DIRECTION 5797904-1EN, REVISION 1 BRIVO CT385 OC UPGRADE INSTRUCTION
Page 2
GE COMPANY
DIRECTION 5797904-1EN, REVISION 1 BRIVO CT385 OC UPGRADE INSTRUCTION
LEGAL NOTES
TRADEMARKS
Adobe, the Adobe logo, Acrobat, the Acrobat logo, Exchange, and PostScript are trademarks of
Adobe Systems Incorporated or its subsidiaries and may be registered in certain jurisdictions.
Microsoft is a registered trademark and Windows is a trademark of Microsoft Corporation.
All other products and their name brands are trademarks of their respective holders.
COPYRIGHTS
All Material, Copyright © 2018 by GE Company, Inc. All rights reserved.
IMPORTANT PRECAUTIONS
LANGUAGE
警告 • 本维修手册仅提供英文版本。
(ZH-CN) • 如果维修服务提供商需要非英文版本,客户需自行提供翻译服务。
• 未详细阅读和完全理解本维修手册之前,不得进行维修。
• 忽略本警告可能对维修人员,操作员或患者造成触电、机械伤害或其他形式
的伤害。
Warning Page 5
GE COMPANY
DIRECTION 5797904-1EN, REVISION 1 BRIVO CT385 OC UPGRADE INSTRUCTION
Page 6 Warning
GE COMPANY
DIRECTION 5797904-1EN, REVISION 1 BRIVO CT385 OC UPGRADE INSTRUCTION
警告 • このサービスマニュアルには英語版しかありません。
(JA) • サービスを担当される業者が英語以外の言語を要求される場合、翻訳作業
はその業者の責任で行うものとさせていただきます。
• このサービスマニュアルを熟読し理解せずに、装置のサービスを行わない
でください。
• この警告に従わない場合、サービスを担当される方、操作員あるいは患者
さんが、感電や機械的又はその他の危険により負傷する可能性があります。
Warning Page 7
GE COMPANY
DIRECTION 5797904-1EN, REVISION 1 BRIVO CT385 OC UPGRADE INSTRUCTION
Page 8 Warning
GE COMPANY
DIRECTION 5797904-1EN, REVISION 1 BRIVO CT385 OC UPGRADE INSTRUCTION
Warning Page 9
GE COMPANY
DIRECTION 5797904-1EN, REVISION 1 BRIVO CT385 OC UPGRADE INSTRUCTION
警告 • 本服務手冊僅提供英文版本。
(ZH-HK) • 倘若客戶的服務供應商需要英文以外之服務手冊,客戶有責任提供翻譯服
務。
• 除非已參閱本服務手冊及明白其內容,否則切勿嘗試維修設備。
• 不遵從本警告或會令服務供應商、網絡供應商或病人受到觸電、機械性或其
他的危險。
警告 • 本維修手冊僅有英文版。
(ZH-TW) • 若客戶的維修廠商需要英文版以外的語言,應由客戶自行提供翻譯服務。
• 請勿試圖維修本設備,除非 您已查閱並瞭解本維修手冊。
• 若未留意本警告,可能導致維修廠商、操作員或病患因觸電、機械或其他危
險而受傷。
Page 10 Warning
GE COMPANY
DIRECTION 5797904-1EN, REVISION 1 BRIVO CT385 OC UPGRADE INSTRUCTION
DAMAGE IN TRANSPORTATION
All packages should be closely examined at time of delivery. If damage is apparent write “Damage
In Shipment” on ALL copies of the freight or express bill BEFORE delivery is accepted or “signed
for” by a GE representative or hospital receiving agent. Whether noted or concealed, damage
MUST be reported to the carrier immediately upon discovery, or in any event, within 14 days after
receipt, and the contents and containers held for inspection by the carrier. A transportation company
will not pay a claim for damage if an inspection is not requested within this 14 day period.
To file a report:
• Call 1-800-548-3366 and use option 6
• Fill out the GIQ workflow for any items missing, damaged, OBF/FOI for in process installs:
http://supportcentral.ge.com/ProcessMaps/
form_new_request.asp?prod_id=268679&form_id=573167&node_id=1916016&map_id=&ref
erence_id=&reference_type
• Contact the local service coordinator.
Warning Page 11
GE COMPANY
DIRECTION 5797904-1EN, REVISION 1 BRIVO CT385 OC UPGRADE INSTRUCTION
IMPORTANT...X-RAY PROTECTION
X-ray equipment if not properly used may cause injury. Accordingly, the instructions herein
contained should be thoroughly read and understood by everyone who will use the equipment
before you attempt to place this equipment in operation. The General Electric Company, Medical
Systems Group, will be glad to assist and cooperate in placing this equipment in use.
Although this apparatus incorporates a high degree of protection against x-radiation other than the
useful beam, no practical design of equipment can provide complete protection. Nor can any
practical design compel the operator to take adequate precautions to prevent the possibility of any
persons carelessly exposing themselves or others to radiation.
It is important that anyone having anything to do with x-radiation be properly trained and fully
acquainted with the recommendations of the National Council on Radiation Protection and
Measurements as published in NCRP Reports available from NCRP Publications, 7910 Woodmont
Avenue, Room 1016, Bethesda, Maryland 20814, and of the International Commission on
Radiation Protection, and take adequate steps to protect against injury.
The equipment is sold with the understanding that the General Electric Company, Medical Systems
Group, its agents, and representatives have no responsibility for injury or damage which may result
from improper use of the equipment.
Various protective materials and devices are available. It is urged that such materials or devices be
used.
CAUTION Danger of explosion if battery is incorrectly replaced. Replace only with the same or
Risk of equivalent type recommended by the manufacturer. Discard used batteries according to the
Explosion manufacturer’s instructions.
Page 12 Warning
GE COMPANY
DIRECTION 5797904-1EN, REVISION 1 BRIVO CT385 OC UPGRADE INSTRUCTION
Revision History
Table of Contents
Chapter 1
Upgrade Overview ................................................................................................. 17
Section 1.0
Flowchart .......................................................................................................... 17
Section 2.0
Material List ...................................................................................................... 18
Chapter 2
Upgrade Procedure ............................................................................................... 19
Section 1.0
Preparation in Site............................................................................................ 19
1.1 Preparation ...................................................................................................................... 19
1.2 Record System INFO Information.................................................................................... 20
1.2.1 Record Power Option ......................................................................................... 20
1.2.2 Record All Remote Host Network Information .................................................... 21
1.2.3 Record Product Network Filters (PNF) Information ............................................ 22
1.2.4 Record AW IP for Direct Information .................................................................. 22
1.3 Save System State .......................................................................................................... 23
Section 2.0
NIO16 Console Upgrade .................................................................................. 24
2.1 Old Console (Z800) Removal .......................................................................................... 24
2.2 New Console (Z820) Installation...................................................................................... 25
2.2.1 Preparative ......................................................................................................... 25
2.2.2 Unpack NIO16 Console with Z820 PC ............................................................... 28
2.2.3 GPU Card Installation ......................................................................................... 29
2.2.4 NIO16 Console Cable Connections.................................................................... 30
2.2.5 Console Cover Installation.................................................................................. 34
Section 3.0
System Configuration ...................................................................................... 34
3.1 Restore System State...................................................................................................... 34
3.2 Option Installation ............................................................................................................ 35
3.2.1 Internal Landmark Option ................................................................................... 35
3.2.2 Power Option Installation.................................................................................... 35
3.2.3 Recon Enhancement Key ................................................................................... 35
3.2.4 Standard Option.................................................................................................. 35
3.2.5 Other Commercial Option ................................................................................... 35
3.3 Configure System Information on Z820 (SUSE).............................................................. 36
3.4 Flash Download .............................................................................................................. 38
3.5 GPU Card Checking ........................................................................................................ 39
Section 4.0
Finalization........................................................................................................ 40
Chapter 1
Upgrade Overview
The Upgrade Kit is only available for Brivo CT385 IB system with NIO16 Console Z800 Host
Computer. The upgrade instruction is to guide FE how to upgrade NIO16 Console Host Computer
from Z800 (App version: 11BW49.xx) to Z820 (App version: 14BW25.xx) on CT-only Brivo CT385
system.
Section 1.0
Flowchart
Start
60mins Finalization
Section 2.0
Material List
Verify that the contents of upgrade kit are correct according to the material list below.
Description Number
CT385 TP/SW Collector B77082CB
NIO Console GEHW IEC3 B76992CB
CT385 IB Upgrade Kit B75132DA
(5797952)
Upgrade Instruction 5797904-1EN
DP to DVI Cable, 3 meter 5408703
Table 1-1 Brivo CT385 IB Upgrade Package (B75122DA)
Description Number
Internal Laser Light Kit B76292CB
Note: This is a Mandatory Option for system without Internal Laser Light.
Table 1-2 CT385 Internal Laser Light (Landmark) Kit
Description Number
ASiR Non ASiR
Console Cover Kit B76562CA B70762RT
System Grade Collector B77042CB B74802CA
Table 1-3 ASiR and Non-ASiR Kit
Chapter 2
Upgrade Procedure
Section 1.0
Preparation in Site
1.1 Preparation
Please confirm Internal Laser Lights, if no, you must order Internal Laser Light Kit (B76292CB) to
install before starting this upgrade.
Figure 2-1 Control Panel without Internal Laser Lights
Write down all of the system INFO information in Console Information Sheets, please refer to
5458022-2EN Service Methods CD.
NOTICE If the System Power Rating Plate and the Power Option Label (40kVA or 30kVA) are attached
on the system, the priority of Option Label (Modified Label) is higher than System Power
Rating Plate.
Figure 2-3 System Power Rating Plate
Refer to the latest service methods (5458022-2EN) to understand the detailed Save State
Procedure and other information saving procedure.
Note: If a Save State was just completed, it can be omitted here.
Note: When performing Save/Restore System State, make sure there is ONLY one USB storage
device plugged in USB ports of both Console and Media Tower simultaneously, otherwise it
may result in incorrect data storage.
1.) Insert the System State DVD-RAM media into the peripheral tower drive or USB media in any
of the console’s USB ports.
2.) Select [Service Desktop] -> [Utilities].
3.) Select [System State] for DVD-RAM media or [System State - USB] for USB media. The
System State Save/Restore screen appears.
4.) Select [All] to select all the cals, characterization, etc.
5.) Select [Save].
6.) If DVD-RAM or USB is ready, select [Yes].
7.) When completed select [Dismiss].
8.) Remove System State DVD-RAM media from the peripheral tower or USB media from USB
port.
Section 2.0
NIO16 Console Upgrade
2.1 Old Console (Z800) Removal
NOTICE Make Sure to follow all Lockout/Tagout requirements while performing this procedure.
1.) Move the cradle all the way out of the gantry and lower to its lowest point.
2.) Shut down the system and power off the Operator Console power switch.
3.) Switch OFF the breaker in the Power Distribution Box (hospital supply).
Note: If there is UPS, switch OFF the UPS output.
4.) Visually verify proper labeling of each cable and disconnect all console rear cables.
5.) Disconnect the cables between the old console and system (Gantry, PDU and local hospital
network).
6.) Remove the old OC aside.
2.2.1 Preparative
Figure 2-6 System Interconnect Diagram
GSCB Gantry
Blue
Green
Gantry Interface
HRST
IN
DB9 DB25
OUT
HOST
COM
AUDIO IN TGP
7
AUDIO OUT
USB2.0
Board
1
KeyBoard
8
USB2.0
USB3.0
2
Mouse
USB2.0
3
Service Key 9 TrackBall
USB2.0
USB3.0
BarCode DCB
4
USB2.0
Reader Board
5
EXT HDD
USB3.0
Figure 2-7 NIO16 OC Interconnect (with Z820)
SWITCH 6
DVD RW
USB3.0
HUB
8XRJ45 RJ45
1 2
ETH3
SRV
DVI-D DVI-D
RX TX RX TX DB9
INJ
DIP
ECG
UPS DP-0 DP-1 DP-2 DP-3
Video
1 2 3 4
Card DVI-D
Page 27
GE COMPANY
DIRECTION 5797904-1EN, REVISION 1 BRIVO CT385 OC UPGRADE INSTRUCTION
d.) Remove the two front cushions from the bottom of the skid.
Figure 2-9 Cushion on Botton of Skid
e.) Lift up on the strap on the front of the step-board (Figure 2-8) to lower the skid. Remove
the step-board.
f.) Ensure the console stabilizers are in line with thee notched portion at the front of the skid.
This will allow enough clearance to smoothly roll the console down the ramps.
4.) Remove the Host PC support bracket from left side of Z820.
Figure 2-11 Host PC Support Bracket
120VAC
GND
# ITEM Description
1 DB9/F Black Host Computer DIP Serial Port
2 DB9/F Gray Host Computer RS232
3 Green Audio Host Computer Audio Out (Green)
4 Blue Audio Host Computer Audio In (Blue)
5 USB Not Used
6 DB9/M Black AC Box J56
7 DB25/M Black TGP Gantry Cable
3.) Connect Media Tower cable connectors to NIO16 Console AC Box and Z820 Host Computer.
4.) Connect the LCD Monitors for Z820 Host Computer.
a.) Remove the old Video cable (5332107-2) from the LCD Monitors.
b.) Connect the new Video cable (5408703, shipped with upgrade kit) between Z820 Host
Computer and LCD Monitor.
Page 32 Chapter 2 - Upgrade Procedure
GE COMPANY
DIRECTION 5797904-1EN, REVISION 1 BRIVO CT385 OC UPGRADE INSTRUCTION
Figure 2-15 Video Cable and Power Cable for LCD Monitor
c.) Connect the LCD Monitor power cables to NIO16 Console AC Box.
- Scan Monitor
Video Cable: from Console Host DP1 to Monitor DVI
Power Cable: from Console AC Box J10 to Monitor
- Image Monitor
Video Cable: from Console Host DP0 to Monitor DVI
Power Cable: from Console AC Box J9 to Monitor
7.) Connect the NIO16 Console power cable and Ground cable to the Console Power Panel.
Section 3.0
System Configuration
3.1 Restore System State
CAUTION If system is 30kVA power, please DON’T install 40kVA power option to avoid power overload.
Note: The Flash Download takes anywhere from 5-30 minutes, depending on which subsystems require
updates.
1.) Perform the Flash Download Utility found on the Common Service Desktop - Utilities Tab,
select [Flash Download].
Figure 18 Common Service Desktop - Utilities Tab, Flash Download
3.) Once the Gantry Hardware Flash Downloads successfully, select [Dismiss].
4.) Close the Common Service Desktop.
Open a shell, then type: cat /proc/ati/1/biosversion. Check the result as below:
Section 4.0
Finalization
1.) Perform Quality Assurance Test of Functional Checks on Service Methods (5457941-2EN),
make sure all of them pass.
2.) Save System State
a.) Insert the Save System State DVD into the Peripheral Tower DVD-RW drive.
b.) Select [SERVICE DESKTOP].
c.) If reloading software, select [UTILITIES].
d.) Select [SYSTEM STATE].
e.) Select [ALL] to select all the cals, characterization, etc.
f.) Select [SAVE].
g.) If applicable, select [OK] when the following message appears: System State Media
Status: Please insert a DVD into the drive and press Save again
h.) When completed select [DISMISS].
i.) Close the Service Desktop window in the upper left corner of the screen.
3.) Old Console Disposal
The old console has components that are needed to cover FRU shortages, so please
RETURN the old console. Follow the normal local return process to return the console.
a.) Package the old console with the shipping material.
b.) Mark type of the old console on the outside of the box.
c.) Return the old console:
- For U.S. sites
Call Schenker Heavyweight Return at 866-229-7877 (Option 25), and ship to:
GE Healthcare Renewable Resources
7624 South 10th Street
Oak Creek, WI. 53154
- For Western Europe sites
Stick the harvest sticker (see below) on the OC box & ship (Heavyweight Return) to:
P.Vinet - Volta Repair Center Manager
Gems, 283 rue de la Minière 78533 Buc
VIA St WITZ