Professional Documents
Culture Documents
Asme V 2013.pdf - 2
Asme V 2013.pdf - 2
Table T-721
Requirements of a Magnetic Particle Examination Procedure
Nonessential
Requirement Essential Variable Variable
Magnetizing technique X ...
Magnetizing current type or amperage outside range specified by this Article or as previously X ...
qualified
Surface preparation X .. .
Magnetic particles (fluorescent/visible, color, particle size, wet/dry) X .. .
Method of particle application X .. .
Method of excess particle removal X .. .
Minimum light intensity X .. .
Existing coatings, greater than the thickness demonstrated X .. .
Nonmagnetic surface contrast enhancement, when utilized X .. .
Performance demonstration, when required X .. .
Examination part surface temperature outside of the temperature range recommended by the X .. .
manufacturer of the particles or as previously qualified
Shape or size of the examination object . . . X
Equipment of the same type . . . X
Temperature (within those specified by manufacturer or as previously qualified) . . . X
Demagnetizing technique . . . X
Post-examination cleaning technique . . . X
Personnel qualification requirements . . . X
173
--`,,,`,`,``,``,,`,``,`,`,`,``,`-`-`,,`,,`,`,,`---
than 3 in. (75 mm) are usually not practical due to banding (d) If the area to be magnetized extends beyond 9 in.
of the particles around the prods. The prod tips shall be (225 mm) on either side of the coil’s center, field adequacy
kept clean and dressed. If the open circuit voltage of the shall be demonstrated using a magnetic field indicator or
magnetizing current source is greater than 25 V, lead, artificial flaw shims per T-764.
steel, or aluminum (rather than copper) tipped prods (e) F o r l a r g e p a r t s d u e t o s iz e a n d s h a p e , t h e
are recommended to avoid copper deposits on the part magnetizing current shall be 1200 ampere-turns
being examined. to 4500 ampere-turns. The field adequacy shall be demon-
strated using artificial flaw shims or a pie-shaped mag-
--`,,,`,`,``,``,,`,``,`,`,`,``,`-`-`,,`,,`,`,,`---
netic field indicator in accordance with T-764. A
T-753 LONGITUDINAL MAGNETIZATION Hall-Effect probe gaussmeter shall not be used with encir-
TECHNIQUE cling coil magnetization techniques.
T-753.1 Magnetizing Procedure. For this technique, T-753.3 Magnetizing Current. The current required
magnetization is accomplished by passing current through to obtain the necessary magnetizing field strength shall
a multi-turn fixed coil (or cables) that is wrapped around be determined by dividing the ampere-turns obtained in
the part or section of the part to be examined. This pro- steps T-753.2(a) or T-753.2(b) by the number of turns
duces a longitudinal magnetic field parallel to the axis of in the coil as follows:
the coil.
If a fixed, prewound coil is used, the part shall be placed
near the side of the coil during inspection. This is of special
importance when the coil opening is more than 10 times
For example, if a 5-turn coil is used and the ampere-
the cross-sectional area of the part.
turns required are 5000, use
ð13Þ T-753.2 Magnetic Field Strength. Direct or rectified
current shall be used to magnetize parts examined by this
technique. The required field strength shall be calculated
based on the length L and the diameter D of the part in ac-
cordance with (a) and (b), or as established in (d) and (e),
below. Long parts shall be examined in sections not to ex- T-754 CIRCULAR MAGNETIZATION TECHNIQUE
ceed 18 in. (450 mm), and 18 in. (450 mm) shall be used T-754.1 Direct Contact Technique.
for the part L in calculating the required field strength. For (a) Magnetizing Procedure. For this technique, magneti-
noncylindrical parts, D shall be the maximum cross- zation is accomplished by passing current through the
sectional diagonal. part to be examined. This produces a circular magnetic
(a) Parts With L/D Ratios Equal to or Greater Than 4. field that is approximately perpendicular to the direction
The magnetizing current shall be within 10% of the of current flow in the part.
ampere-turns’ value determined as follows: (b) Magnetizing Current. Direct or rectified (half-wave
rectified or full-wave rectified) magnetizing current shall
be used.
(1) The current shall be 300 amp/in. (12 A/mm) to
800 amp/in. (31 A/mm) of outer diameter.
(2) Parts with geometric shapes other than round
For example, a part 10 in. (250 mm) long × 2 in. with the greatest cross-sectional diagonal in a plane at
(50 mm) diameter has an L /D ratio of 5. Therefore, right angles to the current flow shall determine the inches
to be used in (1) above.
(3) If the current levels required for (1) cannot be ob-
tained, the maximum current obtainable shall be used and
the field adequacy shall be demonstrated in accordance
with T-764.
(b) Parts With L/D Ratios Less Than 4 but Not Less Than
2. The magnetizing ampere-turns shall be within 10% of T-754.2 Central Conductor Technique.
the ampere-turns’ value determined as follows: (a) Magnetizing Procedure. For this technique, a central
conductor is used to examine the internal surfaces of cy-
lindrically or ring-shaped parts. The central conductor
technique may also be used for examining the outside sur-
faces of these shapes. Where large diameter cylinders are
to be examined, the conductor shall be positioned close to
(c) Parts With L/D Ratios Less Than 2. Coil magnetiza- the internal surface of the cylinder. When the conductor is
tion technique cannot be used. not centered, the circumference of the cylinder shall be
174
Copyright ASME International (BPVC)
Provided by IHS under license with ASME pe.class Licensee=PDVSA - Pequiven site 9/9986712014, User=Gil, Eliecer
No reproduction or networking permitted without license from IHS Not for Resale, 07/23/2013 06:23:05 MDT
2013 SECTION V ARTICLE 7
The distance along the part circumference (exterior) that packs operating as many as three circuits that are ener-
is effectively magnetized shall be taken as four times the gized one at a time in rapid succession. The effect of these
diameter of the central conductor, as illustrated in rapidly alternating magnetizing currents is to produce an
Figure T-754.2.2. The entire circumference shall be in- overall magnetization of the part in multiple directions.
spected by rotating the part on the conductor, allowing Circular or longitudinal magnetic fields may be generated
for approximately a 10% magnetic field overlap. in any combination using the various techniques de-
scribed in T-753 and T-754.
T-755 YOKE TECHNIQUE
For this technique, alternating or direct current electro- T-756.2 Magnetic Field Strength. Only three phase,
magnetic yokes, or permanent magnet yokes, shall be full-wave rectified current shall be used to magnetize
used. the part. The initial magnetizing current requirements
for each circuit shall be established using the previously
described guidelines (see T-753 and T-754). The adequacy
of the magnetic field shall be demonstrated using artificial
flaw shims or a pie-shaped magnetic particle field indica-
tor in accordance with T-764. A Hall-Effect probe
Figure T-754.2.1
Single-Pass and Two-Pass Central Conductor Technique
175
Copyright ASME International (BPVC)
Provided by IHS under license with ASME pe.class Licensee=PDVSA - Pequiven site 9/9986712014, User=Gil, Eliecer
No reproduction or networking permitted without license from IHS Not for Resale, 07/23/2013 06:23:05 MDT
ARTICLE 7 2013 SECTION V
gaussmeter shall not be used to measure field adequacy has been subjected to a major repair, periodic overhaul,
for the multidirectional magnetization technique. An ade- or damage. If equipment has not been in use for a year
quate field shall be obtained in at least two nearly perpen- or more, calibration shall be done prior to first use.
dicular directions, and the field intensities shall be
balanced so that a strong field in one direction does not
overwhelm the field in the other direction. For areas T-764 MAGNETIC FIELD ADEQUACY AND
where adequate field strengths cannot be demonstrated, DIRECTION
additional magnetic particle techniques shall be used to T-764.1 Application. The use of magnetic field indica-
obtain the required two-directional coverage. tors, artificial shims, or Hall-Effect tangential-field probes
are only permitted when specifically referenced by the fol-
lowing magnetizing techniques:
T-760 CALIBRATION (a) Longitudinal (T-753)
T-761 FREQUENCY OF CALIBRATION (b) Circular (T-754)
T-761.1 Magnetizing Equipment. (c) Multidirectional (T-756)
(a) Frequency. Magnetizing equipment with an ammeter T-764.2 Magnetic Field Adequacy. The applied mag-
shall be calibrated at least once a year, or whenever the netic field shall have sufficient strength to produce satis-
equipment has been subjected to major electric repair, factory indications, but shall not be so strong that it
periodic overhaul, or damage. If equipment has not been causes masking of relevant indications by nonrelevant ac-
in use for a year or more, calibration shall be done prior cumulations of magnetic particles. Factors that influence
to first use. the required field strength include the size, shape, and ma-
(b) Procedure. The accuracy of the unit’s meter shall be terial permeability of the part; the technique of magnetiza-
verified annually by equipment traceable to a national tion; coatings; the method of particle application; and the
standard. Comparative readings shall be taken for at least type and location of discontinuities to be detected. When it
three different current output levels encompassing the is necessary to verify the adequacy of magnetic field
usable range. strength, it shall be verified by using one or more of the
--`,,,`,`,``,``,,`,``,`,`,`,``,`-`-`,,`,,`,`,,`---
(c) Tolerance. The unit’s meter reading shall not deviate following three methods.
by more than ±10% of full scale, relative to the actual cur-
(a) Pie-Shaped Magnetic Particle Field Indicator. The in-
rent value as shown by the test meter.
dicator, shown in Figure T-764.2(a), shall be positioned on
T-761.2 Light Meters. Light meters shall be calibrated the surface to be examined, such that the copper-plated
at least once a year or whenever a meter has been re- side is away from the inspected surface. A suitable field
paired. If meters have not been in use for one year or strength is indicated when a clearly defined line (or lines)
more, calibration shall be done before being used. of magnetic particles form(s) across the copper face of the
indicator when the magnetic particles are applied simulta-
T-762 LIFTING POWER OF YOKES neously with the magnetizing force. When a clearly de-
(a) Prior to use, the magnetizing power of electromag- fined line of particles is not formed, the magnetizing
netic yokes shall have been checked within the past year. technique shall be changed as needed. Pie-type indicators
The magnetizing power of permanent magnetic yokes are best used with dry particle procedures.
shall be checked daily prior to use. The magnetizing power (b) Artificial Flaw Shims. One of the shims shown in
of all yokes shall be checked whenever the yoke has been Figure T-764.2(b)(1) or Figure T-764.2(b)(2) whose or-
damaged or repaired. ientation is such that it can have a component perpendicu-
(b) Each alternating current electromagnetic yoke shall lar to the applied magnetic field shall be used. Shims with
have a lifting power of at least 10 lb (4.5 kg) at the maxi- linear notches shall be oriented so that at least one notch
mum pole spacing that will be used. is perpendicular to the applied magnetic field. Shims with
(c) Each direct current or permanent magnetic yoke only circular notches may be used in any orientation.
shall have a lifting power of at least 40 lb (18 kg) at the Shims shall be attached to the surface to be examined,
maximum pole spacing that will be used. such that the artificial flaw side of the shim is toward
(d) Each weight shall be weighed with a scale from a re- the inspected surface. A suitable field strength is indicated
putable manufacturer and stenciled with the applicable when a clearly defined line (or lines) of magnetic particles,
nominal weight prior to first use. A weight need only be representing the 30% depth flaw, appear(s) on the shim
verified again if damaged in a manner that could have face when magnetic particles are applied simultaneously
caused potential loss of material. with the magnetizing force. When a clearly defined line
of particles is not formed, the magnetizing technique shall
T-763 GAUSSMETERS be changed as needed. Shim-type indicators are best used
Hall-Effect probe gaussmeters used to verify magnetiz- with wet particle procedures.
ing field strength in accordance with T-754 shall be cali- NOTE: The circular shims shown in Figure T-764.2(b)(2) illustration
brated at least once a year or whenever the equipment (b) also have flaw depths less and greater than 30%.
176
Copyright ASME International (BPVC)
Provided by IHS under license with ASME pe.class Licensee=PDVSA - Pequiven site 9/9986712014, User=Gil, Eliecer
No reproduction or networking permitted without license from IHS Not for Resale, 07/23/2013 06:23:05 MDT
2013 SECTION V ARTICLE 7
0.002 in.
A
(0.06 mm)
0.005 in.
0.75 in. (0.125 mm)
(20 mm) typical
A 0.0006 in.
(0.015 mm)
0.25 in.
(6 mm) Section A–A
--`,,,`,`,``,``,,`,``,`,`,`,``,`-`-`,,`,,`,`,,`---
Type B
A 0.002 in.
(0.05 mm)
0.75 in.
(20 mm)
A 0.0006 in.
(0.015 mm)
0.5 in.
(12.5 mm) Section A–A
Type C
177
Copyright ASME International (BPVC)
Provided by IHS under license with ASME pe.class Licensee=PDVSA - Pequiven site 9/9986712014, User=Gil, Eliecer
No reproduction or networking permitted without license from IHS Not for Resale, 07/23/2013 06:23:05 MDT
ARTICLE 7 2013 SECTION V
Figure T-764.2(b)(2)
Artificial Flaw Shims
0.75 in. (typ) (19.05 mm) 0.75 in. (typ) (19.05 mm)
0.25 in. 0.25 in.
(6.36 mm) (6.36 mm)
0.75 in. (typ) (19.05 mm) 0.75 in. (typ) (19.05 mm)
0.007 in. (type) 0.007 in. (type)
(0.18 mm) 0.507 in. diam. O.D. (0.18 mm) 0.507 in. diam. O.D.
(12.88 mm) (12.88 mm)
0.79 in. (typ) (20.06 mm) 0.79 in. (typ) (20.06 mm)
0.235 in. (typ) 0.235 in. (typ)
(5.97 mm) (5.97 mm)
230 430
Shim Thickness 0.002 in. (0.051 mm) Shim Thickness 0.004 in. (0.102 mm)
Shim Type CX-230 Shim Type CX4-430
(c)
178
--`,,,`,`,``,``,,`,``,`,`
T-765.3 Contamination. Both fluorescent and non- If the test does not reveal the required number of holes,
fluorescent suspensions shall be checked periodically for the equipment shall be taken out of service and the cause
contaminants such as dirt, scale, oil, lint, loose fluorescent of the loss of sensitivity determined and corrected. This
pigment, water (in the case of oil suspensions), and parti- test shall be run at least once per week.
cle agglomerates which can adversely affect the perfor-
mance of the magnetic particle examination process. The
test for contamination shall be performed at least once T-770 EXAMINATION
per week.
T-771 PRELIMINARY EXAMINATION
(a) Carrier Contamination. For fluorescent baths, the li-
quid directly above the precipitate should be examined Before the magnetic particle examination is conducted,
with fluorescent excitation light. The liquid will have a lit- a check of the examination surface shall be conducted to
tle fluorescence. Its color can be compared with a freshly locate any discontinuity surface openings which may not
made-up sample using the same materials or with an un- attract and hold magnetic particles because of their width.
used sample from the original bath that was retained for
this purpose. If the “used” sample is noticeably more fluor- T-772 DIRECTION OF MAGNETIZATION
escent than the comparison standard, the bath shall be
At least two separate examinations shall be performed
replaced.
on each area. During the second examination, the lines
(b) Particle Contamination. The graduated portion of the
of magnetic flux shall be approximately perpendicular to
tube shall be examined under fluorescent excitation light if
those used during the first examination. A different techni-
the bath is fluorescent and under visible light (for both
que for magnetization may be used for the second
fluorescent and nonfluorescent particles) for striations
examination.
or bands, differences in color or appearance. Bands or
striations may indicate contamination. If the total volume
of the contaminates, including bands or striations exceeds T-773 METHOD OF EXAMINATION
30% of the volume magnetic particles, or if the liquid is The ferromagnetic particles used in an examination
noticeably fluorescent, the bath shall be replaced. medium can be either wet or dry, and may be either fluor-
escent or nonfluorescent. Examination(s) shall be done by
T-766 SYSTEM PERFORMANCE OF HORIZONTAL the continuous method.
UNITS (a) Dry Particles. The magnetizing current shall remain
The Ketos (Betz) ring specimen (see Figure T-766.1) on while the examination medium is being applied and
shall be used in evaluating and comparing the overall per- while any excess of the examination medium is removed.
--`,,,`,`,``,``,,`,``,`,`,`,``,`-`-`,,`,,`,`,,`---
formance and sensitivity of both dry and wet, fluorescent (b) Wet Particles. The magnetizing current shall be
and nonfluorescent magnetic particle techniques using a turned on after the particles have been applied. Flow of
central conductor magnetization technique. particles shall stop with the application of current. Wet
(a) Ketos (Betz) Test Ring Material. The tool steel (Ke- particles applied from aerosol spray cans may be applied
tos) ring should be machined from AISI 01 material in ac- before and/or after magnetizing current is applied. Wet
cordance with Figure T-766.1. Either the machined ring or particles may be applied during the application of magne-
the steel blank should be annealed at 1,650°F (900°C), tizing current if they are not applied directly to the exam-
cooled 50°F (28°C) per hour to 1,000°F (540°C) and then ination area and are allowed to flow over the examination
air cooled to ambient temperature to give comparable re- area or are applied directly to the examination area with
sults using similar rings that have had the same treatment. low velocities insufficient to remove accumulated
Material and heat treatment are important variables. Ex- particles.
perience indicates controlling the softness of the ring by
hardness (90 to 95 HRB) alone is insufficient. T-774 EXAMINATION COVERAGE
(b) U s i n g t h e T e s t R i n g . T h e t e s t r i n g ( s e e All examinations shall be conducted with sufficient field
Figure T-766.1), is circularly magnetized with full-wave overlap to ensure 100% coverage at the required sensitiv-
rectified AC passing through a central conductor with a ity (T-764).
1 in. to 11/4 in. (25 mm to 32 mm) diameter hole located
in the ring center. The conductor should have a length
T-775 RECTIFIED CURRENT
greater than 16 in. (400 mm). The currents used shall be
1400, 2500, and 3400 amps. The minimum number of (a) Whenever direct current is required rectified cur-
holes shown shall be three, five, and six, respectively. rent may be used. The rectified current for magnetization
The ring edge should be examined with either black light shall be either three-phase (full-wave rectified) current, or
or visible light, depending on the type of particles in- single phase (half-wave rectified) current.
volved. This test shall be run at the three amperages if (b) The amperage required with three-phase, full-wave
the unit will be used at these or higher amperages. The rectified current shall be verified by measuring the aver-
amperage values stated shall not be exceeded in the test. age current.
179
Copyright ASME International (BPVC)
Provided by IHS under license with ASME pe.class Licensee=PDVSA - Pequiven site 9/9986712014, User=Gil, Eliecer
No reproduction or networking permitted without license from IHS Not for Resale, 07/23/2013 06:23:05 MDT
ARTICLE 7 2013 SECTION V
Figure T-766.1
Ketos (Betz) Test Ring
7/ in.
8
(22 mm)
Hole 1 2 3 4 5 6 7 8 9 10 11 12
Diame- 0.07 0.07 0.07 0.07 0.07 0.07 0.07 0.07 0.07 0.07 0.07 0.07
ter (1.8 (1.8 (1.8 (1.8 (1.8 (1.8 (1.8 (1.8 (1.8 (1.8 (1.8 (1.8
[Note mm) mm) mm) mm) mm) mm) mm) mm) mm) mm) mm) mm)
(1)]
“D” 0.07 0.14 0.21 0.28 0.35 0.42 0.49 0.56 0.63 0.70 0.77 0.84
[Note (1.8 (3.6 (5.3 (7.1 (9.0 (10.8 (12.6 (14.4 (16.2 (18.0 (19.8 (21.6
(2)] mm) mm) mm) mm) mm) mm) mm) mm) mm) mm) mm) mm)
NOTES:
(1) All hole diameters are ±0.005 in. (±0.1 mm.) Hole numbers 8 through 12 are optional.
(2) Tolerance on the D distance is ±0.005 in. (±0.1 mm).
GENERAL NOTES:
(a) All dimensions are ±0.03 in. (±0.8 mm) or as noted in Notes (1) and (2).
(b) All dimensions are in inches, except as noted.
(c) Material is ANSI 01 tool steel from annealed round stock.
(d) The ring may be heat treated as follows: Heat to 1400°F to 1500°F (760°C to 790°C). Hold at this temperature for one hour. Cool to a mini-
mum rate of 40°F/h (22°C/h) to below 1000°F (540°C). Furnace or air cool to room temperature. Finish the ring to RMS 25 and protect
from corrosion.
180
Copyright ASME International (BPVC)
--`,,,`,`,``,``,
Provided by IHS under license with ASME pe.class Licensee=PDVSA - Pequiven site 9/9986712014, User=Gil, Eliecer
No reproduction or networking permitted without license from IHS Not for Resale, 07/23/2013 06:23:05 MDT
2013 SECTION V ARTICLE 7
indications. The light source, technique used, and light le- (f) The fluorescence excitation light intensity shall be
vel verification is required to be demonstrated one time, measured with a suitable fluorescence excitation light me-
documented, and maintained on file. ter prior to use, whenever the light’s power source is in-
T-777.2 Fluorescent Magnetic Particles With Black terrupted or changed, and at the completion of the
Light. With fluorescent magnetic particles, the process is examination or series of examinations.
essentially the same as in T-777.1, with the exception that
the examination is performed using an ultraviolet light T-778 DEMAGNETIZATION
(i.e., nominal 365 nm), called black light. The examination
When residual magnetism in the part could interfere
shall be performed as follows:
with subsequent processing or usage, the part shall be de-
(a) It shall be performed in a darkened area.
magnetized any time after completion of the examination.
(b) Examiners shall be in a darkened area for at least
5 min prior to performing examinations to enable their
eyes to adapt to dark viewing. Glasses or lenses worn by T-779 POST-EXAMINATION CLEANING
examiners shall not be photochromic or exhibit any
When post-examination cleaning is required, it should
fluorescence.
be conducted as soon as practical using a process that
(c) Black lights shall achieve a minimum of 1000
--`,,,`,`,``,``,,`,``,`,`,`,``,`-`-`,,`,,`,`,,`---
181
Copyright ASME International (BPVC)
Provided by IHS under license with ASME pe.class Licensee=PDVSA - Pequiven site 9/9986712014, User=Gil, Eliecer
No reproduction or networking permitted without license from IHS Not for Resale, 07/23/2013 06:23:05 MDT
ARTICLE 7 2013 SECTION V
182
Copyright ASME International (BPVC)
Provided by IHS under license with ASME pe.class Licensee=PDVSA - Pequiven site 9/9986712014, User=Gil, Eliecer
No reproduction or networking permitted without license from IHS Not for Resale, 07/23/2013 06:23:05 MDT
2013 SECTION V ARTICLE 7
MANDATORY APPENDIX I
MAGNETIC PARTICLE EXAMINATION USING THE AC YOKE
TECHNIQUE ON FERRITIC MATERIALS COATED WITH
NONMAGNETIC COATINGS
--`,,,`,`,``,``,,`,``,`,`,`,``,`-`-`,,`,,`,`,,`---
183
Copyright ASME International (BPVC)
Provided by IHS under license with ASME pe.class Licensee=PDVSA - Pequiven site 9/9986712014, User=Gil, Eliecer
No reproduction or networking permitted without license from IHS Not for Resale, 07/23/2013 06:23:05 MDT
ARTICLE 7 2013 SECTION V
I-723 PROCEDURE/TECHNIQUE The material used for the specimen shall be the same spe-
DEMONSTRATION cification and heat treatment as the coated ferromagnetic
material to be examined. As an alternative to the material
The procedure/technique shall be demonstrated to the
requirement, other materials and heat treatments may be
satisfaction of the Inspector in accordance with the re-
qualified provided:
quirements of the referencing Code Section.
(1) The measured yoke maximum lifting force on the
material to be examined is equal to or greater than the
I-730 EQUIPMENT maximum lifting force on the qualification specimen mate-
rial. Both values shall be determined with the same or
I-730.1 The magnetizing equipment shall be in accor- comparable equipment and shall be documented as re-
dance with Article 7. quired in (c).
ð13Þ I-730.2 When the dry powder technique is used, a (2) All the requirements of (b) through (g) are met
compressed air powder blower shall be utilized for pow- for the alternate material.
der application in any position. Other applicators may be (b) Examine the uncoated specimen in the most unfa-
used if qualified in the same surface position as the exam- vorable orientation expected during the performance of
ination object surface. Applicators qualified for the over- the production examination.
head position may be used in any other position. (c) Document the measured yoke maximum lifting
Applicators qualified for the vertical position may be used power, illumination levels, and the results.
in the horizontal and flat positions.
(d) Measure the maximum coating thickness on the item
I-730.3 Magnetic particles shall contrast with the to be examined in accordance with the requirements of
component background. I-741.
I-730.4 Nonconductive materials such as plastic shim (e) Coat the specimen with the same type of coating,
stock may be used to simulate nonconductive nonmag- conductive or nonconductive, to the maximum thickness
netic coatings for procedure and personnel qualification. measured on the production item to be examined. Alter-
nately, nonconductive shim stock may be used to simulate
nonconductive coatings.
I-740 MISCELLANEOUS REQUIREMENTS (f) Examine the coated specimen in the most unfavor-
able orientation expected during the performance of the
I-741 COATING THICKNESS MEASUREMENT
production examination. Document the measured yoke
The procedure demonstration and performance of ex- maximum lifting power, illumination level, and examina-
aminations shall be preceded by measurement of the coat- tion results.
ing thickness in the areas to be examined. If the coating is (g) Compare the length of the indication resulting from
nonconductive, an eddy current technique or magnetic the longest flaw no longer than the maximum flaw size al-
technique may be used to measure the coating thickness. lowed by the applicable acceptance criteria, before and
--`,,,`,`,``,``,,`,``,`,`,`,``,`-`-`,,`,,`,`,,`---
The magnetic technique shall be in accordance with after coating. The coating thickness is qualified when the
SD-1186, Standard Test Methods for Nondestructive Mea- length of the indication on the coated surface is at least
surement of Dry Film Thickness of Nonmagnetic Coatings 50% of the length of the corresponding indication prior
Applied to a Ferrous Base. When coatings are conductive to coating.
and nonmagnetic, a coating thickness technique shall be (h) Requalification of the procedure is required for a de-
used in accordance with SD-1186. Coating measurement crease in either the AC yoke lifting power or the illumina-
equipment shall be used in accordance with the equip- tion level, or for an increase in the coating thickness.
ment manufacturer’s instructions. Coating thickness mea-
surements shall be taken at the intersections of a 2 in.
(50 mm) maximum grid pattern over the area of examina-
tion and at least one-half the maximum yoke leg separa- I-760 CALIBRATION
tion beyond the examination area. The thickness shall be I-761 YOKE MAXIMUM LIFTING FORCE
the mean of three separate readings within 1/4 in.
(6 mm) of each intersection. The maximum lifting force of the AC yoke shall be deter-
mined at the actual leg separation to be used in the exam-
ination. This may be accomplished by holding the yoke
I-750 TECHNIQUE with a 10 lb (4.5 kg) ferromagnetic weight between the
legs of the yoke and adding additional weights, calibrated
I-751 TECHNIQUE QUALIFICATION on a postage or other scale, until the ferromagnetic weight
(a) A qualification specimen is required. The specimen is released. The lifting power of the yoke shall be the com-
shall be of similar geometry or weld profile and contain bined weight of the ferromagnetic material and the added
at least one surface crack no longer than the maximum weights, before the ferromagnetic weight was released.
flaw size allowed in the applicable acceptance criteria. Other methods may be used such as a load cell.
184
Copyright ASME International (BPVC)
Provided by IHS under license with ASME pe.class Licensee=PDVSA - Pequiven site 9/9986712014, User=Gil, Eliecer
No reproduction or networking permitted without license from IHS Not for Resale, 07/23/2013 06:23:05 MDT
2013 SECTION V ARTICLE 7
I-762 LIGHT INTENSITY MEASUREMENT (b) Examine the coated item in accordance with the qua-
The black light or white light intensity (as appropriate) lified procedure.
on the surface of the component shall be no less than that
used in the qualification test. An appropriate calibrated
black light and/or white light meter shall be used for the I-780 EVALUATION
tests. Minimum white light or black light intensities shall
meet the requirements of T-777.1 or T-777.2 as If an indication greater than 50% of the maximum al-
applicable. lowable flaw size is detected, the coating in the area of
the indication shall be removed and the examination
I-762.1 White Light. The white light intensity shall be repeated.
measured at the inspection surface. The white light inten-
sity for the examination shall be no less than what was
used in the qualification.
I-790 DOCUMENTATION
I-762.2 Black Light. The black light intensity shall be
measured at the distance from the black light in the proce- I-791 EXAMINATION RECORD
dure qualification and at the same distance on the exami- For each examination, the information required in the
nation specimen. The black light intensity shall be no less records section of T-793 and the following information
than that used to qualify the procedure. In addition, the shall be recorded:
maximum white light intensity shall be measured as back-
(a) identification of the procedure/technique
ground light on the inspection surface. The background
white light for the examination shall be no greater than (b) identification of the personnel performing and wit-
what was used in the qualification. nessing the qualification
(c) description and drawings or sketches of the qualifi-
cation specimen, including coating thickness measure-
I-770 EXAMINATION ments and flaw dimensions
(a) Surfaces to be examined, and all adjacent areas with- (d) equipment and materials used
in at least 1 in. (25 mm), shall be free of all dirt, grease, lint, (e) illumination level and yoke lifting power
scale, welding flux and spatter, oil, and loose, blistered, (f) qualification results, including maximum coating
flaking, or peeling coating. thickness and flaws detected.
185
--`,,,`,`,``,``,,`,``,`,`,`,``,`-`-`,,`,,`,`,,`---
MANDATORY APPENDIX II
GLOSSARY OF TERMS FOR MAGNETIC PARTICLE EXAMINATION
186
--`,,,`,`,``,``,,`,``,`,`,`,``,`-`-`,,`,,`,`,,`---
III-760 CALIBRATION
III-761 BLACK LIGHT INTENSITY MEASUREMENT
III-720 GENERAL
The black light intensity on the surface of the compo-
Requirements of Article 7 apply unless modified by this nent shall be no less than that used in the qualification
Appendix. test.
Table III-721
Requirements for an AC or HWDC Yoke Technique With Fluorescent Particles in an Undarkened Area
Nonessential
Requirement Essential Variable Variable
Identification of surface configurations to be examined and product forms (e.g., base material or X ...
welded surface)
Surface condition requirement and preparation methods X . ..
Yoke manufacturer and model X . ..
Particle manufacturer and designation X . ..
Minimum and maximum pole separation X . ..
Identification of steps in performing the examination X . ..
Maximum white light intensity X . ..
Minimum black light intensity X . ..
Personnel qualification requirements ... X
Reference to the procedure qualification records ... X
--`,,,`,`,``,``,,`,``,`,`,`,``,`-`-`,,`,,`,`,,`---
187
Copyright ASME International (BPVC)
Provided by IHS under license with ASME pe.class Licensee=PDVSA - Pequiven site 9/9986712014, User=Gil, Eliecer
No reproduction or networking permitted without license from IHS Not for Resale, 07/23/2013 06:23:05 MDT
ARTICLE 7 2013 SECTION V
188
Copyright ASME International (BPVC)
Provided by IHS under license with ASME pe.class Licensee=PDVSA - Pequiven site 9/9986712014, User=Gil, Eliecer
No reproduction or networking permitted without license from IHS Not for Resale, 07/23/2013 06:23:05 MDT
2013 SECTION V ARTICLE 7
MANDATORY APPENDIX IV
QUALIFICATION OF ALTERNATE WAVELENGTH LIGHT SOURCES
FOR EXCITATION OF FLUORESCENT PARTICLES
IV-710 SCOPE and specific particles. Shim(s) shall be tape sealed to a fer-
romagnetic object’s surface and used as described in
This Appendix provides the methodology to qualify the T-764.1.2 with the notch against the object’s surface.
performance of fluorescent particle examinations using al-
ternate wavelength sources.
IV-752 FILTER GLASSES
If the alternative wavelength light source emits light in
IV-720 GENERAL the visible portion of the spectrum (wavelength of 400 nm
or longer), the examiner shall wear filter glasses that have
Requirements of Article 7 apply unless modified by this been supplied by the manufacturer of the light source to
Appendix. block the reflected visible excitation light while transmit-
ting the fluorescence of the particles.
IV-721 WRITTEN PROCEDURE REQUIREMENTS
IV-721.1 Requirements. The requirements of
Table IV-721 apply to Written Procedure Requirements IV-770 QUALIFICATION EXAMINATIONS
(T-721.1) and when specified by the referencing Code Sec-
tion to Procedure Qualification (T-721.2). IV-771 BLACK LIGHT INTENSITY
The black light intensity on the examination surface
IV-723 PROCEDURE DEMONSTRATION shall be adjusted by varying the distance or power so that
it has a minimum intensity of 1,000 μW/cm2 and a maxi-
The procedure shall be demonstrated to the satisfaction mum intensity of 1,100 μW/cm2.
of the Inspector in accordance with the requirements of
the referencing Code Section.
IV-772 EXAMINATION REQUIREMENTS
The examination parameters for the object chosen shall
IV-750 TECHNIQUE be determined by the rules of T-750 applicable to the ob-
ject chosen and the method of magnetization. Any of the
IV-751 QUALIFICATION STANDARD magnetizing techniques listed in T-751 may be used. The
Slotted shim(s) 0.002 in. (0.05 mm) thick having 30% same indication(s) of the shim discontinuity(ies) shall be
deep material removed as described in T-764.1.2 shall used for both black light and alternate wavelength light
be used to qualify the alternate wavelength light source examinations.
Table IV-721
Requirements for Qualifying Alternate Wavelength Light Sources for Excitation of Specific
Fluorescent Particles
Nonessential
Requirement Essential Variable Variable
Particle manufacturer and designation X . ..
Carrier (water or oil); if oil, manufacturer and type designation X . ..
Alternate wavelength light source manufacturer and model X . ..
Alternate wavelength light source meter, manufacturer, and model X . ..
Filter glasses (if needed) X . ..
Minimum alternative wavelength light intensity X . ..
Qualification records ... X
--`,,,`,`,``,``,,`,``,`,`,`,``,`-`-`,,`,,`,`,,`---
189
Copyright ASME International (BPVC)
Provided by IHS under license with ASME pe.class Licensee=PDVSA - Pequiven site 9/9986712014, User=Gil, Eliecer
No reproduction or networking permitted without license from IHS Not for Resale, 07/23/2013 06:23:05 MDT
ARTICLE 7 2013 SECTION V
IV-772.1 Examination With Black Light. The qualifi- used for magnetic particle examinations. The alternate wa-
cation standard with the attached shim(s) shall be exam- velength light source with at least the minimum intensity
ined with the established parameters and specific qualified shall be used with the specific particle designa-
particles in a darkened area with black light illumination. tion employed in the qualification.
The resulting particle indication(s) shall be photographed.
IV-772.2 Examination With Alternate Wavelength
Light. Using the same particle indication(s) examined in IV-790 DOCUMENTATION
IV-772.1, switch to the alternate wavelength light source IV-791 EXAMINATION RECORD
and adjust the light intensity by varying the distance or
For each examination, the information required in
power, to establish particle indication(s) essentially the
T-793 and the following information shall be recorded:
same as that (those) obtained with the black light above.
The light intensity shall be measured with the alternative (a) alternative wavelength light source, manufacturer,
wavelength light meter. The resulting particle and model
indication(s) shall be photographed using identical photo- (b) alternative wavelength light source meter, manufac-
graphic techniques as used for the black light. However, turer, and model
camera lens filters appropriate for use with the alternate (c) filter glasses, when necessary
wavelength light source should be used for recording (d) fluorescent particle manufacturer and designation
the indication(s), when required. (e) qualification standard identification
(f) technique details
IV-773 QUALIFICATION OF ALTERNATE (g) identification of the personnel performing and wit-
WAVELENGTH LIGHT SOURCE AND nessing the qualification
SPECIFIC PARTICLES (h) equipment and materials used
When the same particle indication(s) as achieved with (i) minimum alternate wavelength light intensity
black light can be obtained with the alternate wavelength (j) black light and alternative wavelength light qualifi-
light source, the alternate wavelength light source may be cation photos, exposure settings, and filters, if used
190 --`,,,`,`,``,``,,`,``,`,`,`,``,`-`-`,,`,,`,`,,`---
--`,,,`,`,``,``,,`,``,`,`,`,``,`-`-`,,`,,`,`,,`---
V-720 GENERAL REQUIREMENTS required, the temperature shall be recorded.
V-733 Magnetic Field Strength. A calibrated gauss-
V-720.1 Requirements. Requirements of Article 7 ap-
meter or artificial shims shall be used to determine the
ply unless modified by this Appendix.
magnetic field strength and direction on surfaces to be ex-
V-720.2 Application. To accommodate the examina- amined. The gaussmeter device shall be equipped with
tion of a variety of surfaces, a liquid polymer containing both transverse and axial field probes. Dial or similar type
ferromagnetic particles is applied to the surface instead calibrated meters of suitable range may be used, providing
of conventional dry or suspended wet particles. During they are capable of making transverse and axial measure-
the cure time, the application of magnetizing fields cause ments. Values for G (kAm-1) or the use of artificial shims
the particles to migrate and form patterns at discontinu- shall be in accordance with T-764.
ities. The polymer cures forming an elastic solid (e.g., a
rubber replica) with indications permanently fixed on its V-734 Magnification. Replica viewing may be aided
surface. by the use of magnification.
191
Copyright ASME International (BPVC)
Provided by IHS under license with ASME pe.class Licensee=PDVSA - Pequiven site 9/9986712014, User=Gil, Eliecer
No reproduction or networking permitted without license from IHS Not for Resale, 07/23/2013 06:23:05 MDT
ARTICLE 7 2013 SECTION V
Table V-721
Requirements for the Magnetic Rubber Examination Procedure
Nonessential
Requirement Essential Variable Variable
Magnetic Rubber Mix Formulations (Manufacturer’s name(s) for material of X —
various viscosities and recommended cure times)
Surface preparation X —
Magnetizing technique X —
Field strength X —
Nonmagnetic coating thickness greater than previously qualified X —
Minimum cure time as recommended by the manufacturer X —
Releasing agent X —
Temperature range as specified by the manufacturer X —
Performance demonstration, when required X —
Number of fields and directions to be applied and magnetizing time for each X —
direction
Demagnetizing — X
Personnel qualification requirements — X
--`,,,`,`,``,``,,`,``,`,`,`,``,`-`-`,,`,,`,`,,`---
Reference to the procedure qualification records — X
192
Copyright ASME International (BPVC)
Provided by IHS under license with ASME pe.class Licensee=PDVSA - Pequiven site 9/9986712014, User=Gil, Eliecer
No reproduction or networking permitted without license from IHS Not for Resale, 07/23/2013 06:23:05 MDT
2013 SECTION V ARTICLE 7
--`,,,`,`,``,``,,`,``,`,`,`,``,`-`-`,,`,,`,`,,`---
193
Copyright ASME International (BPVC)
Provided by IHS under license with ASME pe.class Licensee=PDVSA - Pequiven site 9/9986712014, User=Gil, Eliecer
No reproduction or networking permitted without license from IHS Not for Resale, 07/23/2013 06:23:05 MDT
ARTICLE 7 2013 SECTION V
NONMANDATORY APPENDIX A
MEASUREMENT OF TANGENTIAL FIELD STRENGTH WITH
GAUSSMETERS
194
Copyright ASME International (BPVC)
Provided by IHS under license with ASME pe.class Licensee=PDVSA - Pequiven site 9/9986712014, User=Gil, Eliecer
No reproduction or networking permitted without license from IHS Not for Resale, 07/23/2013 06:23:05 MDT
2013 SECTION V ARTICLE 8
ARTICLE 8
EDDY CURRENT EXAMINATION
--`,,,`,`,``,``,,`,``,`,`,`,``,`-`-`,,`,,`,`,,`---
Code Section. for surface probe eddy current examination for measuring
(c) Definitions of terms for eddy current examination nonconductive-nonmagnetic coating thicknesses.
appear in three places: Mandatory Appendix I to this Arti- (h) Mandatory Appendix VI, Eddy Current Detection
cle; Article 1, Mandatory Appendix I; and Subsection B, and Measurement of Depth of Surface Discontinuities in
Article 30. Nonmagnetic Metals With Surface Probes, provides the re-
(d) Mandatory Appendix II, Eddy Current Examination quirements for surface probe eddy current examination
of Nonferromagnetic Heat Exchanger Tubing, provides for detection of surface connected discontinuities and
the requirements for bobbin coil multifrequency and mul- measuring their depth.
tiparameter eddy current examination of installed nonfer-
romagnetic heat exchanger tubing.
195
Copyright ASME International (BPVC)
Provided by IHS under license with ASME pe.class Licensee=PDVSA - Pequiven site 9/9986712014, User=Gil, Eliecer
No reproduction or networking permitted without license from IHS Not for Resale, 07/23/2013 06:23:05 MDT