Professional Documents
Culture Documents
Publications
5815720-1EN
Revision 1
Proteus XR/f
do not duplicate
WARNING
ENVIRONMENTAL STATEMENT
ON THE LIFE CYCLE OF THE EQUIPMENT OR SYSTEM
This equipment or system contains environmentally dangerous components and materials (such as PCB‘s,
electronic components, used dielectric oil, lead, batteries etc.) which, once the life-cycle of the equipment or
system comes to an end, becomes dangerous and need to be considered as harmful waste according to the
international, domestic and local regulations.
The manufacturer recommends to contact an authorized representative of the manufacturer or an authorized
waste management company once the life-cycle of the equipment or system comes to an end to remove this
equipment or system.
GE Healthcare System Alignment Steering guide for Proteus XR/f
REV 1 5815720-1EN
REVISION HISTORY
This Document is the English original version, edited and supplied by the manufacturer.
The Revision state of this Document is indicated in the code number shown at the bottom of this page.
3
GE Healthcare System Alignment Steering guide for Proteus XR/f
REV 1 5815720-1EN
ADVISORY SYMBOLS
4
GE Healthcare System Alignment Steering guide for Proteus XR/f
REV 1 5815720-1EN
TABLE OF CONTENTS
Section Page
1.1 Step 1: Alignment of Focal spot with Tube Stand, tube and collimator Beam . 9
1.1.1 XR/f ET and AT Tube Stand with Toshiba E7254x . . . . . . . . . . . . . . . . 13
1.1.2 XR/f ET and AT Tube Stand with Toshiba E7884x . . . . . . . . . . . . . . . . 13
1.1.3 XR/f ST Tube Stand with Toshiba E7254x . . . . . . . . . . . . . . . . . . . . . . . 14
1.1.4 XR/f ST Tube Stand with Toshiba E7884x . . . . . . . . . . . . . . . . . . . . . . . 14
1.1.5 Install the Collimator . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 15
1.2 Step 2: Supply the Collimator . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 18
1.2.1 AT and ET Systems . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 18
1.2.2 ST Systems . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 19
1.3 Step 3: wall stand alignment with tube focal spot (using the Collimator Beam) 21
1.4 Step 4: Table alignment with tube focal spot (using the Collimator Beam ) . . 22
5
GE Healthcare System Alignment Steering guide for Proteus XR/f
REV 1 5815720-1EN
6
GE Healthcare System Alignment Steering guide for Proteus XR/f
REV 1 5815720-1EN
This Alignment Steering Guide applies to the Proteus XR/f ET-AT-ST RAD
Room during installation.
Make sure that the Column base is properly anchored and leveled, the
maximum tolerance for leveling is 1.5 mm per meter (0.2 in per 10 feet) as
per the site pre-installation room layout drawing.
Note . Keep in mind that the Tube Stand will be reference from where the
Wall Stand first and then Table will be aligned to.
The following items are factory staged and by design auto aligned, so they do
not need alignment at installation:
1. Alignment of focal spot with Tube Stand, X-ray Tube and Collimator laser
beam (Tube and Collimator Installation).
7
GE Healthcare System Alignment Steering guide for Proteus XR/f
REV 1 5815720-1EN
Illustration 1
Alignment sequence
1
2
3 4
8
GE Healthcare System Alignment Steering guide for Proteus XR/f
REV 1 5815720-1EN
1.1 STEP 1: ALIGNMENT OF FOCAL SPOT WITH TUBE STAND, X-RAY TUBE
AND COLLIMATOR LASER BEAM.
b. the other one removes the shiping safety screws at both sides
9
GE Healthcare System Alignment Steering guide for Proteus XR/f
REV 1 5815720-1EN
X-Ray Tube
Tube-Collimator Support
Spacers
Spacers
Collimator Adaptation
Ring
10
GE Healthcare System Alignment Steering guide for Proteus XR/f
REV 1 5815720-1EN
Illustration 2
Alignment of X-RAY TUBE -- Holes in the Tube Collimator Support and Arm screws/Leveling Rods
7 7
6 6
Tube-Collimator Support
Leveling Rods
Safety Screws
in slotted hole
Leveling Rods
11
GE Healthcare System Alignment Steering guide for Proteus XR/f
REV 1 5815720-1EN
Illustration 3
Alignment of X-RAY TUBE -- Factory Pre-stage alignment marks
The Leveling Rods are factory set and locked with Blue loctite
for each system so it is not necessary to readjust them.
Tube-Collimator Leveling Rods mark
Support mark
Tube-Collimator
Support mark
12
GE Healthcare System Alignment Steering guide for Proteus XR/f
REV 1 5815720-1EN
Position the screws in the 1st and 5th hole of the Tube-Collimator Support and
tighten the 4 Screws M6 (10 Nm) in the slotted holes as indicated below.
Illustration 4
AT & ET with E7254
5
Screw at lowest position
of the slotted hole
Position the screws in the 1st and 4th hole of the Tube-Collimator Support and
tighten the 4 Screws M6 (10 Nm) in the slotted holes as indicated below.
Illustration 5
AT & ET with E7884
13
GE Healthcare System Alignment Steering guide for Proteus XR/f
REV 1 5815720-1EN
Position the screws in the 2nd and 7th hole of the Tube-Collimator Support and
tighten the 4 Screws M6 (10 Nm) in the holes as indicated below.
Illustration 6
ST with E7254X
Position the screws in the 1st and 6th hole of the Tube-Collimator Support and
tighten the 4 Screws M6 (10 Nm) in the holes as indicated below.
Illustration 7
ST with E7884
14
GE Healthcare System Alignment Steering guide for Proteus XR/f
REV 1 5815720-1EN
Illustration 8
Collimator Installation
CAUTION: Two people are required to install the Collimator in the Tube Support.
One person holds the Collimator while the other one tightens the Collimator
screws.
CAUTION: It is not recommended to turn the Tube-Collimator Assembly 180
degrees up side down to perform the Collimator installation because it tends to
turn counterclockwise and the collimator may fall down on the table.
15
GE Healthcare System Alignment Steering guide for Proteus XR/f
REV 1 5815720-1EN
Illustration 8 (continued)
Collimator Installation
While one person holds the collimator with its cone inside
the Collimator Adaptation Ring, the other one tightens the
screws with an Allen 3 mm wrench.
16
GE Healthcare System Alignment Steering guide for Proteus XR/f
REV 1 5815720-1EN
Illustration 8 (continued)
Collimator Installation
Allen wrench 3 mm
17
GE Healthcare System Alignment Steering guide for Proteus XR/f
REV 1 5815720-1EN
Finally, connect the Collimator cables in the Collimator side as shown below.
Collimator cabling
1. Only for installation purposes, install a plug (not supplied -- see image
below) in the Table supply cable and connect the Table Supply Cable:
A6779--01 for ET or A7379--01 for AT Systems to a wall plug (Table can
also be connected to the Room Electrical Cabinet). The Table
Transformer is factory set to destination voltage.
US -- Canada plug
-- not supplied
Europe Plug
-- not supplied
2. Turn the Table red Emergency Switch on (the Circuit breaker at the table
base Table should be on).
18
GE Healthcare System Alignment Steering guide for Proteus XR/f
REV 1 5815720-1EN
3. Double tap on the Up table pedal and raise the table to the top.
1.2.2 ST SYSTEMS
For Systems with Non Telescopic Arm, connect the A3111-04 Collimator cable
to the Lock Board TB7 at Generator.
19
GE Healthcare System Alignment Steering guide for Proteus XR/f
REV 1 5815720-1EN
LOCKS CABLES: Connect wires from the Locks to Terminal Block TB7-5 (+24
VDC) and TB7-6 (0 VDC) of the Lock Board.
Lock Board
Generator
For Systems with Telescopic Arm, connect the A3111-04 cable to TB7 at the
Brake Box and connect the Brake Box to the Room Electrical Cabinet. Refer to
Schematics A8715--01 at the end of the Positioner Service manual
5763452-1EN.
COLLIMATOR CABLE: Connect the wires from the Collimator Lamp to the
Terminal Block TB7-3 (24 VAC) and TB7-4 (0 VAC).
LOCKS CABLES: Connect the wires from the Locks to the Terminal Block
TB7-5/7 (+24 VDC) and TB7-6/8 (0 VDC).
TB7
1 +24VDC
2 0VDC
3 24VAC
4 0VAC
5 +24VDC
6 0VDC
7 +24VDC
8 0VDC
Brake Box
Remove LOTO.
20
GE Healthcare System Alignment Steering guide for Proteus XR/f
REV 1 5815720-1EN
1.3 STEP 3: WALL STAND ALIGNMENT WITH TUBE FOCAL SPOT (USING THE
COLLIMATOR LASER BEAM AND LIGHT)
With the Telescopic Arm of the Tube-Collimator locked in the center of the Arm
movement, point the laser beam to the Wall Stand cover grid, move and level
the Wall Stand base plate so that the laser beam stay in the middle of the Wall
Stand spot for minimum and maximum SID and for minimum and maximum
Wall Stand/Tube Stand heights.
If necessary, level the Wall Stand Column with leveling plates (included in the
Package).
Illustration 9
Wall Stand Alignment
Laser Line
Laser Line
21
GE Healthcare System Alignment Steering guide for Proteus XR/f
REV 1 5815720-1EN
1.4 STEP 4: TABLE ALIGNMENT WITH TUBE FOCAL SPOT (USING THE
COLLIMATOR LASER BEAM AND LIGHT)
1. Turn on the collimator light, align with the bucky handle mark with the
laser beam as shown below.
2. Move the Tabletop frame to the front end (toward the Tube Stand Column
end) and at the same time to the right so the Tabletop frame can be seen
and the centering label is shown on left side of the table chassis (see
image below).
22
GE Healthcare System Alignment Steering guide for Proteus XR/f
REV 1 5815720-1EN
3. Turn on the Collimator light and move the Tube Collimator to the left until
the Collimator light reaches the end of the Table frame.
4. Carefully move the table to Align the label and the Collimator Axis as
shown below. Repeat the procedure at the right side.
Illustration 10
Table lateral alignment adjustment - left and right sides
(Images correspond to XR/f AT or ET systems, XR/f ST system not shown)
Axis of the
Collimator light
Axis of the
Collimator light
Label
Label
23
GE Healthcare System Alignment Steering guide for Proteus XR/f
REV 1 5815720-1EN
The distance from the ST Table chassis to the label center is 261 mm
(10.3”).
261 mm
(10.3 “)
If necessary, level the Table with leveling plates (included in the Package).
Note . Final alignment check will be done with X-ray and acquired images
by FE but no further adjustment would have to be done if above
steps are done properly.
24
GE Healthcare System Alignment Steering guide for Proteus XR/f
REV 1 5815720-1EN
Electrical X System’s Main “A1” Power Disconnect 208 -- 240 VAC, 1 Phase
380 -- 480 VAC, 3 Phase
Pneumatic X
Hydraulic X
Gas/Water/Steam X
Chemical X
Mechanical Motion X
Gravity X
Springs X
Thermal X
Stored Energy X
Air Under Pressure X
Oil Under Pressure X
Water Under Pressure X
Gas Under Pressure X
Steam X
Other X
25
GE Healthcare System Alignment Steering guide for Proteus XR/f
REV 1 5815720-1EN
3. Locate the Main Power Supply (Service) Disconnect (A1) that feeds the
Generator cabinet. Turn off the mains supply Circuit Breaker.
4. Lock Out / Tag Out the Circuit Breaker of the Main Power Supply
Disconnect (A1).
GND Studs
L1 Input Line
L2 Input Line
L3 Input Line
Neutral Stud
(optional)
26
GE Healthcare System Alignment Steering guide for Proteus XR/f
REV 1 5815720-1EN
LED DS1
CHARGE/DISCHARGE
MONITOR BOARDS
d. Verify LOTO by ensuring that the system can not switch on. Then
operate the switches on the equipment sub-modules to confirm
there is not connection. Return the switches/breakers to the OFF
position.
6. After service has been performed, notify the affected personnel that the
equipment will be re-energized and LOTO devices are being removed.
7. Assure the safety of the area and that all safety devices and guards have
been replaced.
10. Locate the Main Power Service Disconnect feeding the system cabinet
and re--apply the power.
11. Locate the Main Power Supply (Service) Disconnect (A1) that feeds the
Generator cabinet and reapply the power by turning on the mains supply
Circuit Breaker.
27
GE Healthcare System Alignment Steering guide for Proteus XR/f
REV 1 5815720-1EN
Follow the steps below to ensure that your system powers up safely, properly,
and reliably.
28
GE Healthcare System Alignment Steering guide for Proteus XR/f
REV 1 5815720-1EN
The System should be powered by the same Room Electrical Cabinet where
the X-ray Generator is connected, that is, the whole System will be powered
from the same Electrical Cabinet.
29
GE Healthcare System Alignment Steering guide for Proteus XR/f
REV 1 5815720-1EN
2. Press the X-Ray Generator Control ON button, the Green Lamp turns
ON.
Illustration 3-1
Start Up sequence
1
Room
Electrical
Cabinet
Workstation
3
ON
30
GE Healthcare System Alignment Steering guide for Proteus XR/f
REV 1 5815720-1EN
2. Turn OFF the Room Electrical Cabinet Switch (with the Emergency
Switch in OFF position).
Note . It is not necessary to turn OFF the Circuit Breaker before turning
off the Room Electrical Cabinet Switch.
Illustration 3-2
Shutdown sequence
1 2
Workstation Room
Electrical
Cabinet
31
GE Healthcare System Alignment Steering guide for Proteus XR/f
REV 1 5815720-1EN
A sudden discharge of static electricity from your finger or other conductor can
destroy static-sensitive devices or microcircuits. Often the spark is neither felt
nor heard, but damage occurs. An electronic device exposed to electrostatic
discharge (ESD) might not appear to be affected at all and can work perfectly
throughout a normal cycle. The device can function normally for a while, but it
has been degraded in the internal layers, reducing its life expectancy.
Networks built into many integrated circuits provide some protection, but in
many cases the discharge contains enough power to alter device parameters
or melt silicon junctions.
Relative Humidity
32
GE Healthcare System Alignment Steering guide for Proteus XR/f
REV 1 5815720-1EN
Heel straps, toe straps, and boot straps can be used at standing
workstations and are compatible with most types of shoes or boots. On
conductive floors or dissipating floor mats, use them on both feet with a
maximum of one--megohm (±10%) resistance between the operator and
ground.
Method Voltage
33
GE Healthcare System Alignment Steering guide for Proteus XR/f
REV 1 5815720-1EN
Turn off the power and input signals before inserting and removing
connectors or test equipment.
Materials and equipment that are recommended for use in preventing static
electricity include:
34