JPPL/WI/LAB/MFI-01

Jayashree polymers pvt Ltd Rev: 00


Rev. Date: 31.03.2022

MFI TESTER (Auto Cutter)


LOCATION :- LAB

1. Press the Main switch for Temp


2. Set temp as per material tc. and wait until
3. Clean barral by using cotton and rod 4. Clean the nozle head with help of cotton 5. Clean the Nozzle with help of screw and
temp came with set temp. and rod. cotton.

100 90

10. After start timer rod will automatically


9. Start timer and verify the shot 8. Select Weight as per Material TC information and 7. After coming temp as per MTC then put material
coming dowon. No need to press from upper (time :- 60 Sec) put on rod inside the Barrle. 6. Set nozzle inside the barrle.
side.

130 140
150

11. After complected of 60 sec of timer then Auto cutter 12. Remove the shot from Machine and 13. then start timer for same process for next 14. Start weight Machine 15. The instrument performs a self-test. Wait
will rotate with cut RM shot and RM will seprate with
machine. store other side. shot. Take 10 shots and check MFI (Weighing machine). until the display shows 0.000 g.

To Switch Off the


Main Switch and Made Report of NOTE DOWN THE ALL
result
READING IN FORMAT
AND CALCULATE THE
MFI
19. Then switch off the all instruments 18. After MFI process done then remove material 16. take weight of these 10 Shot and note down in
inside the barrle and clean the nozzle. the report

Rev. NO. Rev. Date Rev. Description


0 31.03.2022 New Release

Prepared BY:- Jayaganesh Approved By:- Kumarasamy