sm 2401594--100
Revision 2
Proteus XRe
sm
Service Manual
do not duplicate
ENVIRONMENTAL STATEMENT
ON THE LIFE CYCLE OF THE EQUIPMENT OR SYSTEM
This equipment or system contains environmentally dangerous components and materials (such as PCB‘s,
electronic components, used dielectric oil, lead, batteries etc.) which, once the life-cycle of the equipment or
system comes to an end, becomes dangerous and need to be considered as harmful waste according to the
international, domestic and local regulations.
The manufacturer recommends to contact an authorized representative of the manufacturer or an authorized
waste management company once the life-cycle of the equipment or system comes to an end to remove this
equipment or system.
REVISION HISTORY
This Document is the English original version, edited and supplied by the manufacturer.
ADVISORY SYMBOLS
3
GE Medical Systems Proteus XR/e
REV 2 sm 2401594--100
SAFETY SYMBOLS
Ionizing radiation.
Type B equipment.
Dangerous voltage.
Ground.
4
GE Medical Systems Proteus XR/e
REV 2 sm 2401594--100
TABLE OF CONTENTS
Section Page
1 PRE-INSTALLATION CHECKS . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7
2 UNPACKING . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 15
3 INSTALLATION . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 17
3.1 Tools . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 17
1
GE Medical Systems Proteus XR/e
REV 2 sm 2401594--100
Section Page
5 ADJUSTMENTS . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 31
6 PERIODIC MAINTENANCE . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 47
2
GE Medical Systems Proteus XR/e
REV 2 sm 2401594--100
• SLIDING COLUMN
• RADIOGRAPHIC TABLE
7
GE Medical Systems Proteus XR/e
REV 2 sm 2401594--100
• SLIDING COLUMN
Height . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2318 mm
Width . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1090 mm
Length . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2000 mm
Length with additional rail . . . . . . . . . . . . . . . 3000 mm
Weight . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 224 kg
• RADIOGRAPHIC TABLE
Height . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 775 mm
Width . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2000 mm
Length . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 725 mm
Weight . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 80 kg
8
GE Medical Systems Proteus XR/e
REV 2 sm 2401594--100
Height . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2235 mm
Width . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 622 mm
Length . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 345 mm
Weight . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 145 kg
9
GE Medical Systems Proteus XR/e
REV 2 sm 2401594--100
Illustration 1
Dimensions for Installation (Non telescopic Arm)
100
1000 Max.
1231
800
10
GE Medical Systems Proteus XR/e
REV 2 sm 2401594--100
Illustration 2
Dimensions for Installation (Telescopic Arm)
100
1425
800
11
GE Medical Systems Proteus XR/e
REV 2 sm 2401594--100
Illustration 3
Drilling Template
(1215)
(885)
(942)
12
GE Medical Systems Proteus XR/e
REV 2 sm 2401594--100
Illustration 4
Cabling and Room Raceways suggestions
For 9 m High Voltage Cables, install a cable hook on the wall and
position the generator as close as possible to the Column base
Cable Hook
Generator Upper
Cable Entrance
GENERATOR
CABINET
13
GE Medical Systems Proteus XR/e
REV 2 sm 2401594--100
Illustration 4 (cont.)
Cabling and Room Raceways suggestions
GENERATOR
CABINET
Table Base
To Console
Surface Cable Raceways
100
GENERATOR
CABINET
Table Base
To Console
Cable Entrance
(All the Table Base Area can be used for Cable Entrance except when the Generator is installed undertable)
Underfloor Cable Raceways
100
14
GE Medical Systems Proteus XR/e
REV 2 sm 2401594--100
SECTION 2 UNPACKING
1. Place crates close to its final location at room and remove each wood
package part. Do not discard any packing material (envelopes, boxes,
bags) until all parts are identified and listed in the packing list.
15
GE Medical Systems Proteus XR/e
REV 2 sm 2401594--100
16
GE Medical Systems Proteus XR/e
REV 2 sm 2401594--100
SECTION 3 INSTALLATION
3.1 TOOLS
• The minimum distance from the Column Base to its parallel wall is
variable depending on the room size.
17
GE Medical Systems Proteus XR/e
REV 2 sm 2401594--100
1. Position the Column Base on the floor according to the RAD Room plan
distribution. Keep in mind the dimensions and final position of each
component in the System (Column, Wall Bucky, Table, etc.).
The Column Base has to be placed with the Brake Track at the rear side
and perpendicular to the wall where the Vertical Bucky will be installed.
2. Mark the position of the ten (10) anchoring holes on the floor. Move the
Base and drill the anchoring holes. Bolts and hardware are field
provided.
3. Check the leveling in different points on the Base and adjust it by using
leveling plates. Anchor slightly the Base to the floor.
4. Remove the Rail Stop from one of the Base laterals. Rail Stops can be
installed in the front or in the rear rails.
Illustration 5
Sliding Column Base
BRAKE TRACK
ANCHORING HOLES
BUCKY - COLUMN BASE LINK
(OPTIONAL)
COLUMN CARRIAGE
5. With the Column laid on the floor, remove the Upper Cap and Upper Front
Cover (both are secured with two screws). (Refer to Illustration 6).
7. In order to ease the Column installation inside the Base, remove or loose
the Column Foot Brakes located at the rear side of the Column Foot.
(Refer to Illustration 5).
18
GE Medical Systems Proteus XR/e
REV 2 sm 2401594--100
Make sure that the iron cables are properly routed around the
pulleys before standing up the Column.
9. Re-install the Rail Stop in the lateral of the Base and the Column Foot
Brakes.
10. Level the Column vertically on both laterals and front side. Adjust it by
means of the leveling plates in the Column Base. Anchor tightly the Base
to the floor. (Refer to Illustration 6).
Illustration 6
Sliding Column Base
UPPER CAP
LEVELS
VERTICAL CARRIAGE
COLUMN FOOT
BASE
ANCHORING HOLES
19
GE Medical Systems Proteus XR/e
REV 2 sm 2401594--100
11. Install the Tube-Collimator Support to the Column Arm (Coupling Plate)
by using the four screws mounted at the rear side of the Support (upper
and lower holes). The two rods will be used to level the Support when all
the components are assembled. Pass both cables (J1 and J2) in the Arm
through the window of the Support.
Illustration 7
Coupling of Tube-Collimator Support -- Rear View
COUPLING SCREWS
12. Install the Column Control Panel by using the two bolts in the rear side
of the Control Panel.
13. Carefully, remove the Collimator Adaptation Ring from the Collimator.
Avoid the ring screws to fall down into the Collimator.
14. Install the X-ray Tube on its Support using the Collimator Adaptation Ring
and its four Security Screws (Allen).
Illustration 8
X-ray Tube, Collimator and Control Panel
COLLIMATOR
20
GE Medical Systems Proteus XR/e
REV 2 sm 2401594--100
15. Before installing the Collimator, unscrew the same number of turns the
four Centering Adjustment and Security Screws (Allen) to allow the
Collimator installation in the Collimator Adaptation Ring. Adjust the
Collimator Blades to their widest setting and carefully install the
Collimator centering it with the X-ray Tube window.
16. Level horizontally the Tube-Collimator Assembly on the Tube and the
Collimator base. Adjust the level by using the two rods in the rear side
of the Coupling Plate (Arm).
17. Remove the Counterweight Safety Bar (upper front of Column) and the
two Shipping Safety screws close to the Vertical Carriage. The weight
of the Counterweight is equal or slightly greater than the weight of the
Carriage + Arm + Tube + Collimator + Control Panel. It is factory
adjusted.
18. Install the Upper Front Cover and then the Upper Cap (both are secured
with two screws). Install the Foot Cover of the Column pushing it.
Illustration 9
Tube Support Arm Leveling
UPPER CAP
LEVELING RODS
21
GE Medical Systems Proteus XR/e
REV 2 sm 2401594--100
1. Remove the transport screws that secure the Radiographic Table to the
pallet and move away the Table.
2. Remove the front cover screws of the Table in order to access inside the
table.
3. Power ON the Collimator Lamp and align perfectly the Bucky axes with
the Collimator light beam. During alignment, place the Tube-Collimator
Assembly and the Bucky Carriage at different positions to check that it
is correctly aligned.
4. Mark the four anchoring holes on the floor, shift the table and perform the
holes (diam 9).
ANCHORING HOLES
6. Connect a line from Generator TS1 +24VDC, --24VDC and Ground to the
Table Terminal as per schematics 54302065.
7. Connect the Table Bucky Cables as per Bucky schematics. The Bucky
ground cable (yellow/green) must be connected to the Generator ground
point.
8. Optional -- Install the Bucky link part to the rear of the Bucky assembly
(6 allen screws).
ALLEN SCREWS
9. Install the four Table-top bumpers provided at four corners of the table.
22
GE Medical Systems Proteus XR/e
REV 2 sm 2401594--100
10. Install Table-Top, remove first the allen screws located at both sides for
the Table-Top carriage and slide the assembly, then reinstall the removed
screws.
11. Guide cables through table rear outlet and reinstall front cover.
Illustration 10
Wall Bucky Stand
VERTICAL CARRIAGE
BASE
ANCHORING HOLES
23
GE Medical Systems Proteus XR/e
REV 2 sm 2401594--100
3. Decide wether installing the wall anchors with the supplied Small
Spacers, Wall Support or nothing to separate the Wall Bucky Stand from
the wall.
4. At least, mark in the wall two of the upper anchoring holes of the Column
and mark the two anchoring holes on the floor.
Illustration 11
Wall Bucky Stand -- Anchoring Holes and Dimensions
24
GE Medical Systems Proteus XR/e
REV 2 sm 2401594--100
6. Position the Column and secure it to the floor and wall. Check that it is
properly leveled in both laterals and front side. Use the leveling plates
provided to raise and level the Column previous to secure it definitively.
(Refer to Illustration 10).
Note . In case that a Spacer or Wall Support are used, separate the
Column from the wall and assemble the Spacer or Wall Support
at the upper part of the column before marking the anchoring holes
in the wall.
7. Install the Upper Front Cover and then the Upper Cap (both are secured
with two screws). Install the Foot Cover of the Column pushing it.
8. Route the cables along the room to the Generator for their later
connection.
1. Remove the Table front cover. For that unscrew the four cover screws
located at upper and lower front cover.
2. Disconnect ground cable from front cover and take out the front cover.
25
GE Medical Systems Proteus XR/e
REV 2 sm 2401594--100
MICROSWITCH SW1
GENERATOR CARRIAGE
Note . Do not slide the Generator under the table until Configuration and
Calibration procedures are completed. Refer to Generator service
Manual.
26
GE Medical Systems Proteus XR/e
REV 2 sm 2401594--100
1. Connect the Lock cables from the Vertical Carriage (J1-J1) and from the
Column Control Panel (J2-J2) located close to the Tube-Collimator
Support (refer to Interconnection Map IM 341).
2. Connect the Collimator Lamp (if it has not been connected before).
3. Route all cables from the Collimator and X-ray Tube along to the Arm and
secure with the metallic Clamp over the Arm.
CABLES CLAMP
1. Route and connect the Locks cables from the Table to the Terminal Block
TB7 of the Locks Board (on the Top of the Generator).
2. Cable from the Bucky is connected directly to the Terminal Block 3TS1
of the Generator Cabinet. (Refer to Interconnection Map IM 341).
27
GE Medical Systems Proteus XR/e
REV 2 sm 2401594--100
Cables from the Bucky are connected directly to the Terminal Block 3TS1 of the
Generator Cabinet. (Refer to IM 341 ).
28
GE Medical Systems Proteus XR/e
REV 2 sm 2401594--100
Install the Cable Support on the upper right/left side of the Column. Route and
Fasten all the cables to it by using strain reliefs.
TOP OF COLUMN
Position the Sliding Column with the Tube-Collimator Assembly opposite the
Wall Bucky. Measure the SID with the Collimator Metric Tape and determine
the position where the Column has to stop when it reaches the required SID (1
m, 1.8 m, 2 m, etc.) respect to the Wall Bucky. Drill the Base and install the
Centering Plates where the Column has to stop. Adjust the Longitudinal Detent.
on the Column Foot.
Illustration 12
Centering Plate in Base
COLUMN FOOT
CAP OF DETENT
COLUMN BASE
CENTERING PLATE
LONGITUDINAL DETENT
29
GE Medical Systems Proteus XR/e
REV 2 sm 2401594--100
Note . Before re-installing all cabinet covers, perform the rest of the
Calibration procedures if required.
Check that all electrical connections are firm and secure, and all cables are
correctly routed. Remove the ferrite blocks factory clamped to the cables when
it is required to carry out a correct routing, then re-install again the ferrites blocks
at the same place around cables. (Refer to Section 3.7)
YES NO
Re-install the Cabinet covers and connect its internal ground wires. Power line,
High Voltage and Interconnections cables must go through the cover cable
outlet.
30
GE Medical Systems Proteus XR/e
REV 2 sm 2401594--100
SECTION 5 ADJUSTMENTS
The following special Tools have been used for adjustment of the Proteus XR/e
Plus X-ray System (these tools are not included with the System):
2. Turn on Collimator light and center the Collimator in relation to the Bucky.
Horizontal and transverse position of the light axes projected by the
Collimator Lamp must be in line with the axes of the horizontal Bucky.
Illustration 13
Collimator Screws
CENTERING ADJUSTMENT AND SAFETY SCREWS
31
GE Medical Systems Proteus XR/e
REV 2 sm 2401594--100
3. Position the Collimator Test Tool (RMI model 161B) on the Table-Top.
5. Adjust the Light Field of the Collimator Lamp to the rectangle drawn
inside the Collimator Test Tool.
6. Place centered the Beam Alignment Test Tool (RMI model 162A) on the
Collimator Test Tool .
Illustration 14
Alignment Test Tools
7. Load Cassette film Tray with cassette film 24x30 and insert it.
a. Check that the X-ray Field falls just within the image of the inner
rectangle of the Collimator Test Tool.
If an edge of the X-ray Field falls out of the inner rectangle means
a misalignment of the Light Field respect to the X-ray Field. The
maximum misalignment allowed is 2 % of SID (for SID 1m = 2 cm
tolerance).
Refer to Section 5.2.1 for alignment of Light Field with X-ray Field.
32
GE Medical Systems Proteus XR/e
REV 2 sm 2401594--100
-- If the image of the two balls overlap (A) the X-ray Beam is
perpendicular to within 0.5o.
-- If the image of the top ball (larger shadow) intercepts the first
circle (B), the X-ray Beam is about 1.5o away from the
perpendicular.
-- If the image of the top ball (larger shadow) intercepts the second
circle (C), the X-ray Beam is about 3o away from the perpendicular.
A B C
c. Check that the X-ray Beam is properly centered with the Image
Receptor. To determine the center of the Image Receptor, draw
diagonal lines from corner to corner of the X-ray Film. (Alternately,
the film can be folded in half and creased at the center). The two
lines will cross in the center of the Image Receptor (film). Then
draw diagonal lines from the corners of the imaged X-ray Field.
If the center of the X-ray Field and Image Receptor is the same,
the diagonals of both sets of lines should cross at the same point.
The maximum misalignment allowed is 2 % of SID (for SID 1m =
2 cm tolerance).
33
GE Medical Systems Proteus XR/e
REV 2 sm 2401594--100
3. Center the Collimator Test Tool (RMI model 161B) on the Table-Top with
the light axes projected by the Collimator Lamp.
4. Adjust the Light Field of the Collimator Lamp to the rectangle drawn
inside the Collimator Test Tool.
34
GE Medical Systems Proteus XR/e
REV 2 sm 2401594--100
5. Check on the processed film the adjustment required to meet the Light
Field with the X-ray Field. Identify the deviation on the axes imaged on
the film.
PROCESSED FILM
PROCESSED FILM
35
GE Medical Systems Proteus XR/e
REV 2 sm 2401594--100
6. Do not remove the Collimator Test Tool from its original position and
adjust the Light Field by moving the Collimator Unit and/or the Collimator
Lamp.
Illustration 15
Collimator Screws
36
GE Medical Systems Proteus XR/e
REV 2 sm 2401594--100
In case that perpendicularity is out of tolerance (top ball is out of first circle),
adjust perpendicularity as follows:
3. Position the Collimator Test Tool (RMI model 161B) on the Table-Top.
5. Place centered the Beam Alignment Test Tool (RMI model 162A) on the
Collimator Test Tool (fix the Test Tool with adhesive or another product).
Turn on Collimator light and observe if shadow of the Beam Alignment
Test Tool is projected in equal proportion around it.
6. Check on the processed film the adjustment required to center the top
ball mark. Shadow around the Beam Alignment Test Tool can also help
to make a first correction.
7. Loosen slightly the four Safety Screws (M6) at the rear side of the
Coupling Plate (Tube-Collimator Assembly). Then loosen or tighten the
two rods (Allen M5) to adjust horizontally the Tube-Collimator Assembly
with reference to the Horizontal Bucky. (Refer to Illustration 16).
8. Repeat exposure and procedure until the result is satisfactory (top ball
must be inside of the first circle).
37
GE Medical Systems Proteus XR/e
REV 2 sm 2401594--100
Illustration 16
Safety Screws and Rods for Perpendicularity Adjustment
COUPLING PLATE
SAFETY SCREWS
38
GE Medical Systems Proteus XR/e
REV 2 sm 2401594--100
The error between centers of the X-ray Field and the Image Receptor should
not be greater than 2% of the SID. (for SID 1m = 2 cm tolerance).
DEVIATION
1. Check on the processed film the correction required for centering the
X-ray Field with the Image Receptor. Adjustments will be performed as
the same way described for Perpendicularity correction, so only re-adjust
it if really is necessary.
39
GE Medical Systems Proteus XR/e
REV 2 sm 2401594--100
3. Turn on Collimator light and center the Collimator in relation to the Bucky
Table--Top. Horizontal and transverse position of the light axes projected
by the Collimator Lamp must be in line with the axes or film sizes marked
on the Table--Top. Horizontal and vertical position of the light axes
projected by the Collimator Lamp must be in line with the horizontal and
vertical axes of the Table--Top. (Alignment can also be performed by
placing a mirror in the center of the Table--Top and checking the
Collimator light reflection). If needed, readjust position of the Wall Bucky
Stand.
4. Load Cassette film Tray with cassette film 24x30 and insert it.
6. Process film and check that the X--ray Beam is properly centered with the
Image Receptor. To determine the center of the Image Receptor, draw
diagonal lines from corner to corner of the X--ray film. (Alternately, the film
can be folded in half and creased at the center). The two lines will cross
in the center of the Image Receptor (film). Then draw diagonal lines from
the corners of the imaged X--ray Field.If the center of the X--ray Field
and Image Receptor is the same, the diagonals of both sets of lines
should cross at the same point. The maximum misalignment allowed is
2 % of SID (for SID 1m = 2 cm tolerance).
40
GE Medical Systems Proteus XR/e
REV 2 sm 2401594--100
Note . Before starting with the Field Size Indicator Test, the Alignment of
X-Ray Beam Test and the Alignment of Light Field with X-Ray Field
Test should be performed.
4. Turn on the Collimator Light and center the Collimator in relation to the
Horizontal Bucky. Horizontal and Transferral position of the light axes
projected by the Collimator Lamp must be in line with the axes of the
Horizontal Bucky.
Illustration 17
X-ray Field Size
FILM 35x43 cm
WIDTH DIMENSION (W)
PROCESSED FILM
41
GE Medical Systems Proteus XR/e
REV 2 sm 2401594--100
9. The difference between the indicated Field Size and the obtained Field
Size may not exceed 1.5 % (rejection limit) of the SID in either direction.
Therefore the deviation in any direction should not be higher than 15 mm
(rejection limit) for SID at 1 meter.
4. Turn on the Collimator Light and center the Collimator in relation to the
Horizontal Bucky. Horizontal and transverse position of the light axes
projected by the Collimator Lamp must be in line with the axes of
Horizontal Bucky.
5. Place the Light Meter on the Table-Top with sensor directed toward the
Light Source.
42
GE Medical Systems Proteus XR/e
REV 2 sm 2401594--100
6. Place the Light Meter in the center of one quadrant of the light field (refer
to Illustration 18).
Illustration 18
Light Meter on Table-Top
4th QUADRANT
3rd QUADRANT
LIGHT FIELD
LIGHT METER
7. With Collimator Lamp OFF, measure and keep record of the ambient light
level.
8. Do not move the Light Meter. With the Collimator Lamp ON, measure and
keep record of light level of that quadrant.
11. With the measurements obtained on the step-10., the results of Light
Field Intensity in all quadrants should be higher of 170 lux
(16 foot-candles) and among all the quadrants the intensity light field
shall be not differ more than 10%.
12. The average illumination at a distance of 1 meter from the focal spot shall
not be less than 160 lux (15 foot-candles). Rejection limit is 170 lux
(16 foot-candles).
43
GE Medical Systems Proteus XR/e
REV 2 sm 2401594--100
b. If the light level in all quadrants is less than 170 lux, it is necessary
to check the following:
G Check that the Collimator Lamp, the Mirror and the Mylar
window are not dirty or discolored.
3. Open the Collimator Blades to set a Field Size of 25 x 25 cm for SID 1m.
4. Turn ON the Collimator Light and center the Collimator in relation to the
Horizontal Bucky. Horizontal and transverse position of the light axes
projected by the Collimator Lamp must be in line with the axes of
Horizontal Bucky.
5. Place the Light Meter on the Table-Top with sensor directed toward Light
Source.
6. Place the Light Meter in the center of one quadrant of the light field. (Refer
to Illustration 19).
7. Minimize the room lighting. With the Collimator Lamp OFF, measure and
record the ambient light level.
44
GE Medical Systems Proteus XR/e
REV 2 sm 2401594--100
Illustration 19
Light Contrast Calculation
EDGE OF A QUADRANT
LIGHT SENSOR ON THE LEFT EDGE LIGHT SENSOR ON THE RIGHT EDGE
BRH POINT
(LIGHT DROPS 25%)
9. Measure the illumination at a point 3 mm from the edge of the field toward
the center of the field (Light sensor on the left of edge). (Refer to
Illustration 19). Record this as I1.
10. Measure the illumination at a point 3 mm from the edge of the field away
from the center of the field (Light sensor on the right of edge). (Refer to
Illustration 19). Record this as I2.
11. Correct the values of I1 and I2 by subtracting from each value the ambient
light level measured in step-7. Now divide the corrected value of I1 by
I2. This ratio should be 4 or more.
12. Repeat the process from step-8. for all quadrants of Light Field.
13. If the deviation of Light Field Contrast Ratio is out acceptance limits, it is
necessary check the following:
G Check that the Collimator Lamp, the Mirror and the Mylar window
are not dirty or discolored.
G The ambient light level shall be low as it affects the accuracy of the
measurements.
45
GE Medical Systems Proteus XR/e
REV 2 sm 2401594--100
46
GE Medical Systems Proteus XR/e
REV 2 sm 2401594--100
Table 1
Periodic Maintenance
Adjustment and
Cleaning Lubrication Functional Check
calibration
1. Turn OFF generator and mains. Check that there is not voltage at table.
4. Press the mushroom shaped switch and verify that table is disconnected
and the table travels are blocked.
6. Connect Table power and check operator switches (foot switches and
hand switches).
47
GE Medical Systems Proteus XR/e
REV 2 sm 2401594--100
7. Check the status of the motors of the table. (if elevating Table).
10. Release the horizontal Bucky brakes and check correct traveling.
11. Take out the Cassette film tray and check correct functioning.
1. Turn OFF generator and mains. Check that there is not voltage at
Tube--Stand.
4. Turn ON mains and generator and check that the column and
Tube-Collimator Assembly is blocked in all motions.
7. Place the Tube at 0 on the Table--Top and check that X--ray indicator lamp
and the chassis centering indicator are aligned and they turn off after 30
seconds.
8. Check in the Collimator that light beam is centered with the X--ray beam
and the tolerance is not above 1% from focus--film distance.
1. Turn OFF Generator and mains. Check that there is not voltage at
Vertical Bucky.
2. Disassemble Front covers of Wall Bucky Stand and clean steel cable and
brakes.
4. Turn ON mains and generator and check that the Vertical Bucky is
blocked.
6. Press Bucky brake and verify that vertical motion is released and
counterweighted at any point.
7. Take out the Cassette film tray and check correct functioning.
48
GE Medical Systems Proteus XR/e
REV 2 sm 2401594--100
Table 2
Wall Bucky Spare Parts
Item GEMs Ref. MFG Ref. FRU Rep Name QTY Notes
7835--01 2 N Wheel 30 4
7836--01 2 N Wheel 28 4
Table 3
Wall Support Spare Parts
Item GEMs Ref. MFG Ref. FRU Rep Name QTY Notes
7835--01 2 N Wheel 30 4
7836--01 2 N Wheel 28 4
49
GE Medical Systems Proteus XR/e
REV 2 sm 2401594--100
Table 4
Radiographic Table Spare Parts
Item GEMS Ref. MFG Ref. FRU Rep Name QTY Notes
11347-01 2 N Tabletop 1
Table 5
Sliding Column Spare Parts
Item MFG Ref. GEMS Ref. FRU Rep Name QTY Notes
50
GE Medical Systems Proteus XR/e
REV 2 sm 2401594--100
Table 6
Collimator Renewal Parts
Item GEMS Ref. MFG Ref. FRU Rep Name QTY Notes
Table 7
Bucky Renewal Parts
Item GEMS Ref. MFG Ref. FRU Rep Name QTY Notes
Table 8
X-Ray Tubes
Item GEMS Ref. MFG Ref. FRU Rep Name QTY Notes
Table 9
Manuals Renewal Parts
Item GEMS Ref. MFG Ref. FRU Rep Name QTY Notes
51
GE Medical Systems Proteus XR/e
REV 2 sm 2401594--100
52
GE Medical Systems Proteus XR/e
REV 2 sm 2401594--100
SYSTEM INTERCONNECTION
53
GE Medical Systems Proteus XR/e
REV 2 sm 2401594--100
54
A B C D E
RAD TABLE
See schematic 54302065
3 for interconnections SUPPLY CABLE (A7235-XX) 3
TS1 Note 3:
GND CABLE (A6383-XX) AEC CABLE, AEC INTERFACE CABLE AND
Chassis AEC ADAPTATION BOARD NOT USED FOR
Chassis GND RAD ROOM STANDARD (without ION
GND CHAMBER)
BUCKY
GENERATOR CONSOLE
TABLE BUCKY CABLE (A3091-XX) AEC ADAPTATION BOARD
TS1 (A3263-03)
See schematic I/F-001
See schematic IM-015 ATP CONSOLE BOARD
2 2
COMMUNICATION CABLE (A7022-XX)
J3 J3
RAD TABLE
See schematic 54302065
3 for interconnections SUPPLY CABLE (A7235-XX) 3
TS1
GND CABLE (A6383-XX)
Chassis
Chassis GND
GND
BUCKY
TABLE BUCKY CABLE (A3091-XX)
TS1
See schematic I/F-001
ATP CONSOLE BOARD
(LOCKS BOARD) 2
Cable Alimentación Soporte Tubo
(Tube Support Power Supply Cable)
TB7 (A6736-XX) 3
0 VDC LOCKS
6,8 5
Freno Vertical CARRO LONG.
(Vertical Lock) (LONG. CARRIAGE)
6
GND GND
STUD GND Cable STUD
J1
Optional
1
Freno Transversal
(Transverse Lock)
2
J2
1
Freno Rotación
(Rotation Lock)
2
2 2
6
5
4
3
2
1
PANEL CONTROL
TS2
(CONTROL PANEL)
Optional
1 1
MESA (TABLE)
TS2
Bucky Lock Cable (A7096-01)
4 4
1 D2
1N4007
Freno Bucky
2 (Bucky Lock)
Solenoide Transversal
GENERADOR COMPACT
(COMPACT GENERATOR) (Transversal Solenoid)
TS1
Long. Solenoid Cable (A7098-01)
1
TARJETA FRENOS D1 Solenoide Longitudinal
1N5408
(LOCKS BOARD)
Cable Alimentación (Longitudinal Solenoid)
(Power Supply Cable) 2
TB7 (A7235-XX) TS3
Transversal Lock Cable (A7097-01)
+24 VDC LOCKS 1 D3
5,7 1N4007
Freno Transversal
3
0 VDC LOCKS 2 (Transversal Lock)
6,8
3
Foot Switch Cable (A7095-01) 3
4 SW1
GND GND
STUD GND Cable STUD PEDAL (FOOT SWITCH)
Tracking
Bucky Tracking Cable (A7099-01) SW3
2 Nota.- La función de tracking con el SW3 y el Cable Bucky Tracking es opcional. Si no se usa esta opción, hacer puente entre TS1-5 y TS1-6. 2
Note.- The tracking function with the SW3 and the Bucky Tracking Cable is optional. Make jumper between TS1-5 and TS1-6 when that option is not used.
1 1
Shield Shield
Nota 2.– Para sistemas con sólo una sóla Cámara de Ionización
compatible con el Generador, no es necesario instalar la “Tarjeta de
Adaptación AEC” (opcional), conectar directamente la Cámara de
Ionización a 3TS1 del Armario y a J5 de la “Tarjeta ATP Console CPU”
Note 2.– For systems with only one Ion Chamber compatible with the PANEL
Generator, installation of “AEC Adaptation Board” (optional) is not INTERFACE
GENERADOR COMPACT required, connect the Ion Chamber directly to 3TS1 in the Generator
(COMPACT GENERATOR) Cabinet and to J5 in the “ATP Console CPU Board”.
TARJETA FRENOS
(LOCKS BOARD) MANGUERA POTENCIA BUCKY MURAL (ver nota 1) PULSADOR FRENO BUCKY
(WALL BUCKY POWER CABLE – see note 1) TS2 CABLE PULSADOR (BUCKY LOCK PUSH–BUTTON)
TB7 (A6707–01/02/03) (A6706–01)
1
5, 7 +24 VDC LOCKS 2
6, 8 0 VDC LOCKS
3
4
5 FRENOS BUCKY
(BUCKY LOCKS)
GND GND CABLE GND
Nota 1.– Para Generador de Baterias con Millennium, conectarla a la regleta TS8 de
NAME DATE SHEET / OF DWG:
la Mesa (+24 VDC LOCKS a TS8–6, y 0 VDC LOCKS a TS8–5) 54301056
DRAWING F. GARCIA 05/05/01
Note 1.– For Battery Generator with Millennium, connect the cable to TS8 terminal 1/1
REVISED A. DIAZ 05/05/01 B A REV
strip of the Table (+24 VDC LOCKS to TS8–6, and 0 VDC LOCKS to TS8–5).
Shield Shield
GND GND
PANEL
Nota.– Para sistemas con sólo una sóla Cámara de Ionización compatible con el Generador, INTERFACE
no es necesario instalar la “Tarjeta de Adaptación AEC” (opcional), conectar directamente la
Cámara de Ionización a 3TS1 del Armario y a J5 de la “Tarjeta ATP Console CPU”
Note.– For systems with only one Ion Chamber compatible with the Generator, installation of
“AEC Adaptation Board” (optional) is not required, connect the Ion Chamber directly to
3TS1 in the Generator Cabinet and to J5 in the “ATP Console CPU Board”.
GENERADOR COMPACT
(COMPACT GENERATOR)
TARJETA FRENOS
(LOCKS BOARD) MANGUERA POTENCIA SOPORTE MURAL PULSADOR FRENO SOPORTE MURAL
(WALL SUPPORT POWER CABLE) TS2 CABLE PULSADOR (WALL SUPPORT LOCK PUSH–BUTTON)
TB7 (A6707–01/02/03) (A6706–01)
1
5, 7 +24 VDC LOCKS 2
6, 8 0 VDC LOCKS
3
4
5 FRENOS SOPORTE
(SUPPORT LOCKS)
GND GND CABLE GND