Professional Documents
Culture Documents
8000 OTC18 SM - RevE
8000 OTC18 SM - RevE
Table of
Contents
Safety
3 Adjustments
4 Config &
Calibration
5 Trouble-
shooting
6 Component
Replacement
7 Illustrated
Parts List
P/N 8000-OTC18-SM
Revision E
Copyright © 2023, Del Medical Inc. All rights reserved.
This document is the property of Del Medical Inc. and contains confidential and proprietary information
owned by Del Medical Inc. Any unauthorized copying, use or disclosure of this document without the prior
written permission of Del Medical Inc. is strictly prohibited.
Revision History
Table of Contents
Safety and Compliances
Introduction . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . vii
Statement of Liability . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . vii
Safety Conventions Used in this Manual . . . . . . . . . . . . . . . . . . . . . . . . . . . . . viii
Equipment Safety Guidelines . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .ix
Radiation Protection . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . x
Radiation Safety Regulation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .xi
Personnel Radiation Monitoring . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . xi
Radiation Protection Survey . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . xii
Pediatric Use . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . xii
Restrictions on Use . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .xiii
Electromagnetic Compatibility (EMC). . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .xiii
Grounding . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . xv
Product Life and Disposal . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . xv
Hazardous Materials . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . xv
Acceptance Testing for Regulatory Compliance . . . . . . . . . . . . . . . . . . . . . . . . xv
Chapter 1 — Introduction
1.1 Document Information. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1–1
1.2 System Software Versions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1–1
1.3 Specifications . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1–2
1.4 Applicable Standards . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1–3
1.4.1 Equipment Label Symbols . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1–4
1.5 Abbreviations . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1–5
1.6 Glossary . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1–6
Table of Contents i
OTC18M & OTC18T Overhead Tube Crane Service Manual
Chapter 3 — Adjustments
3.1 Introduction . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3–1
3.2 Required Aids and Tools . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3–1
3.3 Parallelism of the X-ray Tube to the Table . . . . . . . . . . . . . . . . . . . . . . . . . 3–1
3.4 Light Field/Radiation Field of Collimator . . . . . . . . . . . . . . . . . . . . . . . . . 3–2
3.5 Support Arm Wobble Adjustment . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3–2
3.6 Tube Unit Drift . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3–4
3.7 RVA, Rotation Around Vertical Axis. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3–5
3.8 RHA, Rotation Around Horizontal Axis. . . . . . . . . . . . . . . . . . . . . . . . . . . 3–6
3.9 Collimator Alignment . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3–7
3.10 Counterbalance Adjustment . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3–7
3.10.1 Counterbalance Test . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3–7
3.10.2 Counterbalance Adjustment . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3–8
3.11 Vertical Axis Brake Gap (RVA). . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3–8
3.11.1 Inspection . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3–8
3.11.2 Adjustment . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3–9
3.12 Horizontal Axis Brake Gap (RHA). . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3–10
3.12.1 Inspection . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3–10
3.12.2 Adjustment . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3–11
3.13 Transverse and Longitudinal Carriage Brake Adjustment. . . . . . . . . . 3–12
3.13.1 Inspection and Adjustment . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3–12
3.14 Rotational Detent Force . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3–13
3.15 Stop Assembly . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3–14
3.15.1 Checking the Installation Position . . . . . . . . . . . . . . . . . . . . . . . . . 3–15
3.15.2 Operation of the Stop Assembly . . . . . . . . . . . . . . . . . . . . . . . . . . . 3–16
3.16 Setting the Longitudinal and Transverse Pots . . . . . . . . . . . . . . . . . . . . 3–17
ii Table of Contents
OTC18M & OTC18T Overhead Tube Crane Service Manual
Chapter 5 — Troubleshooting
5.1 Introduction . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5–1
5.2 Touchscreen Error Codes. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5–1
5.2.1 Communication Errors . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5–1
5.3 Troubleshooting Guide . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5–3
5.3.1 Touchscreen Display does not Power on, Locks do not Release . . 5–4
5.3.2 No Generator Communication . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5–9
5.3.3 No Touchscreen Communication . . . . . . . . . . . . . . . . . . . . . . . . . . . 5–11
5.3.4 No SID Displayed . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5–12
5.3.5 SID does not Change with Table Movement. . . . . . . . . . . . . . . . . . 5–12
5.3.6 Auto Tracking does not Function (OTC18T only) . . . . . . . . . . . . . 5–13
5.4 I/O Screen . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5–13
5.5 Wiring Diagrams. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5–18
5.3.1 OTC18M Tube Crane Wiring Diagram . . . . . . . . . . . . . . . . . . . . . . 5–19
5.3.2 OTC18T Tube Crane Wiring Diagram . . . . . . . . . . . . . . . . . . . . . . . 5–20
5.3.3 Touchscreen Computer 112-5785 Wiring Diagram . . . . . . . . . . . . 5–21
5.3.4 Touchscreen Computer 112-5977 Wiring Diagram . . . . . . . . . . . . 5–22
5.3.5 Manual System (OTC18M Tube Crane) Wiring Diagram - CM Series
Generator . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5–23
5.3.6 Auto Tracking System (OTC18T Tube Crane) Wiring Diagram - CM
Series Generator . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5–24
iv Table of Contents
OTC18M & OTC18T Overhead Tube Crane Service Manual
Table of Contents v
OTC18M & OTC18T Overhead Tube Crane Service Manual
vi Table of Contents
Safety and Compliances
Introduction
Del Medical X-ray equipment meets high standards of performance and
reliability. We enforce strict quality control techniques to eliminate the
potential for defects and hazards in our products.
This equipment provides an adjustable X-ray source for acquiring images of
parts of a patient’s anatomy. Use of this equipment in any other fashion may
lead to serious personal injury.
These safety guidelines are intended to educate the operator on safety issues
to operate and maintain the OTC18M & OTC18T Overhead Tube Crane in a
safe manner.
Warning
Read and fully understand this manual before operating this
equipment. Failure to follow operating instructions can result in
serious injury or death.
Statement of Liability
To prevent excess radiation exposure (primary or secondary) to patients or
staff, this equipment must be operated and serviced by trained personnel
knowledgeable in the required safety precautions. While this equipment is
designed for safe use, improper operation can cause serious injury or
equipment damage. The manufacturer or its agents and representatives
assume no responsibility for the following:
1. Injury or danger to any person from X-ray exposure.
2. Overexposure of patients or staff to X-radiation due to improper
techniques or procedures.
3. Injury or danger from improper use of the adjustable column or tube head
assembly.
4. Problems or hazards resulting from failure to maintain the equipment as
specified in this manual.
Warning Statements
• indicate hazards or unsafe practices that can cause severe injury or death.
• appear in bold type.
• have a triangular symbol enclosing an exclamation point, with the word
Warning.
• precede the step or piece of information to which they refer.
• appear as the following example:
Warning
This text describes special safety precautions to avoid unsafe practices
that can result in severe injury or death.
Caution Statements
• indicate hazards or unsafe practices that can cause minor injury, or
product or property damage.
• appear in bold type.
• have a triangular symbol enclosing an exclamation point, with the word
Caution.
• precede the step or piece of information to which they refer.
• appear as the following example:
Caution
This text describes special safety precautions to avoid unsafe practices
that can result in injury or product or property damage.
Warning
To avoid serious electric shock, disconnect all power and follow the
discharge procedure before servicing, disconnecting high-tension
cables, or removing access panels. Avoid dropping any tools or objects
into the unit.
Warning
To avoid the dangerous and cumulative effects of X-rays, minimize
radiation exposure to any and all patients and staff.
Radiation Protection
Serious detrimental health effects can result from exposure to both short-
term/high-level and long-term/low-level X-rays. Become familiar with the
effects of both types of radiation exposure, and take appropriate measures to
minimize exposure. Some effects of X-radiation are cumulative over a period
of months or years. Most importantly, avoid exposure to the primary beam at
all times.
X-ray facilities typically use lead to shield the radiologist's workstation. The
typical workstation partition is generally effective in reducing residual
radiation to a safe level. Windows in the partition are typically glazed with
lead glass and provide effective protection. Do not leave the protected area
during exposure.
Any object in the path of the primary beam produces scattered radiation,
which can result in a substantial radiation dose to unprotected persons. To
minimize exposure to scattered radiation, use movable lead screens, and
lead-impregnated gloves and aprons. These protective devices must contain a
lead (or equivalent) thickness of at least 0.25 mm. Use such protective devices
for all operators, observers, and/or servicing personnel exposed to radiation
fields of five or more milli-Roentgens per hour.
The X-ray system produces X-rays only when high voltage is applied to the
X-ray tube. When the high voltage is removed, X-ray emission ceases
immediately.
When the distance between the X-ray tube focal spot and the patient’s skin is
decreased, the radiation dose to the patient increases.
Ensure audio and visual communication between operator and patient even
during exposure. This may require video cameras or an intercom.
To minimize risk, ensure that all operating personnel are educated in the
hazards of radiation and understand the safety requirements for X-ray
operation. Review the manuals for each component in the system.
Note: This system may include an anti-scatter grid. For pediatric applications, the
grid can be removed by a qualified service technician. Refer to the service manual for
the pertinent table or wall stand.
Warning
Ensure exposure parameters are set to within safety limits.
Caution
Incorrectly positioning the X-ray tube and collimator can cause the
X-ray field to be misaligned, resulting in an unacceptable image and
the need to re-expose.
Pediatric Use
Exercise special care when imaging patients outside the typical adult size
range, especially smaller pediatric patients whose size does not overlap the
adult size range (e.g., patients less than 110 lb [50 kg] in weight and 59 in [150
cm] in height).
Ionizing radiation exposure is of particular concern in pediatric patients because:
• For certain organs and tumor types, younger patients are more
radiosensitive than adults (i.e., the cancer risk per unit dose of ionizing
radiation is higher for younger patients).
• Use of equipment and exposure settings designed for adults of average
size can result in excessive and unnecessary radiation of smaller patients.
• Younger patients have a longer expected lifetime over which the effects of
radiation exposure may manifest as cancer.
To help reduce the risk of excessive radiation exposure, follow the ALARA (As
Low As Reasonably Achievable) principle and seek to reduce radiation doses to
only the amount necessary for clinically adequate images.
If available on your system, utilize the following features (and their pediatric-
appropriate settings):
• filters
• grid or Bucky
• protocols or exposure indices
• variable focal spot size
• Automatic Exposure Control (AEC)
• DAP meter
Also, ensure that all testing and calibration procedures include measurements
and adjustments for pediatric settings.
For additional guidance, refer to these resources:
• The Image Gently Campaign: http://www.imagegently.org.
Restrictions on Use
The hardware specified for intended applications of the OTC18 has been
selected, tested, and verified by Del Medical to meet applicable regulatory
agency requirements for those countries where it is offered for sale with
respect to its intended applications.
Warning
Do not install components or accessories not intended for use with
this system. Failure to comply can cause injury or equipment damage.
You are responsible for ensuring that the application and use of this
equipment does not compromise the patient contact rating of any equipment
used in the vicinity of or in conjunction with the system.
Caution
The use of accessories and cables other than those specified (or
supplied by Del Medical as replacements) may cause higher emission
or lower immunity levels.
Caution
Avoid using this equipment near, or stacked on, other equipment. If
this is necessary, first verify normal operation for this configuration.
Grounding
Provide protective earth grounding for all components. All associated
components must be grounded to a common bus according to national and
local electrical codes. Ensure that the earth grounding connections between
this equipment and the host system are maintained at all times.
Hazardous Materials
The X-ray tube, PCBs, and collimator used with this equipment contain lead
and oil. Refer to the equipment manuals or contact Del Medical for additional
information regarding use and disposal of hazardous materials.
Introduction
Warning
This document should not be followed by anyone unless the person is
a qualified and trained service engineer. Any service attempt on the
OTC18M & OTC18T Tube Cranes or other system components by any
untrained and unqualified person could result in serious injury or
death.
The OTC18M & OTC18T Tube Cranes are designed to be used with the DM-
OTC18M and DM-OTC18T floor mounted radiographic systems. For
servicing information on other components of the system, please see service
documentation for the component in question.
General information relating to equipment operation, component
compatibility, and labeling can be found in the system operation manuals
8000-DMOTC18M-OP and 8000-DMOTC18T-OP.
System installation information can be found in the system installation
manuals 8000-DMOTC18M-IN and 8000-DMOTC18T-IN.
1 Introduction 1-1
OTC18M & OTC18T Overhead Tube Crane Service Manual
1.3 Specifications
This section describes the specifications for the OTC18M & OTC18T Tube
Cranes and related components. For a full list of system component
specifications see the system operation manual.
1-2 1 Introduction
OTC18M & OTC18T Overhead Tube Crane Service Manual
1 Introduction 1-3
OTC18M & OTC18T Overhead Tube Crane Service Manual
Alternating Current
1-4 1 Introduction
OTC18M & OTC18T Overhead Tube Crane Service Manual
1.5 Abbreviations
AEC Automatic Exposure Control
AP Anteroposterior
APR Anatomically Programmable Radiography
AWG American Wire Gauge
Btu/hr British Thermal Unit/hour
CAN Controller Area Network
CDRH Center for Devices and Radiological Health
1 Introduction 1-5
OTC18M & OTC18T Overhead Tube Crane Service Manual
CE Communautés Européennes
DAC Digital to Analog Converter
DAP Dose Area Product
DR Digital Radiography
OTC18M & OTC18TFloor Mounted Tubestand
ga Gauge
GUI Graphical User Interface
hPa Hecto Pascal
HT High Tension
ICRP International Commission on Radiological Protection
inHg Inches Mercury
IPM Intelligent Power Module
kN kiloNewton
kV kilovolt
kVp peak kilovoltage
kW kilowatt
LAT lateral
mA milliampere
mAs milliampere second
MeV mega-electron Volt
NCRP National Council on Radiation Protection and Measurements
NIST National Institute for Standards and Technology
OBL oblique
PA posteroanterior
PBL Positive Beam Limitation
PID Patient ID
PN Patient Name
RAO Right Anterior Oblique
UL Underwriters Laboratories
1.6 Glossary
Automatic Exposure Control (AEC) — AEC uses an ionization chamber or a
solid-state detector within the Bucky to command the X-ray generator to stop
the exposure when the correct image density is reached.
AEC field — Commonly known as ionizing vector, the AEC field measures the
dose of X-ray in the interested region.
Angle of tube rotation — This is the angle of deviation from the vertical
direction when the tube rotates, and ranges from -180 to 180 degrees.
Bucky/Grid Cabinet — The component that includes an imaging receptor,
AEC, moving or stationary grid, and related components.
1-6 1 Introduction
OTC18M & OTC18T Overhead Tube Crane Service Manual
Collimator — Device that regulates the size and shape of the X-ray beam to
accurately localize the area of interest on the patient, while reducing overall
patient irradiation exposure.
DAP Meter — DAP (Dose Area Product) is a measurement of dose multiplied
by the dose area. A DAP meter is a sensor mounted to the underside of the
collimator that reports dosage information to the X-ray system.
Free Mode — Exposing position independent of a table or wallstand.
Generator — Device that supplies power to and controls the X-ray tube.
N — Newton, the amount of the net force required to accelerate a mass of one
kilogram at a rate of one meter per second.
kVp — Peak kilo-volts, the highest energy of X-rays emitted by an X-ray tube
(equal to the peak applied tube voltage).
LAT — Lateral view position for X-ray exposure.
Longitudinal — Head-to-foot positioning of focal spot.
mAs — Milliampere-seconds. Combined with kVp, it indicates the dose of
X-rays.
OBL — Oblique view position for X-ray exposure.
Platform Tubemount — A platform-style assembly for mounting a tube,
collimator, and console/handlebar assembly to a tubestand. The platform
mount aims the tube in a specific direction with respect to the arm and
console, and allows the collimator to be removed independently.
Portrait Position — The long dimension parallel to the patient axis.
SID — Source to Image Distance; the distance from the source of the X-rays to
the film or detector plane.
Table Mode — Exposing position where the patient lays on a table
Wall Mode — Exposing position where the patient stands in front of a
wallstand.
Technique Parameter — Any of the parameters describing the properties of
an X-ray beam exposure, including beam energy (kVp), beam intensity (mA),
duration (seconds), time-intensity factor (mAs), and at times, Source to Image
Distance (SID).
Transverse — Cross table (front-to-rear) travel of focal spot.
1 Introduction 1-7
OTC18M & OTC18T Overhead Tube Crane Service Manual
1-8 1 Introduction
2
2 Periodic Maintenance
Periodic Maintenance
Note: Due to varying operating conditions, the procedures listed below may be
required at greater or lesser intervals. You may need to adjust intervals according to
your unit's performance.
What to Do When to Do It
Cleaning External Surfaces Weekly or as Required
Longitudinal and Transverse Rails New system: 18 months
Subsequent: 12 months
Vertical Carriage New system: 18 months
Subsequent: 12 months
Support Arm New system: 18 months
Subsequent: 12 months
Verify Displayed SID 12 months or as required
Tube Calibration 12 months or as required
Table 2-1. Periodic maintenance schedule
Procedure
1. Verify that the equipment is turned off.
2. Starting from the top and working downward, use a cleaning wipe and
non-abrasive, hospital-grade disinfectant to clean the external surfaces
and exposed bearing tracks.
Do:
• Do ensure the power has been disconnected and that the emergency
switches have been activated, before starting any cleaning operation.
• Do ensure no liquid gets into the internal parts of the system.
• While cleaning, do inspect for damage such as surface cracks, exposed/
damaged wires, fluid leaks, metal shavings, and finish that has been
scraped off.
• Do practice good hand hygiene to help further reduce the risk of
cross.contamination.
Do Not:
• Do not immerse the equipment, including any components or accessories,
in liquid.
• Do not use water. Water can short circuit the electrical components and
cause corrosion to mechanical parts.
Warning
Before cleaning, ensure that all equipment power sources are
disconnected, to avoid serious injury.
Ensure no liquid gets into any part of the X-ray system. Never
immerse any part of the equipment, including any components or
accessories, in liquid.
2.4.1 Cleaning
During cleaning of the rails, perform the following tasks:
1. Clean the longitudinal rails and transverse bridge rails.
2. Clean the roller bearings on the transverse bridge and the vertical carriage.
3. Remove collected dust inside of components.
4. Remove soiling that is not accessible to the customer's cleaning personnel
during maintenance work.
Warning
The transverse bridge and/or vertical carriage can roll off their rails if
end stops are loosened or removed, causing injury or death.
2. Inspect the mounting and function of the stop detent solenoids in the
transverse bridge and the vertical carriage.
3. Check the function of the brake assemblies in the transverse bridge and
the vertical carriage.
Spacer
Set Screw
Warning
The transverse bridge and/or vertical carriage can roll off their rails if
end stops are loosened or removed, causing injury or death.
Spacer
Set Screws Sleeves
2. If spacer sleeves are absent, you must order and install them (kit part no.
08773681).
3. Inspect the mounting points for the cable guides.
4. Inspect the lateral guide roller set screws for tightness.
5. Inspect the adjustment of the eccentrics.
6. Inspect attached parts for secure installation.
Warning
The transverse bridge and/or vertical carriage can roll off their rails if
end stops are loosened or removed, causing injury or death.
2. Pull and release the corrugated hose to verify that it returns to its original
position
2.5.5 Counterbalance
The counterbalance unit is a spring-loaded mechanism that provides weight
compensation for the tube unit for ease of movement. The counterbalance
requires replacement every 8 years.
Warning
Do not open the counterbalance; death or serious injury can occur.
1. Test the functionality of the counterbalance. You must be able to move the
tube up or down with minimal effort. If not, readjust the weight
compensation with the adjustment screw (Figure 2-6).
Adjustment
Screw
Caution
If the counterbalance needs to be readjusted frequently or if a
reduction in the maximum length of travel occurs, replace the
counterbalance.
is no deflection at all, you must stop using the OTC18M/T and replace the
counterbalance immediately.
Safety Cable
Support Cable
Caution
To avoid hand injury, do not move the tube unit during inspection.
4. Press down on the cable break safety mechanism. You hear a click as the
safety lock engages.
5. Release the cable break safety mechanism and allow it to move back up
against the support cable by itself. You hear another click as the safety lock
disengages.
Caution
If this test fails, replace the counterbalance assembly.
3. Rotate the tube unit around the vertical axis and verify that the vertical
stop lever (2 in Figure 2-11) engages.
4. Inspect the condition and mounting of the six M6 Allen-head screws (3).
These screws are secured with Loctite and must not be loosened. Use a
torque wrench to verify that the screws are tightened to at least 7.4 ft-lb (10
Nm).
5. Inspect the condition and mounting of the end stop (4) for wear.
Warning
To avoid serious injury, tighten all screws to their defined torque.
6. Inspect the shaft nut (5) for twisting (the 4 notches must align).
7. Use a spring balance to test movement and braking forces. Verify that
movement is smooth.
Adjustments
3.1 Introduction
This chapter provides calibration and adjustment procedures for the
OTC18M/OTC18T overhead tube crane.
Before using the OTC18M/OTC18T or if any major repairs have been
completed, perform the calibration procedure. As you follow the steps in this
procedure, you also see the steps in the display text box.
For all measurements, use the collimator tape measure, while compensating
for any differences between the measurement and the actual film plane
location (for example, the film plane may be an additional 1-1/2” beyond the
tabletop or wallstand receptor surface).
3 Adjustments 3-1
OTC18M & OTC18T Overhead Tube Crane Service Manual
3-2 3 Adjustments
OTC18M & OTC18T Overhead Tube Crane Service Manual
Note: As soon as the Allen screw (4) is loose, wobbling changes as a result of the
unit's own weight.
6. Adjust the hex-head screws to tilt the support arm level per the spirit
level.
7. When the support arm is level on the transverse axis, hand tighten all
transverse wobble screws.
8. With a digital spirit level, verify the support arm is level on the
longitudinal axis within ± 0.1 ° (Figure 3-3). If it is not level, continue to the
next step. If it is level, go to step 13.
3 Adjustments 3-3
OTC18M & OTC18T Overhead Tube Crane Service Manual
Warning
If the wobble Allen-head securing screws are not tightened to the
required torque and secured with Loctite 243, serious injury can occur.
3-4 3 Adjustments
OTC18M & OTC18T Overhead Tube Crane Service Manual
RVA Adjustment
1. Remove the cover panels (1 in Figure 3-5) on the rear of the support arm
that covers the column.
2. To adjust the stop disk for the vertical axis, loosen the securing screw (1 in
Figure 3-6).
3. Release the brake by pressing the vertical rotation button on the handlebar
console.
4. While holding the vertical rotation button, turn the eccentric bolt (2) to
adjust.
5. Retighten the securing screw (1).
6. Check the drift and readjust the brake if necessary.
3 Adjustments 3-5
OTC18M & OTC18T Overhead Tube Crane Service Manual
RHA Adjustment
1. Remove the cover panels (Figure 3-5) on the rear of the support arm that
covers the column.
2. Loosen the security screw (1 and 2 in Figure 3-7).
1 3
3-6 3 Adjustments
OTC18M & OTC18T Overhead Tube Crane Service Manual
Procedure
1. Center the tube unit over the table.
2. Switch on the light localizer.
3. The light field must be parallel to the tabletop.
4. Release an exposure to test the adjustment.
5. The collimator plates should be even horizontally and must be parallel in
the X- ray image.
3 Adjustments 3-7
OTC18M & OTC18T Overhead Tube Crane Service Manual
8. When the tube unit begins to move and moves at an even speed, the force
displayed on the spring balance must not exceed 40 N (9.0 lbf).
The optimum force is 25 N (5.62 lbf).
3.11.1 Inspection
1. Remove the cover panels on the rear of the support arm that cover the
column to get access to the suspension of the support arm.
2. Press the button on the handlebar console for the vertical axis movement
direction.
3-8 3 Adjustments
OTC18M & OTC18T Overhead Tube Crane Service Manual
3. Check the brake gap (1 in Figure 3-9) with the feeler gauge.
3.11.2 Adjustment
1. During this adjustment, press the button on the handlebar console for the
vertical axis rotation movement direction.
Note: The screws to adjust the brake gap for the vertical axis are located on the
bottom of the tube unit support arm (2 in Figure 3-10).
2. Loosen the Allen screws (3) and adjust the brake gap with the adjustment
screws (hex screws) (4).
3 Adjustments 3-9
OTC18M & OTC18T Overhead Tube Crane Service Manual
The brake gap must be 0.3 ± 0.1 mm (.012” ±.004”) all around.
Warning
Screws must be tightened with a defined torque (where defined) and
must be secured with Loctite! Otherwise, serious injury can occur.
3.12.1 Inspection
1. Remove the cover panels on the rear of the support arm that cover the
column to get access to the suspension of the support arm.
2. Press the button on the handlebar console for the horizontal rotation axis.
3. Check the brake gap (1 in Figure 3-11) with the feeler gauge.
3-10 3 Adjustments
OTC18M & OTC18T Overhead Tube Crane Service Manual
The brake gap must be 0.3 ± 0.1 mm (.012” ±.004”) all around.
3.12.2 Adjustment
1. During the adjustment, press the button on the handlebar console for
horizontal axis rotation.
The screws to adjust the brake gap for the horizontal axis (1 in Figure 3-12)
are located on the back of the tube unit support arm.
3 Adjustments 3-11
OTC18M & OTC18T Overhead Tube Crane Service Manual
2. Loosen the Allen screws (2) and adjust the brake gap with the adjustment
screws (hex screws) (3).
The brake gap must be 0.3 mm ± 0.1 mm (.012” ±.004”) all around.
The brake gap must be equal all around.
3. Check the rotation movement for:
• Easy rotation
• No scraping
• Easy release and application of the brake
• Reinstall the cover panels
3-12 3 Adjustments
OTC18M & OTC18T Overhead Tube Crane Service Manual
While holding the assembly in place, apply Loctite 243 to the 4 mounting
screws and tighten them. See Figure 3-13.
Mounting
Screws
3. If turning the screw for the vertical axis is insufficient to adjust the detent
force, washers (3 in Figure 3-15) can be placed under the springs.
3 Adjustments 3-13
OTC18M & OTC18T Overhead Tube Crane Service Manual
3-14 3 Adjustments
OTC18M & OTC18T Overhead Tube Crane Service Manual
1. Move the installation shims (1) on the stop assembly all the way forward
up to the mechanical stop (3) so that the spring is tensioned.
2. Secure each shim (1) in place with the nut (4).
3. Ensure that the tabs (2) on the shims (1) are in contact with the rail.
4. Loosen the nuts (4) to allow the shims to return to normal position.
3 Adjustments 3-15
OTC18M & OTC18T Overhead Tube Crane Service Manual
Note: For reliable detection of each individual detent, set a minimal distance of 1.6 in
(40 mm) between stops (equivalent to 5.9 in [150 mm] center-to-center).
Observe the detent states from the touchscreen. Go to Main > Service Panel >
Diagnostics > Display I/O > Input.
Longitudinal
Detent Status
Transverse
Detent Status
Indicators On Status
Y311 Stop Assembly (Longitudinal Axis)
PS1 Long bridge sensor is over warning ramp (1)
PS1 Long PS2 Long bridge sensor is over the stop detent (2)
PS Long PS1 Long PS2 Long bridge stop lock is engaged
Y312 Stop Assembly (Transverse Axis)
PS1 Trans carriage sensor is over warning ramp (1)
PS1 Trans PS2 Trans carriage sensor is over the stop detent (2)
PS Trans PS1 Trans PS2 Trans carriage stop lock is engaged
If the bridge or the carriage travels too fast over a stop detent, the lock does
not engage.
3-16 3 Adjustments
OTC18M & OTC18T Overhead Tube Crane Service Manual
Longitudinal Potentiometer
Toothed Belt
3 Adjustments 3-17
OTC18M & OTC18T Overhead Tube Crane Service Manual
Toothed Belt
Pinion
Spring
3-18 3 Adjustments
4
4 Configuration & Calibration
Configuration &
Calibration
4.1 Introduction
This chapter provides configuration and calibration procedures for the
OTC18M & OTC18T Overhead Tube Crane . Do not perform any calibration
procedures until the OTC18M & OTC18T Overhead Tube Crane is completely
installed and secured.
Before using the OTC18M/OTC18T, perform the following calibration
procedure. As you follow the steps in this procedure, you also see the steps in
the display text box.
For all measurements, use the collimator tape measure, while compensating
for any differences between the measurement and the actual film plane
location (for example, the film plane may be an additional 1-1/2” beyond the
tabletop or wallstand receptor surface).
Tap
SW1
To access the DIP switches in the AL02 collimator, loosen the two Allen head
screws located through the two small access holes on the front cover. See
Figure 4-2. Once the screws are loosened, the front panel swings open.
1 2 3 4
SW1 ON ON OFF ON
SW3 OFF OFF ON OFF
Caution
Adjusting any other DIP switch other than SW1-1 on the SCU board
could cause damage to the equipment.
SW1
Do NOT adjust
SW3
DelWorks or Fuji DR
For DelWorks or Fuji DR systems set
SW1-1 to ON
Note: In the event the SCU board is being replaced, set SW1 identically to the
currently installed board.
Jumper J11
Jumper X25
Jumper J17
Jumper J5
It is important that exactly 2 of these jumpers are in place on the last CAN
node at each end. Table 4-3 describes the CAN termination jumper settings for
various configurations.
Wall Table
Configuration UCU J5 UCU J5 PSI X25 DCU J11
OTC18M / ML03 collimator - jump open jump
OTC18M / AL02 collimator jump open open open
OTC18M / ML03 (no table) - - jump jump
OTC18M / AL02 (no table) jump - open open
OTC18M / ML03 (no wallstand) - jump open jump
OTC18M / AL02 (no wallstand) - jump open open
OTC18T / ML03 collimator jump open open jump
OTC18T / AL02 collimator jump open open open
OTC18T / ML03 (no table) jump - open jump
OTC18T / AL02 (no table) jump - open open
OTC18T / ML03 (no wallstand) - jump open jump
OTC18T / AL02 (no wallstand) - jump open open
Note:
“jump” = jumper is required to be inserted across both pins of the jumper
header.
“open” = jumper is required to be completely removed from the jumper
header.
“-” = UCU board not included with the configuration.
Jumper J3
Jumper J7
Use Figure 4-14 and Table 4-4 to configure the jumper settings for J7 according
the configuration being installed.
LED D7
When an e-stop button is depressed, LED D7 will turn off. At this time the
motor drive for the vertical movement of the OTC18T will be disabled. There
is also an indication on the OTC18T touchscreen.
Tap
Tap
4. In the configuration menu choose the options that pertain to the system.
Table 4-5 defines the fields in the configuration menu.
Function Action
Auto Lamp When activated, automatically turns the collimator lamp
on when moving the OTC18. When deactivated, the
collimator lamp will only turn on when the lamp button
on the collimator is pressed.
Display Inches When activated the SID will be displayed in inches.
When deactivated the SID will be displayed in
centimeters.
Table Rotate Tray Activate only if the table was built with the Del Medical
Rotate Tray Cabinet Assembly. Deactivate if any other
receptor cabinets are installed.
Wall Stand Rotate Tray Activate only if the wall stand was built with the Del
Medical Rotate Tray Cabinet Assembly. Deactivate if
any other receptor cabinets are installed.
Resize on Move When activated, the collimator will return to cassette
size after the OTC18 is moved. When deactivated, the
collimator will retain a selected size when the OTC18 is
moved.
Variable H SID Wall stand receptor only. When activated, SID will be
displayed between the longitudinal detents. When
deactivated, SID will only be displayed when on an
active longitudinal detent.
Table Choose the table model installed with the system.
Collimator Choose the collimator model installed with the system.
System This field is display only and cannot be changed.
Generator Choose the generator model installed with the system.
Wall Stand Choose the wall stand model installed with the system.
Function Action
Orientation Choose the wall in which the wall stand is installed
(looking at the front of the OTC18) - Right Wall, Left
Wall, Front Wall, Back Wall
Lamp Timer Input the desired value (in seconds) that the collimator
lamp will remain lit when turned on.
Column Top Limit Input the desired upper limit ADC value for the tube
crane motor drive. ADC values can be found in Service
> Diagnostics > IO > Column Encoder
Bottom Limit Input the desired lower limit ADC value for the motor
drive. ADC values can be found in Service >
Diagnostics > IO > Column Encoder
SID Offset T Input the measurement between the table receptor and
table top in inches or centimeters. The typical distance
is about 3.5” (9cm)
SID Offset W Input the measurement between the wall stand receptor
and wall stand face in inches or centimeters. The
typical distance is about 1.5” (4cm)
Fixed Size T W Fixed detector installed in table only. Input the width of
the fixed detector in inches or centimeters.
Fixed Size T L Fixed detector installed in table only. Input the length of
the fixed detector in inches or centimeters.
Fixed Size W W Fixed detector installed in wall stand only. Input the
width of the fixed detector in inches or centimeters.
Fixed Size W L Fixed detector installed in wall stand only. Input the
length of the fixed detector in inches or centimeters.
Table 4-5. Configuration fields
4.5 Calibration
Before using the OTC18M/OTC18T, perform the following calibration
procedure. As you follow the steps in this procedure, you also see the steps in
the display text box.
For all measurements, use the collimator tape measure, while compensating
for any differences between the measurement and the actual film plane
location (for example, the film plane will be an additional distance beyond the
tabletop or wallstand receptor surface).
Note: Do not attempt any of these calibration steps until all stop actuators are
installed and all alignments described in this chapter are complete.
Note: Some of the screen images and instruction steps may differ from what is in this
section depending on software version and consideration. Always follow the steps
listed on the display.
2. Tap on Display to enter the display menu. This menu allows adjustment of
the viewable area of the touchscreen.
4. Angle the tube head 90° to the right and press OK.
5. Move the tabletop to the full back and right position. Move the OTC18M/
OTC18T so that the collimator cross hairs line up with the back right
corner of the tabletop and press OK.
6. Move the tabletop to the full front and left position. Move the OTC18M/
OTC18T so that the collimator cross hairs line up with the front left corner
of the tabletop and press OK.
Enter
4. Lower the table to the lowest position and measure the SID to the film
plane (not the table top). Enter the value in the numeric field and press
enter.
The following steps are only applicable when the optional VT300 is
installed with the system. Be sure to save the calibration and skip to
“Tracking Calibration (OTC18T Only)” on page 4-36 if the VS300 is
installed.
6. Tilt the wallstand so it is in the 90° position. Position the OTC18M/
OTC18T centered to the wallstand receptor. Raise the tube vertically to its
absolute highest position. Move the wallstand vertically so the SID is 30”
(76cm) to the film plane. Enter 30 (76) in the numeric field and press enter.
7. Vertically lower the wallstand receptor to its lowest position and measure
the SID to the film plane. Enter the value in the numeric field and press
enter.
8. Lower the tube so that the SID is 20” (50cm) to the film plane. Enter 20 (50)
in the numeric field and press enter.
3. Raise the tube to maximum height and then raise the table to maximum
height. Lower the tube until the collimator rails gently touch the table top
and press OK.
The following steps are only applicable when the optional VT300 is
installed with the system. Be sure to save the calibration and skip to
“Restoring Factory Default Settings” on page 4-41 if the VS300 is
installed.
4. Tilt the wallstand receptor to 90°. Position the OTC18M/OTC18T so that it
is aligned with the wallstand receptor. Move the wallstand receptor to its
lowest position. Lower the tube until the collimator rails gently rest on the
receptor face and press OK.
LCU beam
Calibration is complete
Be sure to save before exiting.
Caution
This task erases all configurations, all calibration data, and resets the
OTC18M/OTC18T to a factory default configuration. Perform this task
only if instructed to do so by your support representative.
Caution
This procedure produces X-ray, take all necessary precautions and
fully close the collimator blades.
From the PC desktop, click on the Windows icon and select the CPI
Canada folder and select GenWare® Generator Utilities to open.
3. The following screen will appear. Select to calibrate the small or large focal
spot.
Troubleshooting
5.1 Introduction
This chapter contains information for troubleshooting various problems and
errors with the OTC18M & OTC18T Overhead Tube Crane. This
troubleshooting chapter will go over various issues which may be presented
on the OTC18M/OTC18T as a component and potential issues at the OTC18M
& OTC18T Overhead Tube Crane System level.
Warning
All service and maintenance, including the procedures described
within this chapter, are to be performed by qualified service
personnel only.
No Collimator
This error indicates that the collimator and SCU are not communicating.
Follow the below suggestions:
• Go into the service mode / configuration and ensure the correct collimator
is selected. For explanation of the service mode, see Chapter 2, “Service
Mode”.
5 Troubleshooting 5-1
OTC18M & OTC18T Overhead Tube Crane Service Manual
• Verify all connections on the collimator are correct. Verify the collimator
has power.
• Verify connections on the back of the touchscreen X8 and X9.
• If this error occurs during a new install, check the DIP switches inside the
collimator. To access the DIP switches in the AL02 collimator, loosen the
two Allen head screws located through the two small access holes on the
front cover. Once the screws are loosened, the front panel swings open.
5-2 5 Troubleshooting
OTC18M & OTC18T Overhead Tube Crane Service Manual
1 2 3 4
SW1 ON ON OFF ON
SW3 OFF OFF ON OFF
No Generator
This error indicates the SCU is not communicating with the generator. This
error is not always produced with the lack of generator communication.
Another symptom of no generator communication is erratic values in the kV,
mA, and / or mAs fields on the display or an “UNDEFINED” error. Follow
the below suggestions:
• Verify serial connections between the PSI board X20 and the generator
control board J21.
• If this occurs during a new installation, follow the steps outlined in the
OTC18M or OTC18T Installation Manuals to verify that all relative license
keys and connections are made.
More information on this condition can be found at “No Generator
Communication” on page 5-9.
No I/O
This error indicates an issue between the MCU and the FPGA on the SCU
board. This is usually caused by configuration. Check the configuration of the
system and verify it is correct.
Warning
Some of the troubleshooting procedures listed in this chapter require
power to be live while parts of the OTC18M/OTC18T are
disassembled. Live power may be exposed. Take all necessary
precautions when troubleshooting with power supplied to the
OTC18M/OTC18T.
5 Troubleshooting 5-3
OTC18M & OTC18T Overhead Tube Crane Service Manual
5.3.1 Touchscreen Display does not Power on, Locks do not Release
1. Locate the PSI box and remove the front cover.
2. View the green LEDs on the PSI board. All of the LEDs shown in Figure 5-
1should be illuminated.
3. LEDs D1-D5 indicate power coming from the fuses F1-F5. These LEDs
should all be illuminated.
4. If D1 is not illuminated and the other LEDs are illuminated, check the
condition of the fuse F1 and replace as necessary.
5. If all of the LEDs are not illuminated, see section “No Power to Tubestand
and PSI LEDs D1-D5 are not Illuminated.” on page 5-5.
5-4 5 Troubleshooting
OTC18M & OTC18T Overhead Tube Crane Service Manual
6. If all of the LEDs are illuminated, see section “No Power to Tubestand and
PSI LEDs D1-D5 are Illuminated” on page 5-7.
Power supply
LED
2. If the power supply LED blinks on and off, see section “No Power to
Tubestand and Power Supply LED Blinks” on page 5-8. If the LED is off,
continue with this section.
3. Verify the PSI input circuit breaker is in the on position.
5 Troubleshooting 5-5
OTC18M & OTC18T Overhead Tube Crane Service Manual
Circuit breaker
4. Verify the SSR (solid state relay) has 2 illuminated LEDs as shown in
Figure 5-4. If the LEDs are off, verify the generator is powered on and the
turn on cable is properly connected to the correct terminal on the
generator. See Table 5-1.
If the LEDs on the SSR are illuminated, check the wiring between the
terminals, circuit breaker, SSR, and power supply.
5-6 5 Troubleshooting
OTC18M & OTC18T Overhead Tube Crane Service Manual
GENERATOR CONNECTION
CM Series Generator Control Board J20
RF Series (Polydoros) D800 Board X82
5 Troubleshooting 5-7
OTC18M & OTC18T Overhead Tube Crane Service Manual
SCU/IO board
2. The SCU/IO board has a series of green and red LEDs. The green LEDs
represent the various power supplies on the board. Verify the green LEDs
are illuminated. If the green LEDs are not illuminated, check the condition
of the cable connecting the SCU/IO board X1 and the PSI board X1.
3. If the green LEDs on the SCU board are illuminated, check the cable
connecting the SCU/IO board X9 and X1 on the touchscreen display.
5-8 5 Troubleshooting
OTC18M & OTC18T Overhead Tube Crane Service Manual
CM Series Generator
1. Verify proper connection with the null model cable between the Generator
Control Board J22 and X20 of the PSI board.
2. Verify the “Digital Interface Options” license key is in the generator. This
is done through the GenWare® MP Generator Utilities software program.
To verify if the license is installed, open the GenWare® Generator Utilities
and click on the About button at the top of the screen.
5 Troubleshooting 5-9
OTC18M & OTC18T Overhead Tube Crane Service Manual
About
Installed options
5-10 5 Troubleshooting
OTC18M & OTC18T Overhead Tube Crane Service Manual
D9
5 Troubleshooting 5-11
OTC18M & OTC18T Overhead Tube Crane Service Manual
Table Z
5-12 5 Troubleshooting
OTC18M & OTC18T Overhead Tube Crane Service Manual
3. Raise and lower the table while viewing the ADC value on the left of the
drop down menu. If the value changes, go through the calibration
procedure at “Calibration” on page 4-21.
4. If the value does not change, look at the following:
• Verify the UCU DIP switch settings found at “UCU Board DIP Switch
Settings” on page 4-4 (this does not apply with version 3.0.0 or later).
• Verify CAN BUS termination jumpers found at “CAN Termination
Jumpers” on page 4-10.
• Verify the leader wire for the vertical table potentiometer is connected to
the underside of the table top and retracts in and out when moving the
table.
• Verify the table vertical potentiometer is properly connected to J16 of the
UCU/TRMD board.
• Verify all CAN communication connections are properly made between
X1/X2 of the UCU/TRMD board to X22/X23 of the PSI board.
5 Troubleshooting 5-13
OTC18M & OTC18T Overhead Tube Crane Service Manual
Inputs
Outputs
Variable ADC
inputs
— I/O inactive
— I/O active
Outputs
Brake RVA — Indicates the column rotation brake is released
Brake RHA — Indicates the tube rotation brake is released
Brake Long — Indicates the longitudinal brake is released
Brake Trans — Indicates the transverse brake is released
5-14 5 Troubleshooting
OTC18M & OTC18T Overhead Tube Crane Service Manual
Inputs
Front Panel 1 - 9 — Indicates a membrane switch is closed, see below:
Front Panel 9
5 Troubleshooting 5-15
OTC18M & OTC18T Overhead Tube Crane Service Manual
Table Cassette — Indicates cassette presence signal from the table receptor
(automatic collimator with size sensing cassette only)
Wall Cassette — Indicates cassette presence signal from the wall receptor
(automatic collimator with size sensing cassette only)
Table Expose Rdy — Not used
Wall Expose Rdy — Not used
Wall Tracking — Indicates wall tracking is enabled and the OTC18M/
OTC18T is in motion tracking to the wallstand.
RS Long — Not used
RS1 Long — Longitudinal detent approach is detected.
RS2 Long — Longitudinal detent is detected.
RS3 Long — Longitudinal detent pin is locked in the detent.
T Rot Land — Indicates the table receptor is inserted in the landscape
orientation (only applicable if a rotate tray cabinet is installed in the EV800T
table).
T Rot Port — Indicates the table receptor is inserted in the portrait orientation
(only applicable if a rotate tray cabinet is installed in the EV800T table).
RS Trans — Not used
RS1 Trans — Transverse detent approach is detected.
RS2 Trans — Transverse detent is detected.
RS3 Trans — Transverse detent pin is locked in the detent.
W Rot Land — Indicates the wallstand receptor is inserted in the landscape
orientation (only applicable if a rotate tray cabinet is installed in the VT300T
wallstand).
W Rot Port — Indicates the wallstand receptor is inserted in the portrait
orientation (only applicable if a rotate tray cabinet is installed in the VT300T
wallstand).
Note: If an option is not installed (such as rotate tray cabinets) the indications may
show as active. This is normal and not an indication of a failure.
5-16 5 Troubleshooting
OTC18M & OTC18T Overhead Tube Crane Service Manual
Vertical
encoder
ADC
Pos Trans — The position of the transverse potentiometer. The full range is 0-
2045 but should never reach these limits as going past a limit will damage the
pot. The value will change when the OTC18M/OTC18T is moved in the
transverse direction.
Pos Long — The position of the longitudinal potentiometer. The full range is
0-2045 but should never reach these limits as going past a limit will damage
the pot. The value will change when the OTC18M/OTC18T is moved in the
longitudinal direction.
Table Z — The position of the table height potentiometer. The full range is 0-
2045 but should never reach these limits as going past a limit will damage the
pot. The value will change when the EV800 is moved vertically.
Table Bucky Cross — The cross value for a size sensing tray. The value will
vary depending on the size of receptor inserted. Not applicable if a rotate tray
or fixed detector cabinet is installed.
Table Bucky Long — The long value for a size sensing tray. The value will
vary depending on the size of receptor inserted. Not applicable if a rotate tray
or fixed detector cabinet is installed.
5 Troubleshooting 5-17
OTC18M & OTC18T Overhead Tube Crane Service Manual
Table Bucky Long Pos — The position of the table receptor positional
potentiometer.The full range is 0-2045 but should never reach these limits as
going past a limit will damage the pot. The value will change when the table
receptor is moved.
Wall Z — The position of the wallstand height potentiometer. The full range
is 0-2045 but should never reach these limits as going past a limit will damage
the pot. The value will change when the VT300T is moved vertically.
Wall Bucky Cross — The cross value for a size sensing tray. The value will
vary depending on the size of receptor inserted. Not applicable if a rotate tray
or fixed detector cabinet is installed.
Wall Bucky Long — The long value for a size sensing tray. The value will
vary depending on the size of receptor inserted. Not applicable if a rotate tray
or fixed detector cabinet is installed.
SCU +5V — An numeric indication of the +5VDC supply on the SCU.
SCU +24V — An numeric indication of the +24VDC supply on the SCU.
5-18 5 Troubleshooting
OTC18M & OTC18T Overhead Tube Crane Service Manual
5 Troubleshooting 5-19
OTC18M & OTC18T Overhead Tube Crane Service Manual
5 Troubleshooting 5-20
OTC18M & OTC18T Overhead Tube Crane Service Manual
5 Troubleshooting 5-21
OTC18M & OTC18T Overhead Tube Crane Service Manual
5 Troubleshooting 5-22
OTC18M & OTC18T Overhead Tube Crane Service Manual
5.3.5 Manual System (OTC18M Tube Crane) Wiring Diagram - CM Series Generator
5 Troubleshooting 5-23
OTC18M & OTC18T Overhead Tube Crane Service Manual
5.3.6 Auto Tracking System (OTC18T Tube Crane) Wiring Diagram - CM Series Generator
5 Troubleshooting 5-24
OTC18M & OTC18T Overhead Tube Crane Service Manual
Note: If the system does not include a table, the wall CANBUS cable connects to
PSI X22
5 Troubleshooting 5-25
OTC18M & OTC18T Overhead Tube Crane Service Manual
5.3.8 Manual System (OTC18M Tube Crane) Wiring Diagram - Polydoros Generator
5 Troubleshooting 5-26
OTC18M & OTC18T Overhead Tube Crane Service Manual
5.3.9 Auto Tracking System (OTC18T Tube Crane) Wiring Diagram - Polydoros Generator
5 Troubleshooting 5-27
OTC18M & OTC18T Overhead Tube Crane Service Manual
5 Troubleshooting 5-28
6
6 Component Replacement
Component
Replacement
6.1 Introduction
This chapter provides instructions for replacing most of the major assemblies
on the OTC18M/OTC18T.
Warning
Before servicing this equipment, turn off the electrical power to the
system at the power source. Also, to avoid injury, ensure that the
power source is locked out and tagged “Equipment Being Serviced”.
Warning
Before servicing this equipment, to avoid serious injury, turn off
electrical power to this equipment at the power source. Also, ensure
that the power source is locked out and tagged “Equipment Being
Serviced”.
6.4 Covers
Removal
Note: All screws for installing the covers are M4x21 Allen paneling screws and
require a 2.5 mm Allen key.
1. Remove the small cover (Figure 6-1) on the back section of the vertical
carriage over the articulated arm.
3. Remove the 4 paneling screws (Figure 6-3) on the vertical carriage front
cover, and remove the front cover.
Assembly
Note: All screws for installing the covers are M4x21 Allen paneling screws and
require a 2.5 mm Allen key.
Removal
Note: All screws for installing the covers are M4x21 Allen paneling screws and
require a 2.5 mm Allen key.
1. Remove the 2 paneling screws at the rear cover (1 in Figure 6-4) and
remove the rear cover.
2. Remove the 3 paneling screws at the lower cover (2) and remove the lower
cover.
3. Remove the 2 common paneling screws at the upper cover (3), and remove
the upper cover.
Note: For easier removal of the upper cover, rotate the tube unit.
Assembly
Note: All screws for installing the covers are M4x21 Allen paneling screws and
require a 2.5 mm Allen key.
1. Install the upper cover with 2 paneling screws (3 in Figure 6-4).
2. Install the lower cover (2) with 3 paneling screws.
3. Install the back cover (1) with 2 paneling screws.
Removal
1. Move the tube unit to a position for easy access to the rear underside of the
transverse bridge.
2. Switch the system OFF.
3. Remove the spring to uncouple the pinion from the toothed belt.
Assembly
1. Mount the new potentiometer on the sensor subassembly.
2. Connect the wires to the new potentiometer using a soldering iron.
3. Install the sensor assembly and secure it with the hex bolt.
Note: Do not engage the pinion in the toothed belt.
4. Switch the system ON.
5. Perform the procedure “Setting the Longitudinal and Transverse Pots” on
page 3-17.
6. Once the potentiometer is properly set, engage the pinion in the toothed
belt.
Caution
Failure to properly set the potentiometer position could damage the
potentiometer.
Removal
1. Switch the system OFF.
2. Unplug the 2 cables connected to the assembly (1 in Figure 6-7).
3. Remove the brake connector (2).
4. Open the shielding clamp.
5. Unplug the GND connection.
6. Remove the 4 Allen screws on the bottom of the carriage plate (3).
7. Remove the notch magnet assembly.
Assembly
1. Install the metal plates on the new notch magnet assembly as shown in
Figure 6-8, with the “nose” (4) pointing downward.
2. Slide the plates all the way to the front end and secure them in position
with the nut.
3. Verify that the spring is already installed.
4. Loosely install the assembly on the carriage plate with the 4 Allen screws.
Do not tighten the screws.
5. Slide the whole assembly toward the rail (8 in Figure 6-9) until the noses of
the plates touch the rail.
6. While firmly holding the assembly in position, tighten the 4 Allen screws
to secure the assembly. Apply Loctite 243.
7. Loosen the nuts of the small metal plates until the springs pull back on the
plates.
Note: These plates have no function other than to define the distance of the notch
magnet assembly to the rails during installation.
8. Install the brake connector (2 in Figure 6-10) and plug in the 2 cables (1).
Note: The magnetic notch assembly is adjusted in the factory, and therefore, no
adjustment is needed.
Removal
1. Switch the system OFF.
2. Unplug the cable (1) and the connector (2) of the brake at the notch magnet
assembly.
3. Tighten the nuts on both sides to secure the brake springs (Figure 6-12).
Turn them towards the assembly.
4. Remove the 4 Allen screws (Figure 6-13) on the bottom of the carriage
plate.
Assembly
1. Install the new brake assembly loosely on the carriage plate with the 4
Allen screws. Do not tighten the screws.
2. Slide the whole assembly to the rail until the brake pad touches the rail.
3. Tighten the 4 Allen screws to secure the assembly (Loctite 243).
4. Loosen the nuts on both sides to release the brake springs in Figure 6-14).
Turn them until they touch the lock washer.
5. Install the brake connector (1 in Figure 6-15) and plug in the cable (2).
Note: The brake gap is adjusted in the factory, and, therefore, no adjustment is
needed.
Removal
1. Switch the system OFF.
2. Unplug the cable (1) and the connector (2) of the brake at the notch magnet
assembly.
3. Tighten the nuts on both sides to secure the brake springs. Turn them
towards the assembly.
4. Remove the 4 Allen screws (Figure 6-18) on the bottom of the carriage
plate.
Assembly
1. Reinstall the brake pad. Install the screws with a torque of 0.7 ft-lb (0.9
Nm) and apply Loctite 243.
2. Install the brake assembly loosely on the carriage plate with the 4 Allen
screws. Do not tighten the screws.
3. Slide the whole assembly to the rail until the brake pad touches the rail.
4. Tighten the 4 Allen screws to secure the assembly and apply Loctite 243.
5. Loosen the nuts on both sides to release the brake springs (Figure 6-20).
Turn them until they touch the lock washer.
6. Install the brake connector (1 in Figure 6-21) and plug in the cable (2).
Removal
1. Determine which bridge bearing set is to be replaced, and roll the vertical
carriage to rest beneath the opposite rail. This opposite rail is also the rail
that will be extended.
2. See instruction included with kit on attaching the rail extension. You may
need to shorten the extension piece.
Note: If needed, temporarily suspend the corrugated hose with cable ties.
3. Release the brake and slowly move the transverse bridge into the
extension until the bearing to be replaced is out of the rail.
Reassembly
1. Insert the replacement roller bearing in the bearing housing.
2. Install the snap ring.
3. Move the roller bearing up to the stop (snap ring) towards the outside.
4. Tighten the bearing set screw to 12 ft-lb (16 Nm).
5. Roll the bridge back in so that the replaced bearing enters the rail.
6. Rear bearings only: Loosen the bridge eccentric bearing set screws
(Figure 6-22).
7. Adjust the eccentric bearing with a 5mm Allen wrench. Ensure that the
bearing is just touching the outside of the longitudinal rail.
8. Retighten the eccentric bearing set screw.
9. Remove the extension.
10. Install the bumper and cover.
Removal
1. Move the tube unit to a position for easy access to the right-side underside
of the vertical carriage.
2. Switch the system OFF.
3. Remove the front cover of the vertical carriage.
5. Remove the hex bolt (Figure 6-26) and remove the complete assembly.
Assembly
1. Mount the new potentiometer on the sensor subassembly.
2. Connect the wires to the new potentiometer using a soldering iron.
3. Install the sensor assembly and secure it with the hex bolt.
Note: Do not engage the pinion in the toothed belt.
4. Switch the system on.
5. Perform the procedure “Setting the Longitudinal and Transverse Pots” on
page 3-17.
6. Once the potentiometer is properly set, engage the pinion in the toothed
belt.
Caution
Failure to properly set the potentiometer position could damage the
potentiometer.
Removal
1. Press the OFF button on the generator console.
2. Remove the cover on the overhead support.
3. Unplug the 2 cables (1 in Figure 6-27) connected to the assembly and the
brake connector (2).
4. Open the shielding clamp.
5. Unplug the GND connection.
6. Remove the 4 Allen screws and remove the notch magnet assembly (3).
Assembly
1. Install the metal plates on the new notch magnet assembly.
Note: The tab (4 in Figure 6-28) must point up. Figure 6-28 shows the tab (4),
the mechanical stop (5), the nut with washer (6) and the sensors (7).
2. Slide the plates all the way to the head end and secure them in position
with the nut.
Note: The spring must already be installed.
3. Install the assembly loosely on the carriage plate (4 Allen screws).
Note: Do not tighten the screws. Do not damage the sensors while installing the
assembly.
4. Slide the whole assembly to the rail until the noses of the plates touch the
rail.
Note: Figure 6-29 shows the rail (8) and the securing screws (9). The Y312 stop
assembly is in installation position.
5. Tighten the 4 Allen screws with a Loctite 243 to secure the assembly.
Note: Ensure that the assembly does not move out of place while tightening the
screws.
6. Loosen the nuts of the small metal plates until the springs pull back the
plates.
Note: The small metal plates function to define the spacing of the notch magnetic
assembly to the rails during installation of the assembly.
7. Install the brake connector (2 in Figure 6-30) and plug in the 2 cables (1).
Removal
1. Switch the system OFF.
2. Remove the cover on the overhead support.
3. Unplug the cable (1) and the brake connector (2 in Figure 6-31) on the
notch magnet assembly.
4. Tighten the nuts on both sides to secure the brake springs (Figure 6-32).
Turn them towards the assembly.
5. Remove the 4 Allen screws (3 in Figure 6-33) and remove the brake
assembly.
Assembly
1. Install the cover mounting Screw into the new brake assembly and tighten
the lock washer.
2. Install the new brake assembly loosely on the vertical carriage with the 4
Allen screws. Do not tighten the screws.
3. Slide the whole assembly to the rail until the brake pad touches the rail.
4. Tighten the 4 Allen screws with a Loctite 243 to secure the brake assembly.
5. Loosen the nuts on both sides to release the brake springs (Figure 6-34).
Turn them until they touch the lock washer.
6. Install the brake connector (2 in Figure 6-35) and plug in the cable (1).
Note: The brake gap is adjusted in the factory, and therefore, no adjustment is
needed.
Removal
1. Switch the system OFF.
2. Remove the cover on the overhead support.
3. Unplug the cable (1) and the brake connector (2 in Figure 6-36) on the
notch magnet assembly.
4. Tighten the nuts on both sides to secure the brake springs (Figure 6-37).
Turn them towards the assembly.
5. Remove the 4 Allen screws (3 in Figure 6-38) and remove the brake
assembly.
6. Remove the lock washer for the cover mounting bolt (4).
7. Press the cover mounting bolt into the brake assembly and remove it.
8. Remove the brake pad (2 screws in Figure 6-39).
Assembly
1. Reinstall the brake pad. Tighten the screws with 0.7 ft-lb (0.9 Nm) of
torque and apply Loctite 243.
2. Install the cover mounting bolt into the new brake assembly and tighten
the lock washer.
3. Install the new brake assembly loosely on the vertical carriage with the 4
Allen screws. Do not tighten the screws.
4. Slide the whole assembly to the rail until the brake pad touches the rail.
5. Tighten the 4 Allen screws with a Loctite 243 to secure the brake assembly.
6. Loosen the nuts on both sides to release the brake springs (Figure 6-40).
Turn them until they touch the lock washer.
7. Install the brake connector (2 in Figure 6-41) and plug in the cable (1).
Removal
1. Remove the covers of the support arm (not the tube covers).
2. Remove the long Allen screw (1) and use it to pull back the notch lever (2)
by putting the screw (1) into the position (3).
Warning
To avoid injury, ensure the spring pulley safety lock is locked.
Removal
1. Remove the 6 support arm Allen-head mounting screws (Figure 6-44).
Caution
Secure the support arm from tilting, to avoid injury and component
damage.
2. Disconnect the brake cable and remove it (make note of how the cable is
laid).
3. Mark the position of the 5 adjusting nuts (1 in Figure 6-45). They are used
to adjust wobbling and may not be turned.
4. Unscrew the 5 Allen screws (2 in Figure 6-45). Support the suspension,
otherwise it will drop.
Assembly
Warning
Lift the telescope tube and unlock the safety lock of the spring pulley.
Do not pull up the telescope tube, because wires might fly out of the
pulley.
Do not unlock the safety lock of the spring pulley. The telescope tube
would be pulled upwards immediately and can cause serious injury.
Note: The brake gap is adjusted in the factory, and, therefore, no adjustment is
needed.
1. Install the new suspension with the 5 Allen screws (2). Tighten the screws
with 7.4 ft-lb (10 Nm) of torque and apply Loctite 243.
Note: The detent stud (3 in Figure 6-46) must be in the 4 o’clock position from
operator's view regardless of the position of the mounting flange.
2. Remove the long Allen screw (4 in Figure 6-47) that pulls back the notch
lever and install it in its former position (5).
Establishing Access
1. Establish the installation location of the rail extension.
2. Determine which carriage bearing is to be replaced--the opposite rail is the
rail to be extended.
3. See instruction included with kit on attaching the rail extension. You may
need to shorten the extension piece.
Note: Make sure the rail extension and the rubber bumper stop are installed
securely. There may be no gap between the rail and the rail extension.
4. Remove the rubber bumper stop on the rail with the defective roller
bearing.
5. Turn the support arm/tube unit so that the weight is exerted on the load
bearing side.
Note: It is important that the accessible roller bearing that is stressed as little as
possible.
6. Release the brake and slowly move the vertical carriage into the extension.
7. To replace the roller bearings, loosen the set screws (Figure 6-48).
Spacer
Set Screws Sleeves
5. Assemble the threaded insert as shown, with the insert (6a) completely
threaded in (6b), the nut (7), the counter nut (8), onto the M8 screw (9).
Assembly
1. Insert the roller bearing in the vertical carriage.
2. Install the snap ring.
3. Move the roller bearing up to the stop (snap ring) towards the outside.
4. Tighten the set screws (5 in Figure 6-49) to 10 ft-lb (13 Nm).
Set Screws
Final Steps
1. Insert the vertical carriage back into the rail.
2. Reinstall the rubber bumper stops.
3. Remove the extension.
Requirements
Preliminary Steps
1. Remove the support arm covers and the tube covers.
2. Remove the long Allen screw (1) and use it on position (3) to pull back the
notch lever (2 in Figure 6-51).
5. Lock the column with the safety lock screw (Figure 6-53) on the spring
pulley.
Warning
To avoid injury, ensure the spring pulley safety lock is locked.
9. Remove the hose from the cable holder to the support arm (Figure 6-54).
10. Remove the cable holder support arm. It is fixed with 4 flathead Allen
screws.
Note: There are important labels on the cable holder, so the cable holder must be
mounted on the new support arm again.
11. Carefully pull out the cable of the potentiometer for vertical rotation. Put
the potentiometer aside. Make a note of how to lay the cable (for
installation).
Removal
1. Remove the 2 Allen screws and the angle plate on the tube assembly (4 in
Figure 6-55).
2. Remove the cable clamps of the high-voltage cables.
3. Remove the 6 Allen screws that secure the tube assembly to the support
arm.
Warning
Do not unlock the safety lock of the spring pulley while the support
arm is removed. If you unlock the safety lock of the spring pulley, the
telescope tube will be pulled upwards suddenly and can cause
component damage and serious injury.
Caution
Secure the support arm to prevent tilting, otherwise, component
damage and serious injury can occur. Place adequate material under
the support arm to avoid tilting. Be careful of the potentiometer.
Caution
Be careful when removing the support arm. The heavy weight of the
support arm can cause injury.
Warning
Wires inside the telescope tube are under mechanical stress and there
is a risk of damaging components and causing serious injury.
Do not pull up the telescope tube.
Assembly
Note: The locking/detent disk (mounted to the telescope tube) is part of the support
arm spare part kit. If the notches are defective, replace the disk.
1. Remove the 8 screws on the bottom of the disk (Figure 6-57).
2. Note the positions of the notches and remove the disk.
3. Install the new disk according to your notes. Tighten the screws with 12.5
ft-lb (17 Nm) of torque and apply Loctite 243.
4. Secure the support arm to the telescope with the 6 Allen-head screws,
torqued to 7.4 ft-lb (10 Nm). Apply Loctite 243.
6. Install the tube assembly on the new support arm. Tighten the 6 screws of
the suspension with a Loctite 243. Afterwards, install the angle plate (4 in
Figure 6-61).
Preliminary Steps
1. Remove the covers of the support arm.
2. Remove the rubber bumper that stops rotation around the vertical axis (1
in Figure 6-63).
3. Turn the tube assembly around the vertical axis until the notch lever
engages in the deep notch (2). The springs are relaxed at their maximum.
6. Measure and note the position of the springs (gap between springs and
mounting plate of the screws) of the horizontal axis notch lever (important
for installation). Then turn the screws (4 in Figure 6-65) to release the
spring tension.
10. Move the tube assembly up until it stops on the safety lock.
11. Adjust the table height, or place adequate material between the collimator
and the table (e.g., wood planks), until the collimator is supported.
12. Switch the system OFF.
Warning
To avoid injury, ensure the spring pulley safety lock is locked.
Removal
1. Remove the cable clamps of the high-voltage cables.
2. Remove the 2 Allen screws and the angle plate on the tube assembly (5 in
Figure 6-67).
3. Remove the 6 Allen screws that secure the tube assembly to the support
arm (6).
Caution
Heavy crane weight presents risk of injury. Ask for assistance.
Note: The brake gap is adjusted in the factory, and, therefore, no adjustment is
needed.
Warning
Tighten all screws with the specified torque and secure them with
Loctite 243. Otherwise, serious injury can occur.
7. Install the new Y315 brake assembly and secure it with the 5 Allen screws
(7 in Figure 6-68). Tighten the screws with 5.9 ft-lb (8 Nm) of torque and
apply Loctite 243.
8. Install the tube assembly on the support arm. First tighten the 6 screws of
the suspension. Tighten the screws with 7.4 ft-lb (10 Nm) of torque and
apply Loctite 243.
9. Mount the angle plate.
10. Install the cable clamps for the high voltage cables.
11. Tighten the screws of the notch lever horizontal rotation according to your
measurements and notes (4 in Figure 6-69).
14. Turn the tube assembly until the notch lever vertical rotation engages in
the deep notch (2) and install the spring (3 in Figure 6-70).
15. Turn back the tube assembly to the 0° position and reinstall the rubber
bumper with Loctite 243.
16. Perform the beam calibration procedures in Chapter 3, “Adjustments”.
Removal
1. Remove the covers of the support arm.
2. Turn the tube assembly to the 0° position.
3. Remove the rubber bumper that stops rotation around the vertical axis (1
in Figure 6-71).
Assembly
1. Install the new rubber bumper.
2. Secure the screws with Loctite 243.
3. Reboot the system and perform an overall function test.
4. Check the rotation of the vertical axis.
5. Install all covers.
Removal
1. Remove the covers of the support arm.
2. Turn the tube assembly to the 0° position.
3. Remove the rubber bumper (located in the housing (1 in Figure 6-72) that
stops rotation around the horizontal axis.
Assembly
1. Install the new rubber bumper.
2. Secure the screws with Loctite 243.
3. Check the rotation of the horizontal axis.
4. Reboot the system and perform an overall function test.
5. Install all covers.
6.6.11 Collimator
Work time in minutes Number of technicians
60 (mechanical work steps) 1
210 (adjustment and startup) 1
Preliminary Steps
1. Move the tube unit into a convenient working position.
2. Switch the system OFF at the generator or circuit breaker.
3. Follow the adjustment instruction of the collimator.
4. Follow the adjustment instruction of the collimator.
5. Perform the beam calibration procedures in Chapter 3, “Adjustments”.
1. Reboot the system and perform an overall function test.
1. Install all covers.
Preliminary Steps
1. Remove the vertical carriage cover.
2. Remove the collimator back cover and disconnect the plug.
3. Move the tube assembly to a suitable working height.
4. Lock the column with the safety lock screw on the spring pulley.
5. Move the tube up until it stops on the safety lock.
6. Cut the cable ties securing the cables on the tube assembly.
7. Loosen the set screws on the HV cable nuts and remove the HV cables.
8. Remove the handlebar console by lifting it and turn it 90 degrees
downwards.
9. Secure the handlebar console to the column and the cable hose to avoid
damaging or falling down.
10. Loosen the 2 set screws (2 in Figure 6-74) that secure the tube against
rotation.
11. Remove the tube assembly (there are 4 Allen screws to remove).
Note: Be careful removing the tube because the High Voltage and Stator cable
come along with it.
Assembly
1. Install the tube assembly with the 4 Allen screws.
Note: Pay attention to the mount ring position and to the cable that goes
through.
2. Install the handlebar console assembly with the 4 Allen screws.
3. Install the HV cables.
4. Secure the HV cables with the set screw.
5. Connect the cable on the anode side of the tube assembly.
Note: Make sure that the warning label (Part No. 4811548) is affixed to the
tube. If not, apply it as on the original tube.
6. Install the collimator.
7. Plug in the collimator connector, install the back cover of the collimator
and plug in the handlebar console connectors if not done yet.
8. Unlock the safety lock on the spring pulley.
9. Pull the tube column all the way down to tighten the ropes of the spring
pulley.
Note: Prior to ordering the spare parts, please check whether or not there is a drop
ceiling. If there is a drop ceiling that enables access to the top of the vertical carriage,
just order a new telescope tube. If the ceiling is not a drop ceiling, the complete tube
unit must be removed and special aids and tools must be ordered.
Preliminary Steps
1. Remove the support arm covers, the tube covers, and the vertical carriage
covers.
2. Lift up the telescope tube all the way.
3. Remove the nut (1) that secures the steel cable assembly (2 in Figure 6-75)
of R303 to the inner section of the column and let the steel cable rewind
slowly.
6. Lock the column with the safety lock screw (Figure 6-76) on the spring
pulley.
Warning
To avoid injury, ensure the spring pulley safety lock is locked.
9. Secure the cable with the clamping piece at the spring pulley (Figure 6-77).
Caution
Secure support arm against tilting to reduce risk of injury and damage
of components.
Put adequate material under the support arm to avoid tilting.
Caution
Heavy crane weight presents risk of injury. Ask for assistance.
12. Remove the locking disk, which is mounted to the telescope tube.
13. Remove the 8 screws on the bottom of the disk (Figure 6-79).
14. Note the positions of the notches and remove the disk.
15. Lift up the telescope column and pull out the load cable suspension with a
hook (e.g. a cable tie formed to a loop).
16. Remove the bolt of the load cable suspension and the spring and release
the cables.
17. Slowly let down the tube until it touches the table. Put some adequate
material between the telescope column and the table to avoid damage.
18. Safely remove the locking plate (Figure 6-81) for the cable release and lift
the cable up and out.
Removal
1. Remove the 4 screws at the top (3 in Figure 6-82).
2. Remove any cables from the cable clamps mounted on the support
brackets.
3. Completely remove 1 support bracket consisting of 6 Allen screws and
washers (4 and 5 in Figure 6-83).
Note: The telescope tube is still secured to the vertical carriage with the other
support bracket.
Caution
Risk of injury is present because the upper tube is heavy, weighing
about 40 kg.
Ask for assistance; 2 people must hold the upper tube of the telescope.
4. Undo the 2 screws (5 in Figure 6-83) from the support bracket that is still
mounted on the vertical carriage.
5. Remove the telescope tube.
Assembly
1. Install the new tube (Figure 6-84)
2. Set the telescope tube assembly down as shown in Figure 6-84 (do not
stand up the smallest tube directly as it is longer than the others).
3. Set up the assembly so that the drill holes and labels point toward the
spring pulley.
4. 2 persons required: Slide the assembly upwards and secure it in the
mounting brackets with the 4 screws.
Warning
Tighten all screws with the specified torque. Otherwise, serious
injury can occur.
5. From the top, insert the 4 screws and tighten them with 5.9 ft-lb (8 Nm) of
torque.
6. Move the tube assembly up completely and pull the cables through.
7. Look into the inside tube to ensure that the cable is not twisted.
8. Insert the bolts with shims in the load cable suspension and insert the
safety cable, including the spring on the right side.
Note: The catch of the load cable suspension (Figure 6-85) must point towards
the operator.
9. Insert the support cable on the left side and slide the bolt through
completely.
10. Insert the spacer and snap ring.
11. Slide the load cable suspension into the telescope column and slowly
lower the column.
12. Remove the clamp at the spring pulley.
13. Install the locking plate for the cable release.
Warning
Tighten all screws with the specified torque and secure them with
Loctite 243. Otherwise, serious injury can occur.
14. Install the locking disk to the telescope column according to your notes,
and tighten the screws. Tighten the screws with 12.5 ft-lb (17 Nm) of
torque and apply Loctite 243.
15. Install the support arm on the telescope and secure it with the 6 Allen
screws. Tighten the screws with 7.4 ft-lb (10 Nm) of torque and apply
Loctite 243.
Note: The detent stud (6 in Figure 6-86) must be in 4 o’clock position from
operator's view regardless of the position of the mounting flange.
16. Remove the long Allen screw (7in Figure 6-87) that pulls back the notch
lever (8) and install it in its former position.
22. If disconnected:
• plug in X4 on the U303 motor controller.
23. Secure the cables with cable ties.
24. Perform the beam calibration procedures in Chapter 3, “Adjustments”.
25. Home Position Adjustment
26. Adjust the End Position
Requirements
Caution
To avoid damage to the telescope tube, perform the following work
steps with precision.
Preliminary Steps
1. Remove the front vertical carriage cover.
2. Fasten the tension belt at the bar of the spring pulley (Figure 6-89).
3. Lash the tensioning belt and hook in the carabiner with the hook.
Removal
1. In each of the telescope tubes (2), (3), and (4) is 1 roll winder. Replace these
3 roll winders.
Note: For the necessary movements, the emergency button can be switched on in
the meantime. After movement, the emergency button must be switched off again.
3. Remove the 4 Allen screws at the edge cap (Figure 6-92) of the
corresponding tube.
4. Hook the hooks between the edge cap and the telescope tube.
5. Pull the telescope tube upwards until the roll winder (7 in Figure 6-93) is
just fully visible.
6. Hang the rope out of the roll winder. Use the notch (8 in Figure 6-94) at the
rear of the roll winder. The rope must not twist.
Top Edge
9. Remove the retaining ring at the end of the roll winder screw. Remove the
roll winder. It has 1 screw at the lower side of the border. Do not damage
the telescope tube.
Caution
Do not damage the telescope tube.
10. Install the new roll winder with the new screw and the new spring.
11. Turn in the screw until the excess length is in accordance with the
corresponding tube (see table).
12. Install a retaining ring (Figure 6-96) at the end of the screw.
13. Reinstall the roll winder on the edge cap at the same position as before.
Note: The roller is installed on a metal plate. This metal plate must point to the
outer side.
14. Hook the rope in the roll winder; use the notch at the rear of the roller to
straddle the rope (9 in Figure 6-97). The rope must not twist.
15. Lower the telescope tube slowly. Make sure that the metal plate of the roll
winder slides into the telescope tube guide.
16. Install the edge cap on the telescope tube.
17. Perform these work steps for tubes 2, 3, and 4.
Assembly
1. Remove the tension belt at the bar of the spring pulley.
2. Reinstall the front vertical carriage cover.
3. Switch on the emergency button.
Mechanical Steps
1. Move the tube unit (Figure 6-98) nearly into the upper position. The
distances between the tubes must be the same.
Note: If the distances between the tubes are different, adjust the distances of the
tubes with the screws of the roll winder at the lower side of the edge cap.
Functionality Tests
1. Move the tube unit several times from the upper into the lower position.
Preliminary Steps
1. Remove the support arm covers and the vertical carriage covers.
2. Lift up the telescope tube all the way.
3. Remove the nut that secures the steel cable assembly of E303 to the inner
section of the column and let the steel cable rewind slowly.
7. Lock the column with the safety lock screw (Figure 6-103) on the spring
pulley.
9. Adjust the table height, or place adequate material between the collimator
and the table (e.g., wood planks), until the collimator is supported.
10. Switch the system OFF.
Warning
To avoid injury, ensure the spring pulley safety lock is locked.
11. Remove the long Allen screw (1) that pulls back the notch lever (2 in
Figure 6-104).
Caution
Place adequate material under the support arm to avoid tilting.
Caution
The support arm is heavy! Perform this step with another person.
13. Remove the locking disk, which is mounted to the telescope tube.
14. Remove the 8 screws on the bottom of the disk (Figure 6-106).
15. Note the positions of the notches and remove the disk.
16. Lift up the telescope column and pull out the load cable suspension with a
hook (e.g. a cable tie formed to a loop).
17. Remove the bolt of the load cable suspension and the spring and release
the cables.
18. Slowly let down the tube until it touches the table. Put some adequate
material between the telescope column and the table to avoid damage.
19. Safely remove the locking plate (Figure 6-108) for the cable release and lift
the cable up and out.
20. Remove all motor and GND connections at the spring pulley.
21. Remove the snap ring on the left side and tap out the bolt to the right.
Support the spring pulley (Figure 6-110).
22. Unscrew the front screw and carefully remove it. Support the spring
pulley and carefully lower it (Figure 6-111).
Warning
Do not open the counterbalance; if not observed, death or serious
injury can occur.
When replacing the counterbalance, do not open the spring housing.
Assembly
1. Ensure that the safety lock of the new spring pulley is locked.
2. A cable clamp is installed to prevent the cables from being pulled in. The
camp must be left in place till the cables are installed.
3. Lift the new spring pulley and secure with the front screw.
4. Insert the mounting bolt through the telescope housing from the right and
secure the bolt with the snap rings.
5. Position the cable over the cable pulley, guide it into the tube, and install
the wire dropping protection plate.
10. Insert the support cable on the left side and slide the bolt through
completely.
11. Insert the spacer and snap ring.
12. Slide the load cable suspension accordingly into the telescope column and
slowly lower the column.
13. Remove the clamp at the spring pulley.
14. Attach the alignment device (Mat. No. 70 44 493) on the tube assembly to
pull it down.
15. 2 people together, pull the tube assembly into the appropriate working
position. The pull force is 81 kg.
16. Before pulling down the tube assembly, a third person must unlock the
safety lock screw at the spring pulley.
17. When the tube assembly reached the appropriate position, a third person
must lock the safety lock screw.
18. Install the locking disk to the telescope column according to your notes
and tighten the screws. Tighten the screws with 12.5 ft-lb (17 Nm) of
torque and apply Loctite 243.
19. Install the support arm on the telescope and secure it with the 6 Allen
screws. Tighten the screws with 7.4 ft-lb (10 Nm) of torque and apply
Loctite 243.
Note: The detent stud (6 in Figure 6-114) must be in four o’clock position from
operator's view regardless of the position of the mounting flange.
20. Remove the long Allen screw (4) that pulls back the notch lever now in (5
in Figure 6-115) and install it in its former position.
22. Pull the tube column all the way down to tighten the ropes of the spring
pulley.
23. Install the encoder with the 4 screws. Use a 3 mm Allen key.
24. Install the steel cable assembly on the inner section of the column.
Note: The small nib must slide along the edge of the metal bar. Use a hand
mirror.
25. Guide the steel cable over the pulley (4 in Figure 6-116).
26. Pull out the cable carefully.
27. If disconnected:
• plug in X4 on the U303 motor controller.
28. Secure the cables with cable ties.
Removal
1. Switch the system OFF.
2. Remove the vertical carriage rear cover.
SCU Board
Assembly
1. Install the replacement SCU board with 4 mounting screws.
2. Reconnect all connectors to the SCU board.
3. Perform the procedure in “Calibration” on page 4-21.
Part or Drawing
Number Description Page
110-5151G1 OTC18M Manual Overhead Tube Crane 7-5
110-5152G1 OTC18T Tracking Overhead Tube Crane 7-7
11146680 Ceiling stand mechanical 3DV 7-9
07046159 Tube Support Arm 7-11
07047389 Vertical Carriage Cover Kit 7-13
10768378 Longitudinal Rails and Transverse Bridge 7-14
10768378 Tube Cables 7-16
112-5977G1/G2 Touchscreen Computer Assembly 7-20
112-5780G1 Touchscreen Display Assembly 7-20
112-5978G1 DCU/BeagleBoard Assembly 7-21
110-5153G1/G2 Power System Integration Box (PSI) 7-22
112-5775G1 Manual Collimator Platform Assembly 7-24
112-5775G2 Auto Collimator Platform Assembly 7-25
112-5776G1 Trunnion Rings Assembly 7-27
ITEM
NO. PART NUMBER DESCRIPTION QTY
1 11146680 (page 7-9) Ceiling stand mechanical 3DV 1
2 07046563 Spring pulley brake 1
5 408-5305P1 Service warning label 1
6 408-5305P2 Service warning label (French) 1
8 124-5177G4 SCU/IO PCB assembly 1
ITEM
NO. PART NUMBER DESCRIPTION QTY
1 11146680 (page 7-9) Ceiling stand mechanical 3DV 1
2 7046787 Motor complete 1
3 126-5576G2 Motor encoder cable assembly 1
6 408-5305P1 Service warning label 1
7 408-5305P2 Service warning label (French) 1
9 124-5177G4 SCU/IO PCB assembly 1
ITEM
NO. PART NUMBER DESCRIPTION
2 07067127 Gas spring
3 07044980 Cable holder set
4 08773681 Split collar kit
5 03071268 Support couplers
6 07047769 Telescope tubestand
7 07044352 Brake assembly
03070815 Brake pad
8 01265214 Expansion spring
9 08687712 Displace sensor
10 07047371 Spring pulley
07044493 Align fixture (order w/ 07047371)
11 11146596 Rope pulley
12 08687712 Vertical encoder
13 08772378 Roller bearing
14 07043735 Suspension swivel arm
15 08772374 Ecc. Roller bearing
16 07044899 Stop assembly
ITEM
NO. PART NUMBER DESCRIPTION
1 07046159 Tube support arm
2 07045037 Vertical axis suspension
3 07045748 Horizontal axis suspension
4 07044972 Stop buffer RVA
9 07042281 Cover for support arm
10 07042398 Scaled cover
12 07044964 Stop buffer RHA
13 07044980 Cable holder set
14 07046852 Small parts set
ITEM
NO. PART NUMBER DESCRIPTION
1 07044071 Top dust cover
2 07046761 Carriage front and rear bottom covers
201 07043560 Set of slide valves
3 08767485 Pointer
04870804 Paneling screws
Figure 7-6. Longitudinal rails and transverse bridge assembly exploded view
ITEM
NO. PART NUMBER DESCRIPTION
1 03115599 Microswitch S323, S324
2 03074510 Mounting accessories
201 08766586 Cap
202 03074650 Colored cover strip (green0
203 07277163 Shim (50x75x1mm)
204 07277155 Spacer plate (50x75x3mm)
205 08764953 Rubber bumper
206 03073959 Tab (125x75x5mm)
3 10501911 Option kit for small rooms
301 08764987 Stop
302 08764953 Rubber bumper
4 03074668 Colored strip (orange)
5 03073652 Rail extension, complete (for replacement of
roller bearing)
6 07044899 Stop assembly
7 07044352 Brake assembly
03070815 Brake pad
8 07044303 Notch-trap collimator tool
9 03073587 ACSS sensor
07755531 Potentiometer (also order 10757601)
11 08772394 Ecc roller bearing
12 08772978 Roller bearing
13 03071268 Replacement support kit
14 08773681 Split collar sleeves (10 gray, 8 white)
15 01995976 SID markers
16 08767477 Pointer
17 07044980 Cable holder set
18 05508689 Toothed belt
07044303 Stop detents
04870804 Paneling screws
Table 7-7. Longitudinal rails and transverse bridge assembly parts list
ITEM
NO. PART NUMBER DESCRIPTION
1 303-5038P1 Housing, touchscreen front
2 303-5039P1 Housing, touchscreen back
3 112-5780G1 Touchscreen display assembly
4 204-5094P1 Touchscreen display, bottom support
5 204-5093P1 Touchscreen display, top support
6 753-40-14203811 Screw, SCKT HD cap #6-32x3/8
7 205-5046P1 Mounting plate, DCU PCB
8 205-5047P1 Cover, DCU PCB
9 210-5147P1 Retainer, LCU PCB
10 124-5175G1 LCU PCB
11 429-5003P1 429-5003P1
13 632-5059P2 Membrane switch, right
14 632-5059P3 Membrane switch, left, tracking
15 408-5335P1 Label, cable ID
16 114-5291G1 Weldment, back cover
17 112-5978G1 (page 7-21) DCU/Beagleboard assembly
18 760-22-1420311 Screw, PPNHMS SEMS 6-32 x 5/16
19 784-12-16200011 Net, hex KEPS 8-32
20 760-20-14203811 Screw, PPNHMS 6-32x1/4
21 755-40-16203811 Screw, HSBHCS #8-32x3/8
22 755-40-19103811 Screw, HSBHCS 10-32x3/8
23 760-20-14202511 Screw, PPNHMS 6-32x1/4
24 644-5008P1 TOF sensor cable assembly
25 404-0045P1 Gasket, touchscreen display
ITEM
NO. PART NUMBER DESCRIPTION QTY
1 623-5027P1 Touchscreen display assembly 1
2 204-5092P2 Top display mounting bracket 1
3 204-5092P1 Bottom display mounting bracket 1
4 760-20-11203811 Screw, PPNHMS 4-40 x 3/8 4
5 784-12-11200011 Nut, KEPS 4-40 8
6 126-5548G1 Backlight cable assembly 1
7 126-5549G1 LVDS cable assembly 1
ITEM
NO. PART NUMBER DESCRIPTION QTY
1 124-5188G1 DCU PCB 1
2 653-5018P1 BeagleBone Black rev C 1
3 755-5006P1 Micro SD card (must be ordered with 1
item #7)
4 NET-CAT5E-6IN Ethernet cable 6” 1
5 407-5105P1 Torx screw 4
7 037-5023P1 BeagleBone software 1
ITEM G1 G2
NO. PART NUMBER DESCRIPTION QTY QTY
1 114-5295G1 PSI box weldment 1 1
2 204-5095P1 PSI box cover 1 1
3 124-5179G1 PSI PCB 1 1
4 112-5782G1 DIN rail assembly 1 1
5 645-5052P1 Solid state relay DPST 1 1
6 629-5013P1 Power supply, 24V 480W 1 1
7 210-5149P1 Power supply mounting plate 1 1
8 253-5013P1 Threaded rod, 10-32 UNF 2 2
9 200-5044P1 Cross bar 1 1
10 204-5012P1 Cable clamp 7 7
ITEM G1 G2
NO. PART NUMBER DESCRIPTION QTY QTY
11 204-5096P1 Cable tie bracket 1 1
12 204-5097P1 Cable opening cover 1 1
13 784-12-19100011 Hex nut, KEPS 10-32 6 6
14 785-11-19000011 Flat washer, #10 narrow 2 2
15 784-12-14200011 Hex nut, KEPS 6-32 18 18
16 760-22-16202511 Screw, PPNHMS SEMS 8-32 x 1/4 4 4
17 705-00-030000501 Screw, HFSC M3 x 5mm 4 4
18 100012P8 Trim screw, 8-32 x 3/8 4 4
19 823-5036G1 Internal wire, PSI box 1 1
20 823-5036G2 Internal wire, PSI box 1 1
21 823-5036G3 Internal wire, PSI box 1 1
22 823-5036G4 Internal wire, PSI box 1 1
23 823-5036G5 Internal wire, PSI box 1 1
24 823-5036G6 Internal wire, PSI box 1 1
25 126-5579G1 AC power cable assembly 1 1
26 126-5340G1 Connection cable 1 1
27 126-5580G1 Generator on cable 1 -
28 126-5325G2 Exposure hold cable 1 -
29 500-5164P2 DB9 null modem cable 25 ft F-F 1 -
31 408-5305P1 Service warning label 1 1
32 408-5305P2 Service warning label (French) 1 1
33 408-5039P1 Earth ground label 1 1
34 408-5039P2 Dangerous voltage warning label 1 1
35 408-5039P3 Attention label 1 1
36 408-0036P1 Fuse warning label 1 1
37 408-0036P2 Fuse warning label (French) 1 1
45 784-43-14200011 Hex nut, NYLOK 6-32 6 6
46 404-0046P1 Shielding Gasket .39’ .39’
47 126-5617G1 Generator on cable - 1
48 126-5616G1 Siemens exposure trigger cables - 1
49 112-5955G1 Siemens generator CANBUS kit - 1
Table 7-11. PSI box parts list
ITEM
NO. PART NUMBER DESCRIPTION QTY
1 114-5290G1 Platform adapter weldment 1
2 114-5289G1 OTC tube platform weldment 1
3 201-5300P1 Cover mounting bracket 1
4 210-5148P2 X-ray tube rear cover 1
5 210-5148P1 X-ray tube front cover 1
6 100012P8 Trim screw, 8-32 x 3/8 6
7 755-40-31205020 Screw, HSBHCS 5/16-18 x 1/2 SS 2
8 701-50-06002001 Hex screw, HD M6 x 1 x 20L 6
9 786-50-25000011 Washer, SPLITLK-STD 1/4 6
10 703-00-06001601 Screw, SHCS M6 x 16 6
ITEM
NO. PART NUMBER DESCRIPTION QTY
11 753-40-19103811 Screw, SHCS 10-32 x 3/8 2
12 784-12-16200011 Hex nut, KEPS 8-32 1
13 701-50-06002501 Hex screw, HD M6 x 1 x 25L 2
14 786-20-31000011 Lock washer, 5/16 INT 2
15 703-00-05001000 Screw, SHCS M5 x 10 2
16 786-50-19000011 Washer, SPLITLK-STD #10 2
Table 7-12. Manual collimator platform assembly parts list
ITEM
NO. PART NUMBER DESCRIPTION QTY
1 114-5290G1 Platform adapter weldment 1
2 114-5289G1 OTC tube platform weldment 1
3 201-5300P1 Cover mounting bracket 1
4 210-5148P2 X-ray tube rear cover 1
5 210-5148P1 X-ray tube front cover 1
6 203-5208P1 X-ray tube spacer 4
7 70 36 168 Collimator flange 1
8 100012P8 Trim screw, 8-32 x 3/8 6
9 755-40-31205020 Screw, HSBHCS 5/16-18 x 1/2 SS 2
10 701-50-06002001 Hex screw, HD M6 x 1 x 20L 6
11 786-50-25000011 Washer, SPLITLK-STD 1/4 6
12 703-00-06001601 Screw, SHCS M6 x 16 6
13 753-40-19103811 Screw, SHCS 10-32 x 3/8 2
14 784-12-16200011 Hex nut, KEPS 8-32 1
15 703-0005001000 Screw, SHCS M5 x 10 2
16 701-50-06002501 Hex screw, HD M6 x 1 x 25L 2
17 755-43-25207510 Screw, HSBHCS 1/4-20 x 3/4lock 4
18 755-43-25210010 Screw, HSBHCS 1/4-20 x 1 lock 4
19 706-03-06002005 Screw, BHSHCS M6- x 20 lock 4
20 706-03-06002505 Screw, BHSHCS M6 x 25 lock 4
21 786-20-31000011 Lock washer, 5/16 INT 2
22 786-50-19000011 Washer, SPLITLK-STD #10 2
ITEM
NO. PART NUMBER DESCRIPTION QTY
1 439-5030P1 Trunnion rings 1
2 201-5301P2 Right gusset 1
3 201-5301P1 Left gusset 1
4 114-5292G1 Touchscreen adapter weldment 1
5 203-5347P1 Collimator cable cover 1
6 201-5300P1 Cover mounting bracket 1
7 210-5148P2 X-ray tube rear cover 1
8 210-5148P1 X-ray tube front cover 1
9 100012P8 Trim screw, 8-32 x 3/8 6
10 786-50-25000011 Washer, SPLITLK-STD 1/4 4
11 703-00-06001601 Screw, SHCS M6 x 16 4
12 785-11-25000011 Flat washer, 1/4 narrow 4
ITEM
NO. PART NUMBER DESCRIPTION QTY
13 784-12-25200011 Hex nut, KEPS 1/4-20 4
14 701-50-06002001 Hex screw, HD M6 x 1 x 20L 6
15 703-00-05001000 Screw, SCHS M5 x 10 2
16 784-12-16200011 Hex nut, KEPS 8-32 1
17 753-40-19103811 Screw, SHCS 10-32 x 3/8 2
18 755-40-31205020 Screw, HSBHCS 5/16-18 x 1/2 SS 2
19 763-20-16201911 Screw, UPFHMS 8-32 x 3/16 2
20 786-20-31000011 Lock washer, 5/16 INT 2
21 786-50-19000011 Washer, SPLITLK-STD #10 2
Table 7-14. Trunnion rings assembly parts list