Test definition ALT 2 Calibr. parameters
Test type Photometric Test in use YES Calibration type NONE Factor -1481 Bias 0
Full name ALT IFCC, 2 Low High Units
reag.system
Online name Test limit 6 1500 U/l Bias corr. in use NO
Result Unit U/l low( Ö ), high
Acceptance AUTOMATIC
Volume (µl) Disp. with Volume (µl) λ 2 (nm) 20 ALT1 Normal 011 7
100 EXTRA 25 380
Disp. with Volume (µl) Disp. with Res. Net. Abs. ALT2 Normal 012 7
EXTRA 15 EXTRA 0.5
Time (sec)
240
Volume (µl) Dilution with Volume (µl)
20 WATER 8
Points & Inter.
Konelab
Wash reagent Wash reagent Wash reagent 60/30/20XT 9/27
Konelab 20 6/42
NONE NONE NONE
Konelab™ ALT (IFCC, two reagents) - code: 981361 (7 x 50 ml), code: 981769 (8 x 20 ml)
Reagent: Code 981361: Add 400 µl of Pyridoxal phosphate 981790 into one bottle of reagent A. Reagent B is ready to use.
Code 981769: Add 160 µl of Pyridoxal phosphate 981790 into one bottle of reagent A. Reagent B is ready to use.
Ö measuring range low
Q.C. sample: Nortrol 981043 & Abtrol 981044. Reconstitute with 5.00 ml of distilled water.