Maintenance Procedure for the

ACQUITY UPLC Binary Solvent Manager


© 2007 Waters Corporation

®
Verify completion of the maintenance procedure for the ACQUITY UPLC binary solvent manager by
completing Form 10. If you require information on how to perform any of the steps in this form, see the
Maintenance Reference Information PDF document for the binary solvent manager on the Waters®
SystemsQT™ Projects/Documentation DVD.

Form 10 Maintenance Procedure for the ACQUITY UPLC Binary Solvent Manager

ACQUITY UPLC Binary Solvent Manager Serial Number

Owner’s System ID

Procedure Completed

1. Replace the pump head seals.


Initials
2. Replace the seal wash spacer.
3. Replace the plungers.
4. Replace the primary and accumulator inlet valve cartridges.
5. Replace the mixer assembly.
6. Replace the solvent reservoir filters.
7. Perform a static decay test to ensure that the binary solvent manager is operating properly.
8. Clean or replace the binary solvent manager air filters.

Comments

Certification

The undersigned performer attests that the maintenance procedure was performed according to the procedure
shown above.

Performer Date
Signature

The undersigned reviewer accepts that the maintenance procedure was performed according to the procedure
shown above.

Reviewer Date
Signature
© 2007 Waters Corporation

715001967 Revision A 1 of 1