FLOWMETER
ACCURACY
W
hen specifying a flowmeter, it is import- FS flow meters are often used in applications that
ant to know the accuracy. The accuracy have a constant flow rate usually at the upper
of a flow meter can be stated one of two range of the flow meter. These meters can also
ways; as a percentage of full scale (FS) or as a be used in fluctuating flow applications where
percentage of reading (RD, also referred to as per- accuracy is less important. These meters are less
centage of rate). It is important to understand the desirable in low flow applications.
meaning of both so that your flow meter performs
as expected and with the desired accuracy. Flow meters that are sometimes
expressed in FS are:
Percentage of Full Scale (FS) Accuracy: • Variable area meters
A flow meter that has an accuracy expressed in • Differential pressure meters
FS has a fixed error band across the flow range of • Insertion impeller meters
the meter. For example, a flow meter with a 1% FS
• Rotameters
accuracy that has a range of 10GPM-100GPM has
an error band +/- 1% of 100GPM, which is equal
Percentage of Reading (RD) Accuracy:
to +/- 1 GPM. Therefore, since the accuracy of
the flow meter is +/- 1 GPM throughout the flow Most flow meter accuracies are expressed as
range, this type of meter is most accurate at full %RD. The accuracy is expressed as a percentage
scale, 100GPM, and becomes less accurate as of flowrate across the range of the flowmeter. The
the flow decreases. When the meter is measuring error band is not a fixed value as we saw with the
flow at 50GPM or 50% FS, the actual flow will FS accuracy. Instead, it is expressed as a percent-
be within +/- 1 GPM, or anywhere from 49GPM- age of rate. For example, a flow meter with a 1%
51GPM. When you do the math, this works out RD accuracy that has a range of 10GPM-100GPM
to be +/- 2% accuracy at 50GPM, 1/50 = 2%. At has an error band is +/- 1% at 100GPM, which is
the bottom end of the flow range, or 10GPM, the equal to +/- 1 GPM. At the lowest end of the flow
accuracy will be +/- 10%, 1/10=10%. range, the error band is +/- 1% of 10GPM, which is
equal to +/- 0.1 GPM. When graphically represent-
ed as a percentage of rate, the %RD is constant, in
this case always 1%.
• Turbine meters
7% Association of
1.0% Energy65Engineers.
6%
1.0% 60
5% 1.0% 55
Accuracy across the range of a 1% RD meter
4% 1.0% 50
3% 1.0% 45
2%
1.0% 40
1.0% 35
1%
1.0% 30
0%
1.0% 25
10 15 20 25 30 35 40 45 50 55 60 65 70 75 80 85 90 95 100
1.0% 20
Flow Rate in GPM
1.0% 15
1.0% 10
Conclusion
It is important to understand the difference be-
tween FS and RD accuracy. FS meters are most
accurate at full scale and the accuracy falls off
across dramatically at lower flow rates. RD meters on the
100 other hand maintain their accuracy throughout
100
the range of the flowmeter.
95
90
85
80
75
70
65
60
55
50
45
40
35 Wichita - Headquarters Sydney - Manufacturing & Sales Mexico - Sales Office
30 Great Plains Industries, Inc. Great Plains Industries Australia Great Plains Industries S De RL DE CV
25 5252 E. 36th St. N. 1/16 Atkinson Road Av. Fray Luis de Leon 8051
20 Wichita, KS 67220-3205 Taren Point, NSW, 2229, Australia Oficina 13 Centro Sur
15 316-686-7361 +61 2 9540 4433 Queretaro, Qro., Mexico 76090
10 +52 (442) 229-3139
Visit FLOMEC.net to access other support materials!
02/18 IND-1095A © 2018 Great Plains Industries, Inc. All Rights Reserved.