Professional Documents
Culture Documents
Carryover RM
Carryover RM
A C Q U I S I TI O N I N F O R M A TI O N
System Name: BSM_SM_TQD Acquired By: sqte
Project Name: SQT_E_BSM_SM_TQD_data Column serial Number:
Sample Set Method: 070 Carryover SSM Sample Kit 1 Expiration Date:
Processing Method SMDO PM Sample Kit 1 Lot Number:
Report Method Name: Carryover RM Build Version: Empower 2 Software Build 2154 SPs
User Name: System Sample Set Id: 2694 Result Set Id 3071
Sample Set Name: Carryover SSM Sample Set Start Date: 1/4/2011 11:31:01 PM EST
In the Carryover qualification, three pre-blank injections are performed to determine if the chromatograph is
clean enough to measure carryover. T hose injections are followed by three Sulfadimethoxine calibration injections
at a concentration of (5 ppm) which is 0.005 percent of the Challenge injection concentration (100000 ppm).
Carryover Carryover
SampleName Peak Type % Carryover
Specification Pass Fail
1 PostBlank1 < or = 0.005% Found 0.003 PASS
A C Q U I S I TI O N I N F O R M A TI O N
System Name: BSM_SM_TQD Acquired By: sqte
Project Name: SQT_E_BSM_SM_TQD_data Column serial Number:
Sample Set Method: 070 Carryover SSM Sample Kit 1 Expiration Date:
Processing Method SMDO PM Sample Kit 1 Lot Number:
Report Method Name: Carryover RM Build Version: Empower 2 Software Build 2154 SPs
User Name: System Sample Set Id: 2694 Result Set Id 3071
Sample Set Name: Carryover SSM Sample Set Start Date: 1/4/2011 11:31:01 PM EST
Intensity
Intensity
4000.0 4000.0 4000.0
PreBlank; Vial: 2:F,8; Inj: 1 PreBlank; Vial: 2:F,8; Inj: 2 PreBlank; Vial: 2:F,8; Inj: 3
10000.0 10000.0 10000.0
Sulfadimethoxine
6000.0 6000.0 6000.0
Intensity
Intensity
Intensity
Std 0.005%; Vial: 2:E,8; Inj: 1 Std 0.005%; Vial: 2:E,8; Inj: 2 Std 0.005%; Vial: 2:E,8; Inj: 3
10000.0 10000.0 10000.0
Sulfadimethoxine
Intensity
Intensity
Stock 100 ng/µL; Vial: 2:F,5; Inj: 1 PostBlank1; Vial: 2:F,6; Inj: 1 PostBlank2; Vial: 2:F,6; Inj: 1
A C Q U I S I TI O N I N F O R M A TI O N
System Name: BSM_SM_TQD Acquired By: sqte
Project Name: SQT_E_BSM_SM_TQD_data Column serial Number:
Sample Set Method: 070 Carryover SSM Sample Kit 1 Expiration Date:
Processing Method SMDO PM Sample Kit 1 Lot Number:
Report Method Name: Carryover RM Build Version: Empower 2 Software Build 2154 SPs
User Name: System Sample Set Id: 2694 Result Set Id 3071
Sample Set Name: Carryover SSM Sample Set Start Date: 1/4/2011 11:31:01 PM EST
10000.0
8000.0
Sulfadimethoxine
6000.0
Intensity
4000.0
2000.0
0.0
1.00 2.00
Minutes
A C Q U I S I TI O N I N F O R M A TI O N
System Name: BSM_SM_TQD Acquired By: sqte
Project Name: SQT_E_BSM_SM_TQD_data Column serial Number:
Sample Set Method: 070 Carryover SSM Sample Kit 1 Expiration Date:
Processing Method SMDO PM Sample Kit 1 Lot Number:
Report Method Name: Carryover RM Build Version: Empower 2 Software Build 2154 SPs
User Name: System Sample Set Id: 2694 Result Set Id 3071
Sample Set Name: Carryover SSM Sample Set Start Date: 1/4/2011 11:31:01 PM EST
Peak Results
Sample Type: Unknow n
SampleName Sample Type Vial Injection Name RT Area Height Peak Type % Carryover
1 PreBlank Unknown 2:F,8 1 Sulfadimethoxine 1.500 Missing
2 PreBlank Unknown 2:F,8 2 Sulfadimethoxine 1.500 Missing
3 PreBlank Unknown 2:F,8 3 Sulfadimethoxine 1.500 Missing
Peak Results
Sample Type: Standard
SampleName Sample Type Vial Injection Name RT Area Height Peak Type % Carryover
1 Std 0.005% Standard 2:E,8 1 Sulfadimethoxine 1.496 245095 98929 Found 0.0050
2 Std 0.005% Standard 2:E,8 2 Sulfadimethoxine 1.494 237294 93166 Found 0.0050
3 Std 0.005% Standard 2:E,8 3 Sulfadimethoxine 1.495 241367 96797 Found 0.0050
Peak Results
Sample Type: Unknow n
SampleName Sample Type Vial Injection Name RT Area Height Peak Type % Carryover
1 PostBlank1 Unknown 2:F,6 1 Sulfadimethoxine 1.495 125041 48541 Found 0.0026
2 PostBlank2 Unknown 2:F,6 1 Sulfadimethoxine 1.504 36321 14433 Found 0.0008
3 PostBlank3 Unknown 2:F,6 1 Sulfadimethoxine 1.500 19908 8028 Found 0.0004