PQ Test Protocol MPA Sample Compartment 20231212 004609
PQ Test Protocol MPA Sample Compartment 20231212 004609
Maximum 100% Line Deviation: 0.5 Measured 100% Line Deviation: 0.28
Energy Test
________________
Page 1 of 5 Page Signature
OVP - PQ Test Protocol
Signal to Noise Test, Spectra
100% Line Test, Spectra
Energy Test, Spectra
101.0
99.5 100.0 100.5
Transmittance [%]
99.0
1 dd53f5e9-aafe-470c-879b-45fbd5c05f47
2 ab17d124-a98a-421c-b888-b95c3b7c396b
3 68ba46ba-9196-460f-b152-5fc18cb72c67
4 bbb231b6-6fe2-4261-a74c-5c6ecce016e2
5 05bcd98b-8b16-42d7-9f30-338c7a3f1eaf
________________
Page 2 of 5 Page Signature
OVP - PQ Test Protocol
Interferogram Peak Test, Spectrum
0.25
0.15
0.05
-0.15 -0.05
dd53f5e9-aafe-470c-879b-45fbd5c05f47
________________
Page 3 of 5 Page Signature
OVP - PQ Test Protocol
Wavenumber Accuracy Test BRM 2065 (37.0 Deg. C, Peak is T. corr.), Spectrum
10245.65
0.8
Absorbance Units
0.4 0.6
0.2
eb55cf71-6649-409e-8a86-88a447ca5225
________________
Page 4 of 5 Page Signature
OVP - PQ Test Protocol
Photometric Reproducibility Test - Glass Filter A, Spectrum
25
Transmittance [%]
10 15 20
5
86767c5b-7bab-463b-a69c-262249c83929
________________
Page 5 of 5 Page Signature