You are on page 1of 3

SOP for Operations & Calibration of Laboratory Glasswares (Volumetric, Micro, Chemical, Analytical)

Objective:

This Standard Operating Procedure (SOP) outlines the procedures for the operation and calibration
of laboratory glassware, including Volumetric, Micro, Chemical, and Analytical glassware.

Scope:

This SOP applies to all laboratory glassware used for various analytical and volumetric purposes at
[Laboratory Name/Location].

Equipment:

Laboratory Glassware (Volumetric, Micro, Chemical, Analytical)

Analytical Balance (for calibration, if applicable)

Calibration standards (for volumetric glassware)

Responsibility:

[List Responsible Personnel Names and Roles]

Procedure for Operation:

Preparation:

a. Ensure that the laboratory glassware is clean and free from contaminants.

b. Verify that glassware is properly labeled for identification.

Usage:

a. Use the appropriate glassware for the intended purpose.

b. Handle glassware with care to avoid breakage or damage.

c. Follow specific procedures for sample preparation, transfer, or measurement as required.

Cleaning:

a. After use, clean glassware thoroughly using suitable cleaning agents.

b. Rinse glassware with distilled water multiple times to remove residues.


c. Allow glassware to dry before storage.

Storage:

a. Store glassware in a clean, dust-free environment.

b. Protect glassware from physical damage or contamination.

Procedure for Calibration (Volumetric Glassware Only):

Preparation:

a. Ensure that the glassware to be calibrated is clean and dry.

b. Verify that calibration standards are available and suitable for the calibration.

Instrument Calibration:

a. Perform the calibration of volumetric glassware following established methods and calculations.

b. Record calibration details, including date, time, and results, in the calibration log.

Calibration Verification:

a. After calibration, perform a verification test using calibration standards.

b. Compare the measured values with the expected values.

c. Record the verification results in the calibration log.

Documentation:

a. Maintain records of instrument calibration, verification, and any deviations.

b. All records should be signed and dated by authorized personnel.

Maintenance:

Regularly inspect laboratory glassware for damage or defects.

Replace damaged or defective glassware promptly.

Follow specific maintenance procedures for specialized glassware if required.

Attachments:
[Include any relevant attachments or reference to attachments]

You might also like