CERTIFICATE OF CALIBRATION

Manufacturer : DLAB Certifcation Number : CAL202102020014


Model : 0-0.2uL Description : Pipettor
Serial Number : YM206AK0000644
Customer Name : ZCMC

Method Used : Comparison Method Temperature : 25°C


Calibration Date : Febuary 2, 2021 Humidity : 60 % RH
Recommended Due Date : Febuary 2, 2022

This calibration was conducted using standards traceable to National Institute of Standards and
Technology (NIST). The user may be obliged to have the product recalibrated at appropriate intervals.
This report applies only to the item identified and shall not be reproduced, except in full, without the
written approval of the calibration laboratory.
This calibration results pertains to the date of calibration. Should the unit be modified or damaged in
any way, or develop inconsistent readings, the result may not be valid and the recalibration of the unit
will be required. The user should determine the suitability of the instrument for its intended use.

Approved by:

_____________________
Iris Clint Q. Caro
Service Supervisor
CALIBRATION REPORT

Manufacturer : DLAB Certifcation Number : CAL202102020014


Model : 0-0.2uL Description : Pipettor
Serial Number : YM206AK0000644 Calibration Date : Febuary 2, 2021

CALIBRATION DATA

REFERENCE
UNIT UNDER REFERENCE STANDARD READING (g)
STANDARD
TEST ERROR
MEAN
SETTING(uL) 1 2 3 4 5
READING (g)
0.2 0.00017 0.00019 0.00018 0.00019 0.00018 0.000183 -0.000017

Calibration Equipment Used:

MANUFACTURER MODEL SERIAL NUMBER CAL DUE DATE TRACEABILITY

BIOBASE BA2204C 039805 March 13, 2022 NIST

Performed by:

___________________
Iris Clint Q. Caro
Service Supervisor