PT100 887857-1
PT100 887857-1
The traceability of measurement results to the SI units is assured by the National standard
Metrological traceability: maintained at Central Organization for standardization and Quality Control through
calibration at :- UME (G1KS-0248)
Environmental conditions
Temp. 32.51 ° C RH.% 35.8%
of calibration:
Results
Set Ref. ( R) UUC (M) ERROR Uncertainty
C° C° C° =(M)-( R) °C ± C°
0 -0.148 0.090 0.238 0.687
50 49.712 50.336 0.624 0.687
100 99.932 100.664 0.733 0.686
150 149.135 150.017 0.883 0.686
Observations , opinions or The Results in the table should be taken in to consideration
recommendations: according to your decision rule
Approved By :
Calibration Specielist : Reveiwed : Ban Omer Farooq
Khalid Naser Hanaa Mohammed Head of Physics Section
6/25/2022 6/25/2022 6/25/2022
This certificate is issued accordance to the laboratory accreditation requirements. It provides traceability of measurement to recognized national standards and to
the (SI) unit of measurement realized at the COSQC or other recognized national standards laboratories. This certificate may not be reproduced other than
photographic process. This certificate refers only to the particular item submitted for calibration
1 of 1
Reading Of REF. ˚ C Set ˚ C Reading Of UUC ˚ C
DOWN
-0.0840
0 UP
0.088
DOWN
0.09
-0.0840 0.088 0.09
-0.0840 0.088 0.09
-0.0840 0.088 0.09
-0.0840 0.088 0.09
-0.0750 0.09 0.091
-0.0750 0.09 0.091
-0.0750 0.09 0.091
-0.0750 0.09 0.091
-0.0750 0.09 0.091
-0.0770 0.09 0.09
-0.0770 0.09 0.09
-0.0770 0.09 0.09
-0.0770 0.09 0.09
-0.0770 0.09 0.09
-0.0790 0.089 0.09
-0.0790 0.089 0.09
-0.0790 0.089 0.09
-0.0790 0.089 0.09
-0.0790 0.089 0.09
-0.7600 0.091 0.091
-0.7600 0.091 0.091
-0.7600 0.091 0.091
-0.7600 0.091 0.091
-0.7600 0.091 0.091
-0.1476 Average 0.0896 0.0904
00 + READER) Res.
0.001
Reading Of REF
READER) Ref. Error UP
0.000 -0.078
-0.078
857) -0.078
Uncertainty -0.078
0.000 -0.078
0.038 -0.088
0.035 -0.088
0.022 -0.088
0.040 -0.088
0.040 -0.088
-0.079
-0.079
الحمام
-0.079
الكحولي
Set. Stability Uniformity -0.079
-40 0.002 0.039 -0.079
-30 0.002 0.032 -0.08
-20 0.002 0.008 -0.08
-10 0.002 0.026 -0.08
0 0.002 0.048 -0.08
10 0.002 0.019 -0.08
20 0.003 0.016 -0.076
30 0.003 0.005 -0.076
0 0.002 0.022 -0.076
-0.076
-0.076
-0.1476
Bath Stability Ref. Uncertainty (PT100 + READER
0.003 0.100
Bath Uniformity Drift (PT100 + READER)
0.028 0.300
PT100 ( 887857)
SET Uncertainty
-40 0.000
0 0.038
50 0.035
100 0.022
150 0.040
200 0.040
PT100 ( 887857)
rift و تاتي من الشهاد الجديدة و القديمة SET Uncertainty
87857 ( المرجعي الثانوي ذو التسلسل -40 0.000
0 0.038
50 0.035
100 0.022
150 0.040
200 0.040
ام المائي
Set.
10
20
30
40
50
60
70
80
85
50
Reading Of REF. ˚ C Set ˚ C Reading Of UUC ˚ C
DOWN
99.9670
100 UP
100.663
DOWN
100.668
99.9670 100.663 100.668
99.9670 100.663 100.668
99.9670 100.663 100.668
99.9670 100.663 100.668
99.9680 100.662 100.667
99.9680 100.662 100.667
99.9680 100.662 100.667
99.9680 100.662 100.667
99.9680 100.662 100.667
99.9690 100.663 100.667
99.9690 100.663 100.667
99.9690 100.663 100.667
99.9690 100.663 100.667
99.9690 100.663 100.667
99.9680 100.662 100.662
99.9680 100.662 100.662
99.9680 100.662 100.662
99.9680 100.662 100.662
99.9680 100.662 100.662
99.9670 100.665 100.663
99.9670 100.665 100.663
99.9670 100.665 100.663
99.9670 100.665 100.663
99.9670 100.665 100.663
99.9317 Average 100.663 100.6654
100 + READER) Res.
0.001
Reading Of REF
READER) Ref. Error UP
0.000 99.961
99.961
7857) 99.961
Uncertainty 99.961
0.000 99.961
0.038 99.63
0.035 99.63
0.022 99.63
0.040 99.63
0.040 99.63
99.962
99.962
الحمام الكحولي 99.962
Set. Stability Uniformity 99.962
-40 0.002 0.039 99.962
-30 0.002 0.032 99.962
-20 0.002 0.008 99.962
-10 0.002 0.026 99.962
0 0.002 0.048 99.962
10 0.002 0.019 99.962
20 0.003 0.016 99.963
30 0.003 0.005 99.963
100 0.002 0.022 99.963
99.963
99.963
99.9317
Bath Stability Ref. Uncertainty (PT100 + READER)
0.005 0.100
Bath Uniformity Drift (PT100 + READER)
0.010 0.300
PT100 ( 887857)
SET Uncertainty
-40 0.000
0 0.038
50 0.035
100 0.022
150 0.040
200 0.040
PT100 ( 887857)
SET Uncertainty
-40 0.000
0 0.038
50 0.035
100 0.022
150 0.040
200 0.040
PT100 ( 887857)
SET Uncertainty
-40 0.000
0 0.038
50 0.035
100 0.022
150 0.040
200 0.040
PT100 ( 887857)
SET Uncertainty
-40 0.000
0 0.038
50 0.035
100 0.022
150 0.040
200 0.040
SET 100 mm 80 mm 60 mm 40 mm 20 mm
UP Ref. UUC Ref. UUC Ref. UUC Ref. UUC Ref. UUC
-0.078 0.088 -0.088 0.090 -0.079 0.090 -0.080 0.089 -0.076 0.091 Bath Stability
-0.078 0.088 -0.088 0.090 -0.079 0.090 -0.080 0.089 -0.076 0.091 0.003
0
-0.078 0.088 -0.088 0.090 -0.079 0.090 -0.080 0.089 -0.076 0.091
-0.078 0.088 -0.088 0.090 -0.079 0.090 -0.080 0.089 -0.076 0.091 Bath Uniformity
Bath :مـائـي -0.078 0.088 -0.088 0.090 -0.079 0.090 -0.080 0.089 -0.076 0.091 0.028
NO.1
SET 20 mm 40 mm 60 mm 80 mm 100 mm
DOWN Ref. UUC Ref. UUC Ref. UUC Ref. UUC Ref. UUC Drift PT
-0.084 0.090 -0.075 0.091 -0.077 0.090 -0.079 0.090 -0.760 0.091 0.3
-0.084 0.090 -0.075 0.091 -0.077 0.090 -0.079 0.090 -0.760 0.091
0 -0.084 0.090 -0.075 0.091 -0.077 0.090 -0.079 0.090 -0.760 0.091 Uncertainty PT
-0.084 0.090 -0.075 0.091 -0.077 0.090 -0.079 0.090 -0.760 0.091 0.1
QF-7.8-02-C-1 : وثيقة رقم
(الجهاز المركزي للتقييس والسيطرة النوعيةCOSQC ) 01 : رقم االصدار
قسم المقاييس/ دائرة التقييس 2022-5-15 : تاريخ االصدار
شعبة القياسات الفيزياوية 2022-6-1 : تاريخ التفعيل
Data Sheet For PT100
مختبر الحرارية/ شعبة القياسات الفيزياوية Certificate No. : PHT /2023 Date of issue : 6/25/2022
Customer :
بغداد- العراق Date of calibration : 6/13/2023 Validation : One Year
Description : PT100 With Chub ( Fluke ) Calibration method : QP-7.2-01-C
Manufacturer : Fluke calibration quantity : Temperature ° C
Model : 5615 Standard ref. : Super Daq (1586 A) With PRT PT100 (3407)
Serial No. : 887857 Place of calibration : PH LAB . ( 1 )
Range : (-200 ----- 660) ˚C Date of Reception : / /2022 Order no. : ـــــــــــــ
Digital
0 : √ Analog : Res. : ° C 0.001 Calibration conditions : Temp. : 32.51 ° C RH.% 35.8%
-0.084 0.090 -0.075 0.091 -0.077 0.090 -0.079 0.090 -0.760 0.091
BATHS PT 1 PT 2 PT 3 3408 887861 887857
Drift PT
Alcohol Water Oil (HETO) Oil (Fluke ) 0.22 0.21 0.01 0.3 0.3
Stability 0.004 0.003 0.005 0.003 PT 1 PT 2 PT 3 3408 887861 887857
Un. PT
Uniformity 0.020 0.028 0.010 0.004 0.29 0.28 0.06 0.2 0.1
SET 100 mm 80 mm 60 mm 40 mm 20 mm
UP Ref. UUC Ref. UUC Ref. UUC Ref. UUC Ref. UUC
49.713 50.330 49.712 50.334 49.710 50.333 49.711 50.335 49.713 50.336 Bath Stability
49.713 50.330 49.712 50.334 49.710 50.333 49.711 50.335 49.713 50.336 0.003
50
49.713 50.330 49.712 50.334 49.710 50.333 49.711 50.335 49.713 50.336
49.713 50.330 49.712 50.334 49.710 50.333 49.711 50.335 49.713 50.336 Bath Uniformity
Bath :مـائـي 49.713 50.330 49.712 50.334 49.710 50.333 49.711 50.335 49.713 50.336 0.028
NO.2
SET 20 mm 40 mm 60 mm 80 mm 100 mm
DOWN Ref. UUC Ref. UUC Ref. UUC Ref. UUC Ref. UUC Drift PT
49.710 50.337 49.713 50.337 49.712 50.338 49.710 50.337 49.712 50.338 0.3
49.710 50.337 49.713 50.337 49.712 50.338 49.710 50.337 49.712 50.338
50 49.710 50.337 49.713 50.337 49.712 50.338 49.710 50.337 49.712 50.338 Uncertainty PT
QF-7.8-02-C-1 : وثيقة رقم
(الجهاز المركزي للتقييس والسيطرة النوعيةCOSQC ) 01 : رقم االصدار
قسم المقاييس/ دائرة التقييس 2022-5-15 : تاريخ االصدار
شعبة القياسات الفيزياوية 2022-6-1 : تاريخ التفعيل
Data Sheet For PT100
مختبر الحرارية/ شعبة القياسات الفيزياوية Certificate No. : PHT /2023 Date of issue : 6/25/2022
Customer :
بغداد- العراق Date of calibration : 6/13/2023 Validation : One Year
Description : PT100 With Chub ( Fluke ) Calibration method : QP-7.2-01-C
Manufacturer : Fluke calibration quantity : Temperature ° C
Model : 5615 Standard ref. : Super Daq (1586 A) With PRT PT100 (3407)
Serial No. : 887857 Place of calibration : PH LAB . ( 1 )
Range : (-200 ----- 660) ˚C Date of Reception : / /2022 Order no. : ـــــــــــــ
Digital : √ Analog : Res. : ° C 0.001 Calibration conditions : Temp. : 32.51 ° C RH.% 35.8%
50
49.710 50.337 49.713 50.337 49.712 50.338 49.710 50.337 49.712 50.338 0.1
49.710 50.337 49.713 50.337 49.712 50.338 49.710 50.337 49.712 50.338
100 99.967 100.668 99.968 100.667 99.969 100.667 99.968 100.662 99.967 100.663 Uncertainty PT
99.967 100.668 99.968 100.667 99.969 100.667 99.968 100.662 99.967 100.663 0.1
99.967 100.668 99.968 100.667 99.969 100.667 99.968 100.662 99.967 100.663
150
QF-7.8-02-C-1 : وثيقة رقم
(الجهاز المركزي للتقييس والسيطرة النوعيةCOSQC ) 01 : رقم االصدار
قسم المقاييس/ دائرة التقييس 2022-5-15 : تاريخ االصدار
شعبة القياسات الفيزياوية 2022-6-1 : تاريخ التفعيل
Data Sheet For PT100
مختبر الحرارية/ شعبة القياسات الفيزياوية Certificate No. : PHT /2023 Date of issue : 6/25/2022
Customer :
بغداد- العراق Date of calibration : 6/13/2023 Validation : One Year
Description : PT100 With Chub ( Fluke ) Calibration method : QP-7.2-01-C
Manufacturer : Fluke calibration quantity : Temperature ° C
Model : 5615 Standard ref. : Super Daq (1586 A) With PRT PT100 (3407)
Serial No. : 887857 Place of calibration : PH LAB . ( 1 )
Range
No.4 : (-200 ----- 660) ˚C Date of Reception : / /2022 Order no. : ـــــــــــــ
Digital : √ Analog : Res. : ° C 0.001 Calibration conditions : Temp. : 32.51 ° C RH.% 35.8%
149.135 150.014 149.136 150.010 149.132 150.009 149.133 150.009 149.129 150.008
150 149.135 150.014 149.136 150.010 149.132 150.009 149.133 150.009 149.129 150.008 Uncertainty PT
149.135 150.014 149.136 150.010 149.132 150.009 149.133 150.009 149.129 150.008 0.1
149.135 150.014 149.136 150.010 149.132 150.009 149.133 150.009 149.129 150.008
Signature Signature
Calibration Specielist : Name Khalid Naser Reveiwed : Name Hanaa Mohammed
Date 6/13/2023 Date 6/25/2022