A C Q U I S I TI O N I N F O R M A TI O N
System Name: BSM_SM_TQD Acquired By: sqte
Project Name: SQT_E_BSM_SM_TQD_data Column serial Number:
Sample Set Method: 04b MS Detector Lin Sens SSM Sample Kit 1 Expiration Date:
Processing Method MS Detector Area PM, MS Detector Sample Kit 1 Lot Number:
Report Method Name: MS Detector Lin Sens RM Build Version: Empower 2 Software Build 2154 SPs
User Name: System Sample Set Id: 2512 Result Set Id 3099
Sample Set Name: MS Detector Lin Sens SSM Sample Set Start Date: 1/4/2011 10:06:59 PM EST
T he MS Detector Linearity qualification injects several amounts of Sulfadimethoxine and plots the
Response vs. Sulfadimethoxine Amount.
T he qualification passes if the Linearity R^2 value for Area is > or = 0.990 and the R^2 value for Height
is > or = 0.990.
A C Q U I S I TI O N I N F O R M A TI O N
System Name: BSM_SM_TQD Acquired By: sqte
Project Name: SQT_E_BSM_SM_TQD_data Column serial Number:
Sample Set Method: 04b MS Detector Lin Sens SSM Sample Kit 1 Expiration Date:
Processing Method MS Detector Area PM, MS Detector Sample Kit 1 Lot Number:
Report Method Name: MS Detector Lin Sens RM Build Version: Empower 2 Software Build 2154 SPs
User Name: System Sample Set Id: 2512 Result Set Id 3099
Sample Set Name: MS Detector Lin Sens SSM Sample Set Start Date: 1/4/2011 10:06:59 PM EST
Calibration Plot
Area v s. Amount
1.0x107
8.0x106
6.0x106
Area
4.0x106
2.0x106
0.0
0.00 50.00 100.00 150.00 200.00 250.00 300.00 350.00 400.00 450.00 500.00
Amount
Area R^2: 0.99963
A C Q U I S I TI O N I N F O R M A TI O N
System Name: BSM_SM_TQD Acquired By: sqte
Project Name: SQT_E_BSM_SM_TQD_data Column serial Number:
Sample Set Method: 04b MS Detector Lin Sens SSM Sample Kit 1 Expiration Date:
Processing Method MS Detector Area PM, MS Detector Sample Kit 1 Lot Number:
Report Method Name: MS Detector Lin Sens RM Build Version: Empower 2 Software Build 2154 SPs
User Name: System Sample Set Id: 2512 Result Set Id 3099
Sample Set Name: MS Detector Lin Sens SSM Sample Set Start Date: 1/4/2011 10:06:59 PM EST
Calibration Plot
Height v s. Amount
4x106
3x106
Height
2x106
1x106
0
0.00 50.00 100.00 150.00 200.00 250.00 300.00 350.00 400.00 450.00 500.00
Amount
Height R^2: 0.9992
A C Q U I S I TI O N I N F O R M A TI O N
System Name: BSM_SM_TQD Acquired By: sqte
Project Name: SQT_E_BSM_SM_TQD_data Column serial Number:
Sample Set Method: 04b MS Detector Lin Sens SSM Sample Kit 1 Expiration Date:
Processing Method MS Detector Area PM, MS Detector Sample Kit 1 Lot Number:
Report Method Name: MS Detector Lin Sens RM Build Version: Empower 2 Software Build 2154 SPs
User Name: System Sample Set Id: 2512 Result Set Id 3099
Sample Set Name: MS Detector Lin Sens SSM Sample Set Start Date: 1/4/2011 10:06:59 PM EST
3.5x106
3.0x106
2.5x106
2.0x106
Intensity
1.5x106
1.0x106
5.0x105
0.0
0.20 0.40 0.60 0.80 1.00 1.20 1.40 1.60 1.80 2.00
Minutes