Professional Documents
Culture Documents
Pausch Uroview FD II CPI Indico IQ Operator Manual - UM - 905122 - M
Pausch Uroview FD II CPI Indico IQ Operator Manual - UM - 905122 - M
X-RAY GENERATOR
OPERATOR’S MANUAL
PREPARED BY:
905122-00
OPERATOR’S MANUAL P/N
905122-01
CPI INDICO IQ®
X-RAY GENERATOR
OPERATOR’S MANUAL
MODEL # VZW2558
PROGRAMMING THE
CONSOLE 5
P.N. #905122
GENERATOR EXPOSURE
TABLES 7
MAINTENANCE SCHEDULE 8
DIGITAL FUNCTION
SUPPLEMENTS 10
MAIN FEATURES
• Models available in maximum output power of 50 to 100 kW, with both radiographic and
fluoroscopic capability.
• Full 150 kV capability for all models.
• Smaller, lighter modular design.
• Constant dose output due to kV and mA regulation during exposures.
• User-friendly controls.
• Large LCD panel display for APR and utility routines.
• User-friendly system configuration.
• APR techniques may be modified by the operator.
• ms and mAs limits for AEC may be operator set.
• Provisions for APR / tube downloading.
• Extensive self-diagnostics with operator prompt messages.
• Serial RS232 communications interface.
• Programmable setup, calibration and APR routines (using an external computer).
• All models support applications intended for direct or indirect Radiography and applications
intended for direct or indirect Radioscopy
OPTIONS
• Remote exposure hand switch.
• AEC (Automatic Exposure Control).
• Tomography.
• Dual speed starter.
• Various digital interfaces.
• Dose-Area Product (DAP).
PATIENT POPULATION
• Age: Infant to geriatric.
• Weight:
∗ APR mode: Pediatric, Small, medium, large patients.
∗ Manual mode: Exposure factors are set manually, will allow weights from newborn to
obese adult.
• Health: Patients requiring an X-ray may have conditions ranging from mild trauma to
chronic, life threatening illnesses.
FREQUENCY OF USE
Typical range 20 exposures / day to 1500 exposures / day (1 day = 24 hr period).
For the X-ray generator operating in Radioscopy mode (Continuous Fluoro or Pulsed Fluoro) for direct
or indirect applications, the Duty Cycle Limits require no installer or operator configuration or
monitoring.
SAFETY NOTICE: This manual contains important operating and safety information. An
understanding of this information is critical to the safe operation and
long term reliability of your equipment. Please ensure that you read
the warning notices before using this equipment.
Internal X-ray generator components will heat up during normal use of the generator. This is
similar to X-ray tube heating during normal generator operation. The amount of heat produced
is proportional to the product of kV, mA, and time.
Modern Diagnostic X-ray Generators are designed to operate with the majority of X-ray tubes
over their rated power ranges. They are designed for operating duty cycles consistent with
practical patient examination routines that allow for reasonable cooling intervals
between X-ray exposures. Insufficient cooling time between exposures may lead to
excessive heat buildup in the generator, which may cause serious generator damage.
This X-ray generator has internal duty cycle monitoring to warn of excessive heat build-up. The
generator monitors the applicable exposure parameters (kV, mA, ms, PPS) for the selected X-ray Tube
and displays the calculated, accumulated Anode Heat Units according to the tube characteristics
prescribed by the X-ray Tube Manufacturer. If the generator calculates that the next exposure will
exceed the rated duty cycle limit, a warning message GEN DUTY WARNING meaning “Generator duty
cycle warning” will be displayed. Similarly the Generator will issue a GEN DUTY WARNING should the
accumulated exposure parameters and generator power supply thermal status approach the maximum
limits for safe, reliable operation.
The generator also monitors the X-ray tube’s thermal switch, and will inhibit exposures when
the tube reaches its thermal limit. It is the responsibility of the installer to implement and verify
this interlock.
Use this procedure when the generator is first turned on for the day, or when it has not been
used for several hours. This procedure provides for exposures at medium power before the
tube is used at maximum mA or kV values. This will reduce the possibility of damaging the
anode and high voltage components. No test setup is required.
3. For a 200 to 300 kHU tube, use approximately 160 mAs per exposure. For a 300 to 400
kHU tube, use approximately 200 mAs per exposure. Depending on the X-ray tube
power rating, select either 100 mA or 200 mA.
4. Make one (1) to three (3) exposures (depending on tube loading) at 30 second
intervals.
TERMINOLOGY
X-ray radiation exposure may be damaging to health, with some effects being cumulative and
extending over periods of many months or even years. X-ray operators should avoid any
exposure to the primary beam and take protective measures to safeguard against scatter
radiation. Scatter radiation is caused by any object in the path of the primary beam and may
be of equal or less intensity than the primary beam that exposes the film.
No practical design can incorporate complete protection for operators or service personnel
who do not take adequate safety precautions. Only authorized and properly trained service
and operating personnel should be allowed to work with this X-ray generator
equipment. The appropriate personnel must be made aware of the inherent dangers
associated with the servicing of high voltage equipment and the danger of excessive exposure
to X-ray radiation during system operation.
• Operators should be trained on the use of the equipment within the facility. They should
have the knowledge in radiology, and the skills, attitude, and judgment to safely and
effectively operate the equipment.
• Operators are trained on the use of the X-ray generator by the installer of the X-ray
generator. The operator’s manual may be used as a training aid.
• Wear protective clothing. Protective aprons with an equivalent of a minimum of 1/64”
(0.35 mm) of lead are recommended.
• Repetitive or prolonged exposures may result in local skin dose levels that cause adverse
tissue reactions.
• To protect the patient against radiation, always use radiation protection accessories in
addition to devices which are fitted to the X-ray equipment.
• Keep as large a distance as possible away from the object being exposed and the X-ray
tube assembly.
• This X-ray equipment may only be operated in medical rooms which meet IEC
requirements.
• The operator must not touch any part of the console or X-ray generator and the patient
simultaneously.
• The control console must be located inside an X-ray shielded control booth within the X-
ray room, or outside the X-ray room.
• The control console is intended for fixed mounting. It is not a portable device.
• Never operate this X-ray equipment in areas where there is a risk of explosion.
Detergents and disinfectants, including those used on patients, may create explosive
mixtures of gases. Please observe the relevant regulations.
• This equipment is not suitable for use in an oxygen-rich environment.
• The operator’s console, or anything electrically connected to it, must never be used within
6 ft (1.8 m) of the patient environment.
• Only trained maintenance staff may remove the covers of the generator cabinet and the
control console.
• Any type of item should be kept away from the top of the generator cabinet. The
generator should not be subjected to any kind of external force at all times.
• The X-ray generator’s installation instructions are included in the service manual, which is
a separate item from the Operator’s manual. The designated service organization is
qualified to perform the equipment installation.
Do not connect unapproved equipment to the rear of the console: J5 is for the
interconnect cable to the generator main cabinet, J2 is a serial port for use by
an external computer, and J13 is for connection of an external hand switch
and / or foot switch. INCORRECT CONNECTIONS OR USE OF
UNAPPROVED EQUIPMENT MAY RESULT IN INJURY OR EQUIPMENT
DAMAGE.
CONSULT YOUR SERVICE REPRESENTATIVE IF EXTERNAL
EQUIPMENT IS TO BE CONNECTED TO THIS SYSTEM. PERIPHERAL
DEVICES SHOULD BE IEC 601-1 APPROVED IF LOCATED IN THE
PATIENT AREA, AND IEC 950 APPROVED IF OUTSIDE THE PATIENT
AREA.
APPLICABLE STANDARDS
A) Safety
The Indico IQ family of X-ray Generators is designed to the following regulatory requirements
and design standards:
• FDA Center for Devices & Radiological Health (CDRH) - 21 CFR subchapter J (USA), Part
1010 & 1020, Class I.
• Radiation Emitting Devices Act - C34 (Canada).
• Medical Device Regulations (Canada), Class 3.
• EC Directive 93/42/EEC (amended by 2007/47/EC) concerning Medical Devices
(European Community), Class IIb.
• EU Commission Regulation 207/2012 on electronic labeling of medical devices.
• EN60601-1/IEC 60601-1, CAN/CSA-C22.2 No. 60601-1, ANSI/AAMI ES60601-1.
• IEC 60601-2-54/EN60601-2-54, CAN/CSA-C22.2 NO. 60601-2-54,
− Type of protection against electric shock: Class I equipment.
Warning: To avoid the risk of electric shock, this equipment must only be
connected to a supply mains with protective earth.
− Overvoltage category classification: II.
− Pollution degree classification: 2.
− Degree of protection against electric shock: Not classified.
− Degree of protection against harmful ingress of water: Ordinary equipment.
− Mode of operation: Continuous operation with intermittent loading (standby - exposure).
− Equipment not suitable for use in presence of a flammable anesthetic mixture with air or
with oxygen or nitrous oxide.
• EN 62304/IEC 62304 - Software life-cycle processes.
• EN ISO 14971- Risk Management.
• EN 60601-1-6/IEC 60601-1-6. EN 62366/IEC 62366 - Usability and application of usability
engineering to medical devices.
• EN 60601-1-2/IEC 60601-1-2, applying the following standards:
Electromagnetic Immunity
IEC61000-4-2 Electrostatic Discharge
IEC61000-4-3 Radiated RF field
IEC61000-4-4 Electrical Fast Transient
IEC61000-4-5 Surge
IEC61000-4-6 Conducted RF Immunity
IEC61000-4-8 Magnetic Field Immunity
Electromagnetic Emission:
EN55011 (CISPR Publication II Emissions Standards Group 1, Class A)
The following are not specific design requirements at this time but are objectives for
consideration over the life of the product.
Radiated RF 3 V/m 3 V/m Field strengths outside the shielded location from
IEC 61000-4-3 80MHz to 2.5 80MHz to 2.5 fixed RF transmitters, as determined by an
GHz GHz electromagnetic site survey, should be less than 3
a
V/m .
NOTE 1 These guidelines may not apply all situations. Electromagnetic propagation is affected by absorption
and reflection from structures, objects and people.
NOTE 2 It is essential that the actual shielding effectiveness and filter attenuation of the shielded location be
verified to assure that they meet the minimum specification.
a
Field strengths from fixed transmitters, such as base stations for radio (cellular/cordless) telephones and
land mobile radios, amateur radio, AM and FM radio broadcast and TV broadcast cannot be predicted
theoretically with accuracy. To assess the electromagnetic environment due to fixed RF transmitters, an
electromagnetic site survey should be considered. If the measured field strength in the location in which the
VZW2558 series of X-ray generators is used exceeds the applicable RF compliance level above, the X-ray
generator should be observed to verify normal operation. If abnormal performance is observed, additional
measures may be necessary, such as re-orienting or relocating the X-ray generator.
The TÜV SUD mark with the indicators “C” and “US” signifies that the product
is certified for both U.S. and Canadian markets, to applicable U.S. and
Canadian safety standards.
In accordance with the intended use, this Diagnostic X-ray Generator complies with the
European Council Directive concerning Medical Devices. The CE marking affixed to this
product signifies this. One of the harmonized standards of this Directive defines the permitted
levels of electromagnetic emission from this equipment and its required immunity from the
electromagnetic emissions of other devices.
It is not possible, however, to exclude with absolute certainty the possibility that other high
frequency electronic equipment, which is fully compliant to the EMC regulations, will not
adversely affect the operation of this generator. If the other equipment has a comparatively
high level of transmission power and is in close proximity to the generator, these EMC
concerns (the risk of interference) may be more pronounced. It is therefore recommended that
the operation of equipment of this type such as mobile telephones, cordless microphones and
other similar mobile radio equipment be restricted from the vicinity of this X-ray generator.
OUTPUT PARAMETERS
ENVIRONMENTAL SPECIFICATIONS
OPERATING
This information is provided to help you establish safe operating conditions for both
you and your High Frequency X-ray generator.
Do not operate this X-ray generator except in accordance with information included in this
section, and any additional information provided by the X-ray generator manufacturer and/or
competent safety authorities.
Any questions regarding X-ray generator operation should be directed to the Customer
Support Department as shown on the inside cover page of this manual.
I II III
mAs mA/ms
IND100SP_009.CDR
NOTE: THIS MANUAL REFERS TO THE 100 kHz R & F CONSOLE. FOR RAD ONLY UNITS,
IGNORE REFERENCES TO THE FLUOROSCOPY FUNCTION.
NOTE: ICONS ADJACENT TO BUTTONS ARE SHOWN AS THE CONSOLE IS SHIPPED FROM
THE FACTORY. THESE MAY BE ALTERED BY THE INSTALLER TO MATCH THE
ACTUAL CONFIGURATION OF THE GENERATOR.
Press and hold the prep button to spin the rotor. The
LED adjacent to the prep indicator will light to indicate the
prep state.
While pressing the prep button, press and hold the
exposure button to make an X-ray exposure.
The rotor may continue to spin briefly after the fluoro foot
switch is released (installer programmable).
HAND SWITCH (Optional): Press the hand switch halfway to the PREP position. This will spin
the rotor. Fully depress and hold the hand switch in the EXPOSE position to make an X-ray
exposure.
I II III
mAs mA/ms
IND100_003B.CDR
NOTE: The highest reference Air Kerma rate will be delivered if the following
parameters are set to their highest values: kV, mA, ms, mAs and density.
Additionally, increasing the frame rate will increase the reference Air
Kerma rate per frame.
I II III Press the film / screen select button to select the desired
film / screen combination. The appropriate LED will light to
indicate the desired selection.
• I for film / screen 1. Normally the lowest speed film.
• II for film / screen 2.
• III for film / screen 3. Normally the highest speed film.
NOTE: The rate of scrolling for the kV, mA and time display increases if the buttons
are pressed continuously. To return to the slowest scroll rate, release the
button(s) and start again.
IND100_004B.CDR
Note: Image Receptors (1-6) are depicted as the console is shipped from the factory.
The receptors may have been reprogrammed during installation. As a result,
the receptors may contain installer - applied icons that differ from those shown
in this manual.
Note: Exposure parameters will typically change if the image receptor changes. These
changed parameters will influence the delivered reference Air Kerma / Air Kerma
rate as described above.
Press this button to select over-table X-ray tube with table image
receptor. The adjacent indicator will light.
Used for Bucky / AEC techniques.
Auxiliary
NOTE: The image receptors may have been reprogrammed for your installation.
IND100_005B.CDR
The table below shows the modes of operation that are available in fluoroscopy. The low,
medium and high dose rate settings are only available with the variable ABS dose rate option.
Otherwise, only one dose rate setting is available for the fluoroscopy modes shown.
Default values for the various modes cannot be stated, as they will vary depending on
programming and the last setting of the generator. The ranges of the factors that can be varied
are listed in the table below.
Refer to FL ABS CURVE, under ANATOMICAL PROGRAMMING CONTROLS AND
DISPLAY in this section for details regarding the fluoro kV, fluoro mA and pulse width
selections. The pulse width (pulse fluoro or high-level pulse fluoro modes) is not operator-
adjustable; it is programmable by the installer only.
Fluoro modes Fluoro kV Fluoro mA Pulse width (ms) PPS
Continuous fluoro,
low dose rate
Continuous fluoro,
0.5 to 10 mA N/A N/A
medium dose rate
Continuous fluoro,
high dose rate
Hi-level continuous
fluoro, low dose rate
Minimum: 40 kV
Hi-level continuous
fluoro, med dose Maximum: 125 kV
0.5 to 20 mA N/A N/A
rate
Hi-level continuous The min kV limit
fluoro, high dose may be increased
rate and the max kV
Pulse fluoro, low limit may be
dose rate decreased by the
installer in the
Pulse fluoro,
generator 2 to 40 ms.
medium dose rate
programming. PPS 50 Hz:
Pulse fluoro, high
dose rate 1, 2.5, 6, 12.5, 25
5 to 99 mA
Hi-level pulse fluoro, PPS 60 Hz:
low dose rate 1, 3, 7.5, 15, 30
Hi-level pulse fluoro,
4 to 80 ms.
med dose rate
Hi-level pulse fluoro,
high dose rate
NOTE: The highest reference Air Kerma rate will be delivered under the following conditions:
Pulse Fluoro: Highest kV, mA, PPS and pulse width. Pulse width is not operator-
adjustable, but it is at its highest if the ABS dose rate is set to high.
Continuous Fluoro: Highest kV, mA and ABS dose rate set to high.
Fluoro ON Indicator
Note: The fluoro kV button is only functional when ABS is switched off. The fluoro
mA button is only functional when ABS is switched off, or when the FL ABS
CURVE (ABS dose rate) function is off.
Press to select ABS. The adjacent indicator will light, and the
fluoro kV will indicate auto ( ). The fluoro mA will also indicate
auto if the ABS dose rate selection is set to auto kV and mA
(standard configuration) or if low, normal or high dose rate is
selected (variable ABS dose rate option.
Fluoro kV and fluoro mA cannot be set manually when auto kV or
auto mA is indicated, respectively.
In non-ABS mode, the fluoro kV and mA are controlled
manually. Decreasing the kV and mA will decrease the
delivered reference Air Kerma / Air Kerma rate.
In ABS mode kV and possibly mA (depending on the ABS dose
rate selection) are selected automatically. The kV and / or mA will
follow installer-programmable ABS dose curves.
IND100SP_010.CDR
THE ABOVE IS A TYPICAL DISPLAY. YOUR UNIT MAY DIFFER DEPENDING ON MODEL,
SETUP, AND PROGRAMMING
Refer to items 10, 11, and 12 for AEC LOCK, FL/RAD KV CURVE, and FL ABS
CURVE, respectively
c) Press the button to the left of the patient size icons to select the patient size. The
default is medium.
• = pediatric.
• = small.
• = medium.
• = large.
Decreasing the patient size will decrease the kV and / or mA and or / ms, and as a
result will decrease the delivered reference Air Kerma / Air Kerma rate.
d) You may override the programmed techniques by pressing the appropriate selection
button (the ability to override the / mAs / mA/ms button may be disabled in
programming). This will not change the programmed parameters. To make permanent
changes to the APR, see CHAPTER 6: PROGRAMMING THE CONSOLE.
4. DAP (Dose Area Product) or AK (Air Kerma) display. These features are optional.
If set up to display DAP, displays the accumulated dose - area product (mGycm , µGym )
2 2
or dose - area product rate (mGycm2/s, µGym2/s). Refer to DAP - DOSE AREA
PRODUCT near the end of this chapter for details.
If set up to display air kerma, displays the accumulated air kerma (mGy) or air kerma rate
(mGy/min). Refer to AK – AIR KERMA DISPLAY at the end of this chapter for details.
5. ms/mAs readout.
The significance of the ms/mAs readout varies depending on mode selected:
• In AEC mode, with fixed backup time selected, the AEC backup time is displayed
• In AEC mode, with either variable ms or variable mAs backup time, the backup mAs or
ms value is displayed.
• In mAs mode, the time of the exposure will be displayed.
• In mA/ms mode, the calculated mAs of the exposure will be displayed (based on the
mA and ms selected).
7. MAG display.
In fluoro mode, this displays the magnification value as described under I.I.
magnification select button earlier in this chapter.
a) Press the button when MENU is displayed at the main APR screen. This will open
a menu similar to figure A or figure B, below. The MENU function will only be available
if DAP or AK (Air Kerma) is enabled, or if a digital-enabled or fluoroscopy-enabled
receptor is selected.
• The DAP RESET, DAP TEST, and PRINT functions are described in the section
DAP - DOSE AREA PRODUCT near the end of this chapter. Depending on DAP
programming, these functions may not be available.
• The functions AEC LOCK, FL/RAD KV CURVE, and FL ABS CURVE are described
in steps 10, 11, and 12, respectively.
• The AK RESET and FILTER functions are described in the section AK – AIR
KERMA DISPLAY at the end of this chapter. The FILTER selection will only be
available if the generator has been programmed for added collimator filtration.
b) Press the button when RETURN is displayed to return to the previous menu.
c) Press the button when RESET is displayed to return to the main APR menu.
d) Pressing the button will clear status and error messages.
The DAP function is only available if the DAP function has been enabled in programming.
Any reference made to tube 2 or the ability to switch tubes applies to two tube generators only.
DAP Overview
• The DAP device(s) must be allowed to stabilize when the generator is switched on,
and the first time when switching from tube 1 to tube 2 or from tube 2 to tube 1.
During this time, when the DAP device is “settling”, the DAP display will indicate
“--------“. X-ray exposures may be made during this time, but the DAP function will be
disabled during the settling period. The settling period may last up to a few minutes
after the generator is switched on.
• The DAP device for the tube that is selected when the generator is switched on is
automatically tested by the generator after the settling period. If the DAP device
passes the self-test, the DAP readout will display its last setting. This indicates that
the DAP is functional, and ready to measure the dose-area product.
If the DAP self-test fails, an error message will be presented, and the DAP
readout will display “--------“. This indicates that the DAP is not functional.
• When switching to the other tube for the first time in that session (from tube 1 to
tube 2 or from tube 2 to tube 1), the DAP device associated with that tube will be
tested. As in the previous step, the DAP readout will display its last setting to
indicate that the DAP is functional, or “--------“ to indicate that the DAP is not
functional.
• The installer may program the generator to present an audible warning, along with a
visual alarm via the LCD display on the console, to indicate that preset limits for
accumulated DAP or DAP rate have been reached.
* “DAP ACCUM WARN” will display to indicate that the accumulated dose exceeds
the programmed accumulated dose-area product limit.
* “DAP RATE WARN” will display to indicate that the current dose rate exceeds
the programmed dose rate limit.
NOTE: A reading of 0mGycm2 or 0.0µGym2 indicates that the DAP display has been
manually reset as described later in this section, and that no exposures have
been taken since it was reset.
NOTE: Always check local regulations to determine how frequently the DAP device
must be tested. The procedure to test the DAP follows later in this section.
AK Overview
• The AK function calculates and displays the air kerma in mGy for radiographic exposures
and mGy/min for serial radiographic and fluoroscopic exposures.
• The AK calculator sums the air kerma for radiographic exposures and for fluoroscopic
exposures. For example, if after a radiographic exposure the total air kerma is 400 mGy,
and then a fluoro exposure is made for 30 seconds at 700 mGy/min, 350 mGy will be
added to the previous value of 400 mGy, for a new total of 750 mGy. Another rad
exposure of 300 mGy will then bring the total AK to 1050 mGy.
• The air kerma calculator in the generator takes into account kV and collimator filtration.
KV compensation is automatic; however the operator must ensure that the FILTER
selection in the MENU screen matches the collimator filtration that is actually in use.
• Predefined limits may be set to present an audible warning, along with a visual alarm via
the LCD display on the console, to indicate that the accumulated air kerma or air kerma
rate limits have been reached.
* “AK ACCUM WARNING” indicates that the accumulated air kerma exceeds the programmed limit.
* “AK RATE WARNING” indicates that the current air kerma rate exceeds the rate limit.
The AK / DAP display(s) may be reset to zero, a manual DAP functional test may be
performed when desired, and a label may be printed with the AK/ DAP and patient information
as described below.
1. In order to select the DAP device to test, select a receptor that is programmed for the
desired tube / DAP device (tube 1 or tube 2). This will also select the DAP to reset if
programmed for individual dose-area product readings for each tube.
If the two DAP readings are summed, the reset function may be performed on either the
tube 1 or tube 2 selection. The DAP display is common to both tubes, therefore the DAP
display may be reset at either the tube 1 or tube 2 tube position.
If the generator is programmed for individual DAP measurements for each tube, and you
are intending to reset the DAP display, confirm that the correct tube / DAP device is
selected before continuing.
2. Press the button at the main APR menu (MENU will be displayed at the bottom center
of the LCD screen). This will present:
The AK/DAP RESET / DAP TEST / PRINT selections, if AK and DAP both are enabled.
The DAP RESET / DAP TEST / PRINT selections, if DAP is enabled only.
1.1 Press DAP RESET to reset the DAP display to zero or press AK RESET to reset the
AK display to zero, or press AK/DAP RESET to reset the DAP and AK display to zero.
2.1 Press DAP TEST to test the DAP. The message “DAP TEST: OK” will be presented if
the DAP has passed its test.
3.1 For DAP only, repeat step 1.1 and step 2.1 for the other tube if desired.
4.1 Press PRINT to print a DAP or an AK or a DAP and AK label. PRINTING… will be
displayed at the bottom of the screen while data is being sent to the printer.
If the printer option for DAP or AK or DAP and AK is installed, a label will be printed with the
date and time near the top of the label, and the accumulated value(s) near the middle of the
label. This information is retrieved from the generator at the time the label is printed.
Several headings are printed on the labels; the corresponding information may be written in
the adjacent spaces using indelible ink:
NOTE: THE PAPER OR LABELS USED IN THE PRINTER MUST MEET ALL
APPLICABLE REGULATIONS. MEDICAL GRADE PAPER OR LABELS,
APPROVED FOR MEDICAL RECORDS, MUST NORMALLY BE USED.
• Before proceeding with initial dose-area product measurements, ensure that the DAP
display is 0mGycm2 / 0.0µGym2 and the AK display is mGy. The AK / DAP display may be
manually reset to zero as described in AK/DAP Reset / Test / Print.
• The DAP measures and displays cumulative dose per unit area when in RAD mode. It may
be programmed to measure and display the dose-area product for each tube individually,
or to sum the readings from both tubes. This is installer programmable.
• In fluoroscopy mode:
Screen Condition
• DAP is enabled.
• DAP Rate is enabled.
• AK is setup for a
tube.
• DAP is enabled.
• DAP Rate is enabled.
• AK is setup for a
tube.
• The maximum accumulated reading on the DAP display is 9999999. An error message will
be presented when this limit is reached.
FILTER
The collimator filter mapping must be known by the operator, i.e. which physical filter
corresponds to the FILTER 1, FILTER 2, and FILTER 3 selections. This information
must be obtained from the installer / service engineer.
NOTE: Depending on the installation, the actual number of filter selections may be
less than the maximum shown above.
When kV / mAs stepping is enabled, five kV / mAs sequences will be available, with each
sequence containing six kV / mAs steps. The kV and mAs is programmable for each step as
described later in this section.
1. Select the image receptor that is configured for peripheral angiography. The radiographic
display will indicate a preselected and possibly arbitrary kV and mAs value. The first
exposure in the exposure sequence will not necessarily use those kV and mAs values.
The kV and mAs used for the exposure sequence will be as displayed in steps 1 to
6 in the console LCD display window.
2. The LCD display window will display the first kV - mAs stepping screen (PER. SHIFT1) as
shown. The sample screens below do not show the secondary displays i.e. tube number,
%HU, time etc. Also, the kV and mAs values may differ from the sample screens.
*PER. SHIFT1
1- 80kV / 25mAs 5- 68kV / 8mAs
2- 75kV / 25mAs 6- 64kV / 4mAs
3- 72kV / 16mAs
4- 72kV / 16mAs
>>
3. Press >> to select the second kV - mAs stepping screen (PER. SHIFT2) as shown below.
This may be programmed for different kV / mAs steps from the previous screen.
*PER. SHIFT2*
1- 80kV / 25mAs 5- 68kV / 8mAs
2- 75kV / 25mAs 6- 64kV / 4mAs
3- 72kV / 16mAs
4- 72kV / 16mAs
>>
4. Press >> to select the remaining kV - mAs stepping screens (PER. SHIFT3 to PER.
SHIFT5). These screens may be programmed for different kV / mAs steps from the
previous screens.
5. Press << to scroll backwards through the available kV - mAs stepping screens.
1. Select the desired kV - mAs stepping screen PER. SHIFT1 to PER. SHIFT5 as previously
described. If required, edit the kV / mAs values as described later in this section for the
selected exposure sequence.
2. Press and hold the prep / expose button on the stepping table. This will cause exposure
number 1 in the console LCD display window to be highlighted, and the kV and mAs in
the radiographic display window to be updated. Exposure number 1 will be made when
the generator receives the expose command.
3. After the table shifts, the next exposure will be highlighted in the LCD display. That
exposure will be made using revised kV / mAs values, if appropriate, when the generator
receives the expose command.
4. Continue the exposure sequence by holding the prep / expose button. Do not release the
prep button until finished; doing so will abort the exposure sequence.
5. When finished the exposure sequence, release the prep button. This will reset the
exposure step counter to number 1.
1. Select the kV or mAs value to change by pressing the button adjacent to the desired kV /
mAs step. The desired selection will be framed.
2. Select the new kV or mAs value by pressing the + or - kV or mAs buttons in the
radiographic display window.
3. Press SAVE under the LCD display window to save the changes, or to exit without
saving the changes. Repeat to change the kV or mAs for other steps in the same screen,
or in one of the other kV - mAs stepping screens.
The default techniques associated with the anatomical views in the APR display window may
be edited if desired. To do this, follow the steps in this chapter.
*GENERATOR SETUP*
UTILITY APR EDITOR: DISABLED
EXIT SETUP
Access is now provided to the following groups of functions. These will be discussed in
sequence.
UTILITY MENU
To select the UTILITY menu, press the selection button adjacent to UTILITY in the APR
display window. This allows access to the following functions:
*UTILITY*
SET TIME & DATE
ERROR LOG
STATISTICS
CONSOLE
EXIT
SET TIME & DATE • Allows setting of the time and date
ERROR LOG • Allows the error log to be reviewed
STATISTICS • Allows review of the tube exposure
counter(s), the fluoro exposure counter if
applicable, and the generator accumulated
exposure counter.
CONSOLE • Allows console parameters to be set
EXIT • Returns to the GENERATOR SETUP
menu (previous page)
Error Log
* ERROR LOG *
ERROR # …..1.. OF 3 08-25-2009/10:23
E222: INVERTER TEMPERATURE
HIGH +
-
EXIT
Statistics
*STATISTICS*
TUBE 1 EXP: 0
TUBE 2 EXP: 0
FLUORO HOURS: 0
TOTAL EXP: 0
EXIT
The statistics menu shows the tube exposure count, accumulated fluoro hours, and
accumulated generator exposure count.
1. From the UTILITY menu select STATISTICS.
2. The statistics (tube 1 exposures, tube 2 exposures if applicable, fluoro hours if applicable,
and total generator exposures) may now be viewed.
3. Select EXIT to return to the UTILITY menu.
Console
*CONSOLE*
SLOW KEY REPEAT: 200MS LCD SCREEN
MED. KEY REPEAT: 150MS APR MODE: NO
FAST KEY REPEAT: 75MS +
-
EXIT >>
*CONSOLE*
KEY CLICK VOLUME: 8
WARNING VOLUME: 8
EXP. INDICATOR VOL: 8 +
FLUORO BEEP: REPEAT -
<< >>
*CONSOLE*
LOGO: NO
LANGUAGE: ENGLISH
LOAD CONSOLE DEFAULTS: NO +
-
<< >>
*CONSOLE*
DAP UNITS: mGycm2 , mGycm2/s
AK/DAP PRINTER: DPU414
+
-
<< RETURN
The CONSOLE menus display console parameters that may be changed to suit operator
preferences.
1. From the UTILITY menu select CONSOLE.
2. Select SLOW KEY REPEAT. Press the + or - button to select the speed at which displays
change while the button is pressed for the first 5 counts.
3. Select MED. KEY REPEAT. Press the + or - button to select the speed at which displays
change while the button is pressed for the next 5 counts.
4. Select FAST KEY REPEAT. Press the + or - button to select the speed at which displays
change while the button is pressed after 10 counts.
5. Select LCD SCREEN. Press the button to toggle between normal and reverse display for
the LCD display screen.
6. Select APR MODE. Press the selection button to toggle between NO and YES. Selecting
YES disallows changes to the / mAs / mA/ms selections when in normal operating
mode. This must be set to NO to allow switching between / mAs / mA/ms in normal
operating mode.
7. Press >>.
8. Select KEY CLICK VOLUME. Press the + or - button to set the speaker loudness in the
range 1 to 10 for the key click function.
9. Select WARNING VOLUME. Press the + or - button to set the speaker loudness in the
range 1 to 10 for the warning function. This applies when a warning or error message is
reported.
10. Select EXP. INDICATOR VOL. Press the + or - button to set the speaker loudness in the
range 1 to 10 for the exposure indicator function. This applies when making a rad or
fluoro exposure.
11. Select FLUORO BEEP. Press the button to select SINGLE (the exposure indicator beeps
once at the start of a fluoro exposure, and once when the fluoro exposure ends) or
REPEAT (the exposure indicator beeps continuously during a fluoro exposure).
12. Select LOGO button. Press to toggle between YES to enable a predefined logo to
display when the console starts, or NO (logo will not be displayed).
13. Select LANGUAGE button to change the language used to display the error and operator
messages. Available languages are: ENGLISH, GERMAN, FRENCH, ITALIAN, SWEDISH,
SPANISH, RUSSIAN, and DUTCH.
14. Select LOAD CONSOLE DEFAULTS button to toggle between YES and NO. Selecting
YES initializes the console CPU’s NVRAM to the factory-default settings when the
generator is switched ON. This restores the factory defaults for the APR and the console
settings. The normal setting for this function is NO. Do not set to YES unless you intend
to restore the console factory defaults.
15. Select DAP UNITS button to change the units used to display the DAP readings. Two
different units can be selected: µGym2 , µGym2/s or mGycm2, mGycm2/s. The desired units
can be selected by toggling the DAP UNITS button.
16. Select AK/DAP PRINTER to change the printer selection. Pressing the AK/DAP
PRINTER button will cycle the selection to the following: OFF (no printer in use), SLP200
or DPU414. If a printer is in use, select the appropriate printer for your setup.
17. Press EXIT twice to return to the UTILITY menu.
18. Press EXIT to return to the GENERATOR SETUP menu.
APR EDITOR
*GENERATOR SETUP*
UTILITY APR EDITOR: DISABLED
EXIT SETUP
10. Continue making all desired APR changes within that view.
11. When finished, press the SAVE button. The message SAVING DATA... will briefly
appear.
12. If you do not wish to save the changes, press EXIT to abort the SAVE function.
13. Press << to return to the previous screen.
14. Press RESET to return to the main APR menu.
15. You may now make changes to other APR techniques if desired by repeating the above
steps 5 to 12.
16. It is suggested that the APR editor be set to DISABLED as described earlier in this
section, in order to avoid unintended APR changes in the future.
__________________ __________________
__________________ __________________ MENU 1
__________________ __________________ APR TABLES
__________________ __________________ MAX 16 CHARACTERS
__________________ __________________
__________________ __________________ MENU 2
__________________ __________________ ANATOMICAL VIEW
__________________ __________________ MAX 16 CHARACTERS
DENSITY Tomography
R/F (Fluoro)
Small
Medium
Large
mAs
mA/ms
Field
Film Screen
Select
IND100SP_014.CDR
When an error occurs, an error message will be displayed on the console LCD. Some errors
remain on the display until the RESET button is pressed. These types of errors need to be
cleared before the generator can be operated again and they should be reviewed carefully.
Other errors are displayed for about half a second and are cleared automatically. These
errors are usually used to warn the user of a wrong setting on the console such as when a
parameter reached its limit, or a feature that is not enabled or active.
The figure below shows an example of an error as displayed on the membrane console.
Check if the error message shows in the column of the table below. If found, perform the
corresponding action. If the problem persists after taking the recommended action,
consult your service representative.
2 WORK STATE The subsystem has reported Switch the generator off and
MISMATCH an incorrect work state to the then on again. Consult your
main CPU. service representative if the
error persists.
5 +35 VDC ERROR The +35 VDC rail is out of Consult your service
tolerance. representative.
6 -35 VDC ERROR The -35V DC rail is out of Consult your service
tolerance. representative.
9 FAN 1 SPEED ERROR The fan 1 RPM is out of Consult your service
range. representative.
10 FAN 2 SPEED ERROR The fan 2 RPM is out of Consult your service
range. representative.
11 FAN 3 SPEED ERROR The fan 3 RPM is out of Consult your service
range. representative.
12 FAN 4 SPEED ERROR The fan 4 RPM is out of Consult your service
range. representative.
13 FAN 5 SPEED ERROR The fan 5 RPM is out of Consult your service
range. representative.
14 FAN 6 SPEED ERROR The fan 6 RPM is out of Consult your service
range. representative.
16 BATTERY LOW ERROR The generator detects that Consult your service
the battery voltage is low (< representative.
2.5 V).
17 +12VDC ERROR The +12 VDC rail is out of Consult your service
tolerance. representative.
18 -12VDC ERROR The -12 VDC rail is out of Consult your service
tolerance. representative.
19 +15VDC ERROR The +15 VDC rail is out of Consult your service
tolerance. representative.
20 -15VDC ERROR The -15 VDC rail is out of Consult your service
tolerance. representative.
21 DAP POWER SUPPLY The +15 V DAP supply is out Consult your service
ERROR of tolerance. representative.
22 +24VDC ERROR The +24 VDC rail is out of Consult your service
tolerance. representative.
23 -24VDC ERROR The -24 VDC rail is out of Consult your service
tolerance. representative.
24 220VAC FAILURE The 220 VAC supply for Consult your service
room power is lost. representative.
25 110VAC FAILURE The 110 VAC supply for Consult your service
room power is lost. representative.
26 26VAC FAILURE The 26 VAC supply for room Consult your service
power is lost. representative.
27 18VAC PHASE 1 The 18 VAC supply for room Consult your service
FAILURE power is lost. representative.
28 18VAC PHASE 2 The 18 VAC supply for room Consult your service
FAILURE power is lost. representative.
91 UPDATING ACB The firmware for the ACB is Wait until the firmware
FIRMWARE currently updating. updating process for the ACB
is completed.
92 UPDATING RIOB The firmware for the RIOB is Wait until the firmware
FIRMWARE currently updating. updating process for the RIOB
is completed.
93 UPDATING DSS The firmware for the DSS is Wait until the firmware
FIRMWARE currently updating. updating process for the DSS
is completed.
101 DC BUS CHARGE The generator did not detect Consult your service
TIMEOUT proper charging of the DC representative.
bus capacitors. The main
contactor in the generator
will not close until the DC bus
is charged.
102 DC BUS NOT The DC bus voltage has not Consult your service
CHARGING reached 25% of its nominal representative.
voltage within 500
milliseconds.
104 DC BUS FAULT The DC bus voltage is out of Consult your service
tolerance. representative.
108 TUBE / HIGH VOLTAGE A resonant inductor over 1. Perform a tube warm up
ARC current condition was procedure as per tube
detected. manufacturer’s
recommendations.
109 TUBE / HIGH VOLTAGE An over current condition 1. Perform a tube warm up
ARC 1 was detected on inverter 1. procedure as per tube
manufacturer’s
recommendations.
110 TUBE / HIGH VOLTAGE An over current condition 1. Perform a tube warm up
ARC 2 was detected on inverter 2. procedure as per tube
manufacturer’s
recommendations.
111 TUBE / HIGH VOLTAGE An over current condition 1. Perform a tube warm up
ARC 3 was detected on inverter 3. procedure as per tube
manufacturer’s
recommendations.
112 TUBE ARC ANODE The anode current is too 1. Perform a tube warm up
high. procedure as per tube
manufacturer’s
recommendations.
113 TUBE ARC CATHODE The cathode current is too 1. Perform a tube warm up
high. procedure as per tube
manufacturer’s
recommendations.
120 HIGH KV DURING The generator detected kV in Switch the generator OFF.
STANDBY the standby state. Prevent further use of the
generator. Then consult your
service representative.
124 ADR OPEN CIRCUIT No reading detected from Consult your service
ADR module. representative.
125 ADR SHORT CIRCUIT High reading detected from Consult your service
ERROR ADR module. representative.
127 HEAT EXCHANGER The heat exchanger cooling Consult your service
WATER WARNING fluid level is low. representative.
128 HEAT EXCHANGER Cooling fluid flow not Consult your service
FLOW ERROR detected. representative.
129 GRID NOT READY Generator does not detect Consult your service
the ADR2. representative.
130 GETTER CABLE The generator does not Consult your service
INTERLOCK ERROR detect the Heat Exchanger representative.
cable.
201 ROTOR ACTIVE IN The rotor ready signal is Consult your service
STANDBY present when Prep is not representative.
activated.
202 ROTOR BOOST ERROR Rotor ready signal not Consult your service
detected within the allowed representative.
timeout period after the prep
was activated.
203 ROTOR RUN ERROR Low rotor current detected in Consult your service
the main or shift windings. representative.
204 LF FEEDBACK TOO The large focus filament Consult your service
LOW feedback voltage is too low. representative.
205 SF FEEDBACK TOO The small focus filament Consult your service
LOW feedback voltage is too low. representative.
207 LF FEEDBACK HIGH IN The generator detected that Turn off the generator to
STANDBY the large focus filament prevent potential tube damage.
current is greater than the Consult your service
limits in standby mode. representative.
208 SF FEEDBACK HIGH IN The generator detected that Turn off the generator to
STANDBY the small focus filament prevent potential tube damage.
current is greater than the Consult your service
limits in standby mode. representative.
210 FILAMENT MISMATCH The focus enabled is not the Consult your service
focus selected. representative.
212 DAP NOT READY The optional DAP is in its Allow the DAP chamber
“warm-up” state and not sufficient time to warm up. If
ready to make DAP the problem persists, consult
measurements. your service representative.
213 DAP DEVICE ERROR The DAP device is not 1. Perform DAP Reset.
functional.
2. Perform DAP Test.
214 AEC STOP SIGNAL The AEC PT STOP signal is Consult your service
ERROR present in the Rad ready or representative.
prep state when the AEC
mode is active.
215 AEC REFERENCE The AEC reference voltage 1. Reduce AEC density
ERROR is out of its operating range setting.
in the Rad ready or prep
state when the AEC mode is 2. If the problem persists,
active. consult your service
representative.
216 AEC DEVICE ERROR The generator has detected 1. Check that the X-ray tube is
insufficient or no feedback pointing at the correct
signal from the AEC device. receptor.
217 AEC BACKUP TIME The AEC exposure exceeded 1. Check that the X-ray tube is
the allowed backup time. pointing at the correct
receptor.
218 AEC BACKUP MAS The AEC exposure exceeded 1. Check that the X-ray tube is
the allowed backup mAs. pointing at the correct
receptor.
220 TOMO BACKUP TIME The tomo exposure has 1. Check the exposure
exceeded the programmed technique settings.
tomo backup time.
2. Increase the tomo backup
time, if necessary.
221 NO SYNC PULSE No sync input was detected 1. Check that the imaging
when a synchronized system is powered on and
exposure is required. operational.
222 INVERTER The generator has sensed a 1. Allow the generator to cool
TEMPERATURE HIGH high inverter temperature. down before resuming
operation.
223 RESONANT The generator has sensed a 1. Allow the generator to cool
TEMPERATURE HIGH high resonant temperature. down before resuming
operation.
224 AMBIENT The generator has sensed a 1. Ensure that the ambient
TEMPERATURE HIGH high internal ambient temperature where the
temperature. generator is located is less
than 40°C.
225 GENERATOR DUTY The generator has exceeded Re-evaluate technique factors.
CYCLE LIMIT its calculated duty cycle limit. Allow the generator to cool if
possible. If exposures are
continued, serious generator
damage may result due to
overheating.
226 TUBE THERMAL The thermal interlock switch 1. Wait for X-ray tube 1 to
SWITCH 1 OPEN for X-ray tube 1 is open. cool.
227 TUBE THERMAL The thermal interlock switch 1. Wait for X-ray tube 2 to
SWITCH 2 OPEN for X-ray tube 2 is open. cool.
229 DSS NOT CONFIGURED The DSS Tube Type has not Consult your service
been configured when representative.
attempting to start a tube
rotor.
230 DSS FAILED TO The DSS Tube Type failed to Consult your service
CONFIGURE download. representative.
231 DSS DIP SWITCH The DSS DIP Switches are Consult your service
WRONG set incorrectly. representative.
232 DSS DATA CORRUPTED The DSS 400 VAC/480 VAC Power cycle generator OFF
voltage table has been and ON to reconfigure.
corrupted.
233 DSS BUSY The DSS CPU has been in Power cycle generator OFF
its Busy state for too long. and ON to reconfigure.
234 DSS LOW SPEED NOT Low speed operation not Consult your service
SUPPORTED supported. representative.
235 DSS HIGH SPEED NOT High speed operation not Consult your service
SUPPORTED supported. representative.
236 DSS ROTOR INVERTER Rotor current detected when Consult your service
WRONG the DSS is not enabled. representative.
237 DSS MAIN CURRENT The tube main current is low. Consult your service
LOW representative.
238 DSS MAIN CURRENT The tube Main Current is Consult your service
HIGH high. representative.
239 DSS SHIFT CURRENT The tube Shift current is low. Consult your service
LOW representative.
240 DSS SHIFT CURRENT The tube Shift Current is Consult your service
HIGH high. representative.
241 DSS INVERTER TRIP Detection circuit on DSS Consult your service
tripped due to excessive representative.
current detected during prep.
243 DSS VBUS LOW The VBus reading is low. Consult your service
representative.
244 DSS VBUS MISMATCH The DSS VBus reading does Consult your service
not match ACB DCBus representative.
reading.
245 DSS CAP NOT FOUND The required DSS Consult your service
Capacitance is not found. representative.
246 ROTOR THAWING IN The Rotor Thawing process Wait for the X-ray tube to
PROGRESS is currently in progress. complete the Rotor Thawing
process.
247 DSS PHASE CURRENT The Three Phase Starter Consult your service
LOW phase current reading is too representative.
low.
248 DSS PHASE CURRENT The Three Phase Starter Consult your service
HIGH phase current reading is too representative.
high.
249 DSS FAN SPEED LOW The Three Phase Starter Fan Consult your service
speed is too low. representative.
250 DSS FAN SPEED HIGH The Three Phase Starter Fan Consult your service
speed is too high. representative.
251 DSS CPU The Three Phase Starter Consult your service
TEMPERATURE HIGH processor temperature is too representative.
high.
301 INVALID CORE DATA The core data in the SRAM Consult your service
is corrupt. representative.
302 INVALID TUBE DATA The tube data in the flash Consult your service
memory is corrupt. representative.
303 INVALID STARTER The starter data in the flash Consult your service
DATA memory is corrupt. representative.
304 INVALID GEN LIMITS The generator limit data in Consult your service
the flash memory is corrupt. representative.
305 INVALID REC DATA The receptor data in the flash Consult your service
memory is corrupt. representative.
306 INVALID ROOM I/O The room I/O data in the Consult your service
DATA flash memory is corrupt. representative.
307 INVALID AEC DATA The AEC setup data in the Consult your service
flash memory is corrupt. representative.
308 INVALID AEC CAL DATA The AEC film / digital Consult your service
calibration data in the flash representative.
memory is corrupt.
309 INVALID AEC DENSITY The AEC density data in the Consult your service
DATA flash memory is corrupt. representative.
310 INVALID FLUORO DATA The fluoro setup data in the Consult your service
flash memory is corrupt. representative.
311 INVALID CF DATA The continuous fluoro setup Consult your service
data in the flash memory is representative.
corrupt.
312 INVALID PF DATA The pulsed fluoro setup data Consult your service
in the flash memory is representative.
corrupt.
313 INVALID HLCF DATA The high-level continuous Consult your service
fluoro setup data in the flash representative.
memory is corrupt.
314 INVALID HLPF DATA The high-level pulsed fluoro Consult your service
setup data in the flash representative.
memory is corrupt.
315 INVALID DAP DATA The DAP setup data in the Consult your service
flash memory is corrupt. representative.
316 INVALID AK DATA The Air Kerma setup data in Consult your service
the flash memory is corrupt. representative.
317 INVALID SID DATA The SID setup data in the Consult your service
flash memory is corrupt. representative.
318 INVALID FRTC DATA The fluoro to rad kV transfer Consult your service
curve setup data in the flash representative.
memory is corrupt.
319 INVALID CAL DATA The tube calibration table is Consult your service
invalid. representative to recalibrate
the X-ray tube.
320 INVALID STATISTICS The generator statistics data Acknowledge the error and
in the flash memory is resume use. Error will
corrupt. automatically clear itself and
reset statistics data.
321 INVALID ERROR LOG The generator error log is Consult your service
invalid. representative.
322 INVALID DATA BACKUP The backup data file is Consult your service
FILE invalid. representative.
324 INVALID CINE DATA The Cine Setup configuration Consult your service
data is invalid. representative.
325 INVALID LLCF DATA The Low Level Continuous Consult your service
Fluoroscopy configuration representative.
data is invalid.
326 INVALID LLPF DATA The Low Level Pulsed Consult your service
Fluoroscopy configuration representative.
data is invalid.
327 INVALID EPF SETUP The Enhanced Pulsed Consult your service
DATA Fluoroscopy configuration representative.
data is invalid.
328 INVALID SOUND SETUP The Sound configuration Consult your service
DATA data is invalid. representative.
501 BUCKY CONTACTS 1 The state of INPUT 1 Check that the selected Bucky
ERROR (BUCKY CONTACTS 1) was is in operation.
not satisfied for its
programmed state.
502 BUCKY CONTACTS 2 The state of INPUT 2 Check that the selected Bucky
ERROR (BUCKY CONTACTS 2) was is in operation.
not satisfied for its
programmed state.
503 BUCKY CONTACTS 3 The state of INPUT 3 Check that the selected Bucky
ERROR (BUCKY CONTACTS 3) was is in operation.
not satisfied for its
programmed state.
504 INPUT 4 ERROR The state of INPUT 4 was 1. Check that the device
not satisfied for its connected to INPUT 4 is
programmed state. powered on and functioning
properly.
505 BUCKY CONTACTS 4 The state of INPUT 5 Check that the selected Bucky
ERROR (BUCKY CONTACTS 4) was is in operation.
not satisfied for its
programmed state.
506 INPUT 6 ERROR The state of INPUT 6 was 1. Check that the device
not satisfied for its connected to INPUT 6 is
programmed state. powered on and functioning
properly.
507 INPUT 7 ERROR The state of INPUT 7 was 1. Check that the device
not satisfied for its connected to INPUT 7 is
programmed state. powered on and functioning
properly.
509 ROOM DOOR The state of INPUT 9 Check that the Room Door is
INTERLOCK ERROR (ROOM DOOR closed.
INTERLOCK) was not
satisfied for its programmed
state.
510 II SAFETY ERROR The state of INPUT 10 (II Check that the Image
SAFETY) was not satisfied Intensifier is properly
for its programmed state. positioned.
511 INPUT 11 ERROR The state of INPUT 11 was 1. Check that the device
not satisfied for its connected to INPUT 11 is
programmed state. powered on and functioning
properly.
512 INPUT 12 ERROR The state of INPUT 12 was 1. Check that the device
not satisfied for its connected to INPUT 12 is
programmed state. powered on and functioning
properly.
513 INPUT 13 ERROR Reserved for INPUT 13 1. Check that the device
(REMOTE TOMO SELECT). connected to INPUT 13 is
powered on and functioning
properly.
514 INPUT 14 ERROR Reserved for INPUT 14 1. Check that the device
(MAG MODE INPUT 1). connected to INPUT 14 is
powered on and functioning
properly.
515 INPUT 15 ERROR Reserved for INPUT 15 1. Check that the device
(MAG MODE INPUT 2). connected to INPUT 15 is
powered on and functioning
properly.
516 INPUT 16 ERROR Reserved for INPUT 16 1. Check that the device
(MAG MODE INPUT 3). connected to INPUT 16 is
powered on and functioning
properly.
517 XRAY DISABLE ERROR The state of X-ray Disable 1. Check that the device
Input was not satisfied for its connected to X-RAY
programmed state. Disable INPUT is powered
on and functioning properly.
518 FLUORO RESET ERROR The state of Fluoro Reset 1. Check that the device
Input was not satisfied for its connected to Fluoro Reset
programmed state. INPUT is powered on and
functioning properly.
590 HS INPUT 1 ERROR The state of HS INPUT 1 1. Check that the device
was not satisfied for its connected to HS INPUT 1
programmed state. is powered on and
functioning properly.
591 HS INPUT 2 ERROR The state of HS INPUT 2 1. Check that the device
was not satisfied for its connected to HS INPUT 2
programmed state. is powered on and
functioning properly.
592 HS INPUT 3 ERROR The state of HS INPUT 3 1. Check that the device
was not satisfied for its connected to HS INPUT 3
programmed state. is powered on and
functioning properly.
593 HS INPUT 4 ERROR The state of HS INPUT 4 1. Check that the device
was not satisfied for its connected to HS INPUT 4
programmed state. is powered on and
functioning properly.
594 HS INPUT 5 ERROR The state of HS INPUT 5 1. Check that the device
was not satisfied for its connected to HS INPUT 5
programmed state. is powered on and
functioning properly.
595 HS INPUT 6 ERROR The state of HS INPUT 6 1. Check that the device
was not satisfied for its connected to HS INPUT 6
programmed state. is powered on and
functioning properly.
596 HS INPUT 7 ERROR The state of HS INPUT 7 1. Check that the device
was not satisfied for its connected to HS INPUT 7
programmed state. is powered on and
functioning properly.
597 HS INPUT 8 ERROR The state of HS INPUT 8 1. Check that the device
was not satisfied for its connected to HS INPUT 8
programmed state. is powered on and
functioning properly.
601 COMM ERROR The console failed to If the problem persists, consult
respond to a generator your service representative.
command after 3 attempts.
602 INVALID The message sent to the If the problem persists, consult
COMMUNICATION generator by the control your service representative.
MESSAGE console does not conform to
the generator protocol.
603 MESSAGE NOT The generator has received If the problem persists, consult
SUPPORTED a valid request from the your service representative.
console, but the generator
does not support the
requested function.
604 MESSAGE NOT The generator has received 1. Wait until the generator has
ALLOWED a valid request from the entered the required state.
console, but the generator is
in a state that does not allow 2. If the problem persists,
the requested function to be consult your service
executed. representative.
605 DIGITAL COMM ERROR The external handshake 1. Ensure that the Imaging
signal is missing. system is powered on and
operational.
606 ACB CANBUS COMM The main CPU failed to If the problem persists, consult
ERROR communicate with the CPU your service representative.
on the auxiliary control
board.
607 RIOB CANBUS COMM The main CPU failed to If the problem persists, consult
ERROR communicate with the CPU your service representative.
on the room I/O board.
608 DSS CANBUS COMM The main CPU failed to If the problem persists, consult
ERROR communicate with the CPU your service representative.
on the dual speed starter
board.
609 AEC CANBUS COMM The main CPU failed to If the problem persists, consult
ERROR communicate with the CPU your service representative.
on the AEC board.
610 COMM TIMEOUT The communication with an If the problem persists, consult
external device has timeout. your service representative.
703 INVALID LICENSE The generator received an If the problem persists, consult
invalid license code or file. your service representative
902 FACTORY DEFAULT The factory default switch is If the problem persists, consult
SWITCH CLOSED active at power on. your service representative.
903 PREP SWITCH CLOSED The prep input was active 1. Ensure that the prep switch
during the generator’s power is not activated.
on phase.
2. Clear the error. If the
problem persists, consult
your service representative.
904 X-RAY SWITCH The X-ray input was active 1. Ensure that the X-ray
CLOSED during the generator’s power switch is not activated.
on phase.
2. Clear the error. If the
problem persists, consult
your service representative.
905 FLUORO SWITCH The fluoro input was active 1. Ensure that the fluoro
CLOSED during the generator’s footswitch is not activated.
power-on phase.
2. Clear the error. If the
problem persists, consult
your service representative.
906 PREP TIMEOUT The generator has been in Reduce the length of time in
the Prep state too long. the Prep state.
907 RECEPTOR TIMEOUT The receptor is not Wait for receptor to be ready,
responding. then retry exposure.
909 GENERATOR KV LIMIT Indicates that the generator Increase or decrease the kV as
kV limit has been reached. required.
910 GENERATOR MA LIMIT Indicates that the generator Reduce the mA as required.
mA limit has been reached.
911 GENERATOR MS LIMIT Indicates that the generator Reduce the ms as required.
ms limit has been reached.
912 GENERATOR MAS Indicates that the generator Reduce the mAs as required.
LIMIT mAs limit has been reached.
913 TUBE KW LIMIT Indicates that the tube kW Reduce the kV, mA, mAs or
limit has been reached. ms to reduce the kW.
914 TUBE KV LIMIT Indicates that the tube kV Reduce the kV as required.
limit has been reached.
915 TUBE MA LIMIT Indicates that the tube mA Reduce the mA as required.
limit has been reached.
916 AEC DENSITY LIMIT Indicates that an AEC Reduce the AEC density.
density step is requested that
is outside of the programmed
AEC density range.
917 FLUORO PPS LIMIT Indicates that the fluoro PPS Reduce the PPS as required.
limit has been reached.
920 KV CHANGE NOT Attempts were made to Set the generator to a state
ALLOWED change the kV in the current that allows the kV to be
state. changed.
921 MA CHANGE NOT Attempts were made to Set the generator to a state
ALLOWED change the mA in the current that allows the mA to be
state. changed.
922 MS CHANGE NOT Attempts were made to Set the generator to a state
ALLOWED change the ms in the current that allows the ms to be
state. changed.
923 MAS CHANGE NOT Attempts were made to Set the generator to a state
ALLOWED change the mAs in the that allows the mAs to be
current state. changed.
924 RECEPTOR CHANGE Attempts were made to Set the generator to a state
NOT ALLOWED change the receptor in the that allows the receptor to be
current state. changed.
925 MODE CHANGE NOT Attempts were made to Set the generator to a state
ALLOWED change the mode (exposure that allows the mode (exposure
technique) in the current technique) to be changed.
state.
926 FOCUS CHANGE NOT Attempts were made to Set the generator to a state
ALLOWED change the focus in the that allows the focus to be
current state. changed.
927 TUBE CHANGE NOT An attempt to change the Ensure that generator is in the
ALLOWED tube via an external device proper state to allow tube
was not valid. changes via an external
device.
928 AEC CHANNEL An attempt to change the Ensure that generator is in the
CHANGE NOT AEC channel via an external proper state to allow AEC
ALLOWED device was not valid. channel changes via an
external device.
929 AEC DENSITY CHANGE Attempts were made to Set the generator to a state
NOT ALLOWED change the density in the that allows the density to be
current state. changed.
930 AEC FILM SCREEN Attempts were made to Set the generator to a state
CHANGE NOT change the film screen in the that allows the film screen to
ALLOWED current state. be changed.
931 AEC FIELDS CHANGE Attempts were made to Set the generator to a state
NOT ALLOWED change the AEC fields in the that allows the AEC fields to be
current state. changed.
932 MAG CHANGE NOT Attempts were made to Set the generator to a state
ALLOWED change the mag mode in the that allows the mag mode to
current state. be changed.
933 PPS CHANGE NOT Attempts were made to Set the generator to a state
ALLOWED change the fluoro PPS in the that allows the PPS to be
current state. changed.
934 APR COMMAND NOT Attempts were made to enter Set the generator to a state
ALLOWED an APR parameter in the that allows the APR parameter
current state. to be entered.
935 RECEPTOR NOT The selected receptor is not Consult your service
ENABLED enabled. representative.
936 FLUORO NOT ENABLED Fluoro is not enabled on the Consult your service
selected receptor. representative.
937 ABS NOT ENABLED Fluoro ABS is not enabled. Consult your service
representative.
938 PPS NOT ENABLED The selected PPS is not Consult your service
enabled. representative.
939 DAP NOT ENABLED DAP is not enabled. Consult your service
representative.
940 AK NOT ENABLED Air Kerma is not enabled. Consult your service
representative.
941 AEC NOT ENABLED AEC is not enabled for the Consult your service
selected receptor. representative.
942 AEC FILM SCREEN NOT The selected AEC film Consult your service
ENABLED screen is not enabled. representative.
943 AEC FIELD NOT The selected AEC field is not Consult your service
ENABLED enabled. representative.
944 NO TUBE SELECTED The selected receptor has no Consult your service
tube assigned. representative.
948 NO AEC FILM SCREEN No AEC Film Screen is Consult your service
SELECTED selected for the operation representative.
951 ANODE HEAT WARNING The anode has exceeded its 1. Wait for the tube to cool.
heat warning limit. Continued
exposures risk overheating 2. If problem persists, consult
the anode. your service
representative.
952 ANODE OVER HEAT The anode has exceeded its 1. Wait for the tube to cool.
heat limit.
2. If problem persists, consult
your service
representative.
955 DAP ACCUM WARNING The accumulated DAP value Reset the DAP.
has reached the
programmed limit.
956 DAP RATE WARNING The current DAP rate Reduce the generators dose
exceeds the programmed output via the kV, mA, mAs or
DAP rate limit. ms settings.
957 AK ACCUM WARNING The accumulated Air Kerma Reset the Air Kerma display.
value has reached the
programmed limit.
958 AK RATE WARNING The current Air Kerma rate Reduce the generators dose
exceeds the programmed Air output via the kV, mA, mAs or
Kerma rate limit. ms settings.
959 FLUORO TIMER The fluoro interval timer Reset the fluoro timer.
WARNING exceeds 4.2 minutes in FDA
mode or 4.6 minutes in IEC
mode.
960 FLUORO TIMER LIMIT The fluoro interval timer has Reset the fluoro interval timer.
reached 4.6 minutes in FDA
mode or a fluoro exposure
has been continuously
activated for a period of 9.6
minutes in IEC mode.
961 FLUORO ABS CHANGE An external system will not Consult your service
NOT ALLOWED allow the requested ABS representative.
mode change.
963 MAS TIMEOUT Desired mAs value was not Consult your service
achieved within allocated representative.
time.
964 AEC DOSE LIMIT Requested AEC Dose Consult your service
exceeds the AEC Dose representative.
Range specified in the AEC
Setup
966 FLUORO MAG NOT The selected Fluoro MAG is Consult your service
ENABLED not enabled. representative.
967 FLUORO ABS CURVE The selected Fluoro ABS Consult your service
NOT ENABLED curve is not enabled. representative.
968 CINE MAG CHANGE The selected Cine MAG Consult your service
NOT ALLOWED change is not allowed. representative.
969 CINE FRAME RATE The selected Cine frame rate Consult your service
CHANGE NOT change is not allowed. representative.
ALLOWED
970 CINE ABS CURVE NOT The Cine ABS Curve is Ensure the selected ABS curve
ENABLED disabled. is enabled.
971 CINE RUN TIME LIMIT The Cine Run Time Limit has Reset the Cine Run Time limit.
been exceeded.
972 RAD RUN TIME LIMIT The maximum Rad Run Reset the Rad Run Time limit.
Time has been exceeded.
990 REMOTE PREP SWITCH The remote prep input was 1. Ensure that the prep switch
CLOSED active during the generator’s is not activated.
power on phase.
2. Clear the error. If the
problem persists, consult
your service representative.
991 REMOTE X-RAY The remote X-ray input was 1. Ensure that the X-ray
SWITCH CLOSED active during the generator’s switch is not activated.
power on phase.
2. Clear the error. If the
problem persists, consult
your service representative.
992 REMOTE FLUORO The remote fluoro input was 1. Ensure that the fluoro
SWITCH CLOSED active during the generator’s footswitch is not activated.
power-on phase.
2. Clear the error. If the
problem persists, consult
your service representative.
993 CINE PREP SWITCH The Cine Prep Switch is 1. Ensure that the Prep switch
ACTIVE active during the generator’s is not activated.
power on phase.
2. Clear the error. If the
problem persists, consult
your service representative.
994 CINE X-RAY SWITCH The Cine X-ray switch is 1. Ensure that the Cine X-ray
ACTIVE active during the generator’s switch is not activated.
power on phase.
2. Clear the error. If the
problem persists, consult
your service representative.
In the above example, read down the 100 mA column to the 50 mAs value. Then read across
to the left side of the page to the 500 ms value, this will be the maximum time allowed for the
stated conditions.
Please review the note under AEC FUNCTIONAL CHECK in section 9. This details AEC
backup options that are available.
Time mA Selected
(ms) 500 630 800 1000
1.0 0.5 0.63 0.8 1.0
2.0 1.0 1.2 1.6 2.0
2.5 1.2 1.5 2.0 2.5
3.2 1.6 2.0 2.5 3.2
4.0 2.0 2.5 3.2 4.0
5.0 2.5 3.1 4.0 5.0
6.3 3.1 3.9 5.0 6.3
8.0 4.0 5.0 6.4 8.0
10 5.0 6.3 8.0 10
12.5 6.2 7.8 10 12
16 8.0 10 12 16
20 10 12 16 20
25 12 15 20 25
32 16 20 25 32
40 20 25 32 40
50 25 31 40 50
63 31 39 50 63
80 40 50 64 80
100 50 63 80 100
125 62 78 100 125
160 80 100 128 160
200 100 126 160 200
250 125 157 200 250
320 160 201 256 320
400 200 252 320 400
500 250 315 400 500
Time mA Selected
(ms) 500 630 800 1000
630 315 396 504 630
800 400 504 640 800
1000 500 630 800 1000
1250 625 787 1000
1600 800
2000 1000
2500
3200
4000
5000
6300
NOTE: 1) mA/TIME values are tube rating dependent. For certain tubes, some mA/TIME selections are not available at
higher kV selections.
2) The maximum power output of this family of generators is 50 kW, 65 kW, 80 kW, or 100 kW respectively. The
generator software does not allow a kV and mA combination to be selected that exceeds these maximum values.
TECHNIQUE SELECTION
(3 Point Selection: mA/TIME & mAs per IEC)
Copy for end user
Time mA Selected
(ms) 10 12.5 16 20 25 32 40 50 63 80 100 125 160 200 250 320 400
1.0 0.1 0.12 0.16 0.2 0.25 0.32 0.4
2.0 0.1 0.12 0.16 0.2 0.25 0.32 0.4 0.5 0.64 0.8
2.5 0.1 0.12 0.15 0.2 0.25 0.31 0.40 0.5 0.62 0.8 1.0
3.2 0.1 0.12 0.16 0.2 0.25 0.32 0.40 0.51 0.64 0.8 1.0 1.2
4.0 0.1 0.12 0.16 0.2 0.25 0.32 0.4 0.50 0.64 0.8 1.0 1.2 1.6
5.0 0.1 0.12 0.16 0.2 0.25 0.31 0.4 0.5 0.62 0.8 1.0 1.2 1.6 2.0
6.3 0.1 0.12 0.15 0.2 0.25 0.31 0.39 0.5 0.63 0.78 1.0 1.2 1.5 2.0 2.5
8.0 0.1 0.12 0.16 0.2 0.25 0.32 0.4 0.5 0.64 0.80 1.0 1.2 1.6 2.0 2.5 3.2
10 0.1 0.12 0.16 0.2 0.25 0.32 0.4 0.5 0.63 0.80 1.0 1.2 1.6 2.0 2.5 3.2 4.0
12.5 0.12 0.15 0.2 0.25 0.31 0.4 0.5 0.62 0.78 1.0 1.2 1.5 2.0 2.5 3.1 4.0 5.0
16 0.16 0.2 0.25 0.32 0.4 0.51 0.64 0.8 1.0 1.2 1.6 2.0 2.5 3.2 4.0 5.1 6.4
20 0.2 0.25 0.32 0.4 0.5 0.64 0.8 1.0 1.2 1.6 2.0 2.5 3.2 4.0 5.0 6.4 8.0
25 0.25 0.31 0.4 0.5 0.62 0.8 1.0 1.2 1.5 2.0 2.5 3.1 4.0 5.0 6.2 8.0 10
32 0.32 0.4 0.51 0.64 0.8 1.0 1.2 1.6 2.0 2.5 3.2 4.0 5.1 6.4 8.0 10 12
40 0.4 0.5 0.64 0.8 1.0 1.2 1.6 2.0 2.5 3.2 4.0 5.0 6.4 8.0 10 12 16
50 0.5 0.62 0.8 1.0 1.2 1.6 2.0 2.5 3.1 4.0 5.0 6.2 8.0 10 12 16 20
63 0.63 0.78 1.0 1.2 1.5 2.0 2.5 3.1 3.9 5.0 6.3 7.8 10 12 15 20 25
80 0.8 1.0 1.2 1.6 2.0 2.5 3.2 4.0 5.0 6.4 8.0 10 12 16 20 25 32
100 1.0 1.2 1.6 2.0 2.5 3.2 4.0 5.0 6.3 8.0 10 12 16 20 25 32 40
125 1.2 1.5 2.0 2.5 3.1 4.0 5.0 6.2 7.8 10 12 15 20 25 31 40 50
160 1.6 2.0 2.5 3.2 4.0 5.1 6.4 8.0 10 12 16 20 25 32 40 51 64
200 2.0 2.5 3.2 4.0 5.0 6.4 8.0 10 12 16 20 25 32 40 50 64 80
250 2.5 3.1 4.0 5.0 6.2 8.0 10 12 15 20 25 31 40 50 62 80 100
320 3.2 4.0 5.1 6.4 8.0 10 12 16 20 25 32 40 51 64 80 102 128
400 4.0 5.0 6.4 8.0 10 12 16 20 25 32 40 50 64 80 100 128 160
500 5.0 6.2 8.0 10 12 16 20 25 31 40 50 62 80 100 125 160 200
Time mA Selected
(ms) 500 630 800 1000
1.0 0.5 0.63 0.8 1.0
2.0 1.0 1.2 1.6 2.0
2.5 1.2 1.5 2.0 2.5
3.2 1.6 2.0 2.5 3.2
4.0 2.0 2.5 3.2 4.0
5.0 2.5 3.1 4.0 5.0
6.3 3.1 3.9 5.0 6.3
8.0 4.0 5.0 6.4 8.0
10 5.0 6.3 8.0 10
12.5 6.2 7.8 10 12
16 8.0 10 12 16
20 10 12 16 20
25 12 15 20 25
32 16 20 25 32
40 20 25 32 40
50 25 31 40 50
63 31 39 50 63
80 40 50 64 80
100 50 63 80 100
125 62 78 100 125
160 80 100 128 160
200 100 126 160 200
250 125 157 200 250
320 160 201 256 320
400 200 252 320 400
500 250 315 400 500
Time mA Selected
(ms) 10 12.5 16 20 25 32 40 50 63 80 100 125 160 200 250 320 400
630 6.3 7.8 10 12 15 20 25 31 39 50 63 78 100 126 157 201 252
800 8.0 10 12 16 20 25 32 40 50 64 80 100 128 160 200 800 320
1000 10 12 16 20 25 32 40 50 63 80 100 125 160 200 250 320 400
1250 12 15 20 25 31 40 50 62 78 100 125 156 200 250 312 400 500
1600 16 20 25 32 40 51 64 80 100 128 160 200 256 320 400 512 640
2000 20 25 32 40 50 64 80 100 126 160 200 250 320 400 500 640 800
2500 25 31 40 50 62 80 100 125 157 200 250 312 400 500 625 800 1000
3200 32 40 51 64 80 102 128 160 201 256 320 400 512 640 800
4000 40 50 64 80 100 128 160 200 252 320 400 500 640 800 1000
5000 50 62 80 100 125 160 200 250 315 400 500 625 800 1000
6300 63 78 100 126 160 201 252 315 396 504 630 787
Time mA Selected
(ms) 500 630 800 1000
630 315 396 504 630
800 400 504 640 800
1000 500 630 800 1000
1250 625 787 1000
1600 800
2000 1000
2500
3200
4000
5000
6300
NOTE: 1) mA/TIME values are tube rating dependent. For certain tubes, some mA/TIME selections are not available at
higher kV selections.
2) The maximum power output of this family of generators is 50 kW, 65 kW, 80 kW, or 100 kW respectively. The
generator software does not allow a kV and mA combination to be selected that exceeds these maximum values.
The second part of this section is the recommended maintenance schedule for your generator.
This maintenance schedule is recommended for all VZW2258 series X-ray generators.
The Operator must schedule all recommended and mandatory maintenance calls with the
RESPONSIBLE ORGANIZATION.
The following procedure may be used to verify that the AEC circuits are functioning on
generators equipped with AEC (automatic exposure control).
1. Switch the generator ON, and select an appropriate radiographic image receptor.
2. Align the X-ray tube and the selected image receptor such that the central ray is directly
over the center field of the AEC pickup device. Set the focal spot to film plane distance to
40 in. (1 m).
3. Select AEC mode of operation. Select center field, large focus.
MINIMUM EXPOSURE TIME:
4. With no object in the radiation field, adjust the collimator or beam-limiting device to project
a 10 in. X 10 in. (24 cm X 24 cm) field at the image receptor.
5. Select 80 kVp, 100 mA, and a backup mAs of 50 if it is operator selectable. If this is not
operator selectable, the default AEC backup settings must be used. Refer to the note at
the end of this page to determine the AEC backup mode that has been programmed.
6. Make an exposure and verify that the POST mAs reading is ≤ 2 mAs.
MAXIMUM EXPOSURE TIME:
7. Close the collimator or beam-limiting device completely. Place a folded lead apron over
the image receptor.
8. Select 60 kVp, 100 mA, and a backup mAs of 50 if it is operator selectable. If this is not
operator selectable, the default backup settings must be used. Refer to the note at the
end of this page to determine the AEC backup mode that has been programmed.
9. Make an exposure and verify that the error message AEC BUT ERROR or AEC BU MAS
ERR is indicated in the APR display window after the exposure has terminated. This
confirms that the exposure has continued until it was terminated by the AEC backup
timer.
NOTE: The AEC backup mode is installer programmable. Three modes are available, FIXED,
MAS, and MS. Only the MAS mode allows the operator to set the backup mAs for an AEC
exposure.
FIXED: The generator will determine the maximum AEC backup time, not to exceed preset
AEC backup mAs/ms values or system limits. The characters AEC will be displayed in the time
window of the LED display during AEC operation.
MAS: Allows the operator to adjust the AEC backup mAs, not to exceed preset AEC backup
mAs/ms values or system limits. The mAs value will be displayed in the time window of the
LED display during AEC operation.
MS: Allows the operator to adjust the AEC backup ms, not to exceed preset AEC backup
mAs/ms values or system limits. The ms value will be displayed in the time window of the LED
display during AEC operation.
The following procedure may be used to verify that the ABS circuits are functioning on
generators equipped with ABS (automatic brightness stabilization).
1. Switch the generator ON, and select an appropriate fluoroscopic image receptor.
2. Select ABS mode. The indicator LED adjacent to the ABS selector button on the control
console or on the remote fluoro control will be lit to indicate that ABS is ON.
3. Open the collimator or beam limiting device shutters to maximum. With no object in the
image field, press the fluoro footswitch. Confirm that the collimator or beam-limiting
device is fully open by observing the monitor.
4. While continuing to press the fluoro footswitch, note the fluoro kV display at the control
console or remote fluoro control. This should be less than 55 kV.
5. Close the collimator or beam-limiting device and press the fluoro footswitch. With no
object in the image field, ensure that the radiation is fully blocked by observing the
monitor. If necessary, cover the image intensifier with lead aprons to block all radiation.
6. While continuing to press the fluoro footswitch, note the fluoro kV display at the control
console or remote fluoro control. This should be the maximum available fluoro kV, 110 kV
or 125 kV (depending on whether the upper fluoro kV limit has been set to 110 kV or 125
kV).
7. Place an absorber (20 cm of water, or 1.5 in. (3.8 cm) of pure aluminum, or equivalent) in
the center of the image field.
8. Press the fluoro footswitch. While viewing the absorber on the monitor, open the
collimator or beam-limiting device sufficiently to cover as much of the absorber as
possible. Ensure that the X-ray beam does not extend beyond the sides of the absorber,
as stray radiation will adversely affect the kVp reading in the next step.
9. While continuing to press the fluoro footswitch, note the fluoro kV display at the control
console or remote fluoro control. This should stabilize between 70 and 80 kVp.
The user must routinely check the X-ray equipment for apparent defects or damage. Report
any apparent defects or irregular operation of any equipment to service personnel
immediately, and discontinue use of the suspected faulty equipment until repairs are made.
Continuing operation with faulty equipment may present various safety hazards, including risk
of increased radiation.
1. Check all visual displays (warning and status lights, technique displays and indicators,
etc) for normal operation.
2. Check all audible indicators (speakers, beepers, etc) for normal operation, and check that
the loudness settings are adequate for the environment.
3. Check all interconnect cables and connectors for damage.
Note: The high voltage receptacles on the H.V module must either be filled with insulating oil (preferred)
or the high voltage connectors at the H.V module end must be coated with vapour-proof insulating
compound. The method used at the X-ray tube’s high voltage connectors is at the discretion of the
installer / service engineer, however it must be adequate to prevent arcs / flashover at the X-ray tube
end as tube high voltage arcs will impact the generator’s usability.
CLEANING
• Do not use any liquids or solvents such as flux, water, acetone, ethanol, isopropyl alcohol,
toluene, or ammonia to clean the surface of the R & F console LED display. Use a dry
clean cloth and gentle wipe the LED display to properly clean.
• To clean the console surfaces (without disinfectant) use soapy water. This should be
applied using a lint-free cloth as described below.
• If console surfaces need to be disinfected, use Actichlor or equivalent with a Hypochlorite
concentration of 3 to 5%.
∗ Dilute the disinfectant 50/50 with water.
∗ Spray or dampen a lint-free cloth with the diluted disinfectant. The cloth must be damp
only, not wet.
∗ Gently wipe the console surfaces with the damp cloth.
• Never use any corrosive, solvent or abrasive detergents or polishes.
• Ensure that no water or other liquid can enter any equipment. This precaution prevents
short circuits and corrosion forming on components.
• Methods of disinfection used must conform to legal regulations and guidelines regarding
disinfection and explosion protection.
• If disinfectants are used which form explosive mixtures of gases, these gases must have
evaporated before switching on the equipment again.
• Disinfection by spraying is not recommended because the disinfectant may enter the X-ray
equipment.
• If room disinfection is done with an atomizer, it is recommended that the equipment be
switched OFF, allowed to cool down and covered with a plastic sheet. When the
disinfectant mist has subsided, the plastic sheet may be removed and the equipment be
disinfected by wiping.
The following pages contain data for the X-ray tube(s) used with your X-ray generator at the
time of installation.
NOTE TO INSTALLER: