Aquilion Multi - Intermittent ring artifact


Phenomenon
Occasionally a sharp band ring shape artifact must have been observed.

Observations and interventions


Investigation with DCA, using raw data from the raw data disk, showed that the problem was
located on or near the channels processed by MDDC nr 9.

Swopping the MDDC made no difference.


A test of the MCONT board at another site proved that the MCONT board was OK.

The problem is suspected to originate in the MDDC board connectors at the backplane board in
the detector assembly. Replacement of the detector assembly solved the problem.

Solution
Replacement of the detector.