Calibration of Coarse Sieves


INSTRUMENT Sieve 9.5 mm CALI. CERTIFICATE NO. OEO-19-

LAB ID. NO. ENS-20 DATE PERFORMED 10/12/2019

SERIAL NO. ---- NEXT DUE DATE 10/12/2020

CALIBRATION METHOD OEO-ICP-29 CALIBRATION TEMPERATURE 25.0 °C

TRIAL NO. 1 2 3 4 5 6 7 8 9 10

Horizontal (1) 9.57 9.58 9.56 9.57 9.56 9.56 9.57 9.56 9.54 9.56

Aperture Size
Vertical (2) 9.57 9.56 9.54 9.56 9.54 9.54 9.56 9.57 9.56 9.55
(mm)

Wire diameter (mm) 2.26 2.26 2.26 2.26 2.26 2.26 2.26 2.26 2.26 2.26

TRIAL NO. 11 12 13 14 15 16 17 18 19 20

Horizontal (1) 9.56 9.57 9.57 9.56 9.57 9.56 9.57 9.56 9.57 9.56
Aperture Size
(mm)
Vertical (2) 9.58 9.57 9.58 9.58 9.56 9.57 9.58 9.57 9.56 9.57

Wire diameter (mm) 2.26 2.26 2.27 2.26 2.26 2.27 2.26 2.27 2.26 2.27

TRIAL NO. 21 22 23 24 25 26 27 28 29 30

Horizontal (1) 9.58 9.59 9.58 9.57 9.57 9.56 9.56 9.57 9.57 9.56

Aperture Size
Vertical (2) 9.55 9.56 9.56 9.57 9.56 9.56 9.55 9.56 9.57 9.56
(mm)

Wire diameter (mm) 2.26 2.27 2.26 2.27 2.26 2.27 2.26 2.27 2.26 2.27

9.59
Horizontal (1) 9.56 Tolerance 9.5  0.300 mm Maximum Individual Openings Max 10.16 mm
Average
2.26
Vertical (2) 9.56 Tolerance 9.5  0.300 mm Average Wire Diameter 2.24±0.34 mm

VISUAL CHECKS √ RESULT OF CALIBRATION Pass ed

REFERENCE INSTRUMENTS USED FOR CALIBRATION


Instrument Laboratory No. Calibrated On & Certificate No. Calibrated Due date
Vernier Caliper EKQ-27 15 April, 2019 142462 15 April 2020

Frame diameter ( mm) 203 +0.76-0.00


Frame Height (mm) 50

Laboratory Management System Based on ISO 17025 & ISO9001