FLOW RATE ACCURACY REPORT

A C Q U I S I TI O N I N F O R M A TI O N

System Name: e2695_2475 MSG Sample Set Id: 7837


Acquired By: Melissa Result Set Id: 7846
Project Path: RC2_SQT _e2695_2475 final

Flow Rate Accuracy is determined manually by recording the delivery of a set volume of mobile phase into a
Class A volumetric flask utilizing a calibrated stop watch.

The specification for Flow Rate Accuracy includes stop watch accuracy, the flask accuracy and operator error
allowance.

The Flow Rate Accuracy test passes only when the calculated absolute flow accuracy is within 0.494 to
0.506 mL/min for the 0.5 mL/min test, 1.98 to 2.02 mL/min for the 2.0 mL/min test, and 4.94 to 5.06 mL/min
for the 5.0 mL/min test.

Flow Rate Accuracy 0.5 mL/min Results


Serial Sample Flow Rate Observed Observed Flow Rate Accuracy
Result Id
Number Name Spec (mL/min) Minutes Seconds (mL/min) Pass/Fail
1 K13475622R Alliance Flow Accuracy 0.494 to 0.506 20.0 3.0 0.499 PASS 7854

Flow Rate Accuracy 2.0 mL/min Results


Serial Sample Flow Rate Observed Observed Flow Rate Accuracy
Result Id
Number Name Spec (mL/min) Minutes Seconds (mL/min) Pass/Fail
1 K13475622R Alliance Flow Accuracy 1.98 to 2.02 5.0 2.0 1.99 PASS 7855

Flow Rate Accuracy 5.0 mL/min Results


Serial Sample Flow Rate Observed Observed Flow Rate Accuracy
Result Id
Number Name Spec (mL/min) Minutes Seconds (mL/min) Pass/Fail
1 K13475622R Alliance Flow Accuracy 4.94 to 5.06 4.0 58.0 5.03 PASS 7856

Waters Project Name and Version: SQT_E_2695_2475 v4.00 © 2007 - 2017 Waters Corporation. All rights reserved.
Date: Monday, February 20, 2017
Printed: 11:01:47 AM US/Eastern Page: 1 of 1