Test definition ALT Calibr. parameters
Test type Photometric Test in use YES Calibration type NONE Factor -1481 Bias 0
Full name ALT IFCC, 1 Low High Units
reag.system
Online name Test limit 4 1800 U/l Bias corr. in use NO
Result Unit U/l low( Ö ), high

Number of decimals 0 Initial absorbance 1 2.5 A

Acceptance AUTOMATIC

Dilution 1 + 0 Dilution limit * 300 U/l


Sample type

Secondary dil. 1+ 0 5
ýSerum ýPlasma ¨Urine
¨C.S.F ¨Other Correction factor 1
More
Correction bias 0

Test flow Reagent definition


Blank NO Normal cuvette
Antigen excess NO

Syringe BCR Id. Stable


speed on board
Name (days)
Curve type
Incubation Incubation
Reagent Time (sec) Volume (µl) Time (sec) λ 1 (nm) LINEARCUT
ALT1 300 15 120 340 Nonlinearity
Resp.
(mA/min)

Volume (µl) Disp. with λ 2 (nm) 20 ALT1 Normal 011 14


125 EXTRA 380

Disp. with Volume (µl) Res. Net. Abs.


EXTRA 15 0.5
Time (sec)
240
Volume (µl) Dilution with
20 WATER
Points & Inter.
Konelab
Wash reagent Wash reagent 60/30/20XT 9/27
Konelab 20 6/42
NONE NONE

Konelab™ ALT (IFCC, one reagent) - code: 981361 (7 x 50 ml), code: 981769 (8 x 20 ml)

Reagent: Pour reagent B into reagent A, recap and mix gently.


Ö measuring range low

Q.C. sample: Nortrol 981043 & Abtrol 981044. Reconstitute with 5.00 ml of distilled water.