You are on page 1of 42

Brivo CT385 OC Upgrade Instruction

Service Documentation

5797904-1EN
Rev 1
© 2018, GE Company,
All Rights Reserved.
GE COMPANY
DIRECTION 5797904-1EN, REVISION 1 BRIVO CT385 OC UPGRADE INSTRUCTION

Page 2
GE COMPANY 
DIRECTION 5797904-1EN, REVISION 1 BRIVO CT385 OC UPGRADE INSTRUCTION

LEGAL NOTES

TRADEMARKS
Adobe, the Adobe logo, Acrobat, the Acrobat logo, Exchange, and PostScript are trademarks of
Adobe Systems Incorporated or its subsidiaries and may be registered in certain jurisdictions.
Microsoft is a registered trademark and Windows is a trademark of Microsoft Corporation.
All other products and their name brands are trademarks of their respective holders.

COPYRIGHTS
All Material, Copyright © 2018 by GE Company, Inc. All rights reserved.

Legal Notes Page 3


GE COMPANY 
DIRECTION 5797904-1EN, REVISION 1 BRIVO CT385 OC UPGRADE INSTRUCTION

This page intentionally left blank.

Page 4 Legal Notes


GE COMPANY 
DIRECTION 5797904-1EN, REVISION 1 BRIVO CT385 OC UPGRADE INSTRUCTION

IMPORTANT PRECAUTIONS

LANGUAGE

警告 • 本维修手册仅提供英文版本。
(ZH-CN) • 如果维修服务提供商需要非英文版本,客户需自行提供翻译服务。
• 未详细阅读和完全理解本维修手册之前,不得进行维修。
• 忽略本警告可能对维修人员,操作员或患者造成触电、机械伤害或其他形式
的伤害。

VÝSTRAHA • Tento provozní návod existuje pouze v anglickém jazyce.


• V případě, že externí služba zákazníkům potřebuje návod v jiném jazyce, je
(CS) zajištění překladu do odpovídajícího jazyka úkolem zákazníka.
• Nesnažte se o údržbu tohoto zařízení, aniž byste si přečetli tento provozní
návod a pochopili jeho obsah.
• V případě nedodržování této výstrahy může dojít k poranění pracovníka
prodejního servisu, obslužného personálu nebo pacientů vlivem
elektrického proudu, respektive vlivem mechanických či jiných rizik.

ADVARSEL • Denne servicemanual findes kun på engelsk.


• Hvis en kundes tekniker har brug for et andet sprog end engelsk, er det
(DA) kundens ansvar at sørge for oversættelse.
• Forsøg ikke at servicere udstyret medmindre denne servicemanual har
været konsulteret og er forstået.
• Manglende overholdelse af denne advarsel kan medføre skade på grund af
elektrisk, mekanisk eller anden fare for teknikeren, operatøren eller
patienten.

WAARSCHUWING • Deze onderhoudshandleiding is enkel in het Engels verkrijgbaar.


• Als het onderhoudspersoneel een andere taal vereist, dan is de klant
(NL) verantwoordelijk voor de vertaling ervan.
• Probeer de apparatuur niet te onderhouden voordat deze
onderhoudshandleiding werd geraadpleegd en begrepen is.
• Indien deze waarschuwing niet wordt opgevolgd, zou het
onderhoudspersoneel, de operator of een patiënt gewond kunnen raken als
gevolg van een elektrische schok, mechanische of andere gevaren.

WARNING • This Service Manual is available in English only.


• If a customer's service provider requires a language other than English, it is
(EN) the customer's responsibility to provide translation services.
• Do not attempt to service the equipment unless this service manual has
been consulted and is understood.
• Failure to heed this warning may result in injury to the service provider,
operator, or patient, from electric shock or from mechanical or other hazards.

Warning Page 5
GE COMPANY 
DIRECTION 5797904-1EN, REVISION 1 BRIVO CT385 OC UPGRADE INSTRUCTION

HOIATUS • Käesolev teenindusjuhend on saadaval ainult inglise keeles.


• Kui klienditeeninduse osutaja nõuab juhendit inglise keelest erinevas
(ET) keeles, vastutab klient tõlketeenuse osutamise eest.
• Ärge üritage seadmeid teenindada enne eelnevalt
käesoleva teenindusjuhendiga tutvumist ja sellest aru saamist.
• Käesoleva hoiatuse eiramine võib põhjustada teenuseosutaja, operaatori
või patsiendi vigastamist elektrilöögi, mehaanilise või muu ohu tagajärjel.

VAROITUS • Tämä huolto-ohje on saatavilla vain englanniksi.


• Jos asiakkaan huoltohenkilöstö vaatii muuta kuin englanninkielistä
(FI) materiaalia, tarvittavan käännöksen hankkiminen on asiakkaan vastuulla.
• Älä yritä korjata laitteistoa ennen kuin olet varmasti lukenut ja ymmärtänyt
tämän huolto-ohjeen.
• Mikäli tätä varoitusta ei noudateta, seurauksena voi olla huoltohenkilöstön,
laitteiston käyttäjän tai potilaan vahingoittuminen sähköiskun, mekaanisen
vian tai muun vaaratilanteen vuoksi.

ATTENTION • Ce manuel de service n'est disponible qu'en anglais.


• Si le technicien du client a besoin de ce manuel dans une autre langue que
(FR) l'anglais, c'est au client qu'il incombe de le faire traduire.
• Ne pas tenter d'intervenir sur les équipements tant que le manuel service n'a
pas été consulté et compris
• Le non-respect de cet avertissement peut entraîner chez le technicien,
l'opérateur ou le patient des blessures dues à des dangers électriques,
mécaniques ou autres.

WARNUNG • Diese Serviceanleitung existiert nur in Englischer Sprache.


• Falls ein fremder Kundendienst eine andere Sprache benötigt, ist es
(DE) aufgabe des Kunden für eine Entsprechende Übersetzung zu sorgen.
• Versuchen Sie nicht diese Anlage zu warten,
ohne diese Serviceanleitung gelesen und verstanden zu haben.
• Wird diese Warnung nicht beachtet, so kann es zu Verletzungen des
Kundendiensttechnikers, des Bedieners oder des Patienten durch
stromschläge, Mechanische oder Sonstige gefahren kommen.

ΠΡΟΕΙΔΟΠΟΙΗΣΗ • Το παρόν εγχειρίδιο σέρβις διατίθεται στα αγγλικά μόνο.


• Εάν το άτομο παροχής σέρβις ενός πελάτη απαιτεί το παρόν εγχειρίδιο σε
(EL) γλώσσα εκτός των αγγλικών, αποτελεί ευθύνη του πελάτη να παρέχει
υπηρεσίες μετάφρασης.
• Μην επιχειρήσετε την εκτέλεση εργασιών σέρβις στον εξοπλισμό εκτός εαν
έχετε συμβουλευτεί και έχετε κατανοήσει το παρόν εγχειρίδιο σέρβις.
• Εαν δε λάβετε υπόψη την προειδοποίηση αυτή, ενδέχεται να προκληθεί
τραυματισμός στο άτομο παροχής σέρβις, στο χειριστή ή στον ασθενή από
ηλεκτροπληξία, μηχανικούς ή άλλους κινδύνους.

Page 6 Warning
GE COMPANY 
DIRECTION 5797904-1EN, REVISION 1 BRIVO CT385 OC UPGRADE INSTRUCTION

FIGYELMEZTETÉS • Ezen karbantartási kézikönyv kizárólag angol nyelven érhető el.


(HU) • Ha a vevő szolgáltatója angoltól eltérő nyelvre tart igényt, akkor a vevő
felelőssége a fordítás elkészíttetése.
• Ne próbálja elkezdeni használni a berendezést, amíg a karbantartási
kézikönyvben leírtakat nem értelmezték.
• Ezen figyelmeztetés figyelmen kívül hagyása a szolgáltató, működtető vagy
a beteg áramütés, mechanikai vagy egyéb veszélyhelyzet miatti sérülését
eredményezheti.

AÐVÖRUN • Þessi þjónustuhandbók er eingöngu fáanleg á ensku.


• Ef að þjónustuveitandi viðskiptamanns þarfnast annas tungumáls en ensku,
(IS) er það skylda viðskiptamanns að skaffa tungumálaþjónustu.
• Reynið ekki að afgreiða tækið nema að þessi þjónustuhandbók
hefur verið skoðuð og skilin.
• Brot á sinna þessari aðvörun getur leitt til meiðsla á þjónustuveitanda,
stjórnanda eða sjúklings frá raflosti, vélrænu eða öðrum áhættum.

AVVERTENZA • Il presente manuale di manutenzione è disponibile soltanto in inglese.


• Se un addetto alla manutenzione richiede il manuale in una lingua diversa, il
(IT) cliente è tenuto a provvedere direttamente alla traduzione.
• Si proceda alla manutenzione dell'apparecchiatura solo dopo aver
consultato il presente manuale ed averne compreso il contenuto
• Il non rispetto della presente avvertenza potrebbe far compiere operazioni
da cui derivino lesioni all'addetto, alla manutenzione, all'utilizzatore ed al
paziente per folgorazione elettrica, per urti meccanici od altri rischi.

警告 • このサービスマニュアルには英語版しかありません。
(JA) • サービスを担当される業者が英語以外の言語を要求される場合、翻訳作業
はその業者の責任で行うものとさせていただきます。
• このサービスマニュアルを熟読し理解せずに、装置のサービスを行わない
でください。
• この警告に従わない場合、サービスを担当される方、操作員あるいは患者
さんが、感電や機械的又はその他の危険により負傷する可能性があります。

경고 • 본 서비스 지침서는 영어로만 이용하실 수 있습니다 .


(KO) • 고객의 서비스 제공자가 영어 이외의 언어를 요구할 경우 , 번역 서비스를
제공하는 것은 고객의 책임입니다 .
• 본 서비스 지침서를 참고했고 이해하지 않는 한은 해당 장비를 수리하려고
시도하지 마십시오 .
• 이 경고에 유의하지 않으면 전기 쇼크 , 기계상의 혹은 다른 위험으로부터
서비스 제공자 , 운영자 혹은 환자에게 위해를 가할 수 있습니다 .

Warning Page 7
GE COMPANY 
DIRECTION 5797904-1EN, REVISION 1 BRIVO CT385 OC UPGRADE INSTRUCTION

BRDINJUMS • Šī apkalpes rokasgrāmata ir pieejama tikai angļu valodā.


• Ja klienta apkalpes sniedzējam nepieciešama informācija citā valodā, nevis
(LV) angļu, klienta pienākums ir nodrošināt tulkošanu.
• Neveiciet aprīkojuma apkalpi bez apkalpes rokasgrāmatas izlasīšanas un
saprašanas.
• Šī brīdinājuma neievērošana var radīt elektriskās strāvas trieciena,
mehānisku vai citu risku izraisītu traumu apkalpes sniedzējam, operatoram
vai pacientam.

ĮSPĖJIMAS • Šis eksploatavimo vadovas yra prieinamas tik anglų kalba.


• Jei kliento paslaugų tiekėjas reikalauja vadovo kita kalba – ne anglų,
(LT) numatyti vertimo paslaugas yra kliento atsakomybė.
• Nemėginkite atlikti įrangos techninės priežiūros, nebent atsižvelgėte į šį
eksploatavimo vadovą ir jį supratote.
• Jei neatkreipsite dėmesio į šį perspėjimą, galimi sužalojimai dėl elektros
šoko, mechaninių ar kitų pavojų paslaugų tiekėjui, operatoriui ar pacientui.

ADVARSEL • Denne servicehåndboken finnes bare på engelsk.


• Hvis kundens serviceleverandør trenger et annet språk, er det kundens
(NO) ansvar å sørge for oversettelse.
• Ikke forsøk å reparere utstyret uten at denne servicehåndboken er lest og
forstått.
• Manglende hensyn til denne advarselen kan føre til at serviceleverandøren,
operatøren eller pasienten skades på grunn av elektrisk støt, mekaniske
eller andre farer.

OSTRZEŻENIE • Niniejszy podręcznik serwisowy dostępny jest jedynie w języku angielskim.


• Jeśli dostawca usług klienta wymaga języka innego niż angielski,
(PL) zapewnienie usługi tłumaczenia jest obowiązkiem klienta.
• Nie próbować serwisować wyposażenia bez zapoznania się i zrozumienia
niniejszego podręcznika serwisowego.
• Niezastosowanie się do tego ostrzeżenia może spowodować urazy
dostawcy usług, operatora lub pacjenta w wyniku porażenia elektrycznego,
zagrożenia mechanicznego bądź innego.

ATENÇÃO • Este manual de assistência técnica só se encontra disponível em inglês.


• Se qualquer outro serviço de assistência técnica, que não a gems, solicitar
(PT-BR) estes manuais noutro idioma, é da responsabilidade do cliente fornecer os
serviços de tradução.
• Não tente reparar o equipamento sem ter consultado ecompreendido este
manual de assistência técnica.
• O não cumprimento deste aviso pode por em perigo a segurança do
técnico, operador ou paciente devido a choques elétricos, mecânicos ou
outros.

Page 8 Warning
GE COMPANY 
DIRECTION 5797904-1EN, REVISION 1 BRIVO CT385 OC UPGRADE INSTRUCTION

ATENÇÃO • Este manual de assistência técnica só se encontra disponível em inglês.


• Se qualquer outro serviço de assistência técnica, que não a gems, solicitar
(PT-PT) estes manuais noutro idioma, é da responsabilidade do cliente fornecer os
serviços de tradução.
• Não tente reparar o equipamento sem ter consultado ecompreendido este
manual de assistência técnica.
• O não cumprimento deste aviso pode colocar em perigo a segurança do
técnico, do operador ou do paciente devido a choques eléctricos,
mecânicos ou outros.

ATENŢIE • Acest manual de service este disponibil numai în limba engleză.


• Dacă un furnizor de servicii pentru clienţi necesită o altă limbă decât cea
(RO) engleză, este de datoria clientului să furnizeze o traducere.
• Nu încercaţi să reparaţi echipamentul decât ulterior consultării şi înţelegerii
acestui manual de service.
• Ignorarea acestui avertisment ar putea duce la rănirea depanatorului,
operatorului sau pacientului în urma pericolelor de electrocutare, mecanice
sau de altă natură.

ОСТОРОЖНО! • Данное руководство по обслуживанию предлагается только на


английском языке.
(RU) • Если сервисному персоналу клиента необходимо руководство не на
английском, а на каком-то другом языке, клиенту следует
самостоятельно обеспечить перевод.
• Перед обслуживанием оборудования обязательно обратитесь к
данному руководству и поймите изложенные в нем сведения.
• Несоблюдение требований данного предупреждения может привести к
тому, что специалист по обслуживанию, оператор или пациент получат
удар электрическим током, механическую травму или другое
повреждение.

UPOZORENJE • Ovo servisno uputstvo je dostupno samo na engleskom jeziku.


• Ako klijentov serviser zahteva neki drugi jezik, klijent je dužan da obezbedi
(SR) prevodilacke usluge.
• Ne pokušavajte da opravite uredaj ako niste procitali i razumeli ovo servisno
uputstvo.
• Zanemarivanje ovog upozorenja može dovesti do povredivanja servisera,
rukovaoca ili pacijenta usled strujnog udara ili mehanickih i drugih opasnosti.

UPOZORNENIE • Tento návod na obsluhu je k dispozícii len v angličtine.


• Ak zákazníkov poskytovateľ služieb vyžaduje iný jazyk ako angličtinu,
(SK) poskytnutie prekladateľských služieb je zodpovednosťou zákazníka.
• Nepokúšajte sa o obsluhu zariadenia skôr, ako si neprečítate návod na
obluhu a neporozumiete mu.
• Zanedbanie tohto upozornenia môže vyústiť do zranenia poskytovateľa
služieb, obsluhujúcej osoby alebo pacienta elektrickým prúdom, do
mechanického alebo iného nebezpečenstva.

Warning Page 9
GE COMPANY 
DIRECTION 5797904-1EN, REVISION 1 BRIVO CT385 OC UPGRADE INSTRUCTION

ATENCION • Este manual de servicio sólo existe en inglés.


• Si el encargado de mantenimiento de un cliente necesita un idioma que no
(ES) sea el inglés, el cliente deberá encargarse de la traducción del manual.
• No se deberá dar servicio técnico al equipo, sin haber consultado y
comprendido este manual de servicio.
• La no observancia del presente aviso puede dar lugar a que el proveedor de
servicios, el operador o el paciente sufran lesiones provocadas por causas
eléctricas, mecánicas o de otra naturaleza.

VARNING • Den här servicehandboken finns bara tillgänglig på engelska.


• Om en kunds servicetekniker har behov av ett annat språk än engelska
(SV) ansvarar kunden för att tillhandahålla översättningstjänster.
• Försök inte utföra service på utrustningen om du inte har läst och förstår den
här servicehandboken.
• Om du inte tar hänsyn till den här varningen kan det resultera i skador på
serviceteknikern, operatören eller patienten till följd av elektriska stötar,
mekaniska faror eller andra faror.

DIKKAT • Bu servis kilavuzunun sadece ingilizcesi mevcuttur.


• Eğer müşteri teknisyeni bu kilavuzu ingilizce dişinda bir başka lisandan talep
(TR) ederse, bunu tercüme ettirmek müşteriye düşer.
• Servis kilavuzunu okuyup anlamadan ekipmanlara müdahale etmeyiniz.
• Bu uyariya uyulmamasi, elektrik, mekanik veya diğer tehlikelerden dolayi
teknisyen, operatör veya hastanin yaralanmasina yol açabilir.

ПРЕДУПРЕЖДЕН • Това упътване за работа е налично само на английски език.


• Ако доставчикът на услугата на клиента изиска друг език, задължение
ИЕ на клиента е да осигури превод.
(BG) • Не използвайте оборудването, преди да сте се консултирали и
разбрали упътването за работа.
• Неспазването на това предупреждение може да доведе до нараняване
на доставчика на услугата, оператора или пациент в резултат на токов
удар или механична или друга опасност.

警告 • 本服務手冊僅提供英文版本。
(ZH-HK) • 倘若客戶的服務供應商需要英文以外之服務手冊,客戶有責任提供翻譯服
務。
• 除非已參閱本服務手冊及明白其內容,否則切勿嘗試維修設備。
• 不遵從本警告或會令服務供應商、網絡供應商或病人受到觸電、機械性或其
他的危險。

警告 • 本維修手冊僅有英文版。
(ZH-TW) • 若客戶的維修廠商需要英文版以外的語言,應由客戶自行提供翻譯服務。
• 請勿試圖維修本設備,除非 您已查閱並瞭解本維修手冊。
• 若未留意本警告,可能導致維修廠商、操作員或病患因觸電、機械或其他危
險而受傷。

Page 10 Warning
GE COMPANY 
DIRECTION 5797904-1EN, REVISION 1 BRIVO CT385 OC UPGRADE INSTRUCTION

UPOZORENJE • Ovaj servisni priručnik dostupan je na engleskom jeziku.


(HR) • Ako davatelj usluge klijenta treba neki drugi jezik, klijent je dužan osigurati
prijevod.
• Ne pokušavajte servisirati opremu ako niste u potpunosti pročitali i razumjeli
ovaj servisni priručnik.
• Zanemarite li ovo upozorenje, može doći do ozljede davatelja usluge,
operatera ili pacijenta uslijed strujnog udara, mehaničkih ili drugih rizika.

OPOZORILO • Ta servisni priročnik je na voljo samo v angleškem jeziku.


(SL) • Če ponudnik storitve stranke potrebuje priročnik v drugem jeziku, mora
stranka zagotoviti prevod.
• Ne poskušajte servisirati opreme, če tega priročnika niste v celoti prebrali in
razumeli.
• Če tega opozorila ne upoštevate, se lahko zaradi električnega udara,
mehanskih ali drugih nevarnosti poškoduje ponudnik storitev, operater ali
bolnik.

DAMAGE IN TRANSPORTATION
All packages should be closely examined at time of delivery. If damage is apparent write “Damage
In Shipment” on ALL copies of the freight or express bill BEFORE delivery is accepted or “signed
for” by a GE representative or hospital receiving agent. Whether noted or concealed, damage
MUST be reported to the carrier immediately upon discovery, or in any event, within 14 days after
receipt, and the contents and containers held for inspection by the carrier. A transportation company
will not pay a claim for damage if an inspection is not requested within this 14 day period.
To file a report:
• Call 1-800-548-3366 and use option 6
• Fill out the GIQ workflow for any items missing, damaged, OBF/FOI for in process installs:
http://supportcentral.ge.com/ProcessMaps/
form_new_request.asp?prod_id=268679&form_id=573167&node_id=1916016&map_id=&ref
erence_id=&reference_type
• Contact the local service coordinator.

CERTIFIED ELECTRICAL CONTRACTOR STATEMENT


All electrical Installations that are preliminary to positioning of the equipment at the site prepared for
the equipment shall be performed by licensed electrical contractors. In addition, electrical feeds into
the Power Distribution Unit shall be performed by licensed electrical contractors. Other connections
between pieces of electrical equipment, calibrations and testing shall be performed by qualified GE
Medical personnel. The products involved (and the accompanying electrical installations) are highly
sophisticated, and special engineering competence is required. In performing all electrical work on
these products, GE will use its own specially trained field engineers. All of GE’s electrical work on
these products will comply with the requirements of the applicable electrical codes.
The purchaser of GE equipment shall only utilize qualified personnel (i.e., GE’s field engineers,
personnel of third-party service companies with equivalent training, or licensed electricians) to
perform electrical servicing on the equipment.

Warning Page 11
GE COMPANY 
DIRECTION 5797904-1EN, REVISION 1 BRIVO CT385 OC UPGRADE INSTRUCTION

IMPORTANT...X-RAY PROTECTION
X-ray equipment if not properly used may cause injury. Accordingly, the instructions herein
contained should be thoroughly read and understood by everyone who will use the equipment
before you attempt to place this equipment in operation. The General Electric Company, Medical
Systems Group, will be glad to assist and cooperate in placing this equipment in use.
Although this apparatus incorporates a high degree of protection against x-radiation other than the
useful beam, no practical design of equipment can provide complete protection. Nor can any
practical design compel the operator to take adequate precautions to prevent the possibility of any
persons carelessly exposing themselves or others to radiation.
It is important that anyone having anything to do with x-radiation be properly trained and fully
acquainted with the recommendations of the National Council on Radiation Protection and
Measurements as published in NCRP Reports available from NCRP Publications, 7910 Woodmont
Avenue, Room 1016, Bethesda, Maryland 20814, and of the International Commission on
Radiation Protection, and take adequate steps to protect against injury.
The equipment is sold with the understanding that the General Electric Company, Medical Systems
Group, its agents, and representatives have no responsibility for injury or damage which may result
from improper use of the equipment.
Various protective materials and devices are available. It is urged that such materials or devices be
used.

LITHIUM BATTERY CAUTIONARY STATEMENTS

CAUTION Danger of explosion if battery is incorrectly replaced. Replace only with the same or
Risk of equivalent type recommended by the manufacturer. Discard used batteries according to the
Explosion manufacturer’s instructions.

ATTENTION Il y a danger d’explosion s’il y a replacement incorrect de la batterie. Remplacer uniquement


Danger avec une batterie du même type ou d’un type recommandé par le constructeur. Mettre au
d’Explosion rébut les batteries usagées conformément aux instructions du fabricant.

OMISSIONS & ERRORS


Customers, please contact your GE Sales or Service representatives.
GE personnel, please use the GEMS CQA Process to report all omissions, errors, and defects in
this publication.

Page 12 Warning
GE COMPANY 
DIRECTION 5797904-1EN, REVISION 1 BRIVO CT385 OC UPGRADE INSTRUCTION

Revision History

Revision Date Reason for change


1 03/14/2018 Initial release.

Revision History Page 13


GE COMPANY 
DIRECTION 5797904-1EN, REVISION 1 BRIVO CT385 OC UPGRADE INSTRUCTION

This page intentionally left blank.

Page 14 Revision History


GE COMPANY 
DIRECTION 5797904-1EN, REVISION 1 BRIVO CT385 OC UPGRADE INSTRUCTION

Table of Contents
Chapter 1
Upgrade Overview ................................................................................................. 17
Section 1.0
Flowchart .......................................................................................................... 17
Section 2.0
Material List ...................................................................................................... 18

Chapter 2
Upgrade Procedure ............................................................................................... 19
Section 1.0
Preparation in Site............................................................................................ 19
1.1 Preparation ...................................................................................................................... 19
1.2 Record System INFO Information.................................................................................... 20
1.2.1 Record Power Option ......................................................................................... 20
1.2.2 Record All Remote Host Network Information .................................................... 21
1.2.3 Record Product Network Filters (PNF) Information ............................................ 22
1.2.4 Record AW IP for Direct Information .................................................................. 22
1.3 Save System State .......................................................................................................... 23
Section 2.0
NIO16 Console Upgrade .................................................................................. 24
2.1 Old Console (Z800) Removal .......................................................................................... 24
2.2 New Console (Z820) Installation...................................................................................... 25
2.2.1 Preparative ......................................................................................................... 25
2.2.2 Unpack NIO16 Console with Z820 PC ............................................................... 28
2.2.3 GPU Card Installation ......................................................................................... 29
2.2.4 NIO16 Console Cable Connections.................................................................... 30
2.2.5 Console Cover Installation.................................................................................. 34
Section 3.0
System Configuration ...................................................................................... 34
3.1 Restore System State...................................................................................................... 34
3.2 Option Installation ............................................................................................................ 35
3.2.1 Internal Landmark Option ................................................................................... 35
3.2.2 Power Option Installation.................................................................................... 35
3.2.3 Recon Enhancement Key ................................................................................... 35
3.2.4 Standard Option.................................................................................................. 35
3.2.5 Other Commercial Option ................................................................................... 35
3.3 Configure System Information on Z820 (SUSE).............................................................. 36
3.4 Flash Download .............................................................................................................. 38
3.5 GPU Card Checking ........................................................................................................ 39
Section 4.0 
Finalization........................................................................................................ 40

Table of Contents Page 15


GE COMPANY 
DIRECTION 5797904-1EN, REVISION 1 BRIVO CT385 OC UPGRADE INSTRUCTION

This page intentionally left blank.

Page 16 Table of Contents


GE COMPANY 
DIRECTION 5797904-1EN, REVISION 1 BRIVO CT385 OC UPGRADE INSTRUCTION

Chapter 1
Upgrade Overview
The Upgrade Kit is only available for Brivo CT385 IB system with NIO16 Console Z800 Host
Computer. The upgrade instruction is to guide FE how to upgrade NIO16 Console Host Computer
from Z800 (App version: 11BW49.xx) to Z820 (App version: 14BW25.xx) on CT-only Brivo CT385
system.

Section 1.0
Flowchart

Start

10mins Record System INFO

10mins Save System State

45mins OC Upgrade (Z800 to Z820)

15mins System Configuration

15mins Flash Download

60mins Finalization

Chapter 1 - Upgrade Overview Page 17


GE COMPANY 
DIRECTION 5797904-1EN, REVISION 1 BRIVO CT385 OC UPGRADE INSTRUCTION

Section 2.0
Material List
Verify that the contents of upgrade kit are correct according to the material list below.

Description Number
CT385 TP/SW Collector B77082CB
NIO Console GEHW IEC3 B76992CB
CT385 IB Upgrade Kit B75132DA
(5797952)
Upgrade Instruction 5797904-1EN
DP to DVI Cable, 3 meter 5408703
Table 1-1 Brivo CT385 IB Upgrade Package (B75122DA)

Description Number
Internal Laser Light Kit B76292CB
Note: This is a Mandatory Option for system without Internal Laser Light.
Table 1-2 CT385 Internal Laser Light (Landmark) Kit

Description Number
ASiR Non ASiR
Console Cover Kit B76562CA B70762RT
System Grade Collector B77042CB B74802CA
Table 1-3 ASiR and Non-ASiR Kit

Page 18 Chapter 1 - Upgrade Overview


GE COMPANY 
DIRECTION 5797904-1EN, REVISION 1 BRIVO CT385 OC UPGRADE INSTRUCTION

Chapter 2
Upgrade Procedure

Section 1.0
Preparation in Site
1.1 Preparation

Please confirm Internal Laser Lights, if no, you must order Internal Laser Light Kit (B76292CB) to
install before starting this upgrade.
Figure 2-1 Control Panel without Internal Laser Lights

Figure 2-2 Control Panel with Internal Laser Lights

Chapter 2 - Upgrade Procedure Page 19


GE COMPANY 
DIRECTION 5797904-1EN, REVISION 1 BRIVO CT385 OC UPGRADE INSTRUCTION

1.2 Record System INFO Information

Write down all of the system INFO information in Console Information Sheets, please refer to
5458022-2EN Service Methods CD.

1.2.1 Record Power Option


Please check system power is 40kVA or 30kVA:
• 40kVA power configuration
- Only attach system power rating plate
- Or, attach system power rating plate and 40kVA power option label
• 30kVA power configuration
Attach system power rating plate and 30kVA power option label

NOTICE If the System Power Rating Plate and the Power Option Label (40kVA or 30kVA) are attached
on the system, the priority of Option Label (Modified Label) is higher than System Power
Rating Plate.
Figure 2-3 System Power Rating Plate

GE HANGWEI MEDICAL SYSTEMS CO., LTD.

West Area of Building No.3, No.1 Yongchang North Road,


Beijing Economic and Technological Development Area,
BEIJING 100176 CHINA

MODEL: Brivo CT385


VOLTS: 200/220/240/380/400/420/440/460/480V, 3~
MOMENTARY POWER: 40KVA
CONTINUOUS POWER: 15KVA
FREQUENCY: 50/60 Hz

Figure 2-4 40kVA Power Option Label

Page 20 Chapter 2 - Upgrade Procedure


GE COMPANY 
DIRECTION 5797904-1EN, REVISION 1 BRIVO CT385 OC UPGRADE INSTRUCTION
Figure 2-5 30kVA Power Option Label

1.2.2 Record All Remote Host Network Information


1.) Select [ImageWorks] from the desktop.
2.) Select in sequence: Tools -> Network Configuration.

3.) Select host name to display remote host information.


4.) Record the network information.

Chapter 2 - Upgrade Procedure Page 21


GE COMPANY 
DIRECTION 5797904-1EN, REVISION 1 BRIVO CT385 OC UPGRADE INSTRUCTION

1.2.3 Record Product Network Filters (PNF) Information


1.) Select [Service] from the desktop.
2.) Select in sequence: Configuration -> Configure PNF.
3.) Record PNF information.

1.2.4 Record AW IP for Direct Information


1.) Select [Service] from the desktop.
2.) Select in sequence: Configuration -> Configure AW DirectConnect and record AW IP address.

Page 22 Chapter 2 - Upgrade Procedure


GE COMPANY 
DIRECTION 5797904-1EN, REVISION 1 BRIVO CT385 OC UPGRADE INSTRUCTION
3.) Select in sequence: Configuration -> Configure AW Server Client and record AW server INFO.

1.3 Save System State

Refer to the latest service methods (5458022-2EN) to understand the detailed Save State
Procedure and other information saving procedure.
Note: If a Save State was just completed, it can be omitted here.
Note: When performing Save/Restore System State, make sure there is ONLY one USB storage
device plugged in USB ports of both Console and Media Tower simultaneously, otherwise it
may result in incorrect data storage.
1.) Insert the System State DVD-RAM media into the peripheral tower drive or USB media in any
of the console’s USB ports.
2.) Select [Service Desktop] -> [Utilities].
3.) Select [System State] for DVD-RAM media or [System State - USB] for USB media. The
System State Save/Restore screen appears.
4.) Select [All] to select all the cals, characterization, etc.
5.) Select [Save].
6.) If DVD-RAM or USB is ready, select [Yes].
7.) When completed select [Dismiss].
8.) Remove System State DVD-RAM media from the peripheral tower or USB media from USB
port.

Chapter 2 - Upgrade Procedure Page 23


GE COMPANY 
DIRECTION 5797904-1EN, REVISION 1 BRIVO CT385 OC UPGRADE INSTRUCTION

Section 2.0
NIO16 Console Upgrade
2.1 Old Console (Z800) Removal

NOTICE Make Sure to follow all Lockout/Tagout requirements while performing this procedure.
1.) Move the cradle all the way out of the gantry and lower to its lowest point.
2.) Shut down the system and power off the Operator Console power switch.
3.) Switch OFF the breaker in the Power Distribution Box (hospital supply).
Note: If there is UPS, switch OFF the UPS output.
4.) Visually verify proper labeling of each cable and disconnect all console rear cables.
5.) Disconnect the cables between the old console and system (Gantry, PDU and local hospital
network).
6.) Remove the old OC aside.

Page 24 Chapter 2 - Upgrade Procedure


GE COMPANY 
DIRECTION 5797904-1EN, REVISION 1 BRIVO CT385 OC UPGRADE INSTRUCTION

2.2 New Console (Z820) Installation

2.2.1 Preparative
Figure 2-6 System Interconnect Diagram

Chapter 2 - Upgrade Procedure Page 25


GE COMPANY 
DIRECTION 5797904-1EN, REVISION 1 BRIVO CT385 OC UPGRADE INSTRUCTION

Table 2-1 System Interconnect Cables for Console

No. Description Long Cable Short Cable


53 LVAC Power Cable - PDU to Operator 2343531 2343531-2
Console
56 Ground, Raceway to Console 5441518 5441518-2
101 Signal Cable - Gantry to Console 5419981 5419981-2
102 Signal Cable (Ethernet) - Gantry to Console 5432020
103 Data Cable (Fiber Optic) - Gantry to Console 5432019

Page 26 Chapter 2 - Upgrade Procedure


GE COMPANY

NIO16 Console Table Top


DIRECTION 5797904-1EN, REVISION 1

GSCB Gantry

Blue
Green
Gantry Interface

HRST

IN
DB9 DB25

OUT
HOST
COM
AUDIO IN TGP
7
AUDIO OUT

USB2.0
Board
1
KeyBoard
8

USB2.0

USB3.0
2
Mouse

USB2.0
3
Service Key 9 TrackBall

USB2.0

USB3.0
BarCode DCB
4

USB2.0
Reader Board
5
EXT HDD

USB3.0
Figure 2-7 NIO16 OC Interconnect (with Z820)

SWITCH 6
DVD RW

USB3.0
HUB
8XRJ45 RJ45

1 2
ETH3
SRV
DVI-D DVI-D

Chapter 2 - Upgrade Procedure


3 4
RJ45
Monitor1 Monitor2
RPM
ETH2
5 6
AW
7 8
AUX/PTR

RX TX RX TX DB9
INJ
DIP
ECG
UPS DP-0 DP-1 DP-2 DP-3
Video
1 2 3 4
Card DVI-D

DVI-D Video Splitter Fiber DVI-D


Room
Optic SC Optic SC DVI-D
Monitor
RJ45 RJ45
Ethernet Video Receiver
ETH1* ETH0*
Card
HSP
Option

OPERATING ROOM SCAN ROOM

Rev History: MADE BY: WU WENDONG DESIGN TITLE


REV 2 DESCRIPTION: Update the service key installation DATE: 2014-OCT. 30 FIRST MADE FOR
NIO CONSOLE ASSEMBLY WITHOUT COVER
GE Healthcare
ELECTRONIC RELEASE PROCESS g China REV SIZE
MADE BY Ma Abin DATE Dec. 1 2014 APP BY DATE DESIGN RELEASED
SHEET 1
5411200-9SCH 2 D
BRIVO CT385 OC UPGRADE INSTRUCTION

Page 27

GE COMPANY 
DIRECTION 5797904-1EN, REVISION 1 BRIVO CT385 OC UPGRADE INSTRUCTION

2.2.2 Unpack NIO16 Console with Z820 PC


Please read details about NIO16 installation procedure in Service Manuals.
1.) Unpack console packaging.
a.) Remove all items from the console.
b.) Remove all packaging and discard per local requirement.
c.) Place the step-board under the front edge of the skid and step on it to raise the front edge
of the skid as in
Figure 2-8 Step-Board Used to Raise Front Edge of Skid

d.) Remove the two front cushions from the bottom of the skid.
Figure 2-9 Cushion on Botton of Skid

e.) Lift up on the strap on the front of the step-board (Figure 2-8) to lower the skid. Remove
the step-board.
f.) Ensure the console stabilizers are in line with thee notched portion at the front of the skid.
This will allow enough clearance to smoothly roll the console down the ramps.

Page 28 Chapter 2 - Upgrade Procedure


GE COMPANY 
DIRECTION 5797904-1EN, REVISION 1 BRIVO CT385 OC UPGRADE INSTRUCTION
2.) Move and place the console to the operator room.
3.) Adjust two rubber footer to prevent console from sliding.
Figure 2-10 Rubber Footer Adjustment

4.) Remove the Host PC support bracket from left side of Z820.
Figure 2-11 Host PC Support Bracket

2.2.3 GPU Card Installation


• If customer does not purchase ASiR option, please ignore this section.
• If customer has already purchase ASiR option, please refer to Recon Enhancement Package
Installation Manual (5456615-1EN_r4 or later) to install GPU card (6450000-109, shipped with
B77042CB) in Z820 Host Computer.

Chapter 2 - Upgrade Procedure Page 29


GE COMPANY 
DIRECTION 5797904-1EN, REVISION 1 BRIVO CT385 OC UPGRADE INSTRUCTION

2.2.4 NIO16 Console Cable Connections


Figure 2-12 Z820 Host Computer Connections

Page 30 Chapter 2 - Upgrade Procedure


GE COMPANY 
DIRECTION 5797904-1EN, REVISION 1 BRIVO CT385 OC UPGRADE INSTRUCTION
Figure 2-13 NIO16 Console AC Box Assignments

120VAC

GND

Chapter 2 - Upgrade Procedure Page 31


GE COMPANY 
DIRECTION 5797904-1EN, REVISION 1 BRIVO CT385 OC UPGRADE INSTRUCTION
1.) Connect the Keyboard and Mouse cables to NIO16 Console Z820 Host Computer.
2.) Connect GSCB cable connectors to NIO16 Console AC Box and Z820 Host Computer
according to Figure 2-14 and Table 2-2.
Figure 2-14 GSCB and Cable

Table 2-2 GSCB Cable Connections

# ITEM Description
1 DB9/F Black Host Computer DIP Serial Port
2 DB9/F Gray Host Computer RS232
3 Green Audio Host Computer Audio Out (Green)
4 Blue Audio Host Computer Audio In (Blue)
5 USB Not Used
6 DB9/M Black AC Box J56
7 DB25/M Black TGP Gantry Cable

3.) Connect Media Tower cable connectors to NIO16 Console AC Box and Z820 Host Computer.
4.) Connect the LCD Monitors for Z820 Host Computer.
a.) Remove the old Video cable (5332107-2) from the LCD Monitors.
b.) Connect the new Video cable (5408703, shipped with upgrade kit) between Z820 Host
Computer and LCD Monitor.
Page 32 Chapter 2 - Upgrade Procedure
GE COMPANY 
DIRECTION 5797904-1EN, REVISION 1 BRIVO CT385 OC UPGRADE INSTRUCTION
Figure 2-15 Video Cable and Power Cable for LCD Monitor

c.) Connect the LCD Monitor power cables to NIO16 Console AC Box.

- Scan Monitor
Video Cable: from Console Host DP1 to Monitor DVI
Power Cable: from Console AC Box J10 to Monitor
- Image Monitor
Video Cable: from Console Host DP0 to Monitor DVI
Power Cable: from Console AC Box J9 to Monitor

d.) Route LCD video cable and power cable.


Figure 2-16 Cable Routing and Keeper

5.) Plug cables into Switch Hub on console if request.


6.) Connect other cables to Z820 Host Computer according to Figure 2-12.

Chapter 2 - Upgrade Procedure Page 33


GE COMPANY 
DIRECTION 5797904-1EN, REVISION 1 BRIVO CT385 OC UPGRADE INSTRUCTION
Figure 2-17 Switch Hub Connections

7.) Connect the NIO16 Console power cable and Ground cable to the Console Power Panel.

2.2.5 Console Cover Installation


Refer to Console Cover Installation procedure in 5458022-2EN Service Methods CD.
Note: If customer purchases ASiR option, the Console Covers are black or else they are white.

Section 3.0
System Configuration
3.1 Restore System State

1.) Power on system, and confirm that system started up normally.


2.) Insert a previously saved system state DVD-RAM media into the peripheral tower drive or USB
media in any of the console’s USB ports.
3.) Wait until the media is ready.
4.) Select [Service] -> [Utilities].
5.) Select [System State] for DVD-RAM media or [System State - USB] for USB media. The
System State Save/Restore screen appears.
6.) Select [All].
7.) Select [Restore]. The Restore System State box appears.
8.) Select [Yes]. A message at the end of the System State Log window should display: Restore
System State Completed Successfully.
9.) When completed, select [Cancel].
Note: If Does Check option is installed, Dose Audit tool starts up when Reconfig Info or Protocol is
selected. That shows configuration and Protocol summary. Press [Quit] to continue.
10.) Select [Yes] when the Scan Hardware Reset pop-up appears.
11.) When completed, select [Dismiss].
12.) Remove System State media and place in a safe location.
13.) Reboot system, then verify Restore System State.

Page 34 Chapter 2 - Upgrade Procedure


GE COMPANY 
DIRECTION 5797904-1EN, REVISION 1 BRIVO CT385 OC UPGRADE INSTRUCTION

3.2 Option Installation

3.2.1 Internal Landmark Option


Please install Internal Landmark Option Key (shipped with upgrade kit).

3.2.2 Power Option Installation


• If system power is 30kVA checked in 1.2.1 Record Power Option (40kVA system had been
restricted to 30kVA), DON’T install 40kVA power option per Product Key (shipped with
upgrade kit).
• If system power is 40kVA checked in 1.2.1 Record Power Option, please install 40kVA Power
Option per Product Key (shipped with upgrade kit).

CAUTION If system is 30kVA power, please DON’T install 40kVA power option to avoid power overload.

3.2.3 Recon Enhancement Key


If the customer has purchased ASiR Option, additional, please install Recon Enhancement Key
(shipped with upgrade kit) using Option Install menu.

3.2.4 Standard Option


Please install standard options per Product Key (shipped with upgrade kit).

3.2.5 Other Commercial Option


Please apply new e-license through e-license website (http://elicense.gehealthcare.com/elicese/). 
Refer to option list recorded in 1.2 Record System INFO Information Record System INFO
Information.

Chapter 2 - Upgrade Procedure Page 35


GE COMPANY 
DIRECTION 5797904-1EN, REVISION 1 BRIVO CT385 OC UPGRADE INSTRUCTION

3.3 Configure System Information on Z820 (SUSE)

1.) Configure all remote network information.


a.) Select [ImageWorks] from the desktop.
b.) Select in sequence: Tools -> Network Configuration.

c.) Select [Add].

Page 36 Chapter 2 - Upgrade Procedure


GE COMPANY 
DIRECTION 5797904-1EN, REVISION 1 BRIVO CT385 OC UPGRADE INSTRUCTION
d.) Input the Host Name then press [Enter].

2.) Configure PNF information, please refer to Section 1.2.3.


3.) Configure AW IP for direct information, please refer to Section 1.2.4.

Chapter 2 - Upgrade Procedure Page 37


GE COMPANY 
DIRECTION 5797904-1EN, REVISION 1 BRIVO CT385 OC UPGRADE INSTRUCTION

3.4 Flash Download

Note: The Flash Download takes anywhere from 5-30 minutes, depending on which subsystems require
updates.
1.) Perform the Flash Download Utility found on the Common Service Desktop - Utilities Tab,
select [Flash Download].
Figure 18 Common Service Desktop - Utilities Tab, Flash Download

2.) When the Flash Download Window opens, Select [Update].

Page 38 Chapter 2 - Upgrade Procedure


GE COMPANY 
DIRECTION 5797904-1EN, REVISION 1 BRIVO CT385 OC UPGRADE INSTRUCTION
Figure 19 Flash Download Window5

3.) Once the Gantry Hardware Flash Downloads successfully, select [Dismiss].
4.) Close the Common Service Desktop.

3.5 GPU Card Checking

Open a shell, then type: cat /proc/ati/1/biosversion. Check the result as below:

Figure 2-20 GPU Check Result

Chapter 2 - Upgrade Procedure Page 39


GE COMPANY 
DIRECTION 5797904-1EN, REVISION 1 BRIVO CT385 OC UPGRADE INSTRUCTION

Section 4.0
Finalization
1.) Perform Quality Assurance Test of Functional Checks on Service Methods (5457941-2EN),
make sure all of them pass.
2.) Save System State
a.) Insert the Save System State DVD into the Peripheral Tower DVD-RW drive.
b.) Select [SERVICE DESKTOP].
c.) If reloading software, select [UTILITIES].
d.) Select [SYSTEM STATE].
e.) Select [ALL] to select all the cals, characterization, etc.
f.) Select [SAVE].
g.) If applicable, select [OK] when the following message appears: System State Media
Status: Please insert a DVD into the drive and press Save again
h.) When completed select [DISMISS].
i.) Close the Service Desktop window in the upper left corner of the screen.
3.) Old Console Disposal
The old console has components that are needed to cover FRU shortages, so please
RETURN the old console. Follow the normal local return process to return the console.
a.) Package the old console with the shipping material.
b.) Mark type of the old console on the outside of the box.
c.) Return the old console:
- For U.S. sites
Call Schenker Heavyweight Return at 866-229-7877 (Option 25), and ship to:
GE Healthcare Renewable Resources
7624 South 10th Street
Oak Creek, WI. 53154
- For Western Europe sites
Stick the harvest sticker (see below) on the OC box & ship (Heavyweight Return) to:
P.Vinet - Volta Repair Center Manager
Gems, 283 rue de la Minière 78533 Buc
VIA St WITZ

- For other Europe sites


Please contact your local service representative for approbation.