 11-Dez-2007 cobas e 411 QC Confidence limit field is activated even thogh QC Violation setting is OFF is planned to be solved in -> >V01-02
AE-000044 </=V01-02 Issue ### 11-Dez-2007 cobas e 411 Operation Some of the screens are not updated in "real-time" mode, even though new data are produced is solved in -> >V01-02
AE-000040 </=V01-02 Issue ### 11-Dez-2007 cobas e 411 QC The "Run status" screen shows control data with old and new "Unit" if the test unit is changed is planned to be solved in -> >V01-02
AE-000020 </=V01-02 Issue ### 11-Dez-2007 cobas e 411 Calibration Selection box "Auto Masking" under Utility/Application/Calib is missing improvements are planned in -> >V01-02
AE-000064 </=V01-02 Issue ### 11-Dez-2007 cobas e 411 Alarm Barcode setting for "Read Error Alarm" ON or OFF under Utility has no relevance is planned to be solved in -> >V01-02
AE-000047 </=V01-02 Issue ### 11-Dez-2007 cobas e 411 Configuration Difference in "Sample Reception Mode" settings and implementations improvements are planned in -> >V01-02
AE-000076 V01-01 Issue 24-Jul-2007 ### cobas e 411 Reagent The number of test is displayed to "0" determination, if a new reagent pack was scanned is solved in -> V01-02
AE-000061 Int. V 1.1.8 Hint 07-Mai-2007 ### cobas e 411 Troubleshooting After language file installation an error occurs on the installation report only for information

created by 'Software Information Tool' [V 1.0] for Roche-internal use only ! 5


Software Info DB: cobas_e_411 24th of March 2009 10:54:27 AM

List of ARCHIVED Records for cobas e 411

Date of previous/last Publication: 14-Mai-2008 / 24-Mrz-2009

Color of new Records since previous Publication


Color of modified Records since previous Publication

Sort Order: LastUpdate,DESC

Click the record no. to get the complete information !

Record Version Section Created LastUpdate SysPart Category Title Status Status Ver
AE-000051 V01-01 Issue ### 24-Jul-2007 cobas e 411 Operation The input no. 6 from on the entire keyboard can not be done is solved in -> Int. V 1.1.8
AE-000014 V01-01 Hint ### 03-Jul-2007 cobas e 411 Alarm Alarm information can be deactivated only for information
AE-000009 V01-01 Hint ### ### cobas e 411 Configuration Test applications can not be deleted only for information
AE-000060 V01-01 Issue 07-Mai-2007 ### cobas e 411 Operation The Back up data under Workplace/Data Review does not work correctly is solved in -> Int. V 1.1.8
AE-000003 V01-01 Hint ### 07-Mai-2007 cobas e 411 Calibration Calibrator rack assignment automaticallly only for information
AE-000004 V01-01 Hint ### 07-Mai-2007 cobas e 411 Operation Used reagent packs should used on one system only for information
AE-000001 V01-01 Hint ### 07-Mai-2007 cobas e 411 Operation Start up time from Initialization to Operation status only for information
AE-000015 V01-01 Hint ### 07-Mai-2007 cobas e 411 Operation Restart is necessary if "P.Stop alarm" occurs only for information
AE-000010 V01-01 Hint ### 07-Mai-2007 cobas e 411 Printer Details to the default printer only for information
AE-000043 V01-01 Hint ### 07-Mai-2007 cobas e 411 Calibration Timeout cause for a calibration request is not specified on the cobas e 411 software versions only for information
AE-000002 V01-01 Hint ### 07-Mai-2007 cobas e 411 Reagent Reagent lid has to be open slightly only for information
AE-000046 V01-01 Hint ### 07-Mai-2007 cobas e 411 Troubleshooting To avoid any trouble do not open the "sipper safety cover" during the run only for information
AE-000016 V01-01 Hint ### 07-Mai-2007 cobas e 411 Calibration Only one calibrator lot can be used per quantitative test only for information
AE-000013 V01-01 Hint ### ### cobas e 411 Operation Used reagent packs should used on one system only

created by 'Software Information Tool' [V 1.0] for Roche-internal use only ! 6