ACUSON X300

Adjustment
System
17 Inch Monitor Adjustment

This document is valid for:


ACUSON X300, ACUSON X300 premium edition (PE)
2 Document Version / Disclaimer / Copyright

Document Version
Siemens Healthcare GmbH reserves the right to change its products and services at any
time.
In addition, manuals are subject to change without notice. The hardcopy documents cor‐
respond to the version at the time of system delivery and/or printout. Versions to hard‐
copy documentation are not automatically distributed.
Please contact your local Siemens Healthineers office to order a current version or refer to
our website http://www.siemens-healthineers.com.

Disclaimer
Siemens Healthcare GmbH provides this documentation “as is” without the assumption of
any liability under any theory of law.
The content described herein requires superior understanding of our equipment and may
only be performed by qualified personnel who are specially trained for such installation
and/or service.

Copyright
“© Siemens Healthineers, 2011” refers to the copyright of a Siemens Healthineers entity
such as:
Siemens Healthcare GmbH - Germany
Siemens Shenzhen Magnetic Resonance Ltd. - China
Siemens Shanghai Medical Equipment Ltd. - China
Siemens Healthcare Private Ltd. - India
Siemens Medical Solutions USA Inc. - USA
Siemens Healthcare Diagnostics Inc. - USA and/or
Siemens Healthcare Diagnostics Products GmbH - Germany

ACUSON X300 US11-102.842.01.01.02 Page 2 of 23 © Siemens Healthineers, 2011


08.11
Table of Contents 3

1 17 Inch Monitor Adjustment 4

1.1 Overview . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4
1.2 Control Locations and Functions. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6
1.3 Accessing the Monitor Control Screen . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9
1.3.1 To Adjust the Contrast and Brightness . . . . . . . . . . . . . . . . . . . . . 9
1.3.2 To Access the OSD Menu. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10
1.3.3 To Adjust the Color Mode . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 11
1.3.4 OSD Menu Key Lock and Unlock . . . . . . . . . . . . . . . . . . . . . . . . . . 18
1.3.5 Variable LAMP Intensity. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 19
1.3.6 No SIgnal Message . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 19
1.3.7 Out of Range Message . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 19
1.3.8 SBC Error Message. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 20

2 Changes to Previous Version 21

2.1 Changes to Previous Version . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 21

3 List of Hazard IDs 22

© Siemens Healthineers, 2011 US11-102.842.01.01.02 Page 3 of 23 ACUSON X300


08.11
4 1 17 Inch Monitor Adjustment

1.1 Overview
The ACUSON X300TM and ACUSON X300TM PE ultrasound system comes with an optional
17 inch, SXGA color monitor with 1280 x 1024 pixels. The following figure shows the
monitor.
Fig. 1: 17 Inch Monitor Shown

Operating at 47Hz to 63Hz frequency range, this non-interlaced monitor has four speak‐
ers - two mid/tweeters and two woofers for use with spectral Doppler.
The monitor controls include:
n Degauss
n Brightness
n Contrast
n Test buttons (user serviceable adjustments are located inside the monitor). Refer to
the FPD section in the Users Manual.

ACUSON X300 US11-102.842.01.01.02 Page 4 of 23 © Siemens Healthineers, 2011


08.11
17 Inch Monitor Adjustment 1 5

 CAUTION
The weight of the monitor (6.5 kg: 14.3 lb.) represent a lifting hazard.
The use of improper lifting methods or lifting in an awkward manner can cause
serious bodily injury.
 Use proper lifting methods.

© Siemens Healthineers, 2011 US11-102.842.01.01.02 Page 5 of 23 ACUSON X300


08.11
6 1 17 Inch Monitor Adjustment

1.2 Control Locations and Functions


Fig. 2: 17 Inch Monitor Control Buttons Shown

(1) Menu Button


(2) Up Button
(3) Down Button
(4) Light Button

ACUSON X300 US11-102.842.01.01.02 Page 6 of 23 © Siemens Healthineers, 2011


08.11
17 Inch Monitor Adjustment 1 7

Fig. 3: Connectors on the Rear of the Monitor

Fig. 4: Control Locations on the bottom of the monitor

Tab. 1 Control Locations and Functions

Name Description
Speaker (3W 8 ohm/ 5W 4ohm) X 2
Menu Button Set or Exit the OSD Menu
Down Button Move or Adjust value in OSD Menu

© Siemens Healthineers, 2011 US11-102.842.01.01.02 Page 7 of 23 ACUSON X300


08.11
8 1 17 Inch Monitor Adjustment

Name Description
Microphone and MIC Microphone and Mic LED
LED
Light Button For light intensity of 4 steps as OFF,
Low, Middle and Max
Up Button Move or Adjust value of OSD Menu
On / Off LED Normal: Green, No Signal: Amber
Digital Input Connector 19P HDMI Input
AC Input Connector AC100V -AC240V
Light Hi-luminance LED

ACUSON X300 US11-102.842.01.01.02 Page 8 of 23 © Siemens Healthineers, 2011


08.11
17 Inch Monitor Adjustment 1 9

1.3 Accessing the Monitor Control Screen


Use the following information to assist in accessing the Monitor Control Screen and set‐
tings.

1.3.1 To Adjust the Contrast and Brightness


Do the following:
n Press MENU. The Contrast menu displays, ( Fig. 5 Page 9). Pressing the menu but‐
ton will cycle through the different sub menus.
n Select the adjustment values with the Up and Down buttons.
Fig. 5: Contrast and Brightness

The FPD of ACUSON X300 PE systems has eight video modes which have different con‐
 trast and brightness values. Those are controlled by ultrasound system input signal (refer
to table below).

Tab. 2 Mode 9300K (WW)

Mode Gamma Contrast Brightness


0 X300 r2.0 default gam‐ 73 47
ma
1 Linear FPD gamma 73 47
2 Linear FPD gamma 80 54
3 Linear FPD gamma 80 49
4 Linear FPD gamma 85 50
5 Linear FPD gamma 90 50
6 Linear FPD gamma 95 54
7 Linear FPD gamma 100 54

© Siemens Healthineers, 2011 US11-102.842.01.01.02 Page 9 of 23 ACUSON X300


08.11
10 1 17 Inch Monitor Adjustment

Tab. 3 Mode 11000K (JPN)

Mode Gamma Contrast Brightness


0 X300 r2.0 default gam‐ 82 72
ma
1 Linear FPD gamma 82 72
2 X300 r2.0 default gam‐ 80 61
ma
3 Linear FPD gamma 73 45
4 Linear FPD gamma 78 45
5 Linear FPD gamma 68 42
6 Linear FPD gamma 73 42
7 Linear FPD gamma 78 42

1.3.2 To Access the OSD Menu


To access the OSD Menu, perform the following:
n Press the MENU and UP button simultaneously, about 10 seconds, until the OSD Main
Menu displays.
Fig. 6: OSD Main Menu

ACUSON X300 US11-102.842.01.01.02 Page 10 of 23 © Siemens Healthineers, 2011


08.11
17 Inch Monitor Adjustment 1 11

1.3.3 To Adjust the Color Mode


1. Press Menu and UP simultaneously for about 10 seconds.
Fig. 7: Menu and Up Buttons Shown

(1) Menu Button


(2) Up Button

© Siemens Healthineers, 2011 US11-102.842.01.01.02 Page 11 of 23 ACUSON X300


08.11
12 1 17 Inch Monitor Adjustment

2. The OSD picture displays. The sub-menu PICTURE is not activated.


Fig. 8:

3. Press the Menu button once to activate the PICTURE sub menu.
Fig. 9:

ACUSON X300 US11-102.842.01.01.02 Page 12 of 23 © Siemens Healthineers, 2011


08.11
17 Inch Monitor Adjustment 1 13

4. Press the UP button twice to reach the COLOR MODE.


Fig. 10:

5. Go to the COLOR MODE menu.


- Press MENU button once to go to the COLOR MODE sub menu.
- The sub OSD menu for 9300K, 11000K, SPARE and USER displays.

© Siemens Healthineers, 2011 US11-102.842.01.01.02 Page 13 of 23 ACUSON X300


08.11
14 1 17 Inch Monitor Adjustment

Fig. 11:

6. Select Color Temperature with the Menu button. The default value is 9300K.
- Select 9300K for WW setting
- Select 11000K for SONOVISTA setting

ACUSON X300 US11-102.842.01.01.02 Page 14 of 23 © Siemens Healthineers, 2011


08.11
17 Inch Monitor Adjustment 1 15

7. To change the Color Temperature:


- Press Menu one time to change the Color Temp to 11000K. Each time the Menu
button is pressed, the Color Temp changes.
Fig. 12:

© Siemens Healthineers, 2011 US11-102.842.01.01.02 Page 15 of 23 ACUSON X300


08.11
16 1 17 Inch Monitor Adjustment

8. Select EXIT with the UP button, press MENU to exit from the menu.
Fig. 13:

9. Press the Menu button after each EXIT. The OSD screen will disappear.
Fig. 14:

ACUSON X300 US11-102.842.01.01.02 Page 16 of 23 © Siemens Healthineers, 2011


08.11
17 Inch Monitor Adjustment 1 17

 The OSD Menu will automatically clear if no button is pressed within 10 seconds.

 Adjust Menu is used only for the Analog RGB input signal.

© Siemens Healthineers, 2011 US11-102.842.01.01.02 Page 17 of 23 ACUSON X300


08.11
18 1 17 Inch Monitor Adjustment

Fig. 15: Monitor Menu Tree

1.3.4 OSD Menu Key Lock and Unlock


Do the following to Lock and Unlock the OSD Menu:

ACUSON X300 US11-102.842.01.01.02 Page 18 of 23 © Siemens Healthineers, 2011


08.11
17 Inch Monitor Adjustment 1 19

Locked OSD Menu key,


1. Press the up or down buttons approximately five seconds. The following message dis‐
plays indicating that the OSD menu key is locked.
Fig. 16: OSD Menu Locked

2. Press the up or down buttons approximately five seconds. The following message dis‐
plays indicating that the OSD menu key is unlocked.
Fig. 17: OSD Menu Unlocked

1.3.5 Variable LAMP Intensity


Press the LIGHT button. The intensity of the LED is changed to On. Continue pressing the
LIGHT button to cycle through the different levels of LED intensity, OFF, Low, Middle and
Max.

1.3.6 No SIgnal Message


If the following message displays indicating that there is no signal, check the signal cable.
Fig. 18: No Signal Message

1.3.7 Out of Range Message


The following message displays when the input frequency is out of range.

© Siemens Healthineers, 2011 US11-102.842.01.01.02 Page 19 of 23 ACUSON X300


08.11
20 1 17 Inch Monitor Adjustment

Fig. 19: OSD Out of Range Message

1.3.8 SBC Error Message


The following message displays when there is an SBC error:
Fig. 20: SBC Error Message

ACUSON X300 US11-102.842.01.01.02 Page 20 of 23 © Siemens Healthineers, 2011


08.11
Changes to Previous Version 2 21

2.1 Changes to Previous Version


Change(s) Description
Initial Release N/A

© Siemens Healthineers, 2011 US11-102.842.01.01.02 Page 21 of 23 ACUSON X300


08.11
22 3 List of Hazard IDs

There are no Hazard IDs in this document.

ACUSON X300 US11-102.842.01.01.02 Page 22 of 23 © Siemens Healthineers, 2011


08.11
All documents may only be used for
rendering services on Siemens
Healthcare Products. Any document
in electronic form may be printed
once. Copy and distribution of elec‐
tronic documents and hardcopies is
prohibited. Offenders will be liable
for damages. All other rights are re‐
served.

siemens-healthineers.com

Siemens Healthineers Headquarters


Siemens Healthcare GmbH
Henkestr. 127
91052 Erlangen
Germany
Telephone: +49 9131 84-0
siemens-healthineers.com

Print No.: US11-102.842.01.01.02 | Replaces: n.a.


Doc. Gen. Date: 08.11 | Language: English