You are on page 1of 1

Start

Attention Box
asking the user to
remove any beam
obstruction

Post Message
Convertor FAIL
to run DAS Gain QUIT
Board Check?
Cal

PASS RETRY
Log Error in error log
that user hit retry

Check for Post Message


FAIL asking user to USER Log Error in error log
dirty mylar
retry, continue or Response that user hit cancel
window quit

Log Error in error Ignore


PASS log that user
ignored the msg
QUIT

Is Warmup YES Do Tube Warmup


needed? (Cold, WarmupI
and II) as required

NO

Collimator Cal of
one station

YES Post Message


Processing
that Collimator Cal QUIT
errors?
failed
NO

NO RETRY

Post Message:
Need ZFET YES ZFET Settings Change ZFET
Setting
are being setting
change?
Changed

NO

prep step saved in Unix file


Cal DB update
COLLIMATOR CAL BL OCK
Save history log DIAGRAM

All 8 stations YES Post Message


QUIT
complete? to run Fastcal