You are on page 1of 1

.

Calibration of Coarse Sieves

INSTRUMENT Sieve 4.75 mm CALI. CERTIFICATE NO. OEO-19-

LAB ID. NO. ENS-19 DATE PERFORMED 10/12/2019

SERIAL NO. ---- NEXT DUE DATE 10/12/2020

CALIBRATION METHOD OEO-ICP-29 CALIBRATION TEMPERATURE 25.0 °C

TRIAL NO. 1 2 3 4 5 6 7 8 9 10

Horizontal (1) 4.73 4.77 4.76 4.75 4.75 4.77 4.75 4.76 4.75 4.75
Aperture Size
(mm)
Vertical (2) 4.76 4.76 4.77 4.78 4.76 4.78 4.75 4.76 4.78 4.78

Wire diameter (mm) 1.58 1.58 1.58 1.59 1.59 1.58 1.59 1.58 1.58 1.59

TRIAL NO. 11 12 13 14 15 16 17 18 19 20

Horizontal (1) 4.75 4.76 4.75 4.75 4.76 4.77 4.75 4.78 4.75 4.77
Aperture Size
(mm)
Vertical (2) 4.75 4.75 4.75 4.75 4.75 4.77 4.75 4.76 4.77 4.75

Wire diameter (mm) 1.59 1.59 1.59 1.58 1.59 1.59 1.58 1.59 1.58 1.59

TRIAL NO. 21 22 23 24 25 26 27 28 29 30

Horizontal (1) 4.76 4.77 4.76 4.77 4.75 4.76 4.77 4.78 4.77 4.78
Aperture Size
(mm)
Vertical (2) 4.77 4.78 4.78 4.78 4.77 4.79 4.78 4.78 4.78 4.78

Wire diameter (mm) 1.59 1.59 1.59 1.59 1.59 1.59 1.59 1.59 1.59 1.59

Maximum Individual
Horizontal (1) 4.76 Tolerance 4.75  0.150 mm 4.79 Max 5.14 mm
Openings
Average
Vertical (2) 4.78 Tolerance 4.75  0.150 mm Average Wire Diameter 1.59 1.60±0.24 mm

Frame diameter ( mm) 203 +0.76-0.00


Frame Height (mm) 50

VISUAL CHECKS √ RESULT OF CALIBRATION Passed


REFERENCE INSTRUMENTS USED FOR CALIBRATION
Instrument Laboratory No. Calibrated On & Certificate No. Calibration Due date
Vernier Caliper EKQ-27 15-4-2019 142462 15/4/2020

Laboratory Management System Based on ISO 17025&ISO9001


Document Number Revision No. Effective Date Page Sequence