Professional Documents
Culture Documents
Archiving a lot Before you add a new lot number, archive the current lot number for future use.
number When you archive a lot number, the accompanying sample types with their
respective reference ranges are archived too.
To archive a lot number:
1. From the TAS Main screen menu, choose QC > Lot number to display the
QC select Lot number screen.
2. Select the lot number that you want to archive.
3. Click Archive.
Restoring a lot You can restore an old lot number to use that lot number’s reference ranges when
number viewing samples.
Before restoring an old lot number, you must archive the current lot number. The
Restore Lot number screen allows you to do this at the same time as restoring a
previous lot number.
Note: When you have finished viewing the old lot number’s reference ranges, you
must restore the current lot number.
REFERENCE RANGES
Haemonetics has established manufacturer’s reference ranges that outline the
normal maximum and minimum limits for R, K, Angle, and MA for the commonly
used TEG sample types and quality controls. These values are encoded in the
TEG Analytical Software (TAS) and are used by the software to determine if the
test results come inside or outside the ranges.
Note: Since the manufacturer’s reference ranges are periodically updated, always
refer to the product inserts for the current manufacturer’s reference range values.
Local reference The TEG Hemostasis System reference ranges for TEG sample types are derived
ranges from running samples on a heterogeneous population of healthy donors and
statistically analyzing the results to determine the values. Since patient
populations may differ due to geography, age, diet, etc., sites should verify the
manufacturer’s ranges or establish their own reference ranges to enter into the
software.
To verify the manufacturer’s reference ranges or to establish local reference
ranges, follow the procedures in the TEG 5000 System Validation Guide.
Updating It may be necessary to update the reference ranges in the software when a new
reference ranges lot becomes available for a quality control or when studies are updated.
in TAS If your institution establishes its own ranges, refer to “Establishing reference
ranges for quality controls” on page 11-17 for information on recalculating
ranges for a new master lot. Then, if necessary, change the ranges in the software
to match the calculated ranges.
If your institution uses the manufacturer’s reference ranges, compare the
reference ranges in the product insert to the current values in the software. If
necessary, change the reference ranges in the software to match the inserts.
Caution: New reference ranges added to TAS apply to previously run samples in
the database.
Note: Before updating reference ranges in TAS for a quality control sample,
archive the old values so that a history of the reference ranges is maintained for
the quality control samples (refer to “Archiving a lot number” on page 11-13).
2. In the User Profile Setup screen, click the Normal values tab.
Field Description
5. Click Done.
Reference ranges For sites that choose to run native whole blood samples (whole blood without the
for native addition of any modifiers), the following reference ranges have been established,
samples based on a group sample size of 132 donors:
Native (N) 12.1 – 26.5 3.2 – 12.8 13.6 – 46.4 41.8 – 63.0 3.2 – 7.1
Citrated 9.4 – 27.4 1.9 – 8.9 22.0 – 58.0 44.4 – 63.6 3.6 – 8.5
Native (CN)
Note: Sites wanting to run this type of sample should verify reference ranges or
establish their own reference ranges following the procedure in the TEG 5000
System Validation Guide.
Establishing If your institution chooses to establish its own ranges to monitor quality control
reference ranges (refer to “Establishing quality control ranges” on page 11-8), you can use the
for quality following procedure to obtain the Mean and Standard Deviation (SD) for the
quality controls.
controls
Establishing a new Mean and SD
To use the TEG Analytical Software (TAS) to establish the new Mean and SD for
each level of quality control:
1. Run both levels of quality control a minimum of 20 times (refer to
“Running quality control samples” on page 7-9).
In TAS, in the Patient name field, use the naming convention, “Master Lot
xxxx-xxxx Reference Range Study.” The master lot number can be found
on the control vials.
2. From the TAS Main screen menu, choose QC > Go To to switch to the QC
database.
3. Click .
4. On the Filter criteria screen, do the following:
a. On the Sample type tab, highlight “L1” or “L2.”
b. On the Patient tab, in the Patient name list, highlight “Master Lot xxxx-
xxxx Reference Range Study.”
c. Click Apply filter.
Next steps
After establishing the ranges for a new master lot of quality control, do the
following:
⚫ Archive all previous lots of control. Refer to “Archiving a lot number” on
page 11-13.
Caution: Do not change the normal ranges in TAS until the reference range
study is complete and the new master lot is in use.
About eConsult The eConsult feature allows you to consult electronically with colleagues and
specialists by securely sending patient data through e-mail.
The eConsult feature can send files and messages to e-mail recipients using
various devices, including standard PCs, notebooks, and personal digital
assistants (PDAs) as long as they are equipped to receive e-mail with graphical
attachments.
Depending on how you set up your eConsult profile, recipients can receive a
short text message on their cell phone or pager informing them that they have e-
mail for an electronic consultation.
eConsult In order to use the eConsult feature, the following prerequisites must be met:
prerequisites ⚫ The eConsult option must already be configured through the User profile
setup screen.
Typically, your Site Administrator will have set this up for you. However,
if you need to add a recipient to eConsult or to define a server to send
and receive mail, refer to “eConsult Tab” on page 6-25.
⚫ The computer being used must be on the network with access to the mail
servers.
Sending patient The eConsult feature allows you to send patient data to a recipient’s PC, laptop,
data using cell phone, or PDA. The eConsult wizard takes you through the process by
eConsult providing screens to:
⚫ Identify the patient for whom you are sending information while
maintaining patient confidentiality.
⚫ Select sample(s) to send, enter relevant clinical data and contact
information.
⚫ Specify the type of attachment to send and the recipient.
The steps
To send patient data through eConsult:
1. From the TAS Main screen, click . The Introduction screen appears.
Note: For privacy purposes, the patient’s name is suppressed in the e-mail
and the patient is identified only by the code that you enter.
4. Click Next.
The Select samples screen appears, listing the samples for this patient:
5. Click to select one sample or hold the Ctrl key down while you select
multiple samples.
6. Click Next.
The eConsult wizard (2) screen appears allowing you to enter clinical data
and contact information.
Figure 12-3, eConsult wizard (2)- Clinical and contact form screen
7. Complete the desired fields:
Field Description
Field Description
Lab results Type the available lab results for the
following in their respective fields:
Platelet number
Fibrinogen level
PT
INR
PTT
DDIMER
8. Click Next.
The eConsult wizard (3) screen appears allowing you to select the format
for sending the data.
The Steps
To copy and paste a tracing into another document:
1. From the TAS Main screen, do one of the following:
Figure 12-7, Tracing pasted into a document using the Capture function
Troubleshooting
TROUBLESHOOTING ........................................................... A-2
Login errors . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . A-3
Database errors . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . A-7
eTest errors . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . A-9
Cup and pin errors . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . A-11
Quality control (Level I and Level II) sample errors . . . . . . . . . . . . . . . . A-12
Unexpected tracing results . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . A-16
Blood sample errors . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . A-19
Temperature errors . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . A-20
Reference ranges errors . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . A-21
Remote workstation access errors . . . . . . . . . . . . . . . . . . . . . . . . . . . . . A-22
Printing errors. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . A-23
Accessing Help errors . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . A-26
TROUBLESHOOTING
This chapter contains troubleshooting tips for some of the most common issues
and concerns experienced when using the TEG® analyzer. This chapter helps you
solve problems associated with:
⚫ Logging in to the TEG Analytical Software (TAS)
⚫ Opening databases
⚫ Performing an eTest
⚫ Loading and ejecting cups and pins
⚫ Running quality control samples
⚫ Unexpected tracing results
⚫ Running blood samples
⚫ Temperature readouts
⚫ Viewing reference ranges
⚫ Accessing remote workstations
⚫ Printing
If a problem still exists after following the recommended actions outlined in the
tables in this chapter, contact TEG System Technical Support to arrange for
repairs or return (refer to “Customer Service” on page 1-4).
Login errors
Database errors
eTest errors
K is out of range. Calcium and or/activator were Prepare another sample and rerun
dispensed incorrectly or are the test.
contaminated.
The sample temperature is Check the temperature (refer to
incorrect. “Temperature errors” on page A-
20).
Angle is out of range. The analyzer is out of Contact TEG System Technical
specification. Support.
There was an electrical disruption Rerun the test.
to the analyzer.
The sample temperature is Check the temperature (refer to
incorrect. “Temperature errors” on page A-
20).
R in one channel is very short and The analyzer is not level. Use the bubble level at the top of
the MA is very large, while the MA the analyzer and the leveling feet
in the other channel is very small. to ensure that the analyzer is
level.
Parameter values are out of range For R and K, TAS may be reporting Convert the mm to minutes,
according to the insert, but not the units in millimeters (mm) and remembering that 2mm=1
according to the software. the insert reports the units in minute.
minutes or seconds. Convert the mm to seconds if this
is the protocol at your site.
You may have used a new lot that Enter the current reference ranges
has different reference ranges. from the product insert into TAS.
Unexpected tracing for Level I. Level I control was placed in the Do the following:
sample cup but “L2” was selected
as the sample type in the software. 1. Confirm that the wrong
sample type was selected in
the software.
2. Select the correct sample type.
Sharp drop in MA. The lever was switched to Load If needed parameters were not
before terminating the sample. defined, rerun the test.
No R value. A preclot. There may have been a Prepare another sample and
clot in the sample or there may rerun the test.
have been a delay in starting the
sample.
Temperature errors
Printing errors
When trying to use Quick Print, the No tracing is selected. Highlight the tracing you want to
Print button is disabled. print and click the Quick Print
button.
You attempted to print from the Press the F6 key to send the
Multi screen which has the Print tracing to the default printer.
button disabled.
When using Quick Print, the This is according to design. Press the F6 key to send the
printer must be selected, even tracing to the default printer.
though the default printer is
selected already.
All pages of the report print, You did not specify a particular When you choose the printer,
instead of the current page. subset or page. indicate the page(s) that you want
to print.
SPECIFICATIONS
This section describes the physical, environmental, and electrical specifications
for storing and operating the TEG® analyzer.
Note: For additional information about sensitivity and interference factors, refer
to “Limitations” on page B-4.
Environmental The environmental specifications for the operation and storage of the TEG
Specifications analyzer are as follows:
Conditions Value
Use Indoor use
Operating position Vibration-free; no solar radiation
Operating temperature 50 °F to 95 °F (10 °C to 35 °C)
Operating humidity Relative humidity 20-80% (non-condensing)
Conditions Value
Transport and storage –22 °F to 122 °F (–30 °C to 50 °C)
temperature
Transport and storage Relative humidity 5-95% (non-condensing)
humidity
Pollution degree Pollution degree 2
Electrical The electrical specifications for operating the TEG analyzer are as follows:
Specifications
Characteristics Value
Power Power supply in protection class I, transformer
with thermal cutoff and monitored safety
insulation.
Power supply must be plugged into a properly
grounded outlet.
Power the device using only the power
adapter supplied by Haemonetics for the
TEG analyzer.
Use an Uninterruptible Power Supply (UPS) unit
between the analyzer and the power source.
Overvoltage Overvoltage Category II
Mains supply fluctuations Not to exceed ±10% of the nominal voltage
220V model Operating voltage 230V, 50 Hz,
rated current 0.20 A,
max input power 46W.
Thermal cutout and 1 A 2AG type Slo-Blo fuse
provided.
120V model Operating voltage 120V, 60 Hz,
rated current 0.38 A,
max input power 46W.
Thermal cutout and 1 A PICO II, Very Fast-Acting
fuse provided.
TEG 5000 Input power rating (from power supply) 24 VAC
30 W
Note: For more information about electrical precautions, refer to “Power outlet
connection” on page 3-3.
LIMITATIONS
Caution: The TEG analyzer is for in vitro diagnostic use only.
Users of the TEG analyzer should be properly trained and should be appropriate
medical or other health care professionals.
The TEG analyzer output is for presentation purposes only, and may vary from
time to time; users should use the actual data derived from their own samples and
their own established reference ranges to quantify the output tracing as to the
degree of abnormality.
Results from the TEG analyzer should not be the sole basis for a patient diagnosis
but should be considered along with the patient’s clinical condition and other
laboratory tests.
Sensitivity factors Sensitivity factors that may affect the measurement capability and operation of
the TEG analyzer are listed below.
⚫ The maximum oscillation of the cup in the TEG instrument is
approximately 5 degrees, as described in the Interference section below.
Therefore, the maximum amplitude (MA parameter) cannot be measured
beyond 96 mm. It is very rare for human blood to exceed this limit; if it
should happen, the diagnosis is obvious-extreme hypercoagulability.
⚫ The eTest value of the TEG instrument determines the zero starting point of
the graphical output tracing. Therefore, out of range conditions may
prevent the TEG graph from reaching its maximum amplitude (the MA
parameter may not reach its maximum value). The software issues a
warning if the eTest value is out of range when a sample is started.
⚫ Non-anticoagulated whole blood samples that are placed on the analyzer
later than six minutes after drawing may result in a clotted sample, leading
to erroneous results.
⚫ Testing sensitivity of the TEG analyzer is affected if the environmental and
electrical specifications referenced in Table B-2 and Table B-3 are not met.
Interference Interference factors that may affect the performance or sensitivity of the TEG
factors analyzer are listed below along with information on ways to mitigate the
interference:
⚫ The moving part of the TEG analyzer is a cylindrical cup that oscillates
very slowly through an angle of approximately 5 degrees. This means the
following:
⚫ The TEG analyzer must be level. A leveling bubble and leveling feet are
built into the instrument.
PERFORMANCE CHARACTERISTICS
Through performance testing, the accuracy and precision of the TEG 5000 Series
analyzer is supported. In addition, Haemonetics has identified reference ranges
as well as factors that may affect the measurement capability, performance, and
sensitivity of the analyzer.
A summary of the accuracy, precision, sensitivity, and reference range
characteristics for the TEG 5000 analyzer are discussed below.
Parameter Description
The results are tabulated below, as the means, standard deviations, differences
between the means, standard deviations of the differences, t-values, and the
2- tail probability.
Parameter
DF 78 78 78 78
2-TAIL P-VALUE 0.497 .168 0.341 0.388
Based on the TEG measurement data, the precision of each TEG parameter
expressed as a percentage using the coefficient of variation (CV) is as follows for
each activator:
Parameter
Activator
R K Angle MA
Celite 8.7% 7.6% 2.4% 4.1%
Sensitivity and Refer to “Limitations” on page B-4 for the following information on sensitivity
specificity and specificity:
⚫ Sensitivity factors that may affect the measurement capability and
operation of the TEG analyzer.
⚫ Interference factors that may affect the performance or sensitivity of the
TEG analyzer and ways to mitigate the interference.
Reference ranges Haemonetics provides reference ranges in the TEG system software which were
derived from data provided by a number of hospitals in the United States.
Reference ranges for native whole blood samples are listed in Chapter 11. Refer
to the appropriate product inserts for the reference ranges for other sample types.
Glossary
GLOSSARY ....................................................................... C-2
GLOSSARY
A TEG Activator sample. Constitutes one of the PlateletMapping® assay tests and
is used to produce a clot without platelet contribution. The value of the
activator sample is subtracted from the other samples to determine the platelet
contribution to clot strength.
A/D box A hardware component in the TEG system that converts the analog signal
produced by the analyzer into a digital signal that is read by the TEG Analytical
Software.
aPTT Activated Partial Thromboplastin Time. A test that measures clotting time in
plasma (the liquid portion of blood). It focuses on the intrinsic pathway in the
blood clotting process and uses a contact activator (kaolin, celite, silica, or
elagic acid) to speed up the test time.
activator A substance that causes a reaction or speeds up a process. Also a TEG sample
type (refer to “A” above).
aggregation The formation of a mass of cells due to binding of specific cell receptors. In
platelets, the receptors associated with aggregation are the GPIIb/IIIa receptors.
agonist A drug or other chemical that can combine with a receptor on a cell to
produce a physiologic reaction typical of a naturally occurring substance.
antifibrinolytic A drug or endogenous protein that decreases the breakdown of fibrin (e.g.,
amicar, transexamic acid, plasminogen activator inhibitor-1 (PAI-1)).
arachidonic acid A platelet agonist. Its action stimulates a pathway that results in activation of
the thromboxane (TxA2) receptors on a platelet.
aspirin An irreversible platelet inhibitor. It inhibits the action of the COX-1 enzyme in
the arachidonic pathway leading to inhibition of thromboxane synthesis. TEG
action: this drug will not alter the MA of a thrombin-activated TEG analysis.
The AA PlateletMapping assay is required to observe the effects of this drug.
Most significant TEG effect: decreases MAAA.
cardiopulmonary bypass Diversion of the flow of blood to the heart directly to the aorta, via a pump
oxygenator, avoiding both the heart and the lungs. A form of extracorporeal
circulation used in heart surgery.
citrate A substance used to anticoagulate blood by chelating calcium. Its action can
be overridden by adding calcium. Used as a preservative in blood samples.
clopidogrel (Plavix®) A potent oral antiplatelet agent often used in the treatment of coronary artery
disease, peripheral vascular disease, and cerebrovascular disease. Also known
as Plavix®. TEG action: this drug will not alter the MA of a thrombin-activated
TEG analysis. The ADP PlateletMapping assay is required to observe the effects
of this drug. Most significant TEG effect: decreases MAADP.
coagulation factor A substance in the blood that is essential for blood clot formation.
coagulopathy A disorder in which blood is either too slow or too quick to coagulate (clot).
cryoprecipitate A concentrated product made from plasma that contains Factors VIII, XIII,
fibrinogen, vWF, and fibronectin. TEG effect: decreases R.
D-dimer A laboratory test for measuring cross linked fibrin breakdown. The results help
rule out, diagnose, and monitor diseases and conditions that cause
hypercoagulability. High D-dimer results can suggest that there has been
significant clot (thrombus) formation and breakdown in the body, but it does
not tell the location or cause. An elevated D-dimer may be due to a venous
thrombotic event (VTE) or disseminated intravascular coagulation (DIC), but it
may also be due to a recent surgery, trauma, or infection. Elevated levels are
also seen with liver disease, pregnancy, eclampsia, heart disease, and some
cancers.
dipyridamole (Persantine®) An antiplatelet drug. Also known as Persantine®. TEG action: requires ADP
PlateletMapping assay to demonstrate effects. Most significant TEG effect:
decreases MAADP.
endothelium The single cell layer that lines all blood vessels. The function of the
endothelium includes acting as a barrier between the blood and extravascular
space, hemostasis regulation, and hemodynamic regulation.
eTest An electronic test that is part of the daily maintenance for TEG analyzers
connected via the serial port Analog/Digital interface box.
enzymatic For coagulation, this relates to the coagulation factor reactions and their
related substances.
fibrin An elastic protein derived from fibrinogen by the action of thrombin and forms
an insoluble network that provides a structure for the blood clot.
fibrinogen A protein in the blood plasma that is essential for the coagulation of blood and
is converted to fibrin by thrombin and ionized calcium. Also called Factor I.
fibrinolysis The breakdown of fibrin in a blood clot, usually by the enzymatic action of
plasmin.
fibrinolytic An endogenous protein or drug that breaks down fibrin, either directly or
through the generation of plasmin (i.e. tissue plasminogen activator (tPA),
urokinase).
FFP Fresh frozen plasma. Plasma that is removed from a donor, separated, and
frozen within 6-8 hours of collection.
hemorrhage An escape of blood from the blood vessels, especially when excessive. Also
called hemorrhea.
ischemia (ischemic) A decrease in the blood supply to an organ or tissue resulting in a reduced
oxygen supply.
lysis The dissolution or destruction of cells, such as blood cells or bacteria, as by the
action of a specific lysin (plural: lyses).
PT Prothrombin time. A test that measures the clotting time of plasma (the liquid
portion of the blood). It focuses on the extrinsic pathway in the blood clotting
process and uses tissue factor as an activator.
PTT Partial Thromboplastin Time. A test that measures clotting time in plasma (the
liquid portion of blood). It focuses on the intrinsic pathway in the blood
clotting process. In the 1970s, activators were added to activate contact factors
to speed up the test time and the test became aPTT (see “aPTT”).
plasma The clear, yellowish fluid portion of blood, lymph, or intramuscular fluid in
which cells are suspended.
plasmin A proteolytic enzyme (i.e. breaks down proteins) that hydrolyzes peptides and
esters of arginine and histidine and converts fibrin to soluble products.
plasminogen The inactive precursor to plasmin that is found in body fluids and blood
plasma.
prasugrel A platelet inhibitor developed for patients with acute coronary syndromes who
undergo percutaneous coronary intervention. Also known as Effient ®. TEG
action: this drug will not alter the MA of a thrombin-activated TEG analysis.
The ADP-PlateletMapping assay is required to observe the effects of this drug.
Most significant TEG effect: decreases MAADP.
sepsis The inflammatory condition resulting from the presence of pathogens or their
toxins in blood.
TEG® Thrombelastograph®.
Thrombelastograph® A device used to measure the change in elastic properties during clot
formation.
thrombin The central clotting factor in the hemostasis process that facilitates the
conversion of fibrinogen to fibrin, activates other clotting factors, and is the
most potent activator of platelets. Thrombin generation is the pivotal point in
the hemostasis process. This enzyme is formed in the blood from prothrombin.
thrombus A blood clot. The final product of blood coagulation from the hemostasis
process.
tracing For TEG testing, the graph that is produced that plots clot strength versus time.
von Willebrand Factor A large multimeric glycoprotein present in blood plasma and produced
(vWF) constitutively in endothelium, megakaryocytes (alpha-granules of platelets),
and subendothelial connective tissue. A protein required for platelet adhesion.
Initial Setup
INITIAL SETUP OVERVIEW ..................................................... D-2
UNPACKING THE ANALYZER . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .D-3
SETUP AND INSTALLATION. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .D-4
Positioning the analyzer . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .D-4
Connecting to the interface box and computer . . . . . . . . . . . . . . . . . . . .D-4
Connecting to power . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .D-5
Disconnecting from power . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .D-5
Connecting multiple analyzers . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .D-5
Installing the software . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .D-6
Conducting maintenance checks . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .D-6
ACTIVATING THE CHANNELS FOR A NEW ANALYZER . . . . . . . . . . . . . .D-7
Note: Save the shipping box and molded polystyrene support forms. If the TEG
analyzer needs to be returned for repair or preventive maintenance, it must be
shipped in its original packaging in order to avoid damage. Haemonetics will
charge for any repairs necessary due to improper packaging.
Positioning the The analyzer should be set up in a location that meets the following
analyzer requirements:
⚫ Minimum of one cubic foot of clearance.
⚫ Sturdy and level surface.
⚫ Free of vibration and away from high traffic.
⚫ Room temperature less than 30 °C.
⚫ Out of the direct path of HVAC ducts or vents.
Note: For more information on operation and storage requirements for the
TEG analyzer, refer to “Storage and Handling” on page 3-2 and
“Specifications” on page B-2.
Connecting to the To connect the TEG analyzer to the analog/digital interface box and computer:
interface box and 1. Insert one end of the female-female DB-9 cable into the 9-pin port labeled
computer “I/O” on the back panel of the analyzer.
2. Insert the other end of the female DB-9 cable into the port labeled “TEG 1”
on the interface box.
Note: There must always be one analyzer connected to the TEG1 port and
powered on.
3. Insert one end of the male-male DB-9 cable into the port on the interface
box labeled “COMPUTER.”
4. Insert the other end of the male DB-9 cable into the 9-pin port on the
computer.
5. Tighten the thumb screws on the ends of the plugs to ensure a good
connection.
Warning: Always plug the power cord into the analyzer first, then into the
UPS and then the wall outlet. This prevents the possibility of electrical
shock or power supply damage.
3. Press the green POWER button on the front of the analyzer. The power
button and yellow MOTOR button illuminate when the analyzer is
powered on.
4. Check to ensure the green light on the interface box is illuminated; this
indicates that all the cables are connected properly.
Note: For more information on cautions and electrical requirements for the
TEG analyzer, refer to “Power outlet connection” on page 3-3 and
“Electrical specifications” on page B-3.
Warning: Do not remove the power cord from the back of the analyzer
without first unplugging the UPS from the wall outlet. This prevents the
possibility of electrical shock or power supply damage.
2. Unplug the analyzer power cord from the UPS unit and then from the back
panel of the analyzer.
Connecting You can connect up to four analyzers to a computer using one analog/digital
multiple interface box.
analyzers Make sure you have the following components ready when connecting multiple
analyzers:
⚫ One power supply unit per analyzer (included with analyzer purchase).
⚫ One female-female DB-9 cable per analyzer.
Installing the For instructions on installing and uninstalling the software, refer to the TEG 5000
software System Site Administrator’s Guide.
Note: The fields are completed in an order that is different than the order
they appear on the screen.
Field Description
Note: This step is performed in order to save the information that you
entered on the Setup tab. You are not performing an eTest at this step.
For instructions on performing an eTest, refer to “Performing an eTest” on
page 7-7.
7. Click Done.
Index
INDEX ...........................................................................I-2
INDEX
Symbols assay
Functional Fibrinogen 4-15
§ (sample note) 8-59
PlateletMapping 4-15
¶ (patient note) 8-60
Audit report 8-49
Audit report data screen 8-49
A
A (Activated) sample type 4-13 B
A (Amplitude) parameter 4-7
backing up databases 9-14
A/D interface box
BASE adjustment screw 2-7
connecting
baseline, adjusting eTest values 7-8
multiple analyzers D-5
biological controls. See QC samples
single analyzer D-4
blood collection. See samples, collecting
serial port, description 2-7
Blood Products tab (Edit case screen) 10-13
A30 parameter 4-8
blood products, editing patient case records 10-13
A60 parameter 4-8
blood samples. See sample types; samples
AA (Arachidonic Acid) sample type 4-13
bloodborne pathogens
Access, importing databases into 9-21
accessories, approved use of 11-5 about 3-3
accuracy of TEG analyzer B-6 contaminated materials
disposing of 3-4
ACT (Activated Clotting Time). See RapidTEG
proper handling 3-3
activator
kaolin 4-14
RapidTEG 4-14
active filter 8-21 C
Add Sample type screen 6-12 cables
Add test screen 6-4 carrier ribbon
Add to Guide screen 8-34 checking 7-6
ADP (Adenosine Diphosphate) sample type 4-13 description 2-4
advanced filters power
about 8-21 connecting D-5
case 8-27 description 2-7
channel 8-25 disconnecting D-5
custom 8-28 specification B-3
dates 8-26 warnings 3-3
patient 8-24 requirements for setup D-4
sample type 8-24 serial
test results 8-22 DB-9 female-female D-4, D-5
Amplitude (A) 4-7 DB-9 male-male D-4
Amplitude, 30 mins (A30) 4-8 CAL adjustment screw 2-7
Amplitude, 60 mins (A60) 4-8 calibration
Analog/Digital interface box. See A/D interface box about adjusting 2-7
Analysis report (single-sample, multi-sample) 8-37 preventive maintenance 11-4
Angle calculation, warning 6-18 Capture function, about 12-8
Angle parameter 4-4 Capture options screen 12-8
APR (Aprotinin) sample type 4-13 carrier ribbon cables, checking 7-6
archiving lot numbers 11-13
H L
hardware requirements, setting up analyzer D-4 L parameter 4-11
hazards, electrical shock 3-3, D-5 laboratory techniques
Help menu, about 5-20 avoiding clot activation in samples 11-3
hemostasis profile 4-3 precautions for blood samples 11-3
heparin Level I and Level II controls. See QC samples
about B-5 leveling
avoiding contamination B-5 bubble 2-7
blood sample, collecting 7-20 feet 2-5
neutralizer (heparinase) 4-15 instructions for analyzer 7-6
tubes (Vacutainer®) 7-20 levers
heparinase about 2-4
description 4-15 positions (Load, Test, Eject) 2-4
preparing blood samples 7-21 using
ejecting pins 7-16
performing eTest 7-7
I running samples 7-11, 7-27
icons on toolbar 5-13 Levey Jennings report 8-45, 11-5
illustrations LG parameter 4-10
exterior back 2-6 library, coagulopathy 8-33
exterior front 2-3 limitations for analyzer B-4
Import Text Wizard screen 9-22 lines
importing colored dashed on tracing 8-4
databases into Access 9-21 grid on tracing 8-4
databases into another database 9-7 setting style of 6-22
databases into Excel 9-19 solid on tracing 8-4
Include screen 6-6, 6-14 viewing tracing 8-3
infection control strategies 3-3 Load position, lever 2-4
installing locking patient case records 10-15
analyzer. See setup Locking tab (Edit case screen) 10-15
TEG Analytical Software (TAS) D-6 logging in to TAS
interface box. See A/D interface box how to 5-5
interference factors for analyzer B-4 setting defaults for 6-24
interrelationship of parameters 4-11 system timeout 5-7
Introduction screen 12-4 troubleshooting A-3
Login tab (User profile setup screen) 6-24
Lot Number History report 8-47
lot numbers
K about 11-9
K (clot kinetics) parameter archiving 11-13
about 4-4 creating new 11-9
setting units for 6-17 deleting 11-12
kaolin generating history report 8-47
about 4-12, 4-14 reference ranges for 11-9
preparing samples 7-22 restore old 11-13
sample type (K) 4-12 updating 11-12, 11-15
KH (Kaolin w/Heparinase) sample type 4-12 LTE parameter 4-9
kinetic, strength, and stability parameters 4-9 LY30 parameter 4-5
LY60 parameter 4-8
Lysis Time Estimate (LTE) 4-9
M N
MA (Maximum Amplitude) parameter N (Native) sample type 4-13, 4-14
about 4-5 native blood samples
calculation warning 6-18 about 4-13, 4-14
limitations B-4 preparing 7-21
Maintenance check message 7-5 reference ranges for 11-16
maintenance checks, conducting daily 7-5 needles used to collect samples 7-18
Maintenance filter screen 8-47 No tracing available message 8-13
Maintenance screen 8-47 non-citrated blood samples, collecting 7-18
manually terminate sample run 7-27 normal tracings
Maximized screen displaying 8-12
about 8-4 setting 8-12, 8-55
displaying setting line style 6-23
data for tracing 8-6 Normal Values tab (User profile setup screen) 6-8
multiple tracings 8-7 normal values. See reference ranges
navigating between tracings 8-8 notes
Maximum Amplitude (MA) 4-5 adding or viewing Notes (patient notes) 8-60
menu options, about 5-19 adding or viewing SNotes (sample notes) 8-59
Merge screen 9-18 editing for patient case records 10-14
merging databases 9-17 Notes (sample) screen 8-60
messages Notes icon 8-61
Channel not at equilibrium 7-8, A-9 Notes tab
COM port is in use or does not exist A-6 Detail screen 8-58, 8-59
Compacting could not be completed A-8 Edit case screen 10-14
Database has too many users printing A-23 NWB (Native Whole Blood) sample type 4-13
Delete Lot number 11-12
Disk or network A-7
Does not exist A-4 O
eTest out of range A-9 off-set mode 8-7
eTest value off center 7-8, A-9 operator report, Analyzer Audit 8-49
File is read only A-3 Options menu, about 5-20
Lot number not found A-20 Other tab (Edit case screen) 10-14
No tracing available 8-13
Object or With Block variable not set A-7
QC Alert 7-12
Search key not found A-7
P
Table cannot be locked A-25 parameters
This may not be the best match 8-33 about 4-4, 5-8
Unrecognized database format A-7 flashing 8-3
motor indicator button 2-5 interrelationship of 4-11
MRL parameter 4-11 MA calculation, warning 6-18
MRLG parameter 4-10 primary clot formation and lysis
MRTG parameter 4-11 Angle 4-4
multiple analyzers, setting up D-5 CI (Coagulation Index) 4-5
multi-sample patient report 8-37 K (clot kinetics) 4-4
LY30 (Percent Lysis, 30 mins) 4-5
MA (Maximum Amplitude) 4-5
R 4-4
TEG-ACT 4-6
training 1-4 W
training, using Guide 8-30 Whole Blood Clot Lysis Index
transporting precautions 3-2 30 mins (CL30) 4-8
trimmer adjustment tool 7-9
60 mins (CL60) 4-8
troubleshooting
whole blood samples
about A-2
modified 4-14
blood sample errors A-19
no modifiers 4-14
compacting errors A-8
sodium-citrated 4-14
cup and pin errors A-11
workstation access, troubleshooting A-22
database errors A-7
eTest errors A-9
login errors A-3
printing errors A-23
Z
QC sample errors A-12 zoom level for VCurve 6-19
reference ranges errors A-21
remote workstation access errors A-22
temperature errors A-20
unexpected tracing results A-16
U
Uninterruptible Power Supply (UPS) D-4
unpacking the analyzer D-3
User profile setup screen 6-2
user profiles
about 6-2
accessing User profile setup screen 6-2
eConsult 6-25
login 6-24
reference ranges 6-8
sample types 6-11
software 6-15
tests 6-3
video 6-20
V
Vacutainer® tubes 7-20
VCurve
converting to G values 6-19
parameters 4-10
setting options for 6-19
velocity parameters
See also VCurve
about 4-10
verifying QC sample results 7-13
Video tab (User profile setup screen) 6-20