Professional Documents
Culture Documents
A C Q U I S I TI O N I N F O R M A TI O N
TE S T D E S C R I P TI O N
In the MS Negative Mass Accuracy test, three (3) on column injections of a Sulfadimethoxine solution are made
into a flow of 0.8 mL/min mobile phase - water (0.1 % formic acid) and acetonitrile (0.1 % formic acid).
The qualification passes if the average measured mass of the Sulfadimethoxine ion for the three replicates is
< or = to +/- 0.2 Da (200 mDa) compared to the theoretical [M - H] - m/z.
A C Q U I S I TI O N I N F O R M A TI O N
Sulfadimethoxine - 2.213
200.00 m/z
250.00 300.00 350.00 400.00 450.00 500.00 550.00
309.16
1.2x106
1.0x106
Sulfadimethoxine - 2.213
8.0x105
Intensity
6.0x105
4.0x105
2.0x105
0.0
1.00 1.20 1.40 1.60 1.80 2.00 2.20 2.40 2.60 2.80 3.00 3.20 3.40 3.60 3.80 4.00
Minutes
Injection 1
A C Q U I S I TI O N I N F O R M A TI O N
Sulfadimethoxine - 2.209
150.00 m/z
200.00 250.00 300.00 350.00 400.00 450.00 500.00 550.00
309.16
1.2x106
1.0x106
Sulfadimethoxine - 2.209
8.0x105
Intensity
6.0x105
4.0x105
2.0x105
0.0
1.00 1.20 1.40 1.60 1.80 2.00 2.20 2.40 2.60 2.80 3.00 3.20 3.40 3.60 3.80 4.00
Minutes
Injection 2
A C Q U I S I TI O N I N F O R M A TI O N
Sulfadimethoxine - 2.198
200.00 m/z
250.00 300.00 350.00 400.00 450.00 500.00 550.00
309.16
1.2x106
1.0x106
Sulfadimethoxine - 2.198
8.0x105
Intensity
6.0x105
4.0x105
2.0x105
0.0
1.00 1.20 1.40 1.60 1.80 2.00 2.20 2.40 2.60 2.80 3.00 3.20 3.40 3.60 3.80 4.00
Minutes
Injection 3
A C Q U I S I TI O N I N F O R M A TI O N
Mean 91
Sulfadimethoxine Mass
Accuracy Results
Summary
MS Neg Mass Accuracy (mDa)
1 91