Professional Documents
Culture Documents
All rights are reserved. No part of this publication may be reproduced in any material form (including
photocopying or storing in any medium by electronic means and whether or not transiently or
incidentally to some other use of this publication) without the written permission.
EdLabQuip GmbH
Peiner Str. 157E, 30519 Hannover, Germany
www.edlabquip.com
sales@edlabquip.com
Tel: +49 511 56 37 98 70
FM 250 Flow Measurement Test Bench
CONTENTS
1
FM 250 Flow Measurement Test Bench
for consequences of failure or damage of the equipment in any way if the equipment is not installed,
maintained and operated as outlined in the instruction manual published by the manufacturers or
suppliers.
(c) In order to safeguard the students and operators of the equipment it is vital that all safety aspects as
outlined in the instruction manual are observed.
3. Safety Guidelines
3.1 General Safety Concerns
Before proceeding to install, commission, or operate the equipment described in the instruction manual we
would like to alert you to the dangerous potential hazards that would be present if safety practices were not
performed in accordance with the local standards and governing body regulations.
- Injury would occur to the operational staff of the equipment through misuse, electric shock, rotating
equipment hazards and lack of cleanliness.
To be able to achieve the aim, of “accidents can be avoided” it must be ensured that the equipment is
installed correctly, regularly maintained and operators of the equipment are made aware of the potential
hazards associated with the particular equipment.
We would like to inform our valuable customers of the safety guide lines when using their equipment.
3.2 Awareness of Safety Hazards
(a) Before attempting to work on the equipment the personnel who are going to install, commission or
operate the equipment must be qualified and fully aware of all the manufacturers and suppliers
recommendations and instructions.
(b) Ensure that the all the recommendations specified in the instruction manuals are maintained as
stated in the contents.
4. Electrical Safety
(a) Ensure that the person who works on the equipment is a qualified electrical engineer/technician who
is competent in the safety aspects and operational mode of the equipment.
2
FM 250 Flow Measurement Test Bench
(a) (b). If the electrical supply to the equipment is supplied by means of a portable trailing cable,
protective devices such as an Earth Leakage Circuit Breaker (ELCB) must be installed.
(b) This protective device must have a very high sensitivity (20-30mA).This device is also referred to as a
residual current device(R C D) within the electrical supply circuitry for personnel protection.
(c) (c) The supply cable must be sized accordingly for all fault and physical conditions pertaining to its
use. The supply network must also incorporate a protection device that will disconnect and isolate
the supply voltage in the case of an overload in a specified period of time without causing any
damage to the equipment. (An overload relay)
5. Installation
(a) On receipt of the equipment extreme care should be used to avoid damage to the equipment on
handling and unpacking. If slings are used ensure they are held on a rigid part of the equipment, the
structure. In the case of a mechanical lift such as a fork lift ensure the lifting forks are beneath the
structure framework so that no damage will occur during the lifting operation.
(b) In some cases it is imperative that the equipment be installed on a level and solid foundation
5.1 Electrical Supply Cables
(a) The normal colour code of the power cables supplied on this equipment is as follows:
- Black -----------------------------------------Line
- Grey or white. --------------------------Neutral
- Green-Yellow --------------------------Ground
(b) The three phase power cable has five wires.
- Red, blue and black. ---------------------Line
- Light grey or white - ------------------Neutral
- Green-Yellow ------ ---------------Ground
5.2 General Precautions for Equipment with Water Including Evaporative Cooling Towers
12.12
(a) Any water contained in the system should be drained regularly. If it is left in the system for a long
period of time without circulation it will stagnate.
(b) The equipment should be flushed regularly with clean water.
(c) Impurities in the water will cause scale or algae and must be cleaned on a regular basis. An anti rust
additive such as used in the automobile industry is recommended to inhibit this process.
(d) The water should be at temperature under 45degreesC to maintain effectiveness.
(e) Many of the problems encountered with water contamination can be reduced and prevented by
means of a water treatment programmer being introduced using the expertise available locally or on
site.
3
FM 250 Flow Measurement Test Bench
(c) Safety valves should be calibrated on a regular basis with mandatory service records maintained.
This should also include pressure reducing valves.
(d) Calibration of any instrumentation such as pressure gauges, thermometers and sensors should be
checked regularly.
(e) Visual inspection of the equipment should be regularly observed for leaks of steam etc and any
frameworks or joints should have the hardware checked for tightness.
(f) Always use protective clothes including gloves when carrying out maintenance on the equipment.
5.5 High Temperature Equipment
(a) When using high temperature equipment there are a number of vital precautions which must be
remembered by the operators and maintenance crew and observed when both operating and
performing maintenance schedules. During operation of this equipment the air, gas or water is at a
high temperature and pressure which can have a very damaging and hazardous effect on students if
safety precautions are not observed.
(b) Ensure that critical values of temperature and pressures listed in the instruction manual are
maintained and not exceeded on the equipment.
(c) Calibration of any instrumentation such as, thermometers and sensors must be checked regularly for
safe operation.
6. Maintenance Safety Practices
(a) Always isolate the equipment from the electrical supply when carrying out maintenance on the
equipment.
(b) Ensure that safety notices are placed on the equipment supply advising personnel that the equipment
is being worked on, inspected and should not be operated.
(c) Check the operation of any protective devices, i.e. E L C B. so that it operates in accordance with its
specifications thus ensuring the safety of all operational personnel working on the equipment. Any
12.12
malfunction of the device must be corrected by a qualified electrician before returning the equipment
back to a service condition.
(d) Ensure on completions of the work that the equipment is returned to its original state and that no
covers, panels are left open along with loose screw drivers, spanners are left in the equipment.
(e) If water is used with the equipment then there are certain preventative mandatory regulations that
have to be taken to prevent infection from harmful micro-organisms.
7. General Safety Conditions when Operating or Maintaining the Equipment
(a) When operating or carrying out maintenance on the equipment the Health and Safety of the
students can be safeguarded in many ways by wearing protective clothing.
(b) Loose fitting clothes should never be worn in a laboratory. These clothes can cause a serious
accident if caught in rotating equipment. i.e. tie etc.
(c) Protective gloves must be used if handling toxic materials or where there is a high temperature
present.
(d) Ear protectors should be worn when operating noisy equipment.
(e) Eye protection should always be used when there is a risk to the eyes.
4
FM 250 Flow Measurement Test Bench
1. General Description
Vent valve
Measuring tank
Venturi meter
This is a bench-top, self-contained water circulating unit for studying various flow
measurement tools including primary flow measuring devices.
The flow measuring instruments are connected in series and all except the measuring
tank are arranged on a water-proof panel.
Measurement of flow by primary flow measuring devices is by coefficient of discharge,
which is determined from a pressure drop across the device and a flow rate. Flow rate is
obtained from a water-meter reading and a stop watch.
The pressure drop is measured via differential pressure using manometers. Pressure
tapping is provided with small ball valves with quick connection. Water tanks are at rear
side of the panel. Flow rate is varied by a flow control valve.
5
FM 250 Flow Measurement Test Bench
H2 L 1
L2 H1
Hand
Measuring tank Venturi Orifice H4 air pump
40 L H3 L3 L4 Flow meter
Storage tank
H 1 H2 H3 H4
L 1 L2 L 3 L4
Drain valve From measuring Pump
device Flow control valve
Low pressure manifold L1-L4 H1-H4
To manometer
To manometer
L1 H1
L2 H2
L3 H3
L4 H4 optional
Low V3
V5 V4
High pressure manifold
High
Differential pressure indicator
6
FM 250 Flow Measurement Test Bench
= VD VD
ReD
ν
= Density of the fluid kg/m3
V = Velocity m/s
D = Pipe diameter m
= Dynamic Viscosity of fluied, Pa-s or N-s/m²
= Kinematics Viscosity m2/s.
Flow is laminar when ReD ≤ 2000,
and is turbulent when ReD ≥ 4000 or more.
12.12
7
FM 250 Flow Measurement Test Bench
p p
g
12.12
p
= Pressure head m
p = Pressure N/m2
= Specific weight N/m3
= Density of fluid kg/m3
g = Acceleration due to gravity m/s2
v2
2.2.3 Velocity Head ( )
2g
When liquid flows in a pipe or in an open channel, it has a kinetic energy. Kinetic energy
per unit weight of liquid is called velocity head. Kinetic energy of the object in motion is:
W
K .E 12 mv 2 12 v 2
g
2
K .E v
W 2g
8
FM 250 Flow Measurement Test Bench
P.E. W .Z
P.E .
Z
W
Where, P.E. = potential energy, Nm
W = weight, N
Z = elevation, m
p v2
H Z
2g
12.12
p1 p 2 v12 v22
h f ,12 Z1 Z 2
2 g 2 g
9
FM 250 Flow Measurement Test Bench
h
D D/2
D2
D1
1 2
12.12
p1 v12 p v2
Z1 2 2 Z 2 (1)
2g 2g
Subscripts 1 & 2 mean point 1 and point 2 of a liquid in the pipe respectively. Point 1 is
before the orifice plate, point 2 is at vena-contract of the jet.
Since Z1=Z2, hence,
p1 v12 p v2
2 2
2g 2g
v22 v12 p p2
1
2g
p p2
v22 v12 2 g 1 (2)
From continuity equation:
A1V1 A2 v2
10
FM 250 Flow Measurement Test Bench
A2 v2
v1 (3)
A1
where, A1 is the cross section area of the pipe, and
A2 is the cross-section area of the orifice.
Substituting V1 from equation (3) in equation (2) gives:
2
Av p p2
v 2 2
2
2 g 1
2
A1
Av
2
p p2
v 1 2 2
2
2 g 1
2
A1
1 p p2
v2 2 g 1
A
2
1 2
A1
1 (4)
v2 2 g h
4
D
1 2
D1
Thus the ideal flow rate (theoretical flow rate) is:
Qth A2 v2
12.12
A2
Qth 2 g h (5)
4
D
1 2
D1
Since there is some energy loss between section 1 and section 2, and there is also a
vena-contract after flow through the orifice, then the actual flow rate will be:
Co Ao
Qact 2 g h (6)
4
D
1 o
D1
where, Qact = actual flow rate, m3/s
Co = coefficient of discharge the for orifice
Ao = cross-section area of the orifice, m²
D1 = pipe diameter (m) = 29mm, for this test equipment
Do = Orifice diameter (m) = 20 mm or for this test equipment
h = difference in pressure head, m
From the above equation we obtain,
4
D
1 o
D1 Qact
Co (7)
Ao 2 g h
11
FM 250 Flow Measurement Test Bench
Ao h (8)
ko
4
D
1 o
D1
Equation (6) may be re-written in another form as:
2 2 g h (9)
Qact Co D1
4 D14
4 1
Do
For the apparatus FM 250, given:
Qo = actual flow rate for orifice meter, m³/s
D1 = 0.029m, D2 = 0.020m
2 g h
Qo Co 0.02912
4 0.029 4
0.020 4 1
h
p1
p2
1 2
throat
in out
Figure 5: Venturi meter
12
FM 250 Flow Measurement Test Bench
A2 v2
v1 (12)
A1
Substitute v1 from equation (12) into equation (11) gives the ideal velocity of fluid at
throat.
p p
v2 ideal 1
2 g 1 Z1 2 Z 2 (13)
2
A
1 2
A1
The ideal flow (theoretical flow) through Venturi meter is then,
A2 p p (14)
Qideal 2 g 1 Z1 2 Z 2
2
A
1 2
A1
Since there is some energy loss between point 1 and point 2, then the actual flow will be
less than the above ideal flow, and after applying the coefficient of discharge CV to the
ideal flow, the actual flow will be:
CV A2 p p
Qact 2 g 1 Z1 2 Z 2
2
A
1 2
A1
p
12.12
CV A2 p (15)
Qact 2 g 1 Z1 2 Z 2
4
D
1 2
D1
13
FM 250 Flow Measurement Test Bench
2 2 g h (17)
Qact CV D1
4 D 4
1
1
4
D o
For FM 250 apparatus, QV = actual flow rate for Venturi meter, m³/s,
D1 = 0.029m, D2 = 0.017m
2 g h
QV CV 0.0292
4 0.029 4
1
0.017
h 2 h 1 h
1 2
Applying Bernoulli’s equation between point 1 in the pipe and the stagnation point 2 in
front of the Pitot tube gives:
p1 v12 p v2
Z1 2 2 Z 2 (19)
2g 2g
Since Z1 = Z2 and v2 = 0, thus
p p
Thus v1 2 g 2 1
v1 2 g h (20)
Actual flow rate is:
Qact C P A1 2 g h (21)
14
FM 250 Flow Measurement Test Bench
h
12.12
h1
h2
1 2
D1 D2
Apply Bernoulli’s equation between point 1 and point 2 as before, then the actual flow rate
through the nozzle is:
C N A2
Qact 2 g Δh
2
A
1 2
A
1
C N A2
Qact 2 g Δh
4
D2
1
D
1
15
FM 250 Flow Measurement Test Bench
Qact KA 2 g Δh (23)
2
where,
Qact = actual Flow rate through nozzle, m3/s
CN = coefficient of discharge for nozzle, dimensionless constant
D1 = Pipe diameter (m) 29mm
D2 = Nozzle diameter (m) 21.46 mm
h = pressure head difference, m
K = flow coefficient, dimensionless, = C N / 1 D D 4 2 1
For this equipment, QN = actual flow rate through nozzle, m³/s
D1 = 0.029m, D2 = 0.0215m
2 gh
QN C N 0.0292 2
4 0.029
1
0.0215
0.9242 at Re = 2 x 107
16
FM 250 Flow Measurement Test Bench
When several devices are available for use in determining the flow rate, the most reliable
choice should be:
1. Measuring Tank
2. Water Meter
3. Flow Meter
17
FM 250 Flow Measurement Test Bench
4. Test Procedures
Vent valve
Measuring tank
Venturi meter
18
FM 250 Flow Measurement Test Bench
- Record volume and time for water meter. Start and stop the watch when one of the
meter pointers reads 0.
- Record volume and time for the measuring tank. Start and stop the watch when
water level reaches the pre-determined levels.
- Repeat the test at other flow rates.
NOTES:
1. Several flow measuring devices may be used simultaneously.
2. Flow rate measured with the measuring tank is considered to be more reliable and
accurate than the water meter and is used as the actual flow rate.
12.12
19
FM 250 Flow Measurement Test Bench
5. Results
Data Sheet
FM 250 Flow Measurement Test Bench
Test by………………………… Date……………………………
Volume (l)
Time (s)
Flow rate (l/min)
WATER METER
Volume (l)
Time (s)
Flow rate (l/min)
ORIFICE METER
PITOT TUBE
VENTURI METER
ISA NOZZLE
VARIABLE-AREA FLOW-METER
20
FM 250 Flow Measurement Test Bench
Sample Data
FM 250 Flow Measurement Test Bench
Test by………………………… Date……………………………
MEASURING TANK
Volume (l) 20 20 18 22 28 32 40 36 38 36 38
Time (sec) 120 91.78 68.06 61.81 63.97 60.,47 64.,8 48.81 44.84 37.96 36.88
Flow rate (l/min) 10 13,07 15,86 21,35 26.26 31.75 37.06 44.25 50.84 56.90 61.82
WATER METER
Volume (l) 21 26 21 23 30 32 45 50 60 60 60
Time (sec) 123.6 105.7 74.64 64.58 68.90 60.14 71.71 67.87 70.19 62.88 57.81
Flow rate (l/min) 10.19 14.75 16.,88 21.36 26.12 31.92 37.65 44.20 51.28 57.25 62.27
ORIFICE METER
Water level of manometer, left side 375 409 323 399 496 738 435 970 884 927
Water level of manometer, right side 346 360 249 278 316 457 43 438 165 25
Difference in water level 29 49 74 121 180 281 392 532 719 902
VENTURI METER
Water level of manometer, left side 311 300 166 156 144 203 345 575 823 732 905
Water level of manometer right side 290 265 113 67 11 5 79 193 323 103 158
Difference in water level 21 35 53 89 133 198 266 382 500 629 747
PITOT TUBE
Water level of manometer, left side 360 381 275 334 395 581 845 683 785 870
Water level of manometer, right side 331 330 201 217 220 317 490 188 120 58
12.12
Difference in water level 29 51 74 117 175 264 355 495 665 812
ISA NOZZLE
Water level of manometer, left side 91 108 217 209 400 715 454 599 605 527 568
Water level of manometer, right side 79 86 177 158 321 609 314 410 329 229 206
Difference in water level 12 22 40 51 79 106 140 184 276 298 362
VARIABLE AREA FLOW METER
Water level of manometer, left side 340 470 546 585 658 830 970
Water level of manometer, right side 152 93 81 108 133 93 48
Difference in water level 188 377 465 477 525 737 922
21
FM 250 Flow Measurement Test Bench
22
FM 250 Flow Measurement Test Bench
6. Sample calculations
Typical Data
Flow rate of volumetric tank
Volume = 28 l
Time = 63.97 s
28 l
Flow rate = 0.4377 l/s
63.97 s
Water level of manometer
Left 658mm, Right 133mm, difference = 525 mm.
Level of manometer from Orifice meter
Left 496mm, Right 316mm, difference = 180 mm.
Level of manometer from Venturi meter
Left 395mm, Right 220 mm, difference = 175 mm.
Level of manometer from Pitot tube
Left 144mm, Right 11mm, difference = 133 mm.
Level of manometer Nozzle
Left 400mm, Right 321mm, difference = 79 mm.
l m3
0.4377 1.583 10 3 CO 0.18
s s
0.4377 l
From which, CO 3
= 0.652
1.583 10 0.424264 10 3 l
23
FM 250 Flow Measurement Test Bench
l m3
0.4377 1.256 10 3 C P 0.133
s s
0.4377 l
From which, CP 3
0.956
1.256 10 0.36469 10 3 l
l m3
12.12
24
FM 250 Flow Measurement Test Bench
Typical Graphs
Coefficients of discharge at various rate of flow for different devices may be plotted in
Graphs as below:
1
Coefficient of Discharge
0.9
0.8
0.7
0.6
0.5
0.4
0 10 20 30 40 50 60 70
Flow Rate ( lpm)
12.12
25
FM 250 Flow Measurement Test Bench
APPENDIX 1
12.12
WATER MANOMETER
26
FM 250 Flow Measurement Test Bench
Description :
Water Manometer employs clear acrylic
tubes with a top common chamber. This
chamber has an air relief valve and can
Pressurising valve be pressurised by a hand air pump or may
for hand air pump be reduced by opening a vent valve.
Air bleeding valve Pressure ports are at the bottom.
Range : 0-450 or 950 mm
Graduation : 1 mm
Top chamber
Application : comparison of water
pressures
This manometer uses 2 clear acrylic tubes.
For multiple readings, the number of tubes
may be 4, 6, 8, 10 or more, which are
available as options.
Instruction for use
1. Close ball valve at the top chamber.
2. Connect pressure lines from the
pressure source to the inlet pressure
ports of the manometer. Water levels
will show on the manometer scale.
12.12
NOTE
More than one pair of tubes may be used simultaneously if the average pressure from one
pair is not much different from the other pairs. In this case, downstream average pressure is
always lower than upstream average pressure. Thus, if any one pair of water levels is out of
the manometer range, that pair cannot be used, simply close the inlet valves of that pair or
close the pressure tapping port at the pressure source.
27
FM 250 Flow Measurement Test Bench
APPENDIX 2
12.12
MERCURY MANOMETER
28
FM 250 Flow Measurement Test Bench
MERCURY MANOMETER.
Description:
29
FM 250 Flow Measurement Test Bench
APPENDIX 3
12.12
30
FM 250 Flow Measurement Test Bench
Top plug
12.12
Drain plug
3. Before running the pump, check that the check valve is not stuck and fill the pump up fully
with water.
4. The motor and the pump are aligned at the factory. If they are separated, care must be taken
when reconnecting. Otherwise, the shaft vibration will cause serious damage to the bearings
and the seals.
Pump Seized
Leaving water in the pump without use for along period of time may cause pump shaft seizing.
Turn on the main power supply switch and turn on the pump.
If there is no flow make sure the flow control valve at the rotameter is open.
31
FM 250 Flow Measurement Test Bench
If pump does not turn, the shaft may be seized. Turn off the pump and open the plug at the top of
the pump and turn the shaft with a screw driver. Once turn, close the plug (see figures below).
If the pump turns but no water flow (make sure the flow control valve before the flow meter is
open), there is a chance that air bubble is trapped in the pump.
In the case of Heat Exchanger unit:
Connect the drain valve next to the pump discharge to outside water supply, open the drain
valve and close the flow control valve before the rotameter. The water supply will purge out air
bubbles.
Close the drain valve and remove the outside water connection.
Turn on the pump again to ensure water is circulating.
In the case of Series and Parallel Pump, multi speed:
Connect the pump discharge hose to the Hydraulics Bench discharge.
Remove the check valve on the pump suction hose.
Open all valves and turn on the Hydraulics Bench pump to purge out air bubbles from the multi
speed pump.
After 1-2 minute, turn off the Hydraulics Bench pump and immediately put back the check
valve on the suction hose.
Remove the discharge hose from the Hydraulics Bench and turn on the multi speed pump to
ensure normal flow.
32
FM 250 Flow Measurement Test Bench
Drain valve
Water pump
12.12
Drain plug at the bottom of the pump or a plug at the top of the pump, and use a screw driver to
turn the impeller (hence the shaft).
Medium Pump
For medium size pump, e.g. 2.5kW, 3.75kW is normally without coupling. The hex head bolts as
the above may not be provided, the pump shaft may be turned by removing a drain plug at the
bottom of the pump or a plug at the top of the pump, and use a screw driver to turn the impeller
(hence the shaft).
33
FM 250 Flow Measurement Test Bench
Gear Pump
The gear pump is designed to be used with oil. If it is used with water, it will get rusted and will
seize.
12.12
Reciprocating Pump
Check oil in the oil crank with the dipstick to make sure that there is enough oil.
34