Direction 5397208-8EN
Revision 2
GE Healthcare
Wireless DR Imaging Option System
Pre-Installation
GE HEALTHCARE
DIRECTION 5397208-8EN, REVISION 2 WIRELESS DR IMAGING OPTION SYSTEM
Page 2
Pre-Install.book Page 3 Tuesday, December 13, 2011 10:32 AM
GE HEALTHCARE
DIRECTION 5397208-8EN, REVISION 2 WIRELESS DR IMAGING OPTION SYSTEM PRE-INSTALLATION
LEGAL NOTES
TRADEMARKS
All other products and their name brands are trademarks of their respective holders.
COPYRIGHTS
All Material, Copyright © 2011 by General Electric Company, Inc. All rights reserved. The material
presented and contained herein may not be reproduced in any form or manner, without the written
permission of General Electric Company, Inc.
GE HEALTHCARE
DIRECTION 5397208-8EN, REVISION 2 WIRELESS DR IMAGING OPTION SYSTEM PRE-INSTALLATION
This page is intentionally left blank.
GE HEALTHCARE
DIRECTION 5397208-8EN, REVISION 2 WIRELESS DR IMAGING OPTION SYSTEM PRE-INSTALLATION
Important Precautions
LANGUAGE
ПРЕДУПРЕЖДЕНИЕ • ТО В А УП Ъ ТВ А Н Е ЗА РА Б О ТА Е Н А Л И Ч Н О С А М О Н А А Н ГЛ И Й С КИ Е ЗИ К.
(BG) • А КО Д О С ТА В Ч И КЪ Т Н А УС Л УГА ТА Н А КЛИ ЕН ТА И З И С КА ЕЗИ К,
РА З Л И Ч ЕН О Т А Н ГЛ И Й С КИ , ЗА Д Ъ Л Ж ЕН И Е Н А КЛ И ЕН ТА Е Д А О С И ГУРИ
П РЕ В О Д .
• Н Е И ЗП О Л ЗВ А Й ТЕ О Б О РУД В А Н ЕТО П РЕД И Д А С ТЕ С Е КО Н С УЛ ТИ РА Л И
И РА ЗБ РА ЛИ УП Ъ ТВ А Н ЕТО ЗА РА Б О ТА .
• Н ЕС П А ЗВ А Н ЕТО Н А ТО В А П РЕД УП РЕЖ Д ЕН И Е М О Ж Е Д А Д О В ЕД Е Д О
Н А РА Н Я В А Н Е Н А Д О С ТА В Ч И КА Н А УС ЛУГА ТА , О П ЕРА ТО РА И Л И
П А Ц И ЕН Т В РЕЗУЛ ТА Т Н А ТО КО В УД А Р И Л И М ЕХА Н И ЧН А И Л И Д РУ ГА
О П А С Н О С Т.
警告 • 本维修手册仅提供英文版本。
(ZH-CN) • 如果维修服务提供商需要非英文版本,客户需自行提供翻译服务。
• 未详细阅读和完全理解本维修手册之前,不得进行维修。
• 忽略本警告可能对维修人员,操作员或患者造成触电、机械伤害或其他形式的伤害。
GE HEALTHCARE
DIRECTION 5397208-8EN, REVISION 2 WIRELESS DR IMAGING OPTION SYSTEM PRE-INSTALLATION
GE HEALTHCARE
DIRECTION 5397208-8EN, REVISION 2 WIRELESS DR IMAGING OPTION SYSTEM PRE-INSTALLATION
GE HEALTHCARE
DIRECTION 5397208-8EN, REVISION 2 WIRELESS DR IMAGING OPTION SYSTEM PRE-INSTALLATION
• このサービスマニュアルには英語版しかありません。
(JA) • サービスを担当される業者が英語以外の言語を要求される場合、翻訳作業はその業
者の責任で行うものとさせていただきます。
• このサービスマニュアルを熟読し理解せずに、装置のサービスを行わないでくださ
い。
• この警告に従わない場合、サービスを担当される方、操作員あるいは患者さんが、
感電や機械的又はその他の危険により負傷する可能性があります。
GE HEALTHCARE
DIRECTION 5397208-8EN, REVISION 2 WIRELESS DR IMAGING OPTION SYSTEM PRE-INSTALLATION
(KO) • 고객의 서비스 제공자가 영어 이외의 언어를 요구할 경우, 번역 서비스를 제공하는 것
은 고객의 책임입니다.
GE HEALTHCARE
DIRECTION 5397208-8EN, REVISION 2 WIRELESS DR IMAGING OPTION SYSTEM PRE-INSTALLATION
GE HEALTHCARE
DIRECTION 5397208-8EN, REVISION 2 WIRELESS DR IMAGING OPTION SYSTEM PRE-INSTALLATION
GE HEALTHCARE
DIRECTION 5397208-8EN, REVISION 2 WIRELESS DR IMAGING OPTION SYSTEM PRE-INSTALLATION
DAMAGE IN TRANSPORTATION
All packages should be closely examined at time of delivery. If damage is apparent, write “Damage
In Shipment” on ALL copies of the freight or express bill BEFORE delivery is accepted or “signed
for” by a GE representative or hospital receiving agent. Whether noted or concealed, damage
MUST be reported to the carrier immediately upon discovery, or in any event, within 14 days after
receipt, and the contents and containers held for inspection by the carrier. A transportation
company will not pay a claim for damage if an inspection is not requested within this 14 day period.
Call GEHC Global Parts 1-800-548-3366 and select option 8, immediately after damage is found.
At this time be ready to supply name of carrier, delivery date, consignee name, freight or express
bill number, item damaged and extent of damage.
Complete instructions regarding claim procedure are found in Section S of the Policy And
Procedures Bulletins.
14 July 1993
GE HEALTHCARE
DIRECTION 5397208-8EN, REVISION 2 WIRELESS DR IMAGING OPTION SYSTEM PRE-INSTALLATION
The purchaser of GE equipment shall only utilize qualified personnel (i.e., GE’s field engineers,
personnel of third-party service companies with equivalent training, or licensed electricians) to
perform electrical servicing on the equipment.
IMPORTANT...X-RAY PROTECTION
X-ray equipment, if not properly used, may cause injury. Accordingly, the instructions herein
contained should be thoroughly read and understood by everyone who will use the equipment
before you attempt to place this equipment in operation. The General Electric Company, Healthcare
Group, will be glad to assist and cooperate in placing this equipment in use.
Although this apparatus incorporates a high degree of protection against x-radiation other than the
useful beam, no practical design of equipment can provide complete protection. Nor can any
practical design compel the operator to take adequate precautions to prevent the possibility of any
persons carelessly exposing themselves or others to radiation.
It is important that anyone having anything to do with x-radiation be properly trained and fully
acquainted with the recommendations of the National Council on Radiation Protection and
Measurements as published in NCRP Reports available from NCRP Publications, 7910 Woodmont
Avenue, Room 1016, Bethesda, Maryland 20814, and of the International Commission on
Radiation Protection, and of any other local authorities, and take adequate steps to protect against
injury.
The equipment is sold with the understanding that the General Electric Company, Healthcare
Group, its agents, and representatives have no responsibility for injury or damage which may result
from improper use of the equipment.
Various protective materials and devices are available. It is urged that such materials or devices be used.
GE HEALTHCARE
DIRECTION 5397208-8EN, REVISION 2 WIRELESS DR IMAGING OPTION SYSTEM PRE-INSTALLATION
This page is intentionally left blank.
GE HEALTHCARE
DIRECTION 5397208-8EN, REVISION 2 WIRELESS DR IMAGING OPTION SYSTEM PRE-INSTALLATION
Preface
Publication Conventions
Standardized conventions for representing information is a uniform way of communicating
information to a reader in a consistent manner. Conventions are used so that the reader can easily
recognize the actions or decisions that must be made. There are a number of character and
paragraph styles used in this publication to accomplish this task. Please become familiar with them
before proceeding forward.
It is important that you read and understand hazard statements, and not just ignore them.
Preface Page 15
Pre-Install.book Page 16 Tuesday, December 13, 2011 10:32 AM
GE HEALTHCARE
DIRECTION 5397208-8EN, REVISION 2 WIRELESS DR IMAGING OPTION SYSTEM PRE-INSTALLATION
NOTICE - Indicates information or a company policy that relates directly or indirectly to the
safety of personnel or protection of property. This signal word is associated directly with a
hazard or hazardous situation and is used in place of 'DANGER,' 'WARNING,' or 'CAUTION.'
It can include:
• Destruction of a disk drive
• Potential for internal mechanical damage, such as to a X-ray tube
Page 16 Preface
Pre-Install.book Page 17 Tuesday, December 13, 2011 10:32 AM
GE HEALTHCARE
DIRECTION 5397208-8EN, REVISION 2 WIRELESS DR IMAGING OPTION SYSTEM PRE-INSTALLATION
Preface Page 17
Pre-Install.book Page 18 Tuesday, December 13, 2011 10:32 AM
GE HEALTHCARE
DIRECTION 5397208-8EN, REVISION 2 WIRELESS DR IMAGING OPTION SYSTEM PRE-INSTALLATION
2.2 Page Layout
The current section and its title The current chapter and its title
are always shown in the footer of are always shown in the footer of
the left (even) page. the right (odd) page.
An exclamation point in a triangle is used Paragraphs preceeded by a symbol
to indicate important information to the user. (e.g. bullets) contain information that
Paragraphs preceeded by Alphanumeric has no specific order.
characters (e.g. numbers) contain infor-
mation that must be followed in a specific order.
Headers and footers in this publication are designed to allow you to quickly identify your location.
The document part number and revision number appear in every header on every page. Odd
numbered page footers indicate the current chapter, its title and current page number. Even
numbered page footers show the current section and its title, as well as the current page number.
Page 18 Preface
Pre-Install.book Page 19 Tuesday, December 13, 2011 10:32 AM
GE HEALTHCARE
DIRECTION 5397208-8EN, REVISION 2 WIRELESS DR IMAGING OPTION SYSTEM PRE-INSTALLATION
Example: This paragraph’s font represents computer generated screen “fixed” output.
Fixed Output Its output is fixed from the sense that it does not vary from application
to application. It is the most commonly used style used to indicate
filenames, paths and text that do not change from system to system. The
character style used is a fixed width such as courier.
Example: This paragraph’s font represents computer screen output that is
Variable Output “variable”. It is used to represent output that varies from application
to application or system to system. Variable output is sometimes found
placed between greater-than and less-than operators for clarification. For
example: <variable_ouput> or <3.45.120.3>. In both cases, the < and >
operators are not part of the actual input.
Example: This paragraph’s font represents fixed input. It is computer input that
Fixed Input is typed-in via the keyboard. Typed input that does not vary from
application to application or system to system. Fixed text the user is
required to supply as input. For example: cd /usr/3p
Example: This paragraph’s font represents computer input that can vary from
Variable Input application to application or system to system. With variable text, the
user is required to supply system dependent input or information. Variable
input sometimes is placed between greater-than and less-than operators.
For example: <variable_input>. In these cases, the (<>) operators would
be dropped prior to input. For example: ypcat hosts | grep <3.45.120.3>
would be typed into the computer as
ypcat hosts | grep 3.45.120.3
without the greater-than and less-than operators.
2.4 Buttons, Switches and Keyboard Inputs (Hard & Soft Keys)
Different character styles are used to indicate actions requiring the reader to press either a hard or
soft button, switch or key. Physical hardware, such as buttons and switches, are called hard keys
because they are hard wired or mechanical in nature. A keyboard or on/off switch would be a hard
key. Software or computer generated buttons are called soft keys because they are software
generated. Software driven menu buttons are an example of such keys. Soft and hard keys are
represented differently in this publication.
Example: A power switch ON/OFF or a keyboard key like ENTER is indicated by applying a character style
Hard Keys that uses both over and under-lined bold text. This is a hard key.
Example: Whereas the computer MENU button that you would click with your mouse or touch with your hand
Soft Keys uses over and under-lined regular text. This is a soft key.
Preface Page 19
Pre-Install.book Page 20 Tuesday, December 13, 2011 10:32 AM
GE HEALTHCARE
DIRECTION 5397208-8EN, REVISION 2 WIRELESS DR IMAGING OPTION SYSTEM PRE-INSTALLATION
This page is intentionally left blank.
Page 20 Preface
Pre-Install.book Page 21 Tuesday, December 13, 2011 10:32 AM
GE HEALTHCARE
DIRECTION 5397208-8EN, REVISION 2 WIRELESS DR IMAGING OPTION SYSTEM PRE-INSTALLATION
Table of Contents
Preface
Publication Conventions ...................................................................................... 15
Section 1.0 Safety & Hazard Information ....................................................... 15
1.1 Hazard Messages............................................................................................................ 15
1.2 Text Format of Signal Words ........................................................................................... 15
1.3 Symbols and Pictorials Used ........................................................................................... 16
Section 2.0 Publication Conventions ............................................................. 17
2.1 General Paragraph and Character Styles........................................................................ 17
2.2 Page Layout..................................................................................................................... 18
2.3 Computer Screen Output/Input Text Character Styles .................................................... 18
2.4 Buttons, Switches and Keyboard Inputs (Hard & Soft Keys) ........................................... 19
Chapter 1 - Introduction........................................................................................ 25
Section 1.0 Objective and Scope of this Manual........................................... 25
Section 2.0 Avoiding Unnecessary Expenses and Delays........................... 25
Section 3.0 An Overview of the Pre-Installation Process............................. 25
Section 4.0 Responsibility of Purchaser/Customer ...................................... 26
Section 5.0 Contract Changes ........................................................................ 26
Section 6.0 Responsibilities of the Purchaser .............................................. 27
Section 7.0 What You Will Receive (System Components).......................... 28
GE HEALTHCARE
DIRECTION 5397208-8EN, REVISION 2 WIRELESS DR IMAGING OPTION SYSTEM PRE-INSTALLATION
1.3 Electrical Ducts ............................................................................................................... 35
1.4 Power Distribution ........................................................................................................... 36
Section 2.0 Hospital Network .......................................................................... 36
2.1 Broadband Network Connection ..................................................................................... 36
2.2 Remote Services Broadband Pre-Installation Requirements for Europe ........................ 36
Section 3.0 Master Interconnect System (MIS) .............................................. 37
GE HEALTHCARE
DIRECTION 5397208-8EN, REVISION 2 WIRELESS DR IMAGING OPTION SYSTEM PRE-INSTALLATION
Section 3.0 Preparing the Delivery Route ...................................................... 57
Section 4.0 Networkflow Audit........................................................................ 58
4.1 What is the Networkflow Audit? ....................................................................................... 58
4.2 Facility Information........................................................................................................... 59
4.3 Workflow Analysis............................................................................................................ 59
4.4 The Physical Network ...................................................................................................... 60
4.5 Wireless DR Imaging Option System Parameters........................................................... 61
4.6 Devices & Services Audit................................................................................................. 63
4.7 Dataflow Analysis ............................................................................................................ 65
4.8 What Will Happen Next?.................................................................................................. 67
Section 5.0 Pre-Installation Checklist ............................................................ 68
GE HEALTHCARE
DIRECTION 5397208-8EN, REVISION 2 WIRELESS DR IMAGING OPTION SYSTEM PRE-INSTALLATION
GE HEALTHCARE
DIRECTION 5397208-8EN, REVISION 2 WIRELESS DR IMAGING OPTION SYSTEM PRE-INSTALLATION
Chapter 1 - Introduction
GE HEALTHCARE
DIRECTION 5397208-8EN, REVISION 2 WIRELESS DR IMAGING OPTION SYSTEM PRE-INSTALLATION
Chapter 6 - -
Product System
Characteristics Componets Site Planning
Chapter 7 - - and layout
Room Layout Information
Order Cable Lengths
Chapter 4 - -
Planning Electrical Electrical
Connections Information
Chapter 5 - -
System Facility
Power & Grounds Pre-installation Site Plan
Documentation Drawings
Chapter 2 - - Installation
Target System Tools
Confirmation Checklist
Delivery
Route
Delivery
Survey/Map
and
Installation
GE HEALTHCARE
DIRECTION 5397208-8EN, REVISION 2 WIRELESS DR IMAGING OPTION SYSTEM PRE-INSTALLATION
Section 6.0 Responsibilities of the Purchaser
The purchaser is responsible for completion of “Pre-Installation”. This includes the procurement and
installation of all required materials and services to get the room ready for installation of the product.
This responsibility includes providing:
• A clean and safe work environment for installation of the product (finished floor, ceiling, walls,
and proper room lighting).
• A location suitable for the installation of the product. See Chapter 3 - - Room Requirements.
- Suitable support structures in the floor, walls, or ceiling necessary for the mounting of the
product and/or its components.
- Installation of conduit, ducts and/or raceways necessary to route cables safely. See
Chapter 5 - - System Facility Power & Grounds andChapter 6 - - Product Characteristics
- Electrical power and grounds of specified quality and reliability. See Chapter 5 - - System
Facility Power & Grounds.
• A location suitable for operation of the product. See Chapter 7 - - Room Layout.
• Installation of non-electric services.
GE HEALTHCARE
DIRECTION 5397208-8EN, REVISION 2 WIRELESS DR IMAGING OPTION SYSTEM PRE-INSTALLATION
The Wireless DR Imaging Option System may consist of the following main components (See
Figure 1-2 and Table 1-2):
2 4
6
5 7
9
8
10
GE HEALTHCARE
DIRECTION 5397208-8EN, REVISION 2 WIRELESS DR IMAGING OPTION SYSTEM PRE-INSTALLATION
Item Component Part Number Comment
11 Probe Holder 2280928-2 Standard
12 Flat Field Phantom 2281087-2 Standard
13 OM English Paper Manual 5397214-1EN Standard
14 OM CD 5397214-299 Standard
15 SM Class A CD 5397205-2EN Standard
16 Warning Label for Wallstand 5401827 Standard
17 OS DVD 5399373 Standard
18 Applicaton Software CD 5399375 Standard
19 P500D Configuration Tool CD 5413253 Standard: Only for P500D
20 Proteus XR/a AEC Compensation Patch CD 5408826 Standard: Only for Proteus
21 English Keyboard 5183547-4 Standard
22 System MIS Cable Collector Refer to MIS Map
23 Medium Tether (7m) 5389155-4
24 Short Tether (4m) 5389155 Optional, at least 1 selected
25 Long Tether (10m) 5389155-2
26 100CM Table Grid 5394289 Optional
27 130CM Grid for Table/WS 5396678 Optional
28 100CM Wallstand Grid 5396677 Optional
29 180CM Wallstand Grid 5396679 Optional
30 8:1 Portable Grid 5363606-2 Optional
31 6:1 Portable Grid 5363606 Optional
32 Tabletop Lateral Detector Holder 5396920 Optional
33 Monitor Wall Mount 2371592 Optional
34 Proteus XR/a Exposure Key Disable Board 5397162 Standard: Only for Proteus
35 Precision 500D VENUS Kit 5328038 Optional: For Precision 500D Non-
Aurora
36 Bar code reader 5271574-2 Optional
Table 1-1 Wireless DR Imaging Option System Component Identification
GE HEALTHCARE
DIRECTION 5397208-8EN, REVISION 2 WIRELESS DR IMAGING OPTION SYSTEM PRE-INSTALLATION
This page is intentionally left blank.
GE HEALTHCARE
DIRECTION 5397208-8EN, REVISION 2 WIRELESS DR IMAGING OPTION SYSTEM PRE-INSTALLATION
YES YES
Make sure you have ordered the Make sure you have ordered the
special parts for Proteus Upgrade special parts for P500D Upgrade
YES
YES
Order and install the IUI Console Replacement Kit (Only after the Kit is
installed, P500D can be upgraded by WDR1). If the system has an Odetta
style OTS with no Can Amp, it may also need the OTS lateral detent upgrade
kit which is a prerequisite for the IUI Console Replacement Kit .
2.) Purchaser needs make sure target system works well before installation of Wireless DR
Imaging System. Refer to target system Operator Manual and Service Manual to do functional
check.
3.) For Proteus XR/a, check that system console cable is 2259298-60. If not,replace it before
proceeding.
GE HEALTHCARE
DIRECTION 5397208-8EN, REVISION 2 WIRELESS DR IMAGING OPTION SYSTEM PRE-INSTALLATION
This page is intentionally left blank.
GE HEALTHCARE
DIRECTION 5397208-8EN, REVISION 2 WIRELESS DR IMAGING OPTION SYSTEM PRE-INSTALLATION
Note: STORAGE values only refer to equipment that is still in shipping containers. If the equipment is
partially or completely installed, refer to IN-USE values.
GE HEALTHCARE
DIRECTION 5397208-8EN, REVISION 2 WIRELESS DR IMAGING OPTION SYSTEM PRE-INSTALLATION
1.3 Heat Output
GE HEALTHCARE
DIRECTION 5397208-8EN, REVISION 2 WIRELESS DR IMAGING OPTION SYSTEM PRE-INSTALLATION
High voltage and power cables must be separated from other cables. Use a separate trough in the
duct system, or use a separate conduit. USB cables concealment need follow local regulatory
requirements. Minimize cable length between the line disconnect and the System Cabinet power
unit to reduce voltage regulation problems and wiring costs.
For information about the cables supplied with your system, please refer to Chapter 9 - - System
Cable Information.
1.2 Conduit
Separate conduits must be used for power and signal wires. These wires must be kept separated
from each other.
Using conduit imposes some important considerations when used with this system. Of primary
concern, the majority of cables used are pre-terminated. Pre-termination greatly simplifies
interconnection but makes cable-pulling difficult because of the added dimensions of the
connectors.
Conduit must be large enough to pass the cable and connector through with all other cables already
in the conduit. Also, the size of conduit chosen must allow for future growth. There is the possibility
of additional cables being added later as the system is developed and options are added.
The use of conduit is recommended for cables running overhead between rooms, especially when
a diagonal run provides the shortest cable path.
GE HEALTHCARE
DIRECTION 5397208-8EN, REVISION 2 WIRELESS DR IMAGING OPTION SYSTEM PRE-INSTALLATION
1.4 Power Distribution
Note: For hospital facility power and ground requirements to the Wireless DR Imaging Option System
Additional power unit, refer to: Chapter 5 - - System Facility Power & Grounds.
Reference For Wireless DR Imaging Option System power distribution from the System Cabinet power unit,
Material Exists refer to Wireless DR Imaging Option System MIS Map, see Chapter 9 - - System Cable Information.
GE HEALTHCARE
DIRECTION 5397208-8EN, REVISION 2 WIRELESS DR IMAGING OPTION SYSTEM PRE-INSTALLATION
Note: For P500D upgrade, two Ethernet ports are needed: one for P500D, another for WDR1.
Note: For specific Wireless DR Imaging Option System interconnect maps and connection details, please
Additional refer to Chapter 9 - - System Cable Information.
Reference
Material Exists
GE HEALTHCARE
DIRECTION 5397208-8EN, REVISION 2 WIRELESS DR IMAGING OPTION SYSTEM PRE-INSTALLATION
This page is intentionally left blank.
GE HEALTHCARE
DIRECTION 5397208-8EN, REVISION 2 WIRELESS DR IMAGING OPTION SYSTEM PRE-INSTALLATION
All Wireless DR Imaging Option System component power connections must be made in
accordance with the Wireless DR Imaging Option System MIS Map, see Chapter 9 - - System Cable
Information.
Calculate feeder losses before you begin. Total feeder losses must be calculated to ensure that the
losses are less than those specified in the installation drawings. Calculating the recommended
minimum transformer sizing for feeding a system ensures the transformer losses are less than half
of the maximum regulation for the system.
Regulation is the calculated voltage losses for the entire power distribution system (No-Load
Voltage minus Full-Load Voltage) divided by the no-load voltage minus the system losses (Full-
Load Voltage):
NoLoadVoltage – FullLoadVoltage
Regulation = ---------------------------------------------------------------------------------------------------- × 100
FullLoadVoltage
In the X-ray room, there must be a lockable facility power disconnect. It must be installed electrically
before the equipment, for the purpose of locking out the power. This must be done before service
to the high voltage system is performed.
GE HEALTHCARE
DIRECTION 5397208-8EN, REVISION 2 WIRELESS DR IMAGING OPTION SYSTEM PRE-INSTALLATION
1.2 Electrical Requirements
PARAMETER DISCRIPTION
Input Power Single phase:
100-127vac ±10%
or
200-240vac ±10%
Frequency 50/60Hz ±10%
Power >650W
Table 5-1 System Cabinet Power Specifications
PARAMETER DISCRIPTION
Input Power Single phase:
100-127vac ±10%
or
200-240vac ±10%
Frequency 50/60Hz ±10%
Power >50W
PARAMETER DISCRIPTION
Input Power Single phase:
100-240vac ±10%
Frequency 50/60Hz ±10%
Power >1.5A
GE HEALTHCARE
DIRECTION 5397208-8EN, REVISION 2 WIRELESS DR IMAGING OPTION SYSTEM PRE-INSTALLATION
1.2.2 Wiring Electrical Power and Disconnects
This section provides additional data regarding power circuits the customer must provide, and
internal electrical circuits necessary to supply the correct power to the Wireless DR Imaging Option
System. Figure 5-1 shows the room power supply installed.
. It is supposed to connect the cabinet to a standard electrical outlet. But, if local regulatory requires
that it is mandatory to have this outlet connected to the power distribution box (in case of emergency
to switch off the complete room), please follow local regulatory requirement.
Grounding reliability can only be achieved when the equipment is connected to an equivalent
receptacle marked "Hospital Only" or "Hospital Grade"This tag is for USA only.
GE HEALTHCARE
DIRECTION 5397208-8EN, REVISION 2 WIRELESS DR IMAGING OPTION SYSTEM PRE-INSTALLATION
GE HEALTHCARE
DIRECTION 5397208-8EN, REVISION 2 WIRELESS DR IMAGING OPTION SYSTEM PRE-INSTALLATION
PRODUCT OR DIMENSIONS
References
COMPONENT Width (mm/in) Depth (mm/in) Height (mm/in)
Operator Console:
Workstation 318/12.52 351/13.82 85/3.35 Figure 6-1
LCD Monitor 420/16.54 254/10.00 458/18.03
Cabinet 240/9.45 265/10.40 330/12.99 See Figure 6-2
Tether Interface Box 255/10.04 110/4.33 310/12.20 See Figure 6-4 through
Figure 6-5
Detector Bin 535/21.06 119/4.68 360/14.17 See Figure 6-6 through
Figure 6-7
Dongle Assembly 62.5/2.46 133/5.24 257/10.11 See Figure 6-8 through
Figure 6-9
GE HEALTHCARE
DIRECTION 5397208-8EN, REVISION 2 WIRELESS DR IMAGING OPTION SYSTEM PRE-INSTALLATION
2.2 Dimensioned Figures and Drawings
85 [3.35]
]
31 .8 2
8[ 13
12
.52 1[
] 35
2.2.2 Cabinet
Unit:mm (inch)
GE HEALTHCARE
DIRECTION 5397208-8EN, REVISION 2 WIRELESS DR IMAGING OPTION SYSTEM PRE-INSTALLATION
2.2.3 Cabinet Bracket
Note: Only for the sites which have seismic regulatory requirement.No anchors were supplied with
bracket. Use floor anchors (obtain locally) to bolt affix bracket to floor.
Unit:mm (inch)
GE HEALTHCARE
DIRECTION 5397208-8EN, REVISION 2 WIRELESS DR IMAGING OPTION SYSTEM PRE-INSTALLATION
2.2.4 Tether Interface Box
Unit:mm (inch)
GE HEALTHCARE
DIRECTION 5397208-8EN, REVISION 2 WIRELESS DR IMAGING OPTION SYSTEM PRE-INSTALLATION
2.2.5 Detector BIN
Unit:mm (inch)
GE HEALTHCARE
DIRECTION 5397208-8EN, REVISION 2 WIRELESS DR IMAGING OPTION SYSTEM PRE-INSTALLATION
2.2.6 Dongle Assembly
Unit mm(inch)
GE HEALTHCARE
DIRECTION 5397208-8EN, REVISION 2 WIRELESS DR IMAGING OPTION SYSTEM PRE-INSTALLATION
2.2.7 Touchscreen Monitor and Mount
The relationship between the monitor and the wall mounting plate is given as following drawing.
This is a view looking into the wall.
337 mm
13.25 in
Monitor
13 mm
0.5 in
Wall
Mounting 187.3 m m
Plate 7.375 in
133 mm
7.1 m m 5.25 in
0.281 in
Dia hole
13 mm 13 mm
0.5 in 50.8 mm 0.5 in
2.50 in
The wall that the monitor is mounted on must be able to carry a vertical shear load of 10.9 kg (24
lbs) and a moment of 59 Nm (522 in-lbs). The factor of safety applied over this shall be a minimum
of 4. If a higher factor of safety or seismic loading is required per the local building codes, it shall be
used to determine the mounting.
GE HEALTHCARE
DIRECTION 5397208-8EN, REVISION 2 WIRELESS DR IMAGING OPTION SYSTEM PRE-INSTALLATION
699 mm
27.5 in
317mm
12.5 in
Wall 10.9 Kg
Loading 24 lb
59 Nm
522 in-lb 4.2 Kg 6.7Kg
9.25 lbs 14.8 lbs
Arm Weight Monitor Weight
Wall
730 mm
28.75 in
51 mm 227mm 227mm
2 in 9 in 9 in
409 mm
16.1 in
Hinges
Wall
Note: Fasteners As part of the pre-installation process, the appropriate wall anchor must be installed prior to the
are not shipped arrival of the system. The associated fasteners (i.e., bolts) should also be made available to those
with system. performing the monitor installation. Fasteners are not provided with the system because physical
characteristics of walls will vary from site to site.
GE HEALTHCARE
DIRECTION 5397208-8EN, REVISION 2 WIRELESS DR IMAGING OPTION SYSTEM PRE-INSTALLATION
2.3 Floor Loading and Recommended Mounting Methods
PRODUCT OR WEIGHT LOAD WEIGHT/OCCUPIED RECOMMENDED MOUNTING
COMPONENT (kg/lb) BEARING AREA INFORMATION
AREA
kg/m2
m2
Operator Console: NA
PC Tower 4.5/9.92 NA Table mount (not anchored)
GE HEALTHCARE
DIRECTION 5397208-8EN, REVISION 2 WIRELESS DR IMAGING OPTION SYSTEM PRE-INSTALLATION
GE HEALTHCARE
DIRECTION 5397208-8EN, REVISION 2 WIRELESS DR IMAGING OPTION SYSTEM PRE-INSTALLATION
Default Location
Adjusting Range
Table
Detector insert direction
WS WS
Table Table
WS WS WS
GE HEALTHCARE
DIRECTION 5397208-8EN, REVISION 2 WIRELESS DR IMAGING OPTION SYSTEM PRE-INSTALLATION
1.) Dongle location should face to the detector insert direction in table housing.
2.) Antenna angle is 45 degree pointing to the wall
3.) Antenna height is close to ceiling, but without metal obstacles within 20 cm range.
4.) No obstacles are between antenna and application range of portable detector.
5.) For “Control room outside” default location is center of wall, and adjusting range is 1.5m to left
and 1.5m to right. There is no restriction for WS insert direction (Left or right) and layout (Foot
and Head).
6.) For “Control room inside” default location is corner of shielding wall of control room, and
Adjusting range is 1.5m to left and 1.5m to right (not behind wallstand patient barrier). There
is no restriction for WS insert direction (Left or right) and layout (Foot and Head).
7.) For “Control room inside” If WS is at Head, another default location is center of wall (not
including control room), and Adjusting range is 1.5m to left and 1.5m to right. There is no
restriction for WS insert direction (Left or right).
GE HEALTHCARE
DIRECTION 5397208-8EN, REVISION 2 WIRELESS DR IMAGING OPTION SYSTEM PRE-INSTALLATION
GE HEALTHCARE
DIRECTION 5397208-8EN, REVISION 2 WIRELESS DR IMAGING OPTION SYSTEM PRE-INSTALLATION
The following tools and materials are needed for installation, but are not shipped with the product: Completed
Assorted hardware for termination of electrical connections (solder-less ring lug terminals and butt
splices, AWG 2-18)
Tie wraps, electrical tape and wire markers
Tags for labelling incomplete work in accordance to OSHA and regulatory requirements
Assorted 12-point sockets (SAE and metric), drives, wrenches and torque wrench (Nm and ft.-lbs)
Electric and hammer drill. Assorted masonry and high-speed bits in both metric and SAE sizes
Assorted sizes of tongue and grove pliers, hammers, hex wrenches (metric and SAE), screw drivers
and metal files
Assorted sizes of wire cutters and strippers, ratchet and standard crimpers (42,400 mm2 and
upwards), and a 75 watt soldering iron
Heat and electrical tape
Chalk line
Digital multimeter
Dosemeter
Densitometer
GE HEALTHCARE
DIRECTION 5397208-8EN, REVISION 2 WIRELESS DR IMAGING OPTION SYSTEM PRE-INSTALLATION
Section 3.0 Preparing the Delivery Route
1.) Step One – Sketch out the Route
Begin preparing Route Survey by sketching the area of the hospital or clinic which will receive
the equipment. Include all areas on the delivery route from outside of building to destination.
See sample sketch below.
Reference Numbers Numbers in circles refer to the Route Survey data. The Route
Survey is a form on which site data is listed (step 2).
Floor 1 Floor 2
3
1 2
4 7
5
Ramp 6
8
Elevator
9 10
Destination
GE HEALTHCARE
DIRECTION 5397208-8EN, REVISION 2 WIRELESS DR IMAGING OPTION SYSTEM PRE-INSTALLATION
Understanding how your facility leverages its network investment through our Networkflow process
will help us better integrate the Wireless DR Imaging Option System into your operations. The
following is intended to identify the various ways the Wireless DR Imaging Option System can fit
into your workflow and the ramifications of selecting one path or another. We would like to start at
the beginning, with the patient arriving at your facility, going through registration/admittance/patient
scheduling and proceed all the way to the read images being archived.
This audit is intended to be performed before the system is quoted to you. The audit process will
uncover aspects of your network and workflow that will impact how well the Wireless DR Imaging
Option System will integrate into your facility. With all facts uncovered, GE can prepare a more
accurate quote and minimize "surprises" at the time of install.
You should fill this out with the GE Healthcare representative. They will be able to answer any
questions you may have.
GE HEALTHCARE
DIRECTION 5397208-8EN, REVISION 2 WIRELESS DR IMAGING OPTION SYSTEM PRE-INSTALLATION
4.2 Facility Information
GEHC Auditor:
When the patient arrives in the Wireless DR Imaging Option System room for the exam, how is the patient data entered
into the system?
If the patient information was downloaded from a HIS/RIS system, how would the query be structured? (Pick all that apply)
Once at the start of the shift Several times during a shift Before each patient
Once the digital diagnostic images are acquired, what is your facility's default workflow?
(Pick one)
GE HEALTHCARE
DIRECTION 5397208-8EN, REVISION 2 WIRELESS DR IMAGING OPTION SYSTEM PRE-INSTALLATION
When images are configured for automatic push, what would you like to be sent to PACS/archive/review stations?
When images are printed, on what device is the print command originated? (Pick all that apply)
How soon after the images are acquired is the first image quality check done?
Before the next image is shot Before the patient leaves After patient leaves
Has 100baseT installed Has 10baseT installed Has a different network installed
Will have 100baseT installed Will have 10baseT installed We don't have a network installed
Yes No
Yes No
GE HEALTHCARE
DIRECTION 5397208-8EN, REVISION 2 WIRELESS DR IMAGING OPTION SYSTEM PRE-INSTALLATION
4.5 Wireless DR Imaging Option System Parameters
Figure 8-2 WDR1 Internal Network Topology
Hospital
P500D PC
(Venus PC or Magic PC)
Gryphon PC Detector
Hospital
Note: If the hospital network uses 192.168.x.x, there will be a conflict. If this conflict occurs, you must
contact your GE Service Representative to change the internal IP Addresses used by the
system.No hazardous situations resulting from net work data coupling.
• Connection of the System to a NETWORK/DATA COUPLING that includes other equipment
could result in previously unidentified RISKS to PATIENTS, OPERATORS or third parties.
• The RESPONSIBLE ORGANIZATION should identify, analyze, evaluate and control these
RISKS.
• Subsequent changes to the NETWORK/DATA COUPLING could introduce new RISKS and
require additional analysis.
• Changes to the NETWORK/DATA COUPLING include.
• Changes in NETWORK/DATA COUPLING configuration.
• Connection of additional items to the NETWORK/DATA COUPLING.
• Disconnecting items from the NETWORK/DATA COUPLING.
• Update of equipment connected to the NETWORK/DATA COUPLING.
• Upgrade of equipment connected to the NETWORK/DATA COUPLING.
Host Name:
GE HEALTHCARE
DIRECTION 5397208-8EN, REVISION 2 WIRELESS DR IMAGING OPTION SYSTEM PRE-INSTALLATION
GE HEALTHCARE
DIRECTION 5397208-8EN, REVISION 2 WIRELESS DR IMAGING OPTION SYSTEM PRE-INSTALLATION
4.6 Devices & Services Audit
Use the following narrative to complete Being a remote host server allows the
the form on the previous page. Wireless DR Imaging Option System to
REMOTE HOSTS: Remote hosts are push images to other devices. If you
DICOM devices to which the Wireless want the device to accept this service,
DR Imaging Option System can push check yes and provide the AE title and
an image. Remote hosts can be review port number.
workstations, archival devices, or Being a query/retrieve service class
PACS systems. Please indicate the provider allows the Wireless DR
type of remote host. Imaging Option System to query this
Now indicate the manufacturer and device and retrieve images stored
model name or number. there. If you want this device to provide
these services to the Wireless DR
Compatibility can vary with software
Imaging Option System check yes and
versions, please indicate the version of
fill in the requested items.
device firmware/software the device will
be running. The query/retrieve by study or patient
controls how much the user is able to
List the device's IP address.
retrieve at one time. For study, the user
The answers to the next several items may retrieve studies, series, images.
can be found in the device's DCS For patient, the user may retrieve all of
(DICOM Conformance Statement). the study attributes plus a patient's
Please indicate the highest level of entire image collection.
DICOM conformance for this device. If A storage commitment provider
the device is not DICOM compliant, confirms that images sent by the
please indicate so and move on to the Wireless DR Imaging Option System to
next device. an archival system were received and
If the device does have some level of stored. Note - This option is only
DICOM conformance please return a available when the Wireless DR
copy of the DICOM Conformance Imaging Option System is sending DX
Statement with this completed form. type images. If your device supports
DICOM supports a number of image both DX image types and storage
types. Please indicate if this device commitment check yes and provide the
supports the DX and/or the CR image AE title, the port number and the
types. network (IP) address.
The host name is the name that will The MPPS server receives the
appear on the screen and users will use messages sent by the Wireless DR
to indicate this device. Please list the Imaging Option System. These
host name. messages consist of information such
as when the exam started and closed,
The next four sections address the four
how many images were acquired, dose
services that remote host devices may
information, etc. This information is
offer. Each of the services will have its
then updated on the Hospital
own AE (application entity) title and port
Scheduling system. If the site has an
number. The AE title is the name given
MPPS server, provide the AE Title , IP
to a service or application provided by a
address and port number.
DICOM device. The port number is a
logical designation within the device.
These pieces of information are
available in the device's DCS.
GE HEALTHCARE
DIRECTION 5397208-8EN, REVISION 2 WIRELESS DR IMAGING OPTION SYSTEM PRE-INSTALLATION
Manufacturer/Model:
Software/Firmware Version:
Host Name:
Printer AE Title:
Port Number:
GE HEALTHCARE
DIRECTION 5397208-8EN, REVISION 2 WIRELESS DR IMAGING OPTION SYSTEM PRE-INSTALLATION
Manufacturer/Model:
Software/Firmware Version:
Host Name:
HIS/RIS AE Title:
Port Number:
RIS Systems: As with the remote hosts Fill in the Host name.
please list the manufacturer and the Look in the DCS for the AE title and port
model name/number. The software/ number.
firmware version should also be
Please indicate if this device supports
entered.
the DX and/or the CR image types. This
Indicate the IP address the device is information should also be in the
using as well as the DICOM compliance device's DCS.
level. Please include the DCS for the
RIS with this completed form.
The Wireless DR Imaging Option System is an acquisition-only device. Because of that fact you
will need to move acquired images off the Wireless DR Imaging Option System and into your work/
data flow. Please use the chart below to describe your data flow. As an example, if your facility
reviewed images as the first step after acquisition, the review box would be checked in the first
column of the Task row and the review workstation would be checked in the first column of the
Device row. You should use each of the functions once.
GE HEALTHCARE
DIRECTION 5397208-8EN, REVISION 2 WIRELESS DR IMAGING OPTION SYSTEM PRE-INSTALLATION
1st step after acquisition 2nd step after acquisition 3rd step after acquisition
Printing: It is important to us to understand the path your images follow before they are printed. We
are now looking to answer the question of what road an image most typically travels on its way to
be printed regardless if that is the first step in your process or not. Please try to find in the list below
the path that best describes the path the image takes from acquisition to printing.
GE HEALTHCARE
DIRECTION 5397208-8EN, REVISION 2 WIRELESS DR IMAGING OPTION SYSTEM PRE-INSTALLATION
Image Review: Now let's trace the path from acquisition to image review. Again, pick the item
below that best describes how the image flows from the Wireless DR Imaging Option System to the
radiologist.
Archive: The final part of this triad is archiving images. Pick the item below that best describes the
flow of images to be archived.
GE HEALTHCARE
DIRECTION 5397208-8EN, REVISION 2 WIRELESS DR IMAGING OPTION SYSTEM PRE-INSTALLATION
Equipment:
2.) Floor is strong enough for intended equipment and mounting methods
approved – seismic regulatory codes considered?
3.) Delivery route accommodates all intended equipment?
8.) Support structures installed for floor, and wall mounted equipment?
Interconnections Completed
4.) Flexible, stranded wire provided for System input power connection?
5.) System “feeder” power cables pulled and sufficient length available at
disconnect box for connections?
6.) Interconnecting cables continuity checked, and labeled?
GE HEALTHCARE
DIRECTION 5397208-8EN, REVISION 2 WIRELESS DR IMAGING OPTION SYSTEM PRE-INSTALLATION
General Completed
1.) Walls, and floor clear of all obstructions?
Comments:
Inspection Date(s):
GE HEALTHCARE
DIRECTION 5397208-8EN, REVISION 2 WIRELESS DR IMAGING OPTION SYSTEM PRE-INSTALLATION
GE HEALTHCARE
DIRECTION 5397208-8EN, REVISION 2 WIRELESS DR IMAGING OPTION SYSTEM PRE-INSTALLATION
GE HEALTHCARE
DIRECTION 5397208-8EN, REVISION 2 WIRELESS DR IMAGING OPTION SYSTEM PRE-INSTALLATION
GE HEALTHCARE
DIRECTION 5397208-8EN, REVISION 2 WIRELESS DR IMAGING OPTION SYSTEM PRE-INSTALLATION
Item Cable Name PN. Voltage Length Diameter Connector Plug Size Connector Plug Size
(V) (m/in.) (mm/in.) One End One End The Other The Other
WidthXHeigh End End
t (mm/in.) WidthXHeight
(mm)
Precision 5395954
500D
Upgrade MIS
cable
Collector
14 Handswitch 5393581 12VDC 2/ 4/0.16 Cabinet J1 30X10 Handswitch 10X10
Input Cable 78.74 /1.18X0.39 /0.39X0.39
15 Handswitch 5393580 12VDC 3/ 6.4/0.25 Cabinet J2 30X10 Precision 30X10
Output For 118.11 /1.18X0.39 500D RCIM /1.18X0.39
Precision J2
500D
16 Sniffer 5393577 12VDC 3/ 6.3/0.25 PC J3 10X10 Precision 10X10
Ethernet 118.11 /0.39X0.39 500D PC: /0.39X0.39
Cable J5 for
Aurora, J11
for others
17 Precision 5393578 12VDC 3/ 8.4/0.33 Cabinet J6 20X10 End1. 30X10
500D Power 118.11 /0.79X0.39 Precision /1.18X0.39
Sense Cable 500D PC
Power Input
End2:Precis
ion 500D
PC Power
Cable
18 Table Bucky 5394068 110VAC 0.1/ J3(Precisio 30X10 No NA
Simulation 3.94 n 500D /1.18X0.39 Connection
Cable Table
Base)
19 Wallstand 5394070 110VAC 0.1/ J210(Preci 30X10 No NA
Bucky 3.94 sion 500D /1.18X0.39 Connection
Simulation Positioner
Cable Cabinet)
Note: System main power cable 5393571 is pre-terminated with US standard plug. This plug is
replaceable. For other countries, customer need prepare the plug matching local electrical outlet
and regulatory requirement.
GE HEALTHCARE
DIRECTION 5397208-8EN, REVISION 2 WIRELESS DR IMAGING OPTION SYSTEM PRE-INSTALLATION
5393573
5393581 2259298-60
5393579 5393574 (Proteus XR/a cable)
J5 for Aurora, J11 for others Hand Switch Power Input System RFP3 J210
Hospital
RCIM
Magic PC(Aurora) PC Cabinet A5 J3
Positioner Electrical
IUI PC(Aretha/Odetta) Table Cabinet Outlet
P500D System
GE HEALTHCARE
DIRECTION 5397208-8EN, REVISION 2 WIRELESS DR IMAGING OPTION SYSTEM PRE-INSTALLATION
This page is intentionally left blank.
GE HEALTHCARE
DIRECTION 5397208-8EN, REVISION 2 WIRELESS DR IMAGING OPTION SYSTEM PRE-INSTALLATION
GE HEALTHCARE
DIRECTION 5397208-8EN, REVISION 2 WIRELESS DR IMAGING OPTION SYSTEM PRE-INSTALLATION
If the customer needs seismic calculations they should speak to their local GE Project Manager or
Installation Specialist to obtain.
GE HEALTHCARE
DIRECTION 5397208-8EN, REVISION 2 WIRELESS DR IMAGING OPTION SYSTEM PRE-INSTALLATION
This page is intentionally left blank.
GE HEALTHCARE
DIRECTION 5397208-8EN, REVISION 2 WIRELESS DR IMAGING OPTION SYSTEM PRE-INSTALLATION
GE HEALTHCARE
DIRECTION 5397208-8EN, REVISION 2 WIRELESS DR IMAGING OPTION SYSTEM PRE-INSTALLATION
ay
Xr
Pre-Install.book Page 82 Tuesday, December 13, 2011 10:32 AM
GE HEALTHCARE
NO.1, YONGCHANG NORTH ROAD
ECONOMIC & TECHNOLOGICAL DEVELOPMENT ZONE
FAX: 86-10-67881850
TELE: 86-10-58068888
82
BEIJING, P. R. CHINA 100176