Professional Documents
Culture Documents
Supporting
Proposed Inspection Date: 24-Oct-2023 Inspection time: 10am
Documents
CTX0152 CT Scanner, Multi-Slice, 64-320 Slice SIEMENS SOMATOM Drive ATTACHED LIST 2
SUB-CONTRACTOR CONTRACTOR
Engineer Engineer / CM QA/QC Representative Project Director
Requested by: Requested/Verified by: Verified by: Approved by:
Signature:
Company: Ali Bin Ali Medical Signature: Signature: Signature:
Name: Paulino Flores Name: Mr. Ibrahim Name: Mr. Rakesh NK Name: Mr. Alaaddin Aktan
Date: 23-Oct-2023 Date: 23-Oct-2023 Date: 23-Oct-2023 Date: 23-Oct-2023
EMPLOYER
Comments / Recommendation:
Employer Action Code: A (Approved) B (Approved with Comments) C (Revise & Resubmit) D (Rejected)
Note: To attach the Employer's comment response sheet after the inspection.
The approval and comments shall not in any ways relieve or absolve the Contractor from compliance of any obligations or liability under or in connection with the Contract, Project
Requirements and Specification.
MMC-HBK-QA-F024 Rev.0, Dated 12.09.2023 Page 1 of 1
MEDICAL EQUIPMENT
COMMISSIONING REPORT
MEQ INSTALLATION AND TEST REPORT DETAILS (to attached approval copy)
Installation and Test Report No. (ITR): MMC-HBK-AP-IR-MD-00087 Rev. No.: 0 Date: 23-Oct-23
COMMISSIONING
STATUS
SN ACTIVITIES
PASSED FAILED N/A
1 Software Integration
Consumables Check
2
(to be applied to the Consumables NOT inspected during the Testing phase)
Legend: Passed (Successfully Performed) Failed (Not Successfully Performed) N/A (Not Applicable)
DOCUMENTATION
Remarks:
Enclosures
6.a Verification
6.b Image QA
6.c Functional and operational test as per the manufacture standards
6.d Test #1 not done in the RFIT:
6.e Test #2 not done in the RFIT:
Integration with other Medical equipment
7
Tag No. Item No. Item Description
7.a
7.b
7.c
7.d
7.e
8 Tags
Notes
Supporting
Proposed Inspection Date: 28-Sep-2023 Inspection time: 02:00pm
Documents
Location: Building: Zone: Level: Approved Drawing
B01 Z01 L0
Material Approval
MST / ITP
Room No.: Attached List Room Name: CT Room Others
MEQ Inspection of 0
SUB-CONTRACTOR CONTRACTOR
Engineer Engineer / CM QA/QC Representative Project Director
Requested by: Requested/Verified by: Verified by: Approved by:
Signature:
Company: Ali Bin Ali Medical Signature: Signature: Signature:
Name: Mr. Paulino Flores Name: Mr. Ibrahim Name: Mr. Rakesh NK Name: Mr. Alaaddin Aktan
Date: 28-Sep-2023 Date: 28-Sep-2023 Date: 28-Sep-2023 Date: 28-Sep-2023
EMPLOYER
Comments / Recommendation:
Employer Action Code: A (Approved) B (Approved with Comments) C (Revise & Resubmit) D (Rejected)
Note: To attach the Employer's comment response sheet after the inspection.
The approval and comments shall not in any ways relieve or absolve the Contractor from compliance of any obligations or liability under or in connection with the Contract, Project
Requirements and Specification.
MMC-HBK-QA-F007 Rev.7, Dated 08.10.2022 Page 1 of 1
MATERIAL INSPECTION REQUEST (MIR)
Proposed Inspection Date: 9-Jul-2023 Inspection Time: 10:00 AM Location: HBK Site Store (On-site)
Supporting Documents: Material Approval Delivery Note Country of Origin Test Reports Others
Country of
Cad ID MEQ Item Description Manufacturer Model Quantity
Origin
SUB-CONTRACTOR CONTRACTOR
Engineer Store QA/QC Representative Project Director
Requested by: Requested/Verified by: Verified by: Approved by:
EMPLOYER
Comments / Recommendation:
1- Refer the attached Comments from HBK/SECO
2- Initially accepted as Boxes with SNs and PNs (Not assembled)
3- Final approval is subjected to final installation, testing & commissioning.
Employer Action Code: A (Approved) B (Approved with Comments) C (Revise & Resubmit) D (Rejected)
Note: To attach the Employer's comment response sheet after the inspection.
This document shall not relieve Contractor of his liabilities under the Contract or constitute authorization of any change to Contract documents.
001
234
234
56789 ÿ76
ÿ.-#52ÿ4224#,2/ÿ(*+
(Fÿ99689 +
ÿ//4@
ÿ9879ÿ,782
279ÿ88679 ÿ60968ÿ789ÿ61989 ÿ879ÿG2,-.ÿ/2422ÿÿ/01123ÿ688763ÿ29876 ÿ4ÿ#77676 ÿÿ#25$*%)ÿ#2ÿ/#4
67769ÿ"ÿ/8ÿ.7893ÿ9ÿ#2ÿÿ:*"$"."$)*ÿ6 ÿ:*"$"."$)%Gÿÿ787ÿ879ÿ76 ÿ908ÿ?
HIJKÿMNOPKQJMMORÿSOKÿOTMUQOMJVORRWÿXYPYNOMYZ[
E
N
8 9
B07 B08
S
N
LA
.63
W
QIB
252
B14
P2.F Notes on preparations for installation B05
CORRIDOR
B09
B02 B03
B01
Contracts for performing and supervising on-site installation preparations should be concluded with technically
CONTRAST
B06
competent companies by the customer. The customer is responsible for timely and proper completion and B12 B13
B11
Z1-L0-IMC-004 AGENT supervision of all preparations for installation at the construction site in observance of all applicable legal regulations B04
B10
(e.g. X-ray regulations, radiation protection regulations) and all applicable general recognized rules of technology
PREP (e.g. VDE regulations, DIN standards).
Execution and supervision of installation preparations at the construction site and later observance of the standard
14.04 m² EQUIPMENT AT operating conditions are not included in our duties. The customer is responsible for checking the static calculations
CHANGE ROOM and, where applicable, the air conditioning in the building to be equipped.
Z1-L0-IMD-024 Z1-L0-IMD-027
Z1-L0-IMD-022 (M) GENERAL NOTES
Z1-L0-IMD-026 7.08 m² 1. PROJECT REFERENCE ELEVATION ±0.00 IS AT +27.50 QATAR NATIONAL DATUM. ELEVATIONS
SHOWN ARE WITH RESPECT TO THE PROJECT REFERENCE ELEVATION OF ±0.00 AT GROUND FLOOR.
2. ALL DIMENSIONS ARE IN MM UNLESS NOTED OTHERWISE.
P2.G
KEY PLAN 1
10 10.1 17 17.1
26
1046
A A
Room lighting
Z1 Z3
SBA
B Ambient lighting in rooms with diagnostics or with workstations must comply with the respective local and national
P2.1
F Z2 F
P3.1
ISO CENTER
2.09
P2.A P3.A
EAT AT General requirements like the needed intensity of illumination - adjustable, reproducible, flicker-free or a limitation of
Z1-L0-IMD-025
2481
2480
dazzlings and reflections etc. have to be observed (EN 12464-1, DIN 5035-7).
440
P2.F P3.F
AT P2.1 P3.1
Z4 Z5
ISO CENTER ISO CENTER
5161 2760 V V
5421
MMC-HBK-AP-DD-BAP-MEQ-L0-APO-1800 2
2500
WORK / 2.07
Approved as
Commented
ISO CENTER
Re-Submit
A Approved
This is to avoid the risk of injury.
D Rejected
Revise &
± 0.00 Authority Signature Date
2940
If safety distances are not maintained appropriate on-site safety measures have to be put in place.
-0.10
C
B
Clear visible markings according to national guidelines, e.g. crushing warning signs, hazard warning
AT 2.10 MMC-PMO
tape, hazard area cordon, safety mats, may be required.
2953
A
SW
3131 1
1100
2.05 The approval and comments shall not in any ways relieve or absolve the
Contractor from compliance of any obligations or liability under or in connection
Room height with the contract, project requirements and specifications.
Measured from the highest point of the finished floor (with covering) to the lowest point of the ceiling
2.08 2.04 2.03
00 ISSUED FOR CONSTRUCTION 31/10/2022 ABAM ABAM PR SG
REVISION DETAILS DATE DWN DES CHD APD
M Lead shielded wall EMPLOYER
Transport dimensions Length (mm) Width (mm) Height (mm) Weight (kg) HBK CONTRACTING CO. WLL
P.O. BOX 1362 DOHA, QATAR
PDC A with transport pallet - without UPS 1120 900 2070 ≤600 TEL: (974) 44451111 FAX: (974) 44451000
1 CT SCAN ROOM - EQUIPMENT ARRANGEMENT PLAN
MEQ-3130 SCALE: 1:30 PDC B with transport pallet 1120 900 2070 ≤450
Floor- and building vibrations LEAD CONSULTANT:
Fixpoints IRSmx4d tower with transport pallet 600 800 690 51.9
SOMATOM Drive - Equipment Legend
Floor- and building vibrations can reduce image quality! Therefore the gantry and the patient table must be
installed in an environment free of vibrations.
Fixpoint Components
MEINHARDT
IRSmx5d tower 645 175 450 25 14TH FLOOR AMWAL TOWER WESTBAY
Sources that produce vibrations are, e.g.: P.O. BOX 1362 DOHA, QATAR
Weight (kg), Heat dissipation to the air (W) Permissible acceleration for continuous vibration TEL: (974) 44117840 FAX: (974) 4411 7845
Railroad routes, subways, roads, road work and GR Gantry PHS 4n - with packing
construction sites, hospital power plants, mines, 2580 875 1570 700
Permissible acceleration rms-value [m/s²]
min. 1999
2 Covers must removed for mounting the transport device 2 SUBMISSION STAGE DATE
1
Smart Remote Services (SRS) Radiation protection
2100
2100
WORKING DRAWING
4 TIPPING HAZARD! 3
Smart Remote Services (SRS) is used for remote diagnostics as well DRAWN DESIGN CHECKED
1022
E
N
B07 B08
S
N
LA
.63
W
QIB
252
B14
B09 B05
8 9 P2.G M B02
B01
B03
B06
B12 B13
B11
+5.50 +5.50 B04
B10
+5.40 +5.40
CONCRETE SCREED
VINYL FLOOR FINISHED
5mm THK MS STEEL GENERAL NOTES
PLATE TRUNKING COVER 1. PROJECT REFERENCE ELEVATION ±0.00 IS AT +27.50 QATAR NATIONAL DATUM. ELEVATIONS
SHOWN ARE WITH RESPECT TO THE PROJECT REFERENCE ELEVATION OF ±0.00 AT GROUND FLOOR.
FLOOR CABLE 2. ALL DIMENSIONS ARE IN MM UNLESS NOTED OTHERWISE.
TRUNKING 3. EQUIPMENT DIMENSIONS ARE SUBJECT TO CHANGE BASED ON MATERIAL SUBMISSION APPROVAL.
4. CONTRACTOR TO COORDINATE AS PER SITE CONDITIONS WHEREVER APPLICABLE.
FFL: ±0.00 300
5. SHOP DRAWINGS ARE ISSUED IN COORDINATION WITH APPROVED DETAILED DESIGN FF&E AND
MEDICAL EQUIPMENT SELECTIONS. SOME AREAS ARE SUBJECT TO CHANGE AFTER MATERIAL
SUBMITTALS APPROVAL .
80
-0.10
KEY PLAN 1
10 10.1 17 17.1
26
+3.00
1 5 FLOOR TRUNKING DETAIL
A A
F Z2 F
1050
Z1-L0-IMD-025 POWER DISTRIBUTION
5160 2760 BOX BY HBK
P2.F P3.F
P2.1 P3.1
Fixpoints Z4 Z5
CT ROOM Fixpoint Components
AT
EMERGENCY STOP
SWITCH
Z1-L0-IMD-025 V V
GR Gantry
3000
3000
1000
1000
Operating console (controlbox, monitor, keyboard) 6 13 13.1 21
2.01 2.04
2.01 Image Control System (Tower PC)
REFERENCE DOCUMENTS
2480
1860 Power Distribution Cabinet A (incl. REPO contact) DOCUMENT NO TITLE REV
2.02
1950
1950
BUILDINGS 1 - ALL ZONES - ALL PARTS - OVERALL -GROUND FLOOR -
1800
1000
AT Emergency off button with locking mechanism
995
FLOOR TRUNKING
WALL OUTLET
CV On-off button with pilot lamp
@+300mm aFFL
300 1900
FFL: ±0.00 FFL: ±0.00
2.05 SW Radiation warning lamp MILITARY MEDICAL CITY PROJECT - AL SAILIYAH
Approval Status
Approved as
Radiation ready indicator
Commented
Re-Submit
A Approved
D Rejected
Revise &
CV CARE Vision Authority Signature Date
C
B
FK 200x80mm The system cable connections (power supply) between
FLOOR TRUNKING MMC-PMO
Gantry and PDC must have a minimum length of 7 m.
Place excess lengths in the cable duct.
C CT SCAN ROOM - ELEVATION C Cable diameter 30 mm, minimum bending radius 148 mm.
A CT SCAN ROOM - ELEVATION A MEQ-3131 SCALE: 1:30
MEQ-3131 SCALE: 1:30
The approval and comments shall not in any ways relieve or absolve the
Contractor from compliance of any obligations or liability under or in connection
with the contract, project requirements and specifications.
WORKING DRAWING REVISION NOTES
1 : AS PER COMMENTS FROM OPERATOR; DRY WALL SHIFTED TO 400mm TO GAVE SPACE ON THE CT SCAN MACHINE AND TO RECESS ON THE WALL THE SUB
EQUIPMENT (PDCa and PDCb) OF THE CT SCAN MACHINE..
9 8 P2.G M
QATAR ARMED FORCES
P2.G MILITARY MEDICAL CITY HOSPITAL PROJECT
MANAGEMENT OFFICE P.O. BOX 2075 DOHA,
QATAR
+5.50 +5.50 +5.50
+5.40 +5.40 +5.40 MAIN CONTRACTOR:
LEAD CONSULTANT:
MEINHARDT
14TH FLOOR AMWAL TOWER WESTBAY
P.O. BOX 1362 DOHA, QATAR
TEL: (974) 44117840 FAX: (974) 4411 7845
+3.00
2486 4993
Z1-L0-IMD-025 ENGINEERING, PROCUREMENT AND
1000
1050
SW
POWER DISTRIBUTION
BOX BY HBK
WORK / CONTROL POWER DISTRIBUTION EMERGENCY STOP
AT SWITCH
EMERGENCY STOP AT CONSTRUCTION (EPC/TURNKEY) FOR
BOX BY HBK SWITCH
PDC A CABINET
WITH UPS
PDC B
CABINET
EMERGENCY STOP
SWITCH
AT SBA
Z1-L0-IMD-023 MILITARY MEDICAL CITY HOSPITAL
CT ROOM (MMC) AT TAREK BIN ZEYAD CAMP
Z1-L0-IMD-025 AT IN SAILIYAH (SOUTH OF DOHA)
3000
1000
1000
50mmØ OPEN PVC PIPE PROJECT NO. AA-103-01
PROVISION FOR INJECTOR
2.08 THRU CONTROL ROOM EAT
ISO CENTER DRAWING TITLE:
2.03 2.04 1860 2480
2.02 BUILDING 01-ZONE 01, -PART 06 - GROUND FLOOR -
2050
1950
1800
1800
1800
1000
1000
COMPLIANCE WITH THE CONTRACT DOCUMENTS. ALL DIMENSIONS AND FIELD CONDITIONS
130 130
560
500
300500
2.05
300
STATUS:
DESIGNED SCALE A1
MEINHARDT 1:30
MMC-HBK-P1-WD-B01-MEQ-L0-P06-3131 0
E
N
B07 B08
S
N
LA
.63
W
QIB
252
8 9 I-Control with wall bracket (schematic) not to scale
SOMATOM Drive - Equipment Legend B14
B09 B05
B02 B03
4 Weight (kg), Heat dissipation to the air (W) B01
400
Pos. Description kg W Remark
B06
B12 B13
B11
1.01 Gantry 2580 1000 #1, isocenter 1020 mm B04
B10
1.02 Patient table PHS 4n (2000 mm scannable range) 500 * * included in 1.01
130 1.03 PDC A cabinet with UPS 610 2000
15
10 3
1.04 PDC B cabinet 400 2000
1.05 IRSmx5d Tower PC (Image Reconstruction System) 25 500
00
0
x. 1
1.06 Control unit with monitor, control box, keyboard 10 100
119
90
2
ma
9 4 1.07 ICS Tower 13 PC (syngo Acquisition Workplace) 25 500 GENERAL NOTES
1300
1. PROJECT REFERENCE ELEVATION ±0.00 IS AT +27.50 QATAR NATIONAL DATUM. ELEVATIONS
1 1.08 Power Distributor by customer, not in scope SHOWN ARE WITH RESPECT TO THE PROJECT REFERENCE ELEVATION OF ±0.00 AT GROUND FLOOR.
of delivery 2. ALL DIMENSIONS ARE IN MM UNLESS NOTED OTHERWISE.
6
1531
1530
3. EQUIPMENT DIMENSIONS ARE SUBJECT TO CHANGE BASED ON MATERIAL SUBMISSION APPROVAL.
1 4. CONTRACTOR TO COORDINATE AS PER SITE CONDITIONS WHEREVER APPLICABLE.
800 - 900
7 159 5. SHOP DRAWINGS ARE ISSUED IN COORDINATION WITH APPROVED DETAILED DESIGN FF&E AND
5 MEDICAL EQUIPMENT SELECTIONS. SOME AREAS ARE SUBJECT TO CHANGE AFTER MATERIAL
SUBMITTALS APPROVAL .
2663 8
1400 1263 1910
Floor- and building vibrations
Floor- and building vibrations can reduce image quality! Therefore the gantry and the patient table must be
3347 installed in an environment free of vibrations.
563
1 Wall bracket (thickness 35 mm) with wall mounting holes Ø 6.6 mm 2 Locking lever for locking/unlocking
KEY PLAN 1
10 10.1 17 17.1
26
Z2
P3.1
1790
F F
Input Output
1900
any other sources of striking vibrations.
787
- Voltage: 100 V AC - 240 V AC - Voltage: +24 V DC ± 5 % P2.A P3.A
- Line frequency: 50-60 Hz - Current: 1.25 A The CT system is not sensitive to common vibrations. If
- Current: 0.7 A - Cable length: 1.5 m
- Stand-by: 0.5 W the CT is away from vibrational sources, or the CT is P2.F P3.F
WORK / replacing a CT system that to date has not shown image P2.1 P3.1
728
6 7 Z4 Z5
quality problems due to vibrations, it is usually not
CONTROL
549
550 8 IVP control cable 9 Strain relief for power supply cable 10 Connection for power supply cable
necessary to execute vibrational measurements.
Z1-L0-IMD-023 V V
If there are any doubts, the following thresholds have to be
105
Delivery volume: - Control unit and a wall bracket for wall mounting (standard installation)
verified by measurement:
cabling
- Drilling template
1 10 100
- Installation material: 3x Fischer dowels 6 13 13.1 21
water
(Type: UX 8x50; nominal drill hole diameter: 8 mm; Drill hole depth: min. 60 mm
- 3x screws M6x80, 3x screw head caps grey REFERENCE DOCUMENTS
Patient Handling System PHS) must not exceed the
thresholds as described in the above shown diagram. DOCUMENT NO TITLE REV
1990
Z1-L0-IMD-025 4 3
The floor construction has to be performed solid and free of
vibration, e.g. concrete flooring C20/25 to C50/60 corresponding to
It is the customer's responsibility to contract a qualified specialist. The specialist must implement site
Center of DIN EN 206-1, according to the maximum values as specified in the
modifications to meet the specific limits, and to design structural solutions in case of deviations.
gravity textblock “Floor and building vibrations”.
256
Concrete fill-up, concrete class It is recommended to test the weight capacity of the concrete or MILITARY MEDICAL CITY PROJECT - AL SAILIYAH
1 2
C20/25 to C50/60 composite flooring by a stress analyst.
Fastening the gantry on the floor is possible but only necessary in
Approval Status
Approved as
Commented
Join with the solid concrete.
Re-Submit
A Approved
94 countries prone to earthquakes or according to local regulations.
D Rejected
Revise &
Securing the patient table to the floor is mandatory.
Authority Signature Date
PDC - Power Distribution Cabinet not to scale
C
B
Total weight of Gantry: 2580 kg : F stat total 25 kN
MMC-PMO
FRONT VIEW ILLUSTRATION for the cable inlet (Isometry)
Partial load on gantry foot 1 2 3 4
M
Nominal static load after levelling:
5.7 7.6 7.9 6.7
F stat nom [kN]
During gantry installation and leveling, the maximum possible load on one gantry foot can be 14 kN (the gantry is
1 CT SCAN ROOM - CONCRETE FILL-UP PLAN standing on two diagonal feet). Design access floors for a weight capacity of min. 400 kg per slab/plate. During
1850
1950
gantry transport, the load may be higher at certain individual points (3-point load, e.g. due to uneven flooring).
MEQ-3132 SCALE: 1:30 00 ISSUED FOR CONSTRUCTION 31/10/2022 ABAM ABAM PR SG
REVISION DETAILS DATE DWN DES CHD APD
EMPLOYER
1 Adjustable gantry feet (Ø 80 mm); mounting holes (Ø 17 mm) ICS Tower 13 not to scale
MAIN CONTRACTOR:
100
2
Adjustable patient table feet with mounting holes (inner: Ø 20.5 mm); minimum extraction force: 2.03 kN at the two mounting points bearing
the greatest load at maximum patient travel path and a 230-kg load. According to IEC 60601-1, it is mandatory that the 4-fold safety factor be observed. 840 488
FRONT VIEW SIDE VIEW
7
TOP VIEW
2
570 0
HBK CONTRACTING CO. WLL
50 43
900 P.O. BOX 1362 DOHA, QATAR
2 Alternative floor mounting holes (Ø 18 mm), if position 2a is not usable 1252 4
2 TEL: (974) 44451111 FAX: (974) 44451000
120
1070 5 2 6
LEAD CONSULTANT:
3 Floor-inlet for cable/hose connection: 49/97 x 488 mm
890 TOP VIEW ILLUSTRATION for floor mounting
3
MEINHARDT
420
175 420 (top view of pedestal)
4 No floor cut-out permitted in this defined area (Ø 190 mm) around each adjustable gantry foot 1435 900 500 900 14TH FLOOR AMWAL TOWER WESTBAY
8 P.O. BOX 1362 DOHA, QATAR
Alternative to 3 : 689.5 745.5
40
TEL: (974) 44117840 FAX: (974) 4411 7845
5 Core hole: max. Ø 100 mm for HV cables and other system cabling 5 6 3
395
2 1
3 1 ARCHITECT & INTERIOR DESIGNER:
8
6 Core hole: max. Ø 80 mm for system cabling
10 3 3
1
691.5
650
570
7 Core hole: max. Ø 160 mm for water hoses 1 1 4
25
8 Gantry footprint 4
1 Air intake 2 Air exhaust (primary) 3 Air exhaust (secondary) 3 3 14 MEDICAL PLANNER: MEDICAL EQUIPMENT : MEP SUBCONTRACTOR : SUBCONTRACTOR :
7 Ø
4 Air must not exhaust directly to the air intake 5 e.g. a wall
9 "Electronics cover" for cabling between gantry and patient table (in scope of delivery)
100
40
6 Minimum distance for cable routing and air circulation 7 Minimum distance for air circulation
100 700 100
9 7
Floor requirements: Gantry and patient table orientation point
PROJECT TITLE:
CT
The floor construction has to be performed solid and free of vibration, e.g. concrete flooring C20/25 to C50/60 scan plane
950
corresponding to DIN EN 206-1, according to the maximum values as specified in the textblock
ILLUSTRATION for the cable inlet
(top view of pedestal)
ENGINEERING, PROCUREMENT AND
service area
“Floor and building vibrations”. Anti-static flooring is recommended for the examination room. ESD flooring is also suitable. 1 1 900 CONSTRUCTION (EPC/TURNKEY) FOR
NOTE: A solid and vibration-free floor is needed to avoid a negative influence on the vibration characteristic of the gantry. IRSmx5 tower not to scale
An incorrect vibration characteristic of the gantry can cause problems with balancing and image quality.
MILITARY MEDICAL CITY HOSPITAL
6 (MMC) AT TAREK BIN ZEYAD CAMP
max.120
Permissible tolerance for floor levelness is at maximum 10 mm, based on the area of the adjustable feet of the
4
gantry and the patient table. 9 175
SIDE VIEW of socle IN SAILIYAH (SOUTH OF DOHA)
orientation point
650
25
On-site subconstruction have to fulfill the flooring conditions. It is recommended to test the weight capacity of 5
PHS 4n
82.5
175 120
the concrete or composite flooring by a stress analyst. 645 PROJECT NO. AA-103-01
Floor mounting, as well as principally every kind of CT system component mounting, must be performed in 2 2 1 4 3 DRAWING TITLE:
accordance with country-specific guidelines in countries prone to earthquakes.The gantry and the patient
645
table have to be on the same (isocentric) plane.
55 790 55 BUILDING 01-ZONE 01, -PART 06 - GROUND FLOOR -
450
450
2a 2a CT SCAN ROOM (Z1-L0-IMD-025) - SOMATOM Drive
An "Installation Template" with footprint information for the gantry and patient table is optional available. 9 650 32.5
2 PRE-INSTALLATION CONCRETE FILL-UP PLAN
May not be used as a drilling template. A drilling template is not necessary. 7
150
205 5 THIS DRAWING HAS BEEN CHECKED FOR COMPLETENESS, ACCURACY, COORDINATION, AND
COMPLIANCE WITH THE CONTRACT DOCUMENTS. ALL DIMENSIONS AND FIELD CONDITIONS
Floor mounting Gantry 1 System cables HAVE BEEN VERIFIED.
- Possible through four mounting holes, but not required in principle (only in areas prone to earthquakes). FRONT VIEW SIDE VIEW TOP VIEW
2a 2a Cable inlets (cut-outs in the base) when cables are routed on the floor; if not used, the cut-outs are covered
Mounting materials must be provided on-site. 2
with dummy plates; feed lines are possible from all four sides
1 Front side with grid for air intake (cooling) 2 Support feet 3 Air intake 4 Air exhaust
- Leveling tolerance: in Z-direction ± 0.2 deg, in X-direction ± 0.5 deg Mr. ALAADDIN AKTAN 20 Sept 2022
2 2 3 Floor mounting holes (all Ø=14 mm) 4 PDC base (footprint) CONTRACTOR'S REPRESENTATIVE: DATE:
- Leveling with four adjustable feet. 5 Minimum distance for air circulation 6 Air must not exhaust directly to the air intake
STATUS:
Ø 5 Cable inlet in PDC base; opening is adjustable 6 Air exhaust for system cooling; must not be covered
7 Sliding door or similar ISSUED FOR CONSTRUCTION
For example, the sliding door of a wall For example, a wall; no safety distance for
Floor mounting patient table 280 The safety distances for air circulation (system cooling) must be maintained. Openings for air circulation in the 7 closet, and so on 8
air circulation (cooling) required SUBMISSION STAGE DATE
393 tower housing must not be covered. Air must not exhaust directly into the air intake (a closed air loop).
- The patient table MUST be secured to the floor (4 mounting holes). 9 Safety distance for air circulation (cooling) required 10 Top exhaust for system cooling; must not be covered 17/10/2022
WORKING DRAWING
- Installation material for floor mounting in concrete is in scope of delivery. Therefor min.140 mm thickness of concrete is required:
Anchor bolts type 1: 4x HSL-4G M 10/60 heavy-duty sleeve (used without bonded screed) or DRAWN DESIGN CHECKED
Anchor bolts type 2: 4x HSL-4G M 10/120 heavy-duty sleeve (used with up to 60 mm thick bonded screed) PRABHU
ABAM ABAM
- Leveling with four adjustable feet.
- Remove the floor cover in the entire footprint area of the table base.The support plate is equal to the footprint DESIGNED SCALE A1
of the patient table base and is used as a help to cut out the floor covering. MEINHARDT 1:30
MMC-HBK-P1-WD-B01-MEQ-L0-P06-3132 0
E
N
B07 B08
8 9
S
AGENT Line impedance related to measured line voltage
N
LA
.63
W
QIB
252
B14
14.04 m²
B06
B12
CHANGE
B13
300 B11
(M) 250
Z1-L0-IMD-026 7.08 m² 250
The maximum allowed line impedance
depends on the actual measured lowest line
4.77 m² 200
voltage.
SW We recommend monitoring the voltage over
TS GENERAL NOTES
150 a period of 24 hours preferable on work days 1. PROJECT REFERENCE ELEVATION ±0.00 IS AT +27.50 QATAR NATIONAL DATUM. ELEVATIONS
to specify the minimum. 2.
SHOWN ARE WITH RESPECT TO THE PROJECT REFERENCE ELEVATION OF ±0.00 AT GROUND FLOOR.
ALL DIMENSIONS ARE IN MM UNLESS NOTED OTHERWISE.
3. EQUIPMENT DIMENSIONS ARE SUBJECT TO CHANGE BASED ON MATERIAL SUBMISSION APPROVAL.
100 85 4. CONTRACTOR TO COORDINATE AS PER SITE CONDITIONS WHEREVER APPLICABLE.
5. SHOP DRAWINGS ARE ISSUED IN COORDINATION WITH APPROVED DETAILED DESIGN FF&E AND
MEDICAL EQUIPMENT SELECTIONS. SOME AREAS ARE SUBJECT TO CHANGE AFTER MATERIAL
P2.G SUBMITTALS APPROVAL .
50
0 1
SBA
CT ROOM 360 380 400 420 440
KEY PLAN
+1600mm aFFL EAT AT +1800mm aFFL Z1-L0-IMD-025 1
10 10.1 17 17.1
26
2481
A A
+1800mm aFFL AT
P2.1
Z1 Z3 P3.1
CONCRETE SCREED F Z2 F
300
Z4 Z5
D GR Gantry FFL: ±0.00
3131
Operating console (controlbox, monitor, keyboard) V V
ICS
80
Image Control System (Tower PC) 6 13 13.1 21
-0.10
WORK / Power Distribution Cabinet A (incl. REPO contact)
GR REFERENCE DOCUMENTS
CONTROL
50mmØ OPEN PVC PIPE Power Distribution Cabinet B DOCUMENT NO TITLE REV
Z1-L0-IMD-023 PROVISION FOR INJECTOR 1911
MMC-HBK-AP-DD-BAP-MEQ-L0-APO-1800
BUILDINGS 1 - ALL ZONES - ALL PARTS - OVERALL -GROUND FLOOR -
2
MEDICAL EQUIPMENT PLAN
*2 THRU CONTROL ROOM DRAIN PIPE FOR IRSmx (Image Reconstruction System Cabinet)
cabling
MMC-HBK-P1-DD-B01-MEQ-L0-P06-1806 BUILDING 1 - ZONE 01- PART 06- GROUND FLOOR MEDICAL EQUIPMENT PLAN 2
CONDENSATION OF THE 3 SECTION DETAIL
water
OPC MACHINE BY HBK AT Emergency off button with locking mechanism MEQ-3133 SCALE: 1:5
2615
IK 200x60mm OK FFL
FK 300x80mm
On-off button with pilot lamp
to @3000mm aFFL
AT +1800mm aFFL 2 MILITARY MEDICAL CITY PROJECT - AL SAILIYAH
1775
WATER COOLING SW Radiation warning lamp
3 Approval Status
Approved as
Commented
3133
Re-Submit
A Approved
Radiation ready indicator
1300
D Rejected
Revise &
6050 FK 300x80
135
Authority Signature Date
C
B
SW 1204
2529 3321 MMC-PMO
85
above The system cable connections (power supply) between
door Cut-out in cable duct
80
FK 200x80mm Gantry and PDC must have a minimum length of 7 m.
Place excess lengths in the cable duct.
300
280
WD *3
h= 450mm
IK 200x60mm Cable diameter 30 mm, minimum bending radius 148 mm.
849
305
OK FFL to *2
845
The approval and comments shall not in any ways relieve or absolve the
@300mm AFFL 485 Contractor from compliance of any obligations or liability under or in connection
with the contract, project requirements and specifications.
549
M
Gantry contour
CV IRS PDCA PDCB Notes
IK 200x60mm
OK FFL to UK PDB 150x300 300x300 300x300 For service purposes; only required if there is no existing
WALL OUTLET
300mm AFFL *3 socket or none is planned by the electrician
00 ISSUED FOR CONSTRUCTION 31/10/2022 ABAM ABAM PR SG
(consult the Siemens Healthineers project management) Scan plane REVISION DETAILS DATE DWN DES CHD APD
M.1 EMPLOYER
*4 Grounded socket for I-control unit.
QATAR ARMED FORCES
2 DETAIL - A MILITARY MEDICAL CITY HOSPITAL PROJECT
MANAGEMENT OFFICE P.O. BOX 2075 DOHA,
MEQ-3133 SCALE: NTS QATAR
2 : ADJUSTMENT ON FLOOR TRUNKING AS PER THE NEW DRY WALL LOCATION Cable set Water hose set
cut-out, Ø in mm (i.e. in suspended ceiling) LEAD CONSULTANT:
* 10 m 10 m
6
16 □
Cu
1.4 13 13 13 21 T1 T2 3VA9988-0AA12 CONSTRUCTION (EPC/TURNKEY) FOR
K2 K3 K1 K1 80% 20% Side mounted motor per Circuit breaker 3VA 24V DC grounded socket, switchable, installation according to national guidelines
PA 1.1 14 14 14 22
10s 1s 7
3VA9117-0HB10 intern 1m 2
max. 2,5 mm² flexible
2.4 MILITARY MEDICAL CITY HOSPITAL
1 K K
Undervoltage release for MCCB per Circuit breaker Siemens Drain pipe provision 25-50mm Ø
6 33 K4 9 10
8
3VA9908-0BB11
max.
250 VAC / 2 A
24 VDC / 2 A
2.3
T2 K1
~
24 23
34 15 44 2.2 network socket
2
5 18 K6 10 Timing relay Siemens 3RP2505-1AB30-0AC1 intern 1m
max. 2,5 mm² flexible
T1
4
1.1 1.4
12 Contactor relay 24V DC 3RH2131-2BB40
max.
250 VAC / 2 A
2.1
UK bottom edge OK top edge PROJECT NO. AA-103-01
15 AUX 3 VA 19 24 VDC / 2 A
DRAWING TITLE:
6A - Contactor relay 24V DC 3RH2131-2BB40 bottom edge suspended ceiling OK FFL top edge finished floor
13 UK ABGD
4 3 B1 F1 E1 K6 intern 1m 2 1.6 BUILDING 01-ZONE 01, -PART 06 - GROUND FLOOR -
5 Priorität 1 20 14a Impulse switch Eltako ES 12Z-200-UC
max. 2,5 mm² flexible
Q3 Q2 Q1 MODE 1
M 3VA <U T1 T2 K2 K3 ZE + ZA K4 K1 + max. 1.5
UK RHD
bottom edge bare ceiling OK RFB top edge unfinished floor CT SCAN ROOM (Z1-L0-IMD-025) - SOMATOM Drive
- 250 VAC / 2 A
4 32A 4 125A 4 125A
7
7
8 9 10 12 13
A2 E2
16 16 Contactor relay 24V DC 3RH2131-2BB40 24 VDC / 2 A PRE-INSTALLATION ELECTRICAL PLAN
RHD bare ceiling RFB unfinished floor level
3a 3f 3f - 13 THIS DRAWING HAS BEEN CHECKED FOR COMPLETENESS, ACCURACY, COORDINATION, AND
K8 19 Auxiliary switch changeover contacts Siemens 3VA9988-0AA12 Power Distribution Cabinet (PDC) ABGD suspended ceiling FFL finished floor level COMPLIANCE WITH THE CONTRACT DOCUMENTS. ALL DIMENSIONS AND FIELD CONDITIONS
3R 3R K6 HAVE BEEN VERIFIED.
+ EQUIPMENT
K8 23 14 20 Contactor relay 24VDC 3RT1017-1BB44-3MA0 (K6) aABGD above suspended ceiling AFFL above finished floor level
ROOM
X20 5.2
X20 2.3
X20 2.4
E
N
B07 B08
S
14.04 m²
N
LA
.63
W
Connection of cooling system (schematic)
QIB
252
not to scale
CHANGE Scope of delivery - Cooling system B14
Z1-L0-IMD-024 1. Standard:
B09
B02 B03
B05
B01
- Flexible water hoses (Ø 34 mm) with thermal insulation (Ø 58 mm); available lenghts 10, 20 and 30 m;
Z1-L0-IMD-026 7.08 m² min. bending radius: 150 mm, shorten excess cable lengths; preordering is possible
B06
B12 B13
- Thermal insulation of water hoses: B11
B10
• Length delivered depends on water hose length ordered.
4.77 m²
B04
Recommended: • Suitable for cable ducts up to 60 mm height and on-site water temperature in the range of ≥ 8° C to 16° C.
SW Please note:
TS Shut-off valve necessary. In this case the outer diameter of the thermal insulation won't fit the standard cable
duct heights of 60 mm.
- Accessories: contains all required parts for the water connection:
Pressure regulator • 2x flexible fabric hoses for water connection (Gantry side) with preinstalled shut-off valves
• Adapters for M36x2 metric to G1" inside thread (see also "Water connection pieces with shut-off valve") GENERAL NOTES
1. PROJECT REFERENCE ELEVATION ±0.00 IS AT +27.50 QATAR NATIONAL DATUM. ELEVATIONS
- Water drain for condensation (produced from the water-cooled gantry during system cooling): SHOWN ARE WITH RESPECT TO THE PROJECT REFERENCE ELEVATION OF ±0.00 AT GROUND FLOOR.
p Diff. pressure gauge • Length 10 m; (outer diameter Ø 10 mm; inner diameter Ø 6 mm) 2. ALL DIMENSIONS ARE IN MM UNLESS NOTED OTHERWISE.
• Min. bending radius: 30 mm 3. EQUIPMENT DIMENSIONS ARE SUBJECT TO CHANGE BASED ON MATERIAL SUBMISSION APPROVAL.
4. CONTRACTOR TO COORDINATE AS PER SITE CONDITIONS WHEREVER APPLICABLE.
• Shorten excess cable length to size.
5. SHOP DRAWINGS ARE ISSUED IN COORDINATION WITH APPROVED DETAILED DESIGN FF&E AND
TI Thermometer • Extension up to 20 m possible; extension material must be provided on-side. MEDICAL EQUIPMENT SELECTIONS. SOME AREAS ARE SUBJECT TO CHANGE AFTER MATERIAL
TI TI • Minimum flow rate: 2 l/h at 75% relative humidity SUBMITTALS APPROVAL .
SBA
CT ROOM On-site 2. Option
cooling Feed water flow meter - Fixed piping kit:
EAT AT Z1-L0-IMD-025 p p water supply • includes the installation material to connect the delivered flexible cooling hoses with on-site fixed piping
• 4x connection pieces (DKOL DN25 VA)
2481
AT • 4x adaptors M36x2 metric to G1" internal thread
KEY PLAN 1
10 10.1 17 17.1
26
A A
2760 F Z2 F
Water connection pieces with shut-off valve not to scale P2.A P3.A
on-site
(in scope of delivery)
P2.F P3.F
Z4 Z5
P3.1
WORK /
150
2 V V
CONTROL GR
Z1-L0-IMD-023 1911 Adapter fitting for hose G1" internal thread
1 6 13 13.1 21
in scope of delivery
*2 DRAIN PIPE FOR
cabling
CONDENSATION OF THE m REFERENCE DOCUMENTS
in.
water
OPC MACHINE BY HBK 80 DOCUMENT NO TITLE REV
2615
534
3 MMC-HBK-AP-DD-BAP-MEQ-L0-APO-1800
BUILDINGS 1 - ALL ZONES - ALL PARTS - OVERALL -GROUND FLOOR -
MEDICAL EQUIPMENT PLAN
2
FK 300x80mm
MMC-HBK-P1-DD-B01-MEQ-L0-P06-1806 BUILDING 1 - ZONE 01- PART 06- GROUND FLOOR MEDICAL EQUIPMENT PLAN 2
4
AT
1775
WATER COOLING The cooling water hoses Ø 34 mm
(with insulation Ø 58 mm) from the gantry
1300
Approved as
Commented
1
Re-Submit
A Approved
D Rejected
Revise &
FK 200x80mm Condensed water hose to Authority Signature Date
557
e.g. a sink or a siphon 1 Hexagon coupling nut 2 Shut-off valve 3 Flexible fabric hose 4
C
B
280
WD
h= 450mm
MMC-PMO
Suitable for installation in surface mounted trunking, cable duct or false floor.
845
GR
6
1 DRAIN and COOLING WATER LOCATION 0 ISSUED FOR CONSTRUCTION
REVISION DETAILS
31/10/2022
DATE
ABAM ABAM
DWN DES
PR
CHD
SG
APD
MEQ-3134 SCALE: 1:30
1 EMPLOYER
max.
QATAR
4m
Utilisation of the CT system [%] 100 75 50 25 0 7 4 5 MAIN CONTRACTOR:
(Max. power) (stand-by)
Technical data of the interface heat exchanger 3
HBK CONTRACTING CO. WLL
Heat dissipation to cooling water [kW] 17 14 11 8 5
Primary circuit components: Shut-off devices, filters, flow indicators, fill and drain valves, plate heat P.O. BOX 1362 DOHA, QATAR
Heat dissipation (Gantry and patient table) to air [kW] 1 1 1 1 1 exchanger, pressure gauge 1 E.g. a sink 2 Water drain; flow rate capability: min. 20 l/h 3 E.g. cable duct TEL: (974) 44451111 FAX: (974) 44451000
Operating pressure nominal: 2 bar to 6 bar, max. 10 bar Flow rate: 4 m³/h 6 Gantry rear view 7 Max. installation height (for hose routing; water drain connection etc.)
NOTE: To guarantee the limit of the maximum water pressure of 10 bar, Pressure loss: 0.5 bar MEINHARDT
the on-site cooling water system must provide a suitable safety device, e. Operating pressure: max. 10 bar 14TH FLOOR AMWAL TOWER WESTBAY
P.O. BOX 1362 DOHA, QATAR
TEL: (974) 44117840 FAX: (974) 4411 7845
g. a pressure relief valve.
Secondary circuit components: Fill and drain valve, shut-off device, diaphragm-type expansion vessel, GR Gantry - connection water, power and datalines
Water quality Potable water quality with particle size ≤ 0,25mm safety valve, circulating pump with fill and drain valve, permanent bleed, ARCHITECT & INTERIOR DESIGNER:
Anti-freeze may be added to the on-site circuit. Max. mixing ratio up to -25°C: 40 % Connections: M36x2 / DN25
Adding anti-freeze reduces the cooling output and requires higher cooling water flow rate. Heat up brine to at least Internal pressure loss: ca. 0.5 bar 3 MEDICAL PLANNER: MEDICAL EQUIPMENT : MEP SUBCONTRACTOR : SUBCONTRACTOR :
+4 °C if the internal cooling system is connected to an existing on-site cooling system! Only then on-site water can Operating voltage: 3/N/PE ~400 VAC / ±10% - 50 Hz / ±1%
flow through the Gantry integrated cooling system. 3/N/PE ~460 VAC / ±10% - 60 Hz / ±1% 1 6 7
Switching off when the CT system is not in operation Power consumption, overall system: max. 0.6 kVA at 50 Hz
For purposes of saving energy the on-site water circulation can be switched off with: max. 0.9 kVA at 60 Hz 2 4 5 PROJECT TITLE:
- On-site installation of a solenoid valve or switch-off feed pump via contactor Fuse protection (provided by client): 10 A
- On-site feed pump 1 Water hose 2 M36 x 2 external thread 3 Adapter (in scope of delivery) ENGINEERING, PROCUREMENT AND
- Optional water/air-split cooling system
General information 4 G1" internal thread 5 G1" external thread 6 On-site water connection 7 e. g. wall CONSTRUCTION (EPC/TURNKEY) FOR
A floating relay contact assembly is included in the PDC:
Maximum 277 VAC/2 A; utilization category AC-14/15 For the gantry-integrated cooling system a on-site cooling water system has to be provided. MILITARY MEDICAL CITY HOSPITAL
Maximum 24 VDC/2 A; utilization category DC-12/13 The cooling water system will be connected via:
- Flexible water hoses (in scope of delivery, orderable lengths 10, 20 or 30 m) and special water connection pieces.
(MMC) AT TAREK BIN ZEYAD CAMP
Minimum flow rate - On-site cooling circuit Piping material
- On-site fixed pipes (PVC, copper, high-grade steel). Therefore a "Fixed piping kit" can be ordered. For vibration
Recommended Not recommended
IN SAILIYAH (SOUTH OF DOHA)
reduction the flexible water connection pieces have to be installed between the fixed piping and the water
Barometric pressure 700 to 1060 hPa
Pressure loss - On-site cooling circuit 2 connection points of the gantry. The direct connection of fixed piping and gantry water connection points is not
Stainless steel (V2A; V4A) Aluminium
permitted!
Non-ferrous metal (e.g. copper, brass) Iron, carbon steel
PROJECT NO. AA-103-01
Barometric pressure 700 to 1060 hPa 3500 l/h For service and filling procedure a water faucet with screw connection and a water drain (e.g. a sink or a siphon) in
2 3 Synthetic materials, plastics Zinc-plated steel
DRAWING TITLE:
4 3200 vicinity of the gantry is necessary.
3 3000 BUILDING 01-ZONE 01, -PART 06 - GROUND FLOOR -
3000 l/h Brazing solder, hard solder Zinc
2.2 bar
CT SCAN ROOM (Z1-L0-IMD-025) - SOMATOM Drive
2.0 bar Fitting solder types 3 and 4 Standard steel pipes PRE-INSTALLATION ELECTRICAL PLAN
2.05
1.8 bar 2050 THIS DRAWING HAS BEEN CHECKED FOR COMPLETENESS, ACCURACY, COORDINATION, AND
2000 l/h COMPLIANCE WITH THE CONTRACT DOCUMENTS. ALL DIMENSIONS AND FIELD CONDITIONS
HAVE BEEN VERIFIED.
1.6 bar 1800
1.45 1600
1.4 bar
1120
1300 1300 Legend
1.2 bar 970 1030 Mr. ALAADDIN AKTAN 20 Sept 2022
1000 l/h CONTRACTOR'S REPRESENTATIVE: DATE:
0.95 740 850
1.0 bar 650 cut-out, LxW (i. e. in FK or raised floor) STATUS:
1500 l / h
500 l / h
3000 l / h
2000 l / h
2500 l / h
DESIGNED SCALE A1
MEINHARDT 1:25
!"
!"
#ÿ$%
ÿ )99 >)
811711 7177
01231245ÿ72893ÿ
!"ÿ#$"ÿ%&
'ÿÿÿ$ÿÿÿ()'ÿ*) +ÿ#,, ÿ-.7/01/2 3/1274ÿ/ÿ*, ', ÿ3$ÿ#5/2+*' 00 '2,+ÿ/ÿ,#62+7
3# 2+7*(ÿ"!"ÿ"ÿÿ!"8ÿÿ8
9:;<ÿ>?@A<B;>>@CÿD@<ÿ@E>FB@>;G@CCHÿIJAJ?@>JKL
811711 2177
01231245ÿ72893ÿ
7*ÿÿ+%),ÿÿ
ÿÿ)%
+%%ÿÿ%ÿ-ÿ !
. ÿÿÿÿ
%ÿ-ÿ ÿ. ÿ-
)%ÿ2ÿÿ
ÿ%ÿ/%ÿ+,ÿ)
(ÿ420ÿ%ÿ1 2
*%ÿ)ÿÿ(
"1ÿ. ÿ%ÿÿ3%ÿ
!"ÿ#ÿ$% &ÿ! '()ÿ)ÿ" *%ÿ/,,ÿ)ÿ
%ÿ-ÿ ÿ. ÿ-
)%ÿ4ÿ.45ÿ. ÿ5
1.ÿ%ÿÿÿ%ÿ-
ÿÿ-ÿ)%ÿ9
1ÿ%%,ÿ-
)%ÿ+%ÿÿ
,ÿ)ÿ+,
+%),ÿ6ÿ*ÿ%
3%ÿ1%,ÿÿ&"
ÿ%%,ÿ)
ÿ%ÿ-ÿ ÿ
7ÿÿÿ
ÿ%ÿ)ÿ
ÿÿÿ%
!"ÿ#ÿ*"11$7& ÿ8%% '()ÿ)ÿ" %(ÿ+,ÿ2ÿ1
.%ÿÿ%ÿ-
9ÿÿ%ÿ,
ÿÿÿ
7ÿ'-%ÿÿ%
ÿÿ
%9ÿ-ÿ2ÿ
%ÿÿ5ÿ%
ÿ#ÿ%(ÿ) )%ÿÿÿ ÿ
!"ÿ#ÿ% ÿ'%: " 4ÿ$;ÿ
%9ÿÿ%ÿ
),ÿ9ÿ(ÿ%,ÿÿ)
ÿÿ)ÿ,ÿ6
< <"ÿÿÿ-ÿ=
* ÿ ÿ5ÿ24"ÿ
"ÿ#ÿ.% .ÿ.%% ÿ#ÿ%(ÿ) 7#ÿ"%%ÿÿ%
" -%ÿÿ $ ÿ-%
%;%ÿÿ)%
(9ÿ2#
"%ÿ%;%
)ÿ--%
811711 4177
01231245ÿ72893ÿ
ÿ !ÿ
"ÿ#ÿ#ÿ
ÿ$#ÿ$ÿ
ÿ4%ÿ&ÿ
ÿ$ÿ
ÿ 'ÿ
(ÿÿ
!#ÿ
&ÿ%ÿ)! )ÿ *" ÿÿ&
811711 177
01231245ÿ72893ÿ
7ÿ(0ÿ,
!"#$!7!%&!#'7! "#$ ÿ!ÿ +ÿ,-- +./ÿ/ÿ ÿÿÿÿÿ,ÿ/
ÿ
()!*'!'!4723 1
ÿ,ÿÿ
7!ÿÿ/,ÿ
/ÿÿÿ.ÿ
3ÿ1ÿÿ1,
/ÿ# ÿ2!ÿ+ÿ3,
,1ÿÿ/ÿÿ
ÿ1,ÿ/ÿ# ÿ.
ÿÿ4717'12'22ÿ4!
"#$ ÿ!ÿ ÿ2/ +./ÿ/ÿ .,,.ÿ1
1
ÿÿ
3,,ÿÿ
,ÿ1ÿ9!
ÿ/,ÿ
ÿ,ÿ/,
ÿ,0ÿ-
# ÿ/,ÿÿ,ÿ
,.ÿ1
75ÿÿ&,/1ÿÿ
ÿÿ/,
&,,ÿÿ,ÿ3ÿ"#$
6 ÿ#ÿÿÿ
,ÿ3ÿ# ÿ6 ÿ3
/,ÿ2ÿÿ
ÿ,ÿ7,ÿ&1ÿ/
.ÿ42'ÿ,ÿ8 9
5,ÿ/ÿÿ.
8ÿ6 ÿ,ÿÿ:,ÿ
"#$ ÿ!ÿ(, 4ÿ$ +./ÿ/ÿ 5,ÿ711ÿ/ÿ
,ÿ3ÿ# ÿ6 ÿ3
/,ÿ4ÿ6%5ÿ6# ÿ;
86ÿ,ÿÿÿ,ÿ3
# ÿÿ3ÿ/,ÿ9
8ÿ,,1ÿ3
/,ÿ&,ÿÿ
1ÿ/ÿ&1
&,/1ÿ<ÿ5ÿ,
:,ÿ8,1ÿÿ4
ÿ,,1ÿ/
ÿ,ÿ3ÿ# ÿ
"#$ ÿ!ÿ581(*4 ÿ2,, +./ÿ/ÿ 7ÿÿÿ
ÿ,ÿ/ÿ
ÿÿÿ,
,.ÿ&1ÿ2ÿ8
811711 177
01231245ÿ72893ÿ
ÿÿÿ!
"ÿÿÿ#
ÿÿÿ
7ÿ)!ÿÿ
ÿÿ
"ÿ!ÿ2ÿ
ÿÿ-ÿ
%ÿ(ÿ+ÿ, ,ÿÿÿ% ÿ
$%&'ÿ(ÿ ÿ)* ' 4ÿ./ÿ
"ÿÿÿ
,#ÿ9ÿ+ÿ#ÿÿ,
ÿÿ,ÿ#ÿ0
1 1'ÿÿÿ!ÿ2
%3 ÿ ÿ-ÿ45'ÿ
7(ÿ'ÿÿ
!ÿÿ . ÿ!
/ÿÿ,
+"ÿ2(
'ÿ/
,ÿ!!
'ÿ(ÿ ÿ %ÿ(ÿ+ÿ, ÿ,#ÿ
' +ÿ"ÿ "ÿ
ÿ!"ÿ!ÿ
ÿ4(ÿ'ÿ
ÿ!ÿ
,ÿ $ÿ
/ÿÿ
#"ÿ
'ÿ(ÿ%# %ÿ )+,ÿ,ÿ'
811711 177
01231245ÿ72893ÿ
7ÿ""ÿÿ"
&"ÿÿ ' ÿ&"
"("ÿÿ%"
$)ÿ2
"ÿ"("
%ÿ&&"
ÿÿ!" !ÿ!"" #ÿÿ"$ÿ% ÿ"%*ÿ
$ÿ)ÿ")ÿ
"ÿ&)ÿ&ÿ
ÿ4ÿ""ÿ
"ÿ&ÿ
%ÿ +ÿ"
"("ÿÿ"
"*")ÿ"
ÿÿ#"*" #ÿ ,$%ÿ%ÿ
811711 0177
01231245ÿ72893ÿ
7+ÿÿ,&*-ÿÿ
ÿÿ*&
,&&ÿÿ&ÿ.ÿ!"
/ ÿ!ÿÿÿ
&ÿ.ÿ! ÿ/ ÿ.
*&ÿ2ÿÿ
ÿ&ÿ0&ÿ,-ÿ*
)ÿ42ÿ&ÿ1 2
+&ÿ*ÿÿ)
#1ÿ/ ÿ&ÿÿ3&ÿ
!"#ÿ$ÿ%& 'ÿ" ()*ÿ*ÿ# +&ÿ0--ÿ*ÿ
&ÿ.ÿ! ÿ/ ÿ.
*&ÿ4ÿ/45ÿ/! ÿ5
1/ÿ&ÿÿÿ&ÿ.
! ÿÿ.ÿ*&ÿ9
1ÿ&&-ÿ.
*&ÿ,&ÿÿ
-ÿ*ÿ,-
,&*-ÿ6ÿ+ÿ&
3&ÿ1&-ÿÿ'#
ÿ&&-ÿ*
ÿ&ÿ.ÿ! ÿ
7ÿÿÿ
ÿ&ÿ*ÿ
ÿÿÿ&
!"#ÿ$ÿ+#11%7' ÿ8&& ()*ÿ*ÿ# &)ÿ,-ÿ2ÿ1
/&ÿÿ&ÿ.
9ÿÿ&ÿ-
ÿÿÿ
7ÿ(.&ÿÿ&
ÿÿ
&9ÿ.ÿ2ÿ
&ÿÿ5ÿ&
!ÿ$ÿ&)ÿ* *&ÿÿÿ! ÿ
!"#ÿ$ÿ& ÿ(&: # 4ÿ%;ÿ
&9ÿÿ&ÿ
*-ÿ9ÿ)ÿ&-ÿÿ*
ÿÿ*ÿ-ÿ6
< <#ÿÿÿ.ÿ=
!+ ÿ ÿ5ÿ24#ÿ
#ÿ$ÿ/& /ÿ/&& !ÿ$ÿ&)ÿ* 7$ÿ#&&ÿÿ&
# .&ÿÿ % ÿ.&
&;&ÿÿ*&
)9ÿ2$
#&ÿ&;&
*ÿ..&
811711 7177
01231245ÿ72893ÿ
ÿ !ÿ
"ÿ#ÿ#ÿ
ÿ$#ÿ$ÿ
ÿ4%ÿ&ÿ
ÿ$ÿ
ÿ 'ÿ
(ÿÿ
!#ÿ
&ÿ%ÿ)! )ÿ *" ÿÿ&
811711 77177
SITE PLAN
E
7
N
Z1-L0-IMD-020
CHANGE B07 B08
S
18.65 m² (F)
N
7.59 m²
LA
.63
W
(M)
QIB
252
Z1-L0-IMD-022
B14
4.78 m² Contracts for performing and supervising on-site installation preparations should be concluded with technically
B01
4.77 m²
B06
competent companies by the customer. The customer is responsible for timely and proper completion and B12 B13
7.16 m² SW supervision of all preparations for installation at the construction site in observance of all applicable legal regulations
B11
B04
B10
TS (e.g. X-ray regulations, radiation protection regulations) and all applicable general recognized rules of technology
(e.g. VDE regulations, DIN standards).
Execution and supervision of installation preparations at the construction site and later observance of the standard
operating conditions are not included in our duties. The customer is responsible for checking the static calculations
and, where applicable, the air conditioning in the building to be equipped.
P2.G
GENERAL NOTES
1. PROJECT REFERENCE ELEVATION ±0.00 IS AT +27.50 QATAR NATIONAL DATUM. ELEVATIONS
AT SHOWN ARE WITH RESPECT TO THE PROJECT REFERENCE ELEVATION OF ±0.00 AT GROUND FLOOR.
2. ALL DIMENSIONS ARE IN MM UNLESS NOTED OTHERWISE.
3. EQUIPMENT DIMENSIONS ARE SUBJECT TO CHANGE BASED ON MATERIAL SUBMISSION APPROVAL.
1486
1550
Room dimensioning 4.
5.
CONTRACTOR TO COORDINATE AS PER SITE CONDITIONS WHEREVER APPLICABLE.
SHOP DRAWINGS ARE ISSUED IN COORDINATION WITH APPROVED DETAILED DESIGN FF&E AND
SBA MEDICAL EQUIPMENT SELECTIONS. SOME AREAS ARE SUBJECT TO CHANGE AFTER MATERIAL
SUBMITTALS APPROVAL .
The indicated room dimensions have to be checked on site. The planning department has to be informed about
EAT possible deviations. Otherwise we cannot assume any guarantee for the accurate implementation of the dimensions
2760 5161 indicated in the planning documents.
AT
A
2981
2981
3126
± 0.00
KEY PLAN
-0.10
10 10.1 17 17.1
1 26
A A
1.12 Ambient lighting in rooms with diagnostics or with workstations must comply with the respective local and national Z2
Z1-L0-IMD-021
1860
D regulations. P2.A
F F
P3.A
3126
General requirements like the needed intensity of illumination - adjustable, reproducible, flicker-free or a limitation of
1.07 dazzlings and reflections etc. have to be observed (EN 12464-1, DIN 5035-7).
2500
P2.F P3.F
1.06
WORK /
5821
P2.1 P3.1
1.02 C Z4 Z5
± 0.00
3126
-0.10 CONTROL V V
1929 1.01
Z1-L0-IMD-023 Display screen workstations
6 13 13.1 21
For setting up display screen workstations, take account of the guidelines in the Display Screen Workstation
directive as well as any national regulations (e.g. EN ISO 9241-5).
REFERENCE DOCUMENTS
1.09
DOCUMENT NO TITLE REV
1.10 AT MMC-HBK-AP-DD-BAP-MEQ-L0-APO-1800
BUILDINGS 1 - ALL ZONES - ALL PARTS - OVERALL -GROUND FLOOR -
MEDICAL EQUIPMENT PLAN
2
2440
MMC-HBK-P1-DD-B01-MEQ-L0-P06-1806 BUILDING 1 - ZONE 01- PART 06- GROUND FLOOR MEDICAL EQUIPMENT PLAN 2
2840
B Safety distances
1 3126
1435
SW Distances from moving parts of the medical device to walls, furniture and other equipment have to be kept to avoid
injuries by crushing in compliance with local regulations, e.g. a minimum distance of 50 cm according to EN 349.
1871
1.05 It is the customer's responsibility to ensure the above requirements are followed.
Approval Status
Approved as
Commented
Re-Submit
A Approved
This is to avoid the risk of injury.
D Rejected
Revise &
Authority Signature Date
If safety distances are not maintained appropriate on-site safety measures have to be put in place.
C
B
1.04 1.03 Clear visible markings according to national guidelines, e.g. crushing warning signs, hazard warning
1.08 AT tape, hazard area cordon, safety mats, may be required.
MMC-PMO
The approval and comments shall not in any ways relieve or absolve the
Contractor from compliance of any obligations or liability under or in connection
M.1
Room height with the contract, project requirements and specifications.
1.01 Gantry 2580 1000 #1, isocenter 1020 The CT system is not sensitive to common vibrations. If
the CT is away from vibrational sources, or the CT is Power Distribution Cabinet B PDC B with transport pallet 1120 900 2070 ≤450
mm LEAD CONSULTANT:
1.02 Patient table PHS 4n (2000 mm scannable range) 500 * * included in 1.01 replacing a CT system that to date has not shown image IRSmx4d tower with transport pallet 600 800 690 51.9
quality problems due to vibrations, it is usually not
IRSmx (Image Reconstruction System Cabinet)
MEINHARDT
necessary to execute vibrational measurements. IRSmx5d tower 645 175 450 25 14TH FLOOR AMWAL TOWER WESTBAY
1.03 PDC A cabinet with UPS 610 2000 AT Emergency off button with locking mechanism P.O. BOX 1362 DOHA, QATAR
TEL: (974) 44117840 FAX: (974) 4411 7845
1.04 PDC B cabinet 400 2000 If there are any doubts, the following thresholds have to be PHS 4n - with packing 2580 875 1570 700
verified by measurement: On-off button with pilot lamp ARCHITECT & INTERIOR DESIGNER:
1.05 IRSmx5d Tower PC (Image Reconstruction System) 25 500 1 10 100
MPT 2n - with packing 2530 850 1430 850
1.06 Control unit with monitor, control box, keyboard 10 100 SW Radiation warning lamp
1.07 ICS Tower 13 PC (syngo Acquisition Workplace) 25 500 Patient Handling System PHS) must not exceed the
1.08 Power Distributor BY HBK, not in thresholds as described in the above shown diagram. Radiation ready indicator
MEDICAL PLANNER: MEDICAL EQUIPMENT : MEP SUBCONTRACTOR : SUBCONTRACTOR :
scope of delivery Any transient vibration (shocks) has to be less than 0.5 m/s² peak-to-peak in the time domain.The vibrations have to CV CARE Vision
be measured with a sampling rate of 1000 Hz.
Climatic conditions for transport and storage
It is the customer's responsibility to contract a qualified specialist. The specialist must implement site The system cable connections (power supply) between
Gantry and PDC must have a minimum length of 7 m. Transport / storage Temperature -20 to +50 °C
modifications to meet the specific limits, and to design structural solutions in case of deviations.
Place excess lengths in the cable duct. all components Relative humidity 10 to 90 % PROJECT TITLE:
Cable diameter 30 mm, minimum bending radius 148 mm. Barometric pressure 700 to 1060 hPa
Transport Gantry not to scale Notes on preparations for installation Temperature gradient 10 K/h (non-condensing) ENGINEERING, PROCUREMENT AND
Max. storage time 2 months
Standard transport Transport through narrow passages 4
Contracts for performing and supervising on-site installation preparations should be concluded with technically CONSTRUCTION (EPC/TURNKEY) FOR
competent companies by the customer. The customer is responsible for timely and proper completion and The data for transporting and storing the CT system apply only if the system is shipped free of damage in the
Rollers swivelled out Rollers swivelled in supervision of all preparations for installation at the construction site in observance of all applicable legal regulations Environmental conditions for operating transport packaging provided by the manufacturer. All components have to be stored in the original transport
MILITARY MEDICAL CITY HOSPITAL
1220 1220
(e.g. X-ray regulations, radiation protection regulations) and all applicable general recognized rules of technology
Entire CT System Air temperature range 18 to 28 °C
packaging. If the original packaging is no longer available, you have to use an equivalent packaging for intermediate (MMC) AT TAREK BIN ZEYAD CAMP
(e.g. VDE regulations, DIN standards). storage. After moving the CT system from a cold environment into a warm room, make sure that the entire CT
1
1256 1
1256 Execution and supervision of installation preparations at the construction site and later observance of the standard Relative humidity 20 to 75 % system is dry before switching it ON.
IN SAILIYAH (SOUTH OF DOHA)
operating conditions are not included in our duties. The customer is responsible for checking the static calculations Barometric pressure 800 to 1060 hPa
and, where applicable, the air conditioning in the building to be equipped. Temperature gradient 6 K/h (non-condensing)
PROJECT NO. AA-103-01
min. 1999
min. 1999
Maintain the temperature range required to meet the needs of changing seasons.
3 3 Gantry with folded-up covers not to scale If it is not possible, an on-site room heating and air conditioning system is required. DRAWING TITLE:
When supplying air from the outside (fresh air), we recommend the on-site use of Network Integration BUILDING 01-ZONE 01, -PART 06 - GROUND FLOOR -
FRONT VIEW SIDE VIEW coarse Filters, Filter class EU3 to EU4 for filtering dust particles up to > 10 μm CT SCAN ROOM (Z1-L0-IMD-021) - SOMATOM Drive
The Siemens Healthineers components are using TCP/IP Protocol, a 100/1000 Mbit/s switched Ethernet network
(DIN 24185/Part 2). PRE-INSTALLATION EQUIPMENT ARRANGEMENT PLAN
2 2 and static IP addresses.
264 3331 4 4 The required network cabling (min. CAT 5 TP) has to be provided on site. Media converters, which are needed for THIS DRAWING HAS BEEN CHECKED FOR COMPLETENESS, ACCURACY, COORDINATION, AND
5 1801 COMPLIANCE WITH THE CONTRACT DOCUMENTS. ALL DIMENSIONS AND FIELD CONDITIONS
using fibre optic cabling, are not in scope of delivery. HAVE BEEN VERIFIED.
6
5 To prepare the implementation of the new system into the existing network environment, the availability
1 With touch panels on the rear (optional) Smart Remote Services (SRS) of the needed network data at least two weeks before starting the installation is mandatory.
This is the only way to ensure a seamless integration of the new system into the workflow of the department.
Mr. ALAADDIN AKTAN 20 Sept 2022
CONTRACTOR'S REPRESENTATIVE: DATE:
2 Covers must removed for mounting the transport device 2 Smart Remote Services (SRS) is used for remote diagnostics as well
1 as remote service to provide highest system availability. STATUS:
2100
2100
PRABHU
ABAM ABAM
Gantry with transport device = 2865 kg, Transport device = 285 kg
1 Height of the gantry 2 Height with folded-up cover (for service) 3 Iso center WORKING DRAWING REVISION NOTES DESIGNED SCALE A1
If there is false floor in transportation route please check bearing capacity of this floor on-site!
MEINHARDT 1:30
If necessary, cover the transport route with metal sheets to distribute the load. 1 : AS PER COMMENTS FROM OPERATOR; DRY WALL SHIFTED TO 400mm TO GAVE SPACE ON THE CT SCAN MACHINE AND TO RECESS ON THE WALL THE SUB
4 e.g. CARE Vision 5 Folded-up cover "hood right" 6 Folded-up cover "front ring"
The maximum possible floor load (two-point load) per roller that may arise during gantry transport is 16 kN. EQUIPMENT (PDCa and PDCb) OF THE CT SCAN MACHINE.. DRAWING NO. REVISION NO.
The door must have a final clearance of 1250 mm if bed entrance to the CT examination room is requested. Note the height of the folded-up covers if planning any technical equipment above the gantry ! MMC-HBK-P1-WD-B01-MEQ-L0-P06-3125 1
E
N
7
M P2.G B07 B08
S
+5.50
N
LA
.63
+5.40
W
+5.50
QIB
252
B14
B02 B03
B01
B06
B12 B13
B11
B10
B04
GENERAL NOTES
1. PROJECT REFERENCE ELEVATION ±0.00 IS AT +27.50 QATAR NATIONAL DATUM. ELEVATIONS
SHOWN ARE WITH RESPECT TO THE PROJECT REFERENCE ELEVATION OF ±0.00 AT GROUND FLOOR.
1 2.
3.
ALL DIMENSIONS ARE IN MM UNLESS NOTED OTHERWISE.
EQUIPMENT DIMENSIONS ARE SUBJECT TO CHANGE BASED ON MATERIAL SUBMISSION APPROVAL.
4. CONTRACTOR TO COORDINATE AS PER SITE CONDITIONS WHEREVER APPLICABLE.
5. SHOP DRAWINGS ARE ISSUED IN COORDINATION WITH APPROVED DETAILED DESIGN FF&E AND
MEDICAL EQUIPMENT SELECTIONS. SOME AREAS ARE SUBJECT TO CHANGE AFTER MATERIAL
SUBMITTALS APPROVAL .
1200
PDC A CABINET Z1-L0-IMD-021
A A
WITH UPS
P2.1
Z1 Z3 P3.1
F Z2 F
POWER DISTRIBUTION
1050
AT
SWITCH SWITCH AT
BOX BY HBK
3000
P2.F P3.F
EMERGENCY STOP AT PDC A CABINET PDC B PDB - POWER 1.01 P2.1 P3.1
SWITCH WITH UPS CABINET DISTRIBUTION Z4 Z5
BOX
CT ROOM 2980
V V
EMERGENCY STOP
Z1-L0-IMD-021
1800
1800
1800
AT
SWITCH
IK 200x60mm
1.02
6 13 13.1 21
OK FFL to UK PDB
3000
1000
WALL TRUNKING
1000
50mmØ OPEN PVC PIPE REFERENCE DOCUMENTS
PROVISION FOR INJECTOR
THRU CONTROL ROOM 5.07 1.05
DOCUMENT NO
MMC-HBK-AP-DD-BAP-MEQ-L0-APO-1800 2
BUILDING 1 - ZONE 01- PART 06- GROUND FLOOR MEDICAL EQUIPMENT PLAN 2
1.02
1950
1800
1800
IK 200x60mm
OK FFL to FK 200x80mm
UK PDB
1.05 FLOOR TRUNKING
1000
1000
MILITARY MEDICAL CITY PROJECT - AL SAILIYAH
Approval Status
Approved as
Commented
C CT SCAN ROOM - ELEVATION C
Re-Submit
A Approved
D Rejected
500
494
Revise &
500
CV
MEQ-3126 SCALE: 1:30 Fixpoints Authority Signature Date
FFL: ±0.00
Fixpoint Components
C
B
MMC-PMO
GR Gantry
80
REVISION DETAILS DATE DWN DES CHD APD
SW Radiation warning lamp
EMPLOYER
-0.10
+5.50 Radiation ready indicator
QATAR ARMED FORCES
+5.40 CV CARE Vision
MILITARY MEDICAL CITY HOSPITAL PROJECT
MANAGEMENT OFFICE P.O. BOX 2075 DOHA,
The system cable connections (power supply) between QATAR
Gantry and PDC must have a minimum length of 7 m.
+3.00
PROJECT TITLE:
2510
1200
2050
1800
1000
1600
CONTRACTOR'S REPRESENTATIVE: DATE:
1035
50mmØ OPEN STATUS:
PVC PIPE
PROVISION FOR ISSUED FOR CONSTRUCTION
130 130
INJECTOR
SUBMISSION STAGE DATE
FFL: ±0.00
560
WORKING DRAWING 09/08/2022
300
FFL: ±0.00 FFL: ±0.00
DRAWN DESIGN CHECKED
PRABHU
ABAM ABAM
E
N
B07 B08
S
N
LA
.63
W
QIB
252
I-Control with wall bracket (schematic) not to scale
SOMATOM Drive - Equipment Legend B14
B09 B05
B02 B03
4 Weight (kg), Heat dissipation to the air (W) B01
7.16 m² 400
Pos. Description kg W Remark
B06
B12 B13
B11
1.01 Gantry 2580 1000 #1, isocenter 1020 mm B04
B10
1.02 Patient table PHS 4n (2000 mm scannable range) 500 * * included in 1.01
130 1.03 PDC A cabinet with UPS 610 2000
15
10 3
1.04 PDC B cabinet 400 2000
1.05 IRSmx5d Tower PC (Image Reconstruction System) 25 500
00
0
x. 1
1.06 Control unit with monitor, control box, keyboard 10 100
119
90
2
ma
9 4 1.07 ICS Tower 13 PC (syngo Acquisition Workplace) 25 500 GENERAL NOTES
1300
1. PROJECT REFERENCE ELEVATION ±0.00 IS AT +27.50 QATAR NATIONAL DATUM. ELEVATIONS
1 1.08 Power Distributor by customer, not in scope SHOWN ARE WITH RESPECT TO THE PROJECT REFERENCE ELEVATION OF ±0.00 AT GROUND FLOOR.
of delivery 2. ALL DIMENSIONS ARE IN MM UNLESS NOTED OTHERWISE.
6 3. EQUIPMENT DIMENSIONS ARE SUBJECT TO CHANGE BASED ON MATERIAL SUBMISSION APPROVAL.
1 4. CONTRACTOR TO COORDINATE AS PER SITE CONDITIONS WHEREVER APPLICABLE.
800 - 900
7 159 5. SHOP DRAWINGS ARE ISSUED IN COORDINATION WITH APPROVED DETAILED DESIGN FF&E AND
1821
5 MEDICAL EQUIPMENT SELECTIONS. SOME AREAS ARE SUBJECT TO CHANGE AFTER MATERIAL
SUBMITTALS APPROVAL .
1 Wall bracket (thickness 35 mm) with wall mounting holes Ø 6.6 mm 2 Locking lever for locking/unlocking
KEY PLAN 1
10 10.1 17 17.1
26
cabling
550
Z2
P3.1
728
549
-0.10 CONTROL - Line frequency: 50-60 Hz - Current: 1.25 A The CT system is not sensitive to common vibrations. If
- Current: 0.7 A - Cable length: 1.5 m
Z1-L0-IMD-023 - Stand-by: 0.5 W the CT is away from vibrational sources, or the CT is P2.F P3.F
800
1790
replacing a CT system that to date has not shown image P2.1 P3.1
6 7 Z4 Z5
quality problems due to vibrations, it is usually not
8 IVP control cable 9 Strain relief for power supply cable 10 Connection for power supply cable
necessary to execute vibrational measurements.
V V
Delivery volume: - Control unit and a wall bracket for wall mounting (standard installation) If there are any doubts, the following thresholds have to be
- Drilling template verified by measurement:
550
1 10 100
- Installation material: 3x Fischer dowels 6 13 13.1 21
(Type: UX 8x50; nominal drill hole diameter: 8 mm; Drill hole depth: min. 60 mm
- 3x screws M6x80, 3x screw head caps grey REFERENCE DOCUMENTS
Patient Handling System PHS) must not exceed the
thresholds as described in the above shown diagram. DOCUMENT NO TITLE REV
1263 1400 3347 Statics not to scale MMC-HBK-P1-DD-B01-MEQ-L0-P06-1806 BUILDING 1 - ZONE 01- PART 06- GROUND FLOOR MEDICAL EQUIPMENT PLAN 2
be measured with a sampling rate of 1000 Hz.
The floor construction has to be performed solid and free of It is the customer's responsibility to contract a qualified specialist. The specialist must implement site
1490
1490
1911 4 3 vibration, e.g. concrete flooring C20/25 to C50/60 corresponding to
Center of DIN EN 206-1, according to the maximum values as specified in the
modifications to meet the specific limits, and to design structural solutions in case of deviations.
gravity textblock “Floor and building vibrations”.
256
It is recommended to test the weight capacity of the concrete or MILITARY MEDICAL CITY PROJECT - AL SAILIYAH
1 2 composite flooring by a stress analyst. Approval Status
Fastening the gantry on the floor is possible but only necessary in
Approved as
Commented
Re-Submit
A Approved
94 countries prone to earthquakes or according to local regulations.
D Rejected
Revise &
Securing the patient table to the floor is mandatory.
Authority Signature Date
PDC - Power Distribution Cabinet not to scale
C
B
Total weight of Gantry: 2580 kg : F stat total 25 kN
MMC-PMO
FRONT VIEW ILLUSTRATION for the cable inlet (Isometry)
Lead shielded wall Partial load on gantry foot 1 2 3 4
Floor contact area gantry foot [cm²] 52.1 52.1 52.1 52.1
During gantry installation and leveling, the maximum possible load on one gantry foot can be 14 kN (the gantry is
1 CT SCAN ROOM - CONCRETE FILL-UP PLAN 1 ISSUED FOR CONSTRUCTION 31/10/2022 ABAM ABAM PR SG
standing on two diagonal feet). Design access floors for a weight capacity of min. 400 kg per slab/plate. During
1850
1950
gantry transport, the load may be higher at certain individual points (3-point load, e.g. due to uneven flooring).
MEQ-3127 SCALE: 1:30 00 ISSUED FOR CONSTRUCTION 20/09/2022 ABAM ABAM PR SG
REVISION DETAILS DATE DWN DES CHD APD
EMPLOYER
1 Adjustable gantry feet (Ø 80 mm); mounting holes (Ø 17 mm) ICS Tower 13 not to scale
MAIN CONTRACTOR:
100
2
Adjustable patient table feet with mounting holes (inner: Ø 20.5 mm); minimum extraction force: 2.03 kN at the two mounting points bearing
the greatest load at maximum patient travel path and a 230-kg load. According to IEC 60601-1, it is mandatory that the 4-fold safety factor be observed. 840 488
FRONT VIEW SIDE VIEW
7
TOP VIEW
2
570 0
HBK CONTRACTING CO. WLL
50 43
900 P.O. BOX 1362 DOHA, QATAR
2 Alternative floor mounting holes (Ø 18 mm), if position 2a is not usable 1252 4
2 TEL: (974) 44451111 FAX: (974) 44451000
120
1070 5 2 6
LEAD CONSULTANT:
3 Floor-inlet for cable/hose connection: 49/97 x 488 mm
890 TOP VIEW ILLUSTRATION for floor mounting
3
MEINHARDT
420
175 420 (top view of pedestal)
4 No floor cut-out permitted in this defined area (Ø 190 mm) around each adjustable gantry foot 1435 900 500 900 14TH FLOOR AMWAL TOWER WESTBAY
8 P.O. BOX 1362 DOHA, QATAR
Alternative to 3 : 689.5 745.5
40
TEL: (974) 44117840 FAX: (974) 4411 7845
5 Core hole: max. Ø 100 mm for HV cables and other system cabling 5 6 3
395
2 1
3 1 ARCHITECT & INTERIOR DESIGNER:
8
6 Core hole: max. Ø 80 mm for system cabling
10 3 3
1
691.5
650
570
7 Core hole: max. Ø 160 mm for water hoses 1 1 4
25
8 Gantry footprint 4
1 Air intake 2 Air exhaust (primary) 3 Air exhaust (secondary) 3 3 14 MEDICAL PLANNER: MEDICAL EQUIPMENT : MEP SUBCONTRACTOR : SUBCONTRACTOR :
7 Ø
4 Air must not exhaust directly to the air intake 5 e.g. a wall
9 "Electronics cover" for cabling between gantry and patient table (in scope of delivery)
100
40
6 Minimum distance for cable routing and air circulation 7 Minimum distance for air circulation
100 700 100
9 7
Floor requirements: Gantry and patient table orientation point
PROJECT TITLE:
CT
The floor construction has to be performed solid and free of vibration, e.g. concrete flooring C20/25 to C50/60 scan plane
950
corresponding to DIN EN 206-1, according to the maximum values as specified in the textblock
ILLUSTRATION for the cable inlet
(top view of pedestal)
ENGINEERING, PROCUREMENT AND
service area
“Floor and building vibrations”. Anti-static flooring is recommended for the examination room. ESD flooring is also suitable. 1 1 900 CONSTRUCTION (EPC/TURNKEY) FOR
NOTE: A solid and vibration-free floor is needed to avoid a negative influence on the vibration characteristic of the gantry. IRSmx5 tower not to scale
An incorrect vibration characteristic of the gantry can cause problems with balancing and image quality.
MILITARY MEDICAL CITY HOSPITAL
6 (MMC) AT TAREK BIN ZEYAD CAMP
max.120
Permissible tolerance for floor levelness is at maximum 10 mm, based on the area of the adjustable feet of the
4
gantry and the patient table. 9 175
SIDE VIEW of socle IN SAILIYAH (SOUTH OF DOHA)
orientation point
650
25
On-site subconstruction have to fulfill the flooring conditions. It is recommended to test the weight capacity of 5
PHS 4n
82.5
175 120
the concrete or composite flooring by a stress analyst. 645 PROJECT NO. AA-103-01
Floor mounting, as well as principally every kind of CT system component mounting, must be performed in 2 2 1 4 3 DRAWING TITLE:
accordance with country-specific guidelines in countries prone to earthquakes.The gantry and the patient
645
table have to be on the same (isocentric) plane.
55 790 55 BUILDING 01-ZONE 01, -PART 06 - GROUND FLOOR -
450
450
2a 2a CT SCAN ROOM (Z1-L0-IMD-021) - SOMATOM Drive
An "Installation Template" with footprint information for the gantry and patient table is optional available. 9 650 32.5
2 PRE-INSTALLATION CONCRETE FILL-UP PLAN
May not be used as a drilling template. A drilling template is not necessary. 7
150
205 5 THIS DRAWING HAS BEEN CHECKED FOR COMPLETENESS, ACCURACY, COORDINATION, AND
COMPLIANCE WITH THE CONTRACT DOCUMENTS. ALL DIMENSIONS AND FIELD CONDITIONS
Floor mounting Gantry 1 System cables HAVE BEEN VERIFIED.
- Possible through four mounting holes, but not required in principle (only in areas prone to earthquakes). FRONT VIEW SIDE VIEW TOP VIEW
2a 2a Cable inlets (cut-outs in the base) when cables are routed on the floor; if not used, the cut-outs are covered
Mounting materials must be provided on-site. 2
with dummy plates; feed lines are possible from all four sides
1 Front side with grid for air intake (cooling) 2 Support feet 3 Air intake 4 Air exhaust
- Leveling tolerance: in Z-direction ± 0.2 deg, in X-direction ± 0.5 deg Mr. ALAADDIN AKTAN 20 Sept 2022
2 2 3 Floor mounting holes (all Ø=14 mm) 4 PDC base (footprint) CONTRACTOR'S REPRESENTATIVE: DATE:
- Leveling with four adjustable feet. 5 Minimum distance for air circulation 6 Air must not exhaust directly to the air intake
STATUS:
Ø 5 Cable inlet in PDC base; opening is adjustable 6 Air exhaust for system cooling; must not be covered
7 Sliding door or similar ISSUED FOR CONSTRUCTION
For example, the sliding door of a wall For example, a wall; no safety distance for
Floor mounting patient table 280 The safety distances for air circulation (system cooling) must be maintained. Openings for air circulation in the 7 closet, and so on 8
air circulation (cooling) required SUBMISSION STAGE DATE
393 tower housing must not be covered. Air must not exhaust directly into the air intake (a closed air loop).
- The patient table MUST be secured to the floor (4 mounting holes). 9 Safety distance for air circulation (cooling) required 10 Top exhaust for system cooling; must not be covered 09/08/2022
WORKING DRAWING
- Installation material for floor mounting in concrete is in scope of delivery. Therefor min.140 mm thickness of concrete is required:
Anchor bolts type 1: 4x HSL-4G M 10/60 heavy-duty sleeve (used without bonded screed) or DRAWN DESIGN CHECKED
Anchor bolts type 2: 4x HSL-4G M 10/120 heavy-duty sleeve (used with up to 60 mm thick bonded screed) PRABHU
ABAM ABAM
- Leveling with four adjustable feet.
- Remove the floor cover in the entire footprint area of the table base.The support plate is equal to the footprint DESIGNED SCALE A1
of the patient table base and is used as a help to cut out the floor covering. MEINHARDT 1:30
MMC-HBK-P1-WD-B01-MEQ-L0-P06-3127 1
E
7
N
B07 B08
S
N
7.16 m²
LA
.63
W
QIB
252
B14
SW TS B09 B05
1 Minimum measured line voltage [ VAC ]
2 B02
B01
B03
B06
B12
300 B11
B13
250
250
800 The maximum allowed line impedance
P2.G depends on the actual measured lowest line
200
voltage.
+1800mm aFFL AT We recommend monitoring the voltage over
DRAIN PIPE FOR 150 a period of 24 hours preferable on work days GENERAL NOTES
1. PROJECT REFERENCE ELEVATION ±0.00 IS AT +27.50 QATAR NATIONAL DATUM. ELEVATIONS
CONDENSATION OF THE to specify the minimum.
1640
1670
SHOWN ARE WITH RESPECT TO THE PROJECT REFERENCE ELEVATION OF ±0.00 AT GROUND FLOOR.
440
2. ALL DIMENSIONS ARE IN MM UNLESS NOTED OTHERWISE.
1842
MACHINE BY HBK SBA 100 85 3. EQUIPMENT DIMENSIONS ARE SUBJECT TO CHANGE BASED ON MATERIAL SUBMISSION APPROVAL.
4. CONTRACTOR TO COORDINATE AS PER SITE CONDITIONS WHEREVER APPLICABLE.
300 EAT
5. SHOP DRAWINGS ARE ISSUED IN COORDINATION WITH APPROVED DETAILED DESIGN FF&E AND
MEDICAL EQUIPMENT SELECTIONS. SOME AREAS ARE SUBJECT TO CHANGE AFTER MATERIAL
1910 5710 50
SUBMITTALS APPROVAL .
WATER COOLING AT
150 ± 0.00 0 1
1760 360 380 400 420 440
-0.10 2 x BK 130x70mm
CT ROOM @300mm aFFL KEY PLAN
500
10 10.1 17 17.1
water
1 26
300
FK 300x80
cabling
A A
Z1-L0-IMD-021 P2.1
Z1 Z3 P3.1
CONCRETE SCREED F Z2 F
2
549
840
728
2220
TRUNKING P2.1 P3.1
80
Image Control System (Tower PC) 6 13 13.1 21
-0.10
MMC-HBK-P1-DD-B01-MEQ-L0-P06-1806 BUILDING 1 - ZONE 01- PART 06- GROUND FLOOR MEDICAL EQUIPMENT PLAN
3128 to @3000mm aFFL 3 SECTION DETAIL
180 AFFL AT AT Emergency off button with locking mechanism MEQ-3128 SCALE: 1:5
1530
Approved as
1900 3100
Commented
CV @ +300mm aFFL
Re-Submit
A Approved
Radiation ready indicator
D Rejected
Revise &
AT +1800mm aFFL SW FK 300x80
135
Authority Signature Date
CV CARE Vision
C
B
MMC-PMO
85
FK 200x80mm The system cable connections (power supply) between
Cut-out in cable duct
80
*4 Gantry and PDC must have a minimum length of 7 m.
IK 200x60mm Place excess lengths in the cable duct.
300
849
305
1.08 @300mm aFFL
*3 The approval and comments shall not in any ways relieve or absolve the
275
549
M
Gantry contour
Lead shielded wall
Notes
For service purposes; only required if there is no existing 1 ISSUED FOR CONSTRUCTION 31/10/2022 ABAM ABAM PR SG
* 10 m 10 m
cut-out, Ø in mm (i.e. in suspended ceiling)
16 m 20 m LEAD CONSULTANT:
25 m 30 m
Power supply SOMATOM Drive *For the shown installation the selected options have to be used.
KB core hole, Ø in mm (minimum inner clearance required)
MEINHARDT
Legend electrical installation BD floor opening, top view LxW (break-throughs and openings, which delimit the
* The length of the water hoses depends on the position of the cooling water connection. 14TH FLOOR AMWAL TOWER WESTBAY
P.O. BOX 1362 DOHA, QATAR
Mains TN-S 3/N/PE AC 50/60 Hz ± 2Hz Line impedance Power consumption 1 To external conductive parts
The final lenghts must be determined on site with the Siemens project manager! fireproof area, must be closed on site after the installation is finished). TEL: (974) 44117840 FAX: (974) 4411 7845
Voltage 400 V ± 10 % Only copper cables are allowed DD ceiling opening in ceiling above LxW (break-throughs and openings, which ARCHITECT & INTERIOR DESIGNER:
2 Test jack 4 mm pin plug Additional electrical connectivity for the customer at delimite the fireproof area, must be closed on site after the installation is finished).
Line to line unbalanced max. 2 % Connection value 173kVA Stand-by 16kVA
WD wall break WxH (break-throughs and openings, which delimit the
Cable cross section is to be determined by national regulations. System off 1kVA 3f Circuit breaker 125A Siemens 3VA1112-4ED46-0AA0 Siemens Customer
Healthineers fireproof area, must be closed on site after the installation is finished).
The size of the terminals in the PDC are designed for. 25 / 95 mm² for time up to < 2 s 293kVA
3a Circuit breaker 32A Siemens 3VA1132-4EE46-0AA0 name of a fixpoint or a connection point
PE N L1 L2 L3 Electrical installation according to IEC 60364-7-710, national regulations must be observed intern 1m 2 4.4
MEDICAL PLANNER: MEDICAL EQUIPMENT : MEP SUBCONTRACTOR : SUBCONTRACTOR :
6
16 □
Cu
1.4 13 13 13 21 T1 T2 3VA9988-0AA12 24 VDC / 2 A
CONSTRUCTION (EPC/TURNKEY) FOR
K2 K3 K1 K1 80% 20% Side mounted motor per Circuit breaker 3VA 24V DC grounded socket, switchable, installation according to national guidelines
PA 1.1 14 14 14 22
10s 1s 7
3VA9117-0HB10
2 2.4
MILITARY MEDICAL CITY HOSPITAL
1 K K
Undervoltage release for MCCB per Circuit breaker Siemens intern 1m Drain pipe provision 32-50mm Ø
6 33 K4 9 10
8
3VA9908-0BB11 max.
max. 2,5 mm² flexible
2.3 warning lamp / indicator,
(MMC) AT TAREK BIN ZEYAD CAMP
18 43
TP telephone socket
K1 2 9 Timing relay 3RP2505-1AB30-0AC1
250 VAC / 2 A
ceiling mounted, wall mounted IN SAILIYAH (SOUTH OF DOHA)
PE
L1
L2
L3
T2 K1 24 VDC / 2 A
~
24 23
34 15 44 network socket
5 18 K6 10 Timing relay Siemens 3RP2505-1AB30-0AC1
T1
4
1.1 1.4
12 Contactor relay 24V DC 3RH2131-2BB40
intern 1m 2
max. 2,5 mm² flexible
2.2
UK bottom edge
OK top edge PROJECT NO. AA-103-01
15 AUX 3 VA 19 2.1
max.
DRAWING TITLE:
6A - Contactor relay 24V DC 3RH2131-2BB40
250 VAC / 2 A bottom edge suspended ceiling OK FFL top edge finished floor
13 24 VDC / 2 A UK ABGD
4 3 B1 F1 E1 K6 BUILDING 01-ZONE 01, -PART 06 - GROUND FLOOR -
5 Priorität 1 20 14a Impulse switch Eltako ES 12Z-200-UC
Q3 Q2 Q1 MODE 1
M 3VA <U T1 T2 K2 K3 ZE + ZA K4 K1 +
UK RHD
bottom edge bare ceiling OK RFB top edge unfinished floor CT SCAN ROOM (Z1-L0-IMD-021) - SOMATOM Drive
- 2 1.6
4 32A 4 125A 4 125A
7
7
8 9 10 12 13
A2 E2
16 16 Contactor relay 24V DC 3RH2131-2BB40
intern 1m
max. 2,5 mm² flexible PRE-INSTALLATION ELECTRICAL PLAN
1.5
RHD bare ceiling RFB unfinished floor level
3a 3f 3f - 13
max.
THIS DRAWING HAS BEEN CHECKED FOR COMPLETENESS, ACCURACY, COORDINATION, AND
250 VAC / 2 A
K8 19 Auxiliary switch changeover contacts Siemens 3VA9988-0AA12 24 VDC / 2 A ABGD suspended ceiling FFL finished floor level COMPLIANCE WITH THE CONTRACT DOCUMENTS. ALL DIMENSIONS AND FIELD CONDITIONS
3R 3R K6 HAVE BEEN VERIFIED.
+ EQUIPMENT
K8 23 14 20 Contactor relay 24VDC 3RT1017-1BB44-3MA0 (K6) aABGD above suspended ceiling AFFL above finished floor level
ROOM
Power Distribution Cabinet (PDC)
1 23 Contactor relay 24V DC 3RH2131-2BB40 (K8)
aD on ceiling / suspended ceiling aFFL on finished floor / raised floor CONTRACTOR'S REPRESENTATIVE:
Mr. ALAADDIN AKTAN
DATE:
20 Sept 2022
X20 5.2
X20 2.3
X20 2.4
E
N
B07 B08
S
N
LA
.63
W
Connection of cooling system (schematic)
QIB
252
not to scale
B14
B09 B05
Scope of delivery - Cooling system B02
B01
B03
1. Standard:
B06
B12
800 - Flexible water hoses (Ø 34 mm) with thermal insulation (Ø 58 mm); available lenghts 10, 20 and 30 m; B11
B13
min. bending radius: 150 mm, shorten excess cable lengths; preordering is possible B04
B10
1640
1670
necessary. In this case the outer diameter of the thermal insulation won't fit the standard cable
1842
150 ± 0.00 TI TI
TI Thermometer
• Shorten excess cable length to size. MEDICAL EQUIPMENT SELECTIONS. SOME AREAS ARE SUBJECT TO CHANGE AFTER MATERIAL
SUBMITTALS APPROVAL .
1760 On-site • Extension up to 20 m possible; extension material must be provided on-side.
-0.10 cooling Feed water flow meter • Minimum flow rate: 2 l/h at 75% relative humidity
CT ROOM 2. Option
500
water supply
water
p p - Fixed piping kit:
300
FK 300x80
cabling
• includes the installation material to connect the delivered flexible cooling hoses with on-site fixed piping
Z1-L0-IMD-021 • 4x connection pieces (DKOL DN25 VA)
• 4x adaptors M36x2 metric to G1" internal thread
KEY PLAN 1
10 10.1 17 17.1
26
A A
549
840
Water return Water inlet Z1 Z3
728
P2.1 P3.1
Z2
FK 300x80mm
F F
2220
P2.A P3.A
GR
on-site
Water connection pieces with shut-off valve not to scale P2.F P3.F
Supply provided by customer G1" external thread (in scope of delivery) P2.1
Z4 Z5
P3.1
3093
2310
150
Adapter fitting for hose G1" internal thread 2 6 13 13.1 21
in scope of delivery
1960
1
REFERENCE DOCUMENTS
DOCUMENT NO TITLE REV
AT m BUILDINGS 1 - ALL ZONES - ALL PARTS - OVERALL -GROUND FLOOR -
in. MMC-HBK-AP-DD-BAP-MEQ-L0-APO-1800
MEDICAL EQUIPMENT PLAN
2
80
150x300mm IRS MMC-HBK-P1-DD-B01-MEQ-L0-P06-1806 BUILDING 1 - ZONE 01- PART 06- GROUND FLOOR MEDICAL EQUIPMENT PLAN 2
534
3
2810
The cooling water hoses Ø 34 mm 4
1530
1900 3100 (with insulation Ø 58 mm) from the gantry
for water connection are in scope of
AT CV delivery and available in 10, 20 or 30 m.
PDCB PDCA
MILITARY MEDICAL CITY PROJECT - AL SAILIYAH
Approval Status
Approved as
Commented
FK 200x80mm
Re-Submit
A Approved
D Rejected
Revise &
Condensed water hose to Authority Signature Date
e.g. a sink or a siphon 1
C
B
300
MMC-PMO
1 Hexagon coupling nut 2 Shut-off valve 3 Flexible fabric hose 4
275
GR Suitable for installation in surface mounted trunking, cable duct or false floor.
Water connection pieces The approval and comments shall not in any ways relieve or absolve the
with shut-off valve Contractor from compliance of any obligations or liability under or in connection
with the contract, project requirements and specifications.
Example for installation of water drain for condensed water not to scale
(schematic)
max.
Technical data of the interface heat exchanger
4m
Heat dissipation to cooling water [kW] 17 14 11 8 5 7 4 5
HBK CONTRACTING CO. WLL
Primary circuit components: Shut-off devices, filters, flow indicators, fill and drain valves, plate heat P.O. BOX 1362 DOHA, QATAR
Heat dissipation (Gantry and patient table) to air [kW] 1 1 1 1 1 exchanger, pressure gauge 3
TEL: (974) 44451111 FAX: (974) 44451000
Anti-freeze may be added to the on-site circuit. Max. mixing ratio up to -25°C: 40 % Connections: M36x2 / DN25 On-site water connection (schematical) not to scale
Adding anti-freeze reduces the cooling output and requires higher cooling water flow rate. Heat up brine to at least Internal pressure loss: ca. 0.5 bar MEDICAL PLANNER: MEDICAL EQUIPMENT : MEP SUBCONTRACTOR : SUBCONTRACTOR :
+4 °C if the internal cooling system is connected to an existing on-site cooling system! Only then on-site water can Operating voltage: 3/N/PE ~400 VAC / ±10% - 50 Hz / ±1% 3
flow through the Gantry integrated cooling system. 3/N/PE ~460 VAC / ±10% - 60 Hz / ±1%
Power consumption, overall system: max. 0.6 kVA at 50 Hz 1 6 7
Switching off when the CT system is not in operation
For purposes of saving energy the on-site water circulation can be switched off with: max. 0.9 kVA at 60 Hz PROJECT TITLE:
- On-site installation of a solenoid valve or switch-off feed pump via contactor Fuse protection (provided by client): 10 A
2 4 5
- On-site feed pump ENGINEERING, PROCUREMENT AND
- Optional water/air-split cooling system 1 Water hose 2 M36 x 2 external thread 3 Adapter (in scope of delivery)
General information CONSTRUCTION (EPC/TURNKEY) FOR
A floating relay contact assembly is included in the PDC:
Maximum 277 VAC/2 A; utilization category AC-14/15 For the gantry-integrated cooling system a on-site cooling water system has to be provided.
4 G1" internal thread 5 G1" external thread 6 On-site water connection 7 e. g. wall
MILITARY MEDICAL CITY HOSPITAL
Maximum 24 VDC/2 A; utilization category DC-12/13 The cooling water system will be connected via:
- Flexible water hoses (in scope of delivery, orderable lengths 10, 20 or 30 m) and special water connection pieces.
(MMC) AT TAREK BIN ZEYAD CAMP
Minimum flow rate - On-site cooling circuit - On-site fixed pipes (PVC, copper, high-grade steel). Therefore a "Fixed piping kit" can be ordered. For vibration IN SAILIYAH (SOUTH OF DOHA)
reduction the flexible water connection pieces have to be installed between the fixed piping and the water Piping material
Barometric pressure 700 to 1060 hPa
Pressure loss - On-site cooling circuit 2 connection points of the gantry. The direct connection of fixed piping and gantry water connection points is not Recommended Not recommended
permitted!
Stainless steel (V2A; V4A) Aluminium
PROJECT NO. AA-103-01
Barometric pressure 700 to 1060 hPa 3500 l/h For service and filling procedure a water faucet with screw connection and a water drain (e.g. a sink or a siphon) in
2 3 Non-ferrous metal (e.g. copper, brass) Iron, carbon steel DRAWING TITLE:
4 3200 vicinity of the gantry is necessary.
3 3000 Synthetic materials, plastics Zinc-plated steel
BUILDING 01-ZONE 01, -PART 06 - GROUND FLOOR -
3000 l/h
2.2 bar CT SCAN ROOM (Z1-L0-IMD-021) - SOMATOM Drive
Brazing solder, hard solder Zinc PRE-INSTALLATION CHILLED WATER and DRAIN DETAILS
2.0 bar
2.05
Fitting solder types 3 and 4 Standard steel pipes
1.8 bar 2050 THIS DRAWING HAS BEEN CHECKED FOR COMPLETENESS, ACCURACY, COORDINATION, AND
2000 l/h COMPLIANCE WITH THE CONTRACT DOCUMENTS. ALL DIMENSIONS AND FIELD CONDITIONS
HAVE BEEN VERIFIED.
1.6 bar 1800
1.45 1600
1.4 bar
1300 1300
1.2 bar 970
1120 Legend
1030 Mr. ALAADDIN AKTAN 20 Sept 2022
1000 l/h CONTRACTOR'S REPRESENTATIVE: DATE:
0.95 740 850
1.0 bar 650 cut-out, LxW (i. e. in FK or raised floor) STATUS:
1500 l / h
500 l / h
3000 l / h
2000 l / h
2500 l / h
DESIGNED SCALE A1
MEINHARDT 1:30
Equipment
Standard IEC 61223-2-6 (Ed.2) Customer Name Customer
Tester Name sm Customer ID Id
Manufacturer Siemens Healthcare GmbH Hospital Name 105781
Product Name SOMATOM Drive Street Street StreetNo
Software Version syngo CT VB20A Zip Code ZIP
Serial Number 105781 City City
Handed over 2023-01-25 District District
Country QAT
Customer Phone Number Phone
Serial Number
Component Serial Number Revision Level
Tube Assembly A 691022371 -
Tube Assembly B 684512271 -
Description
For the test of the lightmarkers the system phantom is centered and aligned, so that the lightmarkers hit the engraved
notches on the phantom. From image of the phantom, scanned with narrow collimation, the congruence of the
engraved notches with the scan plane as well as the saggital and coronal planes are tested.
Last IEC Constancy Reference measurement: 2023-10-11 09:27
(values stated in brackets)
Slice phantom -
Test:
Test Results OK
Result [mm]
0.00
Tolerance
-1.00 ... 1.00
Equipment
Standard IEC 61223-2-6 (Ed.2) Customer Name Customer
Tester Name sm Customer ID Id
Manufacturer Siemens Healthcare GmbH Hospital Name 105781
Product Name SOMATOM Drive Street Street StreetNo
Software Version syngo CT VB20A Zip Code ZIP
Serial Number 105781 City City
Handed over 2023-01-25 District District
Country QAT
Customer Phone Number Phone
Serial Number
Component Serial Number Revision Level
Tube Assembly A 691022371 -
Tube Assembly B 684512271 -
Description
For the test of the lightmarkers the system phantom is centered and aligned, so that the lightmarkers hit the engraved
notches on the phantom. From image of the phantom, scanned with narrow collimation, the congruence of the
engraved notches with the scan plane as well as the saggital and coronal planes are tested.
Last IEC Constancy Reference measurement: 2023-10-11 09:30
(values stated in brackets)
Slice phantom -
Test:
Test Results OK
Result
Sagittal Light Marker Coronal Light Marker
Result 0.00 mm 0.00 mm
Tolerance -1.00 ... 1.00 mm -1.00 ... 1.00 mm
Equipment
Standard IEC 61223-2-6 (Ed.2) Customer Name Customer
Tester Name sm Customer ID Id
Manufacturer Siemens Healthcare GmbH Hospital Name 105781
Product Name SOMATOM Drive Street Street StreetNo
Software Version syngo CT VB20A Zip Code ZIP
Serial Number 105781 City City
Handed over 2023-01-25 District District
Country QAT
Customer Phone Number Phone
Serial Number
Component Serial Number Revision Level
Tube Assembly A 691022371 -
Tube Assembly B 684512271 -
Description
In the Topogram image of the system phantom the position of an tomogram scan with narrow collimation is selected
for the engraved notches in the phantom. From the recorded tomogram image of the phantom the congruence of
the engraved notches with the scan plane is inspected.
Last IEC Constancy Reference measurement: 2023-10-11 09:31
(values stated in brackets)
Slice phantom -
Test:
Test Results OK
Result [mm]
0.00
Tolerance
-2.00 ... 2.00
Equipment
Standard IEC 61223-2-6 (Ed.2) Customer Name Customer
Tester Name sm Customer ID Id
Manufacturer Siemens Healthcare GmbH Hospital Name 105781
Product Name SOMATOM Drive Street Street StreetNo
Software Version syngo CT VB20A Zip Code ZIP
Serial Number 105781 City City
Handed over 2023-01-25 District District
Country QAT
Customer Phone Number Phone
Serial Number
Component Serial Number Revision Level
Tube Assembly A 691022371 -
Tube Assembly B 684512271 -
Description
The tomographic section thickness is evaluated by measuring the width of the image of an inclined aluminium ramp
at the intersection of the ramp with the slice. The width is defined as the Full Width at Half-Maximum (FWHM) of the
baseline corrected CT-value profile. The influence of the reconstruction kernel and the influence of the cone angle
on the measured width are taken into account. The measurement is performed for two sets of scan parameters,
representing a typical head- and a typical body mode. The evaluation is performed for at least the both outer and
one central slice. The measurement and evaluation is performed separately for system A and B.
Last IEC Constancy Reference measurement: 2023-10-11 09:33
(values stated in brackets)
Slice phantom -
Common Parameters
Scan Type Sequence
Voltage 120 kV
UHR no
Shaped filter Wide
Tin filter no
Window width 500 HU
Window center 50 HU
FoV 120 mm
Result
System A
Slice 1 4.69 mm (4.62)
Tolerance 3.80 ... 5.62 mm
Result
System A
Slice 1 5.11 mm (5.05)
Tolerance 4.05 ... 6.00 mm
Result
System A System B
Slice 1 5.08 mm (5.05) 5.02 mm (5.04)
Tolerance 4.05 ... 6.00 mm 4.04 ... 6.00 mm
Equipment
Standard IEC 61223-2-6 (Ed.2) Customer Name Customer
Tester Name sm Customer ID Id
Manufacturer Siemens Healthcare GmbH Hospital Name 105781
Product Name SOMATOM Drive Street Street StreetNo
Software Version syngo CT VB20A Zip Code ZIP
Serial Number 105781 City City
Handed over 2023-01-25 District District
Country QAT
Customer Phone Number Phone
Serial Number
Component Serial Number Revision Level
Tube Assembly A 691022371 -
Tube Assembly B 684512271 -
Description
The measurement is performed for three sets of scan parameters, representing a typical head-mode, a typical body
mode and a body-dual-source mode. For the dual-source mode the evaluation is performed separately for system A
and B. The water value and homogeneity is evaluated by measuring the mean CT number in five Regions Of Interest
(ROI) in the image of a cylindrical water phantom with diameter 20 cm. The ROI are located in the center and near
the edge of the water phantom at 12 o'clock, 3 o'clock, 6 o'clock and 9 o'clock. The water value is the CT-number of
the central ROI, the homogeneity is the difference between the outer ROI and the central ROI each.
Last IEC Constancy Reference measurement: 2023-10-11 09:33
(values stated in brackets)
Water phantom -
Water phantom large -
Common Parameters
Scan Type Sequence
Voltage 120 kV
UHR no
Shaped filter Wide
Tin filter no
Window width 500 HU
Window center 0 HU
FoV 250 mm
Slice 2 1.75 HU (1.71) 0.32 HU (0.44) 0.27 HU (0.33) 0.31 HU (0.42) -0.12 HU (-0.10)
Tolerance -2.29 ... 4.00 HU -1.56 ... 2.44 HU -1.67 ... 2.33 HU -1.58 ... 2.42 HU -2.10 ... 1.90 HU
Slice 3 2.10 HU (2.04) -0.24 HU (-0.12) -0.23 HU (0.15) -0.16 HU (-0.12) -0.16 HU (-0.31)
Tolerance -1.96 ... 4.00 HU -2.12 ... 1.88 HU -1.85 ... 2.15 HU -2.12 ... 1.88 HU -2.31 ... 1.69 HU
Slice 4 2.47 HU (2.06) -0.54 HU (-0.36) -0.43 HU (0.03) -0.37 HU (0.22) -0.60 HU (-0.09)
Tolerance -1.94 ... 4.00 HU -2.36 ... 1.64 HU -1.97 ... 2.03 HU -1.78 ... 2.22 HU -2.09 ... 1.91 HU
Slice 5 1.85 HU (1.63) 0.29 HU (0.47) 0.18 HU (0.35) 0.32 HU (0.52) 0.03 HU (0.35)
Tolerance -2.37 ... 4.00 HU -1.53 ... 2.47 HU -1.65 ... 2.35 HU -1.48 ... 2.52 HU -1.65 ... 2.35 HU
Slice 6 1.88 HU (1.61) 0.06 HU (0.14) 0.14 HU (0.37) 0.18 HU (0.38) -0.31 HU (0.13)
Tolerance -2.39 ... 4.00 HU -1.86 ... 2.14 HU -1.63 ... 2.37 HU -1.62 ... 2.38 HU -1.87 ... 2.13 HU
Slice 7 2.03 HU (1.59) -0.33 HU (-0.09) -0.13 HU (0.34) -0.12 HU (0.06) -0.45 HU (-0.09)
Tolerance -2.41 ... 4.00 HU -2.09 ... 1.91 HU -1.66 ... 2.34 HU -1.94 ... 2.06 HU -2.09 ... 1.91 HU
Slice 2 1.84 HU (1.86) 0.47 HU (0.59) 0.81 HU (0.40) 0.79 HU (0.34) 0.97 HU (1.21)
Tolerance -4.14 ... 6.00 HU -3.41 ... 4.59 HU -3.60 ... 4.40 HU -3.66 ... 4.34 HU -2.79 ... 5.21 HU
Slice 3 2.80 HU (3.49) -0.06 HU (-0.93) -0.25 HU (-1.22) -0.38 HU (-1.01) 0.08 HU (-0.11)
Tolerance -2.51 ... 6.00 HU -4.93 ... 3.07 HU -5.22 ... 2.78 HU -5.01 ... 2.99 HU -4.11 ... 3.89 HU
Slice 4 3.00 HU (3.11) -0.52 HU (-0.37) -0.42 HU (-0.25) -0.11 HU (-0.54) -0.09 HU (-0.09)
Tolerance -2.89 ... 6.00 HU -4.37 ... 3.63 HU -4.25 ... 3.75 HU -4.54 ... 3.46 HU -4.09 ... 3.91 HU
Slice 5 3.45 HU (3.08) -0.63 HU (-0.45) -0.62 HU (-0.20) -0.51 HU (-0.56) -0.50 HU (0.09)
Tolerance -2.92 ... 6.00 HU -4.45 ... 3.55 HU -4.20 ... 3.80 HU -4.56 ... 3.44 HU -3.91 ... 4.09 HU
Slice 6 3.09 HU (3.20) -0.38 HU (-0.80) -0.37 HU (-0.85) -0.79 HU (-0.93) -0.36 HU (-0.49)
Tolerance -2.80 ... 6.00 HU -4.80 ... 3.20 HU -4.85 ... 3.15 HU -4.93 ... 3.07 HU -4.49 ... 3.51 HU
Slice 7 2.33 HU (2.44) -0.22 HU (-0.41) 0.16 HU (-0.41) 0.12 HU (-0.42) 0.54 HU (0.36)
Tolerance -3.56 ... 6.00 HU -4.41 ... 3.59 HU -4.41 ... 3.59 HU -4.42 ... 3.58 HU -3.64 ... 4.36 HU
Slice 2 1.01 HU (1.16) 1.51 HU (1.52) 1.53 HU (1.47) 1.48 HU (1.21) 1.84 HU (2.21)
Tolerance -4.84 ... 6.00 HU -2.48 ... 5.52 HU -2.53 ... 5.47 HU -2.79 ... 5.21 HU -1.79 ... 6.00 HU
Slice 3 1.09 HU (0.87) 1.52 HU (1.90) 1.61 HU (1.82) 1.34 HU (1.90) 2.13 HU (2.05)
Tolerance -5.13 ... 6.00 HU -2.10 ... 5.90 HU -2.18 ... 5.82 HU -2.10 ... 5.90 HU -1.95 ... 6.00 HU
Slice 4 0.55 HU (0.84) 2.06 HU (1.96) 1.96 HU (2.09) 2.19 HU (1.89) 2.56 HU (2.31)
Tolerance -5.16 ... 6.00 HU -2.04 ... 5.96 HU -1.91 ... 6.00 HU -2.11 ... 5.89 HU -1.69 ... 6.00 HU
Slice 5 0.15 HU (0.81) 2.11 HU (1.65) 2.69 HU (1.72) 2.20 HU (1.61) 2.95 HU (2.42)
Tolerance -5.19 ... 6.00 HU -2.35 ... 5.65 HU -2.28 ... 5.72 HU -2.39 ... 5.61 HU -1.58 ... 6.00 HU
Slice 6 -0.10 HU (1.16) 2.58 HU (1.39) 2.66 HU (1.65) 2.75 HU (1.42) 3.06 HU (1.98)
Tolerance -4.84 ... 6.00 HU -2.61 ... 5.39 HU -2.35 ... 5.65 HU -2.58 ... 5.42 HU -2.02 ... 5.98 HU
Slice 7 0.32 HU (0.51) 2.18 HU (1.90) 2.10 HU (2.01) 2.22 HU (2.15) 2.52 HU (2.12)
Tolerance -5.49 ... 6.00 HU -2.10 ... 5.90 HU -1.99 ... 6.00 HU -1.85 ... 6.00 HU -1.88 ... 6.00 HU
Slice 2 1.58 HU (1.47) 0.88 HU (1.07) 0.77 HU (1.04) 0.67 HU (0.95) 0.46 HU (0.67)
Tolerance -2.53 ... 4.00 HU -0.93 ... 3.07 HU -0.96 ... 3.04 HU -1.05 ... 2.95 HU -1.33 ... 2.67 HU
Slice 3 1.84 HU (1.59) 0.44 HU (0.58) 0.46 HU (0.98) 0.55 HU (0.87) 0.14 HU (0.63)
Tolerance -2.41 ... 4.00 HU -1.42 ... 2.58 HU -1.02 ... 2.98 HU -1.13 ... 2.87 HU -1.37 ... 2.63 HU
Equipment
Standard IEC 61223-2-6 (Ed.2) Customer Name Customer
Tester Name sm Customer ID Id
Manufacturer Siemens Healthcare GmbH Hospital Name 105781
Product Name SOMATOM Drive Street Street StreetNo
Software Version syngo CT VB20A Zip Code ZIP
Serial Number 105781 City City
Handed over 2023-01-25 District District
Country QAT
Customer Phone Number Phone
Serial Number
Component Serial Number Revision Level
Tube Assembly A 691022371 -
Tube Assembly B 684512271 -
Description
The measurement is performed for three sets of scan parameters, representing a typical head-mode, a typical body
mode and a body dual-source mode. For the dual-source mode the evaluation is performed separately for system
A and B. The image noise is evaluated measuring the standard deviation of CT numbers in a Region Of Interest
(ROI) in the image of a cylindrical water phantom with diameter 20 cm. The ROI is located in the phantom center.
To suppress the influence of image inhomogeneity and ring artifacts, the evaluation is performed for the difference
image of two succeeding scans.
Last IEC Constancy Reference measurement: 2023-10-11 09:35
(values stated in brackets)
Water phantom -
Water phantom large -
Common Parameters
Scan Type Sequence
Voltage 120 kV
UHR no
Shaped filter Wide
Tin filter no
Window width 500 HU
Window center 0 HU
FoV 250 mm
Result
Results for System A
Slice 1 3.29 HU (3.31)
Tolerance 2.98 ... 3.54 HU
Result
Results for System A
Slice 1 4.72 HU (4.80)
Tolerance 4.32 ... 5.23 HU
Result
Results for System A Results for System B
Slice 1 4.73 HU (4.68) 4.71 HU (4.80)
Tolerance 4.25 ... 5.15 HU 4.32 ... 5.19 HU
Equipment
Standard IEC 61223-2-6 (Ed.2) Customer Name Customer
Tester Name sm Customer ID Id
Manufacturer Siemens Healthcare GmbH Hospital Name 105781
Product Name SOMATOM Drive Street Street StreetNo
Software Version syngo CT VB20A Zip Code ZIP
Serial Number 105781 City City
Handed over 2023-01-25 District District
Country QAT
Customer Phone Number Phone
Serial Number
Component Serial Number Revision Level
Tube Assembly A 691022371 -
Tube Assembly B 684512271 -
Description
The High Contrast Resolution is determined by evaluating the image of a thin tungsten wire placed in a plastic
rod, centered in the scan plane and aligned along the scanner axis. From the images of the wire the point spread
function is obtained and the Modulation Transfer Function (MTF) is calculated as the Fourier Transformation of the
point spread function. The 50% and 10% value of the MTF are evaluated. The measurement is performed for four
to five sets of scan parameters, representing a typical head-mode, a typical body mode, a dual-source mode, the
sharpest mode without UHR comb and, if available, the sharpest mode with UHR-comb. For the dual-source mode
the evaluation is performed separately for system A and B.
Last IEC Constancy Reference measurement: 2023-10-11 09:36
(values stated in brackets)
Wire phantom -
Wire Phantom Type Wire in air
Common Parameters
Scan Type Sequence
Voltage 120 kV
UHR no
Shaped filter Wide
Tin filter no
Window width 100 HU
Window center 30 HU
FoV 50 mm
Equipment
Standard IEC 61223-2-6 (Ed.2) Customer Name Customer
Tester Name sm Customer ID Id
Manufacturer Siemens Healthcare GmbH Hospital Name 105781
Product Name SOMATOM Drive Street Street StreetNo
Software Version syngo CT VB20A Zip Code ZIP
Serial Number 105781 City City
Handed over 2023-01-25 District District
Country QAT
Customer Phone Number Phone
Serial Number
Component Serial Number Revision Level
Tube Assembly A 691022371 -
Tube Assembly B 684512271 -
Description
The patient table is automatically moved forward and backward over 30 cm in continuous movement and stepwise
movement. After each series of movements the distance is measured with a ruler.
Last IEC Constancy Reference measurement: 2023-10-11 09:39
(values stated in brackets)
Test:
Test Results OK
Equipment
Standard IEC 61223-2-6 (Ed.2) Customer Name Customer
Tester Name sm Customer ID Id
Manufacturer Siemens Healthcare GmbH Hospital Name 105781
Product Name SOMATOM Drive Street Street StreetNo
Software Version syngo CT VB20A Zip Code ZIP
Serial Number 105781 City City
Handed over 2023-01-25 District District
Country QAT
Customer Phone Number Phone
Serial Number
Component Serial Number Revision Level
Tube Assembly A 691022371 -
Tube Assembly B 684512271 -
Description
The CTDIfreeAir is evaluated from the measured Dose-Length-Product (over 10 cm) in a typical head-mode, a typical
body mode (A-System) and a scan with the typical body mode parameter in the B-system.
Last IEC Constancy Reference measurement: 2023-10-12 09:48
(values stated in brackets)
Dosimeter -
CT-Ionisation Chamber -
Common Parameters
Scan Type Sequence
UHR no
Test:
Test Parameters Test Results OK
System A DLP [mGycm] Result [mGy/100mAs]
Test:
Test Parameters Test Results OK
System B DLP [mGycm] Result [mGy/100mAs]
Equipment
Standard IEC 61223-2-6 (Ed.2) Customer Name Customer
Tester Name AR Customer ID Id
Manufacturer Siemens Healthcare GmbH Hospital Name 105779
Product Name SOMATOM Drive Street Street StreetNo
Software Version syngo CT VB20A Zip Code ZIP
Serial Number 105779 City City
Handed over 2023-01-26 District District
Country QAT
Customer Phone Number Phone
Serial Number
Component Serial Number Revision Level
Tube Assembly A 647422271 -
Tube Assembly B 633412271 -
Description
For the test of the lightmarkers the system phantom is centered and aligned, so that the lightmarkers hit the engraved
notches on the phantom. From image of the phantom, scanned with narrow collimation, the congruence of the
engraved notches with the scan plane as well as the saggital and coronal planes are tested.
Last IEC Constancy Reference measurement: 2023-10-12 08:53
(values stated in brackets)
Slice phantom -
Test:
Test Results OK
Result [mm]
0.00
Tolerance
-1.00 ... 1.00
Equipment
Standard IEC 61223-2-6 (Ed.2) Customer Name Customer
Tester Name AR Customer ID Id
Manufacturer Siemens Healthcare GmbH Hospital Name 105779
Product Name SOMATOM Drive Street Street StreetNo
Software Version syngo CT VB20A Zip Code ZIP
Serial Number 105779 City City
Handed over 2023-01-26 District District
Country QAT
Customer Phone Number Phone
Serial Number
Component Serial Number Revision Level
Tube Assembly A 647422271 -
Tube Assembly B 633412271 -
Description
For the test of the lightmarkers the system phantom is centered and aligned, so that the lightmarkers hit the engraved
notches on the phantom. From image of the phantom, scanned with narrow collimation, the congruence of the
engraved notches with the scan plane as well as the saggital and coronal planes are tested.
Last IEC Constancy Reference measurement: 2023-10-12 09:04
(values stated in brackets)
Slice phantom -
Test:
Test Results OK
Result
Sagittal Light Marker Coronal Light Marker
Result 0.00 mm 0.00 mm
Tolerance -1.00 ... 1.00 mm -1.00 ... 1.00 mm
Equipment
Standard IEC 61223-2-6 (Ed.2) Customer Name Customer
Tester Name AR Customer ID Id
Manufacturer Siemens Healthcare GmbH Hospital Name 105779
Product Name SOMATOM Drive Street Street StreetNo
Software Version syngo CT VB20A Zip Code ZIP
Serial Number 105779 City City
Handed over 2023-01-26 District District
Country QAT
Customer Phone Number Phone
Serial Number
Component Serial Number Revision Level
Tube Assembly A 647422271 -
Tube Assembly B 633412271 -
Description
In the Topogram image of the system phantom the position of an tomogram scan with narrow collimation is selected
for the engraved notches in the phantom. From the recorded tomogram image of the phantom the congruence of
the engraved notches with the scan plane is inspected.
Last IEC Constancy Reference measurement: 2023-10-12 08:55
(values stated in brackets)
Slice phantom -
Test:
Test Results OK
Result [mm]
0.00
Tolerance
-2.00 ... 2.00
Equipment
Standard IEC 61223-2-6 (Ed.2) Customer Name Customer
Tester Name AR Customer ID Id
Manufacturer Siemens Healthcare GmbH Hospital Name 105779
Product Name SOMATOM Drive Street Street StreetNo
Software Version syngo CT VB20A Zip Code ZIP
Serial Number 105779 City City
Handed over 2023-01-26 District District
Country QAT
Customer Phone Number Phone
Serial Number
Component Serial Number Revision Level
Tube Assembly A 647422271 -
Tube Assembly B 633412271 -
Description
The tomographic section thickness is evaluated by measuring the width of the image of an inclined aluminium ramp
at the intersection of the ramp with the slice. The width is defined as the Full Width at Half-Maximum (FWHM) of the
baseline corrected CT-value profile. The influence of the reconstruction kernel and the influence of the cone angle
on the measured width are taken into account. The measurement is performed for two sets of scan parameters,
representing a typical head- and a typical body mode. The evaluation is performed for at least the both outer and
one central slice. The measurement and evaluation is performed separately for system A and B.
Last IEC Constancy Reference measurement: 2023-10-12 08:56
(values stated in brackets)
Slice phantom -
Common Parameters
Scan Type Sequence
Voltage 120 kV
UHR no
Shaped filter Wide
Tin filter no
Window width 500 HU
Window center 50 HU
FoV 120 mm
Result
System A
Slice 1 4.72 mm (4.73)
Tolerance 3.80 ... 5.73 mm
Result
System A
Slice 1 5.12 mm (5.17)
Tolerance 4.17 ... 6.00 mm
Result
System A System B
Slice 1 5.01 mm (5.02) 5.08 mm (5.20)
Tolerance 4.02 ... 6.00 mm 4.20 ... 6.00 mm
Equipment
Standard IEC 61223-2-6 (Ed.2) Customer Name Customer
Tester Name AR Customer ID Id
Manufacturer Siemens Healthcare GmbH Hospital Name 105779
Product Name SOMATOM Drive Street Street StreetNo
Software Version syngo CT VB20A Zip Code ZIP
Serial Number 105779 City City
Handed over 2023-01-26 District District
Country QAT
Customer Phone Number Phone
Serial Number
Component Serial Number Revision Level
Tube Assembly A 647422271 -
Tube Assembly B 633412271 -
Description
The measurement is performed for three sets of scan parameters, representing a typical head-mode, a typical body
mode and a body-dual-source mode. For the dual-source mode the evaluation is performed separately for system A
and B. The water value and homogeneity is evaluated by measuring the mean CT number in five Regions Of Interest
(ROI) in the image of a cylindrical water phantom with diameter 20 cm. The ROI are located in the center and near
the edge of the water phantom at 12 o'clock, 3 o'clock, 6 o'clock and 9 o'clock. The water value is the CT-number of
the central ROI, the homogeneity is the difference between the outer ROI and the central ROI each.
Last IEC Constancy Reference measurement: 2023-10-12 08:57
(values stated in brackets)
Water phantom -
Water phantom large -
Common Parameters
Scan Type Sequence
Voltage 120 kV
UHR no
Shaped filter Wide
Tin filter no
Window width 500 HU
Window center 0 HU
FoV 250 mm
Slice 2 0.08 HU (-0.11) 0.35 HU (0.70) 0.87 HU (1.10) 0.37 HU (0.72) 0.49 HU (0.45)
Tolerance -4.00 ... 3.89 HU -1.30 ... 2.70 HU -0.90 ... 3.10 HU -1.28 ... 2.72 HU -1.55 ... 2.45 HU
Slice 3 0.86 HU (0.59) -0.35 HU (-0.21) -0.35 HU (-0.33) -0.61 HU (-0.02) -0.83 HU (-0.42)
Tolerance -3.41 ... 4.00 HU -2.21 ... 1.79 HU -2.33 ... 1.67 HU -2.02 ... 1.98 HU -2.42 ... 1.58 HU
Slice 4 0.65 HU (0.28) 0.09 HU (-0.04) 0.00 HU (0.36) -0.20 HU (0.31) -0.39 HU (-0.26)
Tolerance -3.72 ... 4.00 HU -2.04 ... 1.96 HU -1.64 ... 2.36 HU -1.69 ... 2.31 HU -2.26 ... 1.74 HU
Slice 5 0.58 HU (0.12) 0.19 HU (0.47) 0.19 HU (0.46) -0.01 HU (0.69) -0.02 HU (0.37)
Tolerance -3.88 ... 4.00 HU -1.53 ... 2.47 HU -1.54 ... 2.46 HU -1.31 ... 2.69 HU -1.63 ... 2.37 HU
Slice 6 0.62 HU (0.46) -0.30 HU (0.02) -0.22 HU (0.37) -0.10 HU (0.04) -0.19 HU (-0.23)
Tolerance -3.54 ... 4.00 HU -1.98 ... 2.02 HU -1.63 ... 2.37 HU -1.96 ... 2.04 HU -2.23 ... 1.77 HU
Slice 7 0.58 HU (0.15) -0.22 HU (0.53) 0.33 HU (0.48) -0.03 HU (0.37) -0.19 HU (0.20)
Tolerance -3.85 ... 4.00 HU -1.47 ... 2.53 HU -1.52 ... 2.48 HU -1.63 ... 2.37 HU -1.80 ... 2.20 HU
Slice 2 1.54 HU (1.23) -0.30 HU (0.18) -0.08 HU (0.37) -0.41 HU (0.31) 0.21 HU (0.24)
Tolerance -4.77 ... 6.00 HU -3.82 ... 4.18 HU -3.63 ... 4.37 HU -3.69 ... 4.31 HU -3.76 ... 4.24 HU
Slice 3 1.69 HU (1.79) -0.71 HU (-0.64) -0.68 HU (-0.45) -0.63 HU (-0.81) -0.57 HU (-0.44)
Tolerance -4.21 ... 6.00 HU -4.64 ... 3.36 HU -4.45 ... 3.55 HU -4.81 ... 3.19 HU -4.44 ... 3.56 HU
Slice 4 2.20 HU (1.57) -1.11 HU (-0.15) -1.13 HU (-0.21) -1.05 HU (-0.21) -1.01 HU (0.09)
Tolerance -4.43 ... 6.00 HU -4.15 ... 3.85 HU -4.21 ... 3.79 HU -4.21 ... 3.79 HU -3.91 ... 4.09 HU
Slice 5 1.94 HU (2.02) -0.59 HU (-0.47) -0.84 HU (-0.33) -0.86 HU (-0.70) -0.32 HU (-0.29)
Tolerance -3.98 ... 6.00 HU -4.47 ... 3.53 HU -4.33 ... 3.67 HU -4.70 ... 3.30 HU -4.29 ... 3.71 HU
Slice 6 1.94 HU (2.02) -0.87 HU (-0.84) -0.62 HU (-0.75) -0.77 HU (-0.78) -0.81 HU (-0.25)
Tolerance -3.98 ... 6.00 HU -4.84 ... 3.16 HU -4.75 ... 3.25 HU -4.78 ... 3.22 HU -4.25 ... 3.75 HU
Slice 7 1.46 HU (1.30) -0.50 HU (-0.28) -0.40 HU (-0.19) -0.50 HU (-0.43) -0.42 HU (0.09)
Tolerance -4.70 ... 6.00 HU -4.28 ... 3.72 HU -4.19 ... 3.81 HU -4.43 ... 3.57 HU -3.91 ... 4.09 HU
Slice 2 -0.55 HU (-0.07) 2.55 HU (2.32) 2.46 HU (2.36) 2.44 HU (1.85) 2.75 HU (2.10)
Tolerance -6.00 ... 5.93 HU -1.68 ... 6.00 HU -1.64 ... 6.00 HU -2.15 ... 5.85 HU -1.90 ... 6.00 HU
Slice 3 -0.19 HU (-0.14) 1.83 HU (1.90) 2.14 HU (2.43) 2.09 HU (1.98) 2.42 HU (2.28)
Tolerance -6.00 ... 5.86 HU -2.10 ... 5.90 HU -1.57 ... 6.00 HU -2.02 ... 5.98 HU -1.72 ... 6.00 HU
Slice 4 -1.39 HU (-0.75) 2.92 HU (2.40) 3.23 HU (2.55) 2.81 HU (2.12) 3.42 HU (2.54)
Tolerance -6.00 ... 5.25 HU -1.60 ... 6.00 HU -1.45 ... 6.00 HU -1.88 ... 6.00 HU -1.46 ... 6.00 HU
Slice 5 -1.01 HU (-1.04) 2.57 HU (2.67) 3.02 HU (3.26) 2.69 HU (2.63) 3.16 HU (2.80)
Tolerance -6.00 ... 4.96 HU -1.33 ... 6.00 HU -0.74 ... 6.00 HU -1.37 ... 6.00 HU -1.20 ... 6.00 HU
Slice 6 -0.71 HU (-0.48) 2.36 HU (2.05) 2.43 HU (2.36) 2.33 HU (2.04) 2.89 HU (2.38)
Tolerance -6.00 ... 5.52 HU -1.95 ... 6.00 HU -1.64 ... 6.00 HU -1.96 ... 6.00 HU -1.62 ... 6.00 HU
Slice 7 -0.54 HU (-1.22) 1.74 HU (2.06) 1.74 HU (2.84) 1.82 HU (2.52) 2.11 HU (2.96)
Tolerance -6.00 ... 4.78 HU -1.94 ... 6.00 HU -1.16 ... 6.00 HU -1.48 ... 6.00 HU -1.04 ... 6.00 HU
Slice 2 -0.53 HU (-0.28) 1.08 HU (0.82) 1.23 HU (0.96) 1.16 HU (1.08) 0.92 HU (0.40)
Tolerance -4.00 ... 3.72 HU -1.18 ... 2.82 HU -1.04 ... 2.96 HU -0.92 ... 3.08 HU -1.60 ... 2.40 HU
Slice 3 0.14 HU (0.24) 0.71 HU (0.43) 0.46 HU (0.52) 0.52 HU (0.37) 0.07 HU (0.29)
Tolerance -3.76 ... 4.00 HU -1.57 ... 2.43 HU -1.48 ... 2.52 HU -1.63 ... 2.37 HU -1.71 ... 2.29 HU
Equipment
Standard IEC 61223-2-6 (Ed.2) Customer Name Customer
Tester Name AR Customer ID Id
Manufacturer Siemens Healthcare GmbH Hospital Name 105779
Product Name SOMATOM Drive Street Street StreetNo
Software Version syngo CT VB20A Zip Code ZIP
Serial Number 105779 City City
Handed over 2023-01-26 District District
Country QAT
Customer Phone Number Phone
Serial Number
Component Serial Number Revision Level
Tube Assembly A 647422271 -
Tube Assembly B 633412271 -
Description
The measurement is performed for three sets of scan parameters, representing a typical head-mode, a typical body
mode and a body dual-source mode. For the dual-source mode the evaluation is performed separately for system
A and B. The image noise is evaluated measuring the standard deviation of CT numbers in a Region Of Interest
(ROI) in the image of a cylindrical water phantom with diameter 20 cm. The ROI is located in the phantom center.
To suppress the influence of image inhomogeneity and ring artifacts, the evaluation is performed for the difference
image of two succeeding scans.
Last IEC Constancy Reference measurement: 2023-10-12 08:58
(values stated in brackets)
Water phantom -
Water phantom large -
Common Parameters
Scan Type Sequence
Voltage 120 kV
UHR no
Shaped filter Wide
Tin filter no
Window width 500 HU
Window center 0 HU
FoV 250 mm
Result
Results for System A
Slice 1 3.20 HU (3.19)
Tolerance 2.90 ... 3.51 HU
Result
Results for System A
Slice 1 4.86 HU (4.86)
Tolerance 4.37 ... 5.23 HU
Result
Results for System A Results for System B
Slice 1 4.88 HU (4.91) 4.81 HU (4.84)
Tolerance 4.42 ... 5.19 HU 4.36 ... 5.19 HU
Equipment
Standard IEC 61223-2-6 (Ed.2) Customer Name Customer
Tester Name AR Customer ID Id
Manufacturer Siemens Healthcare GmbH Hospital Name 105779
Product Name SOMATOM Drive Street Street StreetNo
Software Version syngo CT VB20A Zip Code ZIP
Serial Number 105779 City City
Handed over 2023-01-26 District District
Country QAT
Customer Phone Number Phone
Serial Number
Component Serial Number Revision Level
Tube Assembly A 647422271 -
Tube Assembly B 633412271 -
Description
The High Contrast Resolution is determined by evaluating the image of a thin tungsten wire placed in a plastic
rod, centered in the scan plane and aligned along the scanner axis. From the images of the wire the point spread
function is obtained and the Modulation Transfer Function (MTF) is calculated as the Fourier Transformation of the
point spread function. The 50% and 10% value of the MTF are evaluated. The measurement is performed for four
to five sets of scan parameters, representing a typical head-mode, a typical body mode, a dual-source mode, the
sharpest mode without UHR comb and, if available, the sharpest mode with UHR-comb. For the dual-source mode
the evaluation is performed separately for system A and B.
Last IEC Constancy Reference measurement: 2023-10-12 09:00
(values stated in brackets)
Wire phantom -
Wire Phantom Type Wire in air
Common Parameters
Scan Type Sequence
Voltage 120 kV
UHR no
Shaped filter Wide
Tin filter no
Window width 100 HU
Window center 30 HU
FoV 50 mm
Equipment
Standard IEC 61223-2-6 (Ed.2) Customer Name Customer
Tester Name AR Customer ID Id
Manufacturer Siemens Healthcare GmbH Hospital Name 105779
Product Name SOMATOM Drive Street Street StreetNo
Software Version syngo CT VB20A Zip Code ZIP
Serial Number 105779 City City
Handed over 2023-01-26 District District
Country QAT
Customer Phone Number Phone
Serial Number
Component Serial Number Revision Level
Tube Assembly A 647422271 -
Tube Assembly B 633412271 -
Description
The patient table is automatically moved forward and backward over 30 cm in continuous movement and stepwise
movement. After each series of movements the distance is measured with a ruler.
Last IEC Constancy Reference measurement: 2023-10-12 09:02
(values stated in brackets)
Test:
Test Results OK
Equipment
Standard IEC 61223-2-6 (Ed.2) Customer Name Customer
Tester Name AR Customer ID Id
Manufacturer Siemens Healthcare GmbH Hospital Name 105779
Product Name SOMATOM Drive Street Street StreetNo
Software Version syngo CT VB20A Zip Code ZIP
Serial Number 105779 City City
Handed over 2023-01-26 District District
Country QAT
Customer Phone Number Phone
Serial Number
Component Serial Number Revision Level
Tube Assembly A 647422271 -
Tube Assembly B 633412271 -
Description
The CTDIfreeAir is evaluated from the measured Dose-Length-Product (over 10 cm) in a typical head-mode, a typical
body mode (A-System) and a scan with the typical body mode parameter in the B-system.
Last IEC Constancy Reference measurement: 2023-10-12 09:23
(values stated in brackets)
Dosimeter -
CT-Ionisation Chamber -
Common Parameters
Scan Type Sequence
UHR no
Test:
Test Parameters Test Results OK
System A DLP [mGycm] Result [mGy/100mAs]
Test:
Test Parameters Test Results OK
System B DLP [mGycm] Result [mGy/100mAs]
ASIM
email=asim.rashid@siemens-
healthineers.com
Date: 2023.10.16 12:54:23 +03'00'
105779
105779 16/10/2023
105779 16/10/2023
105779 16/10/2023
105779 16/10/2023
105779 16/10/2023
105779 16/10/2023
105779 16/10/2023
ASIM
email=asim.rashid@siemens-
healthineers.com
Date: 2023.10.16 15:36:15 +03'00'
Miled
Digitally signed by Miled samir
DN: cn=Miled samir, c=DE,
o=Siemens,
samir
email=samir.miled@siemens-
healthineers.com
Date: 2023.10.15 23:06:30 +03'00'
Miled
Digitally signed by Miled samir
DN: cn=Miled samir, c=DE,
o=Siemens,
samir
email=samir.miled@siemens-
healthineers.com
Date: 2023.10.15 23:08:02 +03'00'
This is to confirm that
Prebin Alakadan
CT Basic Certification
Contents:
• X-Ray fundamentals
• CT basic knowledge
• CT basic application
• Basic Tune up
• Preventive maintenenace
• Safety issues
This certificate is automatically created by system on August 01, 2017 and requires no signature
www.siemens.com/healthineers
CERTIFICATE *
of Attendance
Siemens Healthineers
Global Training Center
Intranet: https://intranet.training.healthcare.siemens.com/
Internet: https://training.healthcare.siemens.com/
CERTIFICATE *
of Attendance
Siemens Healthineers
Global Training Center
Intranet: https://intranet.training.healthcare.siemens.com/
Internet: https://training.healthcare.siemens.com/
CT BASICS
CT1CTBASIC
Siemens Healthineers
Global Training Center
Intranet: https://intranet.training.healthcare.siemens.com
Internet: https://training.healthcare.siemens.com
www.siemens.com/healthineers
CERTIFICATE *
of Attendance
CT BASICS
CT1CTBASIC
from March 14, 2011 to March 25, 2011
at Erlangen, Training Center
Siemens Healthineers
Global Training Center
Intranet: https://intranet.training.healthcare.siemens.com/
Internet: https://training.healthcare.siemens.com/
Imaging & Dental Radiology Imaging Diagnostic Ground CT Scanner, Multi-Slice, 64-
CT ROOM Z1-L0-IMD-021 CTX0152 1 1-Fixed ABAM SIEMENS SOMATOM Drive 105781
Package Package Packages Imaging Floor 320 Slice
Imaging & Dental Radiology Imaging Diagnostic Ground CT Scanner, Multi-Slice, 64-
CT ROOM Z1-L0-IMD-025 CTX0152 1 1-Fixed ABAM SIEMENS SOMATOM Drive 105779
Package Package Packages Imaging Floor 320 Slice
Data sheet
SOMATOM Drive
Data sheet for syngo CT VB20
siemens-healthineers.com/somatom-drive
International version.
Not for distribution or use in the U.S.
SOMATOM Drive
2
syngo CT VB20 · SOMATOM Drive
3
SOMATOM Drive · syngo CT VB20
Highlights
110
90
IR 130
70 140
A B
70 & 80 kV
4
syngo CT VB20 · SOMATOM Drive
Highlights
Superfast Scanning helps eliminate motion artifacts The Adaptive 4D Spiral achieves whole-brain and
without increasing the dose, while breathing artifacts whole-organ coverage for perfusion imaging and
are removed with the most precise dose and the best enables low-dose CT angiography and dynamic CT.
in-plane temporal resolution. This also helps dose- Harness the potential to improve patient outcomes
sensitive patients, who need very low doses – but it does with 4D imaging that exceeds the detector width and
so without limiting scan speed and temporal resolution. allows you to act faster and be more confident than
ever in your treatment decisions. The new Adaptive
4D Spiral visualizes motion with a dynamic scan range
up to 480 mm on SOMATOM Drive.
In virtually all reconstructions, the FAST IRS (image The FAST 3D Camera captures the patient’s shape,
reconstruction system) can reduce wait time. For position, and height using three-dimensional data.
example, filtered back projection (FBP) reconstructions Thanks to infrared measurement, it also recognizes
are calculated with up to 70 fps (frames per second) and body contours: for example, patients are covered with
iterative reconstructions (IR) are calculated with 50 fps. blankets patient gowns. Algorithms utilize this data to
Using FAST IRS, the often numerous reconstructions in calculate the correct patient positioning. The zeroclick
polytrauma and emergency exams can be performed solution runs unique algorithms to support accurate
fast, when time till diagnosis really counts. and reproducible positioning: With FAST Isocentering,
you get the right dose modulation and consistent
images; FAST Range supports scanning the right body
region with no truncation; FAST Direction helps
safeguard the right scan direction.
5
SOMATOM Drive · syngo CT VB20
System Configuration
6
syngo CT VB20 · SOMATOM Drive
System Configuration
1
CARE Contrast III supports injectors classes up to CANOpen standard class IV
2
In clinical practice, the use of ADMIRE may reduce CT patient dose depending on the clinical task, patient size, anatomical location, and clinical practice.
A radiologist and a physicist should be consulted to determine the appropriate dose to obtain diagnostic image quality for the particular clinical task.
3
Turbo Flash Cardio is the high pitch spiral mode (formerly called Flash Spiral Cardio). Availability of the mode depends on the system configuration
4
deconfigurable
7
SOMATOM Drive · syngo CT VB20
System Hardware
Gantry
Aperture
78 cm / 30.71”
Scan field
• 50 cm / 19.69”
• 78 cm / 30.71”
Generator
Rotation time
0.28, 0.33, 0.5, 1.0 s Max. power
200 kW (2 x 100 kW)
Temporal resolution
75 ms temporal resolution independent of the IEC Requested kV Accuracy
heart rate (down to 37.5 ms using 2-segment ± 10%
reconstruction)
Typical kV Accuracy
Three laser light markers ± 1%
Coronal, sagittal, transversal laser light showing
the isocenter position of the scan plane
Ring light
A gantry ring, mood light (LED) in different, preset
and adjustable colors. Create a relaxing atmosphere
for your patients.
1
2 standard, 4 optional
8
syngo CT VB20 · SOMATOM Drive
System Hardware
z-Sharp technology
Over-
sampling
0.6 mm
64-row detector
128-slice acquisition electronics
2 x Straton MX Sigma X-ray tubes and Sigma
generators
The unique Straton X-ray tube utilizes an electron
Tube current range
beam that is accurately and rapidly deflected,
• Single source 20–800 mA
creating two precise focal spots alternating 4,608
• Dual source 40–1600 mA
times per second. This doubles the X-ray projections
High Power 70 reaching each detector element. The two overlapping
650 mA (@ 70 kV) projections result in an oversampling in the z-direction
and allow acquiring twice the number of slices per
High Power 80 detector row. The resulting measurements interleave
750 mA (@ 80 kV) half a detector slice width, doubling the scan
Tube voltage information without a corresponding increase
• 70 –140 kV in 10 kV steps in dose. This provides a scan speed independent of
• Automatically selected through CARE kV based spatial resolution in the z-direction of up to 0.30 mm
on patient body habitus and examination type and a corresponding reduction of spiral artifacts in
the daily clinical routine at any position within the
Tube anode heat storage capacity scan field.
0 MHU (0.6 MHU capacity combined with
7.3 MHU/min (5,400 kJ/min) cooling rate is Multifan principle with Flying Focal Spot
comparable to the performance of a conventional Computer-controlled monitoring of anode
tube with approximately 50 MHU (37,000 kJ) temperature
anode heat storage capacity)
Cooling rate
7.3 MHU/min
9
SOMATOM Drive · syngo CT VB20
System Hardware
Dose Shield
Additional protocol-dependent filtration
• 0.5 mm Al / Tube A and B
• Tin Filter / Tube A and B
10
syngo CT VB20 · SOMATOM Drive
System Hardware
X-ray
11
SOMATOM Drive · syngo CT VB20
System Hardware
1
Acquisition modes driven by z-Sharp technology
2
Turbo Flash Cardio is the high pitch spiral mode (formerly called Flash Spiral Cardio). Availability of the mode depends on the system configuration.
12
syngo CT VB20 · SOMATOM Drive
System Hardware
Table feed speed Turbo Flash mode1 Table feed speed Turbo Flash mode1
Up to 400 mm/s / 15.75 inch/s Up to 400 mm/s / 15.75 inch/s
Table feed speed ECG-triggered Turbo Flash mode1 Table feed speed ECG-triggered Turbo Flash mode1
Up to 458 mm/s / 18.03 inch/s Up to 458 mm/s / 18.03 inch/s
Distance between gantry front and table front Distance between gantry front and table front
386 mm / 15.2” 386 mm / 15.20”
Distance between Touch Panel front and Distance between Touch Panel front
table front and table front
348 mm / 13.7” 387 mm / 15.24”
Additional exchangeable
tabletops3
• High-capacity patient and trauma tabletop
• RTP tabletop
1
Turbo Flash Cardio is the high pitch spiral mode (formerly called Flash Spiral Cardio). Availability of the mode depends on the system configuration.
2
Optional with high-capacity tabletop
3
Not available for standard 2,000 mm patient table
13
SOMATOM Drive · syngo CT VB20
Workplace Overview
Image Reconstruction
Real-time display
Real-time image display (512 x 512) during
spiral acquisition
Raw Data
Slice thickness
0.4, 0.5 , 0.6–20 mm Capacity
(20 mm using dynamic multiscan) 4 TB
Recon field External USB 3.0 ports for quick and easy raw data
• 5–50 cm / 1.97”–16.69” storage are supported
• 5–78 cm / 1.97”–30.71”
Recon time
• Up to 70 fps with weighted filtered 3D
backprojection (WFBP) and z-Sharp technology
at full image quality
• Up to 50 fps1 with integrated Iterative
Reconstruction with z-Sharp technology
at full image quality
Recon matrix
512 x 512
HU scale
– 1,024 to + 3,071
Extended HU scale
– 10,240 to + 30,710
1
Optional with FAST IRS
14
syngo CT VB20 · SOMATOM Drive
syngo Workplace
Technical Specifications
Workplace AWP Image storage
• 3 x 1,000 GB (1 TB) physical
High-performance computer
• ~ 2,100,000 images
Quad Core 3.60 GHz1
Additional storage
Graphics accelerator
• DVD DICOM drive:
NVIDIA Quadro K1200 for fast 3D postprocessing
4.7 GB DVD media
Standard monitor 8,000 images
• 19” (48 cm) flat screen • Write-RW/+RW/-DL/Read
• 1,280 x 1,024 resolution • CD-R: 700 MB
• 1,024 x 1,024 image display matrix 1,100 images
• 0.29 mm pixel size • External USB 3.0 ports for quick and easy raw data
storage are supported. External USB memory stick
Additional monitor2 for image data
Yes
DICOM viewer
Dual monitor3 Included on each CD; automatically started on the
Yes viewer’s PC
RAM storage
16 GB
1
Or equivalent
2
Optional. Additional monitor for replication of primary monitor at remote location. Distance from host up to 30 m.
3
Optional. Dual monitor enables the simultaneous display of two scans on two monitors within the 3D task card, ideally used for comparison of follow-up studies
or native and contrast-enhanced scans.
15
SOMATOM Drive · syngo CT VB20
16
syngo CT VB20 · SOMATOM Drive
Reduction of data volume, since virtually all Evaluation of contrast enhancement in organs
diagnostic information is captured in 3D slices and tissues
Calculation of
• Time-density curves (up to 5 ROIs)
• Peak-enhancement images
• Time-to-peak images
1
Optional
2
Deconfigurable
3
Turbo Flash Cardio is the high pitch spiral mode (formerly called Flash Spiral Cardio). Availability of the mode depends on the system configuration.
17
SOMATOM Drive · syngo CT VB20
Window width and center freely selectable Parallel evaluation of more than
10 Regions of Interest
Single window
• Circle
Double window (e.g., bone/soft tissue) • Irregular
• Polygonal
Multiple window settings for multi-image display
Statistical evaluation
Organ-specific window settings, e.g., for soft tissue • Area/volume
and bones • Standard deviation
2D Postprocessing • Mean value
• Min./max. values
Image zoom and pan • Histogram
Image manipulations Profile cuts
• Averaging, subtraction • Horizontal
• Reversal of grayscale values • Vertical
• Mirroring • Oblique
Advanced image algorithms Distance measurement
• LCE: Low Contrast Enhancement for improving
low-contrast detectability Angle measurement
• HCE: High Contrast Enhancement for increased Online measurement of a 5 x 5 pixel size RoI
sharpness of high-contrast structures
• ASA: Advanced Smoothing Algorithm Freely selectable positioning of coordinate system
edge-preserving smoothing filter, dedicated
Crosshair
to cardiac exams
Image annotation and labeling
18
syngo CT VB20 · SOMATOM Drive
1
Most recent scanner software has to be installed
2
If the security patch needs to be installed on site by Siemens Healthineers charges may result. Service conditions apply
19
SOMATOM Drive · syngo CT VB20
20
syngo CT VB20 · SOMATOM Drive
DistinCT Imaging
Real-time topogram
CARE Dose4D
Scan with
constant mA
Reduced
dose level
based
on topogram
Direct scan parameter adjustment at the push
of a button Real-time
angular dose
FAST Scan Assistant modulation
CARE kV
CARE Filter
21
SOMATOM Drive · syngo CT VB20
DistinCT Imaging
• Dedicated pediatric CT imaging, including 70 kV Visualization of activated dose reduction features and
scan modes and specific CARE Dose4D curves technologies for each scan range of an examination
and protocols
• Special clinical protocols with 70 or 80 kV X-CARE
selection and a wide range of mAs settings. Angular tube current modulation to reduce X-ray
The X-ray exposure is adapted to the child’s exposure (organ dose) of peripheral dose-sensitive
(and small adult’s) weight and age, substantially body regions, e.g., the breasts, thyroid gland or
reducing the effective patient dose eye lens
SYNC.
START
Synchronized scanning and contrast injection through Accurate and automatically aligned preparation of
integration of CT scanner and injector facilitate spine recons with just a single click
enhanced CT examinations and improved workflow
iMAR (iterative Metal Artifact Reduction)
23
SOMATOM Drive · syngo CT VB20
FAST DE Results1
1
Optional
24
syngo CT VB20 · SOMATOM Drive
DistinCT Function –
Cardiac1
syngo HeartView CT FAST Phase
syngo HeartView CT with ECG-synchronized true Intuitive solution fully integrated in the cardiac
isotropic volume acquisition using prospective workflow for helping to quickly set up complex and
ECG-triggered or retrospective ECG-gating mode usually time-consuming cardiac procedures reliably
and repeatedly. By measuring of heart rate and
Basis for 3D cardiac scanning and reconstruction,
rhythm the system automatically chooses the most
e.g., CT angiography of the coronary and thoracic
appropriate phase of the heart cycle to scan and
vessels or calcium scoring
later to reconstruct based on fully configurable and
The ECG signal used for gating the CT images is customizable look up tables.
acquired by an integrated ECG device. The ECG signal
Reduction of complexity and standardization of
is displayed on the gantry front cover and the scan
cardiac acquisitions by making them more robust
interface
and less user-dependent
Dual Source acquisition mode with single-segment
syngo Cardio BestPhase Plus
reconstruction enables heart-rate-independent
temporal resolution of 75 ms (factor 2 x higher than Provides a proposal for the heart phase and 3D VRT
single-source acquisition with same parameters) visualization of the heart in several user-selectable
heart phases to help the physician choose the
Down to 37.5 ms temporal resolution combining
desired heart phase for image reconstruction
syngo HeartView acquisition with robust 2-segment
reconstruction (except Turbo Flash mode1)
1
Optional
2
Turbo Flash Cardio is the high pitch spiral mode (formerly called Flash Spiral Cardio). Availability of the mode depends on the system configuration.
25
SOMATOM Drive · syngo CT VB20
1
Optional
26
syngo CT VB20 · SOMATOM Drive
FAST Planning
FAST 3D Align
27
SOMATOM Drive · syngo CT VB20
Artifacts
Noise
FBP
Audit trail
Enables the evaluation of who has conducted or
altered studies, for example deleting parts of exam In modern healthcare, patients deserve the lowest
studies. possible dose, and radiologists desire the best image
quality in their daily routine. This is known as ALARA –
Groups and role the paradigm to deliver diagnostic image quality at
Supports the creation of individual user privileges a dose As Low As Reasonably Achievable. But the big
and restricts scanner usage according to institution’s challenge is to realize this in the daily routine, e.g.,
defined users. time-critical settings like acute care, making many
institutions step away from iterative reconstruction.
Patient groups
Enables the creation of specific patient groups to ADMIRE – Siemens Healthineers’ Advanced Modeled
protect sensitive patient information (e.g. data Iterative Reconstruction – can reduce patient dose,
of VIP patients). and images will benefit from a higher resolution at
organ borders and improved delineation of edges,
Trusted nodes
e.g., to better localize lesions. Moreover, ADMIRE
Reduces the risk of data manipulation by
smoothly integrates iterative reconstruction into the
authenticating predefined secure medical devices in
daily routine. Thick slices are now reconstructed with
the institution’s network (for example, PACS servers).
an improved image impression, even from ultra-low
dose scans. With this, iteratively reconstructed low-
dose datasets can now easily be stored in PACS or
on film. Therefore, ADMIRE sets a new benchmark in
iterative reconstruction instead of limiting its potential
in the clinical routine.
1
In clinical practice, the use of ADMIRE may reduce CT patient dose depending on the clinical task, patient size, anatomical location, and clinical practice.
A radiologist and a physicist should be consulted to determine the appropriate dose to obtain diagnostic image quality for the particular clinical task.
28
syngo CT VB20 · SOMATOM Drive
Virtual Endoscopy software enabling visualization of Spiral and sequential non-fluoroscopic interventional
vessels, airways, and the intestines procedures
29
SOMATOM Drive · syngo CT VB20
Image Quality
Phantom x/y-plane
CATPHAN (20 cm) 2% MTF 16.4 lp/cm (± 10%)
10% MTF 14.0 lp/cm (± 10%)
Object size
50% MTF 11.4 lp/cm (± 10%)
5 mm
Technique
Contrast difference
Wire phantom
3 HU
Acquisition: 12 x 1.2 mm, 120 kV,
CTDIvol (32 cm) 320 mAs, 1.0 s, Kernel H70h
8.0 mGy at 120 eff. mAs
z-direction
Technique 2% MTF 16.9 lp/cm (± 10%)
10 mm slice, 120 kV (spiral body mode), IR 10% MTF 13.1 lp/cm (± 10%)
50% MTF 6.0 lp/cm (± 10%)
Spatial Resolution (UHR)
Technique
Isotropic spatial resolution in all three planes 128 x 0.6 mm, gold platelet, IR strength 5, 0.5 mm slice
(x, y, and z)
Homogeneity
x/y-plane
2% MTF 24.0 lp/cm (± 10%) Cross-field uniformity in a typical mode,
10% MTF 23.3 lp/cm (± 10%) 20 cm water phantom centered
50% MTF 18.7 lp/cm (± 10%) Max. ± 4 HU
Typ. ± 2 HU
Technique
Wire phantom
Acquisition: 16 x 0.6 mm, Kernel U95u
z-direction1
2% MTF 24.3 lp/cm (± 10%)
10% MTF 17.5 lp/cm (± 10%)
50% MTF 6.5 lp/cm (± 10%)
Technique
Gold platelet phantom
Acquisition: 16 x 0.3 mm, slice width 0.4 mm
1
Optional
30
syngo CT VB20 · SOMATOM Drive
Image Quality
z-axis
0.28 mm
0.29 mm
0.31 mm
0.33 mm
0.36 mm
z-axis
0.33 mm
0.36 mm
0.38 mm
0.42 mm
1
Optional with z-UHR
31
SOMATOM Drive · syngo CT VB20
Installation
Components
Gantry, including water/ ≤ 1,990 / 78.3 ≤ 2,400 / 94.5 ≤ 1,300 / 51.2 ≤ 2,300 / 5,071
water cooling system
Patient table ≤ 850 / 33.5 ≤ 800 / 31.5 ≤ 2,480 / 97.6 ≤ 500 / 1,102
Multi-purpose table 1
≤ 1,000 / 39.4 ≤ 800 / 31.5 ≤ 2,480 / 97.6 ≤ 600 / 1,323
Power cabinet ≤ 1,960 / 77.2 ≤ 900 / 35.4 ≤ 700 / 27.6 ≤ 600 / 1,323
Image Recon. System ≤ 550 / 21.7 ≤ 505 / 18.9 ≤ 765 / 30.1 ≤ 100 / 220
syngo Workplaces
syngo.via1
1
Optional
2
Optional split cooling available
32
syngo CT VB20 · SOMATOM Drive
Installation
Nominal power connection water/air split cooling Disabling of Dual Source mode for single-source
system1/2 acquisition
16 kVA
Electromagnetic Compatibility
Power Consumption
This product is in compliance with IEC 60601-1-2
Computer on and fulfills CISPR 11 Class A
3.0 kVA
Fluctuation
System on standby
Nominal voltage
6.0 kVA
+ 10 / –10%
System scanning
Nominal frequency
285 kVA
± 2 Hz
Water/air cooling when system scanning1
Electromagnetic Compatibility
295 kVA
This product is in compliance with IEC 60601-1-2
Water/air cooling on standby1
and fulfils CISPR 11 Class A
16 kVA
Emissions class according to IEC 601-1-2
1
Optional
2
Power consumption – Note: If pretransformer needed, at least 10% more power
33
SOMATOM Drive · syngo CT VB20
Installation
Cooling
Room Environment
Temperature range
18–28 °C
34
International version.
Not for distribution or use in the U.S.
On account of certain regional limitations of sales rights Note: Any technical data contained in this document
and service availability, we cannot guarantee that all may vary within defined tolerances. Original images
products included in this brochure are available through always lose a certain amount of detail when reproduced.
the Siemens Healthineers sales organization worldwide.
The statements by Siemens Healthineers’ customers
Availability and packaging may vary by country and is
described herein are based on results that were achieved
subject to change without prior notice. Some/All of the
in the customers’ unique setting. Since there is no
features and products described herein may not be
“typical” hospital and many variables exist (e.g. hospital
available in the United States.
size, case mix, level of IT adoption) there can be no
The information in this document contains general guarantee that other customers will achieve the same
technical descriptions of specifications and options as results.
well as standard and optional features which do not
always have to be present in individual cases.
Published by Siemens Healthcare GmbH · Order No. CT-09591-28T2-7600 · Printed in Germany · 8510 1020 · © Siemens Healthcare GmbH, 2020