Record Version Section Created LastUpdate SysPart Category Title Status Status Ver
AE-000150 </=V02-01 Hint 06-Mar-2009 24-Mar-2009 cobas e 411 Others DVD RAM's can be ordered from Roche directly only for information
AE-000151 V02-01 Hint 09-Mar-2009 24-Mar-2009 cobas e 411 Troubleshooting The system information is included completely in the Online Help only for information
AE-000149 </=V02-01 Hint 06-Mar-2009 24-Mar-2009 cobas e 411 QC Specification of how many reagent lots of one test can be viewed on the "QC Install Edit" screen only for information
AE-000154 V01-02 Issue 17-Mar-2009 24-Mar-2009 cobas e 411 QC Standby bottle QC not displayed correctly is solved in -> V02-01
AE-000155 V01-02 Issue 17-Mar-2009 24-Mar-2009 cobas e 411 QC The control PC THYROD2 is not displayed in the Run status screen is solved in -> V02-01
AE-000157 </=V02-01 Hint 18-Mar-2009 24-Mar-2009 cobas e 411 Reagent Reagent packs/tests have been disappeared from the reagent screen in a exception state only for information
AE-000158 V02-01 Hint 23-Mar-2009 24-Mar-2009 cobas e 411 Specification Rack Assignment are limited to 100 Rack ID's for calibrators and controls only for information
AE-000153 V02-01 Issue 10-Mar-2009 23-Mar-2009 cobas e 411 QC Wrong prioritization of a tube/cup with "control barcode label" as opposed to "position assignment" is planned to be solved in -> >V02-01
AE-000127 V02-01 Issue 29-Jan-2009 17-Mar-2009 cobas e 411 System Overview Exception within the Sample Tracking screen no change planned
AE-000156 V01-02 Issue 17-Mar-2009 17-Mar-2009 cobas e 411 Host Result upload not possible under investigation
AE-000126 V01-02 Issue 29-Jan-2009 12-Mar-2009 cobas e 411 QC QC in status screen not saved is solved in -> V02-01
AE-000133 V02-01 Hint 30-Jan-2009 12-Mar-2009 cobas e 411 Operation Calib Mask bar is displayed in addition with Reagent Mask bar under specific circumstances only for information
AE-000125 </=V02-01 Hint 28-Jan-2009 12-Mar-2009 cobas e 411 Operation Workaround to use the so called "keep function" simular to E2010 only for information
AE-000144 V02-01 Issue 02-Feb-2009 12-Mar-2009 cobas e 411 Alarm If a RP and CalSet of two different lots are placed on board an alarm occurs no change planned
AE-000147 </=V02-01 Hint 06-Mar-2009 12-Mar-2009 cobas e 411 Configuration In total 60 tests can be installed only for information
AE-000148 </=V02-01 Hint 06-Mar-2009 12-Mar-2009 cobas e 411 Reagent Correct handling with SysWash only for information
AE-000039 </=V02-01 Issue 22-Jan-2007 12-Mar-2009 cobas e 411 Dilution If a "diluent ratio" is selected for a test by mistake, no alarm message occurs no change planned
AE-000130 V02-01 Issue 30-Jan-2009 12-Mar-2009 cobas e 411 Configuration Special keyboard setting no change planned
AE-000132 V02-01 Hint 30-Jan-2009 12-Mar-2009 cobas e 411 Operation Only 6 digits for the reagent lot no. can be displayed in the Test Review screen no change planned
AE-000152 V02-01 Issue 10-Mar-2009 12-Mar-2009 cobas e 411 QC Deletion of only 1 "cumulative QC-Chart" causes deletion of all "QC-Charts" from same reagent lot no change planned
AE-000140 V02-01 Issue 02-Feb-2009 12-Mar-2009 cobas e 411 Result Data with Review By Exception flags are displayed with"H" instead of "I" for incomplete is planned to be solved in -> >V02-01
AE-000135 V02-01 Issue 30-Jan-2009 12-Mar-2009 cobas e 411 System Overview The colour of Stand by QC is displayed as valid, although it has not been measured no change planned
AE-000136 V02-01 Issue 02-Feb-2009 12-Mar-2009 cobas e 411 Calibration No "change over" calibration request is displayed for "Stand by" reagent packs no change planned
AE-000053 </=V02-01 Issue 30-Mar-2007 12-Mar-2009 cobas e 411 Alarm The alarm messages will be displayed in all language versions improvements are planned in -> >Int. V 2.1.4
AE-000119 V01-02 Issue 09-Dez-2008 12-Mar-2009 cobas e 411 QC Control results did not appear on screens is planned to be solved in -> V02-01
AE-000141 V02-01 Issue 02-Feb-2009 12-Mar-2009 cobas e 411 Alarm Rack- and position information is displayed and printed incorrectly is planned to be solved in -> >V02-01
AE-000124 V02-01 Hint 28-Jan-2009 12-Mar-2009 cobas e 411 Operation How to "switch off/on the system"... only for information
AE-000117 V01-02 Issue 25-Sep-2008 12-Mar-2009 cobas e 411 Result Test order is mixed up in Data Review screen after changes from "Name" to "Test No." settings is planned to be solved in -> V02-01
AE-000121 </=V02-01 Hint 27-Jan-2009 12-Mar-2009 cobas e 411 Troubleshooting In case there is any issue please open a Clarify case as usual with the corresponding information only for information
AE-000116 V01-02 Issue 25-Sep-2008 12-Mar-2009 cobas e 411 Reagent The status for QC values from PC U1/U2 for FT4 were displayed as "Stand by" R/P's incorrectly is planned to be solved in -> V02-01
AE-000134 V02-01 Issue 30-Jan-2009 12-Mar-2009 cobas e 411 Calibration The volume on the calibrators are not calculated if Standby RP's are used no change planned
AE-000122 </=V02-01 Hint 28-Jan-2009 12-Mar-2009 cobas e 411 QC QC Violation settings is not possible for more than one control lot only for information
Record Version Section Created LastUpdate SysPart Category Title Status Status Ver
AE-000011 V02-01 Hint 22-Jan-2007 06-Mar-2009 cobas e 411 Troubleshooting Different options to store system and run data to an internal or external device only for information
AE-000098 </=V02-01 Hint 16-Jan-2008 06-Mar-2009 cobas e 411 Others FBT with Standard Cup Adapter can be used on rack system only only for information
AE-000035 </=V02-01 Hint 22-Jan-2007 06-Mar-2009 cobas e411 Documentation The print out of the "Application Parameter Report" is incorrectly displayed or printed no change planned
AE-000005 V02-01 Hint 22-Jan-2007 06-Mar-2009 cobas e 411 Result The " Rerun" button in Workplace/DataReview screen marks the sample ID only as a rerun only for information
AE-000034 </=V02-01 Hint 22-Jan-2007 06-Mar-2009 cobas e 411 QC The "Control Comment button" under QC/Individual/Chart is limited with 20 characters no change planned
AE-000006 </=V02-01 Hint 22-Jan-2007 06-Mar-2009 cobas e 411 Configuration Yellow colour is displayed for a test if "Reagent Level Check" setting fall short only for information
AE-000080 </=V02-01 Hint 10-Aug-2007 06-Mar-2009 cobas e 411 Host Host communication trace can only be printed no change planned
AE-000106 </=V02-01 Hint 17-Apr-2008 06-Mar-2009 cobas e 411 Troubleshooting Additional information to avoid "Slow Screen Response Time" only for information
AE-000027 </=V02-01 Issue 22-Jan-2007 06-Mar-2009 cobas e 411 Operation The status of an "empty" reagent/pre-treatment pack do not change correctly is planned to be solved in -> >V02-01
AE-000077 </=V02-01 Hint 24-Jul-2007 06-Mar-2009 cobas e 411 Troubleshooting Install the Roche CD Version before starting settings and routine measurements only for information
AE-000056 </=V02-01 Hint 04-Mai-2007 06-Mar-2009 cobas e 411 Operation System water consumptions per day information only
AE-000041 </=V02-01 Hint 22-Jan-2007 06-Mar-2009 cobas e 411 Reagent The "data file" for reagent packs can be manage up to 300 reagent packs only for information
AE-000086 V02-01 Hint 24-Okt-2007 06-Mar-2009 cobas e 411 QC QC results is not displayed in chronological order or they are not printed only for information
AE-000128 V02-01 Hint 30-Jan-2009 26-Feb-2009 cobas e 411 Documentation Reagent lot number is not displayed on the result print out only for information
AE-000093 </=V02-01 Hint 11-Dez-2007 26-Feb-2009 cobas e 411 Configuration The setting "Rack ID Digit 5" should be used for all rack ranges assignment only for information
AE-000137 V02-01 Hint 02-Feb-2009 26-Feb-2009 cobas e 411 Documentation Misleading information on result printout for Calculated tests no change planned
AE-000059 </=V02-01 Issue 04-Mai-2007 26-Feb-2009 cobas e 411 QC Printout for selected QC charts is planned to be solved in -> >V02-01
AE-000089 </=V02-01 Hint 24-Okt-2007 26-Feb-2009 cobas e 411 Dilution The sample volume dilution remains after it has been programmed only for information
AE-000075 </=V02-01 Issue 24-Jul-2007 26-Feb-2009 cobas e 411 QC There is no data link between Workplace/DataReview/QCView and QC-Individual screen no change planned
AE-000103 </=V02-01 Hint 18-Feb-2008 28-Jan-2009 cobas e 411 QC Both systems (Rack and Disk) can not deal with barcode label for Non-Roche controls only for information
AE-000067 </=V02-01 Hint 29-Jun-2007 28-Jan-2009 cobas e 411 Troubleshooting Backup data can not be done, if the external DVD-RAM drive is assigned with the wrong drive letter only for information
AE-000111 </=V02-01 Hint 24-Apr-2008 28-Jan-2009 cobas e 411 Configuration The "Windows Time Zone" should set to the default settings "Tokyo" independents of the countries only for information
AE-000112 </=V02-01 Hint 29-Apr-2008 28-Jan-2009 cobas e 411 Troubleshooting How to ...copy "data files" from the computer to the DVD Writer only for information
AE-000049 </=V02-01 Hint 26-Mar-2007 28-Jan-2009 cobas e 411 Others Abbreviation of cobas only for information
AE-000066 </=V02-01 Hint 25-Jun-2007 28-Jan-2009 cobas e 411 Alarm If the system is rebooted sound setting OFF is changed to ON only for information
AE-000105 </=V02-01 Hint 07-Mar-2008 28-Jan-2009 cobas e 411 Host Manufacturer Specific Records should not be send to Host only for information
AE-000094 </=V02-01 Hint 11-Dez-2007 28-Jan-2009 cobas e 411 Configuration Calculated test screen can be used partly only for information
AE-000073 </=V02-01 Hint 24-Jul-2007 28-Jan-2009 cobas e 411 Reagent Place the Universal Diluent reagent pack/barcode label into correct reagent rotor position only for information
AE-000074 </=V02-01 Hint 24-Jul-2007 28-Jan-2009 cobas e 411 Operation Keyboard appears in help screen only for information
AE-000071 </=V02-01 Hint 03-Jul-2007 28-Jan-2009 cobas e 411 Configuration The setting OFF on the "Automatic Printout" for Routine Sample is more efficiently only for information
AE-000078 </=V02-01 Hint 06-Aug-2007 28-Jan-2009 cobas e 411 Operation Consumption of ProCell during "Finalization" process only for information
AE-000068 </=V02-01 Hint 03-Jul-2007 28-Jan-2009 cobas e 411 Operation Display of "sample status" is changed, if STAT port or routine tray is used wrongly only for information
Record Version Section Created LastUpdate SysPart Category Title Status Status Ver
AE-000058 </=V02-01 Hint 04-Mai-2007 28-Jan-2009 cobas e 411 Operation System reagent consumption in the sample reception mode only for information
AE-000008 </=V02-01 Hint 22-Jan-2007 22-Jan-2009 cobas e 411 Operation Dealing with STAT samples only for information
Record Version Section Created LastUpdate SysPart Category Title Status Status Ver
AE-000101 V01-01 Issue ### ### cobas e 411 Operation Software remains in CF Access while performing sample results backup is solved in -> V02-01
AE-000083 </=V01-02 Issue ### ### cobas e 411 Reagent The "Reagent Load List" button in the System Overview turns yellow is solved in -> V02-01
AE-000036 </=V02-01 Issue ### ### cobas e 411 QC Current QC selection does not work correctly if both Stand by and Current QC's are selected is solved in -> V02-01
AE-000114 V01-02 Issue 29-Mai-2008 ### cobas e 411 QC Manual selection for a control can not be performed on Rack systems only is solved in -> V02-01
AE-000026 </=V01-02 Issue ### ### cobas e 411 Alarm The number of messages on the printout of "Cumulative Alarm Trace is displayed incorrectly is solved in -> V02-01
AE-000109 V01-02 Issue 17-Apr-2008 ### cobas e 411 QC An "unknown error" appeared, if a QC for a test will be installed, but reagent pack is not on board is solved in -> V02-01
AE-000024 </=V01-02 Issue ### ### cobas e 411 Configuration Sample processing works incorrectly, if a test number for a test is not stored in "refer dat file" no change planned
AE-000120 V01-02 Issue ### ### cobas e 411 QC Control was not displayed for Stand by rackpack is solved in -> V02-01
AE-000118 V01-02 Issue 08-Dez-2008 ### cobas e 411 QC QC's are not shown or printed in Cumulative Statistics is solved in -> V02-01
AE-000029 </=V01-02 Issue ### ### cobas e 411 Reagent Status of "Current" diluent reagent pack is changed to "Stand by" is solved in -> V02-01
AE-000115 </=V02-01 Hint ### ### cobas e 411 QC QC chart with the cumulated data (in light blue) still appears after deleting the QC individual Data only for information
AE-000129 V02-01 Issue ### ### cobas e 411 Host The system send a "Cancel" message to the Host in addition to the valid result is planned to be solved in -> >V02-01
AE-000095 </=V01-02 Issue 11-Dez-2007 ### cobas e 411 Operation Unexpected software crash pressing start button from the sampling stop status is solved in -> >V01-02
AE-000007 </=V01-02 Issue ### ### cobas e 411 QC QC after calibration is only valid for a test when the control directly follows the calibrator vials is planned to be solved in -> >V01-02
AE-000113 V01-02 Issue 26-Mai-2008 28-Mai-2008 cobas e 411 QC QC individual chart was printed incorrectly is solved in -> >V01-02
AE-000108 V01-02 Issue 17-Apr-2008 13-Mai-2008 cobas e 411 QC Different control results are displayed under the QC Individual list even only one QC is used is solved in -> >V01-02
AE-000081 </=V01-02 Issue ### 17-Apr-2008 cobas e 411 Maintenance Delete Calibration Data under Service part does not work is solved in -> >V01-02
AE-000107 V01-02 Issue 17-Apr-2008 17-Apr-2008 cobas e 411 QC QC values were displayed incorrectly in the chart with the date "15.02.2008" is solved in -> >V01-02
AE-000102 V01-02 Issue ### 17-Apr-2008 cobas e 411 Operation Unexpected software crash after pressing start button is solved in -> Int. V 1.2.3
AE-000085 </=V01-02 Issue 24-Okt-2007 17-Apr-2008 cobas e 411 Host Screen freeze while uploading results is solved in -> >V01-02
AE-000025 </=V01-02 Issue ### 17-Apr-2008 cobas e 411 QC Sometimes QC Violation is displayed when the setting of the QC-Violation is OFF is planned to be solved in -> >V01-02
AE-000031 </=V01-02 Issue ### 17-Apr-2008 cobas e 411 Operation Position assignment for calibrators /controls is reset after measurment directly improvements are planned in -> >V01-02
AE-000090 </=V01-02 Hint 10-Dez-2007 17-Apr-2008 cobas e 411 Calibration The sorting criteria of calibration causes are according to cobas 6000 specification only for information
AE-000084 V01-02 Issue 24-Okt-2007 17-Apr-2008 cobas e 411 QC Program error after control loading is solved in -> >V01-02
AE-000099 </=V01-02 Issue ### 17-Apr-2008 cobas e 411 Alarm Alarm 41-01-05 "CF has been accessed > 100.000 times" is solved in -> >V01-02
AE-000042 </=V01-02 Issue ### 17-Apr-2008 cobas e 411 QC Individual QC list and chart are displayed incorrectly in same cases is planned to be solved in -> >V01-02
AE-000087 </=V01-02 Issue 24-Okt-2007 17-Apr-2008 cobas e 411 QC Cumulative QC for CA 15-3 TM1and TM2 is solved in -> >V01-02
AE-000096 </=V01-02 Issue 11-Dez-2007 17-Apr-2008 cobas e 411 QC QC violation is indicated after a successful calibration is solved in -> >V01-02
AE-000097 </=V01-02 Issue 11-Dez-2007 17-Apr-2008 cobas e 411 Troubleshooting Restore function does not work for "system specific data" is solved in -> >V01-02
AE-000012 </=V01-02 Issue ### 16-Apr-2008 cobas e 411 Troubleshooting In case of a software trouble following informations are requested for investigations improvements are planned in -> >V01-02
AE-000018 </=V01-02 Issue ### 16-Apr-2008 cobas e 411 Operation Sample Tracking screen is deleted after pressing the Start button improvements are planned in -> >V01-02
AE-000104 V01-02 Issue ### ### cobas e 411 Host A typo error was noticed on the Communication Trace Screen in the software. is planned to be solved in -> V01-02
Record Version Section Created LastUpdate SysPart Category Title Status Status Ver
AE-000100 V01-01 Issue ### ### cobas e 411 QC Software crashed when trying to edit a new control in the QC menu is solved in -> >V01-02
AE-000022 </=V01-02 Issue ### ### cobas e 411 QC The control results with commented and/or excluded status are not displayed with different colors is solved in -> >V01-02
AE-000091 </=V01-02 Hint 11-Dez-2007 12-Dez-2007 cobas e 411 Configuration User Trainings Guide for V.01-01 is available now only for information
AE-000092 </=V01-02 Hint 11-Dez-2007 12-Dez-2007 cobas e 411 Configuration A new printer "LaserJet P2015" is recommended, refer to SB no.6 only for information
AE-000054 </=V01-02 Issue 03-Mai-2007 11-Dez-2007 cobas e 411 QC Non-Roche control values changes automatically when a new RP is placed on board is solved in -> >V01-02
AE-000063 </=V01-02 Issue 21-Mai-2007 11-Dez-2007 cobas e 411 QC Memory capacity of Routine View and QC View in Workplace/DataReview improvements are planned in -> >V01-02
AE-000028 </=V01-02 Issue ### 11-Dez-2007 cobas e 411 Alarm No yellow warning is displayed, if the amount of system reagent set 1 is empty is planned to be solved in -> >V01-02
AE-000052 </=V01-02 Issue ### 11-Dez-2007 cobas e 411 Configuration The button in the "Retry function setup screen" will be changed from "Start" to "Ok" improvements are planned in -> >V01-02
AE-000082 </=V01-02 Issue ### 11-Dez-2007 cobas e 411 Host Communication failure occurrs sporadically is planned to be solved in -> >V01-02
AE-000019 </=V01-02 Issue ### 11-Dez-2007 cobas e 411 Maintenance Maintenance button on the Workflow Guide does not change colour is planned to be solved in -> >V01-02
AE-000048 </=V01-02 Issue ### 11-Dez-2007 cobas e 411 Calibration A "change over" request is not displayed, if registration time of a reagent pack is done earlier is planned to be solved in -> >V01-02
AE-000050 </=V01-02 Hint ### 11-Dez-2007 cobas e 411 QC The 3 SD control ranges are displayed on the QC printout only no change planned
AE-000038 </=V01-02 Issue ### 11-Dez-2007 cobas e 411 Reagent The display of system reagent on the System Overview is display incorrectly improvements are planned in -> >V01-02
AE-000023 </=V01-02 Issue ### 11-Dez-2007 cobas e 411 Operation Indication for a STATsample ID is not implemented on the Disk system. improvements are planned in -> >V01-02
AE-000055 </=V01-02 Hint 04-Mai-2007 11-Dez-2007 cobas e 411 Configuration Application parameter report can only be done for all the tests only for information
AE-000021 </=V01-02 Issue ### 11-Dez-2007 cobas e 411 Operation The test keys colours "dark grey" is not changed after result is available improvements are planned in -> >V01-01
AE-000070 </=V01-02 Issue 03-Jul-2007 11-Dez-2007 cobas e 411 Configuration Incorrect handling of some Documentation and Printer Settings is solved in -> >V01-02
AE-000045 </=V01-02 Issue ### 11-Dez-2007 cobas e 411 Documentation Printout of "Worklist" is not implemented the for Rack system no change planned >V01-02
AE-000079 </=V01-02 Hint ### 11-Dez-2007 cobas e 411 QC QC after Calibration on cobas e411 only for information
AE-000069 </=V01-02 Issue 03-Jul-2007 11-Dez-2007 cobas e 411 Operation During operation it is not possible to change the Operator ID with "Logoff" button is planned to be solved in -> >V01-02
AE-000017 </=V01-02 Issue ### 11-Dez-2007 cobas e 411 Documentation Print View and Print function does not work correctly is planned to be solved in -> >V01-02
AE-000032 </=V01-02 Issue ### 11-Dez-2007 cobas e 411 Result It is not possible to print out the results with "Review by Exception" flags improvements are planned in -> >V01-02
AE-000033 </=V01-02 Issue ### 11-Dez-2007 cobas e 411 Result Reagent information on the Result Report is missing improvements are planned in -> >V01-02
AE-000030 </=V01-02 Issue ### 11-Dez-2007 cobas e 411 Host Cobas host protocol type is not available yet is planned to be solved in -> >V01-02
AE-000037 </=V01-02 Issue ### 11-Dez-2007 cobas e 411 QC Confidence limit field is activated even thogh QC Violation setting is OFF is planned to be solved in -> >V01-02
AE-000044 </=V01-02 Issue ### 11-Dez-2007 cobas e 411 Operation Some of the screens are not updated in "real-time" mode, even though new data are produced is solved in -> >V01-02
AE-000040 </=V01-02 Issue ### 11-Dez-2007 cobas e 411 QC The "Run status" screen shows control data with old and new "Unit" if the test unit is changed is planned to be solved in -> >V01-02
AE-000020 </=V01-02 Issue ### 11-Dez-2007 cobas e 411 Calibration Selection box "Auto Masking" under Utility/Application/Calib is missing improvements are planned in -> >V01-02
AE-000064 </=V01-02 Issue ### 11-Dez-2007 cobas e 411 Alarm Barcode setting for "Read Error Alarm" ON or OFF under Utility has no relevance is planned to be solved in -> >V01-02
AE-000047 </=V01-02 Issue ### 11-Dez-2007 cobas e 411 Configuration Difference in "Sample Reception Mode" settings and implementations improvements are planned in -> >V01-02
AE-000076 V01-01 Issue 24-Jul-2007 ### cobas e 411 Reagent The number of test is displayed to "0" determination, if a new reagent pack was scanned is solved in -> V01-02
AE-000061 Int. V 1.1.8 Hint 07-Mai-2007 ### cobas e 411 Troubleshooting After language file installation an error occurs on the installation report only for information
Record Version Section Created LastUpdate SysPart Category Title Status Status Ver
AE-000051 V01-01 Issue ### 24-Jul-2007 cobas e 411 Operation The input no. 6 from on the entire keyboard can not be done is solved in -> Int. V 1.1.8
AE-000014 V01-01 Hint ### 03-Jul-2007 cobas e 411 Alarm Alarm information can be deactivated only for information
AE-000009 V01-01 Hint ### ### cobas e 411 Configuration Test applications can not be deleted only for information
AE-000060 V01-01 Issue 07-Mai-2007 ### cobas e 411 Operation The Back up data under Workplace/Data Review does not work correctly is solved in -> Int. V 1.1.8
AE-000003 V01-01 Hint ### 07-Mai-2007 cobas e 411 Calibration Calibrator rack assignment automaticallly only for information
AE-000004 V01-01 Hint ### 07-Mai-2007 cobas e 411 Operation Used reagent packs should used on one system only for information
AE-000001 V01-01 Hint ### 07-Mai-2007 cobas e 411 Operation Start up time from Initialization to Operation status only for information
AE-000015 V01-01 Hint ### 07-Mai-2007 cobas e 411 Operation Restart is necessary if "P.Stop alarm" occurs only for information
AE-000010 V01-01 Hint ### 07-Mai-2007 cobas e 411 Printer Details to the default printer only for information
AE-000043 V01-01 Hint ### 07-Mai-2007 cobas e 411 Calibration Timeout cause for a calibration request is not specified on the cobas e 411 software versions only for information
AE-000002 V01-01 Hint ### 07-Mai-2007 cobas e 411 Reagent Reagent lid has to be open slightly only for information
AE-000046 V01-01 Hint ### 07-Mai-2007 cobas e 411 Troubleshooting To avoid any trouble do not open the "sipper safety cover" during the run only for information
AE-000016 V01-01 Hint ### 07-Mai-2007 cobas e 411 Calibration Only one calibrator lot can be used per quantitative test only for information
AE-000013 V01-01 Hint ### ### cobas e 411 Operation Used reagent packs should used on one system only