CAL Reg. No.

Calibration Certificate
Task no. : 121-21014
Certificate no. : 9.8-23905
Page : 1 of 3
CUSTOMER EQUIPMENT DATA: Enclosure : 1 (original data)
Meter type: Ultrasonic meter USM
Size: 24INCH/ANSI600/DN600
Model.: ALTOSONIC V12
No. - Year: A21042625-2021
TAG. no.: 00EKU00CF001
Manufacturer: Krohne- Altometer

CUSTOMER: Krohne Altometer


Kerkeplaat 12,
3313 LC Dordrecht/Netherlands
Netherlands Project: 800002120
att: John Weeda

TEST CONDITIONS:
Test method/medium: Volume flow with natural gas at a static pressure of 49 bar g.

SCOPE OF TEST: 6 point calibration with 3 runs.


Adjustment and 2 verification points with 3 runs.

FOUNDATION: FORCE No. 60.10.4.02 Calibration with 24 INCH HPCL no. C03-002

REMARKS: Technical evaluation: No Remarks.


The meter has been hardware protected and sealed after adjustment.

DATE OF CALIBRATION: 10-05-2021

RESULTS: Final results are given on page 2.


Detailed results are given in enclosure 1.
Gas mixture are given on page 3.

TRACEABILLITY: The laboratory is traceable to the harmonized European cubic meter for natural gas; EuReGa.
Used working standards, and calibration setup are given on page 3.
Version 2.1.13

SIGNATORY:

The certificate is only valid with digital signature from FORCE Technology. The certificate exist as original in FORCE Technology´s database.

FORCE Technology, Navervej 1, 6600 Vejen, phone: +4576961600, Danish, Designated institute (DI) for gas measurement.
The CIPM MRA agreement is valid for a best uncertainty of 0,15%.

The *General Conditions* on the reverse/last page are an integrated part of our services. Extracts from the certificate may only be
reproduced with a written permission from FORCE Technology. The test results only include the tested objects.
Task no. : 121-21014
Certificate no. : 9.8-23905
Page : 2 of 3

Meter data:
Pulse function As found As left
Ultrasonic meter USM Qmax 8538,3 m³/h.
24INCH/ANSI600/DN600 Qt 1707,7 m³/h.
A21042625-2021 control calibration Qmin 375 m³/h. Linearisation
K-factor 1 1
- OK Imp. Factor (HF) 500 500 pulses/m³
- - - - pulses/m³
- - - - pulses/m³
- - - - pulses/m³
- - - - pulses/m³
- - - - pulses/m³
Static calibration pressure 49 49 bar g.

Calibration at HF
Results: (I) As found Forward direction 49 bar g.
Qi/Qmax. % 4,4 12,9 26,9 45,7 71,4 100
Flow [m³/h] 375 1100 2300 3900 6100 8538
Error-avg [%] 0,47 0,35 0,25 0,19 0,13 0,11
±Utot [%] 0,19 0,18 0,16 0,21 0,16 0,15

Error calculated with


manufacturer's
Calculated error As left Forward direction 49 bar g. software
Qi/Qmax. % 4,4 12,9 26,9 45,7 71,4 100
Flow [m³/h] 375 1100 2300 3900 6100 8538
Error-avg [%] 0,00 0,00 0,00 0,00 0,00 0,00
±Utot [%]* 0,19 0,18 0,16 0,21 0,16 0,15

WME: Results: (I) As found 0,18 %


WME: Calculated error As left 0,00 %
WME is calculated according to OIML R137-1&2 point 3.2.5. The Qmax used for calculation of WME
is the used/calibrated Qmax and not necessarily the Qmax of the meter.
The calibration uncertainty (Utot), includes the accredited accuracy (Ucmc) and the repeatability of the meter (Udut).
CMC: (Calibration and Measurement Capability)
As left errors are corrected errors, calculated from As found results by using the meter manufacturer's software.
±Utot [%]* is equal to the uncertainty found before adjustment.
The reported expanded uncertainty ±Utot, is the standard uncertainty of the measurement multiplied by
a coverage factor k=2. The coverage factor corresponds a coverage probability of approximately 95%.
The standard uncertainty of measurement is in accordance with EA-4/02:
Error = (Indicated Volume - Real Volume)/Real Volume * 100%

Error curve as function of flow


2 Results: (I) As
found
1
Error-avg [%]

Results: (II) As left


0

-1

-2
0 2000 4000 6000 8000 10000
Flow [m³/h]
Measure points are connected with smooth lines.
Task no. : 121-21014
Certificate no. : 9.8-23905
Page : 3 of 3

Laboratory control equipment:


Facilities: FORCE no. C03-002 Working standards: Turbine meters

Traceability: The European Harmonized cubic meter; - EuReGa

Used working standards during the calibration:


FORCE ID no. WS1 WS2 WS3 WS4 WS5 WS6 WS7
Flow Flow range [m3/h] 400 - 4000 400 - 4000 400 - 4000 400 - 4000 120 - 1000 8 - 400 2400-25000
m3/h Certificate no. 9.8-22508 9.8-22509 9.8-22510 9.8-22511 9.8-22512 9.8-22513 9.8-22514
375 x
1100 x
2300 x x
3900 x x
6100 x x x x
8538 x x x x

Gas mixture during the calibration:


Analysis by: FORCE Technology
Description Name Unit Results
CH4 Methane mol% 98,77
C2H6 Ethane mol% -
C3H8 Propane mol% 0,13
n-C4H10 n-Butane mol% -
i-C4H10 i-Butane mol% -
n-C5H12 n-Pentane mol% -
i-C5H12 i-Pentane mol% -
n-C6H14 n-Hexane mol% -
CO2 Carbon dioxide mol% 0,34
N2 Nitrogen mol% 0,76

mol% 100,0

Calibration setup:
Up-stream
Meter Down-stream

FORCE DN600; length 10000 mm; ID: 560 mm FORCE DN600; length 10000 mm; ID: 560 mm

Temperature measurement: 3*D down-stream meter


Pressure measurement: Pr. Point

Conditioning: At least 5 min. at 15 m/s


Task No. : 121-21014
Certificate no. : 9.8-23905
Enclosure : 1 of 1
Repea-
Flow Flow Temp. RE Error tability Calibration
nominal actual avg. Number E1 E2 E3 E4 E5 E6 E7 E8 E9 E10 avg. ±udut ±Ucmc Uncertainty
m3/h m3/h °C - % % % % % % % % % % % % % ±Utot (%)

Calibration results. As found Forward direction 49 bar g.

375 374 21,4 6,77E+05 0,47 0,44 0,50 0,47 0,04 0,17 0,19
1100 1102 21,4 2,00E+06 0,34 0,34 0,38 0,35 0,03 0,17 0,18
2300 2296 21,3 4,16E+06 0,26 0,25 0,24 0,25 0,02 0,16 0,16
3900 3889 21,4 7,04E+06 0,13 0,23 0,19 0,19 0,06 0,16 0,21
6100 6096 21,3 1,10E+07 0,09 0,16 0,13 0,13 0,04 0,14 0,16
8538 8509 21,3 1,54E+07 0,12 0,12 0,08 0,11 0,03 0,14 0,15

Calibration results. Verification points Forward direction 49 bar g.

1100 1101 21,3 2,00E+06 0,04 0,02 0,02 0,03 0,01 0,17 0,17
6100 6094 21,3 1,10E+07 -0,01 0,00 -0,02 -0,01 0,01 0,14 0,14

Repeatability (±udut) and calibration uncertainty (±Utot)


The reported repeatability is calcultated in accordance with "WGFF Guidelines for CMC Uncertainty and Calibration Report Uncertainty" by using 95%
confidence level t-value for n-1 (number of runs - 1) degrees of freedom, divided by 2 and assuming k 95=2:

𝑡95 𝑠𝑟𝑒𝑝𝑒𝑎𝑡 𝑈𝐶𝑀𝐶 2


𝑢𝑑𝑢𝑡 = 𝑈𝑡𝑜𝑡 = 2 𝑥 2
+ 𝑢𝑑𝑢𝑡
2 𝑛 2
End of certificate.
Model: ALTOSONIC V12 Printed at: 10-05-2021 14:47:29
Serial No: A21042625 By: 'factory' [FACTORY]

Calibration Parameter Report


Cali- Switch
Nr Parameter Ix Value Unit Changed By bration Protected

1 ALTOSONIC V12/System
1 Vendor Name KROHNE 10-05-2021 09:37:00 Y Y
3 Product Ultrasonic Gas Flow Meter 06-01-2006 04:37:57 Y Y
4 Model ALTOSONIC V12 13-09-2019 13:48:56 developer Y Y
5 Revision 1.0 06-01-2006 04:59:37 factory Y Y
6 Product Code V12 06-01-2006 04:37:57 Y Y
7 Serial Number A21042625 06-04-2021 05:49:35 factory Y Y
8 Tag 00EKU00CF001 06-04-2021 05:49:23 factory Y Y
9 FPGA Version 1.0.0 10-05-2021 09:37:00 system Y
10 SD SW Version 1.0.1 10-05-2021 09:37:00 system Y
11 MB SW Version 1.0.0 10-05-2021 09:37:00 system Y
12 Main Application Name KAFKA_V12 10-05-2021 09:37:00 system Y
13 Main App. Version 2.0.0.3 10-05-2021 09:37:00 system Y
19 Calibrated FORCE Technology 10-05-2021 09:42:46 factory Y Y
20 Calibrator SMEI - Søren meier 10-05-2021 09:42:59 factory Y Y

1.100.100 ALTOSONIC V12/System/Background Jobs/System Integrity Check


1 Check Interval 60 sec 27-07-2018 15:55:02 developer Y Y
2 Items to Check 15 27-07-2018 11:47:03 developer Y Y
3 Reboot on Error 0 27-07-2018 15:44:45 developer Y Y

1.300 ALTOSONIC V12/System/Checksums


100 Exe Checksum (Stored) 942941310 10-05-2021 09:37:06 system Y
101 Exe Checksum (Calculated) 942941310 10-05-2021 09:37:06 system Y
200 Calibration Checksum 3810401343 10-05-2021 14:28:26 system Y
300 Hardware Checksum 3831373500 10-05-2021 09:37:09 system Y
301 PCB Version Checksum 1305597214 10-05-2021 09:37:09 system Y
400 SW Platform Checksum 2986471689 10-05-2021 09:37:09 system Y
401 Flowmeter Application SW Checksum 551375010 10-05-2021 09:37:11 system Y

1.300.500 ALTOSONIC V12/System/Checksums/System Deployment Check


1 Checked Items 7 23-09-2019 07:25:57 Y

4.2 ALTOSONIC V12/Outputs/Frequency & Status Output


100 Freq.1 Output Value ALTOSONIC 22-02-2018 16:10:25 system Y
V12/FlowCalculationModule/
Outputs/Flow_Calcs/Flow/Fl
ow_Proces
101 Freq.1 Output 100% Frequency 3200 Hz 06-04-2021 05:43:28 factory Y Y
102 Freq.1 Output 100% Reference 6,4 ? 06-04-2021 05:43:28 factory Y Y
104 Freq.1 Output Pulses Per Unit 500 06-04-2021 05:43:28 system Y Y
300 Freq.2 Output Value ALTOSONIC 22-02-2018 16:10:25 system Y
V12/FlowCalculationModule/
Outputs/Secondary Flow
Measurement/Flow/Flow_Pr
oces
301 Freq.2 Output 100% Frequency 3200 Hz 06-04-2021 05:43:28 factory Y Y
302 Freq.2 Output 100% Reference 6,4 ? 06-04-2021 05:43:28 factory Y Y
304 Freq.2 Output Pulses Per Unit 500 06-04-2021 05:43:28 system Y Y

5.1.1 ALTOSONIC V12/FlowCalculationModule/Configuration/Flow_Calcs


3 Diameter 0,54 m 06-04-2021 05:43:28 factory Y Y
4 MeterConstant_Fwd 1 24-08-2018 08:19:50 developer Y Y
5 MeterConstant_Rev 1 24-08-2018 08:19:55 developer Y Y

Page 1 of 3
Model: ALTOSONIC V12 Printed at: 10-05-2021 14:47:30
Serial No: A21042625 By: 'factory' [FACTORY]

Calibration Parameter Report


Cali- Switch
Nr Parameter Ix Value Unit Changed By bration Protected

5.1.1.10 ALTOSONIC V12/FlowCalculationModule/Configuration/Flow_Calcs/Channels


1 Ch_PathLength_Cal 1 0,6897911 m 06-04-2021 05:56:58 system Y Y
2 0,9955304 m Y Y
3 1,149682 m Y Y
4 0,9956089 m Y Y
5 0,6897322 m Y Y
6 1,149953 m Y Y
4 Ch_Angle_Measuring 1 70 degree 06-04-2021 05:43:28 factory Y Y
2 70 degree Y Y
3 70 degree Y Y
4 70 degree Y Y
5 70 degree Y Y
6 70 degree Y Y
7 Ch_Weighing_Factor 1 0,248016 06-04-2021 05:43:28 factory Y Y
2 0,412698 Y Y
3 0,535714 Y Y
4 0,412698 Y Y
5 0,248016 Y Y
6 0 Y Y

5.1.1.13 ALTOSONIC V12/FlowCalculationModule/Configuration/Flow_Calcs/Constraints


1 Fwd_MaxFlowRate 6,4 m³/s 06-04-2021 05:43:28 factory Y Y
2 Rev_MaxFlowRate 6,4 m³/s 06-04-2021 05:43:28 factory Y Y
3 Fwd_LowFlowCutoff 0,03 m/s 06-01-2006 04:59:22 factory Y Y
4 Rev_LowFlowCutoff 0,03 m/s 06-01-2006 04:59:22 factory Y Y
5 LowFlowCutoff_Threshold 50 % 06-01-2006 04:59:22 factory Y Y

5.1.1.16 ALTOSONIC V12/FlowCalculationModule/Configuration/Flow_Calcs/Linearisation


1 Linearisation Mode On 10-05-2021 14:18:45 factory Y Y

5.1.1.16.2 ALTOSONIC V12/FlowCalculationModule/Configuration/Flow_Calcs/Linearisation/Table Forward


1 Used Lines 6 10-05-2021 14:23:48 factory Y Y
2 Flow Rate 1 8517,9 m³/h 10-05-2021 14:24:13 factory Y Y
2 6104,164 m³/h Y Y
3 3896,622 m³/h Y Y
4 2301,733 m³/h Y Y
5 1105,933 m³/h Y Y
6 375,4 m³/h Y Y
7 0 m³/h Y Y
8 0 m³/h Y Y
9 0 m³/h Y Y
10 0 m³/h Y Y
3 Deviation 1 0,1085316 % 10-05-2021 14:25:43 factory Y Y
2 0,1268591 % Y Y
3 0,1853966 % Y Y
4 0,2519273 % Y Y
5 0,353765 % Y Y
6 0,4686377 % Y Y
7 0 % Y Y
8 0 % Y Y
9 0 % Y Y
10 0 % Y Y

Page 2 of 3
Model: ALTOSONIC V12 Printed at: 10-05-2021 14:47:30
Serial No: A21042625 By: 'factory' [FACTORY]

Calibration Parameter Report


Cali- Switch
Nr Parameter Ix Value Unit Changed By bration Protected

5.1.1.16.3 ALTOSONIC V12/FlowCalculationModule/Configuration/Flow_Calcs/Linearisation/Table Reverse


1 Used Lines 6 10-05-2021 14:26:02 factory Y Y
2 Flow Rate 1 8517,9 m³/h 10-05-2021 14:27:00 factory Y Y
2 6104,164 m³/h Y Y
3 3896,622 m³/h Y Y
4 2301,733 m³/h Y Y
5 1105,933 m³/h Y Y
6 375,4 m³/h Y Y
7 0 m³/h Y Y
8 0 m³/h Y Y
9 0 m³/h Y Y
10 0 m³/h Y Y
3 Deviation 1 0,1085316 % 10-05-2021 14:28:26 factory Y Y
2 0,1268591 % Y Y
3 0,1853966 % Y Y
4 0,2519273 % Y Y
5 0,353765 % Y Y
6 0,4686377 % Y Y
7 0 % Y Y
8 0 % Y Y
9 0 % Y Y
10 0 % Y Y

5.1.5.3 ALTOSONIC V12/FlowCalculationModule/Configuration/Viscosity/ViscosityKinematic


1 VK_Source Reference value Static 06-01-2006 04:37:57 Y Y

5.1.5.4 ALTOSONIC V12/FlowCalculationModule/Configuration/Viscosity/ViscosityDynamic


1 VD_Source Reference value Static 06-01-2006 04:37:57 Y Y

5.1.5.5 ALTOSONIC V12/FlowCalculationModule/Configuration/Viscosity/Density


1 D_Source Reference value Static 06-01-2006 04:37:57 Y Y

Page 3 of 3
CAL Reg. No. 9

Calibration Certificate
Task no. : 121-21014
Certificate no. : 9.8-23906
Page : 1 of 3
CUSTOMER EQUIPMENT DATA: Enclosure : 1 (original data)
Meter type: Ultrasonic meter USM
Size: 24INCH/ANSI600/DN600
Model.: ALTOSONIC V12
No. - Year: A21042626-2021
TAG. no.: 00EKU00CF002
Manufacturer: Krohne- Altometer

CUSTOMER: Krohne Altometer


Kerkeplaat 12,
3313 LC Dordrecht/Netherlands
Netherlands Project: 800002120
att: John Weeda

TEST CONDITIONS:
Test method/medium: Volume flow with natural gas at a static pressure of 49 bar g.

SCOPE OF TEST: 6 point calibration with 3 runs.


Adjustment and 2 verification points with 3 runs.

FOUNDATION: FORCE No. 60.10.4.02 Calibration with 24 INCH HPCL no. C03-002

REMARKS: Technical evaluation: No Remarks.


The meter has been hardware protected and sealed after adjustment.

DATE OF CALIBRATION: 10-05-2021

RESULTS: Final results are given on page 2.


Detailed results are given in enclosure 1.
Gas mixture are given on page 3.

TRACEABILLITY: The laboratory is traceable to the harmonized European cubic meter for natural gas; EuReGa.
Used working standards, and calibration setup are given on page 3.
Version 2.1.13

SIGNATORY:

The certificate is only valid with digital signature from FORCE Technology. The certificate exist as original in FORCE Technology´s database.

FORCE Technology, Navervej 1, 6600 Vejen, phone: +4576961600, Danish, Designated institute (DI) for gas measurement.
The CIPM MRA agreement is valid for a best uncertainty of 0,15%.

The *General Conditions* on the reverse/last page are an integrated part of our services. Extracts from the certificate may only be
reproduced with a written permission from FORCE Technology. The test results only include the tested objects.
Task no. : 121-21014
Certificate no. : 9.8-23906
Page : 2 of 3

Meter data:
Pulse function As found As left
Ultrasonic meter USM Qmax 8538,3 m³/h.
24INCH/ANSI600/DN600 Qt 1707,7 m³/h.
A21042626-2021 control calibration Qmin 375 m³/h. Linearisation
K-factor 1 1
- OK Imp. Factor (HF) 500 500 pulses/m³
- - - - pulses/m³
- - - - pulses/m³
- - - - pulses/m³
- - - - pulses/m³
- - - - pulses/m³
Static calibration pressure 49 49 bar g.

Calibration at HF
Results: (I) As found Forward direction 49 bar g.
Qi/Qmax. % 4,4 12,9 26,9 45,7 71,4 100
Flow [m³/h] 375 1100 2300 3900 6100 8538
Error-avg [%] -0,57 0,23 0,10 0,03 0,01 -0,01
±Utot [%] 0,20 0,22 0,16 0,17 0,16 0,20

Error calculated with


manufacturer's
Calculated error As left Forward direction 49 bar g. software
Qi/Qmax. % 4,4 12,9 26,9 45,7 71,4 100
Flow [m³/h] 375 1100 2300 3900 6100 8538
Error-avg [%] 0,00 0,00 0,00 0,00 0,00 0,00
±Utot [%]* 0,20 0,22 0,16 0,17 0,16 0,20

WME: Results: (I) As found 0,02 %


WME: Calculated error As left 0,00 %
WME is calculated according to OIML R137-1&2 point 3.2.5. The Qmax used for calculation of WME
is the used/calibrated Qmax and not necessarily the Qmax of the meter.
The calibration uncertainty (Utot), includes the accredited accuracy (Ucmc) and the repeatability of the meter (Udut).
CMC: (Calibration and Measurement Capability)
As left errors are corrected errors, calculated from As found results by using the meter manufacturer's software.
±Utot [%]* is equal to the uncertainty found before adjustment.
The reported expanded uncertainty ±Utot, is the standard uncertainty of the measurement multiplied by
a coverage factor k=2. The coverage factor corresponds a coverage probability of approximately 95%.
The standard uncertainty of measurement is in accordance with EA-4/02:
Error = (Indicated Volume - Real Volume)/Real Volume * 100%

Error curve as function of flow


2 Results: (I) As
found
1
Error-avg [%]

Results: (II) As left


0

-1

-2
0 2000 4000 6000 8000 10000
Flow [m³/h]
Measure points are connected with smooth lines.
Task no. : 121-21014
Certificate no. : 9.8-23906
Page : 3 of 3

Laboratory control equipment:


Facilities: FORCE no. C03-002 Working standards: Turbine meters

Traceability: The European Harmonized cubic meter; - EuReGa

Used working standards during the calibration:


FORCE ID no. WS1 WS2 WS3 WS4 WS5 WS6 WS7
Flow Flow range [m3/h] 400 - 4000 400 - 4000 400 - 4000 400 - 4000 120 - 1000 8 - 400 2400-25000
m3/h Certificate no. 9.8-22508 9.8-22509 9.8-22510 9.8-22511 9.8-22512 9.8-22513 9.8-22514
375 x
1100 x
2300 x x
3900 x x
6100 x x x x
8538 x x x x

Gas mixture during the calibration:


Analysis by: FORCE Technology
Description Name Unit Results
CH4 Methane mol% 98,80
C2H6 Ethane mol% -
C3H8 Propane mol% 0,10
n-C4H10 n-Butane mol% -
i-C4H10 i-Butane mol% -
n-C5H12 n-Pentane mol% -
i-C5H12 i-Pentane mol% -
n-C6H14 n-Hexane mol% -
CO2 Carbon dioxide mol% 0,36
N2 Nitrogen mol% 0,74

mol% 100,0

Calibration setup:
Up-stream
Meter Down-stream

FORCE DN600; length 10000 mm; ID: 560 mm FORCE DN600; length 10000 mm; ID: 560 mm

Temperature measurement: 3*D down-stream meter


Pressure measurement: Pr. Point

Conditioning: At least 5 min. at 15 m/s


Task No. : 121-21014
Certificate no. : 9.8-23906
Enclosure : 1 of 1
Repea-
Flow Flow Temp. RE Error tability Calibration
nominal actual avg. Number E1 E2 E3 E4 E5 E6 E7 E8 E9 E10 avg. ±udut ±Ucmc Uncertainty
m3/h m3/h °C - % % % % % % % % % % % % % ±Utot (%)

Calibration results. As found Forward direction 49 bar g.

375 374 22,6 6,70E+05 -0,61 -0,53 -0,57 -0,57 0,05 0,17 0,20
1100 1102 22,7 1,97E+06 0,20 0,29 0,20 0,23 0,07 0,17 0,22
2300 2295 22,8 4,11E+06 0,10 0,08 0,11 0,10 0,02 0,16 0,16
3900 3891 22,8 6,96E+06 0,01 0,04 0,04 0,03 0,03 0,16 0,17
6100 6098 22,7 1,09E+07 0,01 0,03 -0,02 0,01 0,04 0,14 0,16
8538 8507 22,9 1,52E+07 -0,07 0,04 -0,01 -0,01 0,07 0,14 0,20

Calibration results. Verification points Forward direction 49 bar g.

1100 1102 22,8 1,97E+06 0,11 0,17 0,16 0,15 0,04 0,17 0,19
6100 6097 22,8 1,09E+07 0,03 0,01 0,04 0,03 0,02 0,14 0,15

Repeatability (±udut) and calibration uncertainty (±Utot)


The reported repeatability is calcultated in accordance with "WGFF Guidelines for CMC Uncertainty and Calibration Report Uncertainty" by using 95%
confidence level t-value for n-1 (number of runs - 1) degrees of freedom, divided by 2 and assuming k 95=2:

𝑡95 𝑠𝑟𝑒𝑝𝑒𝑎𝑡 𝑈𝐶𝑀𝐶 2


𝑢𝑑𝑢𝑡 = 𝑈𝑡𝑜𝑡 = 2 𝑥 2
+ 𝑢𝑑𝑢𝑡
2 𝑛 2
End of certificate.
Model: ALTOSONIC V12 Printed at: 10-05-2021 23:43:23
Serial No: A21042626 By: 'factory' [FACTORY]

Calibration Parameter Report


Cali- Switch
Nr Parameter Ix Value Unit Changed By bration Protected

1 ALTOSONIC V12/System
1 Vendor Name KROHNE 10-05-2021 16:27:22 Y Y
3 Product Ultrasonic Gas Flow Meter 06-01-2006 04:37:57 Y Y
4 Model ALTOSONIC V12 13-09-2019 13:48:56 developer Y Y
5 Revision 1.0 06-01-2006 04:59:37 factory Y Y
6 Product Code V12 06-01-2006 04:37:57 Y Y
7 Serial Number A21042626 02-04-2021 06:59:22 factory Y Y
8 Tag 00EKU00CF002 02-04-2021 06:59:10 factory Y Y
9 FPGA Version 1.0.0 10-05-2021 16:27:22 system Y
10 SD SW Version 1.0.1 10-05-2021 16:27:22 system Y
11 MB SW Version 1.0.0 10-05-2021 16:27:22 system Y
12 Main Application Name KAFKA_V12 10-05-2021 16:27:22 system Y
13 Main App. Version 2.0.0.3 10-05-2021 16:27:22 system Y
19 Calibrated Force Technology 10-05-2021 21:17:12 factory Y Y
20 Calibrator Sarmed Hamdi 10-05-2021 21:17:19 factory Y Y

1.100.100 ALTOSONIC V12/System/Background Jobs/System Integrity Check


1 Check Interval 60 sec 27-07-2018 15:55:02 developer Y Y
2 Items to Check 15 27-07-2018 11:47:03 developer Y Y
3 Reboot on Error 0 27-07-2018 15:44:45 developer Y Y

1.300 ALTOSONIC V12/System/Checksums


100 Exe Checksum (Stored) 942941310 10-05-2021 16:27:29 system Y
101 Exe Checksum (Calculated) 942941310 10-05-2021 16:27:29 system Y
200 Calibration Checksum 2832805398 10-05-2021 22:46:00 system Y
300 Hardware Checksum 3171257851 10-05-2021 16:27:32 system Y
301 PCB Version Checksum 1305597214 10-05-2021 16:27:32 system Y
400 SW Platform Checksum 2986471689 10-05-2021 16:27:32 system Y
401 Flowmeter Application SW Checksum 551375010 10-05-2021 16:27:34 system Y

1.300.500 ALTOSONIC V12/System/Checksums/System Deployment Check


1 Checked Items 7 23-09-2019 07:25:57 Y

4.2 ALTOSONIC V12/Outputs/Frequency & Status Output


100 Freq.1 Output Value ALTOSONIC 22-02-2018 16:10:25 system Y
V12/FlowCalculationModule/
Outputs/Flow_Calcs/Flow/Fl
ow_Proces
101 Freq.1 Output 100% Frequency 3200 Hz 02-04-2021 06:57:57 factory Y Y
102 Freq.1 Output 100% Reference 6,4 ? 02-04-2021 06:57:57 factory Y Y
104 Freq.1 Output Pulses Per Unit 500 02-04-2021 06:57:57 system Y Y
300 Freq.2 Output Value ALTOSONIC 22-02-2018 16:10:25 system Y
V12/FlowCalculationModule/
Outputs/Secondary Flow
Measurement/Flow/Flow_Pr
oces
301 Freq.2 Output 100% Frequency 3200 Hz 02-04-2021 06:57:57 factory Y Y
302 Freq.2 Output 100% Reference 6,4 ? 02-04-2021 06:57:57 factory Y Y
304 Freq.2 Output Pulses Per Unit 500 02-04-2021 06:57:57 system Y Y

5.1.1 ALTOSONIC V12/FlowCalculationModule/Configuration/Flow_Calcs


3 Diameter 0,54 m 02-04-2021 06:57:57 factory Y Y
4 MeterConstant_Fwd 1 24-08-2018 08:19:50 developer Y Y
5 MeterConstant_Rev 1 24-08-2018 08:19:55 developer Y Y

Page 1 of 3
Model: ALTOSONIC V12 Printed at: 10-05-2021 23:43:23
Serial No: A21042626 By: 'factory' [FACTORY]

Calibration Parameter Report


Cali- Switch
Nr Parameter Ix Value Unit Changed By bration Protected

5.1.1.10 ALTOSONIC V12/FlowCalculationModule/Configuration/Flow_Calcs/Channels


1 Ch_PathLength_Cal 1 0,6896344 m 02-04-2021 07:24:20 system Y Y
2 0,9951309 m Y Y
3 1,148959 m Y Y
4 0,9948679 m Y Y
5 0,6895329 m Y Y
6 1,149199 m Y Y
4 Ch_Angle_Measuring 1 70 degree 02-04-2021 06:57:57 factory Y Y
2 70 degree Y Y
3 70 degree Y Y
4 70 degree Y Y
5 70 degree Y Y
6 70 degree Y Y
7 Ch_Weighing_Factor 1 0,248016 02-04-2021 06:57:57 factory Y Y
2 0,412698 Y Y
3 0,535714 Y Y
4 0,412698 Y Y
5 0,248016 Y Y
6 0 Y Y

5.1.1.13 ALTOSONIC V12/FlowCalculationModule/Configuration/Flow_Calcs/Constraints


1 Fwd_MaxFlowRate 6,4 m³/s 02-04-2021 06:57:57 factory Y Y
2 Rev_MaxFlowRate 6,4 m³/s 02-04-2021 06:57:57 factory Y Y
3 Fwd_LowFlowCutoff 0,03 m/s 06-01-2006 04:59:22 factory Y Y
4 Rev_LowFlowCutoff 0,03 m/s 06-01-2006 04:59:22 factory Y Y
5 LowFlowCutoff_Threshold 50 % 06-01-2006 04:59:22 factory Y Y

5.1.1.16 ALTOSONIC V12/FlowCalculationModule/Configuration/Flow_Calcs/Linearisation


1 Linearisation Mode On 10-05-2021 22:41:57 factory Y Y

5.1.1.16.2 ALTOSONIC V12/FlowCalculationModule/Configuration/Flow_Calcs/Linearisation/Table Forward


1 Used Lines 0 06-01-2006 04:37:57 Y Y
2 Flow Rate 1 8506,179 m³/h 10-05-2021 22:42:17 factory Y Y
2 6097,993 m³/h Y Y
3 3892,2 m³/h Y Y
4 2297,2 m³/h Y Y
5 1104,822 m³/h Y Y
6 371,6667 m³/h Y Y
7 0 m³/h Y Y
8 0 m³/h Y Y
9 0 m³/h Y Y
10 0 m³/h Y Y
3 Deviation 1 -0,01207046 % 10-05-2021 22:43:36 factory Y Y
2 0,006011812 % Y Y
3 0,03048611 % Y Y
4 0,09742739 % Y Y
5 0,2308387 % Y Y
6 -0,5723608 % Y Y
7 0 % Y Y
8 0 % Y Y
9 0 % Y Y
10 0 % Y Y

Page 2 of 3
Model: ALTOSONIC V12 Printed at: 10-05-2021 23:43:23
Serial No: A21042626 By: 'factory' [FACTORY]

Calibration Parameter Report


Cali- Switch
Nr Parameter Ix Value Unit Changed By bration Protected

5.1.1.16.3 ALTOSONIC V12/FlowCalculationModule/Configuration/Flow_Calcs/Linearisation/Table Reverse


1 Used Lines 0 06-01-2006 04:37:57 Y Y
2 Flow Rate 1 8506,179 m³/h 10-05-2021 22:44:52 factory Y Y
2 6097,993 m³/h Y Y
3 3892,2 m³/h Y Y
4 2297,2 m³/h Y Y
5 1104,822 m³/h Y Y
6 371,6667 m³/h Y Y
7 0 m³/h Y Y
8 0 m³/h Y Y
9 0 m³/h Y Y
10 0 m³/h Y Y
3 Deviation 1 -0,01207046 % 10-05-2021 22:46:00 factory Y Y
2 0,006011812 % Y Y
3 0,03048611 % Y Y
4 0,09742739 % Y Y
5 0,2308387 % Y Y
6 -0,5723608 % Y Y
7 0 % Y Y
8 0 % Y Y
9 0 % Y Y
10 0 % Y Y

5.1.5.3 ALTOSONIC V12/FlowCalculationModule/Configuration/Viscosity/ViscosityKinematic


1 VK_Source Reference value Static 06-01-2006 04:37:57 Y Y

5.1.5.4 ALTOSONIC V12/FlowCalculationModule/Configuration/Viscosity/ViscosityDynamic


1 VD_Source Reference value Static 06-01-2006 04:37:57 Y Y

5.1.5.5 ALTOSONIC V12/FlowCalculationModule/Configuration/Viscosity/Density


1 D_Source Reference value Static 06-01-2006 04:37:57 Y Y

Page 3 of 3