Professional Documents
Culture Documents
UV Detector Linearity & Sensitivity Report
UV Detector Linearity & Sensitivity Report
ACQUI SI TI ON I NFORMATI ON
The Detector Linearity &Sensitivity qualification injects several amounts of Caffeine and plots the Response vs.
Caffeine Amount.
It also computes the Sensitivity (PeakArea/Amount) and (Peak Height/Amount) for each injection.
The qualification passes if the Linearity R^2 value for Area is > or = 0.9990 and the R^2 value for Height
is > or = 0.990. The %RSD of the Sensitivity must be < or = 3.5%for Area and < or = 4.5%for Height.
ACQUI SI TI ON I NFORMATI ON
Calibration Plot
Area vs. Amount
2.5x106
2.0x106
1.5x106
Area
1.0x106
5.0x105
0.0
0.00 0.02 0.04 0.06 0.08 0.10 0.12 0.14 0.16
Amount
Area R^2: 0.99988
ACQUI SI TI ON I NFORMATI ON
Calibration Plot
Height vs. Amount
2.5x106
2.0x106
Height
1.5x106
1.0x106
5.0x105
0.0
0.00 0.02 0.04 0.06 0.08 0.10 0.12 0.14 0.16
Amount
Height R^2: 0.9997
ACQUI SI TI ON I NFORMATI ON
350.00
300.00
250.00
200.00
AU
150.00
100.00
50.00
0.00
0.00 0.10 0.20 0.30 0.40 0.50 0.60 0.70 0.80 0.90 1.00
Minutes
ACQUI SI TI ON I NFORMATI ON