FLOW RATE ACCURACY REPORT

A C Q U I S I TI O N I N F O R M A TI O N

System Name: 2695_2998_ISM_QDa72 Acquired By: ryan


Processing Method: Flow Rate Acc PM Sample Kit 2 Lot Number: 1245
Project Name: SQT_E_2695_2998_QDa_P_ISM_RC2 Sample Kit 2 Exp. Date: 12/22/2014 12:00:00 AM
Sample Set Id: 4210 Column Serial Number: W1412341AD
Sample Set Name: Flow Rate Acc SSM Result Set Id: 4220
Sample Set Start Date: 12/22/2014 3:20:20 PM EST

TE S T D E S C R I P TI O N

Flow Accuracy is determined by measuring the flow rate of the pump head A and pump head B from the Binary
Solvent Manager and comparing it against the set-point programmed on the instrument.
The test requires a calibrated stopwatch.

The Absolute Flow accuracy should be within 0.5 to 2.0 mL/min +/- 1%.

Flow Rate Accuracy 0.5 mL/min Results


Observed Observed Flow Rate Accuracy
Flow Rate Spec
Minutes Seconds (mL/min) Pass/Fail
1 0.495 to 0.505 mL/min 20 0 0.5000 PASS

Flow Rate Accuracy 2.0 mL/min Results


Observed Observed Flow Rate Accuracy
Flow Rate Spec
Minutes Seconds (mL/min) Pass/Fail
1 1.98 to 2.02 mL/min 5 0 2.000 PASS

Printed 10:11:20 AM US/Eastern 12/30/2014 Page: 1 of 1


Waters Project Name and Version: SQT_E_2695_2998_QDa_P_ISM Version 1.00