Adjustment
System
17 Inch Monitor Adjustment
Document Version
Siemens Healthcare GmbH reserves the right to change its products and services at any
time.
In addition, manuals are subject to change without notice. The hardcopy documents cor‐
respond to the version at the time of system delivery and/or printout. Versions to hard‐
copy documentation are not automatically distributed.
Please contact your local Siemens Healthineers office to order a current version or refer to
our website http://www.siemens-healthineers.com.
Disclaimer
Siemens Healthcare GmbH provides this documentation “as is” without the assumption of
any liability under any theory of law.
The content described herein requires superior understanding of our equipment and may
only be performed by qualified personnel who are specially trained for such installation
and/or service.
Copyright
“© Siemens Healthineers, 2011” refers to the copyright of a Siemens Healthineers entity
such as:
Siemens Healthcare GmbH - Germany
Siemens Shenzhen Magnetic Resonance Ltd. - China
Siemens Shanghai Medical Equipment Ltd. - China
Siemens Healthcare Private Ltd. - India
Siemens Medical Solutions USA Inc. - USA
Siemens Healthcare Diagnostics Inc. - USA and/or
Siemens Healthcare Diagnostics Products GmbH - Germany
1.1 Overview . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4
1.2 Control Locations and Functions. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6
1.3 Accessing the Monitor Control Screen . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9
1.3.1 To Adjust the Contrast and Brightness . . . . . . . . . . . . . . . . . . . . . 9
1.3.2 To Access the OSD Menu. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10
1.3.3 To Adjust the Color Mode . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 11
1.3.4 OSD Menu Key Lock and Unlock . . . . . . . . . . . . . . . . . . . . . . . . . . 18
1.3.5 Variable LAMP Intensity. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 19
1.3.6 No SIgnal Message . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 19
1.3.7 Out of Range Message . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 19
1.3.8 SBC Error Message. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 20
1.1 Overview
The ACUSON X300TM and ACUSON X300TM PE ultrasound system comes with an optional
17 inch, SXGA color monitor with 1280 x 1024 pixels. The following figure shows the
monitor.
Fig. 1: 17 Inch Monitor Shown
Operating at 47Hz to 63Hz frequency range, this non-interlaced monitor has four speak‐
ers - two mid/tweeters and two woofers for use with spectral Doppler.
The monitor controls include:
n Degauss
n Brightness
n Contrast
n Test buttons (user serviceable adjustments are located inside the monitor). Refer to
the FPD section in the Users Manual.
CAUTION
The weight of the monitor (6.5 kg: 14.3 lb.) represent a lifting hazard.
The use of improper lifting methods or lifting in an awkward manner can cause
serious bodily injury.
Use proper lifting methods.
Name Description
Speaker (3W 8 ohm/ 5W 4ohm) X 2
Menu Button Set or Exit the OSD Menu
Down Button Move or Adjust value in OSD Menu
Name Description
Microphone and MIC Microphone and Mic LED
LED
Light Button For light intensity of 4 steps as OFF,
Low, Middle and Max
Up Button Move or Adjust value of OSD Menu
On / Off LED Normal: Green, No Signal: Amber
Digital Input Connector 19P HDMI Input
AC Input Connector AC100V -AC240V
Light Hi-luminance LED
The FPD of ACUSON X300 PE systems has eight video modes which have different con‐
trast and brightness values. Those are controlled by ultrasound system input signal (refer
to table below).
3. Press the Menu button once to activate the PICTURE sub menu.
Fig. 9:
Fig. 11:
6. Select Color Temperature with the Menu button. The default value is 9300K.
- Select 9300K for WW setting
- Select 11000K for SONOVISTA setting
8. Select EXIT with the UP button, press MENU to exit from the menu.
Fig. 13:
9. Press the Menu button after each EXIT. The OSD screen will disappear.
Fig. 14:
The OSD Menu will automatically clear if no button is pressed within 10 seconds.
Adjust Menu is used only for the Analog RGB input signal.
2. Press the up or down buttons approximately five seconds. The following message dis‐
plays indicating that the OSD menu key is unlocked.
Fig. 17: OSD Menu Unlocked