Project Name: SQT_E_BSM_FTN_TUV_ELSD Column serial Number: 017139167157 42 Sample Set Method: 030 Wavelength Accuracy SSM Sample Kit 1 Expiration date: 1/25/2012 12:00:00 AM EST Processing Method Wavelength Accuracy PM Sample Kit 1 Lot Number: W04040902 Report Method Name: Wavelength Accuracy RM Build Version: Empower 2 Software Build 2154 SPs Installed: User Name: Lisa Busher (lbusher) Sample Set Id: 6491 Result Set Id 6502 Sample Set Name: Wavelength Accuracy SSM Sample Set Start Date: 11/30/2011 8:14:00 PM EST
Wavelength accuracyis determined byidentifying the 2 lambda max values of caffeine.
The spectrumof caffeine is well defined with a lambda max of 205.0 nmand 273.0 nm. Empower measures deviations fromthese wavelengths to determine the accuracy of the detector. The detector wavelength accuracymust be within ±2 nmof each Caffeine lambda max to pass.
Wavelength Accuracy Results
Specification 1 Lambda Max Lambda Max Pass Fail
1 203nm-207nm 204 PASS
Specification 2 Lambda Max Lambda Max Pass Fail
1 271nm-275nm 273 PASS
Printed 6:26:12 PM US/Eastern 1/16/2012 Page: 1 of 3
WAVELENGTH ACCURACY REPORT ACQUI SI TI ON I NFORMATI ON
Channel ACQUITY TUV ChA 271nm Channel ACQUITY TUV ChA 272nm Channel ACQUITY TUV ChA 273nm Channel ACQUITY TUV ChA 274nm Channel ACQUITY TUV ChA 275nm Channel ACQUITY TUV ChA 276nm
Printed 6:26:12 PM US/Eastern 1/16/2012 Page: 2 of 3
WAVELENGTH ACCURACY REPORT ACQUI SI TI ON I NFORMATI ON