You are on page 1of 23

2013 SECTION V ARTICLE 7

Table T-721
Requirements of a Magnetic Particle Examination Procedure
Nonessential
Requirement Essential Variable Variable
Magnetizing technique X ...
Magnetizing current type or amperage outside range specified by this Article or as previously X ...
qualified
Surface preparation X .. .
Magnetic particles (fluorescent/visible, color, particle size, wet/dry) X .. .
Method of particle application X .. .
Method of excess particle removal X .. .
Minimum light intensity X .. .
Existing coatings, greater than the thickness demonstrated X .. .
Nonmagnetic surface contrast enhancement, when utilized X .. .
Performance demonstration, when required X .. .
Examination part surface temperature outside of the temperature range recommended by the X .. .
manufacturer of the particles or as previously qualified
Shape or size of the examination object . . . X
Equipment of the same type . . . X
Temperature (within those specified by manufacturer or as previously qualified) . . . X
Demagnetizing technique . . . X
Post-examination cleaning technique . . . X
Personnel qualification requirements . . . X

(b) Prior to magnetic particle examination, the surface T-750 TECHNIQUE


to be examined and all adjacent areas within at least
1 in. (25 mm) shall be dry and free of all dirt, grease, lint, T-751 TECHNIQUES
scale, welding flux and spatter, oil, or other extraneous One or more of the following five magnetization techni-
matter that could interfere with the examination. ques shall be used:
(c) Cleaning may be accomplished using detergents, or- (a) prod technique
ganic solvents, descaling solutions, paint removers, vapor (b) longitudinal magnetization technique
degreasing, sand or grit blasting, or ultrasonic cleaning (c) circular magnetization technique
methods. (d) yoke technique
(d) If nonmagnetic coatings are left on the part in the (e) multidirectional magnetization technique
area being examined, it shall be demonstrated that indica-
tions can be detected through the existing maximum coat- T-752 PROD TECHNIQUE
ing thickness applied. When AC yoke technique is used, the
demonstration shall be in accordance with T-752.1 Magnetizing Procedure. For the prod techni-
Mandatory Appendix I of this Article. que, magnetization is accomplished by portable prod type
electrical contacts pressed against the surface in the area
T-741.2 Nonmagnetic Surface Contrast Enhance- to be examined. To avoid arcing, a remote control switch,
ment. Nonmagnetic surface contrasts may be applied by which may be built into the prod handles, shall be pro-
the examiner to uncoated surfaces, only in amounts suffi- vided to permit the current to be applied after the prods
cient to enhance particle contrast. When nonmagnetic sur- have been properly positioned.
face contrast enhancement is used, it shall be
demonstrated that indications can be detected through T-752.2 Magnetizing Current. Direct or rectified
the enhancement. Thickness measurement of this non- magnetizing current shall be used. The current shall be
magnetic surface contrast enhancement is not required. 100 (minimum) amp/in. (4 amp/mm) to
125 (maximum) amp/in. (5 amp/mm) of prod spacing
NOTE: Refer to T-150(a) for guidance for the demonstration required
in T-741.1(d) and T-741.2.
for sections 3/4 in. (19 mm) thick or greater. For sections
less than 3/4 in. (19 mm) thick, the current shall be
90 amp/in. (3.6 amp/mm) to 110 amp/in.
(4.4 amp/mm) of prod spacing.
T-752.3 Prod Spacing. Prod spacing shall not exceed
8 in. (200 mm). Shorter spacing may be used to accommo-
date the geometric limitations of the area being examined
or to increase the sensitivity, but prod spacings of less

173
--`,,,`,`,``,``,,`,``,`,`,`,``,`-`-`,,`,,`,`,,`---

Copyright ASME International (BPVC)


Provided by IHS under license with ASME pe.class Licensee=PDVSA - Pequiven site 9/9986712014, User=Gil, Eliecer
No reproduction or networking permitted without license from IHS Not for Resale, 07/23/2013 06:23:05 MDT
ARTICLE 7 2013 SECTION V

than 3 in. (75 mm) are usually not practical due to banding (d) If the area to be magnetized extends beyond 9 in.
of the particles around the prods. The prod tips shall be (225 mm) on either side of the coil’s center, field adequacy
kept clean and dressed. If the open circuit voltage of the shall be demonstrated using a magnetic field indicator or
magnetizing current source is greater than 25 V, lead, artificial flaw shims per T-764.
steel, or aluminum (rather than copper) tipped prods (e) F o r l a r g e p a r t s d u e t o s iz e a n d s h a p e , t h e
are recommended to avoid copper deposits on the part magnetizing current shall be 1200 ampere-turns
being examined. to 4500 ampere-turns. The field adequacy shall be demon-
strated using artificial flaw shims or a pie-shaped mag-

--`,,,`,`,``,``,,`,``,`,`,`,``,`-`-`,,`,,`,`,,`---
netic field indicator in accordance with T-764. A
T-753 LONGITUDINAL MAGNETIZATION Hall-Effect probe gaussmeter shall not be used with encir-
TECHNIQUE cling coil magnetization techniques.

T-753.1 Magnetizing Procedure. For this technique, T-753.3 Magnetizing Current. The current required
magnetization is accomplished by passing current through to obtain the necessary magnetizing field strength shall
a multi-turn fixed coil (or cables) that is wrapped around be determined by dividing the ampere-turns obtained in
the part or section of the part to be examined. This pro- steps T-753.2(a) or T-753.2(b) by the number of turns
duces a longitudinal magnetic field parallel to the axis of in the coil as follows:
the coil.
If a fixed, prewound coil is used, the part shall be placed
near the side of the coil during inspection. This is of special
importance when the coil opening is more than 10 times
For example, if a 5-turn coil is used and the ampere-
the cross-sectional area of the part.
turns required are 5000, use
ð13Þ T-753.2 Magnetic Field Strength. Direct or rectified
current shall be used to magnetize parts examined by this
technique. The required field strength shall be calculated
based on the length L and the diameter D of the part in ac-
cordance with (a) and (b), or as established in (d) and (e),
below. Long parts shall be examined in sections not to ex- T-754 CIRCULAR MAGNETIZATION TECHNIQUE
ceed 18 in. (450 mm), and 18 in. (450 mm) shall be used T-754.1 Direct Contact Technique.
for the part L in calculating the required field strength. For (a) Magnetizing Procedure. For this technique, magneti-
noncylindrical parts, D shall be the maximum cross- zation is accomplished by passing current through the
sectional diagonal. part to be examined. This produces a circular magnetic
(a) Parts With L/D Ratios Equal to or Greater Than 4. field that is approximately perpendicular to the direction
The magnetizing current shall be within 10% of the of current flow in the part.
ampere-turns’ value determined as follows: (b) Magnetizing Current. Direct or rectified (half-wave
rectified or full-wave rectified) magnetizing current shall
be used.
(1) The current shall be 300 amp/in. (12 A/mm) to
800 amp/in. (31 A/mm) of outer diameter.
(2) Parts with geometric shapes other than round
For example, a part 10 in. (250 mm) long × 2 in. with the greatest cross-sectional diagonal in a plane at
(50 mm) diameter has an L /D ratio of 5. Therefore, right angles to the current flow shall determine the inches
to be used in (1) above.
(3) If the current levels required for (1) cannot be ob-
tained, the maximum current obtainable shall be used and
the field adequacy shall be demonstrated in accordance
with T-764.
(b) Parts With L/D Ratios Less Than 4 but Not Less Than
2. The magnetizing ampere-turns shall be within 10% of T-754.2 Central Conductor Technique.
the ampere-turns’ value determined as follows: (a) Magnetizing Procedure. For this technique, a central
conductor is used to examine the internal surfaces of cy-
lindrically or ring-shaped parts. The central conductor
technique may also be used for examining the outside sur-
faces of these shapes. Where large diameter cylinders are
to be examined, the conductor shall be positioned close to
(c) Parts With L/D Ratios Less Than 2. Coil magnetiza- the internal surface of the cylinder. When the conductor is
tion technique cannot be used. not centered, the circumference of the cylinder shall be

174
Copyright ASME International (BPVC)
Provided by IHS under license with ASME pe.class Licensee=PDVSA - Pequiven site 9/9986712014, User=Gil, Eliecer
No reproduction or networking permitted without license from IHS Not for Resale, 07/23/2013 06:23:05 MDT
2013 SECTION V ARTICLE 7

examined in increments. Field strength measurements in


Figure T-754.2.2
accordance with T-764 shall be used, to determine the ex-
The Effective Region of Examination When
tent of the arc that may be examined for each conductor
Using an Offset Central Conductor
position or the rules in (c) below may be followed. Bars
or cables, passed through the bore of a cylinder, may be
used to induce circular magnetization. Central conductor
Effective
(b) Magnetizing Current. The field strength required region 4d
shall be equal to that determined in T-754.1(b) for a
single-turn central conductor. The magnetic field will in-
crease in proportion to the number of times the central
conductor cable passes through a hollow part. For exam-
d
ple, if 6000 amperes are required to examine a part using
a single pass central conductor, then 3000 amperes are re-
quired when 2 passes of the through-cable are used, and
1200 amperes are required if 5 passes are used (see
Figure T-754.2.1). When the central conductor technique
is used, magnetic field adequacy shall be verified using a
magnetic particle field indicator in accordance with T-764.
(c) Offset Central Conductor. When the conductor pas-
sing through the inside of the part is placed against an in-
side wall of the part, the current levels, as given in T-756 MULTIDIRECTIONAL MAGNETIZATION
T-754.1(b)(1) shall apply, except that the diameter used TECHNIQUE
for current calculations shall be the sum of the diameter T-756.1 Magnetizing Procedure. For this technique,
of the central conductor and twice the wall thickness. magnetization is accomplished by high amperage power
--`,,,`,`,``,``,,`,``,`,`,`,``,`-`-`,,`,,`,`,,`---

The distance along the part circumference (exterior) that packs operating as many as three circuits that are ener-
is effectively magnetized shall be taken as four times the gized one at a time in rapid succession. The effect of these
diameter of the central conductor, as illustrated in rapidly alternating magnetizing currents is to produce an
Figure T-754.2.2. The entire circumference shall be in- overall magnetization of the part in multiple directions.
spected by rotating the part on the conductor, allowing Circular or longitudinal magnetic fields may be generated
for approximately a 10% magnetic field overlap. in any combination using the various techniques de-
scribed in T-753 and T-754.
T-755 YOKE TECHNIQUE
For this technique, alternating or direct current electro- T-756.2 Magnetic Field Strength. Only three phase,
magnetic yokes, or permanent magnet yokes, shall be full-wave rectified current shall be used to magnetize
used. the part. The initial magnetizing current requirements
for each circuit shall be established using the previously
described guidelines (see T-753 and T-754). The adequacy
of the magnetic field shall be demonstrated using artificial
flaw shims or a pie-shaped magnetic particle field indica-
tor in accordance with T-764. A Hall-Effect probe

Figure T-754.2.1
Single-Pass and Two-Pass Central Conductor Technique

175
Copyright ASME International (BPVC)
Provided by IHS under license with ASME pe.class Licensee=PDVSA - Pequiven site 9/9986712014, User=Gil, Eliecer
No reproduction or networking permitted without license from IHS Not for Resale, 07/23/2013 06:23:05 MDT
ARTICLE 7 2013 SECTION V

gaussmeter shall not be used to measure field adequacy has been subjected to a major repair, periodic overhaul,
for the multidirectional magnetization technique. An ade- or damage. If equipment has not been in use for a year
quate field shall be obtained in at least two nearly perpen- or more, calibration shall be done prior to first use.
dicular directions, and the field intensities shall be
balanced so that a strong field in one direction does not
overwhelm the field in the other direction. For areas T-764 MAGNETIC FIELD ADEQUACY AND
where adequate field strengths cannot be demonstrated, DIRECTION
additional magnetic particle techniques shall be used to T-764.1 Application. The use of magnetic field indica-
obtain the required two-directional coverage. tors, artificial shims, or Hall-Effect tangential-field probes
are only permitted when specifically referenced by the fol-
lowing magnetizing techniques:
T-760 CALIBRATION (a) Longitudinal (T-753)
T-761 FREQUENCY OF CALIBRATION (b) Circular (T-754)
T-761.1 Magnetizing Equipment. (c) Multidirectional (T-756)
(a) Frequency. Magnetizing equipment with an ammeter T-764.2 Magnetic Field Adequacy. The applied mag-
shall be calibrated at least once a year, or whenever the netic field shall have sufficient strength to produce satis-
equipment has been subjected to major electric repair, factory indications, but shall not be so strong that it
periodic overhaul, or damage. If equipment has not been causes masking of relevant indications by nonrelevant ac-
in use for a year or more, calibration shall be done prior cumulations of magnetic particles. Factors that influence
to first use. the required field strength include the size, shape, and ma-
(b) Procedure. The accuracy of the unit’s meter shall be terial permeability of the part; the technique of magnetiza-
verified annually by equipment traceable to a national tion; coatings; the method of particle application; and the
standard. Comparative readings shall be taken for at least type and location of discontinuities to be detected. When it
three different current output levels encompassing the is necessary to verify the adequacy of magnetic field
usable range. strength, it shall be verified by using one or more of the

--`,,,`,`,``,``,,`,``,`,`,`,``,`-`-`,,`,,`,`,,`---
(c) Tolerance. The unit’s meter reading shall not deviate following three methods.
by more than ±10% of full scale, relative to the actual cur-
(a) Pie-Shaped Magnetic Particle Field Indicator. The in-
rent value as shown by the test meter.
dicator, shown in Figure T-764.2(a), shall be positioned on
T-761.2 Light Meters. Light meters shall be calibrated the surface to be examined, such that the copper-plated
at least once a year or whenever a meter has been re- side is away from the inspected surface. A suitable field
paired. If meters have not been in use for one year or strength is indicated when a clearly defined line (or lines)
more, calibration shall be done before being used. of magnetic particles form(s) across the copper face of the
indicator when the magnetic particles are applied simulta-
T-762 LIFTING POWER OF YOKES neously with the magnetizing force. When a clearly de-
(a) Prior to use, the magnetizing power of electromag- fined line of particles is not formed, the magnetizing
netic yokes shall have been checked within the past year. technique shall be changed as needed. Pie-type indicators
The magnetizing power of permanent magnetic yokes are best used with dry particle procedures.
shall be checked daily prior to use. The magnetizing power (b) Artificial Flaw Shims. One of the shims shown in
of all yokes shall be checked whenever the yoke has been Figure T-764.2(b)(1) or Figure T-764.2(b)(2) whose or-
damaged or repaired. ientation is such that it can have a component perpendicu-
(b) Each alternating current electromagnetic yoke shall lar to the applied magnetic field shall be used. Shims with
have a lifting power of at least 10 lb (4.5 kg) at the maxi- linear notches shall be oriented so that at least one notch
mum pole spacing that will be used. is perpendicular to the applied magnetic field. Shims with
(c) Each direct current or permanent magnetic yoke only circular notches may be used in any orientation.
shall have a lifting power of at least 40 lb (18 kg) at the Shims shall be attached to the surface to be examined,
maximum pole spacing that will be used. such that the artificial flaw side of the shim is toward
(d) Each weight shall be weighed with a scale from a re- the inspected surface. A suitable field strength is indicated
putable manufacturer and stenciled with the applicable when a clearly defined line (or lines) of magnetic particles,
nominal weight prior to first use. A weight need only be representing the 30% depth flaw, appear(s) on the shim
verified again if damaged in a manner that could have face when magnetic particles are applied simultaneously
caused potential loss of material. with the magnetizing force. When a clearly defined line
of particles is not formed, the magnetizing technique shall
T-763 GAUSSMETERS be changed as needed. Shim-type indicators are best used
Hall-Effect probe gaussmeters used to verify magnetiz- with wet particle procedures.
ing field strength in accordance with T-754 shall be cali- NOTE: The circular shims shown in Figure T-764.2(b)(2) illustration
brated at least once a year or whenever the equipment (b) also have flaw depths less and greater than 30%.

176
Copyright ASME International (BPVC)
Provided by IHS under license with ASME pe.class Licensee=PDVSA - Pequiven site 9/9986712014, User=Gil, Eliecer
No reproduction or networking permitted without license from IHS Not for Resale, 07/23/2013 06:23:05 MDT
2013 SECTION V ARTICLE 7

Figure T-764.2(a) Figure T-764.2(b)(1)


Pie-Shaped Magnetic Particle Field Indicator Artificial Flaw Shims

0.002 in.
A
(0.06 mm)
0.005 in.
0.75 in. (0.125 mm)
(20 mm) typical

A 0.0006 in.
(0.015 mm)
0.25 in.
(6 mm) Section A–A

--`,,,`,`,``,``,,`,``,`,`,`,``,`-`-`,,`,,`,`,,`---
Type B

A 0.002 in.
(0.05 mm)

0.75 in.
(20 mm)

A 0.0006 in.
(0.015 mm)
0.5 in.
(12.5 mm) Section A–A
Type C

2 in. (50 mm) 0.002 in.


Defect Division (0.05 mm)

(c) Hall-Effect Tangential-Field Probe. A gaussmeter and 0.4 in.


(10 mm)
Hall-Effect tangential-field probe shall be used for measur- 0.005 in.
ing the peak value of a tangential field. The probe shall be (0.125 mm)
positioned on the surface to be examined, such that the 0.2 in. typical
(5 mm)
maximum field strength is determined. A suitable field 0.0006 in.
strength is indicated when the measured field is within Type R (0.015 mm)
the range of 30 G to 60 G (2.4 kAm−1 to 4.8 kAm−1) while
GENERAL NOTE: Above are examples of artificial flaw shims used in
the magnetizing force is being applied. See Article 7,
magnetic particle inspection system verification (not drawn to scale).
Nonmandatory Appendix A. The shims are made of low carbon steel (1005 steel foil). The artificial
T-764.3 Magnetic Field Direction. The direction(s) of flaw is etched or machined on one side of the foil to a depth of 30% of
magnetization shall be determined by particle indications the foil thickness.
obtained using an indicator or shims as shown in
Figure T-764.2(a), Figure T-764.2(b)(1), or 0 min to ensure thorough mixing of all particles which
Figure T-764.2(b)(2). When a clearly defined line of parti- could have settled on the sump screen and along the sides
cles are not formed or bottom of the tank.
(a) in the desired direction, or
(b) in at least two nearly perpendicular directions for T-765.1 Concentration. Take a 100-mL portion of the
the multidirectional technique suspension from the hose or nozzle, demagnetize and al-
low it to settle for approximately 60 min with petroleum
the magnetizing technique shall be changed as needed.
distillate suspensions or 30 min with water-based suspen-
T-765 WET PARTICLE CONCENTRATION AND sions before reading. The volume settling out at the bot-
CONTAMINATION tom of the tube is indicative of the particle concentration
in the bath.
Wet Horizontal Units shall have the bath concentration
and bath contamination determined by measuring its set- T-765.2 Settling Volumes. For fluorescent particles,
tling volume. This is accomplished through the use of an the required settling volume is from 0.1 to 0.4 mL in a
ASTM Test Method D 96 pear-shaped centrifuge tube with 100 mL bath sample and from 1.2 to 2.4 mL per 100 mL
a 1-mL stem (0.05-mL divisions) for fluorescent particle of vehicle for nonfluorescent particles unless otherwise
suspensions or a 1.5-mL stem (0.1-mL divisions) for non- specified by the particle manufacturer. Concentration
fluorescent suspensions. Before sampling, the suspension checks shall be made at least every eight hours.
should be run through the recirculating system for at least

177
Copyright ASME International (BPVC)
Provided by IHS under license with ASME pe.class Licensee=PDVSA - Pequiven site 9/9986712014, User=Gil, Eliecer
No reproduction or networking permitted without license from IHS Not for Resale, 07/23/2013 06:23:05 MDT
ARTICLE 7 2013 SECTION V

Figure T-764.2(b)(2)
Artificial Flaw Shims

0.75 in. (typ) (19.05 mm) 0.75 in. (typ) (19.05 mm)
0.25 in. 0.25 in.
(6.36 mm) (6.36 mm)

0.507 in. diam. O.D. 0.507 in. diam. O.D.


(12.88 mm) (12.88 mm)

0.007 in. (typ) 0.007 in. (typ)


(0.18 mm) (0.18 mm)
Notches: Notches:
Depth: 30% 0.0006 in. Depth: 30% 0.0012 in.
(0.015 mm) (0.030 mm)
230 Shim thickness: 430 Shim thickness:
0.002 in. (0.05 mm) 0.004 in. (0.10 mm)
Shim Type CX-230 Shim Type CX-430
(a)

0.75 in. (typ) (19.05 mm) 0.75 in. (typ) (19.05 mm)
0.007 in. (type) 0.007 in. (type)
(0.18 mm) 0.507 in. diam. O.D. (0.18 mm) 0.507 in. diam. O.D.
(12.88 mm) (12.88 mm)

0.383 in. diam. O.D. 0.383 in. diam. O.D.


(9.73 mm) (9.73 mm)
0.258 in. diam. O.D. 0.258 in. diam. O.D.
(6.55 mm) (6.55 mm)

Notch depth: Notch depth:


20% 0.0004 in. 20% 0.0004 in.
(0.010 mm) O.D. (0.010 mm) O.D.
4-234 30% 0.0006 in. 2-234 30% 0.0006 in.
(0.015 mm) center (0.015 mm) center
Shim Type 3C4-234 40% 0.0008 in. Shim Type 3C2-234 40% 0.0008 in.
Shim Thickness 0.004 in. (0.102 mm) (0.020 mm) I.D. Shim Thickness 0.002 in. (0.05 mm) (0.020 mm) I.D.
(b)

0.79 in. (typ) (20.06 mm) 0.79 in. (typ) (20.06 mm)
0.235 in. (typ) 0.235 in. (typ)
(5.97 mm) (5.97 mm)

0.255 in. diam. O.D. 0.255 in. diam.


(6.48 mm) O.D. (6.48 mm)

0.006 in. (typ) 0.006 in. (typ)


(0.152 mm) (0.152 mm)
0.395 in. (typ) 0.20 in. (typ) 0.395 in. (typ) 0.20 in. (typ)
(10.03 mm) (5.08 mm) (10.03 mm) (5.08 mm)

Notch depth: Notch depth:


30% 0.0006 in. 30% 0.0012 in.
(0.015 mm) (0.030 mm)

230 430

Shim Thickness 0.002 in. (0.051 mm) Shim Thickness 0.004 in. (0.102 mm)
Shim Type CX-230 Shim Type CX4-430
(c)

178
--`,,,`,`,``,``,,`,``,`,`

Copyright ASME International (BPVC)


Provided by IHS under license with ASME pe.class Licensee=PDVSA - Pequiven site 9/9986712014, User=Gil, Eliecer
No reproduction or networking permitted without license from IHS Not for Resale, 07/23/2013 06:23:05 MDT
2013 SECTION V ARTICLE 7

T-765.3 Contamination. Both fluorescent and non- If the test does not reveal the required number of holes,
fluorescent suspensions shall be checked periodically for the equipment shall be taken out of service and the cause
contaminants such as dirt, scale, oil, lint, loose fluorescent of the loss of sensitivity determined and corrected. This
pigment, water (in the case of oil suspensions), and parti- test shall be run at least once per week.
cle agglomerates which can adversely affect the perfor-
mance of the magnetic particle examination process. The
test for contamination shall be performed at least once T-770 EXAMINATION
per week.
T-771 PRELIMINARY EXAMINATION
(a) Carrier Contamination. For fluorescent baths, the li-
quid directly above the precipitate should be examined Before the magnetic particle examination is conducted,
with fluorescent excitation light. The liquid will have a lit- a check of the examination surface shall be conducted to
tle fluorescence. Its color can be compared with a freshly locate any discontinuity surface openings which may not
made-up sample using the same materials or with an un- attract and hold magnetic particles because of their width.
used sample from the original bath that was retained for
this purpose. If the “used” sample is noticeably more fluor- T-772 DIRECTION OF MAGNETIZATION
escent than the comparison standard, the bath shall be
At least two separate examinations shall be performed
replaced.
on each area. During the second examination, the lines
(b) Particle Contamination. The graduated portion of the
of magnetic flux shall be approximately perpendicular to
tube shall be examined under fluorescent excitation light if
those used during the first examination. A different techni-
the bath is fluorescent and under visible light (for both
que for magnetization may be used for the second
fluorescent and nonfluorescent particles) for striations
examination.
or bands, differences in color or appearance. Bands or
striations may indicate contamination. If the total volume
of the contaminates, including bands or striations exceeds T-773 METHOD OF EXAMINATION
30% of the volume magnetic particles, or if the liquid is The ferromagnetic particles used in an examination
noticeably fluorescent, the bath shall be replaced. medium can be either wet or dry, and may be either fluor-
escent or nonfluorescent. Examination(s) shall be done by
T-766 SYSTEM PERFORMANCE OF HORIZONTAL the continuous method.
UNITS (a) Dry Particles. The magnetizing current shall remain
The Ketos (Betz) ring specimen (see Figure T-766.1) on while the examination medium is being applied and
shall be used in evaluating and comparing the overall per- while any excess of the examination medium is removed.
--`,,,`,`,``,``,,`,``,`,`,`,``,`-`-`,,`,,`,`,,`---

formance and sensitivity of both dry and wet, fluorescent (b) Wet Particles. The magnetizing current shall be
and nonfluorescent magnetic particle techniques using a turned on after the particles have been applied. Flow of
central conductor magnetization technique. particles shall stop with the application of current. Wet
(a) Ketos (Betz) Test Ring Material. The tool steel (Ke- particles applied from aerosol spray cans may be applied
tos) ring should be machined from AISI 01 material in ac- before and/or after magnetizing current is applied. Wet
cordance with Figure T-766.1. Either the machined ring or particles may be applied during the application of magne-
the steel blank should be annealed at 1,650°F (900°C), tizing current if they are not applied directly to the exam-
cooled 50°F (28°C) per hour to 1,000°F (540°C) and then ination area and are allowed to flow over the examination
air cooled to ambient temperature to give comparable re- area or are applied directly to the examination area with
sults using similar rings that have had the same treatment. low velocities insufficient to remove accumulated
Material and heat treatment are important variables. Ex- particles.
perience indicates controlling the softness of the ring by
hardness (90 to 95 HRB) alone is insufficient. T-774 EXAMINATION COVERAGE
(b) U s i n g t h e T e s t R i n g . T h e t e s t r i n g ( s e e All examinations shall be conducted with sufficient field
Figure T-766.1), is circularly magnetized with full-wave overlap to ensure 100% coverage at the required sensitiv-
rectified AC passing through a central conductor with a ity (T-764).
1 in. to 11/4 in. (25 mm to 32 mm) diameter hole located
in the ring center. The conductor should have a length
T-775 RECTIFIED CURRENT
greater than 16 in. (400 mm). The currents used shall be
1400, 2500, and 3400 amps. The minimum number of (a) Whenever direct current is required rectified cur-
holes shown shall be three, five, and six, respectively. rent may be used. The rectified current for magnetization
The ring edge should be examined with either black light shall be either three-phase (full-wave rectified) current, or
or visible light, depending on the type of particles in- single phase (half-wave rectified) current.
volved. This test shall be run at the three amperages if (b) The amperage required with three-phase, full-wave
the unit will be used at these or higher amperages. The rectified current shall be verified by measuring the aver-
amperage values stated shall not be exceeded in the test. age current.

179
Copyright ASME International (BPVC)
Provided by IHS under license with ASME pe.class Licensee=PDVSA - Pequiven site 9/9986712014, User=Gil, Eliecer
No reproduction or networking permitted without license from IHS Not for Resale, 07/23/2013 06:23:05 MDT
ARTICLE 7 2013 SECTION V

Figure T-766.1
Ketos (Betz) Test Ring

3/ in. (19 mm)


4
Typ.
125
1 2
3
4
5
11/4 in. 6 5 in.
(32 mm) (125
7 mm)
8
12 9
D 11 10

7/ in.
8
(22 mm)

Hole 1 2 3 4 5 6 7 8 9 10 11 12
Diame- 0.07 0.07 0.07 0.07 0.07 0.07 0.07 0.07 0.07 0.07 0.07 0.07
ter (1.8 (1.8 (1.8 (1.8 (1.8 (1.8 (1.8 (1.8 (1.8 (1.8 (1.8 (1.8
[Note mm) mm) mm) mm) mm) mm) mm) mm) mm) mm) mm) mm)
(1)]
“D” 0.07 0.14 0.21 0.28 0.35 0.42 0.49 0.56 0.63 0.70 0.77 0.84
[Note (1.8 (3.6 (5.3 (7.1 (9.0 (10.8 (12.6 (14.4 (16.2 (18.0 (19.8 (21.6
(2)] mm) mm) mm) mm) mm) mm) mm) mm) mm) mm) mm) mm)

NOTES:
(1) All hole diameters are ±0.005 in. (±0.1 mm.) Hole numbers 8 through 12 are optional.
(2) Tolerance on the D distance is ±0.005 in. (±0.1 mm).
GENERAL NOTES:
(a) All dimensions are ±0.03 in. (±0.8 mm) or as noted in Notes (1) and (2).
(b) All dimensions are in inches, except as noted.
(c) Material is ANSI 01 tool steel from annealed round stock.
(d) The ring may be heat treated as follows: Heat to 1400°F to 1500°F (760°C to 790°C). Hold at this temperature for one hour. Cool to a mini-
mum rate of 40°F/h (22°C/h) to below 1000°F (540°C). Furnace or air cool to room temperature. Finish the ring to RMS 25 and protect
from corrosion.

(c) The amperage required with single-phase (half- T-777 INTERPRETATION


wave rectified) current shall be verified by measuring The interpretation shall identify if an indication as false,
the average current output during the conducting half cy- nonrelevant, or relevant. False and nonrelevant indica-
cle only. tions shall be proven as false or nonrelevant. Interpreta-
(d) When measuring half-wave rectified current with a tion shall be carried out to identify the locations of
direct current test meter, readings shall be multiplied by indications and the character of the indication.
two.
T-777.1 Visible (Color Contrast) Magnetic Particles.
T-776 EXCESS PARTICLE REMOVAL Surface discontinuities are indicated by accumulations of
Accumulations of excess dry particles in examinations magnetic particles which should contrast with the exami-
shall be removed with a light air stream from a bulb or nation surface. The color of the magnetic particles shall be
syringe or other source of low pressure dry air. The exam- sufficiently different than the color of the examination sur-
ination current or power shall be maintained while remov- face. A minimum light intensity of 100 fc (1000 Lx) is re-
ing the excess particles. quired on the surface to be examined to ensure adequate
sensitivity during the examination and evaluation of

180
Copyright ASME International (BPVC)
--`,,,`,`,``,``,

Provided by IHS under license with ASME pe.class Licensee=PDVSA - Pequiven site 9/9986712014, User=Gil, Eliecer
No reproduction or networking permitted without license from IHS Not for Resale, 07/23/2013 06:23:05 MDT
2013 SECTION V ARTICLE 7

indications. The light source, technique used, and light le- (f) The fluorescence excitation light intensity shall be
vel verification is required to be demonstrated one time, measured with a suitable fluorescence excitation light me-
documented, and maintained on file. ter prior to use, whenever the light’s power source is in-
T-777.2 Fluorescent Magnetic Particles With Black terrupted or changed, and at the completion of the
Light. With fluorescent magnetic particles, the process is examination or series of examinations.
essentially the same as in T-777.1, with the exception that
the examination is performed using an ultraviolet light T-778 DEMAGNETIZATION
(i.e., nominal 365 nm), called black light. The examination
When residual magnetism in the part could interfere
shall be performed as follows:
with subsequent processing or usage, the part shall be de-
(a) It shall be performed in a darkened area.
magnetized any time after completion of the examination.
(b) Examiners shall be in a darkened area for at least
5 min prior to performing examinations to enable their
eyes to adapt to dark viewing. Glasses or lenses worn by T-779 POST-EXAMINATION CLEANING
examiners shall not be photochromic or exhibit any
When post-examination cleaning is required, it should
fluorescence.
be conducted as soon as practical using a process that
(c) Black lights shall achieve a minimum of 1000
--`,,,`,`,``,``,,`,``,`,`,`,``,`-`-`,,`,,`,`,,`---

does not adversely affect the part.


μW/cm 2 on the surface of the part being examined
throughout the examination.
(d) Reflectors, filters, glasses, and lenses should be
checked and, if necessary, cleaned prior to use. Cracked T-780 EVALUATION
or broken reflectors, filters, glasses, or lenses shall be re-
(a) All indications shall be evaluated in terms of the ac-
placed immediately.
ceptance standards of the referencing Code Section.
(e) The black light intensity shall be measured with a
(b) Discontinuities on or near the surface are indicated
black light meter prior to use, whenever the light’s power
by retention of the examination medium. However, loca-
source is interrupted or changed, and at the completion of
lized surface irregularities due to machining marks or
the examination or series of examinations.
other surface conditions may produce false indications.
T-777.3 Fluorescent Magnetic Particles With Other (c) Broad areas of particle accumulation, which might
Fluorescent Excitation Wavelengths. Alternatively to mask indications from discontinuities, are prohibited,
the requirements in T-777.2, the examinations may be and such areas shall be cleaned and reexamined.
performed using alternate wavelength light sources which
cause fluorescence in specific particle coatings. Any alter-
nate light wavelength light sources and specific particle
designations used shall be qualified30 in accordance with T-790 DOCUMENTATION
Mandatory Appendix IV. The examination shall be per- T-791 MULTIDIRECTIONAL MAGNETIZATION
formed as follows: TECHNIQUE SKETCH
(a) It shall be performed in a darkened area.
(b) Examiners shall be in a darkened area for at least A technique sketch shall be prepared for each different
5 min prior to performing examinations to enable their geometry examined, showing the part geometry, cable ar-
eyes to adapt to dark viewing. Glasses or lenses worn by rangement and connections, magnetizing current for each
examiners shall not be photochromic or exhibit any circuit, and the areas of examination where adequate field
fluorescence. strengths are obtained. Parts with repetitive geometries,
(c) If the fluorescence excitation light source emits visi- but different dimensions, may be examined using a single
ble light intensities greater than 2 fc (20 Lx), the examiner sketch provided that the magnetic field strength is ade-
shall wear fluorescence-enhancing filter glasses approved quate when demonstrated in accordance with T-755.2.
by the light source manufacturer for use with that light
source. T-792 RECORDING OF INDICATIONS
(d) The fluorescence excitation light source shall
T-792.1 Nonrejectable Indications. Nonrejectable in-
achieve at least the minimum light intensity on the surface
dications shall be recorded as specified by the referencing
of the part throughout the examination as qualified in the
Code Section.
tests of Mandatory Appendix IV.
(e) Reflectors, filters, glasses, and lenses should be T-792.2 Rejectable Indications. Rejectable indica-
checked and, if necessary, cleaned prior to use. Cracked tions shall be recorded. As a minimum, the type of indica-
or broken reflectors, filters, glasses, or lenses shall be re- tions (linear or rounded), location and extent (length or
placed immediately. diameter or aligned) shall be recorded.

181
Copyright ASME International (BPVC)
Provided by IHS under license with ASME pe.class Licensee=PDVSA - Pequiven site 9/9986712014, User=Gil, Eliecer
No reproduction or networking permitted without license from IHS Not for Resale, 07/23/2013 06:23:05 MDT
ARTICLE 7 2013 SECTION V

T-793 EXAMINATION RECORDS (d) examination personnel identity and if required by


For each examination, the following information shall referencing Code Section, qualification level
be recorded: (e) map or record of indications per T-792
(a) procedure identification and revision (f) material and thickness
(b) magnetic particle equipment and type of current (g) lighting equipment
(c) magnetic particles (visible or fluorescent, wet or (h) date of examination
dry)
--`,,,`,`,``,``,,`,``,`,`,`,``,`-`-`,,`,,`,`,,`---

182
Copyright ASME International (BPVC)
Provided by IHS under license with ASME pe.class Licensee=PDVSA - Pequiven site 9/9986712014, User=Gil, Eliecer
No reproduction or networking permitted without license from IHS Not for Resale, 07/23/2013 06:23:05 MDT
2013 SECTION V ARTICLE 7

MANDATORY APPENDIX I
MAGNETIC PARTICLE EXAMINATION USING THE AC YOKE
TECHNIQUE ON FERRITIC MATERIALS COATED WITH
NONMAGNETIC COATINGS

I-710 SCOPE I-721.2 Procedure Qualification/Technique Valida-


tion. When procedure qualification is specified, a change
This Appendix provides the Magnetic Particle examina- of a requirement in Table T-721 or Table I-721 identified
tion methodology and equipment requirements applicable as an essential variable from the specified value, or range
for performing Magnetic Particle examination on ferritic of values, shall require requalification of the written pro-
materials with nonmagnetic coatings. cedure and validation of the technique. A change of a re-
quirement identified as an nonessential variable from
the specified value, or range of values, does not require re-
qualification of the written procedure. All changes of es-
I-720 GENERAL sential or nonessential variables from the value, or range
of values, specified by the written procedure shall require
Requirements of Article 7 apply unless modified by this
revision of, or an addendum to, the written procedure.
Appendix.
I-722 PERSONNEL QUALIFICATION
Personnel qualification requirements shall be in accor-
I-721 WRITTEN PROCEDURE REQUIREMENTS dance with the referencing Code Section.
I-721.1 Requirements. Magnetic Particle examination
shall be performed in accordance with a written proce-
dure which shall, as a minimum, contain the requirements
listed in Tables T-721 and I-721. The written procedure
shall establish a single value, or range of values, for each
requirement.

Table I-721 ð13Þ


Requirements of AC Yoke Technique on Coated Ferritic Component
Nonessential
Requirement Essential Variable Variable
Identification of surface configurations to be examined, including coating materials, maximum X …
qualified coating thickness, and product forms (e.g., base material or welded surface)
Surface condition requirements and preparation methods X …
Manufacturer and model of AC yoke X …
Manufacturer and type of magnetic particles X …
Minimum and maximum pole separation X …
Identification of the steps in performing the examination X …
Minimum lighting intensity and AC yoke lifting power requirements [as measured in accordance with X …
Technique Qualification (I-721.2)]
Methods of identifying flaw indications and discriminating between flaw indications and false or X …
nonrelevant indications (e.g., magnetic writing or particles held by surface irregularities)
Instructions for identification and confirmation of suspected flaw indications X …
Applicator other than powder blower X …
Method of measuring coating thickness … X
Recording criteria … X
Personnel qualification requirements unique to this technique … X
Reference to the procedure qualification records … X

--`,,,`,`,``,``,,`,``,`,`,`,``,`-`-`,,`,,`,`,,`---
183
Copyright ASME International (BPVC)
Provided by IHS under license with ASME pe.class Licensee=PDVSA - Pequiven site 9/9986712014, User=Gil, Eliecer
No reproduction or networking permitted without license from IHS Not for Resale, 07/23/2013 06:23:05 MDT
ARTICLE 7 2013 SECTION V

I-723 PROCEDURE/TECHNIQUE The material used for the specimen shall be the same spe-
DEMONSTRATION cification and heat treatment as the coated ferromagnetic
material to be examined. As an alternative to the material
The procedure/technique shall be demonstrated to the
requirement, other materials and heat treatments may be
satisfaction of the Inspector in accordance with the re-
qualified provided:
quirements of the referencing Code Section.
(1) The measured yoke maximum lifting force on the
material to be examined is equal to or greater than the
I-730 EQUIPMENT maximum lifting force on the qualification specimen mate-
rial. Both values shall be determined with the same or
I-730.1 The magnetizing equipment shall be in accor- comparable equipment and shall be documented as re-
dance with Article 7. quired in (c).
ð13Þ I-730.2 When the dry powder technique is used, a (2) All the requirements of (b) through (g) are met
compressed air powder blower shall be utilized for pow- for the alternate material.
der application in any position. Other applicators may be (b) Examine the uncoated specimen in the most unfa-
used if qualified in the same surface position as the exam- vorable orientation expected during the performance of
ination object surface. Applicators qualified for the over- the production examination.
head position may be used in any other position. (c) Document the measured yoke maximum lifting
Applicators qualified for the vertical position may be used power, illumination levels, and the results.
in the horizontal and flat positions.
(d) Measure the maximum coating thickness on the item
I-730.3 Magnetic particles shall contrast with the to be examined in accordance with the requirements of
component background. I-741.
I-730.4 Nonconductive materials such as plastic shim (e) Coat the specimen with the same type of coating,
stock may be used to simulate nonconductive nonmag- conductive or nonconductive, to the maximum thickness
netic coatings for procedure and personnel qualification. measured on the production item to be examined. Alter-
nately, nonconductive shim stock may be used to simulate
nonconductive coatings.
I-740 MISCELLANEOUS REQUIREMENTS (f) Examine the coated specimen in the most unfavor-
able orientation expected during the performance of the
I-741 COATING THICKNESS MEASUREMENT
production examination. Document the measured yoke
The procedure demonstration and performance of ex- maximum lifting power, illumination level, and examina-
aminations shall be preceded by measurement of the coat- tion results.
ing thickness in the areas to be examined. If the coating is (g) Compare the length of the indication resulting from
nonconductive, an eddy current technique or magnetic the longest flaw no longer than the maximum flaw size al-
technique may be used to measure the coating thickness. lowed by the applicable acceptance criteria, before and
--`,,,`,`,``,``,,`,``,`,`,`,``,`-`-`,,`,,`,`,,`---

The magnetic technique shall be in accordance with after coating. The coating thickness is qualified when the
SD-1186, Standard Test Methods for Nondestructive Mea- length of the indication on the coated surface is at least
surement of Dry Film Thickness of Nonmagnetic Coatings 50% of the length of the corresponding indication prior
Applied to a Ferrous Base. When coatings are conductive to coating.
and nonmagnetic, a coating thickness technique shall be (h) Requalification of the procedure is required for a de-
used in accordance with SD-1186. Coating measurement crease in either the AC yoke lifting power or the illumina-
equipment shall be used in accordance with the equip- tion level, or for an increase in the coating thickness.
ment manufacturer’s instructions. Coating thickness mea-
surements shall be taken at the intersections of a 2 in.
(50 mm) maximum grid pattern over the area of examina-
tion and at least one-half the maximum yoke leg separa- I-760 CALIBRATION
tion beyond the examination area. The thickness shall be I-761 YOKE MAXIMUM LIFTING FORCE
the mean of three separate readings within 1/4 in.
(6 mm) of each intersection. The maximum lifting force of the AC yoke shall be deter-
mined at the actual leg separation to be used in the exam-
ination. This may be accomplished by holding the yoke
I-750 TECHNIQUE with a 10 lb (4.5 kg) ferromagnetic weight between the
legs of the yoke and adding additional weights, calibrated
I-751 TECHNIQUE QUALIFICATION on a postage or other scale, until the ferromagnetic weight
(a) A qualification specimen is required. The specimen is released. The lifting power of the yoke shall be the com-
shall be of similar geometry or weld profile and contain bined weight of the ferromagnetic material and the added
at least one surface crack no longer than the maximum weights, before the ferromagnetic weight was released.
flaw size allowed in the applicable acceptance criteria. Other methods may be used such as a load cell.

184
Copyright ASME International (BPVC)
Provided by IHS under license with ASME pe.class Licensee=PDVSA - Pequiven site 9/9986712014, User=Gil, Eliecer
No reproduction or networking permitted without license from IHS Not for Resale, 07/23/2013 06:23:05 MDT
2013 SECTION V ARTICLE 7

I-762 LIGHT INTENSITY MEASUREMENT (b) Examine the coated item in accordance with the qua-
The black light or white light intensity (as appropriate) lified procedure.
on the surface of the component shall be no less than that
used in the qualification test. An appropriate calibrated
black light and/or white light meter shall be used for the I-780 EVALUATION
tests. Minimum white light or black light intensities shall
meet the requirements of T-777.1 or T-777.2 as If an indication greater than 50% of the maximum al-
applicable. lowable flaw size is detected, the coating in the area of
the indication shall be removed and the examination
I-762.1 White Light. The white light intensity shall be repeated.
measured at the inspection surface. The white light inten-
sity for the examination shall be no less than what was
used in the qualification.
I-790 DOCUMENTATION
I-762.2 Black Light. The black light intensity shall be
measured at the distance from the black light in the proce- I-791 EXAMINATION RECORD
dure qualification and at the same distance on the exami- For each examination, the information required in the
nation specimen. The black light intensity shall be no less records section of T-793 and the following information
than that used to qualify the procedure. In addition, the shall be recorded:
maximum white light intensity shall be measured as back-
(a) identification of the procedure/technique
ground light on the inspection surface. The background
white light for the examination shall be no greater than (b) identification of the personnel performing and wit-
what was used in the qualification. nessing the qualification
(c) description and drawings or sketches of the qualifi-
cation specimen, including coating thickness measure-
I-770 EXAMINATION ments and flaw dimensions
(a) Surfaces to be examined, and all adjacent areas with- (d) equipment and materials used
in at least 1 in. (25 mm), shall be free of all dirt, grease, lint, (e) illumination level and yoke lifting power
scale, welding flux and spatter, oil, and loose, blistered, (f) qualification results, including maximum coating
flaking, or peeling coating. thickness and flaws detected.

185
--`,,,`,`,``,``,,`,``,`,`,`,``,`-`-`,,`,,`,`,,`---

Copyright ASME International (BPVC)


Provided by IHS under license with ASME pe.class Licensee=PDVSA - Pequiven site 9/9986712014, User=Gil, Eliecer
No reproduction or networking permitted without license from IHS Not for Resale, 07/23/2013 06:23:05 MDT
ARTICLE 7 2013 SECTION V

MANDATORY APPENDIX II
GLOSSARY OF TERMS FOR MAGNETIC PARTICLE EXAMINATION

II-710 SCOPE black light intensity: a quantitative expression of ultravio-


let irradiance.
This Mandatory Appendix is used for the purpose of es-
tablishing standard terms and definition of terms which magnetic flux: the concept that the magnetic field is flow-
appear in Article 7, Magnetic Particle Examination. ing along the lines of force suggests that these lines are
therefore “flux” lines, and they are called magnetic flux.
II-720 GENERAL REQUIREMENTS The strength of the field is defined by the number of flux
lines crossing a unit area taken at right angles to the direc-
(a) The Standard Terminology for Nondestructive Ex- tion of the lines.
aminations (ASTM E1316) has been adopted by the Com-
mittee as SE-1316. rectified magnetic current: by means of a device called a
(b) SE-1316 Section 7 provides the definitions of terms rectifier, which permits current to flow in one direction
listed in II-730(a). only, alternating current can be converted to unidirec-
(c) For general terms, such as Indication, Flaw, tional current. This differs from direct current in that
Discontinuity, Evaluation, etc., refer to Article 1, the current value varies from a steady level. This variation
Mandatory Appendix I. may be extreme, as in the case of the half-wave rectified
(d) Paragraph II-730(b) provides a list of terms and de- single phase AC, or slight, as in the case of three-phase rec-
finitions, which are in addition to SE-1316 and are Code tified AC.
specific.
half-wave rectified current AC: when a single-phase alter-
nating current is rectified in the simplest manner, the re-
II-730 REQUIREMENTS
verse of the cycle is blocked out entirely. The result is a
(a) The following SE-1316 terms are used in conjunc- pulsating unidirectional current with intervals when no
tion with this Article: ampere turns, black light, central current at all is flowing. This is often referred to as “half-
conductor, circular magnetization, demagnetization, dry wave” or pulsating direct current.
powder, full-wave direct current, half-wave current, long-
itudinal magnetization, magnetic field, magnetic field full-wave rectified current: when the reverse half of the cy-
strength, magnetic particle examination, magnetic particle cle is turned around to flow in the same direction as the
field indicator, magnetic particles, multidirectional magne- forward half. The result is full-wave rectified current.
tization, permanent magnet, prods, sensitivity, suspension, Three-phase alternating current when full-wave rectified
yoke. is unidirectional with very little pulsation; only a ripple
(b) The following Code terms are used in conjunction of varying voltage distinguishes it from straight DC
with this Article: single-phase.

186
--`,,,`,`,``,``,,`,``,`,`,`,``,`-`-`,,`,,`,`,,`---

Copyright ASME International (BPVC)


Provided by IHS under license with ASME pe.class Licensee=PDVSA - Pequiven site 9/9986712014, User=Gil, Eliecer
No reproduction or networking permitted without license from IHS Not for Resale, 07/23/2013 06:23:05 MDT
2013 SECTION V ARTICLE 7

MANDATORY APPENDIX III


MAGNETIC PARTICLE EXAMINATION USING THE YOKE
TECHNIQUE WITH FLUORESCENT PARTICLES IN AN
UNDARKENED AREA

III-710 SCOPE III-750 TECHNIQUE


This Appendix provides the Magnetic Particle examina- III-751 QUALIFICATION STANDARD
tion methodology and equipment requirements applicable A standard slotted shim(s) as described in T-764.1.2
for performing Magnetic Particle examinations using a shall be used as the qualification standard.
yoke with fluorescent particles in an undarkened area.

III-760 CALIBRATION
III-761 BLACK LIGHT INTENSITY MEASUREMENT
III-720 GENERAL
The black light intensity on the surface of the compo-
Requirements of Article 7 apply unless modified by this nent shall be no less than that used in the qualification
Appendix. test.

III-762 WHITE LIGHT INTENSITY MEASUREMENT


III-721 WRITTEN PROCEDURE REQUIREMENTS The white light intensity on the surface of the compo-
III-721.1 Requirements. The requirements of Tables nent shall be no greater than that used in the qualification
T-721 and III-721 apply. test.

III-721.2 Procedure Qualification. The requirements


of Tables T-721 and III-721 apply. III-770 EXAMINATION
The qualification standard shall be placed on a carbon
steel plate and examined in accordance with the proce-
III-723 PROCEDURE DEMONSTRATION
dure to be qualified and a standard procedure that has
The procedure shall be demonstrated to the satisfaction previously been demonstrated as suitable for use. The
of the Inspector in accordance with the requirements of standard procedure may utilize a visible or fluorescent
the referencing Code Section. technique. The flaw indications shall be compared; if the

Table III-721
Requirements for an AC or HWDC Yoke Technique With Fluorescent Particles in an Undarkened Area
Nonessential
Requirement Essential Variable Variable
Identification of surface configurations to be examined and product forms (e.g., base material or X ...
welded surface)
Surface condition requirement and preparation methods X . ..
Yoke manufacturer and model X . ..
Particle manufacturer and designation X . ..
Minimum and maximum pole separation X . ..
Identification of steps in performing the examination X . ..
Maximum white light intensity X . ..
Minimum black light intensity X . ..
Personnel qualification requirements ... X
Reference to the procedure qualification records ... X

--`,,,`,`,``,``,,`,``,`,`,`,``,`-`-`,,`,,`,`,,`---
187
Copyright ASME International (BPVC)
Provided by IHS under license with ASME pe.class Licensee=PDVSA - Pequiven site 9/9986712014, User=Gil, Eliecer
No reproduction or networking permitted without license from IHS Not for Resale, 07/23/2013 06:23:05 MDT
ARTICLE 7 2013 SECTION V

indication obtained under the proposed conditions ap- III-790 DOCUMENTATION


pears the same or better than that obtained under stan-
dard conditions, the proposed procedure shall be III-791 EXAMINATION RECORD
considered qualified for use. For each examination, the information required in
T-793 and the following information shall be recorded:
III-777 INTERPRETATION (a) procedure identification and revision qualified
For interpretation, both black and white light intensity (b) standard procedure identification and revision
shall be measured with light meters. (c) qualification standard identification
(d) identification of the personnel performing and wit-
nessing the qualification
(e) equipment and materials used
(f) illumination levels (white and black light)
(g) qualification results
--`,,,`,`,``,``,,`,``,`,`,`,``,`-`-`,,`,,`,`,,`---

188
Copyright ASME International (BPVC)
Provided by IHS under license with ASME pe.class Licensee=PDVSA - Pequiven site 9/9986712014, User=Gil, Eliecer
No reproduction or networking permitted without license from IHS Not for Resale, 07/23/2013 06:23:05 MDT
2013 SECTION V ARTICLE 7

MANDATORY APPENDIX IV
QUALIFICATION OF ALTERNATE WAVELENGTH LIGHT SOURCES
FOR EXCITATION OF FLUORESCENT PARTICLES

IV-710 SCOPE and specific particles. Shim(s) shall be tape sealed to a fer-
romagnetic object’s surface and used as described in
This Appendix provides the methodology to qualify the T-764.1.2 with the notch against the object’s surface.
performance of fluorescent particle examinations using al-
ternate wavelength sources.
IV-752 FILTER GLASSES
If the alternative wavelength light source emits light in
IV-720 GENERAL the visible portion of the spectrum (wavelength of 400 nm
or longer), the examiner shall wear filter glasses that have
Requirements of Article 7 apply unless modified by this been supplied by the manufacturer of the light source to
Appendix. block the reflected visible excitation light while transmit-
ting the fluorescence of the particles.
IV-721 WRITTEN PROCEDURE REQUIREMENTS
IV-721.1 Requirements. The requirements of
Table IV-721 apply to Written Procedure Requirements IV-770 QUALIFICATION EXAMINATIONS
(T-721.1) and when specified by the referencing Code Sec-
tion to Procedure Qualification (T-721.2). IV-771 BLACK LIGHT INTENSITY
The black light intensity on the examination surface
IV-723 PROCEDURE DEMONSTRATION shall be adjusted by varying the distance or power so that
it has a minimum intensity of 1,000 μW/cm2 and a maxi-
The procedure shall be demonstrated to the satisfaction mum intensity of 1,100 μW/cm2.
of the Inspector in accordance with the requirements of
the referencing Code Section.
IV-772 EXAMINATION REQUIREMENTS
The examination parameters for the object chosen shall
IV-750 TECHNIQUE be determined by the rules of T-750 applicable to the ob-
ject chosen and the method of magnetization. Any of the
IV-751 QUALIFICATION STANDARD magnetizing techniques listed in T-751 may be used. The
Slotted shim(s) 0.002 in. (0.05 mm) thick having 30% same indication(s) of the shim discontinuity(ies) shall be
deep material removed as described in T-764.1.2 shall used for both black light and alternate wavelength light
be used to qualify the alternate wavelength light source examinations.

Table IV-721
Requirements for Qualifying Alternate Wavelength Light Sources for Excitation of Specific
Fluorescent Particles
Nonessential
Requirement Essential Variable Variable
Particle manufacturer and designation X . ..
Carrier (water or oil); if oil, manufacturer and type designation X . ..
Alternate wavelength light source manufacturer and model X . ..
Alternate wavelength light source meter, manufacturer, and model X . ..
Filter glasses (if needed) X . ..
Minimum alternative wavelength light intensity X . ..
Qualification records ... X

--`,,,`,`,``,``,,`,``,`,`,`,``,`-`-`,,`,,`,`,,`---
189
Copyright ASME International (BPVC)
Provided by IHS under license with ASME pe.class Licensee=PDVSA - Pequiven site 9/9986712014, User=Gil, Eliecer
No reproduction or networking permitted without license from IHS Not for Resale, 07/23/2013 06:23:05 MDT
ARTICLE 7 2013 SECTION V

IV-772.1 Examination With Black Light. The qualifi- used for magnetic particle examinations. The alternate wa-
cation standard with the attached shim(s) shall be exam- velength light source with at least the minimum intensity
ined with the established parameters and specific qualified shall be used with the specific particle designa-
particles in a darkened area with black light illumination. tion employed in the qualification.
The resulting particle indication(s) shall be photographed.
IV-772.2 Examination With Alternate Wavelength
Light. Using the same particle indication(s) examined in IV-790 DOCUMENTATION
IV-772.1, switch to the alternate wavelength light source IV-791 EXAMINATION RECORD
and adjust the light intensity by varying the distance or
For each examination, the information required in
power, to establish particle indication(s) essentially the
T-793 and the following information shall be recorded:
same as that (those) obtained with the black light above.
The light intensity shall be measured with the alternative (a) alternative wavelength light source, manufacturer,
wavelength light meter. The resulting particle and model
indication(s) shall be photographed using identical photo- (b) alternative wavelength light source meter, manufac-
graphic techniques as used for the black light. However, turer, and model
camera lens filters appropriate for use with the alternate (c) filter glasses, when necessary
wavelength light source should be used for recording (d) fluorescent particle manufacturer and designation
the indication(s), when required. (e) qualification standard identification
(f) technique details
IV-773 QUALIFICATION OF ALTERNATE (g) identification of the personnel performing and wit-
WAVELENGTH LIGHT SOURCE AND nessing the qualification
SPECIFIC PARTICLES (h) equipment and materials used
When the same particle indication(s) as achieved with (i) minimum alternate wavelength light intensity
black light can be obtained with the alternate wavelength (j) black light and alternative wavelength light qualifi-
light source, the alternate wavelength light source may be cation photos, exposure settings, and filters, if used

190 --`,,,`,`,``,``,,`,``,`,`,`,``,`-`-`,,`,,`,`,,`---

Copyright ASME International (BPVC)


Provided by IHS under license with ASME pe.class Licensee=PDVSA - Pequiven site 9/9986712014, User=Gil, Eliecer
No reproduction or networking permitted without license from IHS Not for Resale, 07/23/2013 06:23:05 MDT
2013 SECTION V ARTICLE 7

ð13Þ MANDATORY APPENDIX V


REQUIREMENTS FOR THE USE OF MAGNETIC RUBBER
TECHNIQUES

V-710 SCOPE V-730 EQUIPMENT


This Appendix provides the methodology and equip- V-731 Magnetizing Apparatus. A suitable means for
ment requirements applicable for performing magnetic producing the magnetic field orientation and strength in
particle examinations using magnetic rubber techniques the part shall be employed, using direct or rectified cur-
in place of wet or dry magnetic particles. The principal ap- rent except where coatings are involved. Fields generated
plications for this technique are by alternating current electromagnetic yokes shall not be
(a) limited visual or mechanical accessibility, such as used except where nonmagnetic coatings are used on ex-
bolt holes ternal surfaces. Gaussmeters or artificial shims shall be
(b) coated surfaces used for field strength and direction determination.
(c) complex shapes or poor surface conditions V-732 Magnetic Rubber Materials. The material shall
(d) discontinuities that require magnification for detec- be in the form of a vulcanizing polymer (rubber) liquid or
tion and interpretation semiliquid, containing ferromagnetic particles. The mate-
(e) permanent record of the actual inspection rial shall be utilized at the temperature range as recom-
mended by the manufacturer. When demonstration is

--`,,,`,`,``,``,,`,``,`,`,`,``,`-`-`,,`,,`,`,,`---
V-720 GENERAL REQUIREMENTS required, the temperature shall be recorded.
V-733 Magnetic Field Strength. A calibrated gauss-
V-720.1 Requirements. Requirements of Article 7 ap-
meter or artificial shims shall be used to determine the
ply unless modified by this Appendix.
magnetic field strength and direction on surfaces to be ex-
V-720.2 Application. To accommodate the examina- amined. The gaussmeter device shall be equipped with
tion of a variety of surfaces, a liquid polymer containing both transverse and axial field probes. Dial or similar type
ferromagnetic particles is applied to the surface instead calibrated meters of suitable range may be used, providing
of conventional dry or suspended wet particles. During they are capable of making transverse and axial measure-
the cure time, the application of magnetizing fields cause ments. Values for G (kAm-1) or the use of artificial shims
the particles to migrate and form patterns at discontinu- shall be in accordance with T-764.
ities. The polymer cures forming an elastic solid (e.g., a
rubber replica) with indications permanently fixed on its V-734 Magnification. Replica viewing may be aided
surface. by the use of magnification.

V-721 WRITTEN PROCEDURE REQUIREMENTS


V-740 MISCELLANEOUS REQUIREMENTS
V-721.1 Requirements. Magnetic rubber techniques
shall be performed in accordance with a written proce- V-741 SURFACE PREPARATION
dure that shall, as a minimum, contain the requirements (a) Prior to the magnetic particle examination, the sur-
listed in Table V-721. The written procedure shall estab- face(s) to be examined and adjacent areas within at least
lish a single value, or range o f values, for each 1
/2 in. (13 mm) of the area of interest shall be dry and free
requirement. of all dirt, oil, grease, paint, lint, scale and welding flux, and
V-721.2 Procedure Qualification. When procedure other extraneous material that could restrict particle
qualification is specified by the referencing Code Section, movement and interfere with the examination by prevent-
a change of a requirement in Table V-721 identified as ing cure or extending the curing time. Nonmagnetic sur-
an essential variable shall require requalification of the face coatings need not be removed for techniques using
written procedure by demonstration. A change of a re- an alternating current electromagnetic yoke.
quirement identified as a nonessential variable does not (b) When nonmagnetic coatings are left on the part in
require requalification of the written procedure. All the area being examined, it shall be demonstrated with
changes of essential or nonessential variables from those an alternating current electromagnetic yoke that the indi-
specified within the written procedure shall require revi- cations can be detected through the existing maximum
sion of, or an addendum to, the written procedure. coating thickness per Appendix I.

191
Copyright ASME International (BPVC)
Provided by IHS under license with ASME pe.class Licensee=PDVSA - Pequiven site 9/9986712014, User=Gil, Eliecer
No reproduction or networking permitted without license from IHS Not for Resale, 07/23/2013 06:23:05 MDT
ARTICLE 7 2013 SECTION V

Table V-721
Requirements for the Magnetic Rubber Examination Procedure
Nonessential
Requirement Essential Variable Variable
Magnetic Rubber Mix Formulations (Manufacturer’s name(s) for material of X —
various viscosities and recommended cure times)
Surface preparation X —
Magnetizing technique X —
Field strength X —
Nonmagnetic coating thickness greater than previously qualified X —
Minimum cure time as recommended by the manufacturer X —
Releasing agent X —
Temperature range as specified by the manufacturer X —
Performance demonstration, when required X —
Number of fields and directions to be applied and magnetizing time for each X —
direction
Demagnetizing — X
Personnel qualification requirements — X

--`,,,`,`,``,``,,`,``,`,`,`,``,`-`-`,,`,,`,`,,`---
Reference to the procedure qualification records — X

V-742 TAPING AND DAMMING V-743 RELEASE TREATMENT


Tape, putty, plugs, and other suitable means shall be Areas where the liquid polymer has been in contact with
used to form dams or encapsulations that will provide a the examination or other surfaces may result in a tempor-
reservoir or containment to hold the liquid or semi-liquid ary adhesion of the rubber. To avoid this condition, the
polymer in contact with the area of interest during magne- area where the liquid polymer will be in contact shall be
tization and until curing is complete. The construction of treated with a Teflon-type release agent prior to the appli-
the containment will depend on the geometry of the mate- cation of the liquid polymer. The release treatment agent
rial and the area of interest. Some examples are as follows: shall not contain silicones.
(a) Horizontal Through–holes. Place adhesive tape over
one side of the hole, making a pinhole in the tape at the
top of the hole for release of air during pouring. A cup, V-750 TECHNIQUES
open on the top side and fabricated from heavy aluminum V-751 TECHNIQUES
foil, may be attached with tape or putty to the opposite
side of the hole to serve as a funnel during pouring of Magnetization techniques used are comparable to those
the liquid polymer. described in T-750. Direct current electromagnetic yokes
are the preferred magnetizing device.
(b) Flat Surface. Putty dams may be constructed around
the area of interest to contain the liquid polymer after
V-752 APPLICATION OF MAGNETIC FIELD
pouring.
(c) Inverted Surfaces. A putty reservoir may be placed Flaws are displayed more vividly when a discontinuity
beneath the examination area and pressure fill the area is oriented perpendicular to the magnetic lines of force.
with liquid polymer allowing trapped air to escape by pla- Magnetism shall be applied in a minimum of two or more
cing a small vent hole in the dam next to the area of inter- directions, where two of the magnetic lines of force are ap-
est. Inverted holes may be filled by pressure feeding the proximately perpendicular to each other and at least one
liquid polymer at the upper side of the dammed hole. Place of the lines of force are perpendicular to suspected
a small tube, open at each end, next to the fill tube with discontinuities.
one end at the same location as the end of the fill tube.
Pressure feed until the polymer overflows from the sec- V-764 MAGNETIC FIELD ADEQUACY AND
ond tube. Remove tubes when fill is completed and plug DIRECTION
access holes. The field strength shall be measured using a gauss-
meter. The area to be examined shall be checked in two di-
rections by placing the gaussmeter probe in the hole or on
the surface to be inspected and noting the field strength
and direction of the magnetic field. Artificial flaw shims,

192
Copyright ASME International (BPVC)
Provided by IHS under license with ASME pe.class Licensee=PDVSA - Pequiven site 9/9986712014, User=Gil, Eliecer
No reproduction or networking permitted without license from IHS Not for Resale, 07/23/2013 06:23:05 MDT
2013 SECTION V ARTICLE 7

as described in T-764.2(b), may also be used when acces- V-780 EVALUATION


sibility allows, to determine the field strength and direc-
tion of magnetization using wet or dry particles. (a) All indications shall be evaluated in terms of the ac-
ceptance standards of the referencing Code Section.
(b) Following removal, the replicas shall be examined
V-770 EXAMINATION visually in order to detect any damage to the surface of
the replica. When the area of interest shows damage or
V-773 Application of Liquid Polymer- Magnetic Par-
lack of fill or contact with the examination surface, the ex-
ticle Material. Following the initial steps of preparation, a
amination shall be repeated.
freshly prepared polymer-magnetic particle mix shall be
(c) When dimensional data is required, an illuminating-
cast or molded into/onto the prepared area. The magnetic
magnifying device capable of making measurements shall
field, previously determined to have the required mini-
be used.
mum field strength recommended by the polymer-particle
manufacturer, shall be applied to the area of interest. A
minimum of two fields 90 deg apart shall be maintained V-790 DOCUMENTATION
for an equal amount of time during the cure time of the li-
quid polymer-particle mix used. When more than two V-793 Examination Records. For each examination,
fields are to be applied, a minimum time in the first direc- the following information shall be recorded:
tion shall be allowed before magnetization in the next di- (a) date of the examination
rection is applied and the same minimum time used for (b) procedure identification and revision
each subsequent magnetization. The cure time applied to (c) m a g n e t i c r u b b e r m i x – m a n u f a c t u r e r a n d
each direction shall be based on the mix’s cure time di- identification
vided by the number of magnetic fields applied. (d) examination personnel, if required by the referen-
cing Code Section
V-774 Movement During Cure. During the cure time
(e) map or record of indications for evaluation, per
of the liquid polymer-particle mix, movement of the item
T-792
shall be avoided to ensure indications are not distorted.
(f) use, type and power of magnification
V-776 Removal of Replicas. Replicas shall be re- (g) material and thickness
moved as soon as practical after cure by careful use of a (h) magnetic particle equipment and type of current
tool or compressed air. Additional time must be allowed (i) gaussmeter; manufacturer, model, serial number, or
if the polymer is not fully cured or sticks to the examina- artificial shims used
tion area. (j) field strength (if gaussmeter is used), duration and
total time of application
(k) when more than two fields are applied, number and
sequencing of the applications
(l) temperature

--`,,,`,`,``,``,,`,``,`,`,`,``,`-`-`,,`,,`,`,,`---

193
Copyright ASME International (BPVC)
Provided by IHS under license with ASME pe.class Licensee=PDVSA - Pequiven site 9/9986712014, User=Gil, Eliecer
No reproduction or networking permitted without license from IHS Not for Resale, 07/23/2013 06:23:05 MDT
ARTICLE 7 2013 SECTION V

NONMANDATORY APPENDIX A
MEASUREMENT OF TANGENTIAL FIELD STRENGTH WITH
GAUSSMETERS

A-710 SCOPE part at the location of measurement to within 5 deg. This


may be difficult to accomplish by hand orientation. A jig or
This Nonmandatory Appendix is used for the purpose of fixture may be used to ensure this orientation is achieved
establishing procedures and equipment specifications for and maintained.
measuring the tangential applied magnetic field strength. The direction and magnitude of the tangential field on
the part surface can be determined by placing the
A-720 GENERAL REQUIREMENTS Hall-Effect tangential field probe on the part surface in
the area of interest. The direction of the field can be deter-
Personnel qualification requirements shall be in accor- mined during the application of the magnetizing field by
dance with Article 1. rotating the tangential field probe while in contact with
Gaussmeters and related equipment shall be calibrated the part until the highest field reading is obtained on the
in accordance with T-763 of Article 7. Gaussmeter. The orientation of the probe, when the high-
Definitions: standard terminology for magnetic particle est field is obtained, will indicate the field direction at that
examinations is presented in SE-1316. point. Gaussmeters cannot be used to determine the ade-
quacy of magnetizing fields for multidirectional and coil
magnetization techniques.
A-730 EQUIPMENT
Once adequate field strength has been demonstrated
Gaussmeter having the capability of being set to read with artificial flaw shims, Gaussmeter readings may be
peak values of field intensity. The frequency response of used at the location of shim attachment on identical parts
the gaussmeter shall be at least 0 Hz to 300 Hz. or similar configurations to verify field intensity and
The Hall-Effect tangential field probe should be no lar- direction.
ger than 0.2 in. (5 mm) by 0.2 in. (5 mm) and should have
a maximum center location 0.2 in. (5 mm) from the part
surface. Probe leads shall be shielded or twisted to pre- A-790 DOCUMENTATION/RECORDS
vent reading errors due to voltage induced during the Documentation should include the following:
large field changes encountered during magnetic particle (a) equipment model and probe description;
examinations. (b) sketch or drawing showing where measurements
are made; and
(c) field intensity and direction of measurement.
A-750 PROCEDURE
Care shall be exercised when measuring the tangential
applied field strengths specified in T-764.1.3. The plane
of the probe must be perpendicular to the surface of the
--`,,,`,`,``,``,,`,``,`,`,`,``,`-`-`,,`,,`,`,,`---

194
Copyright ASME International (BPVC)
Provided by IHS under license with ASME pe.class Licensee=PDVSA - Pequiven site 9/9986712014, User=Gil, Eliecer
No reproduction or networking permitted without license from IHS Not for Resale, 07/23/2013 06:23:05 MDT
2013 SECTION V ARTICLE 8

ARTICLE 8
EDDY CURRENT EXAMINATION

T-810 SCOPE (e) Mandatory Appendix III, Eddy Current Examination


on Coated Ferritic Materials, provides eddy current re-
When specified by the referencing Code Section, the quirements for eddy current examination on coated ferri-
eddy current examination method and techniques de- tic materials.
scribed in this Article shall be used. (f) Mandatory Appendix IV, External Coil Eddy Current
(a) This Article describes the techniques to be used Examination of Tubular Products, provides the require-
when performing eddy current examinations on ments for external coil eddy current examination of seam-
conductive-nonferromagnetic and coated ferritic less copper, copper alloy, austenitic stainless steel, Ni-
materials. Cr-Fe alloy, and other nonferromagnetic tubular products.
(b) The requirements of Article 1, General Require- (g) Mandatory Appendix V, Eddy Current Measurement
ments, also apply when eddy current examination, in ac- of Nonconductive-Nonmagnetic Coating Thickness on a
cordance with Article 8, is required by a referencing Nonmagnetic Metallic Material, provides the requirements

--`,,,`,`,``,``,,`,``,`,`,`,``,`-`-`,,`,,`,`,,`---
Code Section. for surface probe eddy current examination for measuring
(c) Definitions of terms for eddy current examination nonconductive-nonmagnetic coating thicknesses.
appear in three places: Mandatory Appendix I to this Arti- (h) Mandatory Appendix VI, Eddy Current Detection
cle; Article 1, Mandatory Appendix I; and Subsection B, and Measurement of Depth of Surface Discontinuities in
Article 30. Nonmagnetic Metals With Surface Probes, provides the re-
(d) Mandatory Appendix II, Eddy Current Examination quirements for surface probe eddy current examination
of Nonferromagnetic Heat Exchanger Tubing, provides for detection of surface connected discontinuities and
the requirements for bobbin coil multifrequency and mul- measuring their depth.
tiparameter eddy current examination of installed nonfer-
romagnetic heat exchanger tubing.

195
Copyright ASME International (BPVC)
Provided by IHS under license with ASME pe.class Licensee=PDVSA - Pequiven site 9/9986712014, User=Gil, Eliecer
No reproduction or networking permitted without license from IHS Not for Resale, 07/23/2013 06:23:05 MDT

You might also like