You are on page 1of 1
Page 9-70 PW1660 Service Manual 910630 9.10.2.2 mA Control C SOkV/1ma 2OKV/1ma ttm | 7 son [4 Setting Test point X2 Sotting Test point X2 ran 6 i: 40KV/1mA 4,56V 40KV/4mA 4.68V, SOKV/ImA 4.69V SOKV/4mA 4.79V 9.10.3 HT Generator Trouble Shooting, Many instability of the HT generator is suspected then the checks in Section 9, 10.2 must be carried out. If the generator keeps on switching off or gives unstable measuring results than the following checks can be made to trace the error, WARNING! Always switch the instrument off; before exchanging a compo- nent or fitting a board on an extension board, for measurement purposes. Use the generator screen to monitor the HT and safety circuit information. Use the special test mode to alter the HT settings, Required measuring equipment. ‘Multimeter For example PM2718 or Fluke 87 of equivalent Oscilloscope —:——For example PM3256 or equivalent, Current probe For exampie PM9355 or equivalent, Bo one edeptet > Conmmoted paral wt bi arent probe (

You might also like